Caradyne Whisperflow 2 User manual

Whisperflow 2
CPAP System
User’s Manual
Table of contents
Table of contents ........................................................ ii
Introduction .................................................................1
How it works ...............................................................2
Setting up the Whisperflow 2 System.........................2
Using Whisperflow 2 with Criterion 60........................7
Operating the Whisperflow 2 System .......................14
Monitoring the patient ...............................................15
Maintenance .............................................................15
Environmental specifications ....................................16
Accessories ..............................................................19
Service......................................................................19
Warranty ...................................................................19
Approvals..................................................................19
Manufacturer.............................................................20
Year of manufacture .................................................20
Introduction
The Whisperflow 2 CPAP System delivers continuous positive airway pressure (CPAP) throughout the breathing cycle. It provides CPAP at preset levels throughout inspiration and exhalation, independent of the patient's flow rate. The Whisperflow 2 is intended for use on spontaneously breathing patients, and allows you to adjust FIO
The Whisperflow 2 controls are easy to use, and a swivel elbow outlet port simplifies circuit configuration. An internal safety valve eliminates the need for a separate pressure relief valve and tee. The Whisperflow 2 can be pole or bench mounted, and can be attached to Caradyne’s Criterion monitors (Figure 1) to create a system that delivers CPAP and measures applicable
from 28 to 100%.
2
Criterion 60
Whisperflow 2
parameters.
Figure 1. Whisperflow 2 CPAP System
with Criterion Monitor
The Whisperflow 2 System may be contraindicated for patients with any of these conditions:
Fluid retention
Pneumothorax
Decreased cardiac output and gastric distention
Severe facial injury (noninvasive use)
Hypotension secondary to hypovolemia.
How it works
The Whisperflow 2 CPAP System is a precision venturi device that uses an oxygen supply in conjunction with entrained air to generate an output flow. The Whisperflow 2 System uses a 60 psi (412 kPa) oxygen supply, and can generate flows over 140 L/min with a minimum fractional inspired oxygen (FIO 28%. The Whisperflow 2 CPAP valves use constant force springs to maintain the preset positive pressure at flow rates from 10 to 140 L/min.
2
) of
2 Whisperflow 2 CPAP System User's Manual 8-100482-00 Rev. C
Setting up the Whisperflow 2 System
WARNING:
Always use the Whisperflow 2 System with
an oxygen analyzer to alert you to a loss of oxygen wall pressure or other system faults. For complete control of the CPAP therapy, Caradyne recommends using the Criterion 60 Pressure and Oxygen monitor.
Always ensure that the CPAP valve is
correctly connected to the patient circuit, and that the exhaust port of the CPAP valve is not attached to the circuit.
Use an in-line tee to insert a pressure gauge
or pressure monitor (such as the Criterion 40 or Criterion 60) between the output port of the Whisperflow 2 and the patient interface.
The Whisperflow 2 (as is typical of other
CPAP systems) does not include a non­rebreathing valve for use with tracheal tubes. Oxygen supply failure during tracheal tube connection can result in rebreathing and asphyxiation.
CAUTION: To avoid damage to the Whisperflow 2 System by entraining particles into the device, always install a filter (P/N 8-100471-00) on the fresh gas intake port.
1. Select a sealing face mask or endotracheal (ET) tube as
the patient connection. Ensure that the mask fits comfortably, seals the bridge of the nose, and fully covers the nose and mouth.
2. Choose the CPAP valve (see the parts list for a complete
list of available CPAP valve sizes and part numbers).
3. Assemble the circuit according to Figures 3 through 7,
depending on the mask, oxygen analyzer, and humidifier used. As you assemble the circuit, please note:
The Whisperflow 2 includes an internal safety valve that automatically relieves pressure at approximately 28 cmH
O or above. To relieve pressure at lower
2
levels, use a separate pressure relief valve (use safety valve tee, Caradyne part number 8-202502-00, to insert a separate safety valve into the circuit).
To ensure accurate measurement of the FIO
2
delivered to the patient, use an inline oxygen analyzer upstream of the humidification device (if used).
The Whisperflow 2 System can deliver a minimum
of 28-29% to the patient. However, the longer
FIO
2
the tube lengths used, the higher the minimum FIO (longer tubing increases the resistance of the patient circuit, allowing Whisperflow 2 to entrain less air).
The Whisperflow 2 System's location upstream of the patient and the high flows it produces make contamination and cross-infection highly unlikely. However, if cross contamination is a concern, attach a 22-mm filter to the output port of the Whisperflow 2.
Use the Whisperflow 2 with accessories supplied by Caradyne, Ltd. or authorized representative. Caradyne supplies face masks with an integral one­way valve that prevents exhalation into the supply gas tubing in the event of supply gas failure. However, when used with a tee piece, exhalation into the supply tubing can result in rebreathing and asphyxia.
Caradyne supplies face masks with an integral one- way valve that minimizes the risk of asphyxia in case of supply gas failure. However, a face mask with one­way valve results in an increased work of breathing in the event of loss of oxygen that may not be well tolerated by all patients. Do not use a mask or tracheal tube connection when there is no oxygen flow from the Whisperflow 2.
2
4 Whisperflow 2 CPAP System User's Manual 8-100482-00 Rev. C
Use an external monitor (such as a Criterion monitor) to ensure that the Whisperflow 2 is supplying adequate flow to the patient. If flow is adequate, an in-line pressure gauge or pressure monitor shows a constant pressure approximately equal to the PEEP valve specification. A significant (2 cmH
O or more)
2
pressure drop when the patient inhales indicates inadequate flow: increase flow from the Whisperflow
2.
Use an oxygen monitor with alarms (such as the Criterion 60 Pressure and Oxygen monitor) to set and monitor the FiO
delivered by the Whisperflow 2. An
2
oxygen monitor can signal loss of oxygen flow if the patient continues to draw fresh gas without oxygen through the air inlet of the Whisperflow 2.
The Whisperflow 2 is shipped with covers for the oxygen sensor and pressure ports. Keep these covers in place whenever the ports are not in use. Remove the oxygen sensor cover before installing a Criterion 60 oxygen sensor. Remove the pressure port cover when installing a pressure line to the Whisperflow 2.
When using the Whisperflow 2 pressure port, take into consideration that the pressure reading includes the pressure drop due to the resistance of the patient circuit. For example, if the pressure at the patient’s mask is 5 cmH resistance of the circuit at the current flow is 2 cmH then the monitor pressure reading is 7 cmH
0 and the pressure drop due to the
2
0.
2
0,
2
Because high pressure alarms are typically set 10 cmH
0 or more above peak inspiratory pressures, the
2
Whisperflow 2 pressure port is appropriate for use where the pressure drop due to patient circuit resistance is not significant. At very low patient ventilation pressures when small variations in pressure can be significant, measure pressure at the patient connection to eliminate the effect of patient
circuit resistance. Replace pressure tubing and filter monthly during normal use.
6 Whisperflow 2 CPAP System User's Manual 8-100482-00 Rev. C
Using Whisperflow 2 with Criterion 60
The Whisperflow 2 System is designed to be compatible with all Criterion monitors. Follow these steps to attach the Whisperflow 2 to the Criterion 60 Pressure and Oxygen Monitor (Figure 2):
1. Remove the rubber feet from the base of the Criterion 60.
2. Remove the rubber feet from the base of the Whisperflow
2.
3. If using the pole mounting kit (P/N 8-100315-00), connect the mounting bracket to the Criterion 60 and Whisperflow 2 as described in the kit instructions.
4. Turn the Criterion 60 upside down, set the Whisperflow 2 on top of the Criterion 60, and align the mounting locations.
5. Use the four M3 x 25 screws to connect the Criterion 60 and Whisperflow 2.
6. Replace the rubber feet to the base of the Whisperflow 2.
7. Once the Whisperflow 2 and Criterion 60 are attached, connect the oxygen sensor and pressure line as shown in Figures 3, 5, or 7.
Figure 2. Attaching the Whisperflow 2 to the Criterion 60
8 Whisperflow 2 CPAP System User's Manual 8-100482-00 Rev. C
Figure 3. Whisperflow 2 and Criterion 60, dual-port mask
Oxygen sensor
Oxygen analyzer
Whisperflow 2
Pressure monitor
P
CPAP valve
Figure 4. Whisperflow 2 and oxygen analyzer, dual-port
mask
10 Whisperflow 2 CPAP System User's Manual 8-100482-00 Rev. C
Pressure line
Oxygen sensor
CPAP valve
To ET tube
Figure 5. Whisperflow 2 and Criterion 60, ET tube
connection
Whisperflow 2 CPAP System User's Manual 11 8-100482-00 Rev. C
Oxygen sensor
Oxygen analyzer
CPAP valve
P
Whisperflow 2
Figure 6. Whisperflow 2 and oxygen analyzer, ET tube
connection
12 Whisperflow 2 CPAP System User's Manual 8-100482-00 Rev. C
Humidifier chamber
CPAP valve
Figure 7. Installing a humidifier
Whisperflow 2 CPAP System User's Manual 13 8-100482-00 Rev. C
Operating the Whisperflow 2 System
1. Turn control valves off by turning fully clockwise (finger-
tight only).
2. Connect the oxygen hose to the port on the Whisperflow 2
back panel, then listen for leaks. (Connect directly to the oxygen supply or connect an oxygen hose between the oxygen inlet and the oxygen supply.)
WARNING: Do not connect any gas supply other than oxygen to the Whisperflow 2 System.
3. Turn the ON/OFF knob on. Turn the flow adjustment valve
fully on and check that the output flow from the patient outlet varies accordingly.
4. Turn the oxygen adjustment valve counter-clockwise until
the Criterion 60 or other oxygen analyzer displays the appropriate FIO
5. Leave the oxygen and flow controls as you have just set
for 30 seconds.
2
them, then turn the ON/OFF knob off. Attach the circuit to the patient, then turn the ON/OFF knob on.
6. Watch the CPAP valve to ensure that it remains open
during inspiration. Gradually reduce the circuit flow by turning the flow adjustment valve down to point when you can still feel a slight flow at the CPAP valve when the patient inspires. If you are using a Criterion monitor, watch the bar graph to verify sufficient flow.
7. Monitor the FIO
circuit flow decreases, the FIO reduce the FIO
as you reduce the circuit flow: as the
2
by turning the oxygen adjustment
2
rises slightly. You can
2
clockwise.
14 Whisperflow 2 CPAP System User's Manual 8-100482-00 Rev. C
Monitoring the patient
During operation, be sure to check the following on regular basis:
Ensure that there are no leaks at the patient connection.
Ensure that there is flow from the CPAP valve during
inspiration (which means that Whisperflow 2 is supplying adequate flow to meet patient demand). If you are using the Whisperflow 2 with a Criterion monitor, watch the pressure bar graph: flow is adequate if pressure fluctuations are minimal (± 2 cmH
O).
2
Monitor the patient's arterial blood oxygen saturation
).
(SaO
2
Monitor the patient for signs of dehydration and discomfort in the upper airways.
Use the Criterion 60 or other reliable oxygen analyzer to monitor the FIO
delivered to the patient.
2
Maintenance
Under normal conditions, the Whisperflow 2 System does not require any special maintenance or sterilization. However, the Whisperflow 2 System can be gas-sterilized using ethylene oxide (EtO). Following sterilization, open all the valves fully, connect to an oxygen supply for five minutes, and verify performance before reuse.
CAUTION: Do not autoclave or immerse the Whisperflow 2 System or any of its components in any solution. The circuit, including the inlet filter, is for single patient use.
Follow these steps to test the Whisperflow 2 System's minimum FIO
1. Set the flow adjustment to its maximum, and the oxygen
annually:
2
adjustment valve to its minimum.
2. Attach a 10-cmH
O CPAP valve to the Whisperflow 2
2
using a patient circuit.
3. Ensure that the FIO
at a circuit pressure of 10 cmH2O
2
does not exceed 30%.
Whisperflow 2 CPAP System User's Manual 15 8-100482-00 Rev. C
Environmental specifications
Temperature
Operating: 5 to 45 °C at 15 to 95% relative humidity Storage: -40 to 60 °C at 95% relative humidity
Oxygen supply
40 to 80 psi (275 to 550 kPa) See Figure 11 for flow requirements.
Performance
Figures 8 through 11 characterize Whisperflow 2 performance.
160
155
150
145
140
Flow (L/min)
135
130
125
120
5 7.5 10 12.5 15 17.5 20
Pre ssur e (cm H2 O)(h Pa)
Figure 8. Maximum flow and minimum oxygen
versus pressure
32
31.5
31
30.5
30
29.5
29
28.5
28
Flow O2
% O2
16 Whisperflow 2 CPAP System User's Manual 8-100482-00 Rev. C
36
35
34
33
32
31
% Oxygen
30
29
28
27
26
40 60 80 100 120
Flo w (L/m iin)
Figure 9. Minimum oxygen achievable at different flow
settings
for different PEEP values
2
1.8
1.6
1.4
1.2
1
0.8
Pressure (cmH2O)(hPa)
0.6
0.4
0.2
0
40 60 80 100 120
Flo w (L /m in)
Figure 10. Circuit pressure loss per square meter of
tubing
2.5 5
7.5 10 15
Whisperflow 2 CPAP System User's Manual 17 8-100482-00 Rev. C
120
)
100
80
60
40
Supply Oxygen Consumption (L/min
20
PEEP
40 L/min 60 L/min 80 L/min 100 L/min 120 L/min 140 L/min
0
30 50 70 100
Patient Oxygen %
Figure 11. Oxygen supply consumption at several flow
and oxygen settings
18 Whisperflow 2 CPAP System User's Manual 8-100482-00 Rev. C
Accessories
The following accessories are compatible with the Whisperflow 2 CPAP System:
Humidifier: Fisher and Paykel MR730 and MR480. The MR410 Humidifier can also be used when inspiratory flow rates are set to less than 50 L/min.
Oxygen monitor: You can use any in-line oxygen monitor with the Whisperflow 2 System. Caradyne recommends the Criterion 60 Pressure and Oxygen Monitor.
Please contact your local dealer before using other accessories with the Whisperflow 2 System.
Service
Full service and repair is available from:
Caradyne Parkmore Business Centre Parkmore West Galway, Ireland Telephone: +353 (0)91-709010 Fax: +353 (0)91-758929
Or contact your local dealer.
Caradyne 2055 Executive Drive Indianapolis, IN 46241 USA Telephone: 317-246-9010 Fax: 317-246-9011
Warranty
The Whisperflow 2 System is warranted against defects in material and workmanship for two years from the time of sale.
Approvals
The Whisperflow 2 System complies with the requirements of directive 93/42/EEC concerning medical devices and therefore bears the CE mark.
0050
Whisperflow 2 CPAP System User's Manual 19 8-100482-00 Rev. C
Manufacturer
Caradyne Ltd. Parkmore Business Park, Parkmore West Galway, Ireland Telephone: +353 (0)91-709010 Fax: +353 (0)91-758929
E-mail: info@caradyne.com Internet: http://www.caradyne.com
Year of manufacture
The fifth and sixth digits of the serial number indicate the Whisperflow 2 System's year of manufacture.
Parts list
Description Part number
Whisperflow 2 CPAP Systems
Whisperflow 2 CPAP System, Europe 8-100474-00 Whisperflow 2 CPAP System,
USA/Canada
CPAP valves, 22-mm connection
CPAP valve, 2.5 cmH2O 8-208503-00 CPAP valve, 5 cmH2O 8-208505-00 CPAP valve, 7.5 cmH2O 8-208507-00 CPAP valve, 10 cmH2O 8-208510-00 CPAP valve, 12.5 cmH2O 8-208512-00 CPAP valve, 15 cmH2O 8-208515-00 CPAP valve, 20 cmH2O 8-208520-00
CPAP valves, 30-mm connection
CPAP valve, 2.5 cmH2O 8-209503-00 CPAP valve, 5 cmH2O 8-209505-00 CPAP valve, 7.5 cmH2O 8-209507-00 CPAP valve, 10 cmH2O 8-209510-00
8-100475-00
20 Whisperflow 2 CPAP System User's Manual 8-100482-00 Rev. C
Description Part number
CPAP valve, 12.5 cmH2O 8-209512-00 CPAP valve, 15 cmH2O 8-209515-00 CPAP valve, 20 cmH2O 8-209520-00
CPAP masks, SealFlex
Mask, dual-port, large 8-209000-00 Mask, dual-port, medium 8-209001-00 Mask, dual-port, small 8-209002-00
CPAP masks, AirSeal
Mask, dual-port, large 4-005000-00 Mask, dual-port, medium 4-005001-00 Mask, dual-port, small 4-005002-00
CPAP masks, multi-strap
Mask, SealFlex multi-strap, 30-mm, large 4-005520-00 Mask, SealFlex multi-strap, 30-mm,
4-005521-00
medium Mask, SealFlex multi-strap, 30-mm, small 4-005522-00 Mask, SealFlex multi-strap, 22-mm, large 4-005523-00 Mask, SealFlex multi-strap, 22-mm,
4-005524-00
medium Mask, SealFlex multi-strap, 22-mm, small 4-005525-00
Filter, 22M/22F ports 8-208570-00 Filter, Whisperflow 2 8-100471-00 Connectors
Connector, 22F/22F ports 8-201009-00 Connector, 22M/22M ports 8-201007-00 Connector, 22F/30F ports 8-100264-00 Connector, 22M/30M ports 4-005053-00
Headstraps
Headstrap, multi-strap kit-cap-white, large 4-005572-00 Headstrap, multi-strap kit-cap-black, large 4-005573-00 Headstrap, multi-strap kit-cap-white,
4-005574-00
medium
Whisperflow 2 CPAP System User's Manual 21 8-100482-00 Rev. C
Description Part number
Headstrap, multi-strap kit-cap-black,
4-005575-00
medium Headstrap, multi-strap kit ‘H’ white 4-005576-00 Headstrap, multi-strap kit ‘H’ white 4-005577-00 Head strap, four-point, latex 8-209102-00 Head strap, four-point, non-latex 8-209103-00
Whisperflow 2 CPAP System User’s Manual 8-100482-00 Pole mounting kit 8-100315-00 Patient monitors
Criterion 40 Pressure Monitor, UK/Ireland 8-100210-00 Criterion 40 Pressure Monitor, Europe 8-100324-00 Criterion 40 Pressure Monitor,
8-100325-00
USA/Canada Criterion 60 Pressure/Oxygen Monitor,
8-100352-00
UK/Ireland Criterion 60 Pressure/Oxygen Monitor,
8-100351-XX
Europe Criterion 60 Pressure/Oxygen Monitor,
8-100353-00
USA/Canada
Caradyne continues to develop new accessories. Contact your local dealer for an updated list.
22 Whisperflow 2 CPAP System User's Manual 8-100482-00 Rev. C
Loading...