TheCM3 PYRANOMETER is warranted by CAMPBELL SCIENTIFIC,
INC. to be free from defects in materials and workmanship under normal use
and service for twelve (12) months from date of shipment unless specified
otherwise. Batteries have no warranty. CAMPBELL SCIENTIFIC, INC.'s
obligation under this warranty is limited to repairing or replacing (at
CAMPBELL SCIENTIFIC, INC.'s option) defective products. The customer
shall assume all costs of removing, reinstalling, and shipping defective products
to CAMPBELL SCIENTIFIC, INC. CAMPBELL SCIENTIFIC, INC. will
return such products by surface carrier prepaid. This warranty shall not apply
to any CAMPBELL SCIENTIFIC, INC. products which have been subjected to
modification, misuse, neglect, accidents of nature, or shipping damage. This
warranty is in lieu of all other warranties, expressed or implied, including
warranties of merchantability or fitness for a particular purpose. CAMPBELL
SCIENTIFIC, INC. is not liable for special, indirect, incidental, or
consequential damages.
Products may not be returned without prior authorization. The following
contact information is for US and International customers residing in countries
served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs
for customers wi thin their territories. Please visi t www.campbellsci.com to
determine which Campbell Scientific company serves your country. To obtain
a Returned Materials Authorization (RMA), contact CAMPBELL
SCIENTIFIC, INC., phone (435) 753-2342. After an applications engineer
determines the nature of the problem, an RMA number will be issued. Please
write this number clearly on the outside of the shipping container.
CAMPBELL SCIENTIFIC's shipping address is:
CAMPBELL SCIENTIFIC, INC.
RMA#_____
815 West 1800 North
Logan, Utah 84321-1784
CAMPBELL SCIENTIFIC, INC. does not accept collect calls.
CM3 Pyranometer Table of Contents
PDF viewers note: These page numbers refer to the printed version of this document. Use
the Adobe Acrobat® bookmarks tab for links to specific sections.
1. General Description ....................................................1
5-1. Multipliers Required for Average Flux and Total Flux Density
in SI and English Units..................................................................... 6
i
This is a bla nk page.
CM3 Pyranometer
1. General Description
This manual provides information for interfacing Kipp & Zonen’s CM3
Pyranometer to a CR10(X), CR510, CR23X, CR7 or 21X datalogger.
The CM3 is shipped with an instruction manual provided by Kipp & Zonen that
contains information concerning the CM3’s construction, spectral sensitivity,
cosine response, and a simple sensor check out procedure. Included with the
sensor and manual is a calibration certificate with the sensor calibration
constant and serial number. Cross check this serial number against the serial
number on your CM3 to ensure that the given ca libration co nstant corresponds
to your sensor.
The CM3 pyranometer is designed for continuous outdoor use. Due to its flat
spectral sensitivity from 300 to 3000 nm, it can be used in natural sunlight,
under plant canopies, in green houses or buildings, and inverted to measure
reflected solar radiation. Two CM3s can be used in combination with an
albedometer fixture (K&Z’s CAF 1) to measure albedos. The CM3 can also be
used to measure most types of artificial light (Xenon lamps, Halogen lamps,
etc.).
The CM3 pyranometer consists of a thermopile sensor, a housing, a dome, and
a cable. The thermopile is coated with a black absorbent coating. The paint
absorbs the radiation and converts it to heat. The resultant temperature
difference is converted to a voltage by the copper-constantin thermopile. The
thermopile is encapsulated in the housing in such a way that it has a field of
view of 180 degrees and the angular characteristics needed to fulfill the cosine
response requirements.
2. Specifications
The CM3 is an ISO Second Class pyranometer. While the worst case accuracy
for daily sums given by Kipp & Zonen is +10%, the typical accuracy is +
Tests at Campbell Scientific on one CM3 indicated an accuracy of +2% when
compared to a recently calibrated Eppley PSP.
ISO SPECIFICATIONS:
Response Time 95%:18 seconds
2
Zero offset due to 200 W/m
radiation:< 15 Wm
Zero offset due to temperature change
of 5ºK / hr:< +4 Wm
Temperature Dependence of
sensitivity:+ 6% (-10 to + 40ºC)
Tilt response (+80º) (at 1000 W/m2):< + 2%
OTHER SPECIFICATIONS
Expected accuracy for daily sums:+ 10%
Spectral range (50% points, nm):305-2800 nm
Sensitivity:
Expected signal output in atmospheric
application:0 - 50 mV
10 - 35 µV/Wm
-2
3. Installation
Impedance:
Operating Temperature:-40 to +80ºC
Max. irradiance:2000 Wm
Detector:Copper-constantin multi junction
Cable length:15 feet (5 m)
Level accuracy:1 degree
DIMENSIONS / SHIPPING DIMENSIONS
CM3:3x3x3 in / 6x6x6 in
CM3MT:1x5x5 in / 6x6x6 in
WEIGHT/SHIPPING WEIGHT
CM3:0.8 lbs / 3 lbs
CM3MT:0.6 lbs / 3 lbs
The CM3 should be mounted such that it is never shaded by the tripod/tower or
other sensors.
79 - 200 (Ω)
-2
thermopile
To ensure accurate measurements, the CM3 should be mounted using the
CM3MT ba se/leveling fixture or equivalent. The CM3MT incorporates a
bubble level and three adjustment screws. Mount the CM3 Pyranometer to the
CM3MT mount using the two long screws provided. The screws are slightly
shorter than ideal but should provide adequate holding strength.
2
CM3 Sensor
CM3 Pyranometer
Bubble Level
CM3MT
PN7790(3)
015 Pyranometer Mounting Arm or
025 Pyranometer Crossarm Stand
FIGURE 3-1. CM3 and CM3MT
Install the CM3MT Mount on either the 025 Pyranometer Cross Arm Stand, the
015 Pyranometer Mounting Arm, or the UTKZ (not yet available, please call)
before mounting them to the tower or tripod. This is done by first threading the
screws through the springs and just barely through the mounting plate. This
helps remove any paint that might have gotten into the threaded holes. Second,
remove two of the screws/springs, slide the CM3MT Mount onto the remaining
screw by slightly compressing the spring. Slide the remaining two springs
between the CM3MT and the 015 or 025 mount and install the remaining two
screws. Tighten the screws until all three springs have been compressed about
1/8 inch.
Once the pyranometer mount has been installed on the tripod or tower, tighten
the appropriate screws until the bubble indicates the sensor is level.
015ARM Pyranometer
Mounting Arm
Tripod Mast or Tower Support
FIGURE 3-2. 015 Pyranometer Mounting Arm
3
CM3 Pyranometer
N
N
N
025STAND
Crossarm Stand
019ALU Crossarm
3/4” x 3/4”
U-RAIL
3/4” x 1”
U-RAIL
3/4” x 3/4”
U-RAIL
FIGURE 3-3. 025 Crossarm Stand and 019ALU Crossarm
CM3
UT018 Tower
Mounting Bracket
Tower Support
UTKZ Leveling
Fixture and
Crossarm Mount
(includes CM3MT)
UT018 Crossarm
4. Wiring
4
FIGURE 3-4. UTKZ Leveling Fixture and Crossarm Mount and
UT018 Tower Mounting Bracket and Crossarm
Use Differential Voltage measurement instruction 2 to measure the CM3. The
CM3 wiring diagram is shown in Figure 4-1.
The red lead is connected to the high side (H) of any differential channel. The
blue lead is connected to the corresponding low (L) side of the differential
channel. On a CR10(X) or CR510, the white lead is connected to an analog
ground (AG) and the clear to ground (G). On the CR23X, 21X or CR7 both the
white and clear leads are connected to ground (
).
While a differential measurement is better and preferred, the CM3 can be
measured on a single-ended channel using Instruction 1 if the power induced
voltages discussed in the following paragraph are prevented.
CM3 Pyranometer
If a 21X is used to measure the CM3 and it powers a 12 VDC sensor or 12
VDC radio, the current drawn off the 12 VDC supply may cause a difference in
ground potential between the 21X ground terminals and the reference ground
point in the datalogger. This ground potential results in an offset on single
ended measurements. This offset can be as large as +60 mV. Thus, single
ended measurements should be avoided. The offset does not, however, affect
differential measurements. While the 21X is the logger most susceptible to this
condition, the other dataloggers can be affected if the amount of power is too
large or the power return line is connected to analog ground (AG).
FIGURE 4-1. CM3 Wiring
Color
RedSignalDifferential
BlueSignal ReferenceDifferential
WhiteSignal GroundAG
ClearShieldG
FunctionCR10(X),
CR510
Channel - H
Channel - L
White = Signal Ground
Blue = Signal Reference
Red = Signal
Clear = Shield
21X/CR7CR23X
Differential
Channel - H
Differential
Channel - L
≡≡
≡≡
Differential
Channel - H
Differential
Channel - L
5
CM3 Pyranometer
5. Example Programs
Solar radiation can be reported as an average flux density (W m-2) or daily total
flux density (MJ m
-2
). The appropriate multipliers are listed in Table 5-1.
Programming examples are given for both average and daily total solar
radiation.
-6
The output from the CM3 varies from 10-35 x 10
maximum solar radiation of 1500 W m
-2
, the maximum sensor output voltage
V / W m-2. Given a
will be 15 - 52.5 mV. Example:
-6
(21.87 x 10
V W-1m2) ∗ (1500 W m-2) = 0.03281 V or 32.81 mV
TABLE 5-1. Multipliers Required for Average Flux and
Total Flux Density in SI and English Units
UNITSMULTIPLIERS
-2
W m
MJ m
kJ m
cal cm
cal cm
1
3
C∗
10
-2
-2
-2
-1
min
-2
C∗
C∗
14333
.
C∗
tC∗
t
9
10
t
6
10
6
10
.
002389
6
∗
10
(average)
(total)
(total)
(average)
(total)
C=CM3 calibration: eg. 21.87x10-6V / W m-2
t = datalogger execution interval in seconds
5.1 Average Solar Radiation
Example 1 shows the program instructions used by a CR10X to measure the
signal from the CM3. A sixty-minute average is calculated and stored in final
storage.
3: 1Z Loc [ W_m2 ]
4: End (P95)
5: If time is (P92)
1: 0Minutes (Seconds --) into a
2: 60Interval (same units as above)
3: 10Set Output Flag High (Flag 0)
6: Real Time (P77)
1: 1220Year,Day,Hour/Minute (midnight = 2400)
7: Average (P71)
1: 1Reps
2: 1Loc [ W_m2 ]
*Table 2 Program
02: 0.0000Execution Interval (seconds)
*Table 3 Subroutines
End Program
5.2 Total Solar Radiation
In Example 2 a CR10X is used to record daily total flux density. This total flux
density is in MJ m
5.2.1. Negative values are set to zero before they are added to the running
total.
-2
day-1to avoid the need for high resolution discussed in
7
CM3 Pyranometer
5.2.1 Output Format Considerations
If the solar radiation is totalized in units of kJ m-2, there is a possibility of
over-ranging the output limits. The largest number that the datalogger can
output to final storage is 6999 in low resolution and 99999 in high resolution
(Instruction 78, Set Resolution).
-2
Assume that the daily total flux density is desired in kJ m
irradiance of 0.5 kW m
-2
, the maximum low-resolution output limit will be
. Assume an
exceeded in just under four hours. This value was found by taking the
maximum flux density the datalogger can record in low resolution and dividing
by the total hourly flux density.
2
39
..hr
=
6999
211
−−−
053600
kJm sshr
()()
kJm
−
To circumvent this limitation, record an average flux (see Example 1). Then,
during post processing, multiply the average flux by the number of seconds in
the output interval to arrive at an output interval flux density. Sum the output
interval totals over a day to find a daily total flux density.
Another alternative is to record total flux using the high-resolution format
(Instruction 78, see Datalogger manual for details). The disadvantage of the
high-resolution format is that it requires four bytes of memory per data point,
consuming twice as much memory as low resolution. (Given the larger amount
of memory in today’s dataloggers, this may not matter.)
On a monthly basis the level of the pyranometer should be checked. Any dust
or debris on the sensor window should be removed. The debris can be removed
using water (de-ionized or distilled) or alcohol.
Recalibrati on is suggested every two years. Calibrations can be done in one of
two ways.
One method is to calibrate the sensor in the field by placing a “transfer
standard” (a sensor that has been calibrated against a “secondary standard”)
next to the sensor being calibrated. Preferably daily totals of several days
should be compared. The calibration factor could be corrected if results differ
by more than six percent.
Another method would be to send the sensor to a facility that has a “secondary
standard”. Contact Kipp & Zonen (www.kippzonen.com) for the nearest
calibration facility, or send it to Campbell Scientific and they will have it
recalibrated.
9
CM3 Pyranometer
This is a blank page.
10
This is a bla nk page.
Campbell Scientific Companies
Campbell Scientific, Inc. (CSI)
815 West 1800 North
Logan, Utah 84321
UNITED STATES
www.campbellsci.com
info@campbellsci.com
Campbell Scientific Africa Pty. Ltd. (CSAf)
PO Box 2450
Somerset West 7129
SOUTH AFRICA
www.csafrica.co.za
sales@csafrica.co.za
Campbell Scientific Australia Pty. Ltd. (CSA)
PO Box 444
Thuringo wa Cent ra l
QLD 4812 AUSTRALIA
www.campbellsci.com.au
info@campbellsci.com.au
Campbell Scientific do Brazil Ltda . (CSB)
Rua Luisa Crapsi Orsi, 15 Butantã
CEP: 005543-000 São Paulo SP BRAZIL
www.campbellsci.com.br
suporte@campbellsci.com.br
Campbell Scientific Canada Corp. (CSC)
11564 - 149th Street NW
Edmonton, Alberta T5M 1W7
CANADA
www.campbellsci.ca
dataloggers@campbellsci.ca
Campbell Scientific Ltd. (CSL)
Campbell Park
80 Hathern Road
Shepshed, Loughborough LE12 9GX
UNITED KINGDOM
www.campbellsci.co.uk
sales@campbellsci.co.uk
Campbell Scientific Ltd. (France)
Miniparc du Verger - Bat. H
1, rue de Terre Neuve - Les Ulis
91967 COURTABOEUF CEDEX
FRANCE
www.campbellsci.fr
campbell.scientific@wanadoo.fr
Campbell Scientific Spain, S. L.
Psg. Font 14, local 8
08013 Barcelona
SPAIN
www.campbellsci.es
info@campbellsci.es
Please visit www.campbellsci.com to obtain contact information for your local US or International representative.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.