WHY IS MAINTENANCE IMPORTANT?
Daily maintenance is important to reduce the amount of bacterial build up within the system;
preventing blockages and unpleasant odours.
It is very important that whilst cleaning filters and disinfecting thesuctionsystem, atno time
should any detergents or house holdproductscome into contact with anycomponents.
Only use products recommended for use by the manufacturer. Useof other productsmay cause
damage to the equipment or effect its performance.
Always ensure you are familiar with safety warning labels, productinstructions andsafety data
sheets before performing any processes or procedures.
Before carrying out any maintenance on this system, ensure all appropriate Personal Protective
Equipment (PPE) such as: gloves, mask,full sleeve gown and eye protection are worn.
WARNINGS
- Always refer to the operator’s handbook before carrying out any work
- Manuals are available on the website www.cattaniesam.it
- We recommend their consultation especially for updates about safety
- Prior to carrying out any servicing operation make sure to wear disposable
gloves, goggles, a mask and overalls
- Disconnect the appliance from the power
- Keep the plant room free from flammable and corrosive mixtures
PEACE OF MIND FOR THE USER
Cleaning, disinfecting and system maintenance are imperative to
ensure long life of the suction system.
Failure to properly clean, disinfect or maintain the system as per
the manufacturer’s recommendations may affect the warranty.
Cattani recommend Magnolia cleaning products which are fully
compatible with the system’s components and deliver ultimate
care to the unit. Please follow dilution instructions in this guide.
AMALGAM RECYCLING
Amalgam container replacement should be part of the annual maintenance of the suction system
performed by your engineer.
Please DO NOT place this item in the rubbish bin and/or down the sink.
Amalgam must be collected and disposed of by an authorised licensed company.
PLEASE REFER TO USER MANUAL AND WARRANT Y STATEMENT FOR FURTHER DETAILS.
![](/html/aa/aa57/aa57da4a5639205b58e993828bbd224a3414c925eba2f0dfa3f2ba1913984d81/bg3.png)
MAGNOLIA CARE PRODUCTS
PROTECTING YOUR SYSTEM
Puli Jet Plus
(4 x 1L bottle)
060900
Puli Jet Classic
(4 x 5L bottle)
040715
Sanitising
Antifoaming
Tablets
(50 per box)
040825
Disinfe cts, deod orises and cleans al l dental aspir ators (wet, semi-wet and dr y line), does not contain
aldehydes and phenol. 1 litre makes up to 125 litres of disinfectant or 250 litres of cleaner.
WHEN TO USE:
HOW TO U SE:
GENERAL
CLEANING:
DISINFECTION:
Every evening.
1. Shake bot tle, then unscrew red cap slightly.
2. Squ eeze bottle to distr ibute 10ml of liquid to top container.
3. To mix, fill the ‘P ulse Clea ner Bucket’ with war m water (approx temp 5 0°c) up to
the thir d marker, then uns crew the red c ap on the Puli J et bottle fully and p our
the Pul i Jet Plus into th e ‘Pulse Cl eaner bucket ’.
We recommend that you cleanse the aspiration system every evening, at the end of work.
For this, fill the Pul se Cleaner Bucket up to the thir d marker with warm water (approx temp
50°c) and add 10ml of Puli Jet Plus solution. Pour exc ess down th e spittoon.
We reco mmend to disin fect the asp iration sy stem once a we ek with a higher conce ntrate of
Puli J et Plus. For th is, fill the P ulse Clea ner Bucket up to th e third marker with w arm water
(approx temp 50°c) and add 20ml of Puli Jet Plus solution. Pour exces s down the spittoon.
Clean s, sanitis es and deodor ises denta l suction uni ts. 5 litre s makes up to 250 l itres of clea ner.
WHEN TO USE:
HOW TO USE:
Every evening.
1) Fill the ‘Pulse Cleaner Bucket’ with warm water (approx temp 50°c) up to the
third marker, then measure out 50ml of Puli Jet Classic cleaner and pour it into
the ‘Pulse Cleaner bucket’.
2) Using the terminals, aspirate the mixed Puli Jet Classic solution from the Pulse
Cleaner through the equipment. Pour excess down the spittoon.
Stops foam build up, which can cause an interruption in the aspiration.
APPLICATION:
WHEN TO USE:
HOW TO USE:
Single surgery and centralised suction systems.
Ever y day, after cleaning and disinfection.
At the beginning or end of every working day, after cleaning, place 1 or 2 tablets inside the
tip support filter or filter housing on the pump, tablet action lasts the entire day.
Lubri Jet Spray
(1 x 150ml can)
040735
(4 x 150ml can)
040737
Lubri Jet Drop
(1 x 30ml bottle)
040726
(4 x 30ml bottle)
040727
Pulse Cleaner
040720
Silicone spray for lubricating surfaces including diaphragms, O-rings, plastic materials and
metal, useful for treating clamp hinges and scissors before placing in autoclave. Silicone
lubricant with non-flammable ozone-friendly propellant, contains high-quality silicone oil and
is dust-repellent.
APPLICATION:
WHEN TO USE:
HOW TO USE:
Surfaces exposed to friction, rubber o-rings, plastic and metal materials.
When required.
Ready-to-use, Shake before use, Clean and wipe the parts to be lubricated, then spray.
Silico ne based liqu id for lubric ating surfaces inc luding diaphr agms, O-r ings, plas tic material s
and metal, useful for treating clamp hinges and scissors before placing in autoclave. Put a
drop onto the instrument and wipe off excess before putting in autoclave. Silicone lubricant
for the periodic lubricant of the hinge of pincers, scissors and other instruments of the dental
surgery. Contains high-quality silicone oil and is dust-repellent.
APPLICATION:
WHEN TO USE:
HOW TO USE:
Surgical instruments with the parts subjects to friction (scissors, pincers, etc).
When required.
Put a drop on the parts under friction. To be used on clean and dry instruments before
sterilization.
Pulse Cleaner is a turbulence-maker.
To prepare the Puli Jet solution, pour the concentrated product into warm water (approx temp
50°c) and then place the terminals on the pistons. The cleansing aspiration will start without
any further operation.
The solution left in the container can be used to sanitize the spittoon bowl. Do not rinse the
aspirator; do not turn on the spittoon-rinsing tap: Puli Jet Plus will go on sanitizing and
dissolving blood and mucus all night.
1x16mm & 1x11mm connections, reducers for smaller terminals included.
![](/html/aa/aa57/aa57da4a5639205b58e993828bbd224a3414c925eba2f0dfa3f2ba1913984d81/bg4.png)
DAILY MAINTENANCE
TO BE CARRIED OUT BY INSTRUCTED SURGERY STAFF
CHECK FOR ANY ALARMS ON THE DISPLAY
SEE ALARMS DESCRIPTION BELOW
Before carrying out any daily maintenance check the display on the keypad for your unit for any
possible alarms.
These alarms will indicate any issues such as the Amalgam level and any temperature faults.
Alarm Information:
CODE ID DESCRIPTION SOLUTION
I14 2 Amalgam level > 95% Replace the amalgam container
as soon as possible
I15 3 Amalgam level > 100% Replace the amalgam container
I04 36 Over temperature (temperature has
exceeded max. allowed limit)
GENERAL OBSERVATION
Ensure the area around the suction system is clean and clear of debris that may affect ventilation. Check
Ventilate the plant room
(check the control panel cooling fan)
the display on the keypad for any error messages or warnings and report as necessary. Check for any signs
of fluid leaks or damaged hoses and if suspect, report toyour equipment service technician.