Campbell Scientific KH20 User Manual

KH20
Revision:
Krypton Hygrometer
05/2020
2010 – 2020

Table of Contents

PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to specific sections.
1. Introduction................................................................ 1
2. Precautions ................................................................ 1
3. Initial Inspection ........................................................ 1
3.1 Components .........................................................................................1
4. Overview .................................................................... 1
5. Specifications ............................................................ 2
5.1 Measurements ......................................................................................2
5.2 Electrical ..............................................................................................2
5.3 Physical ................................................................................................2
6. Installation ................................................................. 3
6.1 Siting ....................................................................................................3
6.2 Mounting ..............................................................................................3
6.2.1 Parts and Tools Needed for Mounting ..........................................3
6.2.2 Mounting the KH20 Sensor ..........................................................3
6.2.3 Mounting the Electronics Box ......................................................4
6.3 Wiring ..................................................................................................6
6.4 Data Logger Programming ...................................................................7
6.4.1 KH20 Calibration ..........................................................................7
7. Maintenance and Calibration .................................... 7
7.1 Visual Inspection ..................................................................................8
7.2 Testing the Source Tube .......................................................................8
7.3 Testing the Detector Tube ....................................................................8
7.4 Managing the Scaling of KH20 ............................................................8
7.5 Calibration ............................................................................................9
Appendices
A.
Calibrating KH20 .................................................... A-1
A.1 Basic Measurement Theory ............................................................. A-1
A.2 Calibration of KH20 ........................................................................ A-1
B. Example Program .................................................. B-1
i
Table of Contents
C. EasyFlux® DL CR6KH20 ........................................ C-1
C.1 Introduction ...................................................................................... C-1
C.2 Precautions ....................................................................................... C-2
C.3 Wiring .............................................................................................. C-2
C.3.1 Required Sensors ...................................................................... C-3
C.3.2 Optional Sensors ....................................................................... C-3
C.3.2.1 VOLT116 Module .......................................................... C-3
C.3.2.2 Barometer ....................................................................... C-4
C.3.2.3 Fine Wire Thermocouple ................................................ C-4
C.3.2.4 GPS Receiver ................................................................. C-5
C.3.2.5 Radiation Measurements Option 1 ................................. C-5
C.3.2.6 Radiation Measurements Option 2 ................................. C-6
C.2.3.7 Precipitation Gage .......................................................... C-8
C.2.3.8 Soil Temperature ............................................................ C-9
C.2.3.9 Soil Water Content ......................................................... C-9
C.3.2.10 Soil Heat Flux Plates .................................................... C-10
C.4 Operation ........................................................................................ C-12
C.4.1 Set Constants in CRBasic Editor and Load Program .............. C-12
C.4.2 Enter Site-Specific Variables with Data Logger Keypad or
LoggerNet ............................................................................ C-14
C.4.3 Data Retrieval ......................................................................... C-19
C.4.4 Output Tables .......................................................................... C-19
C.4.5 Program Sequence of Measurement and Corrections ............. C-36
C.5 References ...................................................................................... C-37
D. Equations and Algorithms of Water Vapor
Density and Water Flux in KH20 Eddy-
Covariance Systems ........................................... D-1
D.1 Fundamental Equation .................................................................... D-1
D.2 Working Equation ........................................................................... D-2
D.3 Eddy-Covariance Water Flux .......................................................... D-2
D.4 Frequency Response of KH20 ......................................................... D-3
D.5 References ....................................................................................... D-4
Figures
6-1. Mounting KH20 to a tripod. .................................................................4
6-2. Proper mounting position of the electronics box. .................................5
6-3. Attaching cables to the electronics box. ...............................................6
A-1. KH20 ln(mV) vs. Vapor Density .................................................... A-2
C-1. Example screen from CRBasic Editor showing user-defined
configuration constants ............................................................... C-13
C-2. Custom keypad menu; arrows indicate submenus .......................... C-14
Tables
6-1. Wire Color, Function, and Data Logger Connection............................6
6-2. KH20 Calibration Ranges ....................................................................7
A-1. Linear Regression Results for KH20 ln(mV) vs. Vapor Density .... A-2
A-2. Final Calibration Values for KH20 ................................................. A-3
C-1. Default Wiring for Required Sensors ............................................... C-3
C-2. Default Wiring for CS106 Barometer .............................................. C-4
C-3. Default Wiring for Fine Wire Thermocouple ................................... C-4
ii
C-4. Default Wiring for GPS Receiver .................................................... C-5
C-5. Default Wiring for Radiation Measurement Option 1 ...................... C-5
C-6. Default Wiring for Radiation Measurements Option 2 .................... C-6
C-7. A21REL-12 Wiring .......................................................................... C-8
C-8. CNF4 Wiring .................................................................................... C-8
C-9. Default Wiring for Precipitation Gage ............................................. C-9
C-10. Default Wiring for Soil Thermocouple Probes ................................ C-9
C-11. Default Wiring for Soil Water Content Probes .............................. C-10
C-12. Default Wiring for Non-Calibrating Soil Heat Flux Plates ............ C-10
C-13. Default Wiring for Soil Heat Flux Plates (Self Calibrating) .......... C-11
C-14. Station Variables with Descriptions ............................................... C-15
C-15. microSD Flash Card Fill Times with 10Hz Measurement Rate ..... C-19
C-16. Data Output Tables ........................................................................ C-20
C-17. Data Fields in the Time_Series Data Output Table ........................ C-21
C-18. Data Fields in the Diagnostic Output Table ................................... C-22
C-19. Data Fields in the Monitor_CSAT3B Output Table ....................... C-22
C-20. Data Fields in the Flux_AmeriFluxFormat Output Table .............. C-22
C-21. Data Fields in the Flux_CSFormat Data Output Table .................. C-25
C-22. Data Fields in the Flux_Notes Output Table .................................. C-29
CRBasic Example
B-1. CR3000 Program to Measure Water Vapor Fluctuations ................. B-1
Table of Contents
iii
KH20 Krypton Hygrometer

1. Introduction

The KH20 is a highly sensitive hygrometer designed for measurement of rapid fluctuations in atmospheric water vapor, not absolute concentrations. It is typically used together with a CSAT3B in eddy-covariance systems.

2. Precautions

READ AND UNDERSTAND the Safety section at the back of this
manual.
Although the KH20 is rugged, it should be handled as precision scientific
instrument.

3. Initial Inspection

Upon receipt of the KH20, inspect the packaging and contents for damage.
The model number and cable length are printed on a label at the

3.1 Components

The KH20 sensor consist of a sensor head with 2 m (6 ft) cables and an electronics box. The following are also shipped with the KH20:
File damage claims with the shipping company.
connection end of the cable. Check this information against the shipping documents to ensure the correct product and cable length are received (see Section 3.1, Components
KH20CBL-L25 Power/Signal cable with 8 m (25 ft) length. If a
longer cable is desired, order a KH20CBL-L replacement cable and specify the desired length after -L (for example KH20CBL-L50).
1/2 Unit Desiccant Bag
Rain Shield
Horizontal Mounting Boom (51 cm 20 mm DN (20-inch 3/4 IPS)
threaded aluminum pipe)
3/4 x 3/4 in. Nu-Rail Crossover Fitting
4 mm (5/32 in) Allen Wrench
(p. 1)).

4. Overview

The KH20 is a krypton hygrometer for measuring water vapor fluctuations in the air. The name KH20 (KH-twenty) was derived from KH2O (K-H the sensor has been known with this name since 1985. It is typically used with
O), and
2
1
KH20 Krypton Hygrometer
NOTE
the CSAT3B 3-D sonic anemometer for measuring latent heat flux (LE), using eddy-covariance technique.
The KH20 sensor uses a krypton lamp that emits two absorption lines: major line at 123.58 nm and minor line at 116.49 nm. Both lines are absorbed by water vapor, and a small amount of the minor line is absorbed by oxygen. The KH20 is not suitable for absolute water vapor concentration measurements due to its signal offset drift.
The KH20 heads are sealed and will not suffer damage should they get wet. In addition, the electronics box and the connectors are housed inside a rain shield that protects them from moisture. The KH20 is suitable for long-term continuous outdoor applications.
The KH20 sensor is comprised of two main parts: the sensor head and the electronics box. The sensor head comes with cables that connect the sensor to the electronics, a power/signal cable, and mounting hardware.
Discussion on the principles and theory of measurement is included in Appendix A, Calibrating KH20 (p. A-1).
Features:

5. Specifications

5.1 Measurements

Calibration Range: 1.7 to 19.5 g/m3 (nominal) Frequency Response: 100 Hz Operating Temperature Range: –30 to 50 °C

5.2 Electrical

Supply Voltage: 10 V to 16 VDC Current Consumption: 20 mA max at 12 VDC Power Consumption: 0.24 Watts Output Signal Range: 0 to 5 VDC
High frequency response suitable for eddy-covariance applications
Well-suited for long-term, unattended applications
Compatible with Campbell Scientific CRBasic data loggers: CR6,
CR3000, CR1000X, CR800 Series, CR1000, CR5000, and CR9000(X)

5.3 Physical

Dimensions
Sensor Head: 29 x 23 x 3 cm (11.5 x 9 x 1.25 in) Electronics Box: 19 x 13 x 5 cm (7.5 x 5 x 2 in) Rain Shield with Mount: 29 x 18 x 6.5 cm (11.5 x 7 x 2.5 in) Mounting Pipe: 50 cm (20 in) Carrying Case: 64 x 38 x 18 cm (25 x 15 x 7 in)
2

6. Installation

6.1 Siting

6.2 Mounting

6.2.1 Parts and Tools Needed for Mounting

KH20 Krypton Hygrometer
Weight
Sensor Head: 1.61 kg (3.55 lb) Electronics Box: 0.6 kg (1.4 lb) Rain Shield with Mount: 2.2 kg (4.75 lb) Mounting Pipe with Nu-rail: 0.45 kg (1.0 lb) Carrying Case: 4.3 kg (9.45 lb) Shipping: 9.2 kg (20.15 lb)
When installing the KH20 sensor for latent heat flux measurement in an eddy­covariance application, proper siting, sensor height, sensor orientation and fetch are important.
The following user-supplied hardware is required to mount the KH20 sensor:
1. Tripod (CM115 standard) or tower
2. Campbell Scientific crossarm (CM204 standard)
3. 3/4-inch IPS Aluminum Pipe, 12 inches long
4. 3/4-inch-by-1-inch Nu-Rail Crossover Fitting
5. Small Phillips and flat-head screwdrivers
6. 1/2-inch wrench

6.2.2 Mounting the KH20 Sensor

Mount the KH20 sensor head as follows:
1. Attach the 51 cm (20 in) mounting boom to the KH20.
2. Mount a crossarm to a tripod or tower.
3. Mount the 12-inch-long pipe to a crossarm via 1-inch-by-3/4-inch Nu-Rail
Crossover Fitting.
4. Mount the KH20 onto the 30 cm (12 in) pipe using a 3/4-inch-by-3/4-inch
Nu-Rail Crossover Fitting. Mount the KH20 such that the source tube, the longer of the two tubes, is positioned on top, as shown in FIGURE 6-1. Use cable ties to secure loose cables to the tripod or tower mast.
3
FIGURE 6-1. Mounting KH20 to a tripod.
NOTE
KH20 Krypton Hygrometer

6.2.3 Mounting the Electronics Box

Mount the electronics box as follows:
1. Remove the front cover of the rain shield by loosening the two pan-head
screws on the bottom front of the rain shield, and then pushing the cover all the way up, and sliding it out.
It will be difficult to mount the rain shield to a mast with the front cover on, since the 1/2-inch nut holding the bottom U-bolt is located inside the rain shield.
2. Before mounting the rain shield to a tripod, first mount the electronics box
inside the rain shield. Remove the four pan-head screws from the back panel of the rain shield. Align the electronics box and use the four pan­head screws to secure the electronics box onto the back panel. Make sure the electronics box is pushed all the way up, and the screws are positioned at the bottom of the mounting slot on the electronics box (see FIGURE
6-2). This will provide enough room to attach the connectors to the bottom
of the electronics box later.
4
KH20 Krypton Hygrometer
NOTE
FIGURE 6-2. Proper mounting position of the electronics box.
If the electronics box is not pushed all the way up during mounting, you will not have enough room to attach the connectors to the bottom of the electronics box, as the U-bolt for the rain shield will block the position of the connectors.
3. Mount the rain shield onto the tripod or tower mast using the U-bolt
provided. Make sure that the distance between the KH20 sensor head and the rain shield is within 5 feet so that the cables from the sensor head will be within reach of the electronics box. Also make sure that the rain shield is mounted vertically with an opening pointing downward so that the rain will effectively run down the rain shield and not penetrate inside.
5
KH20 Krypton Hygrometer
TABLE 6-1. Wire Color, Function, and Data Logger Connection
4. Connect the three cables to the bottom of the electronics box around the
U-bolt on the rain shield (see FIGURE 6-3). If there is not enough room for the connectors around the U-bolt, make sure the electronics box is mounted at a highest possible position (see step 2).

6.3 Wiring

FIGURE 6-3. Attaching cables to the electronics box.
5. Place the front cover back on the rain shield and tighten the two pan-head
screws to secure it in place.
6. Gather any loose cables and tie them up, using cable ties, onto the tripod or
tower mast.
Wire Wire Label Data Logger Connection Terminal
1, 2
1
,
,
White Signal
Black (from
white/black set)
Signal
Reference
U configured for differential input
DIFF H (differential high, analog-
voltage input)
U configured for differential input
DIFF L (differential low, analog-
2
voltage input)
Red Power 12V 12V
Black (from
red/black set)
Clear Shield
1
U terminals are automatically configured by the measurement instruction.
2
Jumper to ⏚ with a user-supplied wire.
Power
Ground
G
(analog ground)
6

6.4 Data Logger Programming

TABLE 6-2. KH20 Calibration Ranges
The KH20 sensor outputs 0 to 5 VDC analog signal. These signals can be measured using the VoltDiff instruction on the CRBasic data loggers.
Programming basics for CRBasic data loggers are in the following sections. Complete program examples for select CRBasic data loggers can be found in Appendix B, Example Program

6.4.1 KH20 Calibration

Each KH20 is calibrated over a vapor range of approximately 2 to 19 g/m3. The calibration is performed twice under the following two conditions: window clean, and scaled. The water vapor absorption coefficient for three different vapor ranges are calculated from the collected calibration data: full range, dry range, and wet range. TABLE 6-2 shows a sample of the KH20 vapor ranges over which three different water vapor absorption coefficients are calculated. See Appendix A, Calibrating KH20 calibration.
KH20 Krypton Hygrometer
(p. B-1).
(p. A-1), for more information on KH20
Ranges Vapor Density (g/m3)
Full Vapor Range 2 – 19
Dry Vapor Range 2 – 9.5
Wet Vapor Range 8.25 – 19
Before the water vapor absorption coefficient, k logger program for the KH20, the following decisions must be made:
Will the windows be allowed to scale?
What vapor range is appropriate for the site?
Once the decision is made, the appropriate k calibrations sheet. The calibration sheet also contains the path length, x, for a specific KH20. Using the water vapor absorption coefficient for either the dry or the wet vapor range will produce more accurate measurements than using that for the full range. If the vapor range of the site is unknown, or if the vapor range is on the border line between the dry and the wet vapor ranges, the full range should be used.

7. Maintenance and Calibration

The KH20 sensor is designed for continuous field application and requires little maintenance. The tube ends for the KH20 have been sealed with silicone elastomer using an injection-mold method. Therefore, the tubes are protected from water damage, and the KH20 continues to make measurements under rainy or wet conditions. If the water tends to pool up on the tube window and blocks the signal, turn the sensor head at an angle so as to shed the water off the tube window. The rain shield protects the electronics box and the connectors from moisture.
, is entered into the data
w
can be chosen from the
w
7

7.1 Visual Inspection

NOTE
Make sure the optical windows are clean.
Inspect the cables and connectors for any damage or corrosion. If you see
a discoloration on the white co-axial cable, you may suspect that the cable has water damage.

7.2 Testing the Source Tube

The source tube is the longer of the two tubes. Check to see if the source tube is working properly by performing the following test.
First, make sure the UV light is emitted from the source tube. To do this, you may look into the source tube (the longer of the two tubes), and you should see a bright blue light emitted from it.
Avoid looking into the source tube for an extended period of time when the KH20 is powered on to minimize the prolonged exposure to the UV light.
KH20 Krypton Hygrometer
If you see a faint or flickering blue light, perform the following test.
Check the current drain on the KH20
Typical current drain for the KH20 during normal operation should be 15 ~ 20 mA. The current drain of around 5 mA or less indicates the problem on the source tube. Obtain an RMA from Campbell Scientific and send the unit in for repair.
Check the voltage signal output from the KH20
If the voltage output reading is below 50 mV, you may have problems with either the source tube or the detector tube (Section 7.3, Testing the Detector Tube
(p. 8)).

7.3 Testing the Detector Tube

If the source tube tests fine but the output from KH20 is still in question, perform the following test. Prepare a piece of paper and insert it between the source tube and the detector tube to completely block the optical path. You should see an immediate decrease in the voltage reading, and it should go close to zero. No noticeable change in the voltage output, when the optical path is completely blocked, indicates a problem in the detector tube. If the decrease in the voltage reading takes place but the reading remains below 50 mV, when the paper is removed from the optical path, the source tube may be at fault. Obtain an RMA from Campbell Scientific and send the unit in for repair.

7.4 Managing the Scaling of KH20

The KH20 cannot be used to measure an absolute concentration of water vapor, because of scaling on the source tube windows caused by disassociation of atmospheric continuants by the ultra violet photons (Campbell and Tanner,
8
KH20 Krypton Hygrometer
NOTE
1985 and Buck, 1976). The rate of scaling is a function of the atmospheric humidity. In a high humidity environment, scaling can occur within a few hours. That scaling attenuates the signal and can cause shifts in the calibration curve. However, the scaling over a typical flux averaging period is small. Thus, water vapor fluctuation measurements can still be made with the hygrometer.
To see if the source tube window has been scaled, get a clean, dry cotton swab and slide it across the source tube window. The scale is not visible to the naked eye, but if the window is scaled, you will feel a slight but noticeable resistance while you slide the swab across the window. There will be little resistance if the window is not scaled. If you determine the window is scaled, you can clean it with a wet cotton swab.
Use distilled water and a clean cotton swab to clean the scaled window. After cleaning the window, slide a clean, dry swab across the window to confirm the scale has been removed.
You can use the water vapor absorption coefficient for scaled window from the calibration sheet if the window will be allowed to scale during measurements.

7.5 Calibration

For quality assurance of the measured data, Campbell Scientific recommends the KH20 be recalibrated every two years. Calibrations require a returned material authorization (RMA) and completion of the “Declaration of Hazardous Material and Decontamination” form. Refer to the Assistance page at the end of this manual for more information.
For more information on the calibration process, refer to Appendix A, Calibrating KH20
(p. A-1).
9
ww
xk
eT
ρ
=
ww
xk
e
V
V
ρ
=
0
)ln(ln
1
0
VV
xk
w
w
=
ρ
0
lnln VxkV
ww
+=
ρ

Appendix A. Calibrating KH20

A.1 Basic Measurement Theory

The KH20 uses an empirical relationship between the absorption of the light and the material through which the light travels. This relationship is known as the Beer’s law, the Beer-Lambert law, or the Lambert-Beer law. According to the Beer’s law, the log of the transmissivity is anti-proportional to the product of the absorption coefficient of the material, k, the distance the light travels, x, and the density of the absorbing material, ρ. The KH20 sensor uses the UV light emitted by the krypton lamp: major line at 123.58 nm and the minor line at 116.49. As the light travels through the air, both the major line and the minor line are absorbed by the water vapor present in the light path. This relationship can be rewritten as follows, where k vapor, x is the path length for the KH20 sensor, and ρ density.
A-1
is the absorption coefficient for water
w
is the water vapor
w
If we express the transmissivity, T, in terms of the light intensity before and after passing through the material as measured by the KH20 sensor, V and V respectively, we obtain the following equation.
Taking the natural log of both sides, and solving for the density, ρ following equation.
If the path length, x, and the absorption coefficient for water, k becomes possible to measure the water vapor density ρ signal output, V, from KH20.

A.2 Calibration of KH20

The KH20 calibration process is to find the absorption coefficient of water vapor, k
. To do this, we rewrite the equation A-3, and solve for ln(V).
w
,
0
A-2
, yields the
w
A-3
are known, it
w
, by measuring the
w
A-4
It now becomes obvious from the equation A-4 that there is a linear relationship between the natural log of the KH20 measurement output, lnV, and the water vapor density, ρ after we ran a KH20 over a full calibration vapor range.
. FIGURE A-1 shows the plot of the equation A-4
w
A-1
Appendix A. Calibrating KH20
TABLE A-1. Linear Regression Results for KH20 ln(mV)
KH2O Output (mV)
Vapor Density (g/m3)
8
7.5
7
6.5
6
5.5
5
4.5
4
1.74 3.02 4.17 5.44 6.71 7.95 9.2 10.47 11.69 12.9 14.22 15.46 16.78 18.04 19.25
FIGURE A-1. KH20 ln(mV) vs. Vapor Density
We can perform the linear regression on the plot to obtain the slope for the relationship between the ln(mV) and the vapor density. The slope for the graph is the coefficient, k
x. TABLE A-1 shows the result of linear regression
w
analysis. The slope is the product of the absorption coefficient of water vapor,
, and the KH20 path length, x.
k
w
vs. Vapor Density
Description Values
Slope (xkw) –0.205
Y Intercept (ln(V0) 8.033
If we substitute these values, along with the measured lnV into equation A-3, we can obtain the water vapor density, ρ
. Campbell Scientific performs the
w
calibration twice for each KH20: once with the window cleaned, and again with the window scaled. We then break up the vapor density range into dry and wet ranges, and compute the k range. If you know the vapor density range for your site, it is recommended that you select the coefficient, k
values for each sub range, as well as for the full
w
, that is appropriate for your site, the dry range
w
or the wet range. If the vapor range for the site is unknown, or if the vapor range is on the border line between the dry and the wet ranges, use the value for the full range. TABLE A-2 shows the final calibration values the KH20 calibration certificate contains. The data shown in TABLE A-2 is from an actual KH20.
A-2
Appendix A. Calibrating KH20
TABLE A-2. Final Calibration Values for KH20
Vapor Range
3
)
(g/m
Slope (xkw)
Y Intercept
ln (V0)
Coefficient
(kw)
Full Range 1.74 ~ 19.25 -0.205 3087 -0.144
Dry Range 1.74 ~ 9.20 -0.216 3259 -0.151
Wet Range 7.95 ~ 19.25 -0.201 2899 -0.141
A-3

Appendix B. Example Program

CRBasic Example B-1. CR3000 Program to Measure Water Vapor Fluctuations
'CR3000 Series Data Logger
Units kh_mV = mV
NOTE
'This data logger program measures KH20 Krypton Hygrometer.
'The station operator must enter the constant and the calibration value for the KH20. 'Search for the text string "unique" to find the locations of these constants 'and enter the appropriate values found from the calibration sheet of the KH20.
'*** Unit Definitions ***
'Units Description 'ln_mV ln(mV) (natural log of the KH20 millivolts) 'mV millivolts 'rho_w g/m^3
'*** Wiring ***
'ANALOG INPUT '1H KH20 signal+ (white) '1L KH20 signal- (black) 'gnd KH20 shield (clear)
'EXTERNAL POWER SUPPLY 'POS KH20 power+ (red) ' data logger POWER IN 12 (red) 'NEG KH20 power- (black) ' KH20 power shield (clear) ' data logger POWER IN G (black)
PipeLineMode
'*** Constants ***
'Measurement Rate '10 Hz
Const SCAN_INTERVAL = 100 '100 mSec
'Output period
Const OUTPUT_INTERVAL = 30 'Online flux data output interval in minutes. Const x = 1 'Unique path length of the KH20 [cm]. Const kw = -0.150 'Unique water vapor absorption coefficient [m^3 / (g cm)]. Const xkw = x*kw 'Path length times water vapor absorption coefficient [m^3 / g].
'*** Variables ***
Public panel_temp Public batt_volt Public kh(2) Public rho_w Alias kh(1) = kh_mV Alias kh (2) = ln_kh Units panel_temp = deg_C Units batt_volt = volts
The following example program measures the KH20 at 10Hz, and stores the average values into a data table called ‘stats’, as well as the raw data into a data table called ‘ts_data’.
The KH20 does not monitor absolute water vapor concentration.
B-1
Units ln_kh = ln_mV
EndProg
Appendix B. Example Program
Units rho_w = g/m^3
'*** Data Output Tables *** 'Processed data
DataTable (stats,True,-1)
DataInterval (0,OUTPUT_INTERVAL,Min,10) Minimum (1,batt_volt,FP2,False,False) Average (1,panel_temp,FP2,False) Average (2,kh(1),IEEE4,False)
EndTable
'Raw time-series data.
DataTable (ts_data,True,-1)
DataInterval (0,SCAN_INTERVAL,mSec,100) Sample (1,kh_mV,IEEE4)
EndTable
'*** Program ***
BeginProg
Scan (SCAN_INTERVAL,mSec,3,0)
'data logger panel temperature. PanelTemp (panel_temp,250)
'Measure battery voltage. Battery (batt_volt)
'Measure KH20. VoltDiff (kh_mV,1,mV5000,1,TRUE,200,250,1,0) ln_kh = LOG(kh_mV) rho_w = ln_kh/xkw
CallTable stats CallTable ts_data
NextScan
B-2
NOTE
EasyFlux is

Appendix C. EasyFlux® DL CR6KH20

C.1 Introduction

EasyFlux® DL CR6KH20 is a CRBasic program that enables a CR6 data logger, along with a KH20 and CSAT3B, to collect fully corrected fluxes of latent heat (H EC data using commonly used corrections in the scientific literature. The program can also calculate the ground surface heat flux and energy closure by adding an optional suite of energy balance sensors. Because the energy balance sensors require more analog terminals than the CR6 has, the program supports the addition of a VOLT116 (or CDM-A116) analog terminal expansion module.
Specifically, the program supports data collection and processing from the following sensors.
REQUIRED SENSORS:
KH20 Krypton Hygrometer (qty 1) CSAT3B Sonic Anemometer (qty 1) Temperature/Relative Humidity (RH) Probe (qty 1). Supported Models:
OPTIONAL SENSORS:
CS106 Barometer (qty 0 to 1) FW3 Fine Wire Thermocouple (qty 0 to 1) GPS16X-HVS GPS Receiver (qty 0 to 1) Radiation measurements
TE525MM Rain Gage (qty 0 to 1) TCAV Soil Thermocouple Probe (qty 0 to 3) Soil Water Content Reflectometer (qty 0 to 3)
Soil Heat Flux Plates
O), sensible heat, and momentum. The program processes the
2
o HMP155A o EE181
o Option 1
NR-LITE2 Net Radiometer (qty 0 to 1)
CS301 or CS320 Pyranometer (qty 0 to 1)
CS310 Quantum Sensor (qty 0 to 1)
SI-111 Infrared Radiometer (qty 0 to 1)
o Option 2
SN500SS, or NR01, or CNR4 4-Way Radiometer
(qty 0 to 1; if using CNR4, the CNF4 Ventilation and Heating Unit is also supported)
o CS650 o CS655
o Option 1: HFP01 plates (qty 0 to 3) o Option 2: HFP01SC self-calibrating plates (qty 0 to 3)
It may be possible to customize the program for other sensors or quantities in configurations not described here. Contact Campbell Scientific for more information.
a registered trademark of Campbell Scientific, Inc.
C-1

C.2 Precautions

NOTE
NOTE
NOTE
Appendix C. EasyFlux® DL CR6KH20
The VOLT116 and CDM-A116 are functionally the same, however their OSes are not interchangeable. If updating an OS, make sure it is for the correct model.
EasyFlux DL CR6KH20 requires the CR6 to have operating system (OS)
version 09.02 or newer. If using a VOLT116, it must have OS v.01 or newer, or if using a CDM-A116, it must have v.06 or newer.
The program applies the most common EC corrections to fluxes. However, the
user should determine the appropriateness of the corrections for their site.
Campbell Scientific always recommends saving time-series data in the event
reprocessing of raw data is warranted. Further, the user should determine the quality and fitness of all data for publication, regardless of whether said data were processed by EasyFlux DL CR6KH20 or another tool.
As EasyFlux DL CR6KH20 is not encrypted, users have the ability to view and
edit the code. However, Campbell Scientific does not guarantee the function of an altered program.

C.3 Wiring

When wiring the sensors to the data logger or VOLT116, the default wiring schemes, along with the number of instruments EasyFlux DL CRKH20 supports, should be followed if the standard version of the program is being used. TABLE C-1 through TABLE C-13 present the wiring schemes.
A KH20 and CSAT3B are the only required sensors for the program. The additional sensors described in the following tables are optional, although the CS106 and FW3 are recommended. Many of the optional sensors are wired to a VOLT116 (or CDM-A116) module, which effectively increases the CR6 analog terminals. If one or more of the optional sensors are not used, the data logger or VOLT116 terminals assigned to those sensor wires should be left unwired.
If the standard data logger program is modified, the wiring presented in TABLE C-1 may no longer apply. In these cases, refer directly to the program code to determine proper wiring.
If using an analog expansion module, all wiring and connections are the same whether using a VOLT116 or a CDM-A116. Therefore, throughout this appendix, the wiring terminals are only listed for the VOLT116.
C-2
Loading...
+ 44 hidden pages