Campbell Scientific Q7-1-L User Manual

INSTRUCTION MANUAL
Q-7.1 Net Radiometer
Revision: 5/96
Copyright (c) 1991-1996
Campbell Scientific, Inc.

Warranty and Assistance

The Q-7.1 NET RADIOMETER is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless specified otherwise. Batteries have no warranty. CAMPBELL SCIENTIFIC, INC.'s obligation under this warranty is limited to repairing or replacing (at CAMPBELL SCIENTIFIC, INC.'s option) defective products. The customer shall assume all costs of removing, reinstalling, and shipping defective products to CAMPBELL SCIENTIFIC, INC. CAMPBELL SCIENTIFIC, INC. will return such products by surface carrier prepaid. This warranty shall not apply to any CAMPBELL SCIENTIFIC, INC. products which have been subjected to modification, misuse, neglect, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied, including warranties of merchantability or fitness for a particular purpose. CAMPBELL SCIENTIFIC, INC. is not liable for special, indirect, incidental, or consequential damages.
Products may not be returned without prior authorization. The following contact information is for US and International customers residing in countries served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs for customers wi thin their territories. Please visi t www.campbellsci.com to determine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization (RMA), contact CAMPBELL SCIENTIFIC, INC., phone (435) 753-2342. After an applications engineer determines the nature of the problem, an RMA number will be issued. Please write this number clearly on the outside of the shipping container. CAMPBELL SCIENTIFIC's shipping address is:
CAMPBELL SCIENTIFIC, INC.
RMA#_____ 815 West 1800 North Logan, Utah 84321-1784
CAMPBELL SCIENTIFIC, INC. does not accept collect calls.

Q-7.1 NET RADIOMETER

The Q-7.1 is a high-output thermopile sensor which measures the algebraic sum of incoming and outgoing all-wave radiation (i.e. short-wave and long-wave components). Incoming radiation consists of direct (beam) and diffuse solar radiation plus long-wave irradiance from the sky. Outgoing radiation consists of reflected solar radiation plus the terrestrial long-wave component.

1. SPECIFICATIONS

60-junction thermopile with low electrical resistance (4 ohms nominal) to reduce susceptibility to noise
Nominal calibration factors 9.6 W m (for positive values), 11.9 W m negative values)
Spectral response 0.25 to 60 µm
Uncorrected wind effect: up to 6%
reduction @ 7 m s to 1% reduction @ 7 m s
-1
for positive fluxes, up
-1
for negative
fluxes.
-2
mV
-2 mV-1 1
-
(for
CAUTION: Ensure that the cable is tied to the support arm close to the ball joint. If the wires are allowed to flex where they emerge from the sensor head they will break. Sensors are shipped with an appropriate cable tie already installed.
Time constant: Approximately 30 seconds
Top and bottom surfaces painted black and
protected from convective cooling by hemispherical heavy-duty polyethylene windshields (0.25 mm thick)
Windshields do not require pressurization
O-ring seals for easy windshield
replacement
Desiccant contained in support arm; volume of desiccant tube 45 cm
3
; breather
port on the end of the support arm.
No power required
Size of sensing head 57 x 72 x 177 mm;
support arm 20 mm diameter, 750 mm long

2. INSTALLATION

Attach the square mounting plate to a vertical or horizontal pipe or rod that is less than 38 mm (1.5 inches) in diameter with the two larger U­bolts. Attach the radiometer support arm to the mounting plate with the two smaller U-bolts. The radiometer support arm does not need to be level. Fasten the cable to the pipe or rod with tape or plastic ties to prevent strain on the wires and damage to the instrument.
In the northern hemisphere install the radiometer so that the sensor head is pointing south. Likewise, in the southern hemisphere point the sensor head to the north.

2.1 LEVELING

The Q-7.1 must be level. The bubble level is accurate to ±1° and the bubble should be within the bulls-eye. An error of 5° in leveling may cause a cosine response error of 6% under normal conditions; much greater errors are possible under other conditions (e.g. sunrise, sunset and winter use with low sun angles).
To level the Q-7.1, use a 5/8" wrench on the hexagonal coupling nut on the instrument stem to bend the ball joint between the support arm and the instrument. If the instrument does not stay in position, tighten the large hexagonal nut on the support arm slightly with a 15/16" wrench.
CAUTION: Do not attempt to bend the ball joint by holding the instrument head alone, as this may break the stem.
1
Q-7.1 NET RADIOMETER

2.2 WIRING

If a differential measurement is made, connect the red (+) lead to the high side (e.g. 1H) of any datalogger differential channel and the black (-) lead to the low side (e.g. 1L). Also, connect a jumper wire between the low channel and analog ground to prevent common mode errors. Connect the clear (shield) lead to ground (G on the CR10(X), ground on the 21X and CR7).
If a single-ended measurement is made, connect the red lead to any datalogger single­ended channel (H or L) and the black lead to ground (AG on the CR10(X), ground on the 21X and CR7).
The black lead is negative with respect to the red lead when the net radiometer is mounted with the level up and there is more incoming radiation than outgoing.

3. PROGRAMMING

Measure the output of the Q-7.1 with either Instruction 2 (Differential Volts) or Instruction 1 (Single-Ended Volts). Use the 250 mV range for the CR10(X) and the 500 mV range for the 21X or CR7. The slow integration with 60 Hz rejection yields a more noise-free reading.
For still air, net radiation (Q*) can be computed from the thermopile voltage (V
If V
> 0, Q*(Wm-2) = Vt(mV) x F
t
If Vt < 0, Q*(Wm-2) = Vt(mV) x F
) by:
t
p n
where Fp and Fn are the positive and negative calibration factors respectively.
The calibration factors and serial number are given on a label under the sensor head.
The above formula give the correct readings for conditions of zero wind speed. Additional corrections can be applied to reduce errors for non-zero wind speeds.

3.1 CORRECTING ERRORS CAUSED BY WIND EFFECT

Most sensors which measure long wave radiation are subject to some degree of error caused by convective cooling as air moves past the sensors. The response of the Q-7.1 sensor has been determined by the manufacturer using a specially constructed wind tunnel. Curves fitted to experimental data by the manufacturer are given in Figure 1. These functions show the change in reading in response to increasing wind speed.
FIGURE 1. Percentage Change in Reading as a Function of Wind Speed
2
Q-7.1 NET RADIOMETER
FIGURE 2. Correction Functions as a Function of Wind Speed
The inverse of this data gives the required correction factors to give the correct reading as a function of wind speed (u). Figure 2 shows the manufacturer’s data plus correction curves fitted by Campbell Scientific, using functions which can be easily encoded in a datalogger. Note that the functions fitted and shown in both graphs have been amended slightly to allow for a small under-estimate of the error (about 2% of the error) which is the result of a known error of the calibration technique in the wind tunnel.
The correction function for positive fluxes is:
××
(. . )
Correction factor = 1 +
For negative fluxes a linear function is used: Correction factor = If the wind speed is not measured it is often
sufficient to correct the original calibration constant using a fixed correction factor. The manufacturer suggests using fixed correction factors of 1.045 for positive fluxes and 1.000 for negative fluxes.
0 066 0 2
()
()
0 00174. ×
..
0 066 0 2
u
u
()
u
+ 0.99755
If the wind speed is measured, the functions above can be programmed into the datalogger to give a real-time correction of net radiation for the effects of wind speed. Note that the wind speed is expressed in units of meters per second. If the wind speed is measured in units other than meters per second, convert the wind speed to meters per second before inputting it into the above correction functions.
If AC power is available, a special ventilator (RV2) is available for the Q-7.1 which gives a typical wind speed of 3.3 ms negative fixed corrections are 1.06 and 1.0033 respectively. The fan is also useful for reducing the risk of external condensation (dew) on the windshields.
The above corrections were calculated at sea level under conditions close to standard temperature and pressure. The magnitude of the wind speed error will vary in relation to the air density and the heat capacity of air. These changes are generally small enough to be ignored unless extreme temperatures or atmospheric pressures are experienced.
-1
. Positive and
3
Q-7.1 NET RADIOMETER

3.2 PROGRAMMING EXAMPLE WITH FIXED WIND SPEED CORRECTION

The following CR10(X) program measures the output of a Q-7.1. A static wind correction is applied to each measurement.
Example 1. CR10(X) Program with the
Static Wind Speed Correction
;Measure the Q7.1. ;
01: Volt (Diff) (P2)***
1: 1 Reps 2: 24** ± 250 mV 60 Hz
Rejection Range 3: 1* DIFF Channel 4: 3* Loc [ Rn_W_m2 ] 5: 1 Mult 6: 0 Offset
02: IF (X<=>F) (P89)
1: 3* X Loc [ Rn_W_m2 ] 2: 3 >= 3: 0 F 4: 30 Then Do
;Enter the positive multiplier (p.ppp). ;
03: Z=X*F (P37)
1: 3* X Loc [ Rn_W_m2 ] 2: 3: 3* Z Loc [ Rn_W_m2 ]
04: Z=X*F (P37)
1: 3* X Loc [ Rn_W_m2 ] 2: 1.045 F 3: 3* Z Loc [ Rn_W_m2 ]
05: Else (P94)
p.ppp
F
;Enter the negative multiplier (n.nnn). ;
06: Z=X*F (P37)
1: 3* X Loc [ Rn_W_m2 ] 2: 3: 3* Z Loc [ Rn_W_m2 ]
07: End (P95)
n.nnn
F

3.3 PROGRAMMING EXAMPLE WITH DYNAMIC WIND SPEED CORRECTION

The following CR10(X) program measures an R. M. Young wind sentry anemometer model 03101 and a Q-7.1. The wind speed is used to give a real-time wind correction to net radiation. The wind speed must be measured before the wind corrections are applied.
Example 2. CR10(X) Program with the
Dynamic Wind Speed Correction
;Measure the 03101 wind sentry. ;
01: Pulse (P3)
1: 1 Reps 2: 1* Pulse Channel 1 3: 21 Low Level AC, Output Hz 4: 1* Loc [ ws_m_s ] 5: .75 Mult 6: .2 Offset
02: IF (X<=>F) (P89)
1: 1* X Loc [ ws_m_s ] 2: 1 = 3: .2 F 4: 30 Then Do
03: Z=F (P30)
1: 0 F 2: 0 Exponent of 10 3: 1* Z Loc [ ws_m_s ]
04: End (P95)
;Measure the Q7.1. ;
05: Volt (Diff) (P2)***
1: 1 Reps 2: 24** ± 250 mV 60 Hz
Rejection Range 3: 1* DIFF Channel 4: 2* Loc [ Rn_mV ] 5: 1 Mult 6: 0 Offset
* Proper entries will vary with program, and
datalogger channel and input location assignments.
** On the 21X and CR7 use 500 mV input
range.
*** For a Single-Ended measurement use
Instruction 1.
4
06: IF (X<=>F) (P89)
1: 2* X Loc [ Rn_mV ] 2: 3 >= 3: 0 F 4: 30 Then Do
;Apply the positive calibration and ;wind speed correction. ;
Q-7.1 NET RADIOMETER
07: Do (P86)
1: 1* Call Subroutine 1*
08: Else (P94)
;Apply the negative calibration and ;wind speed correction. ;
09: Do (P86)
1: 2* Call Subroutine 2* 10: End (P95) *Table 3 Subroutines
;Positive calibration and wind ;speed correction. ;
01: Beginning of Subroutine (P85)
1: 1* Subroutine 1 02: Z=X*F (P37)
1: 1* X Loc [ ws_m_s ]
2: .2 F
3: 6* Z Loc [ C ] 03: Z=X*F (P37)
1: 6* X Loc [ C ]
2: .066 F
3: 4* Z Loc [ A ] 04: Z=X+F (P34)
1: 6* X Loc [ C ]
2: .066 F
3: 5* Z Loc [ B ]
09: End (P95)
;Negative calibration and wind ;speed corrections ;
10: Beginning of Subroutine (P85)
1: 2* Subroutine 2*
11: Z=X*F (P37)
1: 1* X Loc [ ws_m_s ] 2: .00174 F 3: 4* Z Loc [ A ]
12: Z=X+F (P34)
1: 4* X Loc [ A ] 2: .99755 F 3: 7* Z Loc [ Corr_Fact ]
;Enter the negative multiplier (n.nnn). ;
13: Z=X*F (P37)
1: 2* X Loc [ Rn_mV ] 2: 3: 3* Z Loc [ Rn_W_m2 ]
14: Z=X*Y (P36)
1: 3* X Loc [ Rn_W_m2 ] 2: 7* Y Loc [ Corr_Fact ]
3: 3* Z Loc [ Rn_W_m2 ] 15: End (P95) * Proper entries will vary with program, and
datalogger channel and input location assignments.
n.nnn
F
05: Z=X/Y (P38)
1: 4* X Loc [ A ] 2: 5* Y Loc [ B ] 3: 7* Z Loc [ Corr_Fact ]
06: Z=Z+1 (P32)
1: 7* Z Loc [ Corr_Fact ]
;Enter the positive multiplier (p.ppp). ;
07: Z=X*F (P37)
1: 2* X Loc [ Rn_mV ] 2: 3: 3* Z Loc [ Rn_W_m2 ]
08: Z=X*Y (P36)
1: 3* X Loc [ Rn_W_m2 ] 2: 7* Y Loc [ Corr_Fact ] 3: 3* Z Loc [ Rn_W_m2 ]
p.ppp
F
** On the 21X and CR7 use 500 mV input
range.
***For a Single-Ended measurement use
Instruction 1.

4. MAINTENANCE

4.1 DESICCANT

The Q-7.1 is supplied with heavy duty polyethylene windshields (approximately 0.25 mm thick); no pressurization is required.
Air spaces inside the windshields are connected to a dryer filled with silica gel (referred to as the desiccant tube) to prevent internal condensation. The desiccant tube is located in the support arm and is accessible by removing the end plug and vinal cap.
5
Q-7.1 NET RADIOMETER
Inspect the silica gel monthly to ensure it is still blue and white. If the color changes to pink and white, replace it with dry silica gel (this may have to be done more frequently in wet weather). Wet (pink and white) silica gel can be dried by removing it from the desiccant tube and baking at 130°C until it returns to a blue and white color. Remove the desiccant tube end cap to remove and replace the silica gel.
CAUTION: After the desiccant tube has been replaced, install the end plug and vinal cap. Be sure the holes in the vinal cap point towards the ground.
If the Q-7.1 is installed at a remote site it may be convenient to have a spare desiccant tube for quick replacement (be sure to remove the outer end caps from both tube ends before installing). Spare desiccant tubes are available from Radiation Energy Balance Systems.

4.2 CONDENSATION

Condensation on net radiometer windshields causes incorrect measurements. This is because water does not transmit longwave energy. An example of this problem can be seen by comparing measured net radiation values from two different net radiometers at night, one on which dew is allowed to form and another on which dew is prevented. Without dew both instruments would indicate a similar
-2
net radiation level, e.g. -50 W m
. However, on the instrument on which dew was allowed to form a net radiation level of about zero would be indicated, while the other radiometer would
-2
maintain the reading of -50 W m
. The RV2 ventilator can be used to prevent dew from forming on the windshields.
4.2.1 Causes of Internal Condensation
If condensation develops inside the domes check the following:
Desiccant
Make sure that the desiccant is dry (i.e. still blue and white).
If it has turned pink, replace it with dry desiccant or bake it until it is dry (see section
4.1 above). If the desiccant is pink only at the tube end nearest the sensor head, this indicates a leak somewhere on the sensor.
Windshields and O-rings
Check that the windshields and O-rings are in good condition and are properly seated against the radiometer frame (see below).
If there is still no obvious reason for the internal condensation, check for possible leaks in the sensor as follows:
1. Remove the desiccant tube.
2. Immerse the radiometer in water and blow gently into the open end of the support arm. A stream of bubbles will indicate the location of any leaks.
3. Dry the instrument with a soft facial tissue. If the windshields get dented, reinflate them by blowing into the end of the support arm.
CAUTION: When drying the windshields, dabbing rather than wiping them will help to prevent scratches.
4.2.2 Possible Internal Condensation
Problems After Installation
There is a small possibility that condensation may occur on the inside of the windshields when the Q-7.1 is first installed in the field. This condensation is caused by evaporation of hydrocarbon solvents used in the manufacturing process and may be more apparent after several hot, sunny days if ambient temperatures are low enough.
If you notice condensation inside the windshields after the first few days of sunshine following installation, remove the windshields as described below. Allow the solvent vapor or condensation to escape or evaporate from the inner air space for 15 minutes then replace the windshields.

4.3 CLEANING

The windshields may be washed with a camel hair brush or with a paper tissue and distilled water.
CAUTION: Coarse paper or cloth will scratch the windshields and should be avoided.
6
Q-7.1 NET RADIOMETER

4.4 WINDSHIELD REPLACEMENT

Polyethylene deteriorates with exposure to solar radiation. Inspect the windshields frequently and replace as required (probably every 3-6 months).
Always check for:
1. Condensation inside the windshields, possibly indicating a leak.
2. Cracking or crazing of the windshields (which usually appears first along the base).
Prompt replacement is essential if any of these conditions occur. Also, if the windshields are removed for any reason after more than one month’s use, replace them with new ones. The windshields should always be replaced as a set. Windshield replacement may cause a slight change in calibration, although this is unlikely to exceed 3%.
CAUTION: There are very fine wires inside the head of the Q-7.1 which are very susceptible to corrosion if subjected to moisture, e.g. standing water. It is essential to carry out the preemptive maintenance mentioned above to prevent serious damage to the sensor.
Replace the windshields as follows:
1. Remove the mounting screws and clamping rings.
2. Replacement windshields are pre-cut and punched with holes. Make sure the O-rings are in their grooves before clamping the new windshields into place as described in step
3.
3. Gently tighten the screws in the following sequence: first, third, fifth, second, fourth, sixth. It does not matter which screw you treat as the first. Continue the sequence until the windshields are evenly in contact with the clamping rings and frame.
4. Tighten the screws in the same sequence until no space remains between either windshield flange and its corresponding mounting ring or frame surface. Make sure both upper and lower windshields are completely seated on the frame around the entire periphery.
CAUTION: Do not attempt to dismantle the instrument body. If internal attention is needed, contact Campbell Scientific, Inc. for repair.
If the windshields become dented, remove the dents as follows:
1. Remove the vinal cap from the support arm.
2. Blow gently into the end plug breather hole to inflate the windshields.
Remember to replace the support arm vinal cap. Orient the vinal cap so the holes point towards the ground.
CAUTION: When replacing the windshields, take great care not to touch or mark the black surfaces of the thermopile as this may affect the performance of the instrument. Take care also to avoid overtightening the mounting screws.
4.4.1 Bird Damage
One other cause of damage to windshields is birds pecking at the domes. A simple measure to prevent this is to attach two or three 8 inches cable ties to the support arm at the end near the sensor head, with the free ends of the ties pointing upwards to form an array of ‘spikes’. This may discourage birds from landing on the arm. If this does not work you may have to resort to more drastic techniques such as installing an artificial snake on the support arm or commercial bird scarers.

4.5 CALIBRATION

Q-7.1 net radiometers show excellent stability over long periods, but for serious use, regular calibration checks are recommended. The calibration can be checked in the field by mounting another net radiometer alongside and comparing results. The instruments should be examined carefully if the results differ by more than a few percent. Under normal circumstances, a calibration check every six months is adequate providing the sensitive thermopile surface has not been marked, damaged or exposed to the weather.
If the calibration can not be checked in the field, Radiation Energy Balance Systems
7
Q-7.1 NET RADIOMETER
recommends recalibration every six months, depending on the climate where the sensors are used.
Each instrument is calibrated by the manufacturer in a temperature-controlled calibration chamber by comparison to a transfer standard. The chamber uses a combination of tungsten-halide source and black body chambers. The transfer standard is calibrated by comparison to a precision pyranometer using a partial shading technique developed by Radiation Energy Balance Systems. The long wave calibration of the standard has also been verified using black body chambers.
Care should, however, be taken when comparing readings with sensors from different manufacturers or even older models from Radiation Energy Balance Systems. This is because in recent years academic studies have reported inadequacies both in the design and calibration of net radiometers from most manufacturers. Radiation Energy Balance Systems aim to keep at the forefront of such developments and have published corrections for the calibration of older sensors. However, these corrections cannot fully bring the instruments to the same level of performance as the new sensors. It is, however, possible to upgrade and recalibrate the sensor heads of older models.
8
This is a bla nk page.
Campbell Scientific Companies
Campbell Scientific, Inc. (CSI)
815 West 1800 North
Logan, Utah 84321
UNITED STATES
www.campbellsci.com
info@campbellsci.com
Campbell Scientific Africa Pty. Ltd. (CSAf)
PO Box 2450
Somerset West 7129
SOUTH AFRICA
www.csafrica.co.za
sales@csafrica.co.za
Campbell Scientific Australia Pty. Ltd. (CSA)
PO Box 444
Thuringo wa Cent ra l
QLD 4812 AUSTRALIA
www.campbellsci.com.au
info@campbellsci.com.au
Campbell Scientific do Brazil Ltda . (CSB)
Rua Luisa Crapsi Orsi, 15 Butantã
CEP: 005543-000 São Paulo SP BRAZIL
www.campbellsci.com.br
suporte@campbellsci.com.br
Campbell Scientific Canada Corp. (CSC)
11564 - 149th Street NW
Edmonton, Alberta T5M 1W7
CANADA
www.campbellsci.ca
dataloggers@campbellsci.ca
Campbell Scientific Ltd. (CSL)
Campbell Park
80 Hathern Road
Shepshed, Loughborough LE12 9GX
UNITED KINGDOM
www.campbellsci.co.uk
sales@campbellsci.co.uk
Campbell Scientific Ltd. (France)
Miniparc du Verger - Bat. H
1, rue de Terre Neuve - Les Ulis
91967 COURTABOEUF CEDEX
FRANCE
www.campbellsci.fr
campbell.scientific@wanadoo.fr
Campbell Scientific Spain, S. L.
Psg. Font 14, local 8
08013 Barcelona
SPAIN www.campbellsci.es info@campbellsci.es
Please visit www.campbellsci.com to obtain contact information for your local US or International representative.
Loading...