Campbell Scientific HMP155A-L User Manual

PRODUCT
The HMP155A provides reliable relative humidity (RH) and temperature measurements for a wide range of applications.
It uses a HUMICAP®180R capacitive thin film polymer sensor to measure RH over the 0 to 100% RH range. A PRT measures temperature over the -80° to +60°C range. This rugged, accurate temperature/RH probe is manufactured by Vaisala.
Well-suited for long-term, unattended applications Accurate and rugged
The HMP155A uses a HUMICAP180R capacitive thin film polymer sensor to measure RH, and a PRT to measure temperature.
To reduce the current drain, power can be supplied to the HMP155A only during measurement when the sensor is
Electromagnetic Compatibility
Complies with EMC standard
EN61326-1 Electromagnetic Filter Description Sintered PTFE Housing Body Material PC Housing Classification IP66
Temperature and Relative Humidity Probe
HMP155A-L
Accurate, Wide Temp Range
Higher-end sensor where higher accuracy is required
Overview
Benefits and Features
Detailed Description
Manufacturer's note: In changing temperature conditions with high humidity, the probe temperature can lag behind the temperature of surrounding air and this can cause risk of condensation on the sensor. A wet probe cannot observe the actual humidity in the ambient air. If the condense water is contaminated, the life span of the probe may shorten and the calibration may change.
Mounts to a mast, crossarm, or user-supplied pole Compatible with most Campbell Scientific data loggers
connected to the datalogger’s switched 12 V terminal. Dataloggers that do not have a switched 12 V terminal, such as the CR510 or CR7, can use the SW12V switched 12 V device to switch power to the sensor only during measurement.
Specifications
Voltage Output 0 to 1 Vdc Average Current
Consumption Operating Voltage 7 to 28 Vdc Settling Time 2 s (at power up)
For comprehensive details, visit: www.campbellsci.com.au/hmp155a
≤ 3 mA (analog output mode)
Field-Replaceable Chip or Recalibrate
Recalibrate
Tip Diameter 1.2 cm (0.5 in.) Length 27.9 cm (11 in.) Head Height 4 cm (1.6 in.) Body Height 2.4 cm (0.9 in.) Body Width 2.0 cm (0.8 in.)
Relative Humidity
Sensing Element HUMICAP 180R Measurement Range 0 to 100% RH (non-condensing) Response Time 60 s (90% step change)
20 s (63% step change) The response time for the RH specification is for the HUMICAP 180R at 20°C in still air with sintered PTFE filter and a 0 to 75% RH step change.
Factory Calibration Uncertainty
The factory calibration uncertainty is defined as ±2 standard deviation limits. Uncertainty is at +20°C. Small variations are possible; see also the calibration certificate. ±1.0% RH 40 to 97% RH
±0.6% RH 0 to 40% RH
Accuracy ± (1.2 + 0.012 × reading) % RH
(at 40° to 60°C) ±1% RH (at 15° to 25°C, 0 to 90% RH)
-NOTE- Accuracy specifications include non­linearity, hysteresis, and repeatability.
±1.7% RH (at 15° to 25°C, 90 to 100% RH) ±(1.0 + 0.008 × reading) % RH (at -20° to +40°C) ±(1.2 + 0.012 × reading) % RH (at -40° to -20°C) ±(1.4 + 0.032 × reading) % RH (at -60° to -40°C)
Air Temperature
Sensing Element PT 100 RTD 1/3 class B IEC 751 Measurement Range -80° to +60°C Accuracy ±(0.226 - 0.0028 x
temperature)°C (-80° to +20°C) ±(0.055 + 0.0057 x temperature)°C (+20° to +60°C)
Entire Temperature Range Refer to graph in probe manual.
For comprehensive details, visit: www.campbellsci.com.au/hmp155a
© 2021 Campbell Scientific, Inc. | 01/13/2021