CAS BioMAT 1 User Manual

Page 1
BioMAT® 1
Safety Cabinet
Page 2
Section Description
Section 1
Your Manual
1.1
1.2
1.3
1.4
1.5
1.6
1.7
Section 2
Using your Cabinet
Section 3
Cabinet Control Panel
Section 4
Filter Replacement
2.1
2.2
2.3
2.4
2.5
2.6
2.7
2.8
3.1
3.2
3.3
3.4
3.5
4.1
4.2
4.3
Switches & Indicators Start Up Procedure Shut Down Procedure Cleaning Procedure Fumigation & Formalin Quantities Fumigation Procedure Airflow Measurements Ultraviolet Radiation
Display Alarm Circuits Electrical Protection Fan Speed Control Engineers Menu
Re-circ HEPA Filters Exhaust HEPA Filter General Notes
Section 6
Service & Spares
6.1
6.2
6.3
6.4
Service Schedule Spares List Drawings (attached) Wiring Diagrams (attached)
Page 3
General Description – 1.1
Your BioMAT 1 Microbiological Safety Cabinet has been designed to provide optimum performance for operator protection. Using the latest in microprocessor and fan technology, it is designed to exceed the performance requirements of the European Microbiological Safety Cabinet Standard EN12469:2000. The safety cabinet carcass and all seals are maintained under negative pressure ensuring air cannot leak out during normal operation.
Class 1 Safety Cabinets offer a high level of operator protection; they ensure any aerosols generated within the cabinet are filtered via high efficiency filtration (HEPA) prior to being dispersed back into the laboratory or to atmosphere. Operator protection is provided by a constant inflow of air through the working aperture of the front screen.
The Class 1 Microbiological Safety Cabinet can be supplied as standard in either of the following two modes:
Ø Re-circulating Type – Exhaust air from the safety cabinet is passed through two
high efficiency filters (HEPA) before being released back into the laboratory. The exhaust air is made up of 100% of the total air volume handled by the safety cabinet.
Ø Exhaust Type – Exhaust air from the safety cabinet is passed through a high
efficiency filter (HEPA) before being extracted to atmosphere through a fan assisted extract system. The exhaust air is made up of 100% of the total air volume handled by the safety cabinet.
Page 4
Quality Assurance – 1.2
Although fully tested before leaving our factory as part of the ISO 9001:2008 Quality Assurance Programme, the specified performance will only be maintained if your cabinet is sited correctly and regularly serviced. CAS can only accept responsibility for correct functioning of your cabinet if: -
Ø
Ø
Ø
used.
Ø
The Safety Cabinet is correctly sited in the laboratory to avoid any adverse conditions within the room that may affect the level of operator protection.
It has been installed and commissioned by CAS trained personnel or approved CAS agents.
Ø
Extension, modification, relocation, repairs or other maintenance is carried out by CAS personnel or persons authorised by CAS or, in the case of electrical work, by qualified electricians.
In the case of repair or maintenance, replacement parts supplied by CAS must be
The electrical installation surrounding the unit and to which it is connected comply with the latest IEC regulations.
Ø
The unit is used and maintained in compliance with the instructions contained in this manual.
CE Declaration of Conformity
CAS declares that the equipment supplied conforms to the following CE directives
Machinery 2006/42/EC
Electro Magnetic Compatibility 2004/108/EC
Low Voltage 2006/95/EC
Page 5
Cabinet Siting – 1.3
The siting of your Safety Cabinet is extremely important. Air currents and the movement of people in the laboratory can adversely affect the performance.
Safety Cabinets should be sited away from;
Ø Doors and windows which open
Ø Draughts caused by ventilation and air conditioning units
Ø Pedestrian traffic routes
Ø Other safety cabinets or fume cupboards
Ø Adjacent fridge & Incubator doors
These points are of particular relevance to class 2 safety cabinets. The diagram shown below indicates some suggested locations for the correct siting of safety cabinets and highlights some situations which should be avoided.
3 4
SUPPLY GRILLE
1
2
HORIZONTAL
LAMINAR FLOW
Position 1 - An acceptable site not affected by disruptive air currents
Position 2 - Well sited
Position 3 - Poorly sited if windows open – If not it should be a safe distance
from the cabinet opposite.
Position 4 - Poorly sited – can be affected by air currents from the opening door,
through traffic and the horizontal laminar flow workstation sited directly opposite.
Page 6
Installation – 1.4
Safety Cabinets are sophisticated items of equipment containing delicate filters which require expertise in their safe handling and installation into laboratories.
For exhaust type safety cabinets the exhaust ductwork route should ideally be surveyed and ductwork installation be carried out by qualified engineers as it forms an integral part of the system relating to the overall performance of the cabinet and is required to conform to various safety standards.
A poorly installed cabinet may compromise the protection provided to both personnel and work being handled and may present a hazard to other occupants of the building and the public.
· Make-up Air
It is important that any make-up air compensating for the air exhausted from the safety cabinet does not cause draughts to the discomfort of the laboratory staff or detriment of the cabinet performance.
Air supply diffusers should be positioned more than 1500mm away from the front of the safety cabinet and have a maximum velocity of no more than 0.30m/sec.
· Commissioning
When any safety cabinet is installed, it is necessary to carry out a number of commissioning checks in order to ensure it is fully operational and that the performance on site satisfies the current standard BS EN 12469:2000. This includes measuring the airflows, testing the HEPA filters with a suitable challenge aerosol and a KI Discus Test (operator protection test) to assess the containment of the cabinet.
CAS employs a team of fully trained installation and commission engineers to carry out all work necessary. This ensures that all new safety cabinets operate to the desired performance.
· Site Surveys
If you have any queries regarding the siting of your safety cabinets we will be only too pleased to arrange a site survey by one of our regionally based technical support staff.
· Periodic Maintenance & Servicing
To maintain safety cabinets at their optimum level of performance and to ensure lifetime operation, regular servicing is necessary. CAS provides a full servicing and maintenance scheme tailored to suit your individual needs. For more information on this please contact our service department on 0161-655-8860.
Page 7
Avoiding Disturbances – 1.5
The sketches below show recommendations for avoiding disturbances for both cabinet operator and safety cabinet performance.
300
1000
1500
Keep pedestrians away from the front
of your safety cabinet
Bench
1500
Position clear of bench opposite Keep clear of structural columns Keep well away from opposite wall
Keep clear of adjacent wall
C
300
Position well away from the door openings
2000
Page 8
Technical Data – 1.6
Cabinet Size 900mm 1200mm 1500mm 1800mm
Dimensions
External Dimensions (w/d/h) 900/630/1250 1200/630/1250 1500/630/1350 1800/630/1350 Internal Dimensions (w/d/h) 895/625/700 1195/625/700 1495/625/700 1795/625/700 Opening to Work Area (w/h) 710/200 1010/200 1310/200 1610/200
Weight Typical Kg 150 200 250 300
Loading Capacity
Work Surface Kg 50 50 50 50
Air Volumes Typical
Exhaust Air Volume m³/sec 0.120 0.150 0.200 0.240 Re-circ Air Volume m³/sec 0.120 0.150 0.200 0.240
Pressure Drop Typical
Exhaust Clean Filters Pa 350 350 350 350 Exhaust Dirty Filters Pa 500 500 500 500
HEPA Filter Data
HEPA Type H14 (EN1822) H14 (EN1822) H14 (EN1822) H14 (EN1822) Efficiency @ 0.3µ 99.999% 99.999% 99.999% 99.999%
Heat Gains Typical
Exhaust Watts 150 150 220 250 Re-circ Watts 150 150 220 250
Noise Typical
Noise Level dB (A) <58 <58 <58 <58
Light Typical
Lighting Level White Lux 900 900 900 900 Lighting Level St/St Lux 800 800 800 800
Duct Connection (Exh) Mm 160 160 200 200
Electrical Data Typical
Voltage Volts
AC Frequency Hz 50 50 50 50 Phase N/A Single Single Single Single Power Consumption kW 0.18 0.18 0.25 0.32 Current A 3 3 5 7 Internal Socket(s) 230v /13 Amp 230v /13 Amp 230v /13 Amp 230v /13 Amp
230 230 230 230
Page 9
Your Manual – 1.7
This user manual has been prepared to provide a basic operating and maintenance instruction. It is intended to supplement existing in-house procedures and codes of practice, and not to replace them. If further advice is required on the use or maintenance of this equipment, the staff of Contained Air Solutions Ltd. will be pleased to assist wherever possible.
Switches & Indicators – 2.1
Page 10
The diagram below shows a typical control panel membrane layout for the BioMAT 1 Safety Cabinet.
BioMAT 1
CLASS 1 MICROBIOLOGICAL SAFETY CABINET COMPLIES WITH BS EN 12469:2000
0
1
AIRFLOW SAFE
BioMAT 1
DESCRIPTION LEGEND COLOUR FUNCTION
A – Cabinet Operation Green
B – Lights Yellow Controlling power to lights.
C – Gas Valve (Optional) Black
D – Alarm Mute
Red
Starts / Stops power supply to fans and control circuits.
To activate solenoid Gas Valve. Will only function when cabinet is in safe working condition.
To mute audible alarm. Fault indication on display will remain until fault is rectified.
E – Fumigate Orange
F – UV Lights (Optional)
G – Key Switch
0
1
Blue
Grey Border
To activate fumigation cycle. Will only operate when cabinet is switched OFF.
To activate power to UV Lights. Will only operate when cabinet and lights are switched OFF.
Key switch for supervisor control, used to prevent cabinet operation during fumigation.
Page 11
Start Up Procedure – 2.2
The following notes are for guidance where local laboratory instructions do not exist or are inappropriate. They should complement, not replace, existing codes of practice issued by your Laboratory Safety Officers.
Ø
being illuminated. The key switch on the front of the cabinet must be turned to position 1= (ON).
Ø
main display will show `AIRFLOW STABILISING` for 60 seconds, once airflows are settled the display will show `AIRFLOW SAFE`. On start up the audible alarm will sound until airflows are safe; this may be muted using the Alarm Mute switch `D.
Ø
cabinet is OFF; this can be used if loading or un-loading the cabinet with equipment prior to use.
Ensure power supply to the cabinet is switched on, as evidenced by the main display
Press the green switch Aon the control panel – this will energise the exhaust fan. The
NOTE: Should the airflows change significantly and fall out of the specifications laid down in EN12469:2000 the cabinet alarms will be automatically activated.
Switch on the interior lighting using switch B. Lighting can be activated when the
Page 12
Shut Down Procedure – 2.3
Ø
accordance with laboratory codes of practice. The cabinet should be left running for a few minutes to clear any residual aerosols.
Ø Switch off the interior light using switch `B`
Ø Switch off the cabinet using switch `A`.
Ø Place the closure panel / Night door over the working aperture.
The work area should be cleared of any apparatus / equipment and cleaned in
Page 13
Cleaning Procedure – 2.4
Regular cleaning is very important to prevent the build-up of dirt and hence potentially infectious material. Routine swabbing of work surfaces with 70% v/v IMS (ethanol) or IPA (Isopropyl Alcohol) is recommended.
For cleaning the work surfaces, swabbing with a mild detergent in warm water is very effective. Phenolic or Cresolic disinfectants should be avoided as they may stain the white surfaces with a brownish colour. If they are used, any spillage should be quickly rinsed with clean water and mopped up with an absorbent tissue. Most of the quaternary ammonium compounds and the Glutaraldehyde based surface disinfectants are suitable.
To facilitate cleaning of the work zone and the interior, the whole front screen / visor may be opened and lowered from the top, supported by the hinges at the lower edge. It is good practice to clean the inside of the viewing screen to ensure adequate visibility of the working zone. Always consult the Laboratory Safety Officer before carrying out this procedure.
Warning
If Hypochlorites are used to clean the stainless steel interior of the safety cabinet they will initially cause rust spots and over time may lead to further damage.
Page 14
Fumigation & Formalin Quantities – 2.5
When handling hazardous materials the air space inside the cabinet should be decontaminated regularly and always before servicing and following any spillages. Fumigation by formaldehyde gas is the recommended decontamination procedure for biological hazards although there are alternative methods available including VHP decontamination and Ozone decontamination.
To facilitate the fumigation a sequence has been incorporated in the cabinet controls, this is detailed under fumigation procedure.
A convenient way of generating sufficient formaldehyde is to boil off Formalin (40% formaldehyde BP or equivalent) in a suitable vessel such as a formalin vaporiser. These are available from CAS.
Always consult your Laboratory Safety Officer prior to fumigation of a safety cabinet. ** If in doubt ask **
Formalin Quantities
The recommended quantity stated in BS EN12469:2000 on Page 40 Annex J, Part 2 is 60ml formaldehyde solution mixed with 60ml distilled water per cubic meter of cabinet volume. However this quantity is now considered in excess of that required to achieve a satisfactory kill. We have therefore produced the following table based on quantities employed by users of large numbers of Safety Cabinets.
If you still consider that the quantities recommended in the British Standard are to be used, you may find on completion of the sterilisation cycle that high quantities of fluids containing formaldehyde are present in the cabinet.
Cabinet Size 900mm 1200mm 1500mm 1800mm
Formalin 15ml @40% 20ml @ 40% 25ml @ 40% 30ml @ 40%
Distilled Water 15ml 20ml 25ml 25ml
Page 15
Fumigation Procedure – 2.6
Ø Switch off the cabinet fans by pressing the green button A on the control panel.
Ø
fitted to the cabinet discharge and ensure that the manual damper is fully closed. For cabinets connected to a duct system proceed as follows:
§
quantities) and screw on the aluminium cap – finger tight, having checked the gasket in the cap is undamaged. If the vaporiser is free standing, place on the cabinet work tray, plug into cabinet internal socket and switch on. Now fit the closure panel / night door, if using a lightweight closure panel additional sealing tape will be required around the outside face to ensure no leakage during fumigation.
§
PROGRESS. A countdown timer will be shown directly below in minutes e.g. `285 MINS`.
For re-circulating type safety cabinets the fumigation extract kit (optional) should be
Fill the formalin vaporiser with the correct amount of Formalin (see section 2.5 for
Now press the fumigation button `Eand the display will show `FUMIGATION IN
Ensure all laboratory personnel are aware that the fumigation is taking place; appropriate warning notices should be put on all doors entering the laboratory and on the cabinet being fumigated.
Warning
Ø The cabinet should be left for a minimum of 6 hours, preferably overnight.
Note: Once the fumigatebutton has been activated DO NOT SWITCH CABINET ON until the cycle has been completed.
Ø On completion the cabinet will display `FUMIGATION COMPLETE` and `READY TO
VENT`. When ready, press the fumigation button to exit the cycle and then press the green button `A` to purge the cabinet of formaldehyde gas.
Page 16
For re-circulating type safety cabinets press the fumigatebutton to complete cycle and open the manual shut-off damper on the fumigation adaptor kit (Optional) having first fitted the extract tubing and placed the discharge point in a location approved by the Laboratory Safety Officer.
** When fumigating a re-circulating type safety cabinet where venting is unobtainable, a mobile carbon filter unit may be required. **
Ø
Switch on the cabinet and then remove the sealing tape. To avoid formaldehyde being drawn out of the cabinet remove the rubber bung when using a 100% sealing closure panel.
Ø
Within the first few minutes of purging, the majority of the formaldehyde gas within the safety cabinet will be removed. However due to the fact that formaldehyde adheres to the surfaces of the cabinet and within the media of HEPA filters, we recommend that the cabinet be run for between 4 and 6 hours before the cabinet is serviced or work re-commences.
Ø
** Any poly-formaldehyde residue in the vaporiser may be removed by heating with water containing a little mild detergent at neutral pH.
Remove the (optional) fumigation kit from the cabinet discharge before commencing to use the cabinet.
General Note
Other methods of generating formaldehyde and other methods of cabinet decontamination can be employed; prior to using alternative methods your Laboratory Safety Officer should be consulted. Contained Air Solutions will be pleased to advise if they are able, but a detailed knowledge of every technique cannot be guaranteed.
If your cabinet is fitted with ports to allow fumigation using Vaporised Hydrogen Peroxide (VHP) please refer to the VHP equipment manufacturers manual for the correct application of this technique.
Page 17
Airflow Measurements – 2.7
It is recommended that the inflow airflow be checked regularly, i.e. weekly or monthly according to cabinet usage. A rotating vane type anemometer with a head diameter of about 100mm is considered satisfactory. On 1200mm wide cabinets, take 5 readings, namely one in each corner and one in the centre on the work area aperture. Readings should be not less than
0.70m/sec or greater than 1.0m/sec. Readings above 1.0m/sec may cause problems with containment, this may be checked by conducting a KI test. Anemometer positions for the 1200, 1500 & 1800mm size of cabinets are illustrated below. The results should be recorded to monitor cabinet performance over time.
1200mm BioMAT 1
1500 & 1800mm BioMAT 1
Page 18
Ultraviolet Radiation (Optional) – 2.8
Ultraviolet Radiation (UV) lamps may be fitted as an optional extra; these can be fitted as new in our factory or retrofitted at a later date on site. If installed the cabinet will have1 or 2 (Depending on cabinet size) short wavelength Ultraviolet (UV) tubes emitting 254 nano metres fitted to the inside work area of the safety cabinet.
As a safety feature the UV tubes are interlinked with the cabinet lights to prevent them being used when the cabinet is in use.
Over time, the effective life of UV tubes is known to reduce; therefore we would recommend UV tubes are replaced on an annual basis to ensure maximum efficiency.
Applications
Many bacteria are quite resistant to UV Radiation, and may require prolonged exposure for sterilisation. Dry and/or protein covered organisms may be protected against UV and may be only slightly affected if at all. However, moist, vegetative cells without too much protein covering are killed with reasonable effectiveness after 3-4 hours exposure.
Warning
Extreme care must be exercised when using UV radiation. Consult your Laboratory Safety Officer prior to use.
UV Radiation can cause burns to unprotected skin and it is very important not to look directly at the illuminated tubes with the naked eye.
Page 19
Display - 3.1
The cabinet display is located in the centre of the control membrane; it incorporates backlighting to ensure all signs can be clearly seen from the operating position.
Typical display during normal operation shown below:
AIRFLOW SAFE
BioMAT 1
In the event of an alarm condition the display will clearly indicate the fault; this will remain displayed until fault is rectified.
Typical display during alarm condition shown below:
LOW AIRFLOW
Should the cabinet produce an alarm condition it may be necessary to arrange for a Service Engineer to attend site, in such cases please contact our Service Department on 0161-655-
6183.
Page 20
Alarm Circuits – 3.2
There are three standard alarm circuits on the BioMAT 1 Re-circulating and Exhaust type Safety Cabinets.
1. INFLOW LOW ALARM
Sensed by an accurate pressure sensor mounted on the main printed circuit board and connected via tubing to the sensing points within the cabinet body, this will sense low airflow caused by fan failure or filter soiling, and transmit the signal to the display meter on the front panel and alarm circuit. In the event the alarm is triggered the display will show `LOW AIRFLOW`, and the audible alarm will sound, this can be muted using the alarm mute button.
2. INFLOW HIGH ALARM
Sensed by an accurate pressure sensor mounted on the main printed circuit board and connected via tubing to the sensing points within the cabinet body, this will sense high airflow and transmit the signal to the display meter on the front panel and alarm circuit. In the event the alarm is triggered the display will show `HIGH AIRFLOW`, and the audible alarm will sound, this can be muted using the alarm mute button.
3. FRONT SCREEN
Front screen position is monitored by a micro-switch fitted to the front screen assembly. When opened the display will show `VISOR OPENand an audible alarm will sound. This can be muted using the alarm mute button.
** The alarm circuit is activated each time the cabinet is switched on.
It is important to ensure that the pressure tubing is not damaged or split where it is connected to the pressure sensor. The pressure sensor, in the event of inadequate airflow in either part of the system, will cause the alarm circuit to be activated with a flashing red indicator on the alarm mute button, an intermittent audible alarm tone, and the alarm system will clearly show on the display the fault. The audible alarm tone can be silenced by depressing the alarm mutebutton on the control panel after which the red light will remain continuously lit until the correct airflow is reinstated or the cabinet is switched off.
Page 21
Electrical Protection – 3.3
Ensure the Safety Cabinet is isolated from the mains supply prior to opening access panels.
Warning
Fuses
There are 11 fuses mounted on the printed circuit board to protect the electrical circuits of the cabinet. The cabinet controls are housed behind the front control panel which is secured with two screws located across the bottom edge. Once removed the control panel will hinge upwards to facilitate access to the main printed circuit board.
For full details on fuse ratings and the circuits they protect please see the wiring diagram attached to this manual.
Page 22
Fan Speed Control – 3.4
The inflow fans can be adjusted via the engineers menu, please see section 3.5 for more information.
To maintain optimum performance during routine planned maintenance it may be necessary to increase cabinet fan speeds to overcome filter soiling. This can be adjusted via the front control panel membrane using the Engineers Menu.
It is important any changes to fan speeds must be made by a CAS service engineer or alternative
Warning
competent service provider, failure to do this may result in the warranty being invalid.
Page 23
Engineers Menu – 3.5
It is important any changes within the Engineers Menu be conducted by a CAS service engineer or
Warning
The buttons located on the front control membrane have a second function once access to the Engineers Menu has been gained.
** In order to gain access to the Engineers Menu the cabinet must be switched OFF **
alternative competent service provider; failure to do so may result in the warranty being invalid.
STANDARD SWITCH PANEL ARRANGEMENT
TO ENTER `ENGINEERS MENU` BUTTON 1 SHOULD HELD DOWN FOR >5 SECONDS TO EXIT `ENGINEERS MENU` BUTTON 1 SHOULD HELD DOWN FOR >5 SECONDS USE BUTTON 2 TO START / STOP THE FANS
SOFT BUTTON ARRANGEMENT
Page 24
Engineers Menu Sequence
LOW ALARM
40 Pa
HIGH ALARM
80 Pa
INFLOW FAN
65 %
U.V. LIGHTING 060 MINUTES
TIME ADJUST
14:54
EXHAUST SENSOR
20 Pa
SERVICE TIMER
RESET
INITIALISE VALUE
INITIALISE (0)
YE S
NO
YE S
NO
YE S
NO
YE S
NO
YE S
NO
YE S
NO
YE S
NO
YE S
Page 25
Re-circ HEPA Filter (Air returned to room) – 4.2
Two HEPA filters are located on the top of the main body of the cabinet separated by a spacer section and held in place by four M8 filter bolts.
IMPORTANT
Cabinet must be decontaminated prior to changing any HEPA filters, see section 2.6
Only replace or examine filters if authorised to do so by the Safety Officer or the person in charge of the laboratory. To remove any filters wear disposable gloves, an apron, overalls
and appropriate face covering, especially eye protection.
Ensure the Safety Cabinet is isolated from the mains supply prior to opening access panels.
Warning
To remove filters, release bolts, lift module until filters and spacer section can be removed. Slide one filter forward and out and dispose of as recommended by site regulations, then remove second filter in same manner.
To replace, first ensure that all surfaces have been cleaned and are free from matter.
Remove new filters from their protective packing.
Inspect all exposed surfaces for signs of damage. Any damage should be reported to the supplier immediately.
Slide filters and spacer carefully into position. Tighten filter bolts, ensuring filters and spacer section are correctly positioned and gaskets are compressed to about half their original thickness.
The filters and their gaskets should be tested for leaks by challenging with DOP or equivalent aerosol and scanning the surfaces with the probe of a forward light scattering photometer.
Page 26
Exhaust HEPA Filter (Air ducted to atmosphere) – 4.3
The HEPA filter is located on the top of the main body of the cabinet held in place by four M8 filter bolts.
IMPORTANT
Cabinet must be decontaminated prior to changing any HEPA filters, see section 2.6
Only replace or examine filters if authorised to do so by the Safety Officer or the person in
charge of the laboratory. To remove any filters wear disposable gloves, an apron, overalls
and appropriate face covering, especially eye protection.
Ensure the Safety Cabinet is isolated from the mains supply prior to opening access panels.
Warning
To remove filter, release bolts, lift module until filter can be removed. Slide the filter forward and out and dispose of as recommended by site regulations.
To replace, first ensure that all surfaces have been cleaned and are free from matter.
Remove new filter from its protective packing.
Inspect both exposed surfaces for signs of damage. Any damage should be reported to the supplier immediately.
Slide filter carefully into position. Tighten bolts, ensuring filter is correctly positioned and gaskets are compressed to about half their original thickness.
The filter and its gaskets should be tested for leaks by challenging with DOP or equivalent aerosol and scanning the surface with the probe of a forward light scattering photometer.
Page 27
General Notes – 4.4
Ø Do not store equipment inside the Microbiological Safety Cabinet. The amount of equipment should be kept to a minimum to reduce the disruption to the airflow patterns within the cabinet.
Ø A Bunsen burner should not be used inside the cabinet.
Ø
cabinet is switched on and there is a satisfactory airflow. Any interruption in the power supply or failure of air supply necessitates manual resetting of the valve by depressing the switch on the control panel marked gas valve.
Ø
prevent them being used when the cabinet is in use.
We recommend cabinet users to consult their own laboratory safety information. Other publications available include:
Science Reviews Ltd, Occupational Hygiene Monograph No.9 British Standard BS EN 12469:2000 & 5726-2000 Part 5 Australian Standard AS2567 American National Sanitation Foundation Standard No. 49
We would also recommend the latest guidelines issued by the UK Advisory Committee on Dangerous Pathogens (ACDP). These documents contain much useful information on the performance, installation, testing, use and limitations of Microbiological Safety Cabinets.
The failsafe solenoid operated gas valve (If fitted) will only allow gas to flow when the
As a safety feature the UV tubes (If fitted) are interlinked with the cabinet lights to
Page 28
SERVICE SCHEDULE 6.1
Class II Microbiological Safety Cabinets
Schedule of work included in service agreement
1 Filter Integrity Test (D.O.P) on all main filter(s) and seals
2 Airflow Profiling ensuring compliance to latest British & European Standards
3 Operator Protection Testing (K.I. Discus) 1 x Per Annum
4 Check and adjust alarm parameters & control functions as required
5 Check and adjust electrical and electronic controls as related to the airflow system
6 Check and adjust operation of exhaust system as required
7 Mechanically inspect cabinet components and duct connections (if fitted)
8 Change main filters as necessary (Supply and fitting subject to additional charge)
9 Prepare computer generated Service Report
Above work to be carried out twice each service year
(Operator Protection Test (K.I. Discus) carried out once per year unless instructed otherwise
Service schedules can be tailored to suit individual needs
For a competitive Quotation please contact our service
Department on
0161-655-8860
Page 29
Spares List 6.2
Shown below are ordering codes for the most common parts used on the BioMAT 1 Safety Cabinets. In addition to these items we stock a vast range of Spares, consumables and optional extras. If you cannot see what you require please give us a call on the telephone number shown below.
** Please note cabinet serial number must be quoted when ordering **
UK: Overseas:
Tel: 0161-655-8860 Tel: +44-161-655-8860
Fax: 0161-655-8865 Fax: +44-161-655-8865
ORDERING CODES
Item
Exhaust HEPA Filter x 1-off FIL 161 FIL 021 FIL 144
Recirc HEPA Filters x 2-off FIL 161 FIL 021 FIL 144
Exhaust Fan FAN 050 FAN 057 FAN 057
Main PCB ELC 325 ELC 325 ELC 325
Control Membrane LAB 114 LAB 114 LAB 114
Light Tube ELC 077 ELC 097 ELC 049
UV Light Tube ELC 061 ELC 061 ELC 061
Key Switch ELC 326 ELC 326 ELC 326
Standard Closure Panel CCM 725 CCM 800 CCM 772
100% Sealing Closure Panel CCM 726 CCM 801 CCM 773
Formalin Vaporiser CCP 083 CCP 083 CCP 083
Fumigation Adaptor Box (Recirc) CCM 111 CCM 802 CCM 112
BioMAT 1
1200mm
BioMAT 1
1500mm
BioMAT 1
1800mm
Loading...