TGA100A Trace Gas Analyzer Overview
The TGA100A Trace Gas Analyzer measures trace gas concentration in an air
sample using tunable diode laser absorption spectroscopy (TDLAS). This
technique provides high sensitivity, speed, and selectivity. The TGA100A is a
rugged, portable instrument designed for use in the field. Common applications
include gradient or eddy covariance flux measurements of methane or nitrous
oxide and isotope ratio measurements of carbon dioxide or water vapor. The
TGA100A concentration measurements can be recorded with a Campbell
Scientific datalogger, output as analog voltages, or sent to a computer through
an Ethernet connection.
OV-1
TGA100A Trace Gas Analyzer Overview
OV1. System Components
Figure OV1-1 illustrates the main system components needed to operate the
TGA100A. These system components include:
• TGA100A Analyzer: The analyzer optics and electronics, mounted in an
insulated fiberglass enclosure.
• Computer: A user-supplied computer to display data and set parameters.
• Optional Datalogger (CR5000 shown): Receives concentration data from
the TGA100A through the SDM interface cable.
• Sample Intake (15838 shown): Filters the air sample and sets its flow rate.
• Sample pump (RB0021-L shown): Pulls the air sample and reference gas
through the analyzer at low pressure.
• Suction hose (7123 shown): Connects the analyzer to the sample pump.
Supplied with RB0021 sample pump.
• Reference gas: tank of reference gas, with pressure regulator (supplied by
user).
Computer
Sample Intake
• Reference gas connection (15837 shown): Flow meter, needle valve, and
tubing to connect the reference gas to the analyzer.
Datalogger
20
19
18
17
1516
1314
1112
910
78
6
10
5
34
9
12
8
7
6
SE
5
HL
4
HL
3
HL
2
HL
1
L
H
HL
HL
HL
HL
HL
CR5000 MICROLOGGER
DIFF
CS I/O
40
39
38
37
36
35
34
33
32
31
2930
28
27
26
25
2324
2122
SE
20
19
18
17
16
15
14
13
12
L
11
H
L
H
L
H
HL
L
H
L
H
L
H
HL
HL
HL
DIFF
CONTROL I/O
VX1
VX3
VX2
VX4
CAO1
CAO2
IX1
IX2
IX3
IX4
IXR
P1
P1
C1C2C3
C4
RS-232
COMPUTER
(OPTICALLY ISOLATED)
G
C6
C5
C7
>2.0V
G
C8
G
<0.8V
5V
5V
G
SDI-12
12V
G
SDM-C1
SDM-C2
SDM-C3
G
G
12V
SW-12
SW-12
G
CAUTION
G
12V
DC ONLY
CONTROL I/O
POWER
POWER OUT
G
12V
GROUND
UP
POWER IN
LUG
11 - 16 VDC
pc card
status
Logan, Utah
'
,
_
A B C
D E F
Hm
CURSOR
PgUp
1
2
3
ALPHA
G H I
J K L
M N O
Graph/
char
4
SHIFT
5
6
P R S
T U V
W X Y
Spc Cap
End
PgDn
7
BKSPC
8
9
- + (
* / )
< = >
$ Q Z
Del
Ins
ESC
ENTER
0
SN:
MADE IN USA
TGA100A Analyzer
SDM Cable
Ethernet Cable
Reference Gas Connection
Suction Hose
OV-2
Reference Gas
Sample Pump
FIGURE OV1-1. TGA100A System Components
OV2. Theory of Operation
OV2.1 Optical System
The TGA100A optical system is shown schematically in Figure OV2.1-1. The
optical source is a lead-salt tunable diode laser that operates between 80 and
140 K, depending on the individual laser. Two options are available to mount
and cool the laser: the TGA100A LN2 Laser Dewar and the TGA100A Laser
Cryocooler System. Both options include a laser mount that can accommodate
one or two lasers (up to four lasers can be installed by adding the optional
second laser mount). The LN2 Laser Dewar mounts inside the analyzer
enclosure. It holds 10.4 liters of liquid nitrogen, and must be refilled twice per
week. The Laser Cryocooler System uses a closed-cycle refrigeration system to
cool the laser without liquid nitrogen. It includes a vacuum housing mounted
inside the analyzer enclosure, an AC-powered compressor mounted outside the
enclosure, and 3.1 m (10 ft) flexible gas transfer lines.
The laser is simultaneously temperature and current controlled to produce a
linear wavelength scan centered on a selected absorption line of the trace gas.
The IR radiation from the laser is collimated and passed through a 1.5 m
sample cell, where it is absorbed proportional to the concentration of the target
gas. A beam splitter directs most of the energy through a focusing lens and
short sample cell to the sample detector, and reflects a portion of the beam
through a second focusing lens and a short reference cell to the reference
detector. A prepared reference gas having a known concentration of the target
gas flows through the reference cell. The reference signal provides a template
for the spectral shape of the absorption line, allowing the concentration to be
derived independent of the temperature or pressure of the sample gas or the
spectral positions of the scan samples. The reference signal also provides
feedback for a digital control algorithm to maintain the center of the spectral
scan at the center of the absorption line. The simple optical design avoids the
alignment problems associated with multiple-path absorption cells. The
number of reflective surfaces is minimized to reduce errors caused by FabryPerot interference.
Reference
Detector
TGA100A Trace Gas Analyzer Overview
Sample
Detector
To Pump
To Pump
Reference Gas In
Sample Cell
Laser
Sample In
FIGURE OV2.1-1. Schematic Diagram of TGA100A Optical System
Dewar
OV-3
TGA100A Trace Gas Analyzer Overview
OV2.2 Laser Scan Sequence
The laser is operated using a scan sequence that includes three phases: the zero
current phase, the high current phase, and the modulation phase, as illustrated
in Figure OV2.2-1. The modulation phase performs the actual spectral scan.
During this phase the laser current is increased linearly over a small range
(typically +/- 0.5 to 1 mA). The laser’s emission wavenumber depends on its
current. Therefore the laser’s emission is scanned over a small range of
frequencies (typically +/- 0.03 to 0.06 cm
During the zero current phase, the laser current is set to a value below the
laser’s emission threshold. “Zero” signifies the laser emits no optical power; it
does not mean the current is zero. The zero current phase is used to measure
the detector’s dark response, i.e., the response with no laser signal.
The reduced current during the zero phase dissipates less heat in the laser,
causing it to cool slightly. The laser’s emission frequency depends on its
temperature as well as its current. Therefore the temperature perturbation
caused by reduced current introduces a perturbation in the laser’s emission
frequency. During the high current phase the laser current is increased above
its value during the modulation phase to replace the heat “lost” during the zero
phase. This stabilizes the laser temperature quickly, minimizing the effect of
the temperature perturbation. The entire scan sequence is repeated every 2 ms
(500 scans per second).
-1
).
High Current Phase
(Temperature
Stabilization)
Zero Current Phase
(Laser Off)
FIGURE OV2.2-1. TGA100A Laser Scan Sequence
Modulation Phase
(Spectral Scan)
Omitted
2 ms
Laser
Current
Used in
Calculation
Detector
Response
OV-4
OV2.3 Concentration Calculation
The reference and sample detector signals are digitized at 50 kHz (100 samples
per scan), corrected for detector offset and nonlinearity, and converted to
absorbance. A linear regression of sample absorbance vs. reference absorbance
gives their ratio. The assumption that temperature and pressure are the same
for the sample and reference gases is fundamental to the design of the
TGA100A. It allows the concentration of the sample, C
TGA100A Trace Gas Analyzer Overview
, to be calculated by:
S
))()((
DLC
C
=
s
RR
AS
)1(
DLL
−+
where C
L
L
L
= concentration of reference gas, ppm
R
= length of the short reference cell, cm
R
= length of the short sample cell, cm
S
= length of the long sample cell, cm
A
D = ratio of sample to reference absorbance
The trace gas concentration is calculated for each scan and then digitally
filtered to reduce noise.
OV3. Trace Gas Species Selection
The TGA100A can measure gases with absorption lines in the 3 to 10 micron
range by selecting appropriate lasers, detectors, and reference gas. Lead-salt
tunable diode lasers have a very limited tuning range, typically 1 to 3 cm
within a continuous tuning mode. In some cases more than one gas can be
measured with the same laser, but usually each gas requires its own laser. The
laser dewar has two laser positions available (four with an optional second
laser mount), allowing selection of up to four different species by rotating the
dewar, installing the corresponding cable, performing a simple optical
realignment, and switching the reference gas.
-1
The standard detectors used in the TGA100A are Peltier cooled, and operate at
wavelengths up to 5 microns. These detectors are used for most gases of
interest, including nitrous oxide (N
). Some gases, such as ammonia (NH3), have the strongest absorption
(CO
2
O), methane (CH4), and carbon dioxide
2
lines at longer wavelengths, and require the optional long wavelength, liquid
nitrogen-cooled detectors.
A prepared reference gas must flow through the reference cell to provide a
spectral absorption template for the target gas. The concentration of this
reference gas is chosen to give approximately 50% absorption at the center of
the absorption line. A 200 ft
year at the recommended flow rate of 10 ml min
3
(5.7 m3) tank of reference gas will last over one
-1
.
OV-5