Carbolite VCF User Manual

CONTENTS
Section page
Installation, Operation &
Maintenance Instructions
1200°C Vertical Chamber Furnaces
VCF models
This manual is for the guidance of operators of the above Carbolite products
and should be read before the furnace is connected to the electricity supply.
Manuals are supplied separately for the furnace controller
(and overtemperature controller when fitted).
Please read the controller manuals before operating the furnace.
MF25 – 3.09
Instrument switch: when the instrument switch is operated the temperature control circuit is energised.
Heat Light: the adjacent light glows or flashes to indicate that power is being supplied to the elements
DANGER of electrical shock– read any warning printed by this symbol.
DANGER – hot surface. Read any warning printed by this symbol. WARNING: all surfaces of a furnace may be hot.
DANGER – read any warning printed by this symbol.
VCF
1.0 SYMBOLS & WARNINGS
1.1 Switches and Lights
1.2 Warning Symbols
2 MF25 – 3.09
VCF
note: throughout this
manual the furnace lid is
referred to as a “door”.
2.0 INSTALLATION
2.1 Unpacking & Handling
When unpacking or moving the furnace always lift it by its
base. Never lift it by the door or by any inset or protruding parts. Use two people to carry the furnace where possible. Remove any packing material from the door gear and furnace chamber before use.
NOTE: This product contains Refractory Ceramic Fibre (better described as Alumino Silicate Wool) for precautions and advice in handling this material see the ‘Repairs and Replacements’ section.
2.2 Siting & Setting Up
Place the furnace in a well ventilated room, away from other sources of heat, and on a surface
which is resistant to accidental spillage of hot materials. Do not mount the furnace on an inflammable surface.
Ensure that there is free space around the furnace. Do not obstruct any of the vents in the control
section: they are needed to keep the controls cool.
Ensure that the furnace is placed in such a way that it can be quickly switched off or disconnected
from the electrical supply - see below.
If the chimney is supplied unfitted, then fit it into the hole in the top of the furnace door. If the
furnace is to be used to heat substances which emit fumes, then a fume extraction duct of about 150mm inlet diameter may be placed directly above the chimney outlet; the duct would have to be removable or flexible. Do not make a sealed connection to the furnace chimney as this causes excessive airflow through the chamber and results in poor temperature uniformity.
CONNECTION DETAILS
supply type
Supply
Terminal label
Cable colour
Live-Neutral
Reversible or Live-Live
1-phase
L
Brown
To live
to either power conductor
N
Blue
To neutral
to the other power conductor
PE
Green/Yellow
To earth (ground)
to earth (ground)
supply
Terminal label
Cable colour
2- or 3-phase
L1
Black
to phase 1
L2
Black
to phase 2
L3
Black
to phase 3 except 2-phase
N
Light Blue
to neutral except delta
PE
Green/Yellow
to earth (ground)
VCF
2.3 Electrical Connections
Connection by a qualified electrician is recommended
All furnaces covered by this manual may be ordered for single phase A.C. supply, which may be
Live to Neutral non-reversible, Live to Neutral reversible or Live to Live. The 23 litre models can also be supplied for two or three phase use, and the 100 litre is normally supplied for three phase. Three phase supplies may be with or without neutral: the furnace must only be connected to the correct type of supply.
Check the furnace rating label before connection. The supply voltage should agree with the voltage
on the label, and the supply capacity should be sufficient for the amperage on the label.
The supply should be fused at the next size equal to or higher than the amperage on the label. A
table of the most common fuse ratings is also given in section 8.1 of this manual. Where a supply
cable is present there are internal supply fuses; customer fusing is preferred but not essential. Furnace with supply cable: either wire directly to an isolator or fit with a line plug. Furnace without supply cable: a permanent connection to a fused and isolated supply should be
made to the internal terminals after temporary removal of the furnace back panel. Connection by line plug: the plug should be within reach of the operator, and should be quickly
removable. Connection to isolating switch: this should operate on both conductors (single phase) or on all live
conductors (three phase), and should be within reach of the operator. The supply MUST incorporate an earth (ground).
4 MF25 – 3.09
VCF
3.0 OPERATION
The instructions for operating the temperature controller are given in a separate manual. If the furnace is fitted with a time switch, see also the supplementary manual MS03. If cascade control is fitted, see the supplementary manual MS07.
3.1 Operating Cycle
The furnace is fitted with an Instrument switch. The switch cuts off power to the controller(s) (and
contactor when fitted). Connect the furnace to the electrical supply. Operate the instrument switch to activate the temperature controller. The controller becomes
illuminated and goes through a short test cycle. Close the furnace door and adjust the temperature controller – see the controller manual. Overtemperature option only. If the overtemperature controller has not yet been set as required, set
it and activate it according to the instructions in the appropriate manual. Unless a time switch is fitted and is off, the furnace starts to heat up. The Heat light(s) glow
steadily at first and then flash as the furnace approaches the desired temperature or a program
setpoint. Overtemperature option only. If the overtemperature trip operates then an indicator in the
overtemperature controller flashes, and the heating elements are isolated. Find and correct the
cause before resetting the overtemperature controller according to the instructions supplied. To turn the furnace off, set the Instrument switch to it’s off position; the controller display will go
blank. If the furnace is to be left off unattended, isolate it the electrical supply.
3.2 General Operating Notes
Heating element life is shortened by use at temperatures close to maximum. Do not leave the
furnace at high temperature when not required. The maximum temperature is shown on the furnace
rating label and on the back page of this manual. When heating large objects, in particular poor conductors, avoid shielding the thermocouple from
the heating elements. The thermocouple is intended to sense the temperature near the heating
element, but if a large object is placed in the chamber it may record the average temperature of the
object and the elements, which can lead to overheating of the elements. Allow large objects to gain
heat at a lower temperature and then reset the controller to a temperature close to the desired
maximum. When heating materials which produce smoke or fumes, the chimney must be correctly fitted and
unobstructed. Otherwise, soot accumulates in the chamber and could possibly cause an electrical
breakdown of the heating element. If the furnace is used to heat materials which emit smoke or fumes, regularly heat it up to
maximum temperature for one hour without load to burn away the soot. Materials such as case hardening compounds and other reactive salts may penetrate the furnace
chamber lining and attack the wire elements, causing premature failure. Use of a hearth tile may be
advisable: please consult our technical department. Lightweight ceramic fibre insulation can easily be marked by accidental contact. Some fine cracks
may be visible on the surface of the insulation, or may develop in the surface of the chamber due
to progressive shrinkage of the insulation materials. Cracks are not usually detrimental to the
functioning or the safety of the furnace.