TheSDM-CD16S 16 CHANNEL SOLID STATE DC CONTROL
MODULE is warranted by CAMPBELL SCIENTIFIC, INC. to be free from
defects in materials and workmanship under normal use and service for twelve
(12) months from date of shipment unless specified otherwise. Batteries have
no warranty. CAMPBELL SCIENTIFIC, INC.'s obligation under this
warranty is limited to repairing or replacing (at CAMPBELL SCIENTIFIC,
INC.'s option) defective products. The customer shall assume all costs of
removing, reinstalling, and shipping defective products to CAMPBELL
SCIENTIFIC, INC. CAMPBELL SCIENTIFIC, INC. will return such
products by surface carrier prepaid. This warranty shall not apply to any
CAMPBELL SCIENTIFIC, INC. products which have been subjected to
modification, misuse, neglect, accidents of nature, or shipping damage. This
warranty is in lieu of all other warranties, expressed or implied, including
warranties of merchantability or fitness for a particular purpose. CAMPBELL
SCIENTIFIC, INC. is not liable for special, indirect, incidental, or
consequential damages.
Products may not be returned without prior authorization. The following
contact information is for US and International customers residing in countries
served by Campbell Scientific, Inc. directly. Affiliate companies handle
repairs for customers within their territories. Please visit
www.campbellsci.com to determine which Campbell Scientific company
serves your country. To obtain a Returned Materials Authorization (RMA),
contact CAMPBELL SCIENTIFIC, INC., phone (435) 753-2342. After an
applications engineer determines the nature of the problem, an RMA number
will be issued. Please write this number clearly on the outside of the shipping
container. CAMPBELL SCIENTIFIC's shipping address is:
CAMPBELL SCIENTIFIC, INC.
RMA#_____
815 West 1800 North
Logan, Utah 84321-1784
CAMPBELL SCIENTIFIC, INC. does not accept collect calls.
SDM-CD16S Table of Contents
PDF viewers note: These page numbers refer to the printed version of this document. Use
the Adobe Acrobat® bookmarks tab for links to specific sections.
1. Datalogger to SDM-CD16S Connections.................................................. 6
2. Switch Position and Addresses.................................................................. 7
ii
SDM-CD16S 16 Channel Solid State DC
Control Module
1. Function
FIGURE 1. SDM-CD16S Face Panel
The SDM-CD16S has 16 DC voltage outputs that can be switched on and off
manually or under datalogger control. Separate inputs for the power to the
outputs (48 VDC max) and the power to SDM-CD16 logic (7 – 48 VDC)
allow the option of powering the logic from the datalogger 12 V while
switching a higher voltage. LEDs allow a visual indicator of active outputs.
The outputs can be controlled by a datalogger or controlled manually with a
manual override toggle switch and individual rocker switches for each of the
outputs.
The toggle switch has three positions: MANUAL, OFF and AUTO. In the
MANUAL position, outputs are controlled by the position of the individual
rocker switches. In the OFF position, all outputs are off. In the AUTO
position, the state of the relays is controlled by the SDM commands from the
datalogger or by the logic control inputs.
The SDM-CD16S is a synchronously addressed datalogger peripheral.
Datalogger control ports 1, 2, and 3 are used to address the SDM-CD16S, then
clock out the desired state of each of the 16 control ports. Up to 16 SDMCD16Ss may be addressed, making it possible to control a maximum of 256
ports from the datalogger’s first three control ports. The SDMCD16AC
instruction is used to control the SDM-CD16S in CRBasic dataloggers. In
1
SDM-CD16S 16 Channel Solid State DC Control Module
Edlog dataloggers that support SDM-CD16 devices, I/O Instruction 104 is
used (some old CR7s may use Instruction 29).
In addition to SDM control, the SDM-CD16S has individual digital control
inputs for each of the outputs. A logic high (5 V) on the control input will set
the associated output high. The control input OR the SDM command can set
the output high. Both the SDM command and the control input must be low
for the port to be low. The logic control inputs may be useful for using one of
the datalogger’s control ports with the pulse width modulation instruction for a
proportional control that changes faster than the datalogger scan rate. (The
pulse width modulation instruction is PWM in CRBasic dataloggers,
Instruction 137 in the CR23X.) Devices other than the datalogger that have a
0V/5V logic output for control might also be used to activate the digital control
inputs.
NOTE
Ensure that the datalogger contains the appropriate instruction
and test your datalogger program before going in the field.
Logic power voltage: 7 to 48 VDC
Logic current drain at 12 VDC: 15 mA quiescent;
Maximum cable length: 20 ft total to all SDM devices.
Toggle switch: MANUAL, OFF, AUTO
Supply voltage for output: 48 Volt Max, DC only
Maximum current per channel: 2 Amps
Maximum current all channels total: 10 Amps Fused
CR1000, CR850, CR800, CR23X,
CR10, 21X, and CR7.
2.5 mA per active LED (manual or
auto)
Consult CSI if longer lengths are
necessary.
Individual dip switches for manual
2
Fuse: 3 AG – 10 A
Actuation/release times:8 μs/200 μs
Operating temperature: -40° to 70°C
* If more detailed output characteristics are required, experienced users should
consult the equivalent circuit diagrams shown in Figure 2, below.
SDM-CD16S 16 Channel Solid State DC Control Module
FIGURE 2. Equivalent Output Driver Circuit
3. Power Considerations
The SDM-CD16S power requirements may be large compared to most CSI
products. For most applications, an external power supply (see Figure 3) is
recommended to power the SDM-CD16S.
For some applications, it may be convenient to use the datalogger supply to
power the SDM-CD16S (see Figure 3). For long-term applications, the lead
acid power supply available with CSI dataloggers should be used, allowing the
batteries to be float charged. It is not recommended that the datalogger
alkaline supply be used to power the SDM-CD16S for long term applications.
If the datalogger lead acid supply is used, the current available from the wall
charger limits the SDM-continuous output current.
3
SDM-CD16S 16 Channel Solid State DC Control Module
EXTERNAL
7 TO 48 VDC
+ —
GND
SDM-CD16S
Connection with External Supply
SDM-CD16S
Connection with Datalogger Supply
FIGURE 3. Connection Block Diagrams
If the 21X power supply is used to power the SDM-CD16S Load Power, all
low level analog measurements (thermocouples, pyranometers, thermopiles,
etc.) must be made differentially. This is a result of slight ground potentials
created along the 21X analog terminal strip when the 12 V supply is used to
power peripherals. This limitation reduces the number of available analog
input channels and may mandate an external supply for the SDM-CD16S.
C1
C2
C3
GND
12 V
C1
C2
C3
DATALOGGER
DATALOGGER
4. Installation
CAUTION
4
•The SDM-CD16S must be installed in an enclosure that provides a
pollution degree 2 environment (normally, only nonconductive pollution.
However, a temporary conductivity caused by condensation may be
expected). All Campbell Scientific enclosures meet this requirement.
•Use copper conductors only.
Cables connecting the terminals of the datalogger and
SDM device should be kept as short as possible to
minimize the risk of corruption of the signals and damage
from induced surges. Where long cable runs (>3 m) are
unavoidable and the cables run outside, some extra
protection may be required for the SDM control terminals.
Please contact Campbell Scientific for further advice.
When connecting wires to the SDM signal terminals,
please ensure they are at ground potential before making
the connection, e.g. by touching them to the earth terminal.
SDM-CD16S 16 Channel Solid State DC Control Module
For datalogger connections, see Table 1.
Multiple SDM-CD16Ss may be wired in parallel by connecting the datalogger
side of one SDM-CD16S to the next. In most installations the total length of
the cables connecting the SDM-CD16S and other SDM peripherals should not
exceed 20 feet.
Total cable lengths in excess of 20 feet may adversely influence
communication performance. (For CR7 dataloggers, the total cable length
should not exceed 600 feet.)
4.1 Wiring
4.1.1 SDM-CD16S Power and Control Connections
Refer to Figure 3 and Table 1 for SDM-CD16S operating power and control
connections to the datalogger.
EXTERNAL
7 TO 48 VDC
SDM-CD16S
SDM-CD16S
FIGURE 3. Connection Block Diagrams
+ —
GND
C1
C2
C3
Connection to External Supply
GND
12 V
C1
C2
C3
Connection to Datalogger Supply
DATALOGGER
DATALOGGER
5
SDM-CD16S 16 Channel Solid State DC Control Module
TABLE 1. Datalogger to SDM-CD16S Connections
SDM-CD16S Datalogger Function
Logic Power 12 V on datalogger or
7 to 48 V external supply
Logic Gnd Gnd Common ground
C1 C1 (Control Port 1) Data
C2 C2 (Control Port 2) Clock
C3 C3 (Control Port 3) Enable
4.1.2 Controlled Device Connections
In most applications, the SDM-CD16S acts as a switch (controllable break) in
the positive supply of the circuit powering the controlled device. Figure 4
shows an example.
Power
DC Power
Supply
+
GND
FIGURE 4. Typical Wiring Application
5. Address Selection Switch
Each SDM-CD16S can have 1 of 16 addresses. Shipped from the factory, the
address is set at 00. Table 2 shows switch position and the corresponding
address.
SDM_CD16
Load
Power
Out
Device
6
SDM-CD16S 16 Channel Solid State DC Control Module
The SDMCD16AC instruction is used to control all versions of the SDMCD16.
SDMCD16AC (Source, Reps, SDMAddress)
Remarks
A port on an SDM-CD16AC is enabled/disabled (turned on or off) by sending
a value to it using the SDMCD16AC instruction. A non-zero value will enable
the port; a zero value disables it. The values to be sent to the SDM-CD16AC
are held in the Source array.
The SDMCD16AC instruction has the following parameters:
Source The Source parameter is an array (dimensioned as Float,
Long, or Boolean) which holds the values that will be sent to
the SDM-CD16AC to enable/disable its ports. An SDMCD16AC has 16 ports; therefore, in most instances the
source array should be dimensioned to 16 times the number
of Repetitions (the number of SDM-CD16AC devices to be
controlled). As an example, with the array CDCtrl(32), the
value held in CDCtrl(1) will be sent to port 1, the value h eld
in CDCtrl(2) will be sent to port 2, etc. The value held in
7
SDM-CD16S 16 Channel Solid State DC Control Module
CDCtrl(32) would be sent to port 16 on the second SDMCD16AC.
If the Source parameter is defined as a Long variable, but it
is dimensioned less than 16 * Reps, Source will act as a
binary control for the instruction whose bits 0..15 will
specify control ports 1..16, respectively. In this instance,
Source(1) will be used for the first rep, Source(2) will be
used for the second, and so on.
Reps The Reps parameter is the number of SDM-CD16AC
devices that will be controlled with this instruction.
SDMAddress The SDMAddress parameter is used to define the address of
the SDM-CD16AC that will be controlled with this
instruction. Valid SDM addresses are 0 through 14.
Address 15 is reserved for the SDMTrigger instruction. If
the Reps parameter is greater than 1, the datalogger will
increment the SDM address for each subsequent device that
it communicates with.
NOTE
CRBasic dataloggers use base 10 when addressing SDM devices.
Edlog programmed dataloggers (e.g., CR10X, CR23X) used
base 4 for addressing (Table 2).
CRBasic dataloggers also have the TimedControl instruction which allows a
timed sequence of settings to be managed by the measurement task avoiding
possible processing delays to cause delayed switching (pipeline mode only).
See an example in Section 8.4.
The CR10/10X, CR23X, CR7 and 21X use instruction 104 to control the
SDM-CD16S. Instruction 29 is used by older CR7s. The Instruction
descriptions are shown below.
Instruction 104 – SDM-CD16S used with CR10/10X, CR23X,
The number of SDM-CD16Ss to be addressed is defined by the Reps
(repetitions) parameter. Each Rep sequentially addresses (00, 01, 02,...32, 33)
SDM-CD16Ss, starting with the address specified in parameter 2 (parameter 3
for Instruction 29).
For each repetition, the 16 ports of the addressed SDM-CD16S are set
according to 16 sequential input locations starting at the input location
specified in parameter 3 (parameter 5 for Instruction 29). Any non-zero value
stored in an input location activates (sets HI 5V) the associated SDM-CD16S
port. A value of zero (0) de-activates the port (sets LO 0V). For example,
assuming two repetitions and a starting input location of 33, outputs 1 to 16 of
the first SDM-CD16S are set according to input locations 33 to 48, and outputs
1 to 16 of the second SDM-CD16S are set according to input locations 49 to
64.
For older CR7s with Instruction 29, the Device (parameter 2) specifies what
type of synchronously addressed peripheral is to be addressed. The Device
code for an SDM-CD16S is 2.
For Instruction 29 only (older CR7s), the Card parameter (parameter 4)
specifies which 725 Excitation Card is being used for the control port signals.
The Reps parameter does not advance beyond the specified Card, requiring
another Instruction 29 for each 725 Excitation Card used.
7. Theory of Operation
The SDM-CD16S is a synchronously addressed peripheral. C2 and C3, driven
high by the datalogger, initiate a cycle. While holding C3 high, the datalogger
drives C2 as a clock line and C1 as a serial data line. The datalogger shifts out
a data bit on C1 (LSB first) on the falling edge of the C2 clock. The SDMCD16S shifts in the C1 data bit on the rising edge of the C2 clock.
The first 8 bits clocked out represent the SDM-CD16S address. If the address
matches the SDM-CD16S's address, the SDM-CD16S is enabled. If enabled,
the next 16 bits are shifted into the SDM-CD16S, each bit controlling one port,
the first of which controls OUT 1.
When the 16 control bits are clocked in, C2 is held high while C3 is pulsed low
then high to latch the control bits. The datalogger then lowers both C3 and C2
to complete the cycle.
9
SDM-CD16S 16 Channel Solid State DC Control Module
8. Program Examples
8.1 Control Temperature and Fans – CR1000
In this example, the SDM-CD16S is used to control the temperature between
23° and 28°C in each of 5 greenhouses. In each greenhouse the SDM-CD16S
controls a heating unit, a refrigerating unit, and an air-mixing fan according to
the following conditions.
Heating unit: Activate when temperature < 23.5°C. Deactivate when
temperature > 25.5°C
Cooling unit: Activate when temperature > 27.5°C. Deactivate when
temperature < 24.5°C
Mixing fan: Activate whenever the heating or cooling units are activated.
Activate for 5 minutes out of every 15 minutes.
The program assumes the temperature measurements have been made, and the
average temperature for each greenhouse is computed and residing in the
appropriate variable
EXAMPLE 1: the actual values used to control
the SDM-CD16: CD16_Output(I), I = 1 to 5 are
for Heat, I = 6 to 10 are for Cooling, I= 11 to 15
are for Fans
EXAMPLE 2: the actual value used to control
SDMCD the CD16_Output bits set the SDM-
CD16S ports. bits 0 to 4 are for ‘Heat, 5 to 9 are
for Cooling, 10 to 14 are for Fans
10
SDM-CD16S 16 Channel Solid State DC Control Module
The Example 1 program uses an array of values to set the SDM-CD16S control
outputs:
Program name: SDMCD16Example1.CR1
'Date written: 6/25/2007
'\\\\\\\\\\\\\\\\\\\\\\\\\ DECLARATIONS /////////////////////////
Public Flag(8) as boolean
Public I
Public Temp(5)
Public Heat(5)
Public Cool(5)
Public Fan(5)
'Note CD16_Output(I), I = 1 to 5 are for Heat, I = 6 to 10 are for ‘Cooli ng, I= 11 to 15 are for Fans
Dim CD16_Output(16)
'\\\\\\\\\\\\\\\\\\\\\\\\\\\ PROGRAM ////////////////////////////
BeginProg
Scan(5,Sec, 3, 0)
For I = 1 to 5
If (Temp(I) < 23.5) Then
Heat(I) = 1
ElseIf (Temp(I) >= 25.5) Then
Heat(I) = 0
EndIf
If (Temp(I) >= 27.5) Then
Cool(I) = 1
ElseIf (Temp(I) < 24.5) Then
Cool(I) = 0
EndIf
If (Heat(I) <> 0) OR (Cool(I) <> 0) Then
Fan(I) = 1
Else
Fan(I) = 0
EndIf
Next I
If TimeInToInterval(10,15,Min) Then Flag(2) = True
If TimeInToInterval(0,15,Min) Then Flag(2) = False
If Flag(2) = True then
For I = 1 to 5
Fan(I) = 1
Next I
EndIf
For I = 1 to 5
CD16_Output(I) = Heat(I)
CD16_Output(I+5) = Cool(I)
CD16_Output(I+10) = Fan(I)
Next I
SDMCD16AC(CD16_Output(), 1, 0)
NextScan
EndProg
11
SDM-CD16S 16 Channel Solid State DC Control Module
The Example 2 program uses an integer instead of an array to set the SDMCD16S control outputs:
Public Temp(5)
Public TimedFanOn as Boolean
Dim I as Long
Dim CD16_Output as Long
'Note: CD16_Output bits set the SDM-CD16S ports. bits 0 to 4 are for ‘Heat,
‘5 to 9 are for Cooling, 10 to 14 are for Fans
'\\\\\\\\\\\\\\\\\\\\\\\\\\\ PROGRAM ////////////////////////////
BeginProg
Scan(5,Sec, 3, 0)
For I = 1 to 5
If (Temp(I) < 23.5) Then 'Set appropriate Heater Bit High:
CD16_Output = CD16_Output OR 2^(I-1)
ElseIf (Temp(I) >= 25.5) Then 'Set appropriate Heater Bit Low:
CD16_Output = CD16_Output AND (&H7FFF - 2^(I-1))
EndIf
If (Temp(I) >= 27.5) Then 'Set appropriate Cooler Bit High:
CD16_Output = CD16_Output OR 2^(I+4)
ElseIf (Temp(I) < 24.5) Then 'Set appropriate Cooler Bit Low:
CD16_Output = CD16_Output AND (&H7FFF - 2^(I+4))
EndIf
Next I
CD16_Output = (CD16_Output AND &H3FF) 'Set all Fan Bits Low
'Turn on Fan Bits for active Heaters or Coolers:
CD16_Output = CD16_Output OR (((CD16_Output*2^5) OR (CD16_Output*2^10)) AND &H7C00)
If TimeInToInterval(10,15,Min) Then TimedFanON = True
If TimeInToInterval(0,15,Min) Then TimedFanON = False
If TimedFanON = True Then CD16_Output = CD16_Output OR &H7C00
SDMCD16AC(CD16_Output(), 1, 0)
NextScan
EndProg
12
8.2 Control Temperature and Fans – CR10X
The example is written for the CR10(X) Measurement and Control Module.
The program concepts presented are the same for the 21X and CR7 dataloggers
with minor program code changes.
The conditions and set points are the same as for example 8.1.
The program assumes the temperature measurements have been made, and the
average temperature for each greenhouse is computed and residing in Input
Locations 1 through 5.
SDM-CD16S 16 Channel Solid State DC Control Module
Input Location assignments are as follows:
Input
Location
Location
Label
Description
1..5 Temp #1..#5 Avg temp, greenhouse 1..5
10..14 Heat #1..#5 Heater control, greenhouse 1..5 SDM-CD16S
Port 1..5
15..19 Cool #1..#5 Cooler control, greenhouse 1..5 SDM-CD16S
Port 6..10
20..24 Fan #1..#5 Fan control, greenhouse 1..5 SDM-CD16S
Port 11..15
1: Beginning of Loop (P87) Master Loop, End
1: 0 Delay Loop at Step 30
2: 5 Loop Count
START HEATER CONTROL LOGIC
2: If X<=>F (P89) If "Heater On"
1: 1-- X Loc threshold is
2: 4 < exceeded
3: 23.5 F
4: 30 Then Do Then
3: Z=F (P30) Put a "1" into Heater
1: 1 F Control Location
2: 0 Exponent of 10
3: 10-- Z Loc :
4: End (P95) End Then Do/End
5: If X<=>F (P89) If Heater
1: 10-- X Loc #1 on (Heater Control
2: 2 <> Location <> 0)
3: 0 F
4: 30 Then Do Then
6: If X<=>F (P89)
1: 1-- X Loc Temp #1 Check Upper Threshold
2: 3 >= to see if heater should
3: 25.5 F be tu rned off
4: 30 Then Do
7: Z=F (P30) If heater should be turned
1: 0 F off, enter a "0" into
2: 0 Exponent of 10 heater control location
3: 10-- Z Loc :
8: End (P95) Else Then Do/End
13
SDM-CD16S 16 Channel Solid State DC Control Module
9: Else (P94) Else, If the heater is off,
10: Z=F (P30)
1: 0 F Enter a "0" into heater
2: 0 Exponent of 10 control location
3: 10-- Z Loc :
11: End (P95) End Then Do/Else/End
END HEATER CONTROL LOGIC
START COOLER CONTROL LOGIC
12: If X<=>F (P89) If "Cooler" on
1: 1-- X Loc threshold is
2: 3 >= exceeded
3: 27.5 F
4: 30 Then Do Then
13: Z=F (P30) Put a "1" into cooler
1: 1 F Control Location
2: 0 Exponent of 10
3: 15-- Z Loc :
14: End (P95) End Then Do/End
15: If X<=>F (P89) If cooler is on
1: 15-- X Loc (Cooler control
2: 2 <> Location <>0)
3: 0 F
4: 30 Then Do Then
16: If X<=>F (P89) Check lower threshold to
1: 1-- X Loc see if cooler should be
2: 4 < turned off
3: 24.5 F
4: 30 Then Do
17: Z=F (P30) If cooler should be turned
1: 0 F off, put a "0" into cooler
2: 0 Exponent of 10 control location
3: 15-- Z Loc :
18: End (P95) End Then Do/End
19: Else (P94) Else if cooler is off
20: Z=F (P30)
1: 0 F Put a "0" into cooler
2: 0 Exponent of 10 control location
3: 15-- Z Loc :
21: End (P95) End Then Do/Else/End
END COOLER CONTROL LOGIC
14
SDM-CD16S 16 Channel Solid State DC Control Module
START FAN CONTROL LOGIC BASED ON HEATER/COOLER
22: If X<=>F (P89) If heater is on
1: 10-- X Loc
2: 2 <>
3: 0 F
4: 11 Set high Flag 1 Set flag 1
23: If X<=>F (P89) If cooler is on
1: 15-- X Loc
2: 2 <>
3: 0 F
4: 11 Set high Flag 1 Set flag 1
24: If Flag/Port (P91) If flag 1 is set
1: 11 Do if flag 1 is high
2: 30 Then Do Then
25: Z=F (P30) Put a "1" into fan
1: 1 F control location
2: 0 Exponent of 10
3: 20-- Z Loc:
26: Else (P94) Else, If flag 1 is reset
27: Z=F (P30) Put a "0" into fan
1: 0 F control location
2: 0 Exponent of 10
3: 20-- Z Loc :
28: End (P95) End Then Do/Else/End
29: Do (P86) Reset flag 1
1: 21 Set low Flag 1
30: End Loop (P95) End Master Loop
END FAN CONTROL LOGIC BASED ON HEATER/COOLER
START FAN CONTROL LOGIC BASED ON TIME
31: If time is (P92) If 5 minutes remain
1: 10 minutes into a out of 15 minute
2: 15 minute interval interval,
3: 12 Set high Flag 2 set flag 2
32: If Flag/Port (P91) If flag 2 is set
1: 12 Do if flag 2 is high
2: 30 Then Do Then
15
SDM-CD16S 16 Channel Solid State DC Control Module
33: Beginning of Loop (P87) Start fan loop
1: 0 Delay
2: 5 Loop Count
34: Z=F (P30) PUT A "1" INTO FAN
1: 1 F CONTROL LOCATION
2: 0 Exponent of 10
3: 20-- Z Loc :
35: End (P95) End fan loop
36: End (P95) End then do
37: If time is (P92)
1: 0 minutes into a Reset flag 2 at the
2: 15 minute interval end of the 15 minute
3: 22 Set low Flag 2
END FAN CONTROL LOGIC BASED ON TIME
INPUT LOCATIONS 10 THROUGH 24 ARE NOW LOADED WITH "1" OR "0" TO SET
PORTS ON THE SDM-CD16S.
38: SDM-CD16S (P104) Send instructions to the
1: 1 Reps SDM-CD16S with address 00
2: 00 Address
3: 10 Loc
39: End Table 1 (P)
8.3 Control Gas Sampling Based on Time – CR1000
The SDM-CD16S is ideal for controlling low power DC operated solenoid
valves such as might be used to control a manifold as part of a gas sampling
system. This example shows one way to do that.
8.3.1 Terminology
Site: A site may be the physical site where the air is sampled (i.e. intake or
chamber) or a connection to a calibration gas. Each site has a unique
SDMCD16S/valve setting that configures the manifold to connect gas from
that site to the sample path of the gas analyzer. There are 8 connections to the
“site selection” manifold. The numbers 1 through 8 identify the sites. The site
numbers are the same as the number of the intake on the manifold and the
number of the SDMCD16 relay controlling that valve. Sites 1 – 6 are
connected to sample intakes to measure a atmospheric profile and sites 7 and 8
are used for calibration gases.
Site Sequence: The sequence in which the sites are measured is determined
by the SiteSequence array. The sites are measured in the sequence that the site
numbers are entered in this array, not necessarily in order of site number.
16
SDM-CD16S 16 Channel Solid State DC Control Module
The SiteSequence array also allows a site to be measured multiple times within
the sequence. For some measurements it is necessary to measure calibration
gases frequently (e.g., the concentration of various isotopes of CO2 with a
trace gas analyzer.)
The sequence in the example, with the six sample intakes and two calibration
gases, is to measure 3 sample intakes, measure the zero and span, measure the
other 3 intakes, and again measure the zero and span. There are 10 sites in the
sequence. (The SiteSequence array is dimensioned to 10 elements.) The
sequence is 1, 2, 3, 7, 8, 4, 5, 6, 7, 8. Thirty seconds is spent on each site the
sequence. Thus it takes 5 minutes to run through the sequence of 10 sites (10
sites x 30 seconds/site). This sequence is repeated again and again.
OmitCounts: When the manifold is switched to a new site, it is necessary to
purge the line before the analyzer measurement represents the new site. The
program includes a counter that is incremented by one each scan. The omit
counts is number of measurements to exclude from an average for the site after
switching to the site.
The program records averages for each time the manifold is set to a site. The
average does not include values measured while the count is less that the omit
count. The count is reset to 0 when the program switches to a new site. In this
program there is a constant omitcounts that is used for all sites. If it was
necessary to set the Omit Counts independently for each site, OmitCounts
could be a variable array instead of a constant.
8.3.2 Data Tables
There are two data tables, RawData and SiteAvg. RawData holds samples of
every measurement. SiteAvg holds averages that are calculated for a site when
the manifold is switched to the next site.
8.3.3 Initialization Section (between BeginProg and Scan)
Manifold/SDMCD16S Settings: This section defines the bit pattern for each
valve setting.
Site Sequence: Edit the SiteSequence array assignment for the sequence
desired. For example, the 10 element SiteSequence array described in the
Terminology section above can be loaded with the code:
Dim SiteSequence(10) As Long 'The sequence consists of 10 valve settings
Dim SiteValve(8) As Long 'There are 8 valve settings in the above sequence
Dim SiteValveCD16 As Long 'This is the variable that is used to set the SDM-CD16
Dim AvgDisable As Boolean
Public EndOmit As Boolean
Public SeqActiveFlag As Boolean 'Set false to disable auto site switching
Public Site As Long
Public SiteOutput As Long
Public Count As Long
Public SeqIndex As Long
'Variables for Measurements
Public GasConc(2)
'Variable that controls valve switching and averaging periods
One=1
SeqIndex=1
Count = 1
SeqActiveFlag = True
Site = SiteSequence(SeqIndex)
SiteOutput = Site
Scan (1,Sec,10,0)
'Instructions to Measure Sensors Inserted Here.
'Set valves:
SDMCD16AC (SiteValveCD16,1,1)
'Call Data Tables:
CallTable RawData
CallTable SiteAvg
Count = Count + 1
SiteOutput=Site
If SeqActiveFlag Then
If TimeIntoInterval (0,30,Sec) Then
SeqIndex = (SeqIndex MOD 10) + 1
Count = 1
19
SDM-CD16S 16 Channel Solid State DC Control Module
EndIf
Site = SiteSequence(SeqIndex)
EndIf
SiteValveCD16 = SiteValve(Site)
EndOmit = Count > OmitCounts
AvgDisable = NOT (EndOmit AND SeqActiveFlag)
NextScan
EndProg
8.4 Control Gas Sampling with Timing in Measurement Task –
CR1000
8.4.1 The Purpose of this Example
This example illustrates an instruction that can be used if:
1) There is a requirement for precisely timed switching in a known timing
sequence.
2) The time required in the datalogger program for processing tasks is close to
the scan interval.
If the processing occasionally takes longer than the scan interval, there could
be skipped measurements (sequential mode) or the measurements could be
made but the switching would take place later than expected (pipeline mode).
This program has the same timing and valve control as the example in Section
8.2, however, instead of the logic in the program determining when to switch
the valves, it makes use of the TimedControl instruction to control when valves
switch as part of the measurement task (pipeline mode only).
8.4.2 Pipeline vs Sequential
The Pipeline mode is more efficient than the Sequential mode. In the
Sequential mode, all processing is completed before advancing to the next
instruction. The Pipeline mode separates the measurement and processing
tasks. Not only does this separation of tasks allow the datalogger to complete
both tasks in less time than the sequential mode, it allows the processing task
to briefly fall behind the measurement task without skipping any
measurements.
In the pipeline mode the results of the measurement task are stored in a buffer
for the processing task. The processing task starts after the first measurement
task is complete and can take place at the same time as the next measurement
task. If the processing task has something that temporarily prevents it from
finishing before the next measurement scan is complete, it can fall behind the
measurement scan. Later when the processing is less intensive, the processing
can run faster than the measurements and process the buffered measurement
data and catch up to the measurement scan. Operations that may temporarily
increase the processing are intensive calculations for final output processing,
extra communications overhead, or initializing a new compact flash card.
20
SDM-CD16S 16 Channel Solid State DC Control Module
In the example in Section 8.3, the logic that controls the SDM-CD16 is in the
processing task. In the pipeline mode the measurements continue to occur at
the proper time. If the processing falls behind the measurement the values
used to set the SDM-CD16 will be updated only when the processing task gets
to that point. This will be later than would be expected if the processing task
were not falling behind.
The TimedControl instruction allows the timing of when to switch SDMCD16
settings to be part of the measurement task rather than the processing task.
8.4.3 TimedControl Instruction
TimedControl: This instruction allows a sequence of fixed values and
durations (number of scans) to be controlled by the measurement task. It
allows a series of SDM-CD16 settings to be defined so that the switching
sequence occurs at the proper times even if processing is lagging behind the
measurement task.
Sync_interval: When the program is compiled and starts running or when the
TimedControl is reset the program will wait until an even multiple of this
interval to start the sequence. Enter 0 to start immediately. See Clock/Reset
Option for more information on what happens prior to the start.
Interval_Units: The time units for the Sync_Interval.
Default_Value: The value to set the output to prior to starting the seq uence.
Index_dest: A variable in which to store the index of the value that is
currently being used in the source_array. This instruction will load 0 fo r the
index if the instruction is waiting for the Sync_interval to start.
Source_Array: A two dimension array that contains the values to use and the
duration (in number of scans) for which that value is to remain in use. For
Example, arrayname(x,2) where the array must be dimensioned to have at least
as many variables in the x dimension as the Size parameter requires. For any
variable pair i, arrayname(i,1) = the value to set, and arrayname(i,2) = the
duration in number of scans to use that value before moving to the next value
in the array.
ClockResetOpt: When the TimedControl instruction occurs before
BeginProg this option is used to set how the instruction behaves when the
dataloggers clock is changed:
1. If the datalogger clock is changed, the instruction behaves as if it were just
started after compile and the input to the SDMCD16 goes to the default.
21
SDM-CD16S 16 Channel Solid State DC Control Module
2. The sequence continues running as if nothing happened until the next
occurrence of the sync interval (synced to the current clock) and then
restarts.
3. Ignore the change in the clock, keep the current count and index
proceeding as if nothing happened.
When the TimedControl instruction is used within the program (i.e, after
BeginProg) to reset or change the sequence this option is used to set what
happens between the time the instruction is executed and the sync interval
occurs. Only option codes 1 and 2 are valid:
1. The instructio n behaves as if it were just started after compile and the
input to the SDMCD16 goes to the default value.
2. The sequence continues running as if nothing happened until the next
occurrence of the sync interval and then restarts.
Dim SiteSequence(10) As Long 'The sequence consists of 10 valve settings
Dim SiteValve(8) As Long 'There are 8 valve settings in the above sequence
Dim SiteValveCD16 As Long 'This is the variable that is used to set the SDM-CD16
Dim SeqProgrammedFlag As Boolean
Dim AvgDisable As Boolean
Public EndOmit As Boolean
Public SeqActiveFlag As Boolean 'Set false to disable auto site switching 'and enable manual control of valves
Public Site As Long
Public SiteOutput As Long
Public Count As Long
Public SeqIndex As Long
Dim ValveTime(10,2) As Long
'Variables for Measurements
Public GasConc(2)
'Variable that controls valve switching and averaging periods
22
SDM-CD16S 16 Channel Solid State DC Control Module
Public OmitCounts(8) As Long
Dim I As Long
Dim One
SDM-CD16S 16 Channel Solid State DC Control Module
'Load Sequence Timing Array
For I = 1 To 10
'Load the Valve and Timing Array with the valve Settings:
ValveTime(I,1) = SiteValve(SiteSequence(I))
'Load the Valve and Timing Array with the number of scans at each setting:
ValveTime(I,2) = 30
Next I
'Initialize Variables
One=1
SeqIndex=1
Count = 1
SeqActiveFlag = True
Site = SiteSequence(SeqIndex)
SiteOutput = Site
Scan (1,Sec,10,0)
'Instructions to Measure Sensors Inserted Here.
'Set valves:
SDMCD16AC (ValveTime(1,1),1,1)
'Call Data Tables:
CallTable RawData
CallTable SiteAvg
Count = Count + 1
SiteOutput=Site
If SiteAvg.Output(1,1) Then Count = 1 'Reset Count when the 'Site average is output
If SeqActiveFlag Then
Site = SiteSequence(SeqIndex) 'Show the site number set by 'the Timed Control If SeqProgrammedFlag = False Then 'Reset the first valve to 'the automatic value
ValveTime(1,1) = SiteValve(1)
ValveTime(1,2) = 30
TimedControl(10,2,Min,1,SeqIndex,ValveTime,2)
TimedControl(10,2,Min,1,SeqIndex,ValveTime,2)
SeqProgrammedFlag = True
EndIf
Else 'Set the first valve setting to that for the current Site 'The Site nuumber may be changed by user to switch sites.
ValveTime(1,1) = SiteValve(Site)
ValveTime(1,2) = 0
If SeqProgrammedFlag = True Then 'Set the timed control to only 'use The first value
TimedControl(1,0,Min,SiteValve(Site),SeqIndex,ValveTime,1)
SeqProgrammedFlag = False
EndIf
EndIf
SiteValveCD16 = SiteValve(Site)
EndOmit = Count > OmitCounts(Site)
AvgDisable = NOT (EndOmit AND SeqActiveFlag)
NextScan
EndProg
24
This is a blank page.
Campbell Scientific Companies
Campbell Scientific, Inc. (CSI)
815 West 1800 North
Logan, Utah 84321
UNITED STATES
www.campbellsci.com
info@campbellsci.com
Campbell Scientific Africa Pty. Ltd. (CSAf)
PO Box 2450
Somerset West 7129
SOUTH AFRICA
www.csafrica.co.za
cleroux@csafrica.co.za
Campbell Scientific Australi
PO Box 444
Thuringowa Central
QLD 4812 AUSTRALIA
www.campbellsci.com.au
info@campbellsci.com.au
Campbell Scientific do Brazil Ltda. (CSB)
Rua Luisa Crapsi Orsi, 15 Butantã
CEP: 005543-000 São Paulo SP BRAZIL
www.campbellsci.com.br
suporte@campbellsci.com.br
Campbell Scientific Canada Corp. (CSC)
11564 - 149th Street NW
Edmonton, Alberta T5M 1W7
CANADA
www.campbellsci.ca
dataloggers@campbellsci.ca
Campbell Scientific Ltd. (CSL)
Campbell Park
80 Hathern Road
Shepshed, Loughborough LE12 9GX
UNITED KINGDOM
www.campbellsci.co.uk
sales@campbellsci.co.uk
Campbell Scientific Ltd. (France)
Miniparc du Verger - Bat. H
1, rue de Terre Neuve - Les Ulis
91967 COURTABOEUF CEDEX
FRANCE
www.campbellsci.fr
info@campbellsci.fr
Campbell Scientific Spain, S. L.
Psg. Font 14, local 8
08013 Barcelona
SPAIN
www.campbellsci.es
info@campbellsci.es
Please visit www.campbellsci.com to obtain contact information for your local US or International representative.
a Pty. Ltd. (CSA)
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.