The calibration lev el screen will then be displayed (Figure H) . Thi s screen is used to set
the exac t ref erence level us ed by the acou sti c calib r a tor. Refer to the cal ibra tio n
cert if ic at e f or th e acou s tic calib ra to r for th e cal ibra tor outpu t l eve l.
The ‘Signal input source’ screen (Figure I) will then be shown. NOTE: th is op tion is for
the use of acous tic laboratories only. This allows the signal input to be routed from e ither
the inbuilt microphone or from an external signal generator connected via the auxiliary
2.5m m headphone socket. Press the right key to switch between microphone
input or Alternating Current input . Th e default power up selection
is for mi crophone input. Press the lef t key to confirm any changes.
Specification
General
Designed to provide functio nality wi th re spect to the f ollowing international standards:
IEC 61672-1 2002-5 (Electro-Acoustics – Sound Level Meters) Group ‘X’
instruments, Performance Class 2.
ANS I S1.4 Type 2A Specific at ion for S ou nd Level Meter s
Ran ge : Disp l a y ran g e: 3 0- 130dB(A) R MS, avail ab le in 2 ranges, 30- 10 0d B and
60-130dB. Linear operating range 10dB ab ove noise f lo or.
RMS Frequency Weightings
A and C fi lter weightings, satisfying IEC 61672-1: 2002 Cl ass 2, ANSI S1.4 Type 2A.
Noi se Floor
Total noise floor typically <33dB(A).
Frequency Response
Over all frequency response as per IEC 61672-1: (2002) Class 2, ANSI S1.4 Type 2A.
Time Weightings
Fast, Slow and Impulsive according to IEC 61672-1: (2002), ANSI S1.4 Type 2A.
Reference Conditions
20
o
C air temper ature, 65% Relativ e Hu midity, 101.325kPa atmospheric pressure.
Nominal ref erence level = 114.0dB at 1k Hz. Free field perpendicular incid ence.
Operating Environmenta l Conditio ns
Humidi t y: 5 to 90% RH in th e absenc e of conden sat ion.
Tempera tur e range : 0 to 40
o
C.
Pressure: 65 to 108kPa