Campbell Scientific Q7-1-L User Manual

INSTRUCTION MANUAL
Q-7.1 Net Radiometer
Revision: 5/96
Copyright (c) 1991-1996
Campbell Scientific, Inc.

Warranty and Assistance

The Q-7.1 NET RADIOMETER is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless specified otherwise. Batteries have no warranty. CAMPBELL SCIENTIFIC, INC.'s obligation under this warranty is limited to repairing or replacing (at CAMPBELL SCIENTIFIC, INC.'s option) defective products. The customer shall assume all costs of removing, reinstalling, and shipping defective products to CAMPBELL SCIENTIFIC, INC. CAMPBELL SCIENTIFIC, INC. will return such products by surface carrier prepaid. This warranty shall not apply to any CAMPBELL SCIENTIFIC, INC. products which have been subjected to modification, misuse, neglect, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied, including warranties of merchantability or fitness for a particular purpose. CAMPBELL SCIENTIFIC, INC. is not liable for special, indirect, incidental, or consequential damages.
Products may not be returned without prior authorization. The following contact information is for US and International customers residing in countries served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs for customers wi thin their territories. Please visi t www.campbellsci.com to determine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization (RMA), contact CAMPBELL SCIENTIFIC, INC., phone (435) 753-2342. After an applications engineer determines the nature of the problem, an RMA number will be issued. Please write this number clearly on the outside of the shipping container. CAMPBELL SCIENTIFIC's shipping address is:
CAMPBELL SCIENTIFIC, INC.
RMA#_____ 815 West 1800 North Logan, Utah 84321-1784
CAMPBELL SCIENTIFIC, INC. does not accept collect calls.

Q-7.1 NET RADIOMETER

The Q-7.1 is a high-output thermopile sensor which measures the algebraic sum of incoming and outgoing all-wave radiation (i.e. short-wave and long-wave components). Incoming radiation consists of direct (beam) and diffuse solar radiation plus long-wave irradiance from the sky. Outgoing radiation consists of reflected solar radiation plus the terrestrial long-wave component.

1. SPECIFICATIONS

60-junction thermopile with low electrical resistance (4 ohms nominal) to reduce susceptibility to noise
Nominal calibration factors 9.6 W m (for positive values), 11.9 W m negative values)
Spectral response 0.25 to 60 µm
Uncorrected wind effect: up to 6%
reduction @ 7 m s to 1% reduction @ 7 m s
-1
for positive fluxes, up
-1
for negative
fluxes.
-2
mV
-2 mV-1 1
-
(for
CAUTION: Ensure that the cable is tied to the support arm close to the ball joint. If the wires are allowed to flex where they emerge from the sensor head they will break. Sensors are shipped with an appropriate cable tie already installed.
Time constant: Approximately 30 seconds
Top and bottom surfaces painted black and
protected from convective cooling by hemispherical heavy-duty polyethylene windshields (0.25 mm thick)
Windshields do not require pressurization
O-ring seals for easy windshield
replacement
Desiccant contained in support arm; volume of desiccant tube 45 cm
3
; breather
port on the end of the support arm.
No power required
Size of sensing head 57 x 72 x 177 mm;
support arm 20 mm diameter, 750 mm long

2. INSTALLATION

Attach the square mounting plate to a vertical or horizontal pipe or rod that is less than 38 mm (1.5 inches) in diameter with the two larger U­bolts. Attach the radiometer support arm to the mounting plate with the two smaller U-bolts. The radiometer support arm does not need to be level. Fasten the cable to the pipe or rod with tape or plastic ties to prevent strain on the wires and damage to the instrument.
In the northern hemisphere install the radiometer so that the sensor head is pointing south. Likewise, in the southern hemisphere point the sensor head to the north.

2.1 LEVELING

The Q-7.1 must be level. The bubble level is accurate to ±1° and the bubble should be within the bulls-eye. An error of 5° in leveling may cause a cosine response error of 6% under normal conditions; much greater errors are possible under other conditions (e.g. sunrise, sunset and winter use with low sun angles).
To level the Q-7.1, use a 5/8" wrench on the hexagonal coupling nut on the instrument stem to bend the ball joint between the support arm and the instrument. If the instrument does not stay in position, tighten the large hexagonal nut on the support arm slightly with a 15/16" wrench.
CAUTION: Do not attempt to bend the ball joint by holding the instrument head alone, as this may break the stem.
1
Q-7.1 NET RADIOMETER

2.2 WIRING

If a differential measurement is made, connect the red (+) lead to the high side (e.g. 1H) of any datalogger differential channel and the black (-) lead to the low side (e.g. 1L). Also, connect a jumper wire between the low channel and analog ground to prevent common mode errors. Connect the clear (shield) lead to ground (G on the CR10(X), ground on the 21X and CR7).
If a single-ended measurement is made, connect the red lead to any datalogger single­ended channel (H or L) and the black lead to ground (AG on the CR10(X), ground on the 21X and CR7).
The black lead is negative with respect to the red lead when the net radiometer is mounted with the level up and there is more incoming radiation than outgoing.

3. PROGRAMMING

Measure the output of the Q-7.1 with either Instruction 2 (Differential Volts) or Instruction 1 (Single-Ended Volts). Use the 250 mV range for the CR10(X) and the 500 mV range for the 21X or CR7. The slow integration with 60 Hz rejection yields a more noise-free reading.
For still air, net radiation (Q*) can be computed from the thermopile voltage (V
If V
> 0, Q*(Wm-2) = Vt(mV) x F
t
If Vt < 0, Q*(Wm-2) = Vt(mV) x F
) by:
t
p n
where Fp and Fn are the positive and negative calibration factors respectively.
The calibration factors and serial number are given on a label under the sensor head.
The above formula give the correct readings for conditions of zero wind speed. Additional corrections can be applied to reduce errors for non-zero wind speeds.

3.1 CORRECTING ERRORS CAUSED BY WIND EFFECT

Most sensors which measure long wave radiation are subject to some degree of error caused by convective cooling as air moves past the sensors. The response of the Q-7.1 sensor has been determined by the manufacturer using a specially constructed wind tunnel. Curves fitted to experimental data by the manufacturer are given in Figure 1. These functions show the change in reading in response to increasing wind speed.
FIGURE 1. Percentage Change in Reading as a Function of Wind Speed
2
Loading...
+ 8 hidden pages