“PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are
warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in
materials and workmanship under normal use and service for twelve (12)
months from date of shipment unless otherwise specified in the corresponding
Campbell pricelist or product manual. Products not manufactured, but that are
re-sold by Campbell, are warranted only to the limits extended by the original
manufacturer. Batteries, fine-wire thermocouples, desiccant, and other
consumables have no warranty. Campbell’s obligation under this warranty is
limited to repairing or replacing (at Campbell’s option) defective products,
which shall be the sole and exclusive remedy under this warranty. The
customer shall assume all costs of removing, reinstalling, and shipping
defective products to Campbell. Campbell will return such products by surface
carrier prepaid within the continental United States of America. To all other
locations, Campbell will return such products best way CIP (Port of Entry)
INCOTERM® 2010, prepaid. This warranty shall not apply to any products
which have been subjected to modification, misuse, neglect, improper service,
accidents of nature, or shipping damage. This warranty is in lieu of all other
warranties, expressed or implied. The warranty for installation services
performed by Campbell such as programming to customer specifications,
electrical connections to products manufactured by Campbell, and product
specific training, is part of Campbell’s product warranty. CAMPBELL
EXPRESSLY DISCLAIMS AND EXCLUDES ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. Campbell is not liable for any special, indirect,
incidental, and/or consequential damages.”
Assistance
Products may not be returned without prior authorization. The following
contact information is for US and international customers residing in countries
served by Campbell Scientific, Inc. directly. Affiliate companies handle
repairs for customers within their territories. Please visit
www.campbellsci.com to determine which Campbell Scientific company serves
your country.
To obtain a Returned Materials Authorization (RMA), contact CAMPBELL
SCIENTIFIC, INC., phone (435) 227-9000. After an application engineer
determines the nature of the problem, an RMA number will be issued. Please
write this number clearly on the outside of the shipping container. Campbell
Scientific’s shipping address is:
CAMPBELL SCIENTIFIC, INC.
RMA#_____
815 West 1800 North
Logan, Utah 84321-1784
For all returns, the customer must fill out a “Statement of Product Cleanliness
and Decontamination” form and comply with the requirements specified in it.
The form is available from our web site at www.campbellsci.com/repair. A
completed form must be either emailed to repair@campbellsci.com or faxed to
(435) 227-9106. Campbell Scientific is unable to process any returns until we
receive this form. If the form is not received within three days of product
receipt or is incomplete, the product will be returned to the customer at the
customer’s expense. Campbell Scientific reserves the right to refuse service on
products that were exposed to contaminants that may cause health or safety
concerns for our employees.
Table of Contents
PDF viewers: These page numbers refer to the printed version of this document. Use the
PDF reader bookmarks tab for links to specific sections.
7-1. Status Icons ........................................................................................ 16
iii
Table of Contents
iv
HS2 and HS2P
1. Introduction
FIGURE 1-1. HS2 HydroSense II System
FIGURE 1-2. HS2P HydroSense II Pole System
1
HS2 and HS2P
NOTE
NOTE
CAUTION
The HydroSense II is an easy-to-use, portable device for measuring volumetric
water content of soil. The major components of the system are the display, the
sensor, and the software. A sensor with 12 cm rods (model CS659 or CS659P)
and a sensor with 20 cm rods (model CS658 or CS658P) are available.
Campbell Scientific offers the HS2, the standard HydroSense II system, and
the HS2P, a HydroSense II system with a strong handle and pole. FIGURE
1-1, HS2 HydroSense II System (p. 1) shows the HS2 and FIGURE 1-2, HS2P
HydroSense II Pole System (p. 1) shows the HS2P.
Throughout the manual, HydroSense II refers to both the HS2 and
HS2P, unless specified otherwise.
2. Cautionary Statements
•The connectors that connect the HydroSense II display to the sensor are a
push-pull type with locators to align the connectors. DO NOT TWIST.
The connectors can be damaged if the user twists them or attempts to
screw or unscrew them.
•Remove the AA batteries before putting the HydroSense II display into
storage.
3. Quickstart
•An unobstructed view of the sky may be required to achieve the most
accurate global position data possible from the HydroSense II GPS.
•Sensor rods must be completely inserted into the soil before making a
measurement.
1. Remove the HS2 display or HS2P from the carry case.
2. Remove the blue protective strip from the display window.
Steps 3 and 4 are for the HS2 only because the HS2P comes fully
assembled. If using an HS2P, go directly to step 5.
3. Remove the HS2 sensor from the carry case.
4. Connect the HS2 sensor to the display by mating the connector on the
cable to the connector at the bottom of the display. DO NOT TWIST the
connectors.
The connector is a push-pull type with locators to align the
connectors. The connectors can be damaged if the user
attempts to screw or unscrew them. To mate the
connectors, simply bring them together and turn until they
are aligned, then applying firm pressure, slide them together
until they click. To disconnect, pull back on the connector
collar with the thumb and forefinger and pull the two
connectors apart.
2
HS2 and HS2P
TABLE 4-1. Comparison of HydroSense II and HydroSense(I)
NOTE
NOTE
5. Turn on the HydroSense II by holding the button for 3 seconds.
6. Establishing a GPS sync may take up to one minute or more. When the
GPS icon is displayed (), synchronization has been successful.
Measurements with GPS data can now be taken. See Appendix A.14, GPS
Settings (p. A-6) for information on toggling the GPS.
The HydroSense II stores positional information (when available),
time, and date with measurements. To reliably take advantage of
the GPS, use the HydroSense II outdoors with a clear view of the
sky. Measurements can be made and data stored without a
) signal; however, the HydroSense II Support Software
(p. 19
(HydroSoft) will be unable to later group and display data
geographically.
7. Insert the sensor rods fully into the soil. The HS2P uses a plastic
“bumper” that covers the rod mounting nuts. Insert the rods into the soil
up to the plastic “bumper”.
Only remove the plastic bumper when replacing soil rods and put
it back on when done.
GPS
4. Overview
8. To take a measurement, press . When the hourglass icon () in the
lower right of the screen is extinguished, the measured VWC and period
are shown on the screen.
9. Press to store the measurement. Assuming the GPS is in sync, follow
the prompts to create a zone.
10. Data are now stored in memory. To retrieve data, first establish a
Bluetooth link with your PC, and then collect data using the HydroSense II
support software. See Section 8.4, Connecting via Bluetooth (p. 18) and
Section 8.5, Collecting Data (p. 19).
The HydroSense II is a portable system for measuring volumetric water content
of soil. It is an improved successor to the HydroSense. TABLE 4-1,
Comparison of HydroSense II and HydroSense(I) compares features of the
HydroSense II and HydroSense (I).
Feature HydroSense II HydroSense (I)
Volumetric water
content and period
Yes, period as μsYes, period as ms
Relative water content
& water deficit with
wet and dry references
Yes (10 profiles) Yes (5 sites)
3
HS2 and HS2P
NOTE
Sensor rod length 20 cm and 12 cm
(NOT interchangeable)
Accuracy 20 cm rods: ±3%
(EC ≤ 4 dS/m)
20 cm and 12 cm
(interchangeable)
±3% VWC
(EC < 2 dS/m)
12 cm rods: ±3% (EC≤
6.5 dS/m)
GPS and geotagging Yes No
GPS zones Yes No
Data storage Yes (1000+ readings) No
Bluetooth Yes No
Date and time Yes (with GPS accuracy
No
when available)
Display 128 x 64 pixel graphic
LCD
2 line x 16 char
alphanumeric LCD
LCD backlight Yes No
Firmware updates User updateable Factory only
Field carry case Yes No
Portable Yes Yes
Battery life 1000+ readings 2000+ readings
Unlike the HydroSense (I), the 12 cm and 20 cm rods of the
HydroSense II are not interchangeable.
The HydroSense II can display relative water content (RWC) based on wet and
dry references set by the user. The unit also displays water deficit, which
indicates how much water is required (in mm of applied water) to return the
soil to the previously saved wet reference.
Measurements are made by fully inserting the sensor rods into the soil and
pressing . The process takes 2 to 3 seconds. The measurement can be
stored and later downloaded to a computer for display and analysis.
Communications between the computer and HydroSense II are via Bluetooth
(p. 18). The HydroSense II includes a GPS (Global Positioning System)
(p. 19) receiver that enables data to be stored with position information
(geotagged (p. 19)).
The HydroSense II has these default power saving features:
• turns off when idle for a configurable period
• adjustable backlight brightness and on-time
4
• backlight automatically disabled in bright conditions
• GPS and Bluetooth can each be forced off.
HS2 and HS2P
TABLE 4-2. HydroSense II Parts
CS658
Wrench
HS2 Display
Loctite
Spare
Holder
Screwdriver
The HydroSense II system consists of the items listed in TABLE 4-2,
HydroSense II Parts. These parts are shown in FIGURE 4-1, HS2 Parts in
Hard Carrying Case, FIGURE 4-2, HS2P Carrying Tote, and FIGURE 4-3,
HS2P Parts.
Qty HS2 Part HS2P Part
1 HydroSense II measurement and display unit
1 HS2P Insertion Pole Assembly
(pn 29153)
1 CS658 (20 cm) or CS659
(12 cm) water content sensor
CS658P (20 cm) or CS659P
(12 cm) water content sensor
4 AA batteries (factory installed inside the display)
1 Spare battery holder (pn 28404); spare batteries not included
1 Wrench (pn 26156) for installing / replacing rods
1 Loctite thread locking compound (pn 13665) for rod replacement
1 Phillips screwdriver (pn 6290)
1 Hard carrying case (pn 27789) Carrying tote (pn 29468) or hard
carrying case (pn 27789)
1 HydroSense II Support Software (HydroSoft) on CD
Battery
FIGURE 4-1. HS2 Parts in Hard Carrying Case
5
HS2 and HS2P
Display
HS2P Pole
HydroSoft CD
CS659P
Loctite
Wrench
Spare Battery Holder
FIGURE 4-2. HS2P Carrying Tote
FIGURE 4-3. HS2P Parts
Below are spare and supporting parts available from Campbell Scientific:
• Spare 20 cm rods (pn 26483) for the CS658 or CS658P sensor
• Spare 12 cm rods (pn 10184) for the CS659 or CS659P sensor
• USB Bluetooth adapter for supporting PCs (pn 28411).
5. Specifications
5.1 Display Specifications
GPS Accuracy: ±5 m (16.4 ft) typical;
Data storage: >1000 records (ring memory)
Zone storage: >100 records (fill and stop)
Display: 128 x 64 pixel graphic LCD
±1 ms time with GPS sync
6
Backlight: blue / white LED; brightness adjustable
Compatibility: Bluetooth (to ≈10 m);
Google Earth (via software)
Weight: 0.34 kg (0.75 lb)