“PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are
warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in
materials and workmanship under normal use and service for twelve (12)
months from date of shipment unless otherwise specified in the corresponding
Campbell pricelist or product manual. Products not manufactured, but that are
re-sold by Campbell, are warranted only to the limits extended by the original
manufacturer. Batteries, fine-wire thermocouples, desiccant, and other
consumables have no warranty. Campbell’s obligation under this warranty is
limited to repairing or replacing (at Campbell’s option) defective products,
which shall be the sole and exclusive remedy under this warranty. The
customer shall assume all costs of removing, reinstalling, and shipping
defective products to Campbell. Campbell will return such products by surface
carrier prepaid within the continental United States of America. To all other
locations, Campbell will return such products best way CIP (Port of Entry)
INCOTERM® 2010, prepaid. This warranty shall not apply to any products
which have been subjected to modification, misuse, neglect, improper service,
accidents of nature, or shipping damage. This warranty is in lieu of all other
warranties, expressed or implied. The warranty for installation services
performed by Campbell such as programming to customer specifications,
electrical connections to products manufactured by Campbell, and product
specific training, is part of Campbell’s product warranty. CAMPBELL
EXPRESSLY DISCLAIMS AND EXCLUDES ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. Campbell is not liable for any special, indirect,
incidental, and/or consequential damages.”
Assistance
Products may not be returned without prior authorization. The following
contact information is for US and international customers residing in countries
served by Campbell Scientific, Inc. directly. Affiliate companies handle
repairs for customers within their territories. Please visit
www.campbellsci.com to determine which Campbell Scientific company serves
your country.
To obtain a Returned Materials Authorization (RMA), contact CAMPBELL
SCIENTIFIC, INC., phone (435) 227-9000. After an applications engineer
determines the nature of the problem, an RMA number will be issued. Please
write this number clearly on the outside of the shipping container. Campbell
Scientific’s shipping address is:
CAMPBELL SCIENTIFIC, INC.
RMA#_____
815 West 1800 North
Logan, Utah 84321-1784
For all returns, the customer must fill out a “Statement of Product Cleanliness
and Decontamination” form and comply with the requirements specified in it.
The form is available from our web site at www.campbellsci.com/repair. A
completed form must be either emailed to repair@campbellsci.com or faxed to
(435) 227-9106. Campbell Scientific is unable to process any returns until we
receive this form. If the form is not received within three days of product
receipt or is incomplete, the product will be returned to the customer at the
customer’s expense. Campbell Scientific reserves the right to refuse service on
products that were exposed to contaminants that may cause health or safety
concerns for our employees.
Table of Contents
PDF viewers: These page numbers refer to the printed version of this document. Use the
PDF reader bookmarks tab for links to specific sections.
CMP-series pyranometers are designed for continuous outdoor monitoring of
solar radiation intensity. A flat spectral sensitivity from 285 to 2800 nm
enables accurate measurements in natural sunlight, under plant canopies, and in
green houses or buildings. When inverted, these pyranometers can measure
reflected solar radiation. Uses include monitoring global horizontal irradiance
(GHI) and plane of array irradiance (POA). Diffuse sky radiation can also be
measured with the use of a shade mechanism.
CMP-series pyranometers are manufactured by Kipp & Zonen, and cabled by
Campbell Scientific.
Before using these pyranometers, please study:
• Section 2, Cautionary Statements
• Section 3, Initial Inspection
• Section 4, Quick Start
More details are available in the remaining sections.
2. Cautionary Statements
•CMP-series pyranometers are rugged, but they should be handled as
precision scientific instruments.
•Care should be taken when opening the shipping package to not damage or
cut the cable jacket. If damage to the cable is suspected, consult with a
Campbell Scientific applications engineer.
3. Initial Inspection
Check the contents of the shipment. If there is a shortage (see Section 3.1,
Ships With), contact Campbell Scientific. If any damage has occurred during
transport, immediately file a claim with the carrier and contact Campbell
Scientific to facilitate repair or replacement.
The model number and cable length are printed on a label at the connection end
of the cable. Check this information against the shipping documents to ensure
the correct product and cable length are received.
1
CMP6-L, CMP11-L, and CMP21-L Pyranometers
3.1 Ships With
(2) Bolts for mounting from original mfg
(1) Instruction Manual from original mfg
(1) Sun Shield from original mfg
(2) Nylon washers from original mfg
3.2 Calibration Certificate
Each pyranometer is shipped with an instruction manual provided by Kipp &
Zonen that contains information concerning its construction, spectral
sensitivity, cosine response, and a simple sensor check out procedure. Included
with the sensor and manual is a calibration certificate with the sensor
sensitivity value and serial number.
NOTE
4. Quick Start
NOTE
4.1 Siting
Cross check this serial number against the serial number on your
pyranometer to ensure that the given sensitivity value
corresponds to your sensor.
Appendix A, CVF3 Heater/Ventilator, provides the installation
procedure for the CVF3 ventilation unit.
The pyranometer is usually installed horizontally for global horizontal
measurements. However, the pyranometer can be installed at any angle for
POA measurements and in the inverted position for reflected measurements. In
all cases it will measure the solar flux incident on the sensor surface.
Site the pyranometer to allow easy access for maintenance while ideally
avoiding any obstructions above the plane of the sensing element. It is
important to mount the pyranometer such that a shadow will not be cast on it at
any time.
If this is not possible, try to choose a site where any obstruction over the
azimuth range between earliest sunrise and latest sunset has an elevation not
exceeding 5°. Diffuse solar radiation is less influenced by obstructions near
the horizon. For instance, an obstruction with an elevation of 5° over the
whole azimuth range of 360° decreases the downward diffuse solar radiation
by only 0.8%.
2
The sensor should be mounted with the cable pointing towards the nearest
magnetic pole (e.g., in the Northern Hemisphere point the cable toward the
North Pole); see FIGURE 4-1 through FIGURE 4-4.
4.2 Mounting
CMP6-L, CMP11-L, and CMP21-L Pyranometers
See Section 7.1, Mounting to a Tripod Tower, for more information.
CM245 Adjustable
Angle Mounting
Stand
CM2XX-Series
Crossarm
FIGURE 4-1. Pyranometer installation
FIGURE 4-2. Pyranometer mounted horizontally for the Northern
Hemisphere (left) and Southern Hemisphere (right)
3
CMP6-L, CMP11-L, and CMP21-L Pyranometers
FIGURE 4-3. Two views of a pyranometer mounted at an angle for the
Northern Hemisphere
4
FIGURE 4-4. Pyranometer mounted at an angle for the Southern
Hemisphere
CMP6-L, CMP11-L, and CMP21-L Pyranometers
4.3 Datalogger Programming / Wiring
The simplest method for programming the datalogger to measure a CMP6 or
CMP11 is to use Campbell Scientific's SCWin Short Cut Program Generator
(see FIGURE 4-5). Wire the pyranometer according to the wiring diagram
generated by Short Cut.
NOTE
The CMP21 is not included in Short Cut. Refer to Section 7,
Installation, for wiring and programming information if not
using Short Cut.
FIGURE 4-5. SCWin Short Cut Program Generator
5
CMP6-L, CMP11-L, and CMP21-L Pyranometers
5. Overview
5.1 Models
CMP-series models differ in accuracy and performance. See Section 6,
Specifications. The CMP21 also includes an internal thermistor allowing
individually optimized compensation of the measurements. The –L portion of
the model number indicates that the pyranometer has a user-specified cable
length. The pyranometers have several cable termination options. Their cables
can terminate in:
•Pigtails that connect directly to a Campbell Scientific datalogger
(cable termination option –PT).
•Connector that attaches to a prewired enclosure (cable termination
option –PW).
•Connector that attaches to a CWS900 Wireless Sensor Interface (cable
termination option –CWS). The CWS900 enables the pyranometer to
be used in a wireless sensor network. Please note that this option is
not available for the CMP21.
5.2 Construction
The pyranometers consist of a thermopile sensor, housing, two glass domes,
and cable. The thermopile is coated with a black absorbent coating. The paint
absorbs the radiation and converts it to heat. The resultant temperature
difference is converted to a voltage by the copper-constantan thermopile. The
thermopile is encapsulated in the housing in such a way that it has a field of
view of 180 degrees and the angular characteristics needed to fulfill the cosine
response requirements.
6
CMP6-L, CMP11-L, and CMP21-L Pyranometers
6. Specifications
6.1 Pyranometers
TABLE 6-1. CMP-series Specifications
Specification CMP6 CMP11 CMP21
ISO Classification First Class Secondary Standard
Maximum irradiance
2000 W•m
Spectral range
(50% points)
Response time (95 %) <18 s <5 s
Expected daily
uncertainty
Zero offset due to thermal
radiation
(200 W•m
–2
)
<15 W•m
Zero offset due to
temperature change
–1
(5 K•hr
)
<4 W•m
Non-stability
(change/year)
Non-linearity
(0 to 1000 W•m
–2
)
Directional error
(up to 80° with 1000
–2
W•m
beam)
<20 W•m
Tilt error
(at 1000 W•m
–2
)
Level accuracy 0.1°
Operating temperature –40° to 80°C
Temperature dependence
of sensitivity
Sensitivity
Typical signal output for
atmospheric applications
<4% (–10° to +40°C) <1% (–10° to +40°C) <1% (–20° to +50°C)
5 to 20 µV / W•m
0 to 20 mV 0 to 15 mV
Weight
Impedance* 20 to 200 Ω 10 to 100 Ω
* Impedance is defined as the total electrical impedance at the radiometer output connector fitted to the housing. It
arises from the electrical resistance in the thermal junctions, wires, and passive electronics within the radiometer.
–2
4000 W•m–2
285 to 2800 nm
<5% <2%
–2
<7 W•m–2
–2
<2 W•m–2
<1 % <0.5%
<1% <0.2%
–2
<10 W•m–2
<1% <0.2%
–2
7 to 14 µV / W•m–2
0.6 kg (1.3 lb) without cable;
0.9 kg (2 lb) with 10 m (33 ft) cable
7
CMP6-L, CMP11-L, and CMP21-L Pyranometers
FIGURE 6-1. Dimensions of the CMP6, CMP11, and CMP21
6.2 CVF3 Ventilation Unit
Compatible Pyanometers: CMP6, CMP11, CMP21
Power supply: 12 Vdc, 1.3 A (with 10 W Heater)
Operating temperature range: –40° to 70°C
Ventilation power: 5 W continuously
Heating power: 5 W and 10 W
Heater induced offset: <1 W•m
Weight without cable: 1.6 kg (3.5 lb)
–2
(with CMP11 Pyranometer)
8
FIGURE 6-2. Dimensions of the CVF3
7. Installation
7.1 Mounting to a Tripod or Tower
CMP6-L, CMP11-L, and CMP21-L Pyranometers
Tools required for installation on a tripod or tower:
Small and medium Phillips screwdrivers
5/16”, 1/2” open end wrenches
5/32” Allen wrench
Tape measure
UV-resistant wire ties
Side-cut pliers
Compass
Step ladder
The pyranometers include a bubble level and two leveling screws, which allow
them to be leveled horizontally without using a leveling base. They mount to a
mast, crossarm, or pole (1.0 in. to 2.1 in. outer diameter) via the CM245
Mounting Stand.
NOTE
If using a CFV3 Ventilation Unit, a different mounting stand, the
27084, is required. Refer to Appendix A, CVF3 Heater/Ventilator, for more information.
The CM245 includes slots that allow it to be adjusted to any angle from
horizontal to vertical. If mounting the pyranometer at an angle, ensure that the
crossarm is leveled horizontally before placing the bracket at its proper angle.
Angle positions are included on the bracket label (see FIGURE 7-1 and
FIGURE 7-2).
Pyranometer
mounts here
First
2.125”
u-bolt
First
Second
2.125”
u-bolt
FIGURE 7-1. CM245 bracket with 2.125” u-bolts positioned to mount
the pyranometer horizontally on a crossarm
2.125”
u-bolt
90
Second
2.125”
u-bolt
9
CMP6-L, CMP11-L, and CMP21-L Pyranometers
First
1.5”
u-bolt
40
Second
1.5”
u-bolt
FIGURE 7-2. CM245 bracket with 1.5” u-bolts positioned to mount
pyranometer at a 40° angle on a vertical pipe
Do the following to level the pyranometer horizontally (see FIGURE 7-3):
1. Attach the mounting stand to the crossarm.
2. Loosely mount the pyranometer on the mounting stand. Do not fully
tighten the two mounting screws.
3. Turn the leveling screws as required to bring the bubble of the level within
the ring.
4. Tighten the mounting screws to secure the assembly in its final position.
Check that the pyranometer is still correctly leveled and adjust as necessary.
5. Attach the white plastic sun screen to the pyranometer.
10
Mounting screws
d
N
Levelling screw
CMP6-L, CMP11-L, and CMP21-L Pyranometers
Sun shield
ylon washers
Bubble level
Pyranometer
mounting st45an
7.2 Wiring
NOTE
CM2
Crossarm
FIGURE 7-3. Exploded view of the pyranometer mounting
Short Cut users should wire the sensor according to the wiring
diagram generated by Short Cut.
The cable of the CMP6 and CMP11 has two conductors and a shield. The
cable of the CMP21 has five conductors and a shield. The additional
conductors on the CMP21’s cable are for connecting its internal thermistor. A
schematic for the CMP6, CMP11, and the thermopile of the CMP21 is
provided in Section 7.2.1, CMP6, CMP11, and CMP21 Thermopile Schematic.
Wiring for the CMP6 and CMP11 is described in Section 7.2.2, CMP6 and
CMP11 Wiring, wiring for the CMP21 is described in Section 7.2.3, CMP21
Wiring.
11
CMP6-L, CMP11-L, and CMP21-L Pyranometers
7.2.1 CMP6, CMP11, and CMP21 Thermopile Schematic
A schematic diagram of a CMP6, CMP11, or CMP21 thermopile is shown in
FIGURE 7-4.
W hite(+)
Red
Black (-)
Blue
Black
Shield
FIGURE 7-4. CMP6, CMP11, and CMP21 thermopile detector
schematic
7.2.2 CMP6 and CMP11 Wiring
NOTE
A CMP6 or CMP11 purchased from Campbell Scientific has
different wiring than a pyranometer purchased directly from
Kipp & Zonen.
The pyranometer is measured using either differential analog channels or
single-ended analog channels.
A differential voltage measurement is recommended because it has better noise
rejection than a single-ended measurement.
Connections to Campbell Scientific dataloggers for a differential measurement
are given in TABLE 7-1. A user-supplied jumper wire should be connected
between the low side of the differential input and ground (AG or
) to keep
the signal in common mode range.
Connections to Campbell Scientific dataloggers for a single-ended
measurement are given in TABLE 7-2.
TABLE 7-1. CMP6 and CMP11 Differential Connections to Campbell Scientific Dataloggers
Color
Description
CR9000(X), CR5000,
CR3000, CR1000,
CR800
CR510, CR500,
CR10(X)
21X, CR7, CR23X
White Signal (+) DIFF Analog High DIFF Analog High DIFF Analog High
Black Signal (–) *DIFF Analog Low *DIFF Analog Low *DIFF Analog Low
Shield Shield
* Jumper to AG or
with user supplied 26 AWG or larger wire.
G
12
CMP6-L, CMP11-L, and CMP21-L Pyranometers
TABLE 7-2. CMP6 and CMP11 Single-Ended Connections to Campbell Scientific Dataloggers
Color
Description
CR9000(X), CR5000,
CR3000, CR1000,
CR800
CR510, CR500,
CR10(X)
21X, CR7, CR23X
White Signal (+) SE Analog SE Analog SE Analog
Black Signal (–)
Clear Shield
AG
G
7.2.3 CMP21 Wiring
NOTE
A CMP21 purchased from Campbell Scientific has different
wiring than a CMP21 purchased directly from Kipp & Zonen.
The CMP21’s pyranometer can be measured using either differential analog
channels or single-ended analog channels. A differential voltage measurement
is recommended because it has better noise rejection than a single-ended
measurement. If a differential channel is not available, a single-ended
measurement can be used.
A single-ended channel and a voltage excitation channel are used to measure
the CMP21’s internal thermistor.
Connections to Campbell Scientific dataloggers for a differential measurement
are given in TABLE 7-3. A user-supplied jumper wire should be connected
between the low side of the differential input and ground (AG or
) to keep
the signal in common mode range. Connections to Campbell Scientific
dataloggers for a single-ended measurement are given in TABLE 7-4.
TABLE 7-3. CMP21 Differential Connections to Campbell Scientific Dataloggers
Wire Color
Wire Label/
Description
CR9000(X),
CR5000, CR3000,
CR1000, CR800
CR510, CR500,
CR10(X)
21X, CR7, CR23X
White Pyranometer Sig DIFF Analog High DIFF Analog High DIFF Analog High
Blue Pyranometer Ref *DIFF Analog Low *DIFF Analog Low *DIFF Analog Low
Yellow Thermistor Volt Excite VX or EX E EX
Black Thermistor Sig Single-ended analog Single-ended analog Single-ended analog
Brown Thermistor Ref
Clear Shield
* Jumper to AG or
with user-supplied wire.
AG
G
13
CMP6-L, CMP11-L, and CMP21-L Pyranometers
TABLE 7-4. CMP21 Single-Ended Connections to Campbell Scientific Dataloggers
Wire Color
Wire Label/
Description
CR9000(X),
CR5000, CR3000,
CR1000, CR800
CR510, CR500,
CR10(X)
21X, CR7, CR23X
White Pyranometer Sig Single-ended analog Single-ended analog Single-ended analog
Blue Pyranometer Ref
AG
Yellow Thermistor Volt Excite VX or EX E EX
Black Thermistor Sig Single-ended analog Single-ended analog Single-ended analog
Brown Thermistor Ref
Clear Shield
AG
G
7.3 Programming
NOTE
This section is for users who write their own datalogger
programs. You do not need to read this section if using our Short
Cut Program Generator, or connecting the pyranometer to a
prewired enclosure or CWS900 Wireless Sensor Interface. Our
prewired enclosures include a datalogger program. Refer to the
Wireless Sensor Manual for programming information if using a
CMP6 or CMP11 with a CWS900.
7.3.1 Solar Radiation Measurements
CAUTION
Solar radiation can be reported as an average flux density (W•m–2) or daily
total flux density (MJ•m
7-5. Programming examples are given for both average and daily total solar
radiation.
The pyranometers output a low level voltage ranging from 0 to a maximum of
up to 20 mV, in natural light, depending on the calibration factor and radiation
level.
This voltage output is measured using either a differential voltage instruction
(VoltDiff() in CRBasic or Instruction 2 (P2) in Edlog) or a single-ended
voltage instruction (VoltSE() in CRBasic or Instruction 1 (P1) in Edlog).
Nearby AC power lines, electric pumps, or motors can be a
source of electrical noise. If the sensor or datalogger is
located in an electrically noisy environment, the
measurement should be made with the 60 or 50 Hz
rejection integration option as shown in the example
programs.
–2
). The appropriate multipliers are listed in TABLE
14
7.3.1.1 Input Range
CMP6-L, CMP11-L, and CMP21-L Pyranometers
The output voltage is usually between 5 and 20 mV per 1000 W•m
–2
. When
estimating the maximum likely value of sensor output a maximum value of
–2
solar radiation of 1100 W•m
horizontal surface. Plane of array irradiances can exceed 1500 W•m
can be used for field measurements on a
–2
.
Select the input range as follows:
7.3.1.2 Multiplier
1. Estimate the maximum expected input voltage by multiplying the
–2)
maximum expected irradiance (in W•m
–2
). Divide the answer by 1000 to give the maximum in millivolt
W•m
by the calibration factor (in µV /
units.
2. Select the smallest input range which is greater than the maximum
expected input voltage. Normally the 50 mV range for the CR3000,
CR5000, CR9000(X), CR7, and CR23X and the 25 mV or 250 mV range
for the CR800, CR850, CR1000, CR510, and CR10(X) will be suitable.
The exact range will depend on the sensitivity of your individual sensor
and the maximum expected reading. With some dataloggers an autorange
option can be used if measurement time is not critical.
The parameter code for the input range also specifies the measurement
integration time. The slow or 60 Hz rejection integration gives a more noisefree reading. The 250 µs (CRBasic) or a fast (Edlog) integration takes less
power and allows for faster throughput.
The multiplier converts the millivolt reading to engineering units. The
sensitivity value supplied by the manufacturer gives the output of the sensor as
–2
µV (micro-volts) / W•m
. As the datalogger voltage measurement
instructions give a default output in mV, the following equation should be used
–2
to calculate the multiplier to give the readings in W•m
:
m = 1000/c
Where,
m = multiplier
c = sensor output in µV / W•m
–2
Other units can be used by adjusting the multiplier as shown in TABLE 7-5.
TABLE 7-5. Multipliers Required for Flux Density and Total Fluxes
Units Multiplier Output Processing
W•m–2
MJ•m–2
kJ•m–2
cal•cm–2
cal • cm
–2
• min–1
W • hr • m–2
M Average
M * t * 0.000001 Totalize
M *
t * 0.001
Totalize
M * t * 0.0239 * 0.001 Totalize
M * 1.434 * 0.001 Average
t / 3600 Totalize
M = calibration factor with units of W•m–2 / mV
t = datalogger program execution interval in seconds
15
CMP6-L, CMP11-L, and CMP21-L Pyranometers
7.3.1.3 Offset
The offset will normally be fixed at zero as the sensor should output no
significant signal in dark conditions. In practice, because of the nature of
thermopile detector sensors, there will be some offset in dark conditions;
sometimes this offset can give negative light readings. This offset varies with
several factors (e.g., rate of change of sensor temperature), so it cannot be
removed with a fixed offset. Some users may wish to remove small negative
readings by including code after the measurement instructions that sets
negative readings to zero.
7.3.1.4 Output Format Considerations
Over-ranging may be an issue if the measurement values are totalized. Overranging can be prevented when using CRBasic by storing the data in the IEEE4
format.
When using Edlog, the largest number the datalogger can store in final storage
is 6999 in low resolution mode (FP2) and 99999 in high resolution mode (if
available). The following example shows how over-ranging can be a problem
for Edlog dataloggers.
Example
Assume that daily total flux is desired, and that the Edlog datalogger scan rate
is 1 second. With a multiplier that converts the readings to units of kJ•m
an average irradiance of 0.5 kW•m
will be exceeded in less than four hours.
Solution 1 – Change the multiplier in the instruction to (m * 0.001). This will
totalize MJ•m
Solution 2 – Record the average flux density and later multiply the result by the
number of seconds in the output interval to arrive at total flux.
Solution 3 – Record the total flux using the high resolution format. The draw
back to high resolution is that it requires four bytes of memory per data point,
consuming twice as much memory as low resolution. Instruction 78 is used to
switch to high resolution in the Edlog dataloggers.
–2
instead of kJ•m–2.
–2
, the maximum low resolution output limit
7.3.2 CMP21 Internal Thermistor Measurement
The thermistor is measured using a half bridge measurement instruction
(BrHalf instruction in CRBasic or Instruction 5 (P5) in Edlog). The value
provided by the half bridge instruction needs to be converted to resistance and
then converted to temperature.
The following equation is used to convert to resistance:
–2
and
16
=
⎜
⎝
⎛
⎜
1000Res.
⎞
V
x
⎟
⎟
V1
−
x
⎠
CMP6-L, CMP11-L, and CMP21-L Pyranometers
Where,
Vx = the value provided by the half bridge instruction
In CRBasic, the conversion to resistance is entered as a mathematical
expression. In Edlog, Instruction P59 (Bridge Transform) does the conversion.
The Steinhart-Hart equation is used to convert resistance to temperature. The
Steinhart-Hart equation for converting resistance to degree Celsius is as
follows:
Temperature = 1/[A + B*LN(resistance) + C*(LN(resistance))^3] - 273.15
Where A, B, and C are coefficients for the Steinhart-Hart equation.
The coefficients for the Steinhart-Hart equation are specific to the thermistor
contained in your CMP21. A calibration certificate that lists these coefficients
is shipped with each CMP21 pyranometer.
In CRBasic, the Steinhart-Hart equation is entered as a mathematical
expression. Edlog dataloggers can use Instruction P200 (requires a newer
datalogger operating system).
7.3.3 Example Programs
7.3.3.1 CR1000 Example Program for Measuring a CMP6
Although this example is for the CR1000, other CRBasic dataloggers are
programmed similarly. The following program measures the CMP6 every
second and converts the millivolt output to W•m
14.33 µV / W•m
program outputs the average and standard deviation of the flux (W•m
measurements.
Wiring for this example is given in TABLE 7-6.
TABLE 7-6. CR1000 Wiring for CMP6 Example Program
Wire Color Description CR1000 Jumper*
White Solar Signal (+) 1H
Black Solar Signal (–) 1L
Clear Shield
* Jumper 1L to
–2
is used for the example program. Every 10 minutes, the
'Measure the Battery Voltage and Panel Temperature
PanelTemp (PTemp,250)
Battery (Batt_Volt)
'Measure the CMP6
VoltDiff (CMP6_Irr,1,mV25C,1,True ,10000,_60Hz,1000/14.33,0)
CallTable TenMin
NextScan
EndProg
7.3.3.2 CR1000 Example Program for Measuring a CMP11
Although this example is for the CR1000, other CRBasic dataloggers are
programmed similarly. The following program measures the CMP11 every
second and converts the millivolt output to W•m
8.55 µV / W•m
–2
is used for the example program. Every 10 minutes, the
program outputs the average and standard deviation of the flux (W•m
measurements.
Wiring for this example is given in TABLE 7-7.
TABLE 7-7. CR1000 Wiring for CMP11 Example Program
'Measure the Battery Voltage and Panel Temperature
PanelTemp (PTemp,250)
Battery (Batt_Volt)
'Measure the CMP11
VoltDiff (CMP11_Irr,1,mV25C,2,True ,10000,_60Hz,1000/8.55,0)
CallTable TenMin
NextScan
EndProg
7.3.3.3 CR1000 Example Program for Measuring a CMP21
Although this example is for the CR1000, other CRBasic dataloggers are
programmed similarly. The following program measures the CMP21 every
second. It converts the pyranometer’s millivolt output to W•m
pyranometer calibration of 8.65 µV / W•m
The resistance of the internal thermistor is converted to degree Celsius and then
to Kelvin. Every 10 minutes, the program outputs the average and standard
deviation of the flux (W•m
–2
) measurements and temperature measurements.
Wiring for this example is given in TABLE 7-8.
TABLE 7-8. CR1000 Wiring for CMP21 Example Program
Wire Color Description CR1000 Jumper*
White Solar Signal (+) 3H
Blue Solar Signal (–) 3L
Yellow Voltage Excitation VX1
Black Temp Signal 15 SE
Brown Signal Reference
Clear Shield
* Jumper 3L to with user-supplied 26 AWG or larger wire.
–2
–2
is used for the example program.
. A
19
CMP6-L, CMP11-L, and CMP21-L Pyranometers
'CR1000 Series Datalogger
Public PTemp
Public Batt_Volt
Public CMP21_Irr
Public CMP21_T_C
Public CMP21_T_K
Dim Rs,Vs_Vx
Units CMP21_Irr = W/m2
Units CMP21_T_C = Degrees C
Units CMP21_T_K = Degrees K
DataTable (TenMin,1,-1)
DataInterval (0,1,Min,8)
Minimum (1,Batt_Volt,FP2,0,False)
Sample (1,PTemp,FP2)
Average (1,CMP21_Irr,FP2,False)
StdDev (1,CMP21_Irr,FP2,False)
Average (1,CMP21_T_C,FP2,False)
StdDev (1,CMP21_T_C,FP2,False)
Average (1,CMP21_T_K,FP2,False)
StdDev (1,CMP21_T_K,FP2,False)
EndTable
BeginProg
Scan (1,Sec,0,0)
‘Measure the Battery Voltage and Panel Temperature
7.3.3.4 CR10X Example Program for Measuring a CMP6
The following program uses a CR10X to measure a CMP6 every 10 seconds
and convert the mV output to W•m
14.33 µV / W•m
hourly average flux (W•m
–2
is used for this example program. The program outputs an
–2
–2
and MJ•m–2. A sensor calibration of
), and a daily total flux density (MJ•m–2).
CMP6-L, CMP11-L, and CMP21-L Pyranometers
Wiring for the example is given in TABLE 7-9.
TABLE 7-9. CR10X Wiring for CMP6 Example Program
Wire Color Description CR10X Jumper*
White Solar Signal (+) 1H
Black Solar Signal (–) 1L
Clear Shield AG
* Jumper 1L to AG terminal on CR10X with user-supplied 26 AWG or larger
wire.
;{CR10X}
*Table 1 Program
01: 10.0000 Execution Interval (seconds)
; CMP6 measurement in W/m2
1: Volt (Diff) (P2)
1: 1 Reps
2: 23 25 mV 60 Hz Rejection Range ;use the 50 mV range for the CR7, 21X and CR23X
3: 1 DIFF Channel ;use the 250 mV range for the CR10X if
4: 3 Loc [ Solar_Wm2 ] ;calibration factor is > 25 µV/Wm-2
5: 69.7837 Multiplier ;1000/14.33
6: 0 Offset
; Set negative values to zero
2: If (X<=>F) (P89)
1: 3 X Loc [ Solar_Wm2 ]
2: 4 <
3: 0 F
4: 30 Then Do
3: Z=F x 10^n (P30)
1: 0 F
2: 0 n, Exponent of 10
3: 3 Z Loc [ Solar_Wm2 ]
4: End (P95)
; Calculate units in MJ, where MJ = m * t * 0.000001.
; m = Solar_Wm2 from above, and t = 10 (scan interval).
5: Z=X*F (P37)
1: 3 X Loc [ Solar_Wm2 ]
2: .00001 F
3: 4 Z Loc [ Solar_MJ ]
6: If time is (P92)
1: 0 Minutes (Seconds --) into a
2: 60 Interval (same units as above)
3: 10 Set Output Flag High (Flag 0)
21
CMP6-L, CMP11-L, and CMP21-L Pyranometers
7: Set Active Storage Area (P80)
1: 1 Final Storage Area 1
2: 101 Array ID
8: Real Time (P77)
1: 1220 Year,Day,Hour/Minute (midnight = 2400)
9: Average (P71)
1: 1 Reps
2: 3 Loc [ Solar_Wm2 ]
10: If time is (P92)
1: 0 Minutes (Seconds --) into a
2: 1440 Interval (same units as above)
3: 10 Set Output Flag High (Flag 0)
11: Set Active Storage Area (P80)
1: 1 Final Storage Area 1
2: 102 Array ID
12: Real Time (P77)
1: 1220 Year,Day,Hour/Minute (midnight = 2400)
At regular intervals, physically inspect the pyranometer to ensure that:
• Dome is free of dirt, condensation, and ice (see Section 8.1, Cleaning
• Desiccant granules are orange and opaque (see Section 8.2, Changing the
• Mounting is secure.
• Pyranometer is level (if mounted horizontally).
• Cables are in good condition.
22
Domes).
Desiccant).
8.1 Cleaning Domes
Clean the outer dome at regular intervals (e.g., every week or so). Remove any
accumulated dust, condensation, or ice from the dome and pyranometer body
using a soft cloth dampened with water or alcohol (see FIGURE 8-1).
FIGURE 8-1. Reading is reduced if dome is not dry or clean
8.2 Changing the Desiccant
A desiccant-filled drying cartridge prevents dew from forming on the inner
sides of the domes; Campbell Scientific part number 27052 is the replacement
desiccant for this cartridge. The optional CVF3 Heater/Ventilator Unit is also
available to keep the pyranometer dome free from ice and dew (see Appendix
A, CVF3 Heater/Ventilator). In some applications, the CVF3 may also reduce
the deposition of dust on the pyranometer dome, and therefore reduce the
cleaning interval frequency.
CMP6-L, CMP11-L, and CMP21-L Pyranometers
The silica gel desiccant granules in the drying cartridge should be orange and
opaque. Replace the desiccant granules when they become translucent
(normally after several months). Refill packs of desiccant are shipped with the
pyranometer and can be purchased from Campbell Scientific. The drying
cartridge uses the content of one refill pack. FIGURE 8-2 shows the
replacement process.
When changing the desiccant, ensure that:
•The surfaces touching the rubber o-ring are clean. Dirt, in combination
with water, can cause corrosion, harming it.
•The rubber o-ring is coated with silicon grease or petroleum jelly. The
grease coating improves the o-ring’s seal.
•The drying cartridge is tightly threaded into the pyranometer’s body.
23
CMP6-L, CMP11-L, and CMP21-L Pyranometers
FIGURE 8-2. Changing the desiccant
8.3 Check Sensor Output
It is also important to check the data returned from the sensor as it will show
the first indication of a fault. When doing this you should be aware of several
expected phenomena that can cause strange measurements. In particular on
clear, windless nights the outer dome temperature of horizontally placed
pyranometers can fall as low as the dew point temperature of the air, due to
infrared radiation exchange with the cold sky. (The effective sky temperature
can be 30°C lower than the ground temperature, which results in an infra-red
emission of –150 W
be precipitated on the top of the outer dome and can stay there for several hours
in the morning. An ice cap on the dome is a strong diffuser and can increase
the pyranometer signal by up to 50% in the first hours after sunrise.
8.4 Recalibration
The calibration of the pyranometer may drift with time and exposure to
radiation. Recalibration every two years is recommended. The sensor should
be returned to Campbell Scientific for recalibration. A Returned Materials
Authorization (RMA) is required (refer to the Assistance page for more
information).
–2
•m
). If this happens, dew, glazed frost or hoar frost can
24
9. Troubleshooting
Symptom: NAN, –9999, or radiation values around 0
1. Check that the sensor is wired to the differential channel specified by the
measurement instruction.
2. Verify that the range code is correct for the datalogger type.
CMP6-L, CMP11-L, and CMP21-L Pyranometers
3. Measure the impedance across the red and blue sensor wires. This should
be around 100 ohms plus the cable resistance (typically 0.1 ohm•m
the resistance is very low, there may be a short circuit (check the wiring).
Resistances somewhat lower than expected could be due to water ingress
into the sensor or enclosure connectors. If the resistance is infinite, there
is a broken connection (check the wiring).
4. Disconnect the sensor cable and check the voltage output from the sensor.
With the sensor located 8” below a 60 W incandescent light bulb the
voltage should be approximately 2.5 mV. No voltage indicates a problem
with the sensor.
Symptom: sensor signal is unrealistically high or low
1. Check that the right calibration factor has been properly entered into the
datalogger program. Please note that each sensor has its own individual
calibration factor.
2. Check the condition of the sensor cable.
Symptom: sensor signal shows unexpected variations
1. Check for the presence of strong sources of electromagnetic radiation
(radar, radio, etc.).
–1
). If
2. Check the condition and the connection of the sensor shield wire.
3. Check the condition of the sensor cable.
25
CMP6-L, CMP11-L, and CMP21-L Pyranometers
26
Appendix A. CVF3 Heater/Ventilator
A.1 General Information
The CVF3 consists of a ventilation unit and heaters. The ventilation unit uses a
fan and inlet filter to draw clean air over the pyranometer’s domes. The fan
runs continuously to reduce dust and dirt settling, to dissipate rain drops, and to
stabilize the dome temperature.
The CVF3 has both a 5 W and a 10 W heater. The 5 W heater raises the
temperature of the dome slightly above ambient temperature to prevent the
formation of dew and frost. The 10 W heater is used for more extreme
climates to melt snow and ice.
The 10 W heater’s current drain is approximately 850 mA at 12 Vdc, and the
5 W heater’s current drain is approximately 420 mA at 12 Vdc. The ventilator
draws an additional 5 W of power at 12 Vdc. These power requirements are
large compared to most Campbell Scientific products. Because of this, the
CVF3 should be connected to the 21326 Power Net 5 A Power Supply and the
7321 Crydom Relay.
A.2 Siting
The Crydom relay allows the heater power to be controlled by the datalogger
program and thus reducing power consumption. For example, the datalogger
program can turn on the heater only when the light level falls below 20 W/m
or, if a measurement of air humidity is available, when the dew point of the air
falls to within 1ºC of the sensor body temperature.
Siting information provided in Section 4.1, Siting, is pertinent when using the
CVF3 heater/ventilation. Additionally, the area directly under the CVF3’s
120-mm diameter hole needs to be free from snow, leaves, or other
obstructions that could inhibit the air flow (see FIGURE A-1).
2
FIGURE A-1. Transparent view of CVF3 shows air flow
A-1
Appendix A. CVF3 Heater/Ventilator
A.3 CVF3 Installation
The CVF3 heater/ventilator unit includes the heater/ventilator unit, white
cover, cable, and mounting hardware.
Tools required for mounting to a tripod or tower are:
• Small and medium Phillips screwdrivers
• 5/16”, 1/2” open end wrenches
• 5/32” Allen wrench
• Tape measure
• UV-resistant wire ties
• Side-cut pliers
• Compass
• Step ladder
To install, do the following:
1. Remove leveling screws from the pyranometer.
2. Fit the pyranometer in the upper plate of the CVF3 (see FIGURE A-2).
3. Use the counter sink screws, nylon rings, and nuts to secure the upper plate
of the CVF3 with the lower portion of the unit (see FIGURE A-2).
Nylon Ring
A-2
FIGURE A-2. Pyranometer mounted to the CVF3
Appendix A. CVF3 Heater/Ventilator
4. Loosely mount the pyranometer on the 27084 mounting stand. Do not
fully tighten the two mounting screws.
5. Turn the CVF3’s leveling screws bringing the bubble of the pyranometer’s
level within the ring (see FIGURE A-3).
FIGURE A-3. CVF3 installed onsite
6. Tighten the mounting screws to secure the assembly in its final position.
Check that the pyranometer is still correctly leveled and adjust as
necessary.
7. Use the cover’s screws to fasten the white cover to the pyranometer (see
FIGURE A-4).
A-3
Appendix A. CVF3 Heater/Ventilator
FIGURE A-4. Fastening cover on CVF3
A.4 Wiring
8. Attach the power cable to the CVF3 connector.
Wiring of the CVF3 is shown in TABLE A-1. Refer to Section 7.2, Wiring, for
information about wiring the pyranometer.
TABLE A-1. CVF3 Wiring
Wire Color Description Connection
Red Ventilator Power +12V on 21326 Power Supply
Brown Ventilator Power +12V on 21326 Power Supply
Blue Ventilator Ground G on 21326 Power Supply
Black Ventilator Ground G on 21326 Power Supply
Gray Ventilator Ground G on 21326 Power Supply
Green 5 W Heater Power +12V on 21326 Power Supply
White 5 W Heater Power +12V on 21326 Power Supply
Clear Shield
Yellow 5 V Tacho Output Control port on datalogger
on 21326 Power Supply
A-4
Appendix A. CVF3 Heater/Ventilator
A.5 CVF3 Heater/Ventilator Maintenance
1. Refer to Section 8, Maintenance, for the pyranometer’s maintenance.
2. Inspect the area directly under the 120 mm diameter hole in the mounting
plate to ensure that it is free from leaves, snow, or other obstructions that
can inhibit air flow.
3. Unclip the CVF3’s filter cover and check the filters (see FIGURE A-5).