3. The centrifuge is now locked and it will not be possible to open the lid until the
rotor has stopped and the amber ‘LID OPEN’ neon indicator has come on.
4. The centrifuge may be subject to some vibration initially, while the rotor accelerates
through a critical stage, but this should cease once it has reached a reasonable
speed. However. if the vibration continues, it is most likely that the centrifuge is
unbalanced. If this is the case, the ‘TIMER’ knob must be turned OFF and the
rotor will slow down and come to a stop. Once the amber ‘LID OPEN’ neon
indicator has come on, the lid should be opened and the balance checked.
COMPLETION
1. Once the run has been completed, the timer will switch off and the rotor will
start to decelerate. The red ‘RUN’ neon indicator will go out and the green
‘POWER’ neon indicator will remain on. Once the rotor has stopped, the
amber ‘LID OPEN’ neon indicator will come on, after a short delay, indicating
that the lid may be opened.
2. Remove the sample tubes and close the lid, securing the catch on the front of the
unit. Once the lid has been closed, the power should be turned off at the rear of
the unit.
WARNING - ALWAYS ENSURE THAT THE POWER IS
ON WHEN ATTEMPTING TO CLOSE THE LID
BREAKAGES
1. In the event of tube breakage,the ‘TIMER’ should be turned OFF, by turning it
in an anti-clockwise direction back to the vertical position. Once the amber ‘LID
OPEN’ neon indicator has come on, the lid should be opened and the rotor
should be removed by unscrewing the knurled nut that secures it to the unit
(see
figure 3). The rotor should then be thoroughly cleaned and washed, being careful
not to allow any broken glass to cause injury.
2. After any breakages, the rotor must be completely free of any glass fragments or
other materials. If this is not the case, it will not be possible to successfully balance
the rotor on future operations.
3. Once the rotor has been successfully cleaned and the imbalance rectified, the
centrifuge may be re-loaded with samples and operated in line with the procedures
set out in the preceding sections.
(Figure 3)
Knurled
nut
3