The CS511,manufactured by Sensorex, is a submersible,
rugged, low-maintenance sensor that senses dissolved
oxygen in a medium (typically water). Itconsists of a self-
In-line thermistor provides automatic temperature
compensation
Compatible with most Campbell Scientific dataloggers
Oxygen diffuses through the CS511’s membrane onto a
cathodethat produces a chemical reaction. An electrical
current isproduced by this reaction, which is converted from
microampsto millivolts by an in-line resistor. An in-line
thermistor providesautomatic temperature compensation.
Because the CS511 is shipped dry, DO electrolyte needs to
beadded before using the sensor. DO electrolyte as well as
sparemembranes are included in the sensor’s shipping kit.
Operating Temperature
Range
0° to 50°C
Measurement Range 0.5 to 50 ppm (for O2)
Accuracy ±2%
Widely
Compatible
Submersible, rugged, lowmaintenance DO sensor
PRODUCT
Overview
Benefits and Features
Detailed Description
polarizing galvanic cell that generates a millivoltsignal
proportional to the amount of oxygen present in
themeasured medium (typically water).
Agitator available that keeps the sensor clean and moves
water across membrane for more accurate readings
Currently, the CS511 is Sensorex’s Model DO6400/T. Prior to
June 2008, the CS511 was Sensorex’s Model DO6200/T.
Programming, wiring, and most specifications are the same
for these two sensors. However, they use different
accessories and look different. Replacement parts remain
available for the older DO6200/T. (See the Ordering Info on
the web page.)
For an image of the older DO6200/T, see the retired version
of the CS511 manual. Contact Campbell Scientific for more
information.
Specifications
Output 33 mV ± 9 mV (100% saturation);
< 2 mV (0% saturation)
Response Time 5 min from 100% to 0% oxygen
For comprehensive details, visit: www.campbellsci.com.au/cs511-l

Minimum Water Velocity
5 cm s-1 (2 in. s-1) across
membrane
Minimum Submersion
Depth
6.35 cm (2.5 in.)
Pressure 0 to 100 psig
Temperature
Compensation
Automatic from 4° to 40°C
Probe Electrolyte NaCl + glycerol (prevents
freezing)
Diameter 5.72 cm (2.25 in.)
Height 17.78 cm (7 in.) from bottom of
sensor to end of cable strain
relief
Shipping Weight 0.8 kg (1.75 lb) including sensor
and shipping kit
For comprehensive details, visit: www.campbellsci.com.au/cs511-l
© 2021 Campbell Scientific, Inc. | 02/15/2021