CEM CP-09 Operating Instruction

OPERATING INSTRUCTION
CP-09
Smaller sized, convenient-to-use AC Current Clamp-on Adaptor
Introducing
Using the Current Clamp Safely
Warning
To prevent electric shock or fire and personal injury, carefully read all safety information before attempting to operate the Current Clamp and follow these
Procedures:
Do not use the clamp on circuits rated higher than 600V in Installation Category II. Use
caution when clamping around uninsulated conductors or bus bars.
Do not use a clamp that is cracked, damaged, or has a defective cable. Such clamps
should be made inoperative by taping the clamp shut to prevent operation.
Check the magnetic mating surfaces of the clamp jaws; these should be free of dust, dirt,
rust, and other foreign matter.
Keep your fingers off the clamp jaws.  Keep your fingers behind the safety barrier as shown in
Safety Information
This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present
Double insulation
Specifications
General Specifications Output Impedance <1kΩ
AC Bandwidth 40 to 400Hz
Jaw opening 16 mm
Operating conditions 32ºF to 86ºF (0ºC to 30ºC) 90%RH; 86ºF to 104ºF (30ºC to 40ºC)
75%RH; 104ºF to 122ºF (40ºC to 50ºC) 45%RH
Storage conditions - 22ºF to 140ºF (-30ºC to 60ºC); < 90% Relative Humidity Altitude Operate at less than 3000 meters Weight 129g Dimensions 111 x 50 x 33mm (HXWXD) Standards IEC1010-1 (1995); EN61010-1 (2001) Category II 600V, Category
III 300V .
Range Specifications
Function
Range
Output
Sensitivity
Accuracy (of
reading)
AC Current (50/60Hz)
0~200A
1mV/A
1A/mV AC
±2.0%±0.5A
Adaptor Description
1, Current sense jaw
2. Clamp trigger
3. Safety barrier
4. Voltage output lead
Using the Current Clamp
To use the Current Clamp, follow these instructions:
1, Connect the test leads to the output shock of the clamp probe and input shock of the DMM
(or other voltage measurement device). 2, Turns on the DMM and set it at ACV 200mV or 400mV 3, Position the Current Clamp perpendicular to and centered around the conductor. 4, Read the measured value from the DMM LCD display.
Example with multimeter for the AC Current Clamp-on Adaptor
Current Clamp sensitivity = 1mV/A. Multimeter displays 100.0mV.
Actual current = display value / sensitivity Current Clamp=100.0mV/1mV/A=100.0A