CARLO GAVAZZI MDI 40 TF - CONFIGURATION SOFTWARE, MDM 40 TF - CONFIGURATION SOFTWARE, MDI 40 TF, MDM 40 TF Instruction Manual

Page 1
Page 2
V
MDM
Digital Meter
al1 al2
RPM
F
Page 3
ENGLISH
MDI40.TF Rate meter and controller
OPERATING INSTRUCTIONS
• General Features 2
• Technical Features 3
• Installation 7
• Preliminary Operations 12
• Front Panel Description 13
• Operating mode 15
Important:
We suggest you keep the original packing for a further shipping of the instrument. In order to guarantee a correct use of the instrument, we recommend the user to carefully read the present instruc­tion manual.
Page 4
3
ENGLISH
2
ENGLISH
•operating mode;
•scaling;
•display;
•commands;
•alarm set-points;
•filter;
•analogue output;
•serial output.
TECHNICAL FEATURES
ACCURACY ± 0.01 % F.S., ± 3 dgt (@ 18°C to 23°C); RESPONSE TIME Time base + 200 msec. TEMPERATURE DRIFT ± 100 ppm/°C. DISPLAY 7-segment, red LED, h 14.2 mm. MAXIMUM INDICATIONS 9999. MINIMUM INDICATIONS -1999. INPUTS 2 measuring inputs (channel "A" and channel "B"); 2 ranges:
range r1: 0.001 to 500Hz; minimum duration of ON signal 500µsec; period measurement: 2ms to 1000s; range r2: 0.1 to 50KHz, minimum duration of ON signal 9µsec; period measurement: 20µs to 10s;
INPUT TYPE
DC: NPN (pins 6 to 7): ON<2VDC; OFF - Open collector, leakage current 1mA. DC: PNP (pins 6 to 7): ON>10VDC; OFF - Open collector, leakage current ≤1mA.
CARLO GAVAZZI Instruments Multi-range
µP-based panel rate meter
MDI40.TF
rev. 0
OPERATING INSTRUCTIONS
Important:
We suggest you keep the original packing for a further shipping of the instrument. In order to guarantee a correct use of the instrument, we recommend the user to carefully read the present instruction manual.
GENERAL FEATURES
Features:
•display of measured value;
•two independent control points;
•scaling capability;
•transmission of measured value.
Programming parameters:
•password;
•three levels of password protection;
•range: 500 Hz or 50 KHz;
•time base;
Page 5
5
ENGLISH
4
ENGLISH
Output type: 2 contacts SPST, NO, 5A/250VAC/VDC 40W/1200VA,
130.000 cycles. Min. response time: < 400 msec. filter excluded. Insulation: see table 1.
DIGITAL FILTER
Operating range: 0 to 9999. Filtering coefficient: from 1 to 255.
ANALOGUE OUTPUT (ON REQUEST)
From 0 to 20mADC / from 0 to 10VDC, programmable within the whole analogue output range. Accuracy: ±0.3% F.S. (@ 18°C to 23°C). Re­sponse time: 500ms (filter excluded). Temperature drift: ±200 ppm/°C. Load: 500 (mA output): 10k (V output). Insulation: see table 1.
SERIAL TRANSMISSION RS485 (ON REQUEST)
Multidrop: uni-directional (STD); bi-directional (on request); 2 or 4 wires; max. distance 1200m; termination and/or line biasing directly on the instrument; 255 programmable addresses; data format: 1 start bit, 8 data bits, no parity, 1 stop bit; baud rate: 1200, 2400, 4800 and 9600 bauds selectable by key-pad; communication protocol according to the stand­ard MODBUS, JBUS. Uni-directional communication: dynamic data (reading only): measurement, valley data, peak data, alarm status; Static data (reading): all programming data; Bi-directional communication: dynamic data (reading only): measurement, valley data, peak data, alarm status; Static data (reading/writing): all programming data, reset of peak and valley data, reset of alarm set-points with latch.
DC: NAMUR: ON 1mADC; OFF 2.2 mADC DC: TTL (pins 6 to 4): ON > 4VDC; OFF 2VDC. DC: Free of voltage contact: ON<1K; OFF>20KΩ. AC: PICK-UP voltage up to 100VAC: ON > 2VAC; OFF < 1VAC. AC: PICK-UP voltage up to 500VAC: ON > 9VAC; OFF < 6VAC.
KEYPAD
4 keys for programming and displaying; "S" for menu selection; "UP" and "DOWN" for value programming/function selection; "F" for special func­tions and "esc" (escape).
PASSWORD
From 0 to 255, numeric code of max. 3 digits, 3 protection levels of the programming data.
FUNCTIONS
• Frequency measurement of both channels A and B.
• Other functions: Fa-Fb, Fa/Fb, [(Fa-Fb)/Fb]*100, [Fb/(Fa+Fb)]*100, 1/ Fa, Fa (Fb for rotation sensing).
• Peak and valley.
ALARM
Type: OFF: out of range alarm; UP: up alarm; DO: down alarm; D.DO: down alarm with disabling at power-on; UP .L: up alarm with latch; DO.L: down alarm with latch. Set point adjustment: in the whole range of visualization. Hysteresis: in the whole range of visualization. Set-point limit adjustment: programmable minimum and maximum limits. Activation time delay: from 0 to 255s De-activation time delay: from 0 to 255s Relay status: programmable normally energized / de-energized.
Page 6
7
ENGLISH
6
ENGLISH
PROTECTION DEGREE / WEIGHT
IP 65 (standard), 470 g. approximately (included analogue output and packing)
AC Input RL1 Output RL2 Output Analogue DC Power Serial
Power Supply Output Supply Output AC Power Supply ––– 4kV 4kV 4kV 4kV ––– 4kV Input 4kV ––– 2kV 2kV 500V 2kV 500V RL1 Output 4kV 2kV ––– 2kV 2kV 2kV 2kV RL2 Output 4kV 2kV 2kV ––– 2kV 2kV 2kV Anal. Output 4kV 500V 2kV 2kV ––– 2kV 500V DC Power S. ––– 2kV 2kV 2kV 2kV ––– 2kV Serial Output 4kV 500V 2kV 2kV 500V 2kV –––
Tab.1: insulation table
INSTALLATION
Overall dimensions and panel cut-out
AAAAAA
A
AAAAAA
A
AAAAAA
A
92 mm
45 mm
124 mm
F
S
al1
al2
96 mm
48 mm
91 mm
112 mm
max. 10 mm
EXCITATION OUTPUT (ON REQUEST)
15VDC / 40mA (12VDC / 60mA) (non-stabilized)
POWER SUPPLY INPUT
230VAC -15% +10% 50/60Hz (standard) (200mAT protection fuse*); 115 VAC -15% +10% 50/60Hz (on request) (200mAT protection fuse*); 48 VAC -15% +10% 50/60 Hz (on request) (630mAT protection fuse*); 24 VAC -15% +10% 50/60 Hz (on request) (630mAT protection fuse*); 120VAC-15% +10% 50/60 Hz (on request) (200mAT protection fuse*); 240VAC-15% +10% 50/60 Hz (on request) (200mAT protection fuse*); 9 to 32VDC galvanic insulation (1AT protection fuse*); 40 to 155VDC galvanic insulation (315mAT protection fuse*).
*note: the mains has to be protected by means of a proper fuse type
SELF CONSUMPTION
8VA
OPERATING TEMPERATURE
From 0 to +50°C (R.H. <90% non-condensing).
STORAGE TEMPERATURE
From -10 to +60°C (R.H. <90% non-condensing).
INSULATION REFERENCE VOLTAGE
300VRMS
INSTALLATION CATEGORY CAT III - 300V (EN61010-1). DIELECTRIC STRENGTH 4000 VRMS for 1 minute. EMC IEC 801-2, IEC 801-3, IEC 801-4 (level 3). SAFETY STANDARDS EN 61010-1, IEC 1010-1, VDE 0411. CONNECTIONS screw-type, detachable. HOUSING SIZE / DIMENSIONS / MATERIAL
1/8 DIN / 48 x 96 x 124 mm / ABS, self-extinguishing: UL 94 V-0.
Page 7
9
ENGLISH
8
ENGLISH
Signal from a 2-wire inductive proximity sensor (NAMUR). NOTE: When using a NAMUR input, the terminals 5 and 6 must be jumpered.
Signal from capacitive or inductive switches, type NPN or PNP, 3 wires. NOTE: when using a NPN or PNP in­put, the terminals 6 and 7 must be jumpered.
Signal from a PNP/NPN encoder
Mounting
Insert the instrument into the panel and fasten it by fixing the two lateral brackets (1) supplied with the instrument to the appropriate location (2), and subsequently locking them by means of the 2 screws (3) supplied with the instrument.
Connections
According to the requirements of EN61010-1 the power supply input of the instrument to be connected to the mains must be protected against short circuit by means of appropriate fuses (see "power supply input").
General connection label for MDI40.TF1
3
2
1
Page 8
11
ENGLISH
10
ENGLISH
AC signals from pick-up up to 100VAC
AC signals up to 500VAC
Double input signal from 2 capacitive or inductive switches, type NPN or PNP, 3 wires. NOTE: when using a NPN or PNP in­put, the terminals 6 and 7 must be jumpered.
General connection label for MDI40.TF2
Page 9
13
ENGLISH
12
ENGLISH
FS
CARLO GAVAZZ
al1 al2
2
1
4
3
FRONT PANEL DESCRIPTION
1.Key-pad: functions available outside the programming phase.
Keys to be pressed:
Displays set-point 1 corresponding to a password between 0
and 255;
Displays set-point 2 corresponding to a password between 0
and 255;
For longer than 2 seconds: modification of set-point 1 (pass-
word between 128 and 255 only);
For longer than 2 seconds: modification of set-point 2 (pass-
word between 128 and 255 only);
The command input (display hold function or key-pad disabling) must be connected across the terminals 3 and 4 (only free of voltage contact). The voltage signals must be connected as follows:
500V , channel A to the terminals 1 and 5, channel B to the terminals 3 and 5.
100V , channel A to the terminals 2 and 5, channel B to the terminals 4 and 5. Note: when using only input channel A, connect the unused B channel terminals 3 and 4 to the terminal 5 The command input (display hold function or key-pad disabling) must be connected across the terminals 5 and 6 (only free of voltage contact).
PRELIMINARY OPERATIONS
Before supplying the instrument, make sure that the power supply voltage correspond to what is shown on the label. Example:
MDI40.TF1.D.2.A.XX.IX
SER.N. 990600/20078 POWER 230 VAC 50/60 Hz INPUT NPN, PNP N. 2 control set points OUTPUT 0/20 mA - 0/10 VDC
Page 10
15
ENGLISH
14
ENGLISH
modification of the selection or programming will not be saved if the
S
key has not been previously pressed)
2.Display
Alphanumeric indication by means of a 7-segment display:
of the measured value;
of the programming parameters;
of the measuring abnormal conditions.
3.LED
Indication of the alarm set-point status.
4.Engineering unit window
To insert the interchangeable engineering unit in the special
window, proceed as follows: remove the front cover by inserting a suitable screw driver in the special slot on the short sides of the front panel; force gently until the front cover is completely removed. Insert the desired engineering unit by means of a pair of tweezers. Replace the front cover by inserting it first in the lower part and then in the upper part of the locking system.
OPERATING MODE
• Power-on
When you switch the unit on, the display shows for approximately 5 seconds the instruments' software revision, for example: r.0.
• Displaying, control and diagnostics
The instrument shows continuously the value of the input variable as defined in the programming phase.
+ Displays highest measured value (peak function); + Displays lowest measured values (valley function); For longer than 2 seconds: acknowledge of an alarm (if the relays have the latch function)
1. Key-pad:
functions available in the programming phase.
Keys to be pressed:
For longer than 2 seconds: programming phase entry and password confirmation;
Menu selection (from the first to the last); Menu selection (from the last to the first);
Confirmation and entry:
in the configuration menus;
in the secondary menus relating to parameters.
In the selected menu / secondary menu:
increase of displayed value
modification of parameter selection;
In the selected menu/secondary menu:
decrease of displayed value
modification of parameter selection;
In the menus: exit from the programming phase (message "End" on the display) and return to the measuring and control function; In the secondary menus: exit and return to the main menu (
the
F
F F F
S
S
Page 11
17
ENGLISH
16
ENGLISH
"PAS" message for the second time; enter the desired numerical code using the "
" or " " keys, then confirm it by pressing the "S" key:
the display will show the first configuration menu ("InP");
if the Password has already been entered, you can modify it following
the procedure described at No. 2); after the "PAS" message has been shown a second time, enter the new numerical code using the "
"
or "
" keys and confirm it by pressing the "S" key: the display will
show the first configuration menu ("InP").
Data protection levels:
if the Password is "0", the configuration data are not protected by
undesired access;
if the Password is a number between "1 and 127", the configuration data
are entirely protected against undesired access;
if the Password is a number between "128 and 255", the configuration
data are protected against undesired access except for the program­ming of the values ("SEt") of set-point 1 and/or 2.
It is possible to reset the Password by entering the number 3584.
All programming/configuration steps of the instrument are shown
in the flow chart on the last page of this manual. The flow chart gives the operator a better understanding of the programming structure of the instrument, indicating the current function with regard to the others. The flow chart also makes it easier to understand the commands used in the programming phase.
See the chapter "Front panel description" for information regarding the use of the key-pad and the relevant main functions.
The value shown on the display is continuously compared with the value of the two set-points and of the other parameters, thus generating the control function by energizing/de-energizing the output relays.
Programming
This phase is identified by the blinking of the decimal point on the right side of the display.
To enter the programming phase, press the "
S
" key until "PAS" is
shown on the display; then "0" is displayed: the correct numerical code (Password) is to be entered. The following conditions may occur:
1) the operator hasn't entered any Password: press the "S" key again to enter the configuration menus of the instrument;
2) the operator has already entered a Password: select the correct Password by means of the "
" key (to increase the value) or " "
key (to decrease it) until the desired value is displayed. Press the "S" key to confirm the value: if the Password is correct, then the display will
show "PAS" again followed by the relating numerical code; press the "S" key once more in order to display the first configuration menu; if the Password is not correct, the display shows "End" and the instrument
goes back to the measuring and control phase. PROGRAMMING OF A NEW PASSWORD AND AUT OMATIC SELEC­TION OF THE PROTECTION LEVEL OF THE CONFIGURATION DATA. To enter the new Password:
if the Password is "0", press the "S" key when the display shows the
Page 12
19
ENGLISH
18
ENGLISH
F1 : frequency, tachometer, rate mode A or B channel
The instrument measures the frequency either on input A or on input B and executes the following calculation: display=Fa/Pu.1*PS.1 (or Fb/ Pu.2*PS.2). It can be used, for example, to measure the rotation speed of a motor, or to display frequency in Hz and kHz, as tachometer mode: RPS, RPM, RPH, MPS, MPM, MPH and as rate meter: flow (m
3
, cm3, mm3,kg/m3, g/cm3, l/s, l/min, l/h, m3/s, m3/min, m3/h, t/h, kg/s) or speed (mm/s, cm/s, m/s, mm/min, cm/min, m/min, mm/h, cm/h, m/h, km/h, in/ s, ft/s, yd,s, in/min, ft/min, yd/min, in/h, ft/h, yd/h).
F2 : A - B, frequency (speed) difference
The instrument measures the frequency of input A and input B and executes the following calculation: display=(Fa/Pu.1*PS.1)-(Fb/
HOLD Input
Input A
Display
HOLD Input
Input A - B
Display
Glossary of displayed symbol:
PAS :
protection key: programmable
from 0 to 255.
InP : input range selection menu: r1 : from 0.001 to 500 Hz or from 2ms to 1000s; r2 : from 0.1 to 50 KHz or from 20µs to 10s
t.ba :
time base programming menu: values programmable from 0.1 to 999.9; it is the updating time of both measure­ment and display.
oPr :
operating mode selection menu: General information: "A" means always channel A input; "B" means always channel B input; the signal on the channel A is managed always according to this formula: (Fa/ Pu.1)*PS.1; the signal on the channel B is managed always according to this formula: (Fa/Pu.2)*PS.2. Where: Fa or Fb is the frequency signal measured on channel A or B; Pu.1 or Pu.2 is the number of pulses per revolution gener­ated by the sensor connected to channel A or B; PS.1 or PS.2 is the prescaler value of the sensor A or B.
Page 13
21
ENGLISH
20
ENGLISH
F4 : A/B, frequency (speed) ratio mode
The instrument measures the frequency of input A and input B and executes the following calculation: display=(Fa/Pu.1*PS.1)/(Fb/ Pu.2*PS.2). This measuring capability is ideal for monitoring the relative speeds of shafts, conveyor belts and other moving machinery.
F5 : B/(A+B)*100
The instrument measures the frequency of input A and input B and executes the following calculation: display=(Fb/Pu.2*PS.2)/[(Fa/ Pu.1*PS.1)+(Fb/Pu.2*PS.2)]*100. This measuring capability is used in all the applications where it is necessary to measure a mixture flow between two liquids. If the instrument is equipped with an analogue output, this signal can be transmitted to a paper recorder to show the
HOLD Input
Input B/(A+B)
Display
HOLD Input
Input A/B
Display
Pu.2*PS.2). This result represents the difference between the input channels and can be used when the speed difference between two conveyor belts has to be as low as possible without considering the reference speed of the first conveyor (in any case within a well known value that can be controlled by the available alarm set-point) in order to avoid any transportation problem of the produced goods.
F3 : (A - B)/B*100, frequency (speed) error ratio
The instrument measures the frequency of input A and input B and executes the following calculation: display=[(Fa/Pu.1*PS.1)-(Fb/ Pu.2*PS.2)]/(Fb/Pu.2*PS.2)*100. This result represents the relative error between the 2 input channels; this measuring capability is used when the speed difference between two conveyor belts has to be as low as possible in any case within a well known value that can be controlled by the available alarm set-points. If the instrument is equipped with the analogue output, this signal can be used to correct the speed of the second conveyor belt.
HOLD Input
Input (A-B)/B
Display
Page 14
23
ENGLISH
22
ENGLISH
if the time base is, for instance, three times the period being measured (see figure below), the updating of the display is made at the end of the time base (end gate time) as average calculation of the measured periods P=(P1+P2+Pn)/n
F7 : A, frequency (speed) clockwise and counter clockwise rotation sensing mode.
This measuring capability is ideal for monitoring the relative speed of shafts, conveyor belts and other moving machinery taking into account the rotation mode, showing the reverse speed by means of a " - " sign. The rotation sensing is detected by means of a phase difference, measured using the two input channels available as standard in the instrument. If we assume that the channel A is the main channel ( see figure on the next page), the channel B is used to detect if a signal arrives "after" (phase displacement) the signal of the main channel (begin gate time) meaning "clockwise" rotation or "before" meaning
mixture deviation that is connected to the quality result of the mixture itself.
F6 : 1/A, period or average period mode
The instrument measures the frequency of input A or input B and executes the following calculation: display=1/(Fa/Pu.1*PS.1). The time period is directly connected to the time base that has been programmed in the instrument (for low frequencies or long time periods):
if the time base is in the range of the period being measured (see figure below), the updating of the display is as fast as possible and any period changing is updated immediately on the display;
Page 15
25
ENGLISH
24
ENGLISH
SCA: scaling factor programming menu. This menu allows to scale the input signals coming from both A and B meas­uring channels in such a way that they can be shown on the display as a rate, a frequency, a period or a speed. The parameters that allow to scale this input signals according to the needs are:
Pu.1 : number of pulses per revolution
of the sensor (proximity switch) con­nected to channel A . Programmable value: 1<Pu.1<9999.
PS.1 : Prescaler channel A stated in
the following way: A1*10
b1
.
A1 : 1<A1<9999. b1 : -9<b1<9. Pu.2 : number of pulses per revolution
of the sensor (proximity switch) con­nected to channel B . Programmable value: 1<Pu.1<9999.
PS.2 : Prescaler channel B stated in
the following way: A2 *10
b2
A2 : 1< A2 < 9999. b2 : -9 < b2 < 9.
"counter clockwise rotation". Such kinds of signals are normally gener­ated by standard encoders.
Page 16
27
ENGLISH
26
ENGLISH
For instance, if the wheel has 30 cogs (30 references) that can be detected by an inductive proximity switch, at the end of its rotation it generates 30 pulses, so the number that has to be programmed as Pu.1 or Pu.2 parameter is 30. Pay attention: if, instead of a speed or a rate, you have to measure and display a frequency or a period, Pu.1 and Pu.2 have to be "1" (pulse per revolution).
First example: if you have to measure a speed of a motor with the following data: motor speed: 800 RPM; pulses per revolution : 1; requested displayed value: 800 RPM; you have to proceed in the following way: calculate the
maximum frequency that is given by: "[max. speed (RPM) * pulses per revolution]/60", that takes in our case: "[800 * 1]/60 = 13.33Hz"
"InP": to select the proper input range that is connected to the maximum frequency. You have to measure, in our case, 13.33Hz, therefore the range is "r1" (from 0.001Hz to 500Hz). With the "r1" range you can easily calculate the minimum speed the instrument can measure and display: "[min. frequency (Hz) * 60] / pulses per revolution" that takes in our case "[0.001 * 60]/1=0.06 RPM".
"t.bA": to select the proper time base in order to exploit the maximum performance of the instrument. If the speed of the motor can be adjusted from 100 RPM to 800 RPM we can calculate the time period corre­sponding to the minimum speed that is given by: "60/[min. speed (RPM)* pulses per revolution]", that takes in our case "60/[100*1]= 0.6
What is the meaning of "prescaler"? It's a multiplier used to multiply the measured frequency by a programmable value in order to achieve the desired value on the display. To allow the scaling capability of the instrument to be as powerful as possible, this parameter has been divided into two sub parameters: the mantissa "A" and the exponent "b". The prescaler is stated as
A * 10
b1
. The calculation capability is from: 9999 * 10
-9
(0.000009999) max. divisor;
1 * 10
-1
(0,1) min. divisor;
9999 * 10
9
(9999000000000) max. multiplier;
1 * 10
0
(1) min. multiplier.
What is the meaning of "pulses per revolution"? It is the number of pulses that are generated by the input connected sensor for every complete revolution (360°) of the wheel or of the drive shaft being detected.
Page 17
29
ENGLISH
28
ENGLISH
Second example:
if you have to measure the speed difference between two wheels, detected by 2 inductive sensors, that have the following data: Wheel number 1 (connected to the channel "A" of the MDI40): max rotational speed: 700 RPM; pulses per revolution: 4. Wheel number 2 (connected to the channel "B" of the MDI40): max rotational speed: 600 RPM; pulses per revolution: 4. you have to proceed in the following way: calculate the
maximum frequency that is given by: "[max. speed (RPM) * pulses per revolution]/60", that makes in our case, for wheel 1: Fa = (700 * 4)/60 = 46.66Hz for wheel 2: Fb = (600 * 4)/60 = 40.00Hz
"InP": to select the proper input range that is connected to the maximum frequency you have to measure; in our case the value is between 40Hz and 50Hz, therefore the range is "r1" (from 0.001Hz to 500Hz). With the "r1" range you can easily calculate the minimum speed the instrument can measure and display: "[min. frequency (Hz) * 60] / pulses per revolution" that makes in our case "[0.001 * 60]/4=0.015 RPM".
"t.bA": to select the proper time base in order to exploit the maximum performance of the instrument. If the speed of the two wheels can be adjusted from 100 RPM to 700 RPM (for the first wheel) and from 100 to 600 (for the second one), we can calculate the time period corre­sponding to the minimum speed that is given by: "60/[min. speed (RPM)* pulses per revolution]", that makes in our case "60/[100*4]=
sec., therefore the "t.bA" would be "0.6", better if "0.7".
"oPr": to perform the measurement of a "speed" you have to choose
the "F1" function
"SCA"
"Pu.1": the number of "pulses per revolution" is in our case "1",
therefore "Pu.1" becomes 1.
"PS.1": the prescaler formed by the two sub-parameters "A1", man­tissa, and "b1", exponent, has to be set in the following way: the maximum read-out capacity of the instrument is 9999, that means the speed can be shown as 800.0 "RPM". We have to measure "800" but to obtain the maximum speed resolution we have to display 8000
(don't consider the decimal point). The calculation made by the instrument according to function selection "F1" is the following: value to be displayed = (Fa * PS.1) / Pu.1 (1)
- "Fa" according to the frequency of 800 RPM is 13.33Hz;
- "Pu.1" is 1;
- "PS.1" can be calculated from the above mentioned formula (1) that means: (value to be displayed * Pu.1) / Fa, that makes :
(800*1) / 13.33 = 60.01, that means: 60.01 = A1*10
b1
therefore A1=6001 and b1 = -2 (10 b1 → 0.01)
"dSY": to show on the display 800.0 we have to move the decimal point
to "111.1";
"Lo": see the proper explanation on the next pages;
"Hi": see the proper explanation on the next pages;
The PS.2 menu is not available for the functions F1, F6 and F7.
Page 18
31
ENGLISH
30
ENGLISH
that means: PS = (value to be displayed * Pu.1) / (Fa - Fb), that makes: (100 * 4) / 6.66 = 60.06, that means: 60.06=A*10
b
therefore A = 6001 and b = -2 (10 -2 → 0.01)
"Pu.2": the number of "pulses per revolution", of channel "B", is in our
case "4", therefore "Pu.2" becomes 4.
"dSY": to show on the display 100.0 we have to move the decimal point
to "111.1";
"Lo": see the proper explanation on the next pages;
"Hi": see the proper explanation on the next pages:
Parameter table (suggested values)
Rotational Speed
FUNCTION
(
oPr
)
MEASUREMENT
RPS
RPM
RPH
UNIT
(
displa
y)
TIME BASE
(t.bA)
PULSE / REVOL.
(PU.1) (Pu.2)
PRESCALER (PS.1) (PS.2)
Linear Speed (circumferential speed)
F1, F2 F3, F4
1 60 60
1 1 1
mm/s
cm/s
m/s mm/min cm/min
m/min
1 1
1 60 60 60
F1, F2 F3, F4
cm/h
km/h MPH
60 60 60
1 1 1
Hz
kHz
1
0,1
Frequency F1
m/h
60 60
mm/h
1 1 1 1 1 1 1 1
n n
1
60
3600
π * d * 3600
π * d * 3.6
π * d * 2.23
1000 * π * d
100 * π * d
π * d
1000 * π * d * 60
100 * π * d * 60
π * d * 60
1000 * π * d * 3600
100 * π * d * 3600
n
n / 1000
0.15 sec., therefore the "t.bA" would be "0.15", better if "0.2".
"oPr": to perform the measurement of a "speed difference" you have
to choose the "F2" function
"SCA"
"Pu.1": the number of "pulses per revolution" of channel "A", is in our
case "4", therefore "Pu.1" becomes 4.
"PS.1": the prescaler of channel "A", formed by the two sub-param­eters "A1", mantissa, and "b1", exponent. It must be equal to the prescaler of channel "B", "PS.2".
"Pu.2": the number of "pulses per revolution", of channel "B", is in our case "4", therefore "Pu.2" becomes 4.
"PS.2": the prescaler of channel "B", formed by the two sub-param­eters "A2", mantissa, and "b2", exponent. It must be equal to the prescaler of channel "A" and can be calculated in the following way: the maximum read-out capacity of the instrument is 9999, that means the minimum speed difference (given subtracting the max. speed of channel "A" and the one of channel "B") can be shown as
100.0 "RPM". We want to measure "100" but to have the maximum speed resolution we have to display 1000 (don't consider the deci­mal point). The calculation made by the instrument according to function selection "F2" is the following: value to be displayed = [(Fa * PS.1) / Pu.1 - (Fb * PS.2) / Pu.2 ] (1)
- PS.1 = PS.2 = PS that means: A1 = A2 = A and b1 = b2 = b
- Fa according to the frequency of 700 RPM is 46.66Hz;
- Fb according to the frequency of 600 RPM is 40.00Hz; the prescaler can be calculated from the above mentioned formula (1)
Page 19
33
ENGLISH
32
ENGLISH
Cnd : auxiliary command selection menu
r1 : display HOLD. r2 : key-pad disabling.
t.in : activation time of the auxiliary
input, programmable value: 20 < t.in < 255 msec.
SP.1 / SP.2 :
set-point programming
menu
tYP : working type:
oFF : abnormal condition notifying. The
output will be activated either when the measurement is out of the measur­ing range or when it cannot be displayed (blank display or "EEE" / "-" indi­cation on the display). Note: if the above mentioned function will be selected, the Lo.S, Hi.S, SP1, HYS parameters will not be consid­ered.
do : down alarm. uP : up alarm. d.do: down alarm with disabling at
power-on. uP.L: up alarm with latch ( the alarm
Where: d = diameter of the circumference n = number of pulses per revolution
dSY :
display parameter programming
menu
d.P : decimal point position selection. Lo : minimum value of the display
scale, programmable value: -1999 < Lo < 9999
Hi : maximum value of the display
scale, programmable value: -1999 < Hi < 9999 These values are referred only to the output limits of the analogue output and to the "OFF" alarm function. For instance: if Lo=100.0 and hi=800.0 and the alarm type "TYP" is "OFF", the display will blink and activate the rel­evant output when the measured variable goes out of the "100.0 to 800.0" limits. If the analogue output is avail­able, the output signals will start from "100.0" and will arrive to "800.0".
Page 20
35
ENGLISH
34
ENGLISH
Example: “do” alarm, if “SEt”=2200 (value of the ON alarm status) and the hysteresis “HYS”=12, the resulting OFF value (end of alarm status) is: 2212 (resulting from 2200 +12). “uP” alarm, if “SEt”=2200 (value of the ON alarm status) and the hysteresis “HYS”=12, the resulting OFF value (end of alarm status) is: 2188 (resulting from 2200 - 12). NOTE: the hysteresis is to be programmed according to the displayed range. This means that the hysteresys must always be much lower than the displayed range.
of.d : OFF time delay; programmable
value: 0 < OF.d < 255.
on.d : ON time delay; programmable
value: 0 < On.d < 255.
rL Y : coil status of the relay in the
normal condition:
nd : normally de-energized coil. nE : normally energized coil.
Note: the normally closed or open sta­tus of the alarm contact is modified by means of an inside selectable solder jumper.
can be reset only if the "F" key will be held down for at least 2 seconds). do.L : down alarm with latch ( the alarm can be reset only if the "F" key will be held down for at least 2 seconds).
Lo.S: value of the lower Set-point limit,
programmable value: Lo(Hi) < LoS < Hi(Lo).
Hi.S : value of the upper Set-point
limit, programmable value: Lo.S < Hi.S < Hi(Lo)
SEt : value of the Set-point; program-
mable value: Lo.S < SEt < Hi.S.
HYS : hysteresis value of the set-point.
The hysteresis is a numerical value included within the range: 0 < HYS < 9999 and represents the difference between the value of the ON alarm status and the value of the OFF alarm status. The hysteresis modifies the value of the OFF alarm status not only with regards to the set alarm value, but also with regards to the alarm type: the hysteresis value is summed to the set value if the alarm type is "do" and subtracted from the set value if the alarm type is "uP".
Page 21
37
ENGLISH
36
ENGLISH
played value, the control set-points and the analogue output. NOTE: for a correct working of the filter, the relative coefficient must sat­isfy the following relationship: 1 < Fi.C < (Fi.S x 8) < 255.
A.ou :
analogue output programming
menu.
Lo.A : value to be expressed as % of
the output range (0/20mA-0/10V) to be generated in correspondence with the minimum measured value. Value pro­grammable within the range: 0.00 < Lo < 99.99.
Hi.A :
value expressed as % of the output range (0/20mA-0/10V) to be gen­erated in correspondence with the maximum measured value (Hi.E/Hi pa­rameters). V alue to be programmed within the range: 0.00 < Hi.A < 99.99.
Example: minimum measured value = 100, that must correspond to a retransmitted signal of 4mA.
Lo.A (%) =
100 * ?mA that in our example corresponds to:
20
FiL : filter parameter programming menu. This function allows you to stabilize the instruments digital display, in order to obtain steady readings and better con­trol.
Fi.S : filter working range. The coeffi-
cient is given in digits and defines the range in which the filter works, pro­grammable value: 0 < Fi.S < 9999 The programmable numerical value represents the fluctuation range of the value which has been measured and displayed by the instrument. In the first configuration phase this value must be 0 and the right value is to be entered only after the verification of the possi­ble fluctuation. Example: if the measured instantane­ous value varies from 10.00 to 10.06, the "Fi.S" value to be entered is 0.06 (10.06 - 10.00).
Fi.C : filtering coefficient value. Pro-
grammable value: 1 < Fi.C < 255. The higher "Fi.C", the higher the filter­ing of the measured value and the longer the updating time of the dis-
Page 22
39
ENGLISH
38
ENGLISH
S.ou: selection menu of the serial communication output.
Add : programmable address value:
1 < Add < 255
b.dr : programming of the data baud
rate: r1: 1200 bps. r2: 2400 bps. r3: 4800 bps. r4: 9600 bps.
100 * 4mA/20 = 20%, therefore enter 20.00. Example: minimum measured value = 100, that must correspond to a
retransmitted signal of 1V.
Lo.A (%) =
100 * ?V that in our example corresponds to:
10
100 * 1V/10 = 10%, therefore enter 10.00. Example: maximum measured value 800, that must correspond to a
retransmitted signal of 18mA.
Hi.A (%) =
100 * ?mA that in our example corresponds to:
20
100 * 18mA/20 = 90%, therefore enter 90.00. Example: maximum measured value 800, that must correspond to a
retransmitted signal of 5V.
Hi.A (%) = 100 * ?V that in our example corresponds to:
10
100 * 5V/10 = 50%, therefore enter 50.00. Also in this case it is possible to invert the scale, that is, a decreasing
value of the retransmitted signal may correspond to an increasing value of the input variable (example: Lo.A > Hi.A).
Page 23
40
ENGLISH
In the secondary menus: press the key to go
back to the main menu (
the selection or programming
modification will not be saved if the
key has not
been previously pressed)
YES
NO
In the menus: press the key in order to exit from the
programming phase: the display shows "End" and then the
measured value
Loading...