Carestream Health 5850 User Manual

KODAK DRYVIEW 5800 Laser Imager
H221_0017
and
CARESTREAM DRYVIEW 5850 Laser Imager
Safety Manual
Carestream Health, Inc.
150 Verona Street
© Carestream Health, Inc., 2009
DRYVIEW is a trademark of Carestream Health, Inc.
The Kodak trademark and trade dress are used under license from Kodak.
Pub No. 2G0734
Rev. C
Table of Contents
Safety and Related Information - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -1
Safety, Warnings, and Cautions - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -1
Safety Labels - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -5
Safety and Health Compliance - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -10
Safety - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10 EMC - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 11 EU Directives - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 16 CE Marking - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 16 Please Note - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 17
Imager Specifications - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 17
February 27, 2009 2G0734 i
Safety and Related Information
RISK OF ELECTRIC SHOCK:
Safety, Warnings, and Cautions
Please read and understand all instructions before using this product.
This equipment is operated with hazardous voltage which can shock, burn, or cause death.
Remove wall plug before servicing equipment. Never pull on
cord to remove from outlet. Grasp plug and pull to disconnect. Do not attempt to service or repair the laser imager yourself to avoid exposure to dangerous voltage, laser beam, or other danger. Always call an Authorized Service Provider of Carestream Health, Inc. products for any service or repair.
Do not operate equipment with a damaged power cord.
Do not use an extension cord to power this equipment.
Do not operate equipment with any of the safety interlocks
overridden.
Position the power cord so it will not be tripped over or pulled.
Connect this equipment to a grounded wall outlet.
Three power cord sets are provided with this equipment:
power cord with plug for use in North America
power cord with plug for use in China
power cord with plug for use in Europe
February 27, 2009 2G0734 1
All other countries must use an Agency-approved power cord with a plug type suitable for the country of use, or contact an authorized Carestream Health, Inc. dealer.
DANGER:
Do not operate equipment with the covers open.
DANGER: This equipment is not contained in a sealed cabinet.
This equipment contains moving parts that may be accessible to the user. Loose clothing, jewelry or long hair may cause personal injury or damage to the equipment.
Do not use this equipment in locations where it can come in contact with liquids, including body fluids.
Safety and Related Information
CAUTION:
CAUTION:
CAUTION:
CAUTION:
CAUTION:
CAUTION:
1.83 m
Double pole/neutral fusing.
Do not use a cell phone within 2 meters of a laser imager. This proximity includes any imager behind a wall adjacent to your location.
Do not use a microwave oven within 4 meters of a laser imager. Electromagnetic radiation from a microwave oven is only an issue if after the oven door is closed and latched, the seal does not maintain an electromagnetic tight fit between the oven door and oven main housing. Determining if the seal has an electromagnetic tight fit requires special detection equipment.
Do not use in the presence of flammable anesthetics, oxygen, or nitrous oxide. This equipment does not have a gas-sealed electronics enclosure and could ignite any flammable or explosive gases present in its environment.
This equipment uses a DICOM network port, and is intended to connect to other medical devices. It is not intended to be connected directly outside the building. Only an Authorized Service Provider of Carestream Health, Inc., products or customer’s qualified service personnel may perform installation and service maintenance.
This device is NOT to be located within the patient environment. Therefore, the equipment must not be located closer than 1.83 meters from a patient bed or chair.
2 2G0734 February 27, 2009
Safety and Related Information
CAUTION:
CAUTION:
CAUTION:
CAUTION:
CAUTION:
CAUTION:
CAUTION:
This device should not be used in close contact with MRI devices, due to possible very high magnetic fields near an MRI unit. The magnetic field in the area where this equipment is installed must be less than 50 Gauss.
Do not substitute or modify any part of this equipment without prior written approval of Carestream Health, Inc.
Federal law prohibits dispensing without a prescription.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the User Guide and other User Documentation, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
Do not use isopropyl alcohol to clean the exterior surfaces of the laser imager. Isopropyl alcohol can dissolve the exterior paint on the laser imager.
In the U.S., exhausted filters are considered to be non-hazardous waste according to the US Environmental Protection Agency Resource Conservation Recovery Act (RCRA). Municipality owned and licensed solid waste management facilities are an appropriate disposal option. Contact your local or state solid waste authorities to determine if additional disposal requirements apply. In other regions, contact local or regional solid waste authorities for proper disposal guidance.
February 27, 2009 2G0734 3
Safety and Related Information
CAUTION:
LASER WARNING:
Lithium batteries should only be replaced by an Authorized Service Provider of Carestream Health, Inc. products. The laser imager uses a lithium battery to power the clock and calendar circuitry. THERE IS A DANGER OF EXPLOSION IF THE BATTERY IS REPLACED INCORRECTLY. The battery must be replaced only with the same or equivalent type. The U.S. EPA’s RCRA does not regulate disposal of this lithium battery. Users should discard spent batteries in municipal trash unless their community offers a battery collection program. In other regions, contact local or regional solid waste authorities for proper disposal guidance.
The Equipment uses a 50-Milliwatt invisible laser. Laser radiation may be present when the machine operates without the rear cover installed. Covers with this label may only be removed by an Authorized Service Provider of Carestream Health, Inc. products. USE OF CONTROLS OR OTHER ADJUSTMENTS, OR OTHER PERFORMANCE OF PROCEDURES OTHER THAN THOSE SPECIFIED HEREIN, MAY RESULT IN EYE DAMAGE.
4 2G0734 February 27, 2009