Cardinal Detecto 6127 Series, Detecto 6337, Detecto 6129 Series, Detecto 6339 Operating Instructions Manual

Page 1
(
DIGITAL PHYSICIAN
Waist-High Scale
Model 6127 Series, 6129 Series
Eye-Level Scale
Operating Instructions
CARDINAL SCALE MFG. CO. 8525-M310-O1 Rev E 06/15
www.detectoscale.com
PO BOX 151
417) 673-4631 FAX (417) 673-5001
PH
WEBB CITY, MO 64870
and
Printed in USA
Technical Support: Ph: 866-254-8261 techsupport@cardet.com
8525-M310-O1 Rev E 612X Series and 633X Series
1
Page 2
2
8525-M310-O1 Rev E 612X Series and 633X Series
Page 3
FCC COMPLIANCE STATEMENT
This equipment generates uses and can radiate radio frequency and if not installed and used in accordance with the instruction manual, may cause interference to radio communications. It has been tested and found to comply with the limits for a Class A computing device pursuant to Subpart J of Part 15 of FCC rules, which are designed to provide reasonable protection against such interference when operated in a commercial environment. Operation of this equipment in a residential area may cause interference in which case the user will be responsible to take whatever measures necessary to correct the interference.
You may find the booklet “How to Identify and Resolve Radio TV Interference Problems” prepared by the Federal Communications Commission helpful. It is available from the U.S. Government Printing Office, Washington, D.C. 20402. Request stock No. 001-000-00315-4.
PROPER DISPOSAL
When this device reaches the end of its useful life, it must be properly disposed of. It must not be disposed of as unsorted municipal waste. Within the European Union, this device should be returned to the distributor from where it was purchased for proper disposal. This is in accordance with EU Directive 2002/96/EC. Within North America, the device should be disposed of in accordance with the local laws regarding the disposal of waste electrical and electronic equipment.
It is everyone’s responsibility to help maintain the environment and to reduce the effects of hazardous substances contained in electrical and electronic equipment on human health. Please do your part by making certain that this device is properly disposed of. The symbol shown to the right indicates that this device must not be disposed of in unsorted municipal waste programs.
COPYRIGHT
All rights reserved. Reproduction or use, without expressed written permission, of editorial or pictorial content, in any manner, is prohibited. No patent liability is assumed with respect to the use of the information contained herein.
DISCLAIMER
While every precaution has been taken in the preparation of this manual, the Seller assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from use of the information contained herein. All instructions and diagrams have been checked for accuracy and ease of application; however, success and safety in working with tools depend to a great extent upon the individual accuracy, skill and caution. For this reason the Seller is not able to guarantee the result of any procedure contained herein. Nor can they assume responsibility for any damage to property or injury to persons occasioned from the procedures. Persons engaging the procedures do so entirely at their own risk.
Serial Number_______________________ Date of Purchase ____________________ Purchased Form_____________________
___________________________________ ___________________________________
RETAIN THIS INFORMATION FOR FUTURE USE
8525-M310-O1 Rev E 612X Series and 633X Series
PRECAUTIONS
Before using this instrument, read this manual and pay special attention to all "WARNING" symbols:
IMPORTANT ELECTRICAL
WARNING
3
I
Page 4
This page intentionally left blank.
4
II
8525-M310-O1 Rev E 612X Series and 633X Series
Page 5
INTRODUCTION
Thank you for your purchase of our Detecto Digital Physician Waist-High or Detecto Digital Physician Eye-Level scale. It has been manufactured with quality and reliability at our factory in Webb City, MO USA. Your scale has been tested before leaving our factory to insure accuracy and dependability for years to come.
This manual will guide you through assembly, installation, and operation of your scale. Please read it thoroughly before attempting to use your scale and keep it handy for future reference.
SPECIFICATIONS
Capacity: 550 lb x .2 lb
(250 kg x .1 kg), selectable
Platform Size: 10 5/8” x 14 1/2”
(270 mm x 368 mm)
Dimensions: 10 7/8” W x 20 7/8” D x 37 7/8” H
(276 mm x 530 mm x 962 mm) Shipping Weight: 34 lb (15.4 kg) 36 lb (16.3 kg) Scale Height: 37 1/2 ” (953 mm) 57 1/8” (1451 mm) Height Rod Measures: 6129, 6129KG, 6339 - 30” to 78” (76 cm to 200 cm) x inch/cm
6129KGM - 30” to 78” (76 cm to 200 cm) by 1/4”/1mm Display: Six digit, Seven segment, 0.7 inch (17.8 mm) high LCD Power Requirements: 6 “AA” size Alkaline, Ni-Cad or NiMH batteries (not included) OR an
optional 100 to 240 VAC 50/60Hz 12 VDC 1A wall plug-in UL/CSA
listed AC power adapter (Cardinal part number 6800-1045). Keyboard: Membrane type with 5 keys and 4 directional arrows
Waist-High Eye-Level
550 lb x .2 lb (250 kg x .1 kg), selectable
10 5/8” x 14 1/2” (270 mm x 368 mm)
10 7/8” W x 20 7/8” D x 57 1/2” H (276 mm x 530 mm x 1461 mm)
8525-M310-O1 Rev E 612X Series and 633X Series
1
Page 6
UNPACKING INSTRUCTIONS
Remove scale from carton by lifting up with equal force from column and platform base. Check for any damage incurred in shipping. If scale has been damaged, place a claim with
the carrier. Use the original carton and shipping material to return the scale.
Remove all plastic wrapping, foam fillers and cardboard material from the scale. Remove and unpack the power supply and cord, if the scale was ordered with this option. Carton contents:
1) Column
2) Indicator
3) Platform base
4) Screws and lock washers
5) Height measuring rod - optional
NOTE! If column bracket is bent, straighten before assembling.
ASSEMBLY INSTRUCTIONS
Refer to Figures No. 1, 2, 3, and 4, then follow the instructions for scale assembly.
1. Carefully remove load cell and draft rod assembly from column.
IMPORTANT! Handle load cell assembly with care! Do not drop, bend or twist.
2. Slip column over column bracket on platform base. Screw column to column brackets (two 1/4 truss head screws in front of column and four in the back). See Figure No. 1.
3. Referring to Figure No. 2, remove both #10-32 Truss Head Screws and lock washers securing the indicator bracket to the column.
4. Lower load cell and draft rod assembly into column. Draft Rod must pass through opening in column bracket. Referring to Figure No 1, make sure open end of lower draft rod hook is facing forward toward platform.
5. Next, referring to Figure No. 3, fasten top hook bolt of load cell assembly (C) to column. Make sure the load cell assembly does not touch the sides of column.
6. Slide grommet away from load cell assembly and slide into slot on column. NOTE! The slot should be on back (wide side) of column.
7. Pull excess load cell cable in column leaving enough slack in cable so as not to have any side pull on load cell.
8. Place digital indicator bracket on column and attach with 2 (#10-32) truss head screws and lock washers removed in step 3. See Figure No. 2.
9. Using a desk or table, lay down scale with column horizontal to floor. Raise draft rod keeping out of lever's way. Push lever. When lever is in position, hook draft rod around pivot. See Figure No. 4.
10. Attach indicator to indicator bracket on column, then connect load cell cable to back of indicator. Refer to the 750 Weight Indicator Owner’s Manual (8555-M483-O1) for additional information.
11. Scale is now ready for operation.
2
8525-M310-O1 Rev E 612X Series and 633X Series
Page 7
ASSEMBLY INSTRUCTIONS, CONT.
Grommet
Load Cell
Cable
Figure No. 1
Figure No. 3
Column
C
Draft Rod
#10-32 Truss Head Screw
Opening
Column Bracket
Raise
Draft
Rod
Push
Lever
Figure No. 4
Indicator Bracket
Lock Washer
Figure No. 2
8525-M310-O1 Rev E 612X Series and 633X Series
3
Page 8
OPERATION
ALWAYS assist the patient in stepping on and off the scale platform to ensure they do not fall. NEVER leave a patient unattended while they are on the scale platform. Failure to maintain control of the patient at all times can result in serious injury to you and/or the patient.
To Weigh
1. Set scale on any hard, level, flat surface or low-cut carpet.
2. Press ON/OFF key to turn on indicator.
3. Press ZERO key to zero weight display. The ZERO and lb or kg annunciator will turn on to show that scale is ready for use.
4. Place patient on scale and read weight display.
5. Remove patient from scale.
Zero Weight Display
1. If the indicator is not showing zero weight on the display, press ZERO key.
2. Weight display will return to zero. ZERO, STABLE  and lb or kg annunciators will turn on to show a stable, center-of-zero weight condition.
Metric Conversion
Press UNITS key to toggle between pounds and kilograms. Note that lb or kg annunciator will turn on to show which weighing unit is active.
4
8525-M310-O1 Rev E 612X Series and 633X Series
Page 9
OPERATION, CONT.
To Weigh and Calculate BMI
1. Press ON/OFF key to turn on indicator.
2. Press ZERO key to zero weight display. The ZERO and lb or kg annunciator will turn on
to show that scale is ready for use.
3. Place patient on scale and read weight display.
4. Press BMI / ENTER key.
5. If pounds is the active weighing unit, display will change to show 5
FEET
6 blinking. (The blinking character is the position to be changed).
a. Press  or  keys until desired height in inches is displayed and then press  or
key to move the blinking character to FEET. b. Press  or  keys until desired height in feet is displayed. c. Proceed to Step 7.
6. If kilograms is the active weighing unit, display will change to show 170 170 blinking. (The blinking character is the position to be changed).
a. Press  or  keys until the desired number is displayed. b. Press  or  key to move to next character to change. c. Repeat steps A and B until the desired height in centimeters is displayed. d. Proceed to Step 7.
7. Press BMI / ENTER key.
8. Read BMI on display.
9. Remove patient from scale.
10. Press BMI / ENTER key to return to weighing operation.
, 6
INCHES
.with the 0 of
cm
with the
8525-M310-O1 Rev E 612X Series and 633X Series
5
Page 10
HEIGHT ROD INSTRUCTIONS
(
)
(Waist-High Models)
Height Rod Hole Location
Please check the Height Rod hole locations before installing the Height Rod.
Some models may require holes to
be drilled.
Please follow the measurements for
drilling holes if necessary.
After holes have been drilled, please
follow installation instructions.
IMPORTANT!
Measure from bottom of column, not floor or scale base.
2in (51mm)
47/64”
29
(75.5 cm)
9/64”
(3.5mm) diameter holes, 1” (25.4mm) from center of column to center of hole
Height Rod Installation
1. Remove the height rod, height rod brackets and hardware pack from the cardboard insert.
2. Attach the height rod brackets to the height rod using the four screws, lock washers and nuts enclosed in the hardware pack. Tighten all four screws.
5
13.3 cm
Height Rod
15/64”
Height Rod Bracket
6
8525-M310-O1 Rev E 612X Series and 633X Series
Page 11
HEIGHT ROD INSTRUCTIONS, CONT.
(Waist-High Models)
3. Place the large end of the slotted holes in the height rod brackets over the four hex head screws on the back of the column.
4. Pull down to set the screws in the small end of the slotted hole and secure the bracket to the column.
5. Use the included wrench to tighten the hex head screws. Do not over-tighten the screws.
Height Rod Bracket
Scale
Column
8525-M310-O1 Rev E 612X Series and 633X Series
7
Page 12
HEIGHT ROD INSTRUCTIONS, CONT.
(Waist-High Models)
Operation
1. Before patient steps onto scale platform, spoon should be rotated to horizontal position, and raised well above patient's apparent height.
A
B
2. Patient may now step onto scale platform. Spoon should be held horizontal and above patient's head.
E
3. Carefully lower spoon, keeping it horizontal, until it rests gently on top of patient's head. If patient is shorter than 3' 4" (101.5 cm), push latch to right while simultaneously pushing down on spoon, until the spoon rests horizontally on top of
Height Rod in “rest” position
patient's head.
4. Read height of patient as follows:
A = Latch B = Spoon
If back of spoon points to outer height
rod, then it points to correct height.
If back of spoon points to inner height
rod, then correct height is read at top of
C = Inner Height Rod D = Outer Height Rod E = Measurement
“Read” Line
outer height rod (see "Read" arrow on outer height rod).
5. While holding spoon horizontally, raise spoon above patient's head. Patient may now step off scale platform. Hold spoon horizontal until patient is clear of height rod.
6. Rotate spoon back to vertical position and adjust height rod back to rest position (i.e. spoon should be locked in place within inner height rod and inner rod should be at its lowest position).
C
D
8
8525-M310-O1 Rev E 612X Series and 633X Series
Page 13
HEIGHT ROD INSTRUCTIONS, CONT.
(Eye-Level Models)
Height Rod Installation
1. Remove cardboard insert from height rod.
2. If not already installed, insert the two hex head screws from the hardware pack into holes in front of column and tighten them with included wrench until hex heads are 1/8" from column.
3. Place both height rod brackets over two pre-installed hex head screws and pull down.
4. Use included wrench to tighten both hex head screws. Do not overtighten screws.
8525-M310-O1 Rev E 612X Series and 633X Series
9
Page 14
HEIGHT ROD INSTRUCTIONS, CONT.
(Eye-Level Models)
1. Before the person steps onto the scale platform, the spoon should be rotated to the horizontal position, and raised well above the person’s apparent height.
2. The person may now step onto the scale platform. The spoon should be held horizontal and above the person’s head.
3. Carefully lower the spoon, while keeping it horizontal, until it rests gently upon the top of the person’s head. If the person is shorter than 101.5 cm/3' 4", then push the latch to the right, while simultaneously pushing down on the spoon, until the spoon rests horizontally upon the top of the person’s head.
4. Read the height of the person as follows:
If back of spoon points to outer height rod, then it points to correct height. If back of spoon points to inner height rod, then correct height is read at the top of
outer height rod (see “READ” arrow on outer height rod).
5. While holding the spoon horizontal, raise the spoon above the person’s head. The person may now step off of the scale platform. Hold the spoon horizontal until the person is clear of the height rod.
6. Rotate the spoon back to the vertical position and adjust the height rod back to the rest position (i.e. the spoon should be locked in place within the inner height rod and the inner rod should be at its lowest position).
CAUTION: Do not raise or lower Height Rod by the spoon tip!
10
8525-M310-O1 Rev E 612X Series and 633X Series
Page 15
ERROR AND STATUS DISPLAYS
Display Meaning
-Err-
-OF-
-trL-
-UnS-
CALib
AdErr ErrAL
General error, invalid keypad entry was attempted. Attempting to display a negative number greater than –9,999 or a
positive number greater than 99,999. Indicates an attempt to zero a weight outside scale zero range. Motion is present when indicator is attempting to perform one of the
following operations: Power Up Zero or Zero Weight Display. Indicates calibration is necessary.
Consult your scale service representative.
ErrAH
OCAP
OFF
Scale weight exceeds scale capacity. Displayed to indicate indicator is turning off.
CARE AND CLEANING OF SCALE
DO NOT subject platform to sudden shocks.  DO NOT submerge indicator in water, pour or spray water directly on indicator.  DO NOT use abrasive cleaners on this scale.  DO NOT use acetone or other volatile solvents for cleaning.  DO clean scale and indicator using a damp soft cloth and mild detergent.  DO avoid areas where scale might be exposed to moisture.
8525-M310-O1 Rev E 612X Series and 633X Series
11
Page 16
PARTS IDENTIFICATION
FINAL ASSEMBLY
ITEM QTY PART NUMBER DESCRIPTION
1 1 3P1201X-1 BASE ASSEMBLY 2 1 430XG3R800 COLUMN ASSEMBLY FOR 6127(KG), 6129(KG)
1 0044-C180-1A COLUMN ASSEMBLY FOR 6337, 6339 3 1 709GR1206 LOAD CELL ASSY 4 1 0033-B254-08 DRAFT ROD FOR 6127(KG), 6129(KG)
1 0033-B256-08 DRAFT ROD FOR 6337, 6339 5 1 3P8068 PLATFORM COVER 6 4 63K1038 BASE LEG 7 1 6650-0087 LABEL, MADE IN USA 8 1 0033-B438-0A MOUNTING BRKT 9 1 391RV204 NUT ELASTIC STOP 1/4-20 Z/P
10 1 6021-1454 SCW HEX HEAD 1/4 -20 X 3/4 11 1 6024-0004 WASHER FLAT 1/4 12 1 0033-B104-0A TOP PLATE ASSY. 13 1 593R986 SERIAL TAG 14 2 6024-0049 WASHER LOCK EXT #10 15 6 6024-0047 WASHER LOCK EXT 1/4 16 6 6021-1423 SCW TRUSS HEAD 1/4-20 X 1/2 17 2 6021-1045 SCW HEX HEAD #10-32 X 1/2 Z/P 18 1 750 DIG. WT. INDICATOR 19* 1 0033-D203-0A HEIGHT ROD ASSEMBLY, INCH/CM FOR 6129(KG) 20* 1 0033-D475-0A HEIGHT ROD ASSEMBLY, 1/4”/1mm FOR 6129KGM 21* 1 0033-D063-1A HEIGHT ROD ASSEMBLY, INCH/CM FOR 6339 22* 1 6800-1045 OPTIONAL AC ADAPTER 100-240VAC/12VDC @ 1A
* NOT SHOWN
12
8525-M310-O1 Rev E 612X Series and 633X Series
Page 17
PARTS IDENTIFICATION, CONT.
BASE ASSEMBLY
ITEM QTY PART NUMBER DESCRIPTION
1 1 3P2011X PLATFORM WELDMENT 2 4 3P2087 PLATFORM LOAD BEARING 3 1 3P1001X BASE WELDMENT 4 2 3P60 CHECK PLATE 5 2 6021-0985 HEX WASHER HD SCW #10 X 1/2 6 4 63K1038 BASE LEG 7 1 2U58 UNION CENTER HANGER 8 1 3P8003X SHORT LEVER ASSEMBLY
9 4 3P8059 UNION HANGER 10 1 3P122 BUMPER 11 1 3P8002X LONG LEVER ASSEMBLY 15 6 6024-0047 WASHER LOCK EXT 1/4 16 6 6021-1423 SCW TRUSS HEAD 1/4-20 X 1/2
8525-M310-O1 Rev E 612X Series and 633X Series
13
Page 18
STATEMENT OF LIMITED WARRANTY
Detecto Scale warrants its equipment to be free from defects in material and workmanship as follows: Detecto warrants to the original purchaser only that it will repair or replace any part of equipment which is defective in material or workmanship for a period of two (2) years from date of shipment. Detecto shall be the sole judge of what constitutes a defect.
During the first ninety (90) days Detecto may choose to replace the product at no charge to the buyer upon inspection of the returned item.
After the first ninety (90) days, upon inspection of the returned item, Detecto will repair or replace it with a remanufactured product. The customer is responsible for paying for t he freight both ways.
This warranty does not apply to peripheral equipment not manufactured by Detecto; this equipment will be covered by certain manufacturer’s warranty only.
This warranty does not include replacement of expendable or consumable parts. This does not apply to any item which has deteriorated or damaged due to wear, accident, misuse, abuse, improper line voltage, overloading, theft, lightning, fire, water or acts of God, or due to extended storage or exposure while in purchaser’s possession. This warranty does not apply to maintenance service. Purchased parts will have a ninety (90) day repair or replacement warranty only.
Detecto may require the suspect product to be ret urned to the factory; item(s) must be properly packed and shipping ch arges prepaid. A return authorization number must be obtained for all returns and marked on the outside of all returned packages. Detecto accepts no responsibility for loss or dam age in transit.
14
8525-M310-O1 Rev E 612X Series and 633X Series
Page 19
STATEMENT OF LIMITED WARRANTY
Conditions Which Void Limited Warranty
This warranty shall not apply to equipment which: A.) Has been tampered with, defaced, mishandled or has had
repairs and modifications not authorized by Detecto. B.) Has had serial number altered, defaced, or removed. C.) Has not been grounded according to Detecto’s recommended
procedure.
Freight Carrier Damage
Claims for equipment damaged in transit must be referred to the freight carrier in accordance with freight carrier regulations.
This warranty sets forth the extent of our liability for breach of any warranty or deficiency in connection with the sale or use of the product. Detecto will not be liable for consequential damages of any nature, including but not limited to, loss of profit, delays or expenses, whether based on tort or contract. Detecto reserves the right to incorporate improvements in material and design without notice and is not obligated to incorporate improvements in equipment previously manufactured.
The foregoing is in lieu of all other warranties, express or implied including any warranty that extends beyond the descripti on of the product including any warranty of merchantability or fitness for a particular purpose. This warranty covers only those Detecto products installed in the forty-eight (48) contiguous continental United States.
Ph. (800) 641-2008
E-mail: detecto@cardet.com
203 E. Daugherty
Webb City, MO 64870
D268-WARRANTY-DET-A
08/14
Printed in USA
8525-M310-O1 Rev E 612X Series and 633X Series
15
Page 20
16
8525-M310-O1 Rev E 612X Series and 633X Series
Loading...