Viscometric cell and exit capillary (cutaway view) ......................................................... 40
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 4
CHAPTER
1
1
Overview
Manual
INTRODUCTION
The CANNON® High-T emperature High-Shear (HTHS) Series II Viscometer is designed to determine the viscosity of engine oils and other
oils under conditions of high shear at high temperatures.
The HTHS Series II is capable of testing at a temperature of 150°C as
specified in ASTM D 5481. For other applications, temperature may be
set as low as 25°C above ambient.
Oil samples are introduced into the viscometric cells and then propelled
through the glass capillary tubes at the test temperature to obtain various
shear rates. The HTHS Series II monitors the test and records the results.
Data can then be analyzed via the computer software and test results can
be displayed and printed.
This manual provides information about:
Components
Functions and operation of the HTHS Series II
HTHS testing method
Basic maintenance and repair
For additional details regarding the High Shear Viscosity Calculator, a
data analysis software
package for W indows
provided with the HTHS
Series II, consult the
software manual.
The HTHS Series II viscometer includes 5 viscometric cells in a
thermostated block with an
internal temperature control
system, a digital pressure
measuring system, automatic timer, vacuum/
pressure controls and waste
receiver.
®
Patents
CANNONCANNON
CANNON
CANNONCANNON
U.S. Patent Numbers:
4,658,636; 4,723,442
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
The CANNON® High Temperature High Shear Series II Viscometer
meets the high shear testing requirements of ASTM D 5481. SAE J300
(DEC99) requires HTHS viscosities from 2.6 to 3.7 mPa•s. Because the
HT75 standard has a nominal viscosity of 2.7 mPa•s and the HT150
standard has a nominal viscosity of of 3.7 mPa•s, CANNON® recommends that the users calculate cell volumes using either of these standards.
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
Please observe the following safety procedures and notices for proper
operation of your HTHS Series II instrument. Deviation from the installation, operation or maintenance procedures described in this manual may
result in a hazardous situation and may void the manufacturer's warranty .
•Make sure that your unit is operated only by qualified personnel
•Make sure that you read and understand all operating instructions
and safety precautions listed in this manual before installing or
operating your unit. If you have questions regarding instrument
operation or documentation, contact CANNON
®
Instrument Com-
pany.
•Transport the unit with care. Sudden jolts or drops may cause damage to components.
•Observe all warning labels.
•Never remove warning labels.
•Never operate damaged or leaking equipment.
•Never operate the equipment with damaged mains power cables.
CANNONCANNON
CANNON
CANNONCANNON
•The instrument power cord should only be connected to a suitable
AC mains power source (with protective earth ground) matching the
specifications of the S/N label on the instrument rear panel.
•Position power cords so that they are not likely to be walked on or
pinched by items placed on or against them. Keep all connections as
neat as possible.
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 7
4
•T o disconnect the power cord, pull it out by the plug. Never pull the
cord itself.
•Always turn off the unit and disconnect the mains cable from the
power source before performing approved service or maintenance
procedures, or before moving the unit.
•Refer all service and repairs to qualified personnel.
•Do not attempt to service the unit beyond the service and/or repair
General Caution
procedures detailed in this manual or other documentation provided
by the manufacturer. Contact CANNON
all additional service/repair needs.
In addition to the cautionary statements listed previously , additional
cautions may be posted throughout this manual. These warnings may be
designated by an appropriate symbol inside an equilateral triangle.
General cautions are indicated with an exclamation point (see diagram,
left). Read and follow these important instructions. Failure to observe
these instructions can result in permanent damage to the unit, significant
property damage, and personal injury .
®
Instrument Company for
Hot Surface Caution
Protective Conductor
WARNING
MAINS
AC Power Input Symbol
( O )
Supply OFF Symbol
Hot surface cautions (see diagram, left) may be attached on or near hot
surfaces of the instrument. Avoid touching these surfaces during instrument operation above 50°C.
The Protective Conductor Terminal symbol is used to indicate required
ground connections for your instrument electrical supply .
When supplying power to this instrument, connect the protective ground
(earth) terminals of the instrument to the protective conductor of the
(supplied) line (MAINS) power cord. The main plug for the power cord
should only be inserted in a socket outlet (receptacle) provided with a
protective ground (earth) contact.
Do not use an extension cord (power
cable) without a protective conductor (grounding).
The ~MAINS symbol indicates instructions or connections for the AC
power supply . The AC Power input must match the electrical specifications listed on the label on the rear panel of the instrument. The supplied
AC Mains power cord must be attached to the connector labelled
~MAINS. This connection serves as a means of disconnect and should be
readily accessible.
The (O) symbol indicates the OFF position for the electrical switches for
your unit (AC Mains or accessories).
CANNONCANNON
CANNON
CANNONCANNON
Do not use this equipment in a manner not specified by the manufacturer.
If you do, the protection provided by the equipment may be impaired,
and you may void the manufacturer warranty .
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 8
Apparatus description
5
WARNING!
Waste-oil receiver
The HTHS has been designed for testing of hot oil samples under high
pressure. Care should be taken to perform sample tests per the procedures described in this manual.
The High-Temperature High-Shear Series II V iscometer consists of a
heating device (an aluminum thermostatic block) with cavities housing
five viscometric cells. The block permits temperature regulation of the
test sample from 30°C to
150°C, ± 0.1°C. Components and accessories
include the waste-oil
receiver, Vacuum Pressure
Unit (built-in), digital timer
mechanism, soak cycle
timing watch, thermal
sensors and safety devices.
The waste-oil receiver (see
photo) consists of a recess
underneath the block with
HTHS with waste oil receiver
apertures for the exit tubes
from the five test cells. A plastic container is placed under the tubes to
catch the oil sample as it is expelled from the viscometer cells. The
container slides into place at the left front of the HTHS Series II. A
hinged door provides protection from hot oil. At the rear, an internal fan
draws vapors into the exhaust manifold.
Cell capillaries
NOTE
Nitrogen supply
CAUTION
CANNONCANNON
CANNON
CANNONCANNON
Each viscometric cell contains a borosilicate-glass capillary sealed with
Viton® gaskets at each end. A capillary support seats the capillary tube in
the cell.
Matching of capillary
lengths from one test
cell to the next within
1 mm will allow the
individual cell
calibrations to be
more closely
matched, making
subsequent runs of identical sample in different cells more repeatable.
Matched capillary sets supplied with the instrument at initial purchase
and are also available for purchase from CANNON®.
The HTHS cell capillary
The user-supplied nitrogen gas cylinders should be fitted with a pressure
regulator to reduce the pressure supplied to the HTHS Series II to no
more than 500 psi. Testing pressures will range from 100 to 500 psi.
Nitrogen is the only gas recommended as a source of pressure for the
HTHS Series II viscometer.
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 9
6
Operation
Oil is introduced into the cells via the syringe-like
charge cylinder(s) (see photo). The oil sample is
filtered through a screen, then warmed to test
temperature in the viscometric cell and finally
propelled under high pressure from the cell through
a narrow exit tube. The oil sample is expelled into
a plastic container in the waste-oil receiver. A
digital timer tracks the duration of the test run.
Safety features:
Protective housing
The HTHS components are protected from damage
by a sturdy metal housing. The housing also
prevents direct operator contact with the heater
block assembly .
Pressure relief
The HTHS ballast tank stores pressurized gas from
the user nitrogen supply for the HTHS. If pressure
in the tank exceeds 700 psi, the internal HTHS
pressure relief rupture disc will vent gas to atmosphere until the gas source is turned off or depleted. A failed rupture disc
must be replaced prior to resuming HTHS operations. See APPENDIX B
for additional information on rupture disc replacement.
Charge cylinder
in Luer fitting
Thermostat
CANNONCANNON
CANNON
CANNONCANNON
HTHS internal rear view including pressure relief rupture disc
assembly
The HTHS Series II is equipped with a thermostat that prevents the
instrument from overheating. If temperature rises to an unacceptable
level (160°C ± 5°C), power will be removed from the HTHS and will not
be restored until the instrument has cooled to a safe state.
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 10
CHAPTER
7
2
Unpacking/inspection
ASSEMBLY
The CANNON® High Temperature-High Shear (HTHS) Series II Viscometer is shipped with only minor assembly required. The five viscometric cells are shipped in place. Several accessories are included and
must be installed prior to HTHS operation.
Follow the directions below to set up the HTHS:
1. Remove all components from the shipping container(s).
2. Remove any and all packing materials (styrofoam, etc.) from the
components.
3. Compare and verify shipped materials by comparing equipment
items with packing/parts list(s). Report missing items to CANNON
Instrument Company immediately .
®
Damaged items
CAUTION
Installation procedure
4. Inspect each component for signs of damage. Report damages to the
shipper and to the CANNON
Retain all packing materials until the instrument is connected and functioning properly. If any component(s) must be returned to CANNON
Instrument Company, the damaged item(s) should be packaged in the
original shipping container. Refer to the final chapter of this manual for
instructions on returning defective equipment. Customers outside the
United States should contact the local CANNON® dealer for procedures
on returning products to CANNON® .
The HTHS is a precision instrument. Great care should be taken in
handling the equipment.
1. Attach the included power cord to the connector marked MAINS on
the lower rear panel of the HTHS . Do NOT plug the power cord into
a power outlet at this time.
2. Place the 2-piece panel collar over the circular opening on top of the
HTHS housing.
®
Instrument Company immediately .
®
CANNONCANNON
CANNON
CANNONCANNON
3. Place the cell cover over the viscometer cells and slide it into place
inside the panel collar.
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 11
8
Connecting the Syringe Hose Adapter
The Syringe Hose Adapter is an adapter plate with a hose connection
which permits introduction of both vacuum and air pressure to the
syringe-like charge cylinders. The adapter is designed to slide over the
upper aperture of the charge cylinder and provide an airtight seal when
the sponge pad on the underside of the adapter is pressed against the lip
of the charge cylinder.
The adapter is shipped with a length of clear
1/16 inch plastic tubing. Follow the procedure below to connect the adapter to the
built-in Vacuum Pressure Unit of the HTHS.
1. Locate the adapter plate and the length
of clear 1/
2. Slide one end of the plastic tube over
the fitting on the adapter plate.
3. Slide the other end of the plastic tube
over the hose fitting on the quickconnect hardware.
16 inch plastic tubing.
4. Push the hardware into the connector
atop the HTHS by pushing downward.
NOTE
To release the hardware, press down on the
knurled portion of the connector.
Connecting the nitrogen supply
Connect the regulated nitrogen supply to the brass intake at the back of
the HTHS unit, using the procedure below:
1. Locate the 10' length of 1/8 inch copper tubing and Swagelok
connectors provided.
2. Secure the tubing to the brass intake at the left rear of the HTHS unit
using the Swagelok
tubing into the fitting, tighten finger-tight and then tighten an additional 3/4 turn with a wrench. The intake is located to the left of the
MAINS power cord connector.
3. Connect the other
end of the copper
tubing to your
nitrogen supply.
Syringe Hose Adapter
installed
®
®
connectors (see photo). To tighten, insert the
CAUTION
®
CANNONCANNON
CANNON
CANNONCANNON
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
The nitrogen supply must be equipped
with a pressure regulator preventing
the release of gas at pressures above 500 psi.
®
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
Instrument Company
Swagelok™ connections
Page 12
Checking the power supply
MAINS
Verifying VPU operation
1. Connect the power cord to an appropriate AC power source (see rear
panel label for power input requirements).
2. T wist the front panel POWER switch to the right to provide power
to the HTHS unit.
The timer and pressure displays will light up.
The exhaust fan will begin operating.
®
The W atlow
temperature.
The instrument will begin heating the block to the target temperature.
Check the operation of the built-in Vacuum Pressure Unit (VPU) as
follows:
1. Turn the front panel PUMP switch ON and toggle the
temperature display will indicate ambient and target
PRESSURE/VACUUM switch to PRESSURE. Y ou should
hear the pump operating.
9
CAUTION
2. Place your finger over the end of the black syringe hose
adapter that is connected to the top of the unit by a clear
plastic tube. You should feel the pressure of the air flow. The
rate of flow is not adjustable.
3. Flip the PRESSURE/VACUUM switch to VACUUM and
check the adapter for a slight suction with your finger.The
VPU has been carefully adjusted at CANNON
®
but it may
be necessary to readjust the force of the vacuum by turning
the ADJUST dial on the front panel of the unit. The amount
of force determines the relative speed at which oil is drawn
into the charge cylinder. You may experiment with an oil
sample to fine-tune the adjustment (see Soaking the sample,
Chapter 5).
T o increase the force of the vacuum, rotate the dial clockwise; to decrease the force of the vacuum, rotate the dial
counterclockwise.
The flow of oil should be regulated in such a way that
the oil is not drawn into the charge cylinder too quickly. If
this happens, the oil may be drawn into the VPU via the
syringe hose adapter, possibly contaminating future
samples and damaging the unit.
CANNONCANNON
CANNON
CANNONCANNON
4. Turn the pump OFF and press the HTHS power switch in to
turn it off.
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 13
10
This page intentionally left blank.
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
Page 14
CHAPTER
11
3
Software features
HIGH SHEAR VISCOSITHIGH SHEAR VISCOSIT
HIGH SHEAR VISCOSIT
HIGH SHEAR VISCOSITHIGH SHEAR VISCOSIT
CC
ALAL
AL
ALAL
CULACULA
CULA
CULACULA
C
CC
The High-Shear Viscosity Calculator is a standalone software package
that calculates viscosity of a sample in mPa·s (cP) using efflux time and
pressure values obtained during testing with the CANNON
T emperature High-Shear Viscometer (HTHS). The Calculator also
provides a routine for ensuring the accuracy of constants used in viscosity calculation.
Designed for use in conjunction with the HTHS, the Calculator incorporates a 15-minute temperature equilibrium timer for samples heating in
the test cells. The Calculator can also recommend a Target Pressure,
based on the user’s best guess for sample viscosity, to achieve the desired
shear rate with the HTHS test—this feature may considerably reduce the
time required to achieve an acceptable shear rate with the HTHS instrument.
®
High-
TT
T
TT
YY
Y
YY
OROR
OR
OROR
Computer
A computer is used to run the software for the High-Temperature HighShear Viscometer, but is not physically attached to the instrument. The
HTHS software program is for data analysis only , and NOT for control-
ling or monitoring HTHS tests.
For more information on using the High Shear Viscosity Calculator with
your HTHS instrument, and for details on the calibration procedure, see
the High Shear Viscosity Calculator Instruction Manual.
The table below provides minimum specifications for successful operation of the High Shear Viscosity Calculator.
®
letnI
®
muitneP
®
swodniWSM
®
swodniW
®
®
TN/0002/89/59
PX/
yromemMARsetybagem23
elbachtiwretnirpelbitapmoc-
)rettebrozHM661(tnelaviuqero
esuomdna)dednemmocerroloc(rotinom,draobyeK
)noitarugifnocmuminim(snoitacificepSretupmoC
NOTE
CANNONCANNON
CANNON
CANNONCANNON
ecapsksiddrahfosetybagem02
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 15
12
If a computer was ordered from CANNON® for the HTHS, the computer will
meet the minimum configuration requirements.
Program Installation
Before installing the High-Shear Viscosity Calculator software, turn on
your computer and make certain that the Windows
performing normally . Close all other software programs before beginning
the installation procedure below:
Checklist
1. Place your High-Shear Viscosity Calculator installation CD-ROM in
your computer’s CD-ROM drive (usually D:).If the installation
program is not initiated automatically , click Start/Run and click theBrowse ... button to select the SETUP program on the CD-ROM
drive from the RUN window . Click Open. Then click OK.
2. Follow the installation program prompts to complete the installation.
The default path for installation is C:\Program Files\CANNONInstrument Company\HTHS.
The setup program will install the High-Shear Viscosity Calculator files
to the drive and directory specified. The program will also create a
shortcut icon on your Desktop.
Initial software configuration
Configuration disk
Your High-Temperature High-Shear Viscometer is shipped with a Configuration floppy disk with initial software calibration information
derived from HTHS factory tests.
®
operating system is
Configuring software
Prior to recording test data with the High-Shear V iscosity Calculator
software, copy the configuration files from the floppy disk to the installation directory for the High-Shear Viscosity Calculator (the default
directory is:
C:\Program Files\Cannon Instrument Company\HTHS).
Then start (or restart) the software to load the configuration data.
For more information on configuring your software to conform to HTHS
characteristics, consult the High Shear Viscosity Calculator InstructionManual.
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 16
CHAPTER
13
4
WARNING!
Checking nitrogen supply
HTHS PRE-OPERATION
CHECKLIST
The HTHS has been designed for testing of hot oil samples under high
pressure. Care should be taken to perform sample tests per the procedures described in this manual.
T o verify that the HTHS is functioning correctly, follow the checklist
below:
1. Turn the HTHS power switch clockwise to turn it on. Ensure that the
regulated nitrogen supply is providing pressurized nitrogen gas to the
HTHS via the rear panel connection (see photo). Listen for signs of
gas leakage.
Checking solenoid valves
®
CANNONCANNON
CANNON
CANNONCANNON
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
HTHS Series II nitrogen connection to internal valve system (panel
removed to provide internal detail)
2. If there is no audible indication of gas leakage, depress and hold the
front panel push button switch labeled INCREASE ( ). This will
actuate the solenoid valve that admits nitrogen from the regulated
nitrogen supply source to the ballast tank. As gas flows into the
ballast tank, the reading on the front panel digital pressure indicator
should steadily increase. Increase pressure to about 100 psi. To stop
the flow of gas, release the INCREASE switch. If the pressure in the
ballast tank rises above the desired level, push and hold the button
solenoid switch labeled DECREASE ( ) to vent gas from the
ballast tank to the atmosphere. Release the button when the desired
pressure has been obtained.
®
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
Instrument Company
Page 17
14
NOTE
CAUTION
There will be a time delay after the switch is released and before the
HTHS digital pressure display stabilizes. This is normal during filling and
venting the ballast tank. You may need to practice working with the valve
settings to establish the desired pressure in a relatively short time.
The internal valve system for the HTHS Series II includes a replaceable
rupture disc as a safety measure to prevent pressure from rising above
700 psi. HTHS pressure settings above 500 psi are not recommended.
3. Ensure that all of the plug valves at the top of the viscometric cells
are closed. When you are done, all valve handles should be in the
horizontal position.
4. Select a test cell using the Cell Selection switch. Then press theStart/Stop button to release nitrogen gas into the selected cell. Listen
for the hiss of nitrogen gas escaping from the exit tube of the viscometric cell. The digital display will indicate a drop in pressure, and
the digital timer will start.
5. Press the Start/Stop button again to stop the flow of nitrogen.
6. Repeat steps 4 and 5, changing the Cell Selection switch to each of
the other four cells to ensure that the valve system is working properly. Note that the digital timer will reset for each test.
7. Check the W atlow
®
temperature display. The heater for the aluminum
thermostatic block is energized when the power switch is turned on. The
display indicates current and target block temperatures. After the power
has been on for a while, the temperature indicator should show that the
temperature of the block is rising.
8. Use the
and buttons on the Watlow
®
temperature meter to
select the target temperature (normally 150°C). Allow the temperature to stabilize at that setting. The temperature can be set for any
value up to 150°C. It is necessary for the HTHS to stabilize temperature
at the desired temperature before proceeding with actual viscosity tests.
This may require more than two hours.
NOTE
NOTE
CANNONCANNON
CANNON
CANNONCANNON
For temperature meter calibration and other meter options unrelated to
HTHS testing, consult the Watlow® meter Operation Manual.
9. Insert a glass charge cylinder onto each of the five Luer fittings on
the top of the viscometric cells and rotate the cylinder clockwise to
lock it in place.
The HTHS will not begin a test unless all plug valves are in the closed
(horizontal) position).
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 18
CHAPTER
15
5
Preliminary checklist
STANDARD HTHS TEST
PROCEDURE
The procedure below should be followed whenever taking measurements
with the HTHS Viscometer . Before meaningful data can be obtained,
however, each of the viscometric cells must be calibrated with the
standards provided (see CALIBRATION chapter). Before calibrating the
HTHS, complete several test measurements with a readily-available oil to
become familiar with the operation of the unit. Frequency of recalibration
is at the user's discretion. Always recalibrate if changing the test temperature (usually 150°C).
Follow the preliminary checklist below to prepare a sample run for the
HTHS:
1. Open the flow valve on your pressurized nitrogen supply to the
HTHS (an internal ballast tank provides a pressure reservoir for
efficient HTHS operation).
CAUTION
2. Make sure that the waste oil container is in its position at the base of
the HTHS under the viscometer cell exit tubes. Cap the HTHS exit
tubes to prevent oil leakage during the preliminary part of the test.
Satisfactory stoppers are available from CANNON
N98; package of 12).
When removing and replacing stoppers, always use a protective
glove rated for protection at 150°C to prevent burns.
3. Attach the syringe tube (charge cylinder), which will contain
the oil sample, to the Luer fitting on the desired viscometer
cell holder at the top of the HTHS.
4. Turn the Cell Selection switch on the HTHS front panel to
the number of the desired test cell. Pour approximately 5 ml
of sample into the charge cylinder. This quantity of oil
sample will be used to flush any oil remaining from the
previous HTHS sample run.
5. Open the plug valve for the desired viscometer cell by
turning the green plug valve handle counterclockwise to a
vertical position.
®
(Catalog #9726-
Plug valve open with Syringe
Hose Adapter installed
®
CANNONCANNON
CANNON
CANNONCANNON
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
6. Place the Syringe Hose Adapter over the top aperture of the
charge cylinder and apply pressure to seal the sponge pad
against the charge cylinder opening (see photo).
®
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
Instrument Company
Page 19
16
7. Locate the 2-way toggle switch for the Pressure/Vacuum pump on
the left side of the HTHS front panel and flip the toggle upward to
PRES(sure). Then turn the PUMP switch ON to force the rinse oil
sample into the viscometric cell.
8. After the oil has been forced into the cell, close the green plug valve
by moving it to the horizontal position.
NOTE
DO NOT PROCEED TO THE NEXT STEP OF THE RUN PROCEDURE
IF THE PLUG VALVE FOR THE TEST CELL IS IN THE OPEN
(HANDLE VERTICAL) POSITION. The HTHS Series II test cycle will not
begin unless the plug valve is closed.
9. Remove the adapter from the charge cylinder .
10. Using the toggle switch, turn the Pressure/Vacuum pump OFF.
11. Using the DECREASE and INCREASE push buttons, adjust the
12. Using a protective glove, carefully remove the stopper from the
13. After the test has been completed, toggle the START/STOP button
14. Using protective gloves, replace the exit tube stopper. You are now
Soaking the test sample
ballast tank pressure to between 100 and 150 psi.
HTHS exit tube for the desired cell. Then press the START/STOP
button to force the oil through the capillary. The sample will be
flushed through the cell. An audible hiss indicates that the oil has
been discharged and gas is flowing through the capillary.
on and off for a few seconds to remove as much excess oil from the
cell as possible.
ready to prepare the sample for HTHS testing.
CANNONCANNON
CANNON
CANNONCANNON
1. To integrate testing with the software analysis package, load the
High-Shear Viscosity Calculator software and select the desired test
cell by clicking the graphic representation or typing the cell number
in the appropriate field. You can also type the Cell Number in the
appropriate field.
2. Fill the charge cylinder with the test sample to approximately the 10
to 12 ml level. Verify that the Cell Selection switch is turned to the
number of the desired test cell.
3. Open the plug valve by turning the green plug valve handle counterclockwise to the vertical position.
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 20
4. Place the Syringe Hose Adapter over the top aperture of the charge
cylinder and apply pressure to seal the sponge pad against the top of
the charge cylinder (see photo).
5. Ensure that the Pressure/Vacuum switch
on the left side of the HTHS front panel is
set to Pressure. Then turn the Pump
switch ON to introduce sample into the
viscometric cell.
6. Close the plug valve.
7. Let the HTHS oil sample soak in the heated
viscometric cell for at least 15 minutes to
achieve test temperature. You can use the
Equilibrium Timer built into the HighShear V iscosity Calculator. From the
Introducing sample
interface window , just click on the Equilibrium Timer button to start and stop the timer . Enter initial test data
(Cell Number and Sample ID Name) in the High-Shear Viscosity
Calculator software program during this time.
17
NOTE
Setting sample volume
8. Check the cell temperature on the front panel display and verify with
your temperature reference to ensure that the HTHS is controlling at
the desired test temperature.
9. (optional) Input the estimated sample viscosity in the field provided
by the High-Shear Viscosity Calculator. The Calculator will display
the optimum gas pressure for achieving the desired shear rate.
10. Adjust the gas pressure to the level required for test (See ASTM D5481 sections 9 and 11 for more information). To increase the gas
pressure, press and hold the Increase button. To decrease pressure,
press and hold the Decrease button. Monitor the current gas pressure reading using the HTHS digital pressure display .
Set the pressure 1.5 to 2 lbs. higher than the desired optimum test
pressure, since the pressure will fall slightly during testing.
11. Wait for the pressure reading to stabilize before proceeding to the
next steps.
Before the HTHS test can be run successfully, the volume of oil in the
viscometer tube must be precisely regulated. This is accomplished by
withdrawing excess sample from the cell immediately prior to the test
run.
CANNONCANNON
CANNON
CANNONCANNON
1. Before adjusting the volume of oil in the cell, recheck the pressure to
ensure that it is still at the desired value. If necessary , adjust the
pressure before proceeding to the next step.
2. Verify that the Cell Selection switch is turned to the number of the
desired test cell. Then open the plug valve for the desired cell by
turning the green plug valve handle to its vertical position.
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 21
18
3. T oggle the Pressure/Vacuum switch to the Vacuum position and
again place the adapter on top of the charge cylinder. Turn the Pump
switch ON to withdraw excess sample from the viscometric cell into
the charge cylinder. If necessary, the intensity of vacuum can be
controlled by using the Adjust dial on the front panel. When nitrogen
gas bubbles appear consistently in the charge cylinder, the sample
has been adjusted to the proper volume.
NOTES
NOTE
Testing sample(s)
At least 0.5 ml of excess oil sample should be drawn into the charge
cylinder. If bubbles appear consistently before this volume has been
attained, not enough oil was used when filling the cell. You will need to
introduce additional sample and begin the soak procedure again.
Empirical testing has suggested that variability in the intensity of vacuum
used to withdraw excess sample from the HTHS test cell prior to testing
may alter the sample volume in the cell. For better determinability and/or
repeatability, avoid adjusting the valve setting from test to test.
4. Immediately turn the pump switch OFF and turn the plug valve to
the CLOSED (horizontal) position. Remove the adapter from the top
of the charge cylinder.
The HTHS will not begin a test unless all plug valves are in the closed
(horizontal) position).
5. Verify cell temperature by checking the front panel temperature
display.
6. Recheck the gas pressure reading and make last-minute adjustments
if necessary , using the INCREASE and DECREASE buttons.
NOTE
CAUTION
®
CANNONCANNON
CANNON
CANNONCANNON
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
The HTHS will not begin a test unless all plug valves are in the closed
(horizontal) position).
1. Verify that the Cell Selection switch is set to the number of the
desired test cell. Using a protective glove, carefully remove the
stopper from the viscometer cell exit tube. Then close the hinged
door and begin the test immediately by pressing the Start/Stop
button.
Make certain to remove the cap from the exit tube before initiating the
sample run. Make certain that the plastic waste container is in position
under the exit tubes.
2. Precisely 10 seconds into the timed sample test, check the current
pressure from the digital pressure display and write it down.
3. Wait for the test run to be completed. The end of the sample flow
period will be indicated by an audible hiss as nitrogen gas escapes
through the capillary tube. The internal timer gauge will trip, stopping the timer. The HTHS should automatically stop the flow of
pressurized gas to the test cell.
®
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
Instrument Company
Page 22
4. Write down the ef flux time for the sample run, as indicated on the
HTHS digital timer.
5. Use the High-Shear Viscosity Calculator to enter the values for
Pressure and Efflux T ime from the test run. Then click the Calculate Viscosity … button. The Calculator will automatically deter-
mine the shear rate from the efflux time and stored cell volume data.
Whenever an acceptable shear rate is obtained, the viscosity data will
be displayed in tabular format in the Calculator display window .
6. If the shear rate is not within the user-determined minimum and
maximum values for the cell (you can click Calibration/Cell Data to
view parameters for each cell), the software will automatically
calculate and display a new recommended value for pressure. Readjust the HTHS instrument to the new test pressure, and repeat the test
until the desired shear rate is obtained.
7. Repeat the test procedure for each cell.
19
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 23
20
This page intentionally left blank.
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
Page 24
CHAPTER
21
6
DAILY MAINTENANCE
Because the HTHS is performing high temperature runs with a variety of
oils, it is possible for the capillary drain tubes to become clogged,
particularly if high viscosity oil remains in the capillary for extended
periods of time.
T o prevent this occurrence, CANNON
flushed at the conclusion of testing each day with approximately 5 ml of
a low viscosity neutral oil (2 mPa·s at 150°C) or commercially-available
Stoddard solvent or Naptha.
®
recommends that all five cells be
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 25
22
This page intentionally left blank.
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
Page 26
CHAPTER
23
7
CANNON standards
CALIBRATION
Occasionally it may be necessary to recalculate the mathematical constants used by the High Shear Viscosity Calculator to calculate viscosity .
A test protocol using CANNON-manufactured oil standards provides the
data necessary to derive the constant values. We will refer to this protocol
as a calibration. The procedure is described in detail in the High ShearViscosity Calculator for Windows® Instruction Manual.
CANNON Instrument Company produces a series of Newtonian viscosity standards designed specifically for calibration of the High Temperature High Shear Viscometers such as the CANNON® High-T emperature
High Shear Viscometer. They work equally well calibrating other instruments at this temperature. These standards cover the nominal range of
2.0 to 7.0 mPa•s at 150°C.
Viscosities for all standards sold by CANNON® are based on the National Institute of Standards and Technology (NIST) value of 1.0016
mPa•s(centipoise) for water at 20°C (68°F). We have been granted ISO
9002 registration for the manufacture and certification of viscosity
standards. Standards are available in 120, 470, and 940 mL bottles as
well as 3.8 and 19 L containers. These sizes correspond to 4 ounce, pint,
quart, gallon and 5-gallon containers.
Capillary specifications
NOTES
Each calibration oil should be run in each cell following the normal test
procedure. For assistance in determination of appropriate test pressures
for each cell, consult the High Shear V iscosity Calculator for WindowsInstruction Manual.
The Cell Calibration Record on page 35 of this manual contains capillary
specifications necessary to complete the calibration. The Record should
be updated by the user whenever a viscometer cell or capillary is replaced. Input current data from the Cell Calibration Record into the
HTHS Calculator per instructions in the software manual.
As per ASTM D 5481, samples are to be run with a shear rate of
1,400,000 reciprocal seconds (1.40 x 106 sec-1). Other shear rates are
obtainable.
Empirical testing has suggested that the volume of sample residue left
on the cell walls following the HTHS test will vary as a function of sample
viscosity. SAE J300 requires HTHS viscosities from 2.6 to 3.7 mPa•s.
Because the HT75 standard has a nominal viscosity of 2.7 mPa•s and
the HT150 standard has a nominal viscosity of of 3.7 mPa•s, CANNON
recommends that the users initially calculate cell volumes using either of
these standards. It is not necessary to recalculate the volume unless the
standards change or the viscometric cell(s) or capillaries are replaced.
®
®
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 27
24
This page intentionally left blank.
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
Page 28
CHAPTER
25
8
MAINTENANCE & REPAIR
Flushing viscometer cells
CANNON® recommends that all five cells be flushed at the conclusion
of testing each day with approximately 5 ml of a low viscosity neutral oil
(2 mPa·s at 150°C) or commercially-available Stoddard solvent. This is
especially important if the HTHS viscometer is maintained at temperature for an extended period without running sample, as many automotive
engine oils will carbonize if left in the cell for long periods at high
temperatures.
Clearing a blocked capillary tube
Erratic HTHS measurements may suggest a blocked or
partially blocked capillary tube. If the capillary
becomes blocked or partially blocked, it will usually
need to be removed for cleaning. A solvent flush
should be attempted first if the symptoms suggest
partial blockage.
Cleaning the capillary
Preliminary disassembly
If the flush fails to remedy the blockage, follow the
checklist below to clean the capillary . You will need
two wrenches, a small screwdriver and a narrow gauge
wire (less than 0.15 mm) to perform portions of the
operation. You will also need two replacement gaskets
and a #400 mesh filter screen (available from CAN-
NON®—see HTHS replacement parts list, Chapter 9):
Prior to cleaning, follow the instructions in APPENDIX A for removal and disassembly of the viscometric
cell.
After removing the capillary , remove and discard the
gasket and filter screen from the capillary support. The
screen and gasket (see photo) may remain in the cell
even after the capillary has been removed. If so, you
will need to use a small tool similar to a dental pick to
remove these items from the interior of the viscometric cell.
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 29
26
Removing obstructions
1. Pass the narrow gauge wire through the capillary several times to
break through any obstructions.
2. Clean the capillary with a suitable solvent, or chromic acid cleaning
solution if necessary , to remove the contamination. Then repeatedly
flush the capillary with distilled water to remove the cleaning agent.
Follow the instructions in APPENDIX A to reassemble and replace the
viscometric cell.
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 30
CHAPTER
27
9
Part NumberDescription
9728-C60KIN VIS THERMOMETER 150C FOR HTHS
P25.4030POWER SUPPLY +/-15/5V .3/.2/2A
P41.2057PUMP VACUUM 115V 50/60HZ
P41.39MUFFLER/FILTER VACUUM PUMP
P61.216FILTER, 60-MICRON,B-2F-60
P61.3030POWER SUPPLY 24VDC 1.1A UNI IN
P80.1111WASTE OIL RECEIVER PLASTIC
P80.1190VALVE NEEDLE
P80.1195FIL TER AIR-LINE MODIFIED
P80.1231TRANSDUCER, PRESSURE SENSOR
P80.1400SCREEN N400 MESH L\FILTER
P80.1603VALVE SHUT-OFF
P80.1604FITTING SYRINGE
P80.1605VISCOMETRIC CELL
P80.1607EXTENSION TUBE BOTTOM
P80.1608FILTER ADAPTOR BRASS
P80.1609SCREEN #180 MESH PRECAP
P80.1615FILL TUBE (7.0ml NOMINAL)
P80.1650GLASS CAPILLARY
P80.1660SYRINGE GLASS
P80.1950ADAPTOR, SYRINGE HOSE
P80.1970TUBING 1/16 CLR
P80.1980TIMER
P80.2210RS232 AT MODEM CABLE
P80.4007DOOR, LEXAN HTHSG2
P80.4009TOP COLLAR
P80.4010COVER, TOP
P80.4015VALVE, 3 WAY SOLENOID
P80.4016VALVE, 7 WAY SWITCHING
P80.4017PANEL, SIDE TRIM
P80.4019ADAPTOR, SYRINGE
P80.4020VALVE 4 WAY TOGGLE MTV-5
P80.4021ASSY, PROBE SENSOR HTHS2
P80.4022FITTING, QUICK DISCON 1/16
P80.4023FITTING, QUICK DISCON INSERT
P80.4024BUMPER BLACK -95 SHORE A
P80.4026PROBE ASSY , RTD HTHS
P80.4029PID CONTROLLER 24VDC
P80.4031RELAY, SOLID STATE 5A 60VDC
P80.4033HEATER, 175W .25D
P80.4034FUSE M250V X 2.5A 1/4 X 1 1/4
P80.4039VALVE MODIFIED FOR HOLDER
P80.4041VISCOMETRIC CELL-UNCALIBRATED
P80.4042SWITCH RSD 2100-1090-002
P80.4043MAGNET RSD 2100-4902-000
P80.4048RUPTURE DISC 700 PSI
P80.4049FITTING, MALE CONN B-200-1-4
SPARE PARTS LIST
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 31
28
Part NumberDescription
P80.4050SWITCH, TOGGLE SPDT
P80.4051SWITCH, PUSH BUTTON
P80.4052SWITCH, MOMENTARY INC/DEC
P80.4053CCA, VALVE JUNCTION
P80.4055CCA, POWER CONVERTOR
P80.4058IC CMOS MICRO 8 BIT PROGRAMMED
P80.4059FITTING, NIPPLE 1/8NPT X 3.0L
P80.4062FAN, MUFFIN 4.5 IN
P80.4063LABEL, START/STOP PHOTO NEG
P80.4064LABEL, INC/DEC SW INSERT
P80.4066CCA, TRIGGER HTHS2
P80.4069FITTING, ELBOW BRASS 1/8NPT FE
P80.4070REDUCER 3/16TUBE X 1/8TUBE
P80.4073GASKET, VITON FOR HTHS CAPILLARY
P80.4073GASKET HTHS VISCOCELL
P80.4077TRANSFORMER 600VA
P80.4079CCA CLOCK/COUNTER HTHS II
P80.4085PRESSURE TRANSDUCER MODIFIED
P80.4990OPERATION MANUAL HTHS II
P80.4995MANUAL HIGH SHEAR VISC CALC
P80.4996TEST HIGH SHEAR VISCO CALC
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 32
CHAPTER
29
10
Products limited warranty
In addition to other manufacturers’ warranties, CANNON® Instrument
Company (“the Company”) warrants all products (other than reagents and
chemicals) delivered to and retained by their original purchasers to be free
from defect in material and workmanship for one year from the date of
the Company’s invoice to the purchaser. For a period of one year from the
date of such invoice, the Company will correct, either by repair or replacement at the Company’s sole election, any defect in material or
workmanship (not including defects due to misuse, abuse, abnormal
conditions or operation, accident or acts of God, or to service or modification of the product without prior authorization of the Company) without
charge for parts and labor. The determination of whether any product has
been subject to misuse or abuse will be made solely by the Company .
WARRANTY/RETURN
INFORMATION
The Company shall not be liable for any special, incidental, or consequential damages, or any damage to plant, personnel, equipment or
products, directly or indirectly resulting from the use or misuse of any
product sold by the Company except as set forth in and limited by the
foregoing warranties. Representations and warranties made by any
person, including dealers and representatives of the Company , which are
inconsistent, in conflict with, or in excess of the terms of this warranty
shall not be binding upon the Company unless placed in writing and
approved by an officer of the Company.
Reagent and chemical warranty
CANNON® Instrument Company (“the Company”) warrants all reagents
and chemicals sold by the Company and delivered to and retained by their
original purchasers to conform to the weight, specifications and standards
stated on the package. The Company will, at its sole option, either replace
or refund the price (net of freight, handling charges and taxes), of any
reagent or chemical sold by the Company which does not conform to such
weight, specifications and standards upon the prompt return of the unused
portion. Except for replacement or refund of the net price, the Company
shall not be liable for any damages occurring as a consequence of the
failure of any reagent or chemical sold by the Company to conform to the
weight, specifications and standards stated on the package.
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 33
30
Returning a product to CANNON
Procedure
Before returning a CANNON® product for repair or service, make every
attempt to identify the problem. If, after careful checking, the problem
remains unidentified or unsolved, telephone CANNON
Company (or the local service agent) to consult with a product specialist.
If the specialist cannot recommend a simple solution or repair, CAN-
®
will authorize the return of the product through the issuance of a
NON
Return Authorization number (RA).
CANNON
CANNON® Fax Number814-353-8007
Products returned to CANNON
to the following address:
CANNON Instrument Company
ATTN: Return Authorization # __________
2139 High Tech Road
State College, PA 16803 USA
Please include the following:
®
®
Instrument
®
Telephone Number814-353-8000
®
must be carefully packed. Ship prepaid
Required information
Hazardous materials
Shipping notification
• The Return Authorization number (RA).
• The name and telephone number of the person at your company to
contact regarding the product.
• Shipping and billing instructions for the return of the product to your
location.
• A detailed explanation of the reason for the return.
If the product is not covered by warranty , the customer will be provided
with an estimate of the repair costs and asked for approval before any
repairs are made. The customer will be required to issue a purchase order
for the cost of the repairs.
Stringent government regulations restrict the shipment of mercury. Please
contact CANNON® before returning a product that could possibly
contain mercury.
Products returned without prior notification (by either telephone or fax),
or without Cannon’s authorization, will not be accepted.
The customer may be billed a testing fee if a product is returned to
CANNON® and found to be working properly.
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 34
CHAPTER
31
A
APPENDIX A—HTHS CAPILLARY
Replacing an HTHS capillary
It may occasionally be necessary to replace a capillary for one or more of
the HTHS test cells. The procedure below describes the steps necessary
to accomplish this task.
Parts/Tools required:
You will need wrenches (sizes 7/16", 9/16" and 1/2") and a small screwdriver to perform portions of this operation. You will also need two
replacement gaskets (P80.4073) and a #400 mesh filter screen (P80.1401).
CAUTION
Do not attempt the capillary replacement procedure unless the instrument has been turned off for several hours to allow cells to cool to a
temperature of 50°C or less.
REPLACEMENT
Removing the viscometric cell
CAUTION
CAUTION
Turn off or disconnect the main gas supply to the instrument before
continuing with this procedure!
1. Make sure that the HTHS nitrogen gas pressure is near 0 and that the
gas line connections to the HTHS are closed. If the cells are pressurized,
press the DEC button on the HTHS front panel to vent pressurized gas
to the atmosphere. If the instrument has been in operation, turn it off
and allow several hours for the test cells to cool to 50°C or less.
Unplug the power cord.
2. Remove the glass charge cylinder(s) from the top of the cell.
3. Remove the cell cover. THen lift and remove the two-piece collar on
top of the HTHS.
4. Remove the 12 screws securing the top panel to the HTHS frame.
Remove the top panel and set it aside.
Do not touch internal HTHS components unless specifically instructed to
do so in these instructions.
CANNONCANNON
CANNON
CANNONCANNON
5. Using a 7/16" wrench, disconnect the Swagelok™ fitting to the
copper gas line connection for the cell requiring capillary replacement (see photo, next page).
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 35
32
Disconnecting cell gas line connection
6. Trace the electrical wiring from the cell to the valve junction connector and detach the connector from the corresponding numbered
junction. Carefully lift the cell(s) out of the thermostating chamber.
Disassembling the cell
CAUTION
Removing the capillary
7. Orient the cell horizontally and hold the cell with
a wrench, using the flats on the brass fitting at the
base of the cell.
8. Use a second wrench to loosen the hex nut attached to the exit tube at the base of the cell and
remove this fitting by turning the hex nut counterclockwise.
9. Remove the exit tube assembly from the cell.
Do not attempt to remove the fill tube and plug
valve assembly from the opposite end of the
viscometric cell. Their removal may cause a
change in volume when reassembled.
10. Maintaining the horizontal position of the cell,
remove the gasket* from the opening at the base
of the cell, being careful that the glass capillary
does not fall out of the cell. Discard the used
gasket.
CANNONCANNON
CANNON
CANNONCANNON
11. Carefully orient the cell in a vertical position over
a soft cloth or paper and allow the glass capillary
to slide out of the tube onto the cloth/paper. If the
capillary does not drop out readily, tap the base of
the cell gently to jar the capillary loose. If the
capillary cannot be dislodged by this procedure,
try heating the base of the cell with a laboratory
hot plate or heat gun.
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 36
Reassembling the cell
33
12. After removing the capillary, remove and discard the gasket and filter
screen (see photo). The gasket and screen may remain in the test cell
even after the capillary has been removed. If so, you will need to use
a small screwdriver or other tool similar to a dental pick to remove
these items from the interior of the viscometric cell.
Follow the procedure below to reassemble the viscometric cell:
Reassembly sequence (left to right)
13. Orient the viscometric cell vertically with the plug valve assembly
down.
14. Place a new filter screen in the cell and gently tamp into place with
the tip of the exit tube assembly. Be careful not to deform the screen.
15. Place a new gasket in the cell and make sure it is seated against the
screen.
16. Insert the capillary into the cell. The capillary is bilaterally symmetrical; either end may be inserted first.
17. Replace the gasket* at the bottom end of the capillary and replace the
exit tube fitting. Rotate the fitting until finger tight and then turn
approximately one full turn with a wrench to seal the viscometric
cell.
18. Carefully orient the assembled cell with the exit tube pointing down;
replace the assembly in the HTHS thermostating chamber.
19. Reconnect the Swagelok™ fitting to the copper gas line connection
for the cell. Tighten the fitting by hand; then secure with a 1/8" turn
of a wrench.
20. Replace the two-piece collar and cell cover.
21. Replace the glass charge cylinder(s) at the top of the cell and plug in
the power cord.
CANNONCANNON
CANNON
CANNONCANNON
22. Reconnect or turn on the main gas supply to the instrument.
23. Turn on the HTHS and increase the gas pressure to approximately
50-100 psi.
24. Set the selector switch for the correct cell and press the START/STOP switch.
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 37
34
25. Verify that gas can be heard dischar ging from the exit tube (a hissing
sound should be clearly audible).
26. When you have verified that the new capillary is free of blockage,
press the ST ART/STOP switch again to stop the flow of gas. You
are now ready to calibrate the cell and resume normal HTHS operations.
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 38
High-Temperature High-Shear Rheometer
Cell Calibration Record
#laireStnemurtsnI
:naicinhceTSHTH:etaD
35
#lleC
)5-1(
lleC
emulov
yrallipaC
htgnelyrallipaC
retemaid
)lm()mm()mm(
1
2
3
4
5
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 39
36
This page intentionally left blank.
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
Page 40
CHAPTER
37
B
RUPTURE DISC REPLACEMENT
Replacing an HTHS rupture disc
WARNING
CAUTION
There is no guarantee of rupture disc life. Disc life span is affected by
corrosion, creep and fatigue, and physical damage. These conditions will
derate the rupture disc to a lower set pressure. The user should be
prepared to handle a premature failure of the rupture disc. Do not allow
any buildup or solidification of media to occur on a rupture disc. This may
increase the pressure setting of the rupture disc.
The ½” STANDARD rupture disc is a precision instrument and must be
handled with extreme care. Rupture discs should be installed only by
qualified personnel familiar with rupture discs and proper piping practices.
APPENDIX B—HTHS
NOTE
Installation
Do not install rupture disc if there is any damage in the dome area. A
damaged rupture disc is any rupture disc with visible nicks or dents in the
dome.
The rupture disc manufacturer does not recommend reinstalling a
rupture disc that has been removed from the holder. Always reinstall the
entire rupture disc assembly, as reinstallation of the disc alone may
adversely affect the joint sealing capabilities and/or performance of the
rupture disc.
The angular seat rupture disc design is rated for up to 700+ psig.
The rupture disc for the HTHS Series II instrument is sold and installed
as an assembly that includes the rupture disc itself, vacuum support and
holddown ring, conical threaded outlet and inlet connector. The rupture
disc assembly is connected inline with the HTHS pressure system via a
threaded connection on the inlet hex (see images, next page).
The SCREW TYPE rupture disc assembly is permanently stamped with a
maximum burst pressure rating. Ensure that the unit being installed is
compatible with current operating conditions. Install the unit "upside
down" with the outlet hex toward the floor of the compartment.In the
event of a rupture disc failure, this will direct the gas pressure away from
more sensitive internal components.
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 41
38
Rupture disc installed inline
1. Before placing the assembly into the system, turn off the gas supply.
Press the DECREASE button until the gas pressure is close to 0.
2. Remove power from the instrument.
3. Remove the 12 screws securing the top panel to the HTHS frame.
Remove the top panel and set it aside.
4. Clean the mating threads of all grit, dirt, or other foreign
material to ensure a proper seal.
5. Make certain that the flow
arrow stamped on the inlet
body points in the proper
system flow direction.
6. Apply PTFE tape or
other appropriate
sealer to the threads.
7. Place wrench on the
thicker INLET HEX
(see figure) and tighten with a
torque wrench to 40 Ft·Lbs (54 N·M).
CAUTION
Do NOT use the narrower OUTLET HEX for tightening the assembly into
the system, as you may damage or unseat the rupture disc.
Preventive maintenance
Replace the rupture disc every year under normal conditions. More
frequent changes may be necessary due to corrosion, fatigue, temperature
or other adverse conditions. These factors should be evaluated through
actual service experience.
T o avoid operation downtime, a spare rupture disc should remain in stock
at all times for each holder in use. The number of spares required will be
determined by actual service conditions.
CAUTION
If the rupture disc is not replaced periodically when exposed to adverse
conditions, premature failure of the rupture disc may occur, thereby
discharging process media.
Once the integrity of the rupture disc has been compromised, gas will
leak from the line until the assembly is replaced.
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 42
CHAPTER
39
C
Plug Valve construction
APPENDIX C–FIGURES &
DIAGRAMS
CANNONCANNON
CANNON
CANNONCANNON
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Page 43
40
Viscometric cell and exit capillary (cutaway view)
(variable)
CANNONCANNON
CANNON
CANNONCANNON
(variable)
(variable)
(variable)
®
High-Temperature High-Shear Viscometer Series II Instruction & Operation Manual
Revision 1.0d—June, 2011;
2139 High Tech Road • State College, PA 16803 • USA
CANNONCANNON
CANNON
CANNONCANNON
®
Instrument Company
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.