CAS CBX, CBW User Manual

Page 1
2
CBX 앞표지
Page 2
3
PRECAUTIONS
Page 3
4
Page 4
5
Other conventions used in this manual include:
Item Description
1) 2) 3)…
Indicates the step number in a procedure or a sequence of changes in the
balance display.
Indicates the display change or action sequence performed by the balance
after an operation has been initiated.
Refer to the recommended sections for further information on a specific topic.
etc.
Indicates the balance display status.
Underlined portions (_) indicate that the character is blinking.
mass display
Indicates the balance is in the weighing mode and mass is displayed in one of
the weighing units.
These sections include information to make using the balance more
convenient.
Menu Map item
[1]
Indicates the Menu Map item to be selected. The number inside the [ ] is the number of the item on the Menu Map.
See Section 2.3 "Menu Map."
Page 5
6
1. BASIC OPERATION.........................................................................................10
1.1 Balance Components......................................................................................10
1.1.1 Operations of the Keys on the Main Unit................................................................. 12
1.1.2 Symbol Display............................................................................................................. 13
1.2 Installation.............................................................................................15
1.2.1 Choosing the Installation Site ...................................................................................15
1.2.2 Balance Installation....................................................................................................... 15
1.3 Turning the Power ON ................................................................................... 17
1.4 Adjusting the Built-in Clock ..........................................................................18
1.5 Span Calibration.............................................................................................18
1.6 Wegighin........................................................................................................18
1.6.1 Changing the Unit Display.......................................................................................... 19
1.6.2 Changing the Minimum Display Digit
............................. 19
1.7 Maintaining and Transporting the Balance..................................................... 19
1.7.1 Cleaning the Balance ......................................................................................................19
1.7.2 Transporting the Balance................................................................................................19
1.8 Specifications ................................................................................................. 20
2. MEMU ITEM SELECTION.............................................................................21
2.1 Introduction....................................................................................................21
2.2 Procedure of Menu Item Selection .................................................................21
2.3 Menu Map ...................................................................................................... 23
2.4 General Menu Operations...............................................................................31
2.4.1 Setting Numeric Values............................................................................................... 31
2.4.2 Setting a Decimal Point................................................................................................ 31
2.4.3 Menu Lock..................................................................................................................... 31
2.4.4 Last Menu Recall.......................................................................................................... 32
2.4.5 Returning to the Default Settings (menu reset)........................................................ 33
2.5 Calibration Execution Menu...........................................................................34
2.5.1 Calibration......................................................................................................................... 34
2.5.2 Span Calibration Using the Built-in Weight,
............. 35
2.5.3 Calibration Check Using the Built-in Weight
................ 35
2.5.4 Span Calibration Using External Weights
........................................ 36
2.5.5 Calibration Check Using External Weights
............................ 37
2.6 Calibration Setting Menu ...............................................................................38
2.6.1 Selecting the Standard Calibration Type (
- ) ........................ 38
2.6.2 PCAL - Calibration of the Built-in Weight
- .................... 38
Page 6
7
2.7 Graphic Display Menu : ...............................................39
2.7.1 Overview.........................................................................................................................39
2.7.2 Full Scale Mode
...................................................................39
2.7.3 Target Mode .........................................................................39
2.7.4 Group Mode
.........................................................................40
2.7.5 Level Mode
..........................................................................40
2.7.6 No Graphic Display ............................................................40
2.8 Environment Menu
: ....................................................41
2.8.1 Overview.........................................................................................................................41
2.8.2 Averaging
................................................................................41
2.8.3 Stability Detection Band
........................................................41
2.8.4 Tracking
....................................................................................42
2.8.5 Pretaring Value ..................................................................42
2.8.6 Whole Lighting Mode
.............................................................43
2.8.7 Taring/Printing
.................................................43
2.9 Application Menu
: .......................................................44
2.9.1 Overview.........................................................................................................................44
2.9.2 Zero Range.....................................................................................................................44
2.9.3 Auto Zero........................................................................................................................45
2.9.4 Auto Print........................................................................................................................45
2.9.5 Peak Hold........................................................................................................................46
2.9.6 Interval Timer.................................................................................................................47
2.9.7 Auto-Memory and Zeroing ...........................................................................................47
2.9.8 Animal Weighing..........................................................................................................48
2.10. Unit Registration Menu and Unit Change....................................................49
2.10.1 Unit Registration Menu................................................................................................49
2.10.2 % Conversion.................................................................................................................49
2.10.3 Piece Counting...............................................................................................................50
2.10.4 Solid Specific Gravity Measurement.........................................................................51
2.10.5 Liquid Specific Gravity Measurement......................................................................52
2.11 System Setting Menu
: .............................................53
2.11.1 Date for the Built-in Clock...........................................................................................53
2.11.2 Time for the Built-in Clock
...................................................53
2.11.3 Display During Standby ........................................................53
2.11.4 Measurement Control System
................................................54
Page 7
8
2.12 Communication Menu ( blinking) ....................... 55
2.12.1 Overview........................................................................................................................ 55
2.12.2 Handshaking - .................................................................... 55
2.12.3 Format
................................................................................ 55
2.12.4 Communication Speed
.................................................. 56
2.12.5 Parity / Bit Length
- .............................................................. 56
2.12.6 Stop Bit
.............................................................................56
2.12.7 Delimiter ........................................................................... 56
3. CONNECTIONG PERIPHERAL INSTRUMENTS.........................57
3.1 EP-60A Electronic Printer.............................................................................. 57
3.2 Personal Computer - RS-232C -..................................................................... 57
3.2.1 Connecting the Cable...................................................................................................57
3.2.2 Data Format ................................................................................................................... 58
3.2.3 Using Command Codes.............................................................................................. 59
3.2.4 Windows® Direct Function .......................................................................................64
3.2.5 Multi-Connection Mode.............................................................................................. 67
3.3 Optional AKB-301 Application Keyboard.....................................................69
3.3.1 Basic Information.......................................................................................................... 70
3.3.2 Example of Use............................................................................................................. 71
3.4 Foot Switches.................................................................................................73
3.5 Hook for Below-Balance Weighing ...............................................................74
4. APPENDIX............................................................................................75
4.1 Part List..........................................................................................................75
4.1.1 Standard Accessories and Maintenance Parts......................................................... 75
4.1.2 Options............................................................................................................................ 75
4.2 Table of Unit Conversion Constants...............................................................76
4.3 Specifications for the RS-232C/AUX Connector...........................................77
4.4 Error Display List........................................................................................... 78
4.4.1 General Display............................................................................................................. 78
4.4.2 Error Display.................................................................................................................. 79
Page 8
9
INTRODUCTION
Thank you for purchasing this medium-sized electronic balance, the CAS CBX-K/ CBW-K series. In addition to its ability to make rapid mass measurements, the CBX-K/ CBW-K series is also well-equipped with a four-mode analog display, clock function, various application measurement functions and a unit conversion function. Furthermore, it is a high performance, multi-functional balance which can be used in a great variety of applications when connected to peripherals. The CBW-K series features a built-in calibration weight which allows easy sensitivity calibration wherever you are, making it even more convenient to use. Before using the balance read this instruction manual carefully to ensure long and fruitful use of the CBX-K/ CBW-K series.
Page 9
10
1. BASIC OPERA TION
1.1 Balance Components
Open the package and sure following items are included. Balance body, pan, AC adaptor, Balance cover, adhesive tape, Instruction manual, inspection certificate, <CBW-K only> Internal weight, Fix screw of the internal weight, cover of the internal weight Fix of the screw of the internal cover
Name Appearance Notes
Balance unit
The serial number of the balance is stated on this nameplate
Pan
Built-in calibration mass
Check that the numerals on this part are the same as the last 5 figures of the serial number on the balance unit (CBWK series only)
Calibration weight stopper screw
(CBWK series only)
Calibration weight cover
(CBWK series only)
Calibration weight cover setscrew
(CBWK series only)
AC adaptor
Page 10
11
Position of the calibration lever (CBW-K series only)
Regardless of whether the calibration lever is in the measurement or calibration position, ensure that both the left and right levers have been turned until they come to a stop.
Calibration lever (2 places, on the left and on the right)
(CBW-K series only)
Level screws
Level
Dis
play
Pan
Balance unit
Ke
y
switch
KEY Connector
RS-232C Connector
DATA I/O
Connector
DCIN Connector
Measurement position ···········································
Calibration position ···········································
Rear of balance
Front of balance
Page 11
12
1.1.1 Operations of the Keys on the Main Unit
The functions of the keys found on the front of the balance are described in the table below.
During Weighing
During Menu Selection (1)
Key
Pressing Once and Releasing
Pressing and Holding for About 3 Seconds
Pressing Once and Releasing
Pressing and Holding for About 3 Seconds
Switches between the operation and standby modes.
Exits the Application Measurement function.
Returns to the previous menu.
Returns to the mass display.
Enters span calibration or menu selection. (*2)
Displays the last Menu Map item that was set. (Last menu recall)
Moves to the next Menu Map item.
Displays the last Menu Map item that was set. (Last menu recall)
Tares the balance. (Displays zero.) (*3)
Displays the Pretare value.
Selects and sets the currently displayed Menu Map item.
No operation.
Changes the selected unit. (*4)
Switches between the 1d and 10d display. (*5)
Increases the numeric value of the blinking digit by 1.
No operation.
Sends the displayed value to a peripheral device.
Sends the date and time to a peripheral device.
Moves to the next digit during numeric value entry.
No operation.
*1 Refer to Section 2. “MENU SELECTION.”
*2 This key is used to set values when percent (%), number (PCS), solid specific
gravity (d), or liquid specific gravity (d) units are displayed.
*3 When a Pretare value is set, zero is not displayed and the [- Pretare value] is
displayed.
*4 Units other than “kg” must be registered and set before they can be used for
measurement. Kirogram (kg) is the only unit is registered before shipment. To register other units, refer to Section 2.10. “REGISTRATION, RELEASE, AND SELECTION OF MEASUREMENT UNITS”.
*5 When the unit is set to 10d, the resolution of the minimum display is decreased by
one decimal place.
Page 12
13
1.1.2 Symbol Display
Sensitivity calibration mark
Stability mark Communication mark
Graphic display Auto print mark Standby mark Unit display
`
An example of “whole lighting” display .
Display Name Description
Stability symbol Indicates that the balance is stable. (*1)
Tare symbol Indicates that a Pretare value has been set.
Zero symbol
Indicates that Auto Zero is ON for the Application Measurement function.
Weight symbol
Blinks when span calibration is necessary. This symbol continues to blink until either a manual or automatic span calibration has been made.
Number symbol Indicates numeric value entry.
Menu symbol Indicates that the menu lock is on.
*
Asterisk Indicates that the displayed numeric value is not a mass value.
Animal symbol
Illuminates when Animal Weighing is ON for the Application
Measurement function. Auto-Memory & Zeroing symbol
Illuminates when Auto-Memory and Zeroing are ON for the Application
Measurement function.
Memory symbol
Indicates that Net Total Weighing is ON for the Application
Measurement function. Communication symbol
Illuminates during communication to external equipment through the
RS-232C or DATA I/O connector.
BATT
Battery symbol
When the balance is operated with the optional battery pack, this
symbol illuminates to indicate that the battery voltage has dropped.
AP
Auto Print symbol
Indicates that Auto Print is ON for the Application Measurement
function.
STAND-BY
Stand-by symbol
Illuminates when the balance power is in the standby mode.
This symbol also illuminates when the Application Measurement
function has entered the standby mode.
Page 13
1
4
Inverse triangle symbol
Illuminates when the solid specific gravity unit is used. This symbol is also used as a substitute for the decimal point.
1 Stability symbol
The displayed value may change while the stability symbol remains illuminated if the load is changing slowly or if the stability detection band has been set to a large value.
Page 14
15
1.2 Installation
1.2.1 Choosing the Installation Site
(1) Power supply
• Select an installation site that is near a power source to ensure that the attached AC adapter is used properly. If this is not possible, an optional battery pack is available as a special accessory.
• Verify that the supply power voltage conforms to that indicated on the AC adapter.
(1) Installation site
Choose an installation site where the balance will be protected from the
following:
• corrosive or flammable gasses.
• dust, wind, vibration, electromagnetic waves, or magnetic fields.
• direct sunlight.
• extreme temperature or humidity.
• rain or the possibility of being splashed with water.
Large capacity balances should be installed on a sturdy floor and table that can
withstand the total load of the balance AND object to be weighed.
1.2.2 Balance Installation
1) Put the balance onto its side and unscrew the 2 transportation screws in accordance with the instruction label on the bottom of the balance. Screw these into the “Release” hole.
2) Place the balance unit as shown in the diagram on
the right. If intending to fit the vinyl cover, fix it at this stage of the procedure. Refer to “Fitting the Vinyl Cover”, mentioned later.
Release hole
Instruction label
Page 15
16
3) Fix the calibration weight cover (1) with the calibration mass cover setscrew (2).
Note
Steps (3) to (5) are only required with the CBW-K series.
4)Put the built-in calibration weight (3) in place and
tighten the calibration weight stopper screw (4).
5 )Turn both the left and right calibration levers of
the pan to the Measurement position.
6) Place the pan on the balance. For the CBW-K series, the pan will be easier to place if it is placed slightly from the rear.
7) The level of the balance is adjusted Note.
Turn the level screws to adjust it so that the air bubble in the level rests inside the red circle. Check that the balance is steady when adjustment is complete.
Note
Adjusting the level can be done with ease if all 4 level screws are at first grounded equally. Next, adjust the front and the back using the 2 front feet, and then adjust left and right using either of the 2 feet on the left and right.
1.2.2.1 Fitting the Vinyl Cover
When using the balance in a location where it is likely to get dirty easily, refer to the diagram below and affix a vinyl cover by following the procedure below. Remove the pan first if fitting the vinyl cover after the balance has been assembled.
1) Cut the supplied tape (2) and stick it onto the 4 points on the case (3) (refer to the diagram below).
2) Fix the vinyl cover (1) onto the double-sided tape (2), ensuring that there are no creases.
3) Fit the pan.
Measurement
Page 16
1
7
4) Place a weight onto the pan which is close to the weighing capacity to check that the pan and vinyl cover are not touching. Accurate measurements cannot be made if there is contact between the two items. If this is the case, re-fix the vinyl cover.
(Note) The vinyl cover (1) is folded up when it is in the packaging.
1.3Turning the Power ON
1) Insert the plug of the AC adapter into the DC IN connector on the rear of the balance, then insert the AC adapter into the power source.
Note: When using the optional battery pack (special accessory), connect the fully charged
battery pack to the DC IN connector of the balance using the cable attached to the battery pack.
2) The balance self-check is activated and the following messages are displayed in the order indicated.
Ver. No.
Whole lighting
(
and
are not displayed for the CBX series.)
3) When the
key is pressed, the whole display illuminates and then the display
changes to indicate the kilogram-display.
Note
If ENVIRONMENT menu 37
has been selected, the display stops in
the whole lighting state. In this case, press the key during whole lighting to
change the display to the kilogram-display.
Page 17
1
8
1.4 Adjusting the Built-in Clock
The CBX-K and CBX series of balances have a built-in battery driven clock. After unpacking and installing the balance, adjust this clock to the appropriate date and time.
Refer to Section 2.11.1 “Date for the Built in Clock.”
Refer to Section 2.11.2 “Time for the Built-in Clock.”
1.5 Sp an Calibration
It is necessary to calibrate the balance after it is moved, or when the ambient temperature has changed. Verify that the balance is stable before performing the span calibration. To achieve a very stable state, ensure that the balance has been turned on for at least one hour, that the temperature is constant, that there are no breezes or vibrations and that the balance is in an area isolated from the normal traffic flow.
Refer to the following sections for more information on the Span Calibration procedure.
For CBX-K series balances, refer to Section 2.5.2 “Span Calibration Using the
Built-in Weight.”
For CBX series balances refer to Section 2.5.4 “Span Calibration Using
ExternalWeights.”
1.6 Weighing
1) Place the weigh vessel (tare) on the pan.
2) Press the
key to zero the display.
Note
If a Pretare value is set, the
symbol is illuminated and zeroing of the display does not occur.
3) Place the object to be weighed on the pan, and read the displayed value after the stability symbol is displayed.
Errors Displayed During Weighing
Overload: Weighing capacity has been exceeded. Negative Overload: The load on the balalnce is too light.
The pan is not adjusted properly.
Display Overload: This display appears if the mass display exceeds
7 digits due to the choice of unit.
May appear after the
key is pressed while the solid specific gravity
(▼d)or liquid specific gravity (d) units are used.
Page 18
19
1.6.1 Changing the Unit Display
Every time the key is pressed, the unit display changes sequentially among
those registered in the UNIT REGISTRATION menu.
Note: Before a unit can be displayed it must be registered in the UNIT
REGISTRATION menu. Refer to “2.10 Unit Registration Menu and Unit Change” for information on how to register units. The unit is displayed sequentially according to the order of the UNIT REGISTRATION menu.
1.6.2 Changing the Minimum Display Digit ( )
It is possible to decrease the resolution of the minimum balance display by one decimal place if necessary.
1) Press and hold the
key for approximately three seconds. is
displayed and the display is decreased by one decimal place.
2) Press and hold the
key for approximately three seconds. is
displayed and the display returns to the original number of decimal places.
1.7 Maintaining and Transporting the Balance
1.7.1 Cleaning the Balance
Use a soft damp cloth containing a neutral detergent to clean the balance.
Avoid using organic solvents, chemicals, or dusting sprays as they may damage the coatings of the balance or the display panel.
Attach the in-use balance cover (standard accessory) when the balance is used in an environment where it is susceptible to being soiled.
The pan can be removed and washed with water. Verify that the pan is completely dry before replacing it on the balance.
1.7.2 Transporting the Balance
To carry the balance, hold the balance firmly with both hands.
To transport the balance, use the shipping carton used to deliver the balance.
Before placing the balance into the shipping carton, verify that the transportation screws at the bottom of the balance have been tightened until they stop.
Page 19
20
1.8 Specification
Type
Weighing
Capacity
(kg)
Minimum
Display (g)
Standard Deviation
(σ)g
Linearity
±g
Range of
the External
Calibration
Weight (kg)
CBX32KS
CBW32KS
32 1 0.6 1 30
CBX52KS
CBW52KS
52 1 0.6 1 50
CBX12KH
CBW12KH
12 0.1 0.1 0.2 10
CBX22KH
CBW22KH
22 0.1 0.1 0.2 20
CBX32KH
CBW32KH
32 0.1 0.12 0.2 30
The CBX-K series is not equipped with a built-in calibration weight. The CBW-K series is equipped with a built-in calibration weight.
Temperature coefficient of sensitivity: H-type = ± 4ppm/°C S-type = ± 5ppm/°C (Range of temperature compensation: 10 to 30°C)
Common Specifications
CBW-K CBX-K
Pan size (mm)
Approx.345 W x 250 D
Dimensions
Approx.360 W x 355 D x 125 H
Weight (kg)
Approx.16.5kg Approx 10.5kg
Operating
Temperature
Range
5 - 40
Power Supply
AC adapter (AC100V) or battery pack (special accessory)
RS-232C connector for RS-232C
DATA I/O connector for Electronic Printer, Printer
I/O Connector
KEY connector for application keyboard AKB-301
Page 20
21
2 MEMU ITEM SELECTION
2.1 Introduction
The CBX-W and CBW-K series balance has many functions that can be selected to meet the requirements of the user. Menu Item selection is used to program these functions. The symbol is displayed during Menu Item selection. Once the menu items have been set based on the installation environment and weighing purpose, it is not necessary to select the menu items each time the balance is used. Once the contents of the menu are set, they are stored even if the balance is turned OFF or if the power is disconnected.
2.2 Procedure of Menu Item Selection
The menu of the CBX-W / CBW-K balance consists of four levels.
1) Press the
key to cycle through the items within a menu level.
( in the table below)
2) Press the
key to choose the current item or move to the
next menu level. ( in the table below)
3) Press the
key to move back one menu level.
4) Press and hold the key to return to the mass display.
This instruction manual identifies each menu item by a number. For example,
, “Anti-Vibration mode” is identified as [25].
Refer to Section 2.3 "Menu Map" or the attached operation explanatory sheet.
Page 21
22
Symbols in the : Display
Press the
key when the following symbols are blinking to enter the
associated menu.
(Calibration) = Enters the menu related to span calibration.
(Graphic display) = Enters the menu related to the graphic
display, target and checkweighing.
(Environment) = Enters the menu related to the installation
environment and general operation of the balance.
(Application) = Enters the menu related to applications and
printing.
(Unit) = Enters the menu related to unit conversion.
(System) = Enters the menu related to the system (clock,
reports, and passwords).
(Communication) = Enters the menu related to communication
with a computer and software.
The stability symbol in the
: display indicates that
, , and are set.
Page 22
23
2.3 Menu Map
To enter the Menu Map: Turn the balance on.
Press the
key during mass display.
(This menu CAN NOT be entered from %, PCS,
, and displays.)
The selected span calibration type is displayed. The possible displays are:
# (CBW only)
(CBW only)
# (CBX or CBW)
(CBX or CBW)
Press the
key again to display the first menu level: :
Press the key repeatedly to cycle through the items or press the
key to select the blinking item.
Mass Weighing Display
# Performs selected standard span calibration.
: Selects the blinking item.
Moves to the next item
# Default settings
Page 23
2
4
Calibration Menu
symbol blinks.
Sets type of span calibration.
blinking Sets I-CAL. (Span calibration using
the built-in weight) # [1]
blinking Sets I-TEST. (Calibration check
Using the built-in weight) [2]
blinking Sets I-CAL. (Span calibration using
the external weight) # [3]
blinking Sets I-TEST. (Calibration check
using the external weight) [4]
blinking Executes calibration of the built-in
calibration weight. [10]
go to next page
Menu items [3] and [4] are not available on the CBX-K series balance.
# Default settings
Page 24
25
Graphic Display Menu
Graphic display blinking
blinking Full Scale mode # [11] blinking Target mode [12]
Target value [13]
Limit value [14]
blinking Group mode [15]
Upper threshold [16] Lower threshold [17]
blinking Level mode [18]
Upper threshold [19] Lower threshold [20]
blinking No graphic display [21]
go to next page
# Default settings
Page 25
26
Environment Menu
blinking
blinking Averaging
Auto # [22] Pouring [23] Standard [24] Anti-vibration [25] Anti-wind [26]
blinking Stability detection band
1 count # [27] 2 count [28] 4 count [29] 8 count [30] 16 count [31] 32 count [32] 64 count [33]
blinking Tracking
ON # [34]
OFF [35]
and blinking
Pretare with all lighting at start
[36]
blinking
Stops. [37] Does not stop #. [38]
blinking Taring/Print
Immediately # [39]
Waiting for stabilization
[40]
go to next page
# Default settings
Page 26
2
7
Applications Menu
blinking
symbol blinking Auto Zero On/Off [41] symbol blinking Auto Print
at + value [42] at - value and + value [43] at 0 and + value [44]
at - value, 0, and + value
[45]
Continuous Output [46]
Stability with GO [47] blinking : Zero Range [48] blinking
Peak Holding On/Off [49]
blinking Interval Timer PRINT
: : Time interval [50]
symbol blinking
Memory Weighing On/Off
[51]
symbol blinking
Auto-Memory and Zeroing
[52]
symbol blinking
Animal Weighing On/Off
[53]
go to next page
# Default Setting
Page 27
2
8
Page 28
29
System Setting Menu
blinking
blinking Current date setting [63] blinking Current time setting [64] blinking Display during standby setting
Time [65] Data [66]
No display # [67]
blinking Report and control setting
CAL report creating
Create [68]
Not create # [69]
Balance ID
: [70]
PCAL Password
: [71]
blinking
Menu reset to default [72]
go to next page
Note: Menu item [71] is not available with the CBX-K balances.
# Default settings
Page 29
30
Communication Settings Menu
blinking
blinking Handshaking
No handshake [73]
Software [74]
Hardware [75]
Timer # [76]
blinking Data Format
EB type # [77]
Old EB type [78]
PR type [79]
IPS type [80]
blinking Baud Rate (Bits/sec)
300 bps [81] 600 bps [82] 1200 bps # [83] 2400 bps [84] 4800 bps [85] 9600 bps [86] 19200 bps [87] 38400 bps [88]
blinking Parity & bit length
None (8 bits) # [89]
Odd (7 bits) [90]
Even (7 bits) [91]
blinking Stop bit
1 bit # [92] 2 bits [93]
blinking Delimiter
CR # [94] LF [95]
CR+LF [96] WindowsDirect down [97] WindowsDirect right [98]
Returns to blinking symbol.
# Default settings
Page 30
31
2.4 General Menu Operations
2.4.1 Setting Numeric V alues
Numeric values may be used to set the threshold of the comparator with the CBX-K and CBW-K series balances. Use the optional AKB-301 Application Keyboard to easily set these values. The values may also be set using only the balance keys.
In a menu used to set numeric values, and are both illuminated and the digit to be input blinks.
Press the
key to increase the value of the blinking digit by one.
→ )
Press the key to move the blinking digit one place to the right.
Press the key to store the displayed value in the balance
memory.
is displayed when the value has been successfully saved.
is displayed when the balance failed to save the value.
Press the
key to stop numeric entry.
is displayed briefly
and the display returns to the menu, one level up.
2.4. 2 Setting a Decimal Point
A decimal point is only used when setting units for solid density weighing, liquid density weighing or when setting the multiplier for the user defined unit. Set the decimal point while setting numerical values as follows.
Press the
key repeatedly until the last digit is blinking. Press the
key once more to initiate decimal point setting mode.
The symbol or current decimal point blinks.
Press the key to move the blinking decimal point one digit at a time to
the desired position.
Press the
key to set the decimal point position.
is displayed briefly to indicate that the setting is completed.
2.4. 3 Menu Lock
The CBX-K and CBW-K series of balances have a "Menu Lock" function that locks the menu selections to avoid changes being made by mistake.
The menu lock is toggled ON and OFF by pressing the
key during
display that appears after power is supplied to the balance.
is displayed when the menu is locked.
is displayed when the menu lock is turned off.
Page 31
32
Menu lock is on. Menu lock is off.
Note: Access is denied and
is displayed when the user attempts to select a locked menu. Use the following procedure to turn off the menu lock function and restore access to the menus.
1) Disconnect power from the balance and wait 10 seconds. Reconnect
power to the balance.
2) When
is displayed, press the key.
3)
is displayed to indicate that the menu lock has been
turned off.
2.4.4 Last Menu Recall
This function is convenient when an application requires frequent changes to a specific Menu Map item.
During mass display or menu selection, press and hold the
key for
approximately three seconds. The last Menu Map item that was changed or set is displayed.
Page 32
33
2.4.5 Returning to the Default Settings (menu reset)
The procedure below describes how to reset the menu and return to the default settings. Default settings are indicated with the # symbol in the Menu Map. (
Section 2.3 "Menu Map"). Select Menu Map item [72] to reset the menu.
1) In the mass display, press the
key repeatedly until the of
: blinks.
2) Press the
key.
The system setting menu is selected .
3) Press the
key repeatedly until the
in is
blinking.
4) Press the
key to display (" " means
"?").
5) Press the
key again.
is displayed to indicate menu reset completion.
6) Press the
key several times (or hold it for approximately 3
seconds) to return to the mass display.
Page 33
3
4
2.5 Calibration Execution Menu
2.5.1 Calibration
Calibration is required to accurately weigh items with an electronic balance. Calibration should be performed:
When the location of the balance is changed, even within the same
room.
When the room temperature changes considerably.
Periodically, according to the quality control plan of the user.
Terms used in this manual:
Span Calibration: Adjustment of the balance to specifications using two
weight values; zero and an appropriate value for the balance capacity.
Calibration Check: Comparing the current calibration mass reading to the
calibration mass reading after the last span calibration.
PSC: Span Calibration is performed automatically using the
built-in calibration mass when the balance detects a temperature change that would affect weighing accuracy.
Clock-CAL: Span Calibration is performed automatically using the
built-in calibration mass at up to 3 specific, preset times during each day. The user selects the times.
PCAL: Procedure to calibrate the built-in calibration mass to an
external calibration mass standard. The PCAL procedure is password protected.
Factory Default Settings:
CBX-K series Calibration using the external weight. ( ) CBW-K series Calibration using built-in weights. ( )
The settings for the type of calibration can be changed with menu selection.
Section 2.6.1 "Selecting Standard Calibration Type."
Note: The following displays may appear in the situations listed.
When it takes too much time to stabilize the balance. When the zero point of the balance appears to have changed greatly. When the balance calibration appears to have changed greatly or the
wrong weight is loaded on the pan during calibration. In all of these situations, span calibration or calibration check is not performed.
Refer to Section 4.4.2 “Error Display.
Page 34
35
2.5.2 Span Calibration Using the Built-in Weight,
1) Verify that the balance is in mass display and that the pan is empty.
2) Press the
key once, is displayed
(If
is not displayed, return to mass display and select Menu Map
item [1].)
3) Press the
key.
When
is displayed, and the stability mark () illuminates
on the left in the display unit, turn both of the calibration levers on the left and right of the pan to the back until they stop. At this point the stability mark will temporarily disappear.
4) Press the
key when the stability mark lights up again.
5) The display will show . Return the calibration levers to the measurement position. The stability mark will disappear temporarily when this is done.
6) Press the
key after the stability mark lights up again.
After awhile the
display will appear, and will be displayed. Span Calibration will be complete once the balance goes to kg display.
2.5.3 Calibration Check Using the Built-in Weight
1) Verify that the balance is in mass display and that the span is empty.
2) Press the
key once,
is displayed.
(If
is not displayed, return to mass display and select Menu Map
item [2].)
3) Press the
key.
When
is displayed, and the stability mark () illuminates
on the left in the display unit, turn both of the calibration levers on the left and right of the pan to the back until they stop. When this is done the stability mark will disappear temporarily.
4) Press the key after the stability mark lights up again.
The display will show
. Return the calibration levers to the measurement position. The stability mark will disappear temporarily when this is done.
5) Press the
key after the stability mark lights up again.
After awhile
xxx will be displayed (xxx will be numerals). This value indicates the extent of the current deviation in the balance sensitivity.
Page 35
36
6) When correcting the value to zero, press the key.
Press the
key if not correcting it to zero
(although pressing the
key will also leave it uncorrected, this will in
effect mean “Calibration Aborted”).
Note: Changing the
value to zero is equivalent to performing span calibration.
7)
is displayed, indicating the completion of the calibration check.
Note: Examples for interpreting the results of a Calibration Check:
Value Actual Mass
Displayed Mass
-0.0003 30kg 29.9997 (32kg/0.1g balance) +0.21 400g 400.21 (420g/0.01g balance)
Error codes that may be displayed:
indicates that the value is 1000 counts or more.
indicates that the value is -1000 counts or less.
2.5.4 Span Calibration Using External Weights
1) Verify that the balance is in mass display and unload the sample from the pan.
2) Press the
key once. is displayed.
(If
is not displayed, return to mass display and select Menu Map
item [3].)
3) Press the
key.
The value of the correct calibration mass to be loaded is displayed and blinks.
Changing the Calibration Mass to be Used
Pressing the key allows changes to the weight value. Modify the value
using the
key and key, then press the key. To
interrupt modification, press the
key.
4) Load the indicated calibration weight and press the key.
5) Shortly, zero display blinks. Unload the weight from the pan and press the
key.
is displayed briefly to indicate completion of span calibration.
Page 36
3
7
2.5.5 Calibration Check Using External Weights
1) Verify that the balance is in mass display and unload the sample from the pan.
2) Press the
key once to display .
(If
is not displayed, select Menu Map item [4].)
3) Press the
key.
The value of the correct calibration mass to be loaded is displayed and blinks.
Changing the Calibration Mass to b e Used
Pressing the key allows changes to the weight value. Modify the value
using the
key and key, then press the key. To
interrupt modification, press the
key.
4) Load the indicated calibration mass and press the key.
The zero display blinks.
5) Unload the weight from the pan and press the
key.
The display changes to the
xxx display. (xxx indicates a numeric value)
6) To change the
value to zero, press the key.
Press the key to avoid changing the value to zero.
Pressing the
key interrupts calibration and does not change this value
to zero.
Note: Changing the
value to zero is equivalent to performing span calibration.
Page 37
3
8
2.6 Calibration Setting Menu
2.6.1 Selecting the St andard Calibration T ype ( )
Menu Map item [1]
blinking Selects "Span calibration
using the built-in weight."
Menu Map item [2] blinking Selects "Calibration check
using the built-in weight."
Menu Map item [3]
blinking Selects "Span calibration
using external weights."
Menu Map item [4]
blinking Selects "Calibration check using
external weights."
2.6.2 PCAL - Calibration of the Built-in Weight -
PCAL is used to calibrate the built-in weight to a standard calibration weight that is correctly adjusted, traceable and/or certified.
Use a correctly controlled, precise calibration mass for this procedure. If it is performed without a correct calibration mass, span calibration and calibration checks using the built-in mass may not be correct in subsequent operations.
1) Unload the sample from the pan and verify a zero mass display.
2) Select the Menu Map item [10].
: is displayed.
3) Enter the PCAL password using the
and keys, then press
the
key.
The default password is 9999, set at shipment or upon menu reset.
4) After is displayed, the value of the standard weight to be
loaded blinks.
Load the standard mass displayed and press the key.
5) Shortly, zero is displayed and blinks. Unload the weight and press the
key.
6) From this point the display will proceed automatically. When
is displayed,
operate the levers on both sides of the pan to put on the built-in calibration weight.
Page 38
39
7) Press the key when the stability symbol re-illuminates after
having disappeared for a moment.
8) When
is displayed, operate the levers on both sides of the pan to take off the
built-in calibration weight.
9) Press the
key when the stability symbol re-illuminates after having
disappeared for a moment.
10) From this point the display will proceed automatically. The procedure is complete when the balance goes to mass display mode.
Note: is displayed during the process. Leave the balance in a
stable state until the mass display appears as in step 6.
In PCAL, the value of the "weight to be loaded" cannot be changed.
Set the PCAL password using Menu Map item [71].
2.7 Graphic Display Menu :
2.7.1 Overview
The CBX-K and CBW-K series balance has a bar graph located on the left side of the display. This graph can be conveniently used for liquid weighing and pass or failure judgment (comparator function) of the sample. The graphic display functions include the Full Scale mode, Target mode, Group mode, Level mode, and no graphic display mode.
Note: Target, limit, upper, and lower values are set as numeric values only. Set the correct numeric
value for the unit that will be used for weighing.
In weighing mode, changing the displayed unit does NOT change the target, limit, upper, or lower numeric values. For example, when the upper limit has been set at 10g, switching the unit from “g” to “kg” does not change the limit to 10kg. It will be 0.01kg.
The decimal point is invisible in the value setting display. Determine its position based on the
resolution of the balance. For example, if the balance’s minimum display is 0.01g, 150g must be set by inputting “15000” in the display.
Numeric values are memorized independently for each graphic display mode.
The graphic display mode can be selected without setting numeric values.
2.7.2 Full Scale Mode
The relative amount of the load on the pan is displayed in the bar graph. This feature helps to prevent errors due to OL (overload) status. Use Menu Map item [11] to select the Full Scale mode. A bar displayed in the lower areas of the scale indicates that the load on the pan is small. A bar displayed in the upper areas of the scale indicates that the load on the pan is close to the weighing capacity.
2.7.3 T arget Mode
Target mode is useful for filling applications or for quality check by weight.
Page 39
40
The value is the numeric value that is the desired amount in the unit that is used for weighing. The
value is the numeric amount above or below the target value that is acceptable. This mode is useful for constant amount weighing of liquid or judgment of excess and shortage.
Menu Map item [12] selects the Target mode. Menu Map item [13] sets the
value, which corresponds to the center line of the
graphic display.
Menu Map item [14] sets the
value, which corresponds to the distance between the
center line and upper or lower triangles.
2.7.4 Group Mode
This is the best mode to determine pass or failure judgment based on the sample weight.
Menu Map item [15] selects the Group mode. Menu Map item [16] sets the upper threshold value, which corresponds to the upper triangle. Menu Map item [17] sets the lower threshold value, which corresponds to the lower triangle.
Note: Determination as follows: Display
< Sample weight HI < Sample weight < GO
Sample weight <
LO
2.7.5 Level Mode
Use this mode for classification based on the sample weight. The display looks like a bar graph, but also includes a comparator function.
Menu Map item [18] selects the Level mode. Menu Map item [19] sets the upper threshold value, which corresponds to the upper triangle. Menu Map item [20] sets the lower threshold value, which corresponds to the lower triangle.
Note: Determination as follows: Display
< Sample weight HI < Sample weight < GO
Sample weight <
LO
2.7.6 No Graphic Display
Menu Map item [21] turns off the graphic display.
Page 40
41
2.8 Environment Menu :
2.8.1 Overview
Settings on the balance can be changed to compensate for the installation environment such as the degree of vibration etc. or for the purpose of weighing a solid, liquid or powder.
2.8.2 Averaging
It is possible to match the stability of the display and the degree of response with the requirements of specific applications.
Menu Map item [22] (Environment/Averaging-AUTOmatic)
The balance automatically performs optimum averaging dynamically while observing the load data. This is the recommended setting and should be used unless special circumstances exist.
Menu Map item [23]
(Environment/Averaging-POURing) This mode is only suitable for constant volume weighing of liquids etc. This mode is very sensitive to wind and vibration.
Menu Map item [24]
(Environment/Averaging-STaNDard) This mode is suitable for weighing in a normal environment. Averaging is fixed and does not change dynamically as in
.
Menu Map item [25] (Environment/Averaging-VIBRation)
Use this mode when the balance is used in a location where there are large vibrations and the display fluctuates in the
mode.
Response time is deteriorated by small mass amount changes.
Menu Map item [26]
(Environment/Averaging-WIND) Use this mode when the balance is used in a location where it is exposed to airflow that causes the display to fluctuate in the
mode. Response time deteriorates further than , but weighing is comparatively stabilized.
Note: If weighing cannot be performed efficiently even with
, change the
installation site of the balance or use the deluxe windbreak or large pan windbreak.
2.8.3 Stability Detection Band
Menu items [27] to [33] are used to determine the conditions for indicating balance stability.
(Environment/Band-1): When the display has remained constant (within one display count), the balance is regarded as stable and the stability symbol illuminates. Stability detection band settings can be selected from to
.
Note: Use
to only when the Auto Print function is used and it is therefore required that the sensitivity to vibration be reduced to allow the balance to be regarded as stable and print the measurement value. The balance may not operate properly if these settings are selected under normal use conditions and environment.
Page 41
42
2.8.4 T racking
Tracking is the function that will maintain the current displayed value as long as possible.
Menu Map item [34] Turns ON this function. Menu Map item [35]
Turns OFF this function.
Zero Tracking Function
When the display is zero, menu functions as "zero tracking" to keep the zero display as long as possible. "Zero tracking" automatically cancels small zero drift.
It is recommended to set to
when measuring slight mass change such as in the
process of drop addition or liquid evaporation.
2.8.5 Pretaring V alue
This function is used to weigh the mass of a sample packed in a container such as a bottle or bag without opening the container. Pretare function should be used only if the mass of the containers does not vary from sample to sample. Instead of zero the pretare value is displayed (as a negative value)
when the
key is pressed. The pretare value is then subtracted from the load on the
balance pan to determine the weight of the sample. Menu Map item [36] opens the Pretare setting screen.
Set the Pretare value using the
key and key, and press the key.
Pretare Value
Cancel the Pretare value by setting the value to zero. When a Pretare value other than zero has been set,
illuminates. The Pretare value is set using the "kg" unit. The maximum value is dependant on the weighing capacity of the balance.
Check the Pretare value by pressing and holding the key for approximately 3 seconds
during weighing. To set a pretare value using the optional AKB-301 Application Keyboard, press the [Pretare value] and
the
key at the same time. Use the AKB-301 if you have to change the Pretare value
frequently.
Page 42
43
2.8.6 Whole Lighting Mode
Press the key to switch from the standby mode to the mass display. The entire display is
illuminated. This function sets whether the display automatically proceeds to mass display or waits for
a
key command. When the mode has been set to , it is easier to
check the display because the display remains illuminated.
Menu Map item [37]
Display stops at whole lighting. Press the
key to proceed to
the mass display.
Menu Map item [38]
After the whole lighting display has appeared for approximately 0.5 sec, the display automatically proceeds to the mass display.
2.8.7 T aring/Printing
Determine if the balance should wait for stability before printing when the key is pressed or
displaying the zero point when the
key is pressed.
When the Application Measurement function is selected, the
key does not wait for stability.
Menu Map item [39]
Print and tare operate without waiting for stabilization.
Menu Map item [40]
Print and tare operate after the stability symbol illuminates.
Note: When the balance is in the
mode and waiting for stability to be determined, -
--- is displayed when the
key is pressed.
Press the
key to disable this function. Taring is not executed after the key is pressed.
Once the
key is pressed, data is not output until the and symbol has illuminated and the
balance is stable. Data is output only after the stability symbol is lit.
Page 43
4
4
2.9 Application Menu :
2.9.1 Overview
The Application Measurement function is the generic name of the functions used to perform measurement work using the balance in more efficient ways. Only one Application Measurement function can be used at a time. When the power is turned ON, the balance will begin in the function mode that was last set.
Cancellation of Application Measurement
To cancel the Application Measurement, press the key while in the mass display and hold it
until
is displayed.
To use the Application Measurement function that was just canceled, use the last menu recall
function by pressing and holding the
key for 3 seconds in the mass display.
Using the Application Measurement Function with a Unit Other Than "kg"
Select the Application Measurement function while in the mass display, and select the unit of
weighing with the
key.
Because the operation of the
differs from the one in normal mass display, the Application
Measurement function cannot be selected while %, PCS, or specific gravity units are displayed.
2.9.2 Zero Range
The "Zero Range" value is used in the Application Measurement function as a reference to identify whether the sample is loaded. For example, if the display is within the Zero Range, the balance determines that a sample is not loaded. If the display is more than five times the Zero Range it determines that a sample is loaded. Application Measurement functions which involves judgment of whether a sample is loaded or not work in accordance with the Zero Range setting. In Menu Map item [48], the "Zero Range" value is set using the number of counts displayed in kilogram-display. The setting range is 01 to 99 with 01 being the default value. Even when weighing will be done in another unit, Zero Range setting is made by only gram value.
Note: When a Pretare value is set, the value to determine that there is "no load" becomes
"- Pretare ± Zero Range" during mass display. For example, the Zero Range function works as expected when attempting to weigh and Auto Print the mass of a bottled sample during mass display.
Page 44
45
2.9.3 Auto Zero
Select Menu Map item [41]. When the displayed value is within the Zero Range and the stability symbol has illuminated, zeroing occurs automatically. The symbol appears in the display when the Auto Zero function is active. Other keys function as expected with the Auto Zero function activated.
2.9.4 Auto Print
Auto Print function allows output of the data automatically without pressing the key for each
sample. The
symbol is illuminated when the Auto Print function is activated.
Six types of Auto Print are possible.
Menu Map item [42] (on positive load)
Load the sample when the value displayed is within the Zero Range. When the stability symbol has illuminated and the positive displayed value is more than 5 times the Zero Range, data is automatically output. The next data output is not performed unless the display has returned to a value within the Zero Range by unloading the sample or pressing the
key.
Menu Map item [43] (on positive or negative load)
Load or unload the sample when the displayed value is within the Zero Range. When the stability symbol has illuminated and the displayed positive or negative value is more than 5 times the Zero Range, data is automatically output. The next data output is not performed unless the display has returned to a value within the Zero Range by unloading the sample or pressing the
key.
Menu Map item [44] (on positive load and on zero)
Load the sample when the value displayed is within the Zero Range. When the stability symbol has illuminated and the positive displayed value is more than 5 times the Zero Range, data is automatically output.
Unload the sample or press the
key. When the displayed
value is within the Zero Range and the stability symbol has illuminated, data is output again.
Menu Map item [45] (on positive or negative load and on zero)
Load the sample when the value displayed is within the Zero Range. When the stability symbol has illuminated and the displayed positive or negative value is more than 5 times the Zero Range, data is automatically output.
Unload the sample or press the
key. When the displayed
value is within the Zero Range and the stability symbol ha illuminated, data is output again.
Page 45
46
Menu Map item [46] (on /continue)
By pressing the
key while the and symbols are lit,
the
symbol goes out, the symbol lights and the displayed
data is continuously output.
Continuous output stops temporarily when the
key is pressed.
Note: During continuous output, the
symbol may appear to remain lit. If the transfer speed of the data output is slow, the display may flash. Increase the transfer speed as much as possible and set the handshake to .
Menu Map item [47]
(on / go) When the graphic display is in Comparator mode (Group mode or Level mode) and the stability symbol lights in the GO range, the data is output once. The next data output is executed after the displayed value is within the Zero Range.
2.9.5 Peak Hold
Menu Map item [49] measures the displayed peak value. The symbol is illuminated when the Peak Hold function is activated. "Peak value" is the highest or lowest stable value displayed after the display has changed beyond five times the Zero Range.
1) In the peak detection standby state with the
and symbols illuminated, press the
key to tare the display.
2) Press the
key.
The
symbol disappears and peak value detection starts.
3)
and * are simultaneously displayed after the peak value is detected, and the data is output.
This display will not change regardless of the load on the pan.
4) Press the
key.
The balance returns to the peak detection standby state in step 1).
Note: Press the
key in the peak detection standby state to initiate the power standby state.
Press the
key during detection of the peak to return to the peak detection standby state.
Polarity of the peak value displayed is "polarity of the displayed value of the first change by five times or more of Zero Range from the display within Zero Range." Usually the peak value is easily measured by setting to
[23]. Depending on the
weighting conditions and the sample type, this setting may not always be successful.
Page 46
4
7
2.9.6 Interval Timer
Automatically outputs the displayed value at preset intervals. The symbol is illuminated when the Interval Timer is activated. The optional AKB-301 Application Keyboard is convenient when the interval is changed frequently.
1) Using Menu Map item [50], set the output interval (00:01 = 1 sec to 99:59 = 99 minutes 59 seconds).
2) In the interval timer standby state when the
and symbols are both illuminated,
press the
key.
The first data is output. Data is automatically output at the set time intervals.
3) To stop output, press the
key.
The balance returns to the interval timer standby state in step 1).
Notes:
Use the key to erase the tare or zero the balance at any time.
Pressing the key while in the interval timer standby state causes the power supply standby
state to be initiated.
To release the interval timer function, keep pressing the key. This does not reset the
interval timer function to zero.
Using the interval timer function to record data over a long period may cause data error due to
balance drift.
Some instruments receiving the data may not operate normally if the set time interval is short. To
correct this, set the time interval to a longer period. When the set time interval is short and the instrument connected to DATA I/O is unknown, it is recommended to set the handshake to a setting other than
[76].
2.9.7 Auto-Me mory and Zeroing
Use Menu Map item [51]. This menu is used to weigh a large number of individual samples. The
(Auto-Memory and
Zeroing) symbol is illuminated when this function is active.
1) Load the weighing vessel and press the
key in the Auto-Memory and Zeroing
standby state (
and symbols are lit).
Zeroing occurs.
2) Press the
key.
The
symbol disappears, and Auto-Memory and Zeroing measurement starts.
3) Load the first sample. Each time the stability symbol is lit and the display is a value five times or
more the Zero Range or the
key is pressed, the displayed value is output and zeroing
occurs.
Page 47
4
8
4) For the next sample, additional weighing is performed without pressing the key.
5) Press the
key.
The balance returns to the Auto-Memory and Zeroing standby state and total mass on the pan without the tare is displayed.
Press the
key to print this value.
Notes:
When the stability symbol is illuminated and the displayed value is within the Zero Range, zeroing occurs automatically.
When the
key is pressed and the displayed value is five times the Zero Range, zeroing occur
after data output. (Manual loading)
When the key is pressed in the Auto-Memory and Zeroing standby state, the power supply
standby state is initiated.
2.9.8 Animal Weighing
Set with Menu Map item [53]. This function is suitable for weighing animals.
(animal) symbol is illuminated when the Animal
Weighing mode is active.
1) Load the weighing vessel and press the
key.
Note: Data may be output when the weighing vessel is loaded. This is not a malfunction.
2) Load the sample (animal etc.) with a mass more than 50 times the Zero Range.
3) When the weighed value is relatively stable, the value is automatically output.
4) Press the key or unload the sample.
5) When the displayed value is stable and less than 10 times the Zero Range, automatic zeroing occurs. Any residue remaining on the pan (excrement or fur) is automatically canceled and zeroing occurs. If zeroing does not occur, increase the value of the Zero Range.
Notes:
Standby state is not available in the Animal Weighing mode.
Press the
key to initiate the power supply standby state.
On the premise of weighing animated objects, the stability detection band is automatically extended in the Animal Weighing mode. Reproducibility of the measurement data is slightly less than with other modes. When the animal being weighed cannot be controlled and Auto Print will not activate, press the
key to output the displayed value. Then unload the animal. Even if the stability symbol lights
before the animal is removed, data is not printed again. By setting a larger stability detection band in the menu, the stability symbol will light more readily. If the balance is slow to return to the zero point set a larger Zero Range value.
Page 48
49
2.10 Unit Registration Menu and Unit Change
2.10.1 Unit Registration Menu
Press the key in mass display to sequentially change the registered units. It is possible to
display units other than "kg" with the CBX-K and CBW-K series of balances. Before weighing, set the display units to be used with the unit setting menu [54] to [62]. (Unit kg, %, number (PCS) were set before shipment.)
Note:
In the unit setting menu, the stability symbol is illuminated to indicate the currently set units.
Set or release the unit by pressing the key when the unit is displayed.
Functional Units: Solid specific gravity unit (▼d): Density of the liquid in which the sample is immersed. Liquid specific gravity unit (d): Volume of the reference weight to be immersed in the sample.
unit (►): Numeric value (multiplier) multiplied by the gram (g)
weight of the unit.
To register Functional Units, set the constant to something other than 0. Setting the constant to 0 turns the function off. For the unit names, refer to
Section 2.3 "Menu Map."
2.10.2 % Conversion
1) Set the % unit with Menu Map item [56]. The % unit is set before shipment.
2) Press the
key several times in the mass display until the % unit is displayed.
Setting the 100% reference
Press the
key to tare the balance.
Load the reference sample that corresponds to the 100%value. This value must be equivalent to 100 counts or more in the "kg" unit.
When the stability symbol illuminates, press the
key.
is displayed briefly and the reference sample weight is displayed as 100%.
Page 49
50
Note: The numbers of digits displayed in the % unit and rounding off of the minimum digit vary
depending on the mass value of the reference sample and the balance model. It is not possible to obtain resolution greater than that in the “g” unit. The weight of subsequent samples are displayed as a percentage of the reference sample weight.
2.10.3 Piece Counting
1) Register the PCS unit with Menu Map item [57]. (The PCS unit is registered before shipment.)
2) Press the
key several times in the mass display until the PCS unit is displayed.
3) Load the container and press the
key to tare the balance.
4) Count exactly five pieces (or 10, 20, 50, 100, or 200 pieces) of sample to be measured and load them on the pan.
5) Press the
key.
6) Every time the
key is pressed, the display sequentially changes as
pcs pcs pcs ... (Piece menu).
The default setting is
pcs.)
Press the
key when the display is equivalent to the number of loaded pieces.
Example: If 50pcs. are loaded, press the
key when pcs is displayed.
This determines the unit weight or average weight per piece. As sample is added or removed, the piece count (number of pieces) is displayed.
Note: Repeat steps 3 through 6 above when the sample or manufacturing lot is changed.
Piece Count Menu Display a Next Setting
In this example, when the key is pressed in the next PCS menu, display starts from
pcs.
Counting Pieces Using the Optional
AKB-301 Application Keyboard
AKB-301 Application Keyboard By using the optional AKB-301 Application Keyboard, piece count based on an arbitrary number of pieces is possible. The unit weight itself, if it is known, can be input using the numeric keys.
Section 3.5 "AKB-301 Application Keyboard".
Page 50
51
2.10.4 Solid Specific Gravity Measurement
Solid specific gravity measurement refers to the measurement of the sample (solid) weight in the air and in a liquid of known density and the calculation of the sample density. The ▼d symbol is used to represent the solid density unit in this balance. The data output unit is DS.
1) Set Menu Map item [60].
Note: Enter the value for the density (g/cm3) of the liquid (water, alcohol etc.) in which the sample is
immersed.
2) Attach the optional weighing hook to the bottom of the balance, attach the hanging pan, and then immerse the hanging pan in the tank filled with the liquid of known density.
3) Press the
key until ▼d is displayed.
4) Press the
key.
5) Load the sample on the balance pan or in the hanging pan in air.
6) After the stability symbol illuminates, press the key.
may be displayed but does not indicate a malfunction.
7) Load the sample on the hanging pan immersed in the liquid. The density of the sample is displayed.
8) Repeat steps 4 through 7 for each additional sample.
Note:
Up to four decimal places are displayed for specific gravity. Since it may not be possible to
stabilize the balance using all 4 places, 1d/10d switching is possible.
When loading the sample on the pan in the liquid, ensure that the entire sample is immersed in the
liquid.
The balance does not re-zero when the key is pressed in this unit display.
When the optional Specific Gravity kit is used, refer to the Instructional Manual of the Specific
Gravity kit for the procedure.
Return the balance to kg display mode first before performing the next menu operation.
Page 51
52
2.10.5 Liquid Specific Gravity Measurement
Liquid specific gravity measurement refers to the measurement of the weight of a reference solid of a known volume in air and in the sample liquid. Specific gravity for the liquid is calculated from these two values of. The display unit for liquid specific gravity is "d." The data output unit is DL.
1) Set Menu Map item [61].
Note: Enter the value for the volume (cm3) of the reference weight.
2) Attach the optional weighing hook to the bottom of the balance, attach the hanging pan, and then immerse the hanging pan in a tank containing the sample liquid.
3) Press the
key until "d" is displayed.
4) Press the
key.
5) Load the reference weight on the pan of the balance.
6) After the stability symbol is illuminated, press the key.
may be displayed but does not indicate a malfunction.
7) Load the reference weight on the hanging pan and immerse it in the sample liquid. The specific gravity of the sample liquid is displayed.
8) Repeat steps 4 through 7 for each additional sample.
Note:
Up to four decimal places are displayed for specific gravity. Since it may not be possible to
stabilize the balance using all 4 places, 1d/10d switching is possible.
When loading the sample on the pan, immerse the entire sample in the liquid.
T release the d, d or units, set the value to zero.
When the optional Specific Gravity kit is used, refer to the Instructional Manual of the Specific
Gravity kit for the procedure.
Page 52
53
2.1 1 System Setting Menu :
The SYSTEM SETTING menu is used to set the items that pertain to or are controlled by the balance.
2.1 1.1 Date for the Built-in Clock
1) Select Menu Map item [63].
2) Press the key.
3) Set the last two figures of the year, month and day, using the
and keys. Then
press the
key.
Example:
February 1st, 1997 Set as
. .
February 29th, 2004 Set as
. .
Note:
The built-in clock corrects for the leap year automatically.
The moment the key is pressed in step 2 above, seconds are set to zero. If the date
is set after setting the time, the second value will be incorrect. It is important to set the date first and then the time, or to correct the seconds value using the ± second correcting function described later.
2.1 1.2 Time for the Built-in Clock
1) Select Menu Map item [64].
2) Press the
key.
3) Use the
key and keys to set the time in the 24 hour system, then press the
key.
Example: 1 o'clock 23 minutes in the afternoon Set as
:
Note: The moment the
key is pressed is 00 seconds.
2.1 1.3 Display During St andby
Determine what is to be displayed during a power supply standby.
Menu Map item [65]
Displays the time during standby. Menu Map item [66] Displays the date during standby. Menu Map item [67]
Displays nothing during standby.
Page 53
5
4
Convenient Functions of Time Display
The following functions are available when the time is displayed during standby. Seconds display function:
Press the
key to enable the display/non-display of the seconds value.
2.1 1.4 Measurement Control System
Items related to the calibration of the balance and those set by the administrator are summarized in this menu.
2.1 1.4.1 Calibration Report
Turns the calibration report function on/off. Use this to generate a calibration report as for GLP, GMP, or ISO9000. An electronic printer (special accessory) is required to print the report.
Menu Map item [68] Calibration report is created. Menu Map item [69]
Calibration report is not created.
2.1 1.4.2 Balance ID
Individual balances can be identified by the serial number on the main body of the balance. The user can add a four-digit ID number to the calibration report.
Set with Menu Map item [70]. Use a 4-digit number from
to .
2.1 1.4.3 PCAL Password
Enter a 4-digit number from to with Menu Map item [71]. It is recommended that the balance administrator set this password to prevent an unauthorized person from incorrectly calibrating the built-in calibration mass.
Section 2.6.4 "PCAL - Calibration of the Built-in Weight"
Note: When the menu is reset, the PCAL password is reset to .
Page 54
55
2.12 Communication Menu ( blinking)
2.12.1 Overview
This menu is used to set the specifications for communication between the balance and a personal computer or electronic printer.
Note: This menu affects both the RS-232C and DATA I/O at the same time. For the instrument to be
connected to the DATA I/O connector of an electronic printer set the specifications for communication of the balance to the default settings, which are
[76], [77],
[83],
[89], [92], and [94].
2.12.2 Handshaking -
Handshaking determines whether the peripheral equipment can receive communication data from the balance. This function does not relay the status of the balance to the peripheral equipment. The balance is able to receive as long as there is space in the receiving buffer of the balance. This function operates once is displayed, operation in other states is not guaranteed. When the balance output is retained by handshaking, the display of the balance is locked. Determine the specifications for handshaking.
Menu Map item [73]
Software handshaking is not performed.
Menu Map item [74]
Software handshaking is performed. After the balance receives X-OFF (13H), the balance output is retained. After the balance receives X-ON (11H), the balance output is initiated.
Menu Map item [75]
Hardware handshaking is performed. When DTR is OFF, the output from the balance is retained. When DTR is ON, the output from the balance is initiated.
Menu Map item [76] Timed hardware handshaking is performed.
2.12.3 Format
Set the balance output format.
Menu Map item [77]
The standard format for the CAS electronic balance.
Menu Map item [78]
The old output format for the CAS electronic balance
The old output format is employed in the following models. EB-500, 5000, 280, 2800, AEL-1600, EB-50K (except -15)
Note: In this format, the number of the lowest place of Menu Map item [70] is assigned
to identify the balance.
Menu Map item [79]
Compatible format for the PR and SR series of CAS
electronic balances. Available commands, functions and responses are limited.
Menu Map item [80]
Compatible format for the IPS series of CAS electronic
balances. Available commands, functions and responses are limited.
Page 55
56
2.12.4 Communication Speed
Select the communication speed (300, 600, 1200, 2400, 4800, 9600, 19200, or 38400 bps). Number of xxx shows bps (bits/second). Baud rate and bps are the same value.
Set with Menu Map items [81] to [88].
2.12.5 Parity / Bit Length -
Select the parity and bit length.
Menu Map item [89]
No parity, 8-bit length
Menu Map item [90]
Odd number parity, 7-bit length
Menu Map item [91]
Even number parity, 7-bit length
2.12.6 Stop Bit
Select the number of stop bits.
Menu Map item [92]
Stop bit 1.
Menu Map item [93]
Stop bit 2.
2.12.7 Delimiter
The "delimiter" is used to separate individual pieces of data or commands. Set the delimiter as
follows:
Menu Map item [94]
Set to CR(0DH).
Menu Map item [95]
Set to LF(0AH). Menu Map item [96] Set to CR+LF(0D0AH). Menu Map item [97]
Transfers the data directly to Microsoft® Windows®. This is equivalent to pressing the Enter key of the personal computer.
Menu Map item [98]
Transfers the data directly to Microsoft® Windows®. This is equivalent to pressing the right cursor key of the personal computer.
Note:
Section 3.2.4 "Windows® Direct". When these Menu Map items are set, it is not possible to send commands to the balance from the peripheral instruments. The personal computer and electronic printer cannot be used at the same time.
Page 56
5
7
3. Connecting Peripheral Instruments
A variety of peripheral instruments are available for use with the CBW and CBX series balance, such as an electronic printer, keyboard or personal computer. This chapter describes how to connect and use peripheral instruments.
3.1 EP-60A Electronic Printer
1) Use the default communication settings for the balance listed below.
Menu [76] [77] [83] [89] [92] [94]
2) Disconnect the power to the balance and EP-60A Electronic Printer.
3) Connect the DATA I/O connector of the balance to the EP-60A using the attached cable of the EP-60A.
4) First turn ON the power to the balance, then to the EP-60A.
3.2 Personal Computer - RS-232C -
3.2.1 Connecting the Cable
Signals other than RS-232C are also output by the CBW and CBX series balances through the RS­232C/AUX connector. If these signal lines are incorrectly connected, damaged may occur to the personal computer or balance. Correctly connect an appropriate cable for communication between the balance and personal computer. All types of personal computers may not operate normally when connected with the optional RS­232C cable according to the diagram.
Page 57
5
8
(1) IBM PC/AT and its compatible personal computers (D-sub 9-pin)
Personal computer side Balance side
FRAME SHIELD FRAME
(This wiring cable is not always necessary.)
(2) IEEE Standard (D-sub 25-pin)
Personal computer side Balance side
1 ------------ 1 2 ------------ 3 3 ------------ 2 4 ------------ 5 5 ------------ 4 6 ------------ 20 7 ------------ 7
20 ------------ 6
3.2.2 Data Format
The following explanation is applied when is selected. For other formats, refer to the data corresponding to the compatible machine.
Note: ٱ indicates space code and <delimiter> indicates delimiter code.
(1) For the measured value` First character: Minus: '-' Non-minus: space 2nd to 11th characters: Numeric values or " [", "]" are flushed right. The position of the decimal point varies depending on the type of instrument. 12th to 13th characters: Unit such as gٱ or kg 14th to 15th characters: Delimiter
Note: When the delimiter is
or , the 13th character is not present.
When printing stability information, the following characters are put in front of the first character
Page 58
59
mentioned above.
Stable time: S Unstable time: D
When the delimiter is
or , the special format is used instead of that
described above.
(2) For of
or
ٱٱٱٱٱ OL ٱٱٱٱ <Delimiter>
- ٱٱٱٱ OL ٱٱٱٱ <Delimiter>
3.2.3 Using Command Codes
Note: If communication conditions are incorrectly set, a communication error is
displayed.
(1) Commands that end with a number, character, or symbol other than [=]: Transmit to the balance
with a delimiter for each command code.
Example 1: PRINT<CR> ... The same operation as pressing the
key
(2) Commands that end with a [=]: Transmit the number to the balance with a delimiter.
Example 2: TIME=1234 <CR> .. 12:34 is set as the current time. Example 3: P.TARE=1.23 <CR> (example of type of the second decimal place).
…1.23g is set as a Pretare value.
Example 4: P.TARE=0.00 <CR> (example of the second decimal place)
…Clears (cancels) the Pretare value.
Note: Number of digits, decimal point, position of decimal point of the numeral transmitted succeeding
to '=' are the same as the case of setting the numeric value using the AKB-301 Application Keyboard. Number of digits below the decimal point is as follows. S-type and H-type: Use the same number of decimal places as in the gram-display.
This restriction does not apply to USER=, SOLID=, and LIQUID=.
Example 5: MENU=4630 <CR> (4-digit number after [=])
This is the same in the following operation.
... from mass display, press the
key four times, the press the
key once and the press the
key six times. This returns to the same display
as if the
key was never pressed.
Press the
key once, the key three times, and the
key once more.
This example indicates that
is selected.
Note: If there is 0 in the four-digit number, the setting is complete at that point and menu selection is
ended. The result of this command varies depending on the type of the balance.
Page 59
60
Example 6: #=2.56 <CR> Example 7: #=12.345.67 <CR>
A personal computer can instruct the weighing and display a specific number on the balance. With the commands in Example 6 &7, [#2.56] and [#12.345.67] are displayed on
the balance. When the operator presses the
key, the character string '2-
56<CR>' and '12-345-67<CR> are output from the balance.
(3) Echo back command
The balance again transmits the character strings of N pieces included between an echo back command '{' or '}' and the delimiter. An unprocessed echo back command is not left in the receiving buffer of the balance, N≤30.
Example 8: ABCDEFG12345<CR>
... After receiving this command, the balance outputs ABCDEFG12345<CR>. The printer can print this character string.
Note: Only capital alphabets and a part of symbols (decimal point, symbol etc.) can be used when
printing with an electronic printer. A maximum of 15 characters per line.
(4) Command codes for
and
(i) Commands related to output
D01 Continuous output D03 Continuous output with stability information D05 Single output D06 Auto Print setting (type of Auto Print is set separately) D07 Single output with stability information D09 Release of continuous output and Auto Print
(ii) Commands related to operation keys
POWER Equivalent to the
key
Q Equivalent to the
key
MENU Equivalent to the
key
TARE Equivalent to the
key
T Equivalent to the
key
RANGE Equivalent to the
key
B Equivalent to the
key
UNIT Equivalent to the
key
PRINT Equivalent to the
key
Page 60
61
POWER+ Equivalent to holding the key for approximately 3 seconds.
MENU+ Equivalent to holding the
key for approximately 3 seconds.
UNIT+ Equivalent to holding the
key for approximately 3 seconds.
PRINT+ Equivalent to holding the
key for approximately 3 seconds.
RECALC Equivalent to the [RECALC] key of the AKB-301Application Keyboard C Equivalent to the [C] key of the AKB-301 Application Keyboard
(iii) Commands related to Application Measurement
PEAK Sets the Peak Hold mode. AZERO Sets the Auto Zero mode ON. INTERVAL Sets the Interval Timer mode. MEMORY Sets the Memory mode. M Immediately operates after setting the Memory mode,. ADDON Sets the Auto-Memory and Zeroing mode. + Immediately operates after setting the Auto-Memory and Zeroing mode,. A Sets the Animal Weighing mode. ANIMAL Sets the Animal Weighing mode. R Releases the Application weighing mode.
(iv) Commands related to unit conversion
g Switches to "g" unit kg Registration of "kg" unit and switching mg Registration of "mg" unit and switching PERCENT Registration of "%" unit and switching % Sets 100% when display is in "%" unit. G kg <- -> % switching PCS Registration of "PCS" unit and switching CT Registration of "karat" unit and switching MOM Registration of "monme" unit and switching SDENSE Registration of "solid density" unit and switching LDENSE Registration of "liquid density" unit and switching CU Switches to "user" unit (Set the conversion coefficient beforehand.) RSTUNIT Returns the default units.
(v) Readout commands of set value
TARGET Readout of target set value LIMIT Readout of limit set value G.LO Readout of lower limit set value in Group mode G.LO Readout of upper limit set value in Group mode L.LO Readout of lower limit set value in Level mode L.UP Readout of upper limit set value in Level mode UW Readout of unit weight set value G/PCS Equivalent to g/PCS key
Page 61
62
CALWT Readout of external weights set value for span calibration ACALT1 Readout of ACAL time 1 ACALT2 Readout of ACAL time 2 ACALT3 Readout of ACAL time 3 P.TARE Readout of Pretare set value ZRNG Readout of Zero Range set value USER Readout of user unit conversion coefficient VOL Readout of reference weight set value DENSE Readout of surrounding liquid density set value I.TIME Readout of Interval Timer set value
Note: The command to read out the numerical value settings can only be given by the personal
computer.
(vi) Commands for numeric value setting
CALWT= Sets external weights value for span calibration. ACALT1= Sets ACAL time 1. ACALT2= Sets ACAL time 2. ACALT3= Sets ACAL time 3. P.TARE= Sets Pretare value. ZRNG= Sets Zero Range value. UW= Sets unit weight. USER= Sets user unit conversion coefficient. VOL= Sets volume of reference weight. SDENSE= Sets surrounding liquid density. I.TIME= Sets interval timer value. DATE= Sets the date. TIME= Sets the time. TARGET= Sets the target value. LIMIT= Sets the limit value. G.LO= Sets the lower limit value of Group Display mode. G.UP= Sets the upper limit value of Group Display mode. L.LO= Sets the lower limit value of Level Display mode. L.UP= Sets the upper limit value of Level Display mode. PCS= Sets the arbitrary loading piece. #= Corresponds to numeral keys of AKB-301 Application Keyboard. ID= Sets ID. PASSSET= Sets PCAL password. PASS= Inputs PCAL password. (vii) Commands of special functions CAL Enters Span Calibration mode. C18 Enters Span Calibration mode. LOCK Sets menu lock. RELEASE Releases menu lock. TIME Readout of date and time
Page 62
63
ADJCLK Adjusts ± 30 seconds. RSTMN Menu reset MENU= Sets arbitrary menu. { Echo back } Echo back []  Sets to Multi-Connection mode. ( represents a lower  -case alphabet
character.)
Note: For Multi-Connection mode, refer to Section 3.3.5 "Multi-Connection Mode"."
(5) Compatible commands with Mettler® Electronic Balances
S One time output at a stable state SI Immediate one time output SIR Continuous output SR Continuous output at a stable state T Taring after stabilized TI Immediate taring Z Zero setting (same as immediate taring)
(6) Compatible command with Sartorius® Electronic Balances
<ESC>P One time output <ESC>T Taring
Note: <ESC> indicates escape code (1BH).
Page 63
6
4
3.2.4 Windows® Direct Function
Note: For Windows®95 Version 4.00.950B, see page 69 “Compatibility Notification Regarding
Linking of “Windows® Direct” Function with Windows®95 Version 4.00.950B.” The CBX-K and CBW-K series balance can transfer data directly to a personal computer running Lotus 1-2-3, Excel, or other applications of Windows®*.
3.2.4.1 Preparation (Inst allation)
Install according to the procedure described below.
1) Supply power to the balance and set the condition for communication as follows.
Menu [74] [77] [81] [89] [92]
Menu [97] [98]
2) Keep pressing the
key and complete selection of the balance menu.
3) Disconnect the power to the personal computer and balance.
4) Connect the personal computer to the balance with the RS-232C cable.
5) Turn ON the power to the personal computer only and start Windows®*.
6) Click "Start", choose "Settings" and "Control Panel" then select "Accessibility Options."
7) Confirm that there are no check marks for any items on tabs including "General."
8) Put a check mark at "Support Serialkey device" in the "General" tag. This is the only check mark on all the tabs.
9) Open "Settings."
Select the serial port corresponding to the RS-232C port of your personal computer. (Serial port: any one of COM1 to 4. Mostly, COM1)
11) Select a "Baud rate" of 300.
12) Click "OK."
13) Click "Apply."
14) Click "Start", point to "Shut Down" then click "Restart the computer?".
15) After Windows®* has completely started, turn on the power supply to the balance.
It is not necessary to perform the Windows®* control panel setting operation every time.
Page 64
65
3.2.4.2 Checking Operations
(1) Open the "Note pad" attached to Windows®* (or the application in use).
(2) Press the key on the balance.
(3)Verify that the numeric value displayed on the balance is displayed on the screen of personal
computer.
(4)End the operation using the standard ending procedure.
Windows®* = Windows 95, Windows 98, Windows Me, Windows 2000
3.2.4.3 Activate Windows
To ensure correct operation, activate Windows®* using the following procedure.
Procedure
(1) Disconnect the balance power supply. (2) Activate Windows®* and wait until activation is complete. (3) Plug the AC adapter into the outlet and turn ON the balance.
Note: Turning ON the balance before Windows®* is completely activated may cause incorrect
operation.
Note: This function may not operate on a PC on which a normal U.S. version of Microsoft
Windows®* does not operate. Some types of personal computers may not be able to use this function or some features may be limited. CAS does not guarantee that this function can be used on all the PCs without any problems currently or in the future.
• CAS is not liable for any direct or indirect problems caused by this function. It is recommended that important data or programs on your PC be backed-up before using this function. For the operation of Windows®* or the PC, refer to commercial tutorials or the appropriate instruction manual.
• It is necessary to have the "Accessibility Options" function of Windows®* installed on the PC. To install "Accessibility Options", select "Start" -> "Setting" -> "Control panel" -> "Add /Remove Programs" and open the "Windows Setup" tag. Place a check mark on "Accessibility Options." For more information, see the Windows®* instruction manual.
• Once the serial key device is made effective, software which uses the RS-232C port on that PC does not operate correctly, unless it is made ineffective again. If an external modem, plotter or etc. is connected, remove the check mark placed on "Using the serial key device" and re-activate the OS after connection to the balance is terminated.
Page 65
66
3.2.4.4 T roubleshooting
When the Function Does Not Operate At All:
• For some notebook PCs, it is possible to exclude the RS-232C port for energy saving purposes. Set the PC so that the RS-232C port can be used.
• Change the settings of COM 1 to 4. Re-start Windows®* after the changes are complete.
• Verify that the correct RS-232C cable is being used.
Windows®* = Windows 95, Windows 98, Windows Me, Windows 2000
When the Function Intermittently Malfunctions:
• Use a communication speed of 300bps. Depending on the processing ability of the PC, this function may operate incorrectly if communication speed is too high.
• Send the next data only after the current one is displayed on the screen. Depending on the processing ability of the PC, this function may operate incorrectly if interval of data transmission is too short.
• Do not touch the keyboard or the mouse while the balance is transmitting data.
Stop the data transmission and confirm that no data is entering the PC before touching the keyboard
or the mouse.
• This operation may be incorrect when the displayed value is not a weight value (i.e. error code or time).
• The unit designation is not transmitted.
• This function may operate incorrectly depending on the settings of various lock keys of the keyboard such as the NUMLOCK or cursor key lock. Change the state of the lock and function keys on the PC keyboard.
• The peripheral devices connected to the DATA I/O such as Electric Printer EP-50 cannot be used with this function.
• When this function is used, a command cannot be sent from the peripheral device to the balance.
• Set the data formats, such as decimal places and units, in each application.
Page 66
6
7
3.2.5 Multi-Connection Mode
A maximum of 26 CBX-K/CBW-K series balances can be connected to one personal computer at the same time. This is called "Multi-Connection mode." To use the balance in this mode, prepare RS-232C cables in the number of balances connected, and the optional IFB-102A RS-232C Interface.
Connecting Method
Example for connecting 10 units of CBW/CBX series balances to one personal computer is shown in the diagram below.
Assignment of Identification Name In this example (10 balances connected to one computer), assign the identification name in lower-case alphabets. Since 10th character of alphabet is "j," assign "j" to the balance closest to the personal computer, "i" to next closest balance, and so on, back to "a." Setting to Multi-Connection Mode
1) Adjust the communication menu settings of all the balances to the specification of the personal computer, and then, turn off the power to all balances.
2) Start supplying the power to the whole system. Wait until all balances display OFF.
3) Send "[α] <delimiter>" from the personal computer.
α is the name of the balance next to the personal computer. In the example above, lower-case "j."
This command is valid only one time after turning the power ON. Respective balances automatically enter the Multi-Connection mode and the mass is displayed. On this process, the name smaller than the sent command by one, i.e. "[i] <delimiter>", is returned.
4) This procedure completes the setting to the Multi-Connection mode.
Command Codes in the Multi-Connection Mode
Only the commands shown below are valid in the Multi-Connection mode. (α is the name of the balances.)
[α]TARE Same as the operation of pressing the
key of the balance "α."
[α]T Same as the operation of pressing the
key of the balance "α."
[α]POWER Same as the operation of pressing the key of the balance "α."
Personal computer IFB IFB
CBW or CBX [j] CBW or CBX [i] CBW or CBX [a]
Page 67
6
8
[α]Q Same as the operation of pressing the key of the balance "α."
[α] PRINT Same as the operation of pressing the
key of the balance "α."
[α]D05 Same as the operation of pressing the
key of the balance "α."
[α]D07 Same as sending the D07 command to the balance "α."
[α]UNIT Same as the operation of pressing the
key of the balance "α."
[α]RANGE Same as the operation of pressing the
key of the balance "α."(Valid
only for D-type)
[α]CAL Same operation as sending the CAL command to the balance "α."
[α]UNIT+ Same as the state of holding down the
key of the balance "α."
Format of Output Data from the Balance
The output data from the balance "α" is performed with the following format. "[α]" data <delimiter> Format of the load data is the same as the normal output form (
). Operation is not guaranteed for the data containing characters, such as printing of date, time, and calibration document etc.
Restricted Items in the Multi-Conne ction Mode
Multi-Connection mode is not designed for each balance to independently send the data. This mode is for sampling the data by control of multiple balances with one PC. This is not the function to support multiple balances for sending the data separately. When multiple number of data and commands exist on the system at the same time, the balance may not operate normally. For example, multiple data may exist simultaneously on the system when using multiple balances in the Application Measurement mode such as Auto Print or Continuous Output, or
pressing the
key. Normal operation cannot be guaranteed.
Peripheral instruments connected to the DATA I/O or IFB-102A connector, such as an EP-50 Electronic Printer, cannot be used. The communication formats can only be used with
. Delimiters cannot be used with
or . It will take an average 0.05 x N seconds for the data from the personal computer to reach balance unit N, even at setting of 38400 bps. Also, it will take almost the same time for the data from the balance in the Nth position to reach the personal computer.
Cancellation of Multi-Connection mode
Multi-Connection mode cannot be canceled unless the power for all balances is turned OFF.
Page 68
69
3.3 Optional AKB-301 Application Keyboard
The optional AKB-301 Application Keyboard is provides numeric keys which are very convenient for inputting numbers. It is possible to perform various numeric settings such as, piece count and threshold value of graphic display (comparator), as if you were operating a pocket calculator. Use of the AKB-301 Application Keyboard is recommended for performing piece count and pass or failure judgment in daily operation.
Page 69
70
3.3.1 Basic Information
3.3.1.1 Connection
1) Disconnect the power to the balance.
2) Connect the AKB-301 cable to the KEY connector on the back of the balance.
3) Turn on the power to the balance.
3.3.1.2 Overview of Use
1) To set numeric values Press a function keys after the number keys.
2) To check the set numeric values
Press only a function key, not a number key.
3) To return to the mass display
Press the C key.
If numeric keys other than C are pressed, the
symbol illuminates to indicate that a number is being
entered.
3.3.1.3 Summary of Function Keys
Function
Function key
When this key is pressed after the numeric keys
If this key is pressed while the numeric keys are not pressed *1
UPPER key *2
Sets
value of graphic display. Displays value of graphic
display.
LOWER key *3
Sets
value of graphic display
or sets
value.
Displays
value of graphic
display or displays
value.
TARG ET k ey *4
For target display, sets value.
For target display, displays
value.
g/PCS key Recalculates the weight with regard to the
present load as the input value (pieces) and saves.
Switches gram-display unit and displays in piece unit.
UNIT WEIGHT key
Stores the numeric values entered as the unit weight.
Displays the unit weight presently stored.
RECALC key Nothing is done. When piece is displayed, divides the
present load by the present displayed
piece count to recalculate the unit weight. PRINT key *5 Outputs the input numerals. Outputs the displayed value. TARE key *5 Sets Pretare value. Performs taring.
*1 When these values are displayed, the * and # symbols are illuminated on the screen. Pressing the C key returns these displays to the mass display. These displays can be output (printed) with the PRINT key.
*2 This key functions only when a comparator is used in Group or Level mode.
is displayed
for other modes.
*3 When a comparator is used in Group or Level mode, is the target value.
Page 70
71
is the target value in the Target mode. is displayed for other modes.
*4 This operation is performed only in Target mode.
is displayed for other modes.
*5 Operation of the TARE and PRINT keys are the same as keys on the main balance body.
3.3.2 Example of Use
3.3.2.1 Using with a Comparator
Note: The decimal key of the AKB is not used for setting comparator values. The decimal place is set
automatically according to the display capabilities of the balance.
Preparation
Set the kind of graphic display you desire to use with menu selection.
1) From the mass display, press the key three times and the key once.
2) Use the
and keys to set the desired graphic display mode.
(1) Target display mode
Setting the target value
Example: Press the numeric keys for 10000 and then the [TARGET]key.
The Target value is set to 10000.
Setting the limit value Example: Press the numeric keys for 5000 and then the [LOWER] key.
The Limit value is set to 5000.
Checking the target value Press the [TARGET] key to display the set value.
Checking the limit value Press the [LOWER] key to display the set value.
After checking, press the g/PCS key to return to the mass display.
Note:
is displayed if the [TARGET] key is pressed without being in the Target display
mode.
(2) In Comparator mode (group display and level display)
Setting the upper limit value
Example: Press the numeric keys for 20000 and then the [UPPER] key.
The upper limit value is set to 20000.
Setting the lower limit value
Example: Press the numeric keys for 10000 and then the [LOWER] key.
The lower limit value is set to 10000.
Checking the upper limit value Press the [UPPER] key to display the set value.
Page 71
72
Checking the lower limit value Press the [LOWER] key to display the set value.
After checking, press the g/PCS key to return to the mass display.
3.3.2.2 Piece Counting
Use the [g/PCS], [UNITWT], and [RECALC] keys.
Counting Pieces
Use the following procedure to count pieces based on a sample with an arbitrary number of pieces. For example, to recalculate the unit weight based on 25 pieces and perform piece counting.
Example Procedur e
1) Load the container and press [TARE] key or zero the balance.
2) Count 25 pieces and load them on the pan.
Note: If in gram-display, the gram value of the 25 piece sample is displayed. If in PCS display, the
value is converted into piece display using the previously stored unit weight. The display may not be 25 pieces.
3) Press the numeric keys for 25 and then the [g/PCS] key.
4) Piece count is now displayed in the new unit weight.
Recalculating the Unit W eight
To recalculate the unit weight with the currently displayed piece count, press the [RECALC] key when the correct piece count is displayed.
Inputting Unit W eight
To input a unit weight directly, (for example, 1.234). Press 1.234 and then press the [UNIT WEIGHT] key. Set the unit weight in "g". A value using the maximum number of digits displayed by the balance can be set. Use the decimal point key on the AKB when entering this value. Recalling the Unit Weight Press the [UNIT WEIGHT] key. Switching from g to PCS Press the [g/PCS] key to switch between the "g" unit and "piece count" unit.
3.3.2.3 Other Uses
Setting the Pretar e Value Example: Enter 1.234 and press the [TARE] key.
1.234 g is set as the Pretare value.
Note: Number of decimal places is set as follows:
Page 72
73
H-type and S-type ... with number of decimal places at 1d gram-display unit
If not correct, is displayed.
Example: To cancel the Pretare value for a balance with three decimal places:
X wrong example: 0 [TARE] Ο correct example: 0.00 [TARE]
Checking the Pretar e value
Press and hold the [TARE] key for approximately 3 seconds. Press the [TARE] key again to return to the mass display.
Cancel a Pretar e V a lue
Cancel a pretare value by setting the value to zero using the correct decimal placement for the balance. Output Keyboard Numbers to Peripheral Devices Arbitrary numbers with a maximum of seven digits can be output using the numeric keys and the [PRINT] key. The sample number can be input at the balance and transferred with the balance output to the personal computer. Example: To input 123-456: Enter 123.456 and press the [PRINT] key.
Setting a Numeric Menu V alue
Use the number keys of the AKB to enter numeric values during menu selection. Numbers are set sequentially starting from the blinking digit. The decimal point can be set using the decimal point on the AKB.
Last Menu Recall function with AKB-301
Last Menu Recall function is helpful when changing the numeric values for the
same Menu Map item. To use this function, press and hold the
key until
the menu is displayed.
3.4 Foot Switches
A foot switch is convenient when the operator must work with both hands. Two types of foot switches are available.
FSB-102T has the same operation as the key of the balance
FSB-102P has the same operation as the
key of the balance
Note: Either one of the foot switches can be connected at a time. The foot switches cannot be used
together with the AKB-301 Application Keyboard.
Page 73
7
4
3.5 Hook for Below-Balance Weighing
This hook is useful for conducting measurements with the sample suspended beneath the balance, such as in specific gravity measurement.
Attachment method
Attach the below-balance weighing fitting by screwing the supplied screws into the screw holes for the transportation screws on the bottom of the balance (2 places), as shown in the diagram on the right, until the screws cannot be tightened anymore.
Note:
Be careful to ensure that no undue force is applied to the fitting when putting down the balance while the fitting is attached. Otherwise, the balance may not operate correctly. In particular, do not put down the balance in a location where there is no hole for below-balance weighing.
Page 74
75
4. APPENDIX
4.1 Part List
4.1.1 Standard Accessories and Maintenan ce Parts
Name Part No. Remarks
Pan ASSY,KW
321-41074 For the CBW-K Pan ASSY 321-40919-01 For the CBX-K Pan support cap 321-40965-04 Vinyl cover 321-40924 Double-sided tape for fixing the vinyl
cover
320-02138
Calibration weight stopper screw 321-41046 For the CBW-K Calibration mass cover setscrew (SUS bind M4x8)
020-37512 For the CBW-K
Transportation screw (SUS Sems P3BK M4x16)
020-46551
Caster ASSY 321-30070-03 AC adapter 321-61610 For AC100V
4.1.2 Options
Name Part No. Remarks
EP-50 Electronic Printer 321-34986 Dot matrix type EP-60A Electronic Printer 321-42008-10 Thermal type RS-232C cable 25P-9S(1.5m
321-60117-01 For PC/AT, DOS/V
RS-232C cable 25P-25P(1.5m)
321-60116-01
For some NEC PC-98 models (Standard IEEE size)
RS-232C cable 25P-14P(2m
321-60118-01
For NEC PC-98 notebooks (Centro-half)
IFB-102A RS-232C Interface
321-41167-10 Required for multiple connections
Pocket computer printer CD­PCE650
320-02144-14
Application measurement key AKB­301
321-53382-01
FSB-102PK Foot Switch
321-60110-11
Equivalent to PRINT key on main
body FSB-102TK Foot Switch 321-60110-12 Equivalent to TARE key on main body Battery PC7200B 321-61691-01 Rechargeable type. For AC100V only Hook for below-balance weighing 321-40962
Note:
Part No. and specification are subject to change without notice. It is not guaranteed that RS-232C cable will conform to all computers.
Page 75
76
4.2 T able of Unit Conversion Const ants
SCROLL
ORDER
MENU UNIT
Conversion
coefficient (g=)
Note
(display)
[54] U-kg kg 0.1
[55] U-mg mg 1000
[56] U-% %
[57] U-pcs Pcs
[58] U-ct ct 5
[59] U-mom Mom 0.266667
[60] U- d  DS
[61] U-d DL
[61a] U-Lb Lb 0.00220462
[61b] U-Oz Oz 0.0352740
[61c] U-Ozt Ozt 0.0321507
[61d] U-HK TL-HK 0.0267165
A triangle symbol
displays
[61e] U-HK TL-HK(Jwl) 0.0267173
Three triangle symbols
display
[61f] U-SporE TL-S’pore 0.0264554
[61g] U-tiwAn TL-Taiwan 0.0266667
A triangle symbol
displays
[61h] U-tiwAn TL-Taiwan 0.0266667
Three triangle symbols
display
[61i] U-mal TL-Malaysia 0.0264600
[61j] U-ChinA TL-China 0.0266071
A triangle symbol
displays
[61k] U-ChinA TL-China 0.0266071
Three triangle symbols
display
[61l] U-dwt dwt 0.643015
[61m] U-GN GN 15.4324
[61n] U-t t 0.0857339
62 U-USEr Set by user
Page 76
7
7
4.3 Specifications for the RS-232C/AUX Connec tor
Pin number Name Function Remarks
1 FG Frame ground
2 TXD Data output
3 RXD Data input
4 RTS Internal connection with CTS
5 CTS Internal connection with RTS
6 DSR Handshake (receiving)
7 SG Signal grounding
8 NC Blank
9 NC Blank
10 NC Blank
11 NC Blank
12 NC Blank
13 NC Blank
14 NC Blank
15 NC Blank
16 NC Blank
17 NC Blank
18 NC Blank
19 NC Blank
20 DTR
Handshake (receiving)
21 NC Blank
22 NC Blank
23 NC Blank
24 NC Blank
25 NC Blank
Note: NC= No connection within the balance.
Page 77
7
8
4.4 Error Display List
4.4.1 General Display
Display Description of message
Wait for next display.
Minimum display resolution is decreased by one decimal place.
Minimum display digit is returned to original state.
Date and time are being output.
Operation was aborted.
Application Measurement was released.
Calibration check detects too large error.
Calibration check detects too large error.
Menu lock is applied.
Menu lock is released.
Menu was reset.
Contents of new setting and coefficient were stored.
Reset by power failure.
Built-in weight is moving. Wait.
All numerals blinking Place the displayed calibration mass.
Page 78
79
4.4.2 Error Display
Error display Description Countermeasure
The load on the pan is unstable at calibration.
Only one of the calibration levers is working.
Avoid wind and vibration.
The drift of zero point is large at calibration. Empty the pan.
The drift is large at the time of PCAL. Use correct weight.
The drift is large at span calibration. Use correct weight.
Weight is wrong. Use correct weight.
Stops when this is displayed. *
Received command code is not correct. Check delimiter etc.
Integer of the displayed unit exceeded 7 digits. Decrease the load.
Failure in the balance. *
PCAL password error.
“___“ is displayed after the error.
Check the password.
Press the
key and
return the balance to mass display mode.
Attempted to set improper numeric value. Review the numeric value
and decimal places.
Necessary conditions and numeric values are not set.
Check graphic display mode.
key was pressed during menu lock.
Release menu lock.
Mode for operation is different. Check graphic display mode.
Cannot be stored, as the supply voltage is abnormal.
Check the supply voltage.
*Contact your CAS representative.
Page 79
80
MEMO
Page 80
81
MEMO
Page 81
82
MEMO
Page 82
83
Page 83
8
4
Loading...