Caradyne Criterion 60 User manual

Criterion 60
Pressure/Oxygen
Monitor
User’s Manual
Table of contents
Table of contents _________________________________iii Introduction _____________________________________ 1
Labels________________________________________ 3
Connecting the Criterion 60 to power _________________ 4 Connecting the Criterion 60 to the patient system________ 5 The Criterion 60 front panel _________________________ 9 Functional test __________________________________11 Normal operation ________________________________ 12
Pressure bar graph ____________________________ 13 Oxygen concentration measurement _______________ 14 Language selection ____________________________ 15 Illegal key press _______________________________ 15 AC power loss ________________________________ 15
Criterion 60 Alarms ______________________________ 16
When an alarm is active_________________________ 19 Adjusting alarm settings _________________________ 20 Alarm silence _________________________________ 20 Alarm presilence ______________________________ 20
Maintenance ___________________________________ 21
Calibration procedures __________________________ 22
Zero calibration: pressure (non-technical user) ____ 23 Gain calibration: pressure (qualified technician only) 24 21% calibration: oxygen (non-technical user)______ 28 Gain calibration: oxygen (non-technical user)______ 29 Battery pack recharge (non-technical user) _______ 31 Battery pack replacement (qualified technician only) 32
Specifications___________________________________ 33 Service _______________________________________ 35 Warranty ______________________________________ 35 Approvals______________________________________ 35 Parts list_________________________________________
_______________________________________ 37
Appendix: Effect of humidity on oxygen measurements __ 38
iii
Figures
Figure 1. Criterion 60 Pressure/Oxygen Monitor __________2
Figure 2. Connecting the Criterion 60 to AC (wall) power ___ 4
Figure 3. Connecting the Criterion 60 to the patient system _6
Figure 4. Connecting the Criterion 60 to a face mask ______ 7
Figure 5. Connecting the Criterion 60
to a tee or wye and ET tube ______________________8
Figure 6. Criterion 60 front panel _____________________ 10
Figure 7. Pressure monitoring screen _________________ 12
Figure 8. Oxygen monitoring screen __________________ 13
Figure 9. Entering service mode _____________________22
Figure 10. Calibration kit and setup ___________________ 25
Figure 11. Pressurizing the system ___________________26
Figure 12. Replacing the battery pack _________________ 32
Figure 13. Saturated water vapor pressure (SVP)
of gas (100% RH) versus temperature _____________ 38
Tables
Table 1. Criterion 60 front panel ______________________ 9
Table 2. Criterion 60 alarms ________________________ 16
Table 3. Selectable alarm limits______________________18
iv
8-100333-00 Rev. B
Introduction
The Criterion 60 Pressure/Oxygen Monitor (Figure 1) monitors patient airway pressure and oxygen concentration throughout the breathing cycle. The monitor uses solid-state pressure transducers to measure circuit and atmospheric pressures, and a galvanic oxygen sensor to measure the partial pressure of oxygen. The oxygen sensor samples gas directly without diverting it from the circuit. The monitor uses the atmospheric pressure and partial pressure of oxygen measurements to derive oxygen concentration.
The Criterion 60 is intended for use with devices that are not capable of monitoring pressure or oxygen concentration (for example, a resuscitation bag or basic ventilator) or as an independent backup for devices that do measure pressure or oxygen.
The Criterion 60 alarms when the airway pressure or oxygen concentration falls outside of the user-selected high and low alarm limits, and displays real-time airway pressure and oxygen concentration.
The Criterion 60 also includes alarm delay (for pressure alarms), alarm silence, and alarm presilence features. Battery backup provides up to 24 hours of backup power in case of AC power failure or when AC power is not available.
NOTE: The Criterion 60 is designed for stationary and intra­hospital transport use only.
Power
Oxyge
supply
Criterion 60 Pressure/
n Monitor
Pressure
Oxygen sensor cable
Oxygen sensor tee
Figure 1. Criterion 60 Pressure/Oxygen Monitor
WARNINGS:
To avoid the risk of fire, do not use this device near
flammable anesthetics.
Federal law (U.S.) restricts the sale of this device to, or by
the order of, a physician.
Do not use this device to monitor pressures when patient
respiratory rate exceeds 60 breaths per minute.
This device is not suitable for use in a MRI environment.
2
8-100333-00 Rev. B
Labels
y
Rear panel label
Manufactured b Caradyne Parkmore Business Centre Parkmore West, Galway, Ireland Tel: + 353 (0) 91 709010 Fax: + 353 (0)91 758929
2055 Executive Dri ve Indianapol is, IN 46241, USA Tel: 317-246- 9010 Fax: 317-246-9011
Email : i nfo@cara dyne.c om www.cara dyne.c om
NOTE: The symbol “P” on the back panel label shows the pressure tubing connection, and the symbol “O
” shows the
2
oxygen sensor connection.
Battery compartment label
Connecting the Criterion 60 to power
The Criterion 60 comes with a power supply that accepts mains power (120 or 230 V AC, depending on model/country) and provides a 12 V DC output. Figure 1 shows how the Criterion 60 connects to AC (wall) power. The ON/OFF switch turns monitoring on and off (when the monitor is connected to AC power, the battery can recharge whether the ON/OFF switch is ON or OFF).
WARNINGS
To avoid personal injury or equipment damage, keep the
power supply cable out of the way of users.
Use the Criterion 60 Monitor only with Caradyne power
supply and battery pack.
Figure 2. Connecting the Criterion 60 to AC (wall) power
4
8-100333-00 Rev. B
Connecting the Criterion 60 to the patient system
The Criterion 60 monitors pressure and oxygen:
The monitor measures pressure using disposable
pressure tubing that includes a hydrophobic filter. The filter protects the monitor from moisture, and reduces the risk of cross contamination.
The monitor measures oxygen concentration using a
reusable oxygen sensor. The sensor is designed for use with the Criterion 60 Pressure/Oxygen Monitor only.
The pressure tubing is typically connected close to the patient airway, while the oxygen sensor is placed in the inspiratory limb upstream of the humidifier. Figure 3, Figure 4, and Figure 5 show how the Criterion 60 can connect to a patient system.
WARNING: Do not disassemble the oxygen sensor as it contains harmful caustic substances.
CAUTIONS The sensor will not operate in 100% condensing humidity as the condensing water will block the O2 supply to the sensor. (It will work in 100% non-condensing O2).
The sensor should not be handled unnecessarily during calibration or use. Body heat can cause disproportional changes in the sensor compared to the gas, which may produce errors in the readings.
To avoid monitor damage due to condensation, position the pressure sensor at least 25 cm (10 in.) lower than the monitor.
Criterion 60 Pressure/
(
Oxygen Monitor
back panel)
Oxygen sensor (4-pin) connector
Pressure (luer)
Oxyge n
Oxygen sensor
From positive pressure
22F/22F straight connector (use
Pressure
Tee for pressure
Figure 3. Connecting the Criterion 60 to the patient
system
6
8-100333-00 Rev. B
Figure 4. Connecting the Criterion 60 to a face mask
Figure 5. Connecting the Criterion 60 to a tee or wye and
ET tube
8
8-100333-00 Rev. B
The Criterion 60 front panel
The Criterion 60 front panel (Figure 6) includes the elements summarized in Table 1.
Table 1. Criterion 60 front panel
andkeys
9 (ACCEPT) key Saves a new setting or proceeds to
8 (CLEAR) key Cancels a change that has not yet
Soft keys Select a setting or option displayed
soft key The Menu soft key, which selects the
(ALARM) indicator
(SILENCE) key Silences the audible portion of an
(RESET) key Resets audible and visual alarm
Scroll up and down to adjust values of selected setting.
next calibration step.
been accepted.
below each key.
pressure or oxygen screen for display. Flashes when an active alarm has not yet been reset (whether or not alarm silence is in effect).
alarm for two minutes following the most recent key press. The SILENCE key LED lights when the silence is active and the monitoring screen shows alarm silence time remaining. Also used for alarm presilence feature, which provides two minutes of silence from the most recent key press if pressed before an alarm becomes active.
indicators, cancels alarm silence, and clears peak pressure displays on the screen (measurement and bar graph).
Loading...
+ 30 hidden pages