CAROLINA BIOLOGICAL 591157, 591158, 591160 Directions For Use Manual

Wolfe Beta Elite Microscopes
Models # 591157, 591158, #591160
DIRECTIONS FOR USE
Be certain the microscope Power Cord is connected to a grounded
electrical service. The use of a surge protector will also help in preventing any possible power surges that could cause damage to the electronics in the microscope.
Rotate the 4X objective into place, perpendicular to the
microscope stage. Provision has been made that the objectives “click” into the correct position.
Using the course-focus adjustment, position the stage so that it is
at the upper limit of its range of travel.
Place a slide on the stage securely, beneath the spring-loaded
clips, or the mechanical stage clip centering the specimen over the hole in the stage.
Adjust the iris diaphragm so that the largest opening coincides
with the opening in the stage. This maximum aperture is generally used at the lowest magnification. With the diaphragm lever you can open or close the Iris to regulate the amount of light coming up through the condenser. In conjunction with the Iris Diaphragm you can regulate the amount of light with the dimmer dial located on the lower right hand side of the base.
While looking through the eyepiece and using the course-focus
adjustment, slowly move the stage/specimen AWAY from the objective until the image is clear.
Rotate the 10X objective into place. While looking through the
eyepiece and using the fine-focus adjustment, clearly focus the image, that now appears enlarged compared to that observed through the 4X scanning objective. You will notice, increasing magnification decreases the field-of-view or relative observable area of the specimen.
At this point, it may be beneficial to diminish the amount of light
entering the specimen. Decreasing the aperture of the iris diaphragm will reduce glare and increase contrast in the image. Use of the iris diaphragm will require experience. It is important to select both the magnification and the degree of contrast that renders the optimum detail in the observed specimen. You may also need to regulate the amount of light with the Dimmer Dial.
Rotate the 40X objective into place and repeat the steps taken for
the initial magnification increase. The objectives of the microscope are parfocal, meaning that the gross image will remain in focus when changing through the objectives. It should only be necessary to use the fine-focus adjustment once the initial image is clearly viewed at low magnification. With the 40X it is necessary to be careful not to touch the slide or any liquid specimen. The 40X is not sealed against liquid and could be damaged. If the specimen will not focus, make sure that you are using the correct thickness of slides.
In order to obtain maximum resolution of the 100x oil immersion
lens, it is necessary to apply immersion oil to the top of the cover glass on the slide. Use the smallest amount possible. A tiny drop goes a long way in use. It is extremely important that the oil from the 100x is not transferred to the other objectives. The oil will damage the non-oil immersion lens.
1. Use a very small amount of Immersion oil. Immersion Oil can
be purchased from Carolina. # 853311 15ml bottle.
2. After viewing with the 100 x lens, clean the oil from the glass
surface and anywhere the oil has come in contact with the objective. It is important to clean all the oil from the slide as well so that the oil does not damage the other objectives.
It will be useful to repeat these procedures until they become
second nature. Your familiarity with them will add to the usefulness of the microscope and to the ease in interpreting the information it provides.
NOTE: If, at any or all magnifications, it is not possible to establish a clear image, verify: that the power supply is on; that all optical surfaces are clean; that the diaphragm is properly set to allow light to pass through the specimen; the objective is in the correct position; and that the stage stop has been correctly set according to the procedure outlined further in this manual.
FEATURES
WIDEFIELD EYEPIECE: The 10X eyepiece, furnished with this microscope, is removable so that it can be cleaned and can be used with or without the rubber eyecups that are provided. Diopter Adjustment: On the Binocular and Trinocular models you will find the round dials located at the base of the eyepiece tubes. These Diopter adjustment
dials are for focusing the user’s eyes for more accuracy. When not in use, leave the dials on the ‘O” position. The Monocular model does not use this correction
feature. MONOCULAR HEAD: Your Wolfe Beta microscope is equipped with an angled head. The inclined angle will promote user comfort when viewing slides for extended periods. The Head can mounted and unmounted by loosening the locking screw found at top right of the Nosepiece. You can also move the position of the head by loosening the screw and moving the head to the desired angle. Always leave the screw tightened when you have made your adjustment. When setting up the scope for viewing, you can view the image with either eye. BINOCULAR HEAD: Your Wolfe Beta microscope is equipped with an angled, Binocular head. The inclined angle will promote user comfort when viewing slides for extended periods. When setting up the scope for viewing, you may need to adjust the distance of the eyepieces for individual use. The distance between the eyepieces will vary from person to person. You can push the eyepieces together or pull them apart to allow the distance to match the user’s eye width distance. Once the distance is aligned, you may also need to adjust the critical focus adjustment on each eyepiece tube. This can be done once you have started viewing a specimen. Normally the diopter
adjustment dials can remain at the “0” setting. It may be necessary to remove eyeglasses in order to get a “combined” image with both eyes.
TRINOCULAR HEAD: Your WolfeBeta Trinocular is much like the Binocular version
with the addition of a top port that comes with a “C” mount camera adapter. This
port is made so that when used, you will need to extend the black tipped rod located on the right side of the prism head. Pushed in the scope will view the image with the eyepieces. When the rod is extended you will be able to bring the image up through the adapter and into the camera source. NOSEPIECE AND SEMI-PLAN OBJECTIVES: The objectives are installed in a revolving, ball bearing nosepiece. By convention, they are arranged in clockwise order. Starting from 4X, clockwise rotation of the nosepiece increases the magnified image. The objectives are corrected to eliminate color distortion in the image that
results from a magnifying lens’s tendency to separate light much like a prism. Because the objectives are the primary “eyes” of your microscope, they should only
Loading...
+ 5 hidden pages