The E-Con EA-200 application makes it possible for you to control a CASIO EA-200 Data
Analyzer from a connected ClassPad. It allows you to perform the following operations from
the ClassPad unit.
• Configuring EA-200 sampling parameters
• Starting and ending an EA-200 sampling operation
• Saving sample data
• Analysis and graphing of sample data
All of the explanations provided here assume that you are already familiar
with the operating precautions, terminology, and operational procedures of
the ClassPad 300/ClassPad 300 PLUS and the EA-200.
This chapter describes the configuration of the E-Con EA-200 application screen, and
provides basic information about its menus and commands.
Starting Up the E-Con EA-200 Application
u ClassPad Operation
(1) On the ClassPad application menu, tap the down arrow in the upper right corner and
then select [Additional] or [All] from the dropdown list.
(2) On the application menu, tap .
This starts up the E-Con EA-200 application and displays a Sensor Setup window.
E-Con EA-200 Application Windows
The E-Con EA-200 application has three windows: a Sensor Setup window, an E-Con Graph
Editor window, and an E-Con Graph window. Only the Sensor Setup window is visible when
you start up the E-Con EA-200 application. You can have any two of the three windows
(Sensor Setup, E-Con Graph Editor, E-Con Graph) displayed at the same time.
In addition to the three windows described above, there is also a Multimeter window, which
appears on the screen during certain operations.
Menu bar
Sensor Setup window
Multimeter window
{
{
E-Con Graph Editor windowE-Con Graph window
Toolbar
20031201
20040201
E-Con EA-200 Application Overview
1-2
k Sensor Setup Window
This window is for selecting a sensor for each of the EA-200 channels to be used for
sampling, and for configuring sampling parameters. The Sensor Setup window has two tabs.
Use this tab:When you want to do this:
SingleConfigure sampling parameters for a single probe
MultipleConfigure parameters for simultaneous sampling with up to three probes
k E-Con Graph Editor Window
This window displays the “current data”, which is data that is stored temporarily for each
channel following a sampling operation. The E-Con Graph Editor window has three tabs.
Use this tab:When you want to do this:
CurrentView current data
Normal
Compare
You can configure separate graph settings for each of the above tabs. After selecting one of
the tabs, you can then change to the E-Con Graph window and graph its data.
Recall saved sample data without losing the current data
k E-Con Graph Window
The E-Con Graph window graphs sampled data in accordance with the settings configured
with the E-Con Graph Editor window. You can use the E-Con Graph window to analyze a
graph be zooming it, by performing Fourier series expansion, by obtaining a periodic
frequency, and more. You can also save the data for a specific range of the graph as list data
or matrix data.
k Multimeter Window
The Multimeter window provides real-time display of all channel measurement values during
sampling. Displaying this window while the [Single] tab is active on the Sensor Setup window
displays the measurement values of the CH1 or SONIC channel.
When the [Multiple] tab is active, this window shows individual values for all channels that
are being used for sampling.
20031201
E-Con EA-200 Application Overview
1-3
E-Con EA-200 Application Menus and Buttons
This section provides an overview of E-Con EA-200 application menu commands and
toolbar buttons.
k Menu Commands and Toolbar Buttons Common to All Windows
Menu/Command
Settings
O
Keyboard
Sensor Setup
E-Con Graph Editor
E-Con Graph
Graph Editor
Graph
List Editor
Main
Close
—
—
Button
—
—
C
c
n
!
$
(
~
—
v
5
Functions
See “Using the O Menu” on page 1-5-4 of your
ClassPad 300/ClassPad 300 PLUS User’s Guide.
Makes the Sensor Setup window active.
Makes the E-Con Graph Editor window active.
Makes the E-Con Graph window active.
Displays the ClassPad Graph Editor window. For
details, see the section about the Graph & Table
application in the ClassPad 300/ClassPad 300
PLUS User’s Guide.
Displays the ClassPad Graph window. For details,
see the section about the Graph & Table
application in the ClassPad 300/ClassPad 300
PLUS User’s Guide.
Displays the ClassPad List Editor window.
For details, see the section about the Statistics
application in the ClassPad 300/ClassPad 300
PLUS User’s Guide.
Displays the ClassPad Main application work area
window. For details, see the section about the
Main application in the ClassPad 300/ClassPad 300
PLUS User’s Guide.
Closes the currently active window. Note,
however, that this command does not close the
Sensor Setup window.
Displays the Multimeter window.
Displays the Variable Manager window. For details,
see the section about the Variable Manager in the
ClassPad 300/ClassPad 300 PLUS User’s Guide.
Tip
• The button for the currently active window is not displayed on the toolbar. If the Sensor Setup
window is active, for example, the C button is not on the toolbar.
20031201
2004020120050601
E-Con EA-200 Application Overview
k Sensor Setup Window Menus
Menu/Command
Setup
SetupStore
Mem
Recall
ToolCustom New
ProbeCustom
Probe
Edit
Custom
Probe
Edit
Ver nier
Sensor
Clear All
Version
Displays the [Sample] tab of the Setup dialog box. The Setup
dialog box has a [Sample] tab for advanced sampling
parameters, a [Trigger] tab for advanced trigger parameters,
and a [Graph] tab for graph settings.
Saves the settings on the Sensor Setup window under a file
name for later recall. Saved data includes [Single] tab and
[Multiple] tab sensor selections, Setup dialog box parameter
settings, and E-Con Graph Editor window settings.
Recalls saved settings to the Sensor Setup window.
Displays a dialog box for configuring a new custom probe.
Displays a dialog box for editing the settings of an existing
custom probe.
Displays a dialog box for creating a new custom probe based
on the settings of a Vernier sensor.
Returns the sensor selections on the [Single] tab and
[Multiple] tab, and all the settings on the Setup dialog box and
E-Con Graph window to their initial defaults.
Displays the version dialog box.
1-4
Functions
k E-Con Graph Editor Window Menus
Menu/Command
Setup
SetupStore
Mem
Recall
ToolClear All
Version
Displays the [Graph] tab of the Setup dialog box. Only [Graph]
tab settings can be configured when you open the Setup
dialog box from this menu.
Saves the settings on the Sensor Setup window under a file
name for later recall. Saved data includes [Single] tab and
[Multiple] tab sensor selections, Setup dialog box parameter
settings, and E-Con Graph Editor window settings.
Recalls saved settings to the Sensor Setup window.
Returns the sensor selections on the [Single] tab and
[Multiple] tab, and all the settings on the Setup dialog box and
E-Con Graph window to their initial defaults.
Displays the version dialog box.
20031201
Functions
E-Con EA-200 Application Overview
1-5
k E-Con Graph Window Menus and Buttons
Menu/Command
ZoomAll Zoom
Auto
yAuto
Full
Original
Analysis
Trace
Period
Fourier
Mem
Store Picture
Recall Picture
Store List
Store
Matrix
Button
b
—
R
—
—
N
m
—
All
Select
All
Select
Zooms all currently displayed graphs.
Resizes the y-axis so the entire graph fits in the
screen. The x-axis is adjusted automatically in
accordance with the number of samples.
Resizes the graph so all of it fits in the screen along
the y-axis, without changing the x-axis.
Resizes the graph so all of it fits in the screen.
Resizes the y-axis in accordance with the sampling
range of the sensor being used for sampling, and
resizes the x-axis to a size that allows display of all
data on the same screen.
Displays a cross-shaped trace cursor. The cursor
keys can be used to move the trace cursor along
the graph and obtain coordinate readings.
Calculates the periodic frequency for a range
specified on a graph.
Uses Fourier series expansion to convert a
displayed waveform graph to a function.
Saves the currently displayed graph as a graphic
image (Store Picture). You can recall a saved graph
image and overlay it on another graph to compare
them (Recall Picture). For details, see the section
about the Graph & Table application in the
ClassPad 300/ClassPad 300 PLUS User’s Guide.
Stores all of the sample data of the currently
displayed graph as list data.
Stores the sample data in the range selected on
the currently displayed graph as list data.
Displays all sample data as matrix data.
Displays the currently selected range of sample
data as matrix data.
Functions
20031201
2004020120050601
E-Con EA-200 Application Overview
1-6
k E-Con Graph Window Menus and Buttons (Continued)
Menu/Command
a
1Move
1Zoom
Speaker Output
Change Axes
Change Unit
Cls
Functions
Starts a process for moving one of the multiple graphs on the
display.
Starts a process for zooming one of the multiple graphs on
the display.
Outputs a specific range of the displayed graph through the
speaker.
When drawing multiple graphs, selecting this command
changes the currently displayed source data name and axes
to those of the next graph.
Changes the unit of the axes that are displayed when the
E-Con Axes item on the [Graph] tab of the Setup dialog box
is turned on.
Clears the graph image recalled using the [Mem] menu’s
[Recall Picture] command.
20031201
20040201
E-Con EA-200 Application Overview
1-7
E-Con EA-200 Application Status Bar
The following shows how the status bar appears for each of the E-Con EA-200 application
windows.
k Sensor Setup Window Status Bar
This item shows the currently selected sampling mode (Normal, Real-Time, Fast,
Extended, or Period). For more information, see “Modes” on page 3-3.
k E-Con Graph Editor Window Status Bar
This item shows “Axes” when the E-Con Axes option is turned on, and nothing when
E-Con Axes is turned off. For details about the E-Con Axes option, see “Configuring
Graph Window Options” on page 3-11.
k E-Con Graph Window Status Bar
•As with ClassPad applications, tapping anywhere on the E-Con graph screen with the
stylus displays the coordinates of the tapped location in the status bar.
•As operations are performed on the graph screen, the status bar also displays guidance
about the next operation.
20031201
Basic Steps for Configuring Sampling Parameters
2-1
2 Basic Steps for Configuring Sampling
Parameters
This chapter explains the basic operations you need to perform when configuring EA-200
sampling parameters from the E-Con EA-200 application. Before performing any of the
procedures in this chapter, be sure to connect the EA-200 to your ClassPad.
Configuring Parameters for Sampling with a Single Sensor
Use the [Single] tab of the Sensor Setup window to configure the parameters for a single
sensor.
• The procedure for configuring sampling settings on the [Single] tab is similar to a Windows
Setup Wizard. As you select a sensor and input setting values on the [Single] tab, settings
are adjusted automatically in order to make it possible to perform sampling properly. Even if
you are configuring settings for a single sensor, you should use the [Multiple] tab when you
want to configure all of the detailed settings yourself.
u To configure parameters for sampling with a single sensor
(1) Start up the E-Con EA-200 application.
• This displays the [Single] tab of the Sensor Setup
window.
• If the E-Con EA-200 application is already running
and another window (besides the Sensor Setup
window) is displayed, tap C to change to the Sensor
Setup window.
20031201
Basic Steps for Configuring Sampling Parameters
2-2
(2) Tap the [Sensor] box.
• This displays a Select Sensor dialog box like the one
shown to the right.
(3) Select the sensor you will use for sampling.
•Tap one of the tabs ([CASIO], [Vernier], [Custom]), and then tap the option button
next to the name of the sensor you want to select. For details about each of the
selectable sensors, see the “10 Sensor List”.
•Tapping the [Custom] tab displays a sheet for configuring the parameters of a userdefined custom probe. For more information, see “Using a Custom Probe” on page
3-12.
(4) After selecting a sensor, tap [Set].
• This closes the Select Sensor dialog box. The Sensor Setup window now shows the
name of the sensor you selected, along with boxes for the sample range and total
elapsed sampling time (Total Sampling Time).
• The example screen below shows what would appear if when the CASIO
Te mperature (°C) sensor is selected.
20031201
Name of channel where the
selected sensor is connected
Basic Steps for Configuring Sampling Parameters
2-3
•Selecting some sensors will cause different parameters from those shown above to
appear. The following table explains you should go for more information about such
these sensors.
For this type of sensor:Go here for more details:
[CASIO] tab, [Microphone-FFT]
See “Microphone-FFT Parameters” on page 2-4
for more information.
[Vernier] tab, [Photogate]See “Photogate Sensor Parameters” on page 2-5
or [Photogate (Pulley)]for more information.
• The [Setup] menu is unavailable while [Speaker (Sample Data)] is selected on the
[CASIO] tab of the Select Sensor dialog box.
•When [Speaker (y = f(x))] is selected on the [CASIO] tab of the Select Sensor dialog
box, sampling is not performed by a sensor. Instead, the sound of the waveform of a
function input on the ClassPad is output from the EA-200’s speaker. See “6
Outputting a Function to the Speaker” for more information.
(5) Tap the [Total Sampling Time] box.
• This displays the Total Sampling Time dialog box.
Use the ClassPad
keyboard to input a
total time value.
Tapping here displays a menu of
units (msec, seconds, minutes,
hours, days) for the total time.
• If you specify a total sampling time value in the range of 10 seconds to 1 day, starting
the sampling operation will automatically enter the Real-Time mode and perform realtime graphing during sampling.
(6) To apply the current settings on the Total Sampling Time dialog box, tap [Set].
• This closes the Total Sampling Time dialog box, and displays the recalculated total
sampling time in the Total Sampling Time box on the Sensor Setup window. See the
“Tip” on the next page for information about how recalculation is performed.
(7) At this point you could configure advanced setup parameters, if you want.
•See “3 Setup” for more information.
(8) You can start sampling immediately or you can save the setup in memory for later
recall.
•To start sampling immediately, tap V. For more information, see “5 Executing a
Sampling Operation”.
•To store the setup in memory, tap [SetupMem] on the menu bar, and then tap [Store].
See “4 Using Setup Memory” for more information.
20031201
20040201
Basic Steps for Configuring Sampling Parameters
2-4
Tip
• When you use the above procedure to configure sampling parameters for a single sensor, the
sampling mode is selected automatically in accordance with the specified total sampling time. In
this case, the parameters of the Setup dialog box’s [Sample] tab and [Trigger] tab are also
configured automatically. The current sampling mode is indicated in the status bar.
For information about the mode indication in the status bar, see “E-Con EA-200 Application
Status Bar” on page 1-7. For information about the Setup dialog box’s [Sample] tab and [Trigger]
tab, see “3 Setup”.
•Tapping [Set] on the Total Sampling Time dialog box automatically configures the sampling
interval in accordance with the total sampling time that you input on the dialog box. The number
of samples value is calculate and rounded to the nearest integer. The number of samples integer
value is used to recalculate the total sampling time ({Number of Samples} × {Sampling Interval}),
and displays the result in the Total Sampling Time box on the Sensor Setup window. All of this
means that the value you input on the Total Sampling Time dialog box may not necessarily the
value that appears on the Sensor Setup window.
k Microphone-FFT Parameters
The [CASIO] - [Microphone-FFT] sensor option selects the EA-200’s built-in microphone for
sampling sound.
In step (3) of the procedure under “To configure parameters for sampling with a single
sensor” on page 2-1, select [Microphone-FFT] on the [CASIO] tab for the sensor. This
causes the parameters described below to appear on the Sensor Setup window that appears
in step (4).
Parameter
Frequency Pitch10Specify the step in Hertz that should be used for sampling.
Frequency Max4000
Initial
Default
Description
Specify the maximum frequency for sampling (minimum is
always 0 Hz).
Tip
• The menu bar’s [Setup] button is unavailable while [Microphone-FFT] is selected.
20031201
20040201
Basic Steps for Configuring Sampling Parameters
2-5
k Photogate Sensor Parameters
Connection of a Vernier Photogate to the EA-200 requires configuration of parameters that
are different from those for other types of sensors.
u To configure a setup for Photogate alone
In step (3) of the procedure under “To configure parameters for sampling with a single
sensor” on page 2-1, select [Photogate] on the [Vernier] tab for the sensor. This causes the
parameters described below to appear on the Sensor Setup window that appears in step (4).
ParameterInitial DefaultDescription
(Channel)SONICCH1:Select this option when the Photogate
SONIC: Select this option when the Photogate
Gate StatusOpen-OpenOpen-Open: Timing starts when the gate opens, and
Open-Close: Timing starts when the gate opens, and
Close-Open: Timing starts when the gate closes,
Close-Close: Timing starts when the gate closes,
Number of Samples 10Input an integer in the range of 1 to 1000 to
specify the number of samples.
Store List FolderEConSampSpecify the variable folder for storage of interval data
Dataand count data.
IntervalIntervalInput a variable name up to eight characters long for
storage of interval data.
CountCountInput a variable name up to eight characters long for
storage of count data.
sensor is connected to CH1.
sensor is connected to SONIC.
continues until it closes and then opens
again.
continues until it closes.
and continues until it opens.
continues until it opens, and then closes
again.
20031201
Basic Steps for Configuring Sampling Parameters
2-6
u To configure a setup for Photogate and Smart Pulley
In step (3) of the procedure under “To configure parameters for sampling with a single
sensor” on page 2-1, select [Photogate (Pulley)] on the [Vernier] tab for the sensor. This
causes the parameters described below to appear on the Sensor Setup window that appears
in step (4).
ParameterInitial DefaultDescription
(Channel)SONICCH1:Select this option when the Photogate
Distance1mInput a value in the range of 0.1 to 4 to specify the
Store List FolderEConSampSpecify the variable folder for storage of time,
Datadistance, velocity, and acceleration data.
TimeTimeInput a variable name up to eight characters long for
Distance DistancInput a variable name up to eight characters long for
Velocity VelocityInput a variable name up to eight characters long for
AccelAccelInput a variable name up to eight characters long for
(Pulley) sensor is connected to CH1.
SONIC: Select this option when the Photogate
(Pulley) sensor is connected to SONIC.
distance in meters.
storage of time data.
storage of distance data.
storage of velocity data.
storage of acceleration data.
Tip
• The menu bar’s [Setup] button is unavailable while [Photogate] or [Photogate (Pulley)] is
selected.
20031201
Basic Steps for Configuring Sampling Parameters
2-7
Configuring Parameters for Sampling with Multiple Sensors
Use the [Multiple] tab to configure parameters for simultaneous sampling with multiple
sensors. The [Multiple] tab lets you select up to three channels for sampling, from among the
EA-200’s CH1, CH2, CH3, and SONIC channels.
u To configure parameters for sampling with multiple sensors
(1) Start up the E-Con EA-200 application.
• This displays the [Single] tab of the Sensor Setup window.
• If the E-Con EA-200 application is already running and another window (besides the
Sensor Setup window) is displayed, tap C to change to the Sensor Setup window.
(2) Tap the [Multiple] tab.
• This displays the [Multiple] tab, which appears as
shown to the right.
(3) Tap inside the box of one of the sensors you want to
select ([CH1], [CH2], [CH3], [SONIC]).
• This causes a Select Sensor dialog box like the one
shown to the right to appear. The dialog box shown
here is for CH1, CH2, and CH3.
20031201
Basic Steps for Configuring Sampling Parameters
2-8
(4) Select the sensors you will use for sampling.
• Depending on the sensors you have connected to each channel, select the [CASIO],
[Vernier], or [Custom] tab and then tap the option button for the applicable sensor
name. For details about the sensors that can be selected for each channel, see the
“10 Sensor List”.
• Tapping the [Custom] tab displays a sheet for configuring the parameters of a userdefined custom probe. For more information, see “Using a Custom Probe” on page
3-12.
• If you do not want to use a particular channel for sampling, select the [None] option
on the Select Sensor dialog box.
(5) After selecting a sensor, tap [Set].
• This closes the Select Sensor dialog box and displays the [Multiple] tab again (step
(2)), with the name of the sensor you selected next to the channel name.
(6) Repeat steps (3) through (5) for all of the channels you want to use for sampling.
• You can select sensors for up to three channels at one time.
• If you select sensors for four channels, an error occurs when you start sampling. To
correct the problem, display the Select Sensor dialog box for one of the channels,
and then select the [None] option button to deselect it.
(7) On the menu bar, tap [Setup].
• This displays the [Sample] tab of the Setup dialog
box.
(8) Use the [Mode] box to select the mode you want.
• For more information about modes, see “Modes” on page 3-3.
• This causes the applicable parameters for the mode to appear on the [Sample] and
[Trigger] tabs of the Setup dialog box.
20031201
20040201
Basic Steps for Configuring Sampling Parameters
2-9
(9) Use the [Sample] and [Trigger] tabs of the Setup dialog box to configure the required
parameters.
• For details about the required parameters in each mode, see “Mode Parameters” on
page 3-5.
(10) To apply the current settings on the Setup dialog box, tap [Set].
• This closes the Setup dialog box.
(11)You can start sampling immediately or you can save the setup in memory for later
recall.
•To start sampling immediately, tap V.
See “5 Executing a Sampling Operation” for more information.
•To store the setup in memory, tap [SetupMem] on the menu bar, and then tap [Store].
See “4 Using Setup Memory” for more information.
20031201
3-1
Setup
3 Setup
This chapter explains the various parameters you can configure with the Setup dialog box.
Important!
•Configuring advanced setup parameters is optional in the case of single-sensor sampling.
See “To configure advanced parameters for a single sensor” on page 3-2.
•With multiple-sensor sampling, advanced setup parameters are part of the normal
configuration procedure. See steps (7) through (10) under “To configure parameters for
sampling with multiple sensors” on page 2-7.
Setup Overview
Tapping [Setup] on the Sensor Setup window menu bar displays the Setup dialog box. The
Setup dialog box has the three tabs shown below.
[Sample] Tab[Trigger] Tab[Graph] Tab
The following describes each of the tabs in detail.
[Sample] Tab and [Trigger] Tab
These tabs let you configure advanced parameters for the tab ([Single] or [Multiple]) that is
currently open on the Sensor Setup window. The [Sample] tab has parameters that control
how sampling is performed, while the [Trigger] tab has parameters that control the event that
triggers sampling. See “Configuring Advanced Sampling Parameters” on page 3-2 for more
information.
[Graph] Tab
This tab is for selecting options that control how sampled data is graphed. See “Configuring
Graph Window Options” on page 3-11 for more information.
20031201
20040201
3-2
Setup
Configuring Advanced Sampling Parameters
This section explains how to configure advanced sampling parameters on the [Sample] and
[Trigger] tabs of the Setup dialog box.
u To configure advanced parameters for a single sensor
(1) Perform steps (1) through (6) under “To configure parameters for sampling with a single
sensor” on page 2-1.
(2) On the menu bar, tap [Setup].
• This displays the [Sample] tab of the Setup dialog
box.
• The initial [Mode] setting is the mode that matches the sensor that was selected on
the [Single] tab.
(3) Use the [Mode] box to select the mode you want.
• For more information about modes, see “Modes” on page 3-3.
• Selecting a mode causes the applicable parameters to appear on the [Sample] and
[Trigger] tabs.
(4) Use the [Sample] and [Trigger] tabs of the Setup dialog box to configure the required
parameters.
• For details about the required parameters in each mode, see “Mode Parameters” on
page 3-5.
(5) After everything is the way you want, tap [Set].
• This applies your settings and closes the Setup dialog box.
20031201
3-3
Setup
(6) You can now start sampling immediately or you can save the setup in memory for later
recall.
• To start sampling immediately, tap V.
See “5 Executing a Sampling Operation” for more information.
• To store the setup in memory, tap [SetupMem] on the menu bar, and then tap [Store].
See “4 Using Setup Memory” for more information.
Modes
The [Mode] box at the top of the [Sample] tab of the Setup dialog box controls the current
mode setting.
There is a selection of five modes, depending on the type of data being sampled and the
sampling method: Normal, Real-Time, Fast, Extended, and Period. The Normal, Real-Time,
Fast, and Extended modes are normally available, while the Period mode is available in
special cases only.
k Normal Mode
This general-purpose mode can be used for a wide range of sampling applications. The
Normal Mode is used for relatively short sampling. When the Normal Mode is selected, you
need to specify the sampling interval, the number of samples, and the timing of the start
trigger.
k Real-Time Mode
Basically, the Real-Time Mode is the same as the Normal Mode, except that it produces a
real-time graph of data as it is sampled. Use this mode when you need to graphically monitor
changes in sampled values as they are being sampled.
k Fast Mode
Use this mode when sampling a sound waveform or when other very fast sampling is
required. This mode can be selected only when CH1 or MIC is being used for sampling. This
mode is selected automatically whenever sampling is performed using the microphone. In
this mode, the sampling interval can be set to a very short interval in the range of 20 애sec to
500 애sec.
20031201
20040201
3-4
Setup
k Extended Mode
The Extended Mode is the opposite of the Fast Mode in that it allows setting of a long
sampling interval. In this mode, the sampling interval can be set in a range of 5 to 240
minutes. This mode is best for sampling data like temperature or humidity over a long period.
k Period Mode
The Period Mode is for sampling data that changes periodically. Samples recorded by the
Period Mode differ from the samples recorded in other modes as shown below.
Sampled
Val ues
Sampling
Interval
Time
Number of Samples
Other ModesPeriod Mode
Unlike the other modes that record time and sampled values, the Period Mode records the
number of samples (1st sample, 2nd sample, and so on) and the intervals between samples.
For details, see “Period Mode Parameters” on page 3-7.
Tip
• If you display the Setup dialog box from the [Multiple] tab of the Sensor Setup window, the [Mode]
box will allow selection of the Normal, Real-Time, or Extended mode only.
• Note that the availability of modes depends on the sensor being used for sampling. Certain
sensors allow selection of particular modes only.
• The currently selected mode is shown in the status bar of the Sensor Setup window.
20031201
3-5
Setup
Mode Parameters
This section explains the parameters that can be configured on the [Sample] tab and
[Trigger] tab of the Setup dialog box, in accordance with the mode selected on the [Sample]
tab.
k Parameters Common to All Modes
The following explains the parameters that normally appear, regardless of the currently
selected mode.
Sampling Interval:Specify a value for the sampling interval. If you specify an interval of
Number of Samples: Specify the number of samples that should be taken. The total
Warm-Up:Specify the warm-up time for the sensor. You can select from among
one second, for example, a sample will be taken every second.
sampling time is calculated in accordance with the specified interval
and number of samples, and displayed on the [Single] tab of the
Sensor Setup window.
the following options for this setting.
Auto:Automatically sets a preset value for each sensor.
Manual:Selecting this option displays a dialog box like the one
shown below. Specify a warm-up time in the range of 1
to 99 seconds, and then tap [Set].
Start Trigger:Specify the event that triggers the start of sampling. You can select
from among the following options for this setting.
Tap Screen: This option specifies that sampling should start when
[OK] is tapped on the screen.
Count Down: This option specifies that sampling should start when the
end of a countdown is reached.
CH1:This option specifies that sampling should start when
CH1 channel input reaches a certain value.
SONIC:This option specifies that sampling should start when
SONIC channel input reaches a certain value.
Microphone: This option specifies that sampling should start when the
EA-200’s microphone detects a sound.
20031201
20050601
3-6
Setup
k Normal Mode Parameters
TabParameterInitial DefaultRange
[Sample]Sampling Interval0.05 sec0.0005 to 299 sec
Number of Samples10010 to 30000
Warm-UpAutoAuto, Manual (1 to 99)
[Trigger]Start TriggerTap ScreenTap Screen, Count Down, CH1, SONIC
• For details about each parameter, see “Parameters Common to All Modes” on page 3-5.
•Except for the [Tap Screen] item, all of the other start trigger options require specification of
additional parameters. For details, see “Additional Start Trigger Parameters” on page 3-9.
k Real-Time Mode Parameters
TabParameterInitial DefaultRange
[Sample]Sampling Interval1 sec0.3 to 299 sec
Number of Samples10010 to 30000
Warm-UpAutoAuto, Manual (1 to 99)
[Trigger]Start TriggerTap ScreenTap Screen, Count Down
• For details about each parameter, see “Parameters Common to All Modes” on page 3-5.
•Except for the [Tap Screen] item, all of the other start trigger options require specification of
additional parameters. For details, see “Additional Start Trigger Parameters” on page 3-9.
k Fast Mode Parameters
TabParameterInitial DefaultRange
[Sample]Sampling Interval20 애sec20 애sec to 500 애sec
When FFT Graph is turned on, the setting range become 10 through 8192.
• For details about parameters other than FFT graph, see “Parameters Common to All
Modes” on page 3-5.
•Except for the [Tap Screen] item, all of the other start trigger options require specification of
additional parameters. For details, see “Additional Start Trigger Parameters” on page 3-9.
20031201
*
Loading...
+ 56 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.