This BMP5 Direct Software Development Kit software, hereinafter referred to
as the BMP5 Direct SDK, is protected by both United States copyright law and
international copyright treaty provisions. The installation and use of this
software constitutes an agreement to abide by the provisions of this license
agreement. The term "developer" herein refers to anyone using this BMP5
Direct SDK.
By accepting this agreement, you acknowledge and agree that Campbell
Scientific may from time-to-time, and without notice, make changes to one or
more components of the SDK or make changes to one or more components of
other software on which the SDK relies. In no instance will Campbell
Scientific be responsible for any costs or liabilities incurred by you or other
third parties as a result of these changes.
The core operational files included with this BMP5 Direct SDK (hereinafter
referred to as “BMP5 Direct Binaries”) include the files: SimplePB.DLL and
coralib3d.dll. Developer may distribute or sell their software including the
BMP5 Direct Binaries subject to the terms hereafter set forth.
RELATIONSHIP
Campbell Scientific, Inc. hereby grants a license to use BMP5 Direct Binaries
in accordance with the license statement above. No ownership in Campbell
Scientific, Inc. patents, copyrights, trade secrets, trademarks, or trade names is
transferred by this Agreement. Developer may use these BMP5 Direct
Binaries to create as many applications as desired and freely distribute them.
Campbell Scientific, Inc. expects no royalties or any other compensation.
Developer is responsible for supporting applications created using the BMP5
Direct Binaries.
RESPONSIBILITIES OF DEVELOPER
The Developer agrees:
•To provide a competent programmer familiar with Campbell Scientific,
Inc. datalogger programming to write the applications.
•Not to sell or distribute documentation on use of the BMP5 Direct
Binaries.
•Not to sell or distribute the applications that are provided as examples in
the BMP5 Direct SDK.
•Developers may copy and paste portions of the code into their own
applications, but their applications are expected to be unique creations.
•This Agreement does not give Developer the right to sell or distribute any
other Campbell Scientific, Inc. Software (e.g., PC200W, VisualWeather,
LoggerNet or any of their components, files, documentation, etc.) as part
of Developer's application. Distribution of any other Campbell Scientific,
Inc. software requires a separate distribution agreement.
i
•Not to develop applications that compete directly with any application
developed by Campbell Scientific, Inc. or its affiliates.
•To assure that each application developed with BMP5 Direct Binaries
clearly states the name of the person or entity that developed the
application. This information should appear on the first window the user
will see.
WARRANTY
There is no written or implied warranty provided with the BMP5 Direct SDK
software other than as stated herein. Developer agrees to bear all warranty
responsibility of any derivative products distributed by Developer.
TERMINATION
Any license violation or breach of Agreement will result in immediate
termination of the developer's rights herein and the return of all BMP5 Direct
SDK materials to Campbell Scientific, Inc.
MISCELLANEOUS
Notices required hereunder shall be in writing and shall be given by certified or
registered mail, return receipt requested. Such notice shall be deemed given in
the case of certified or registered mail on the date of receipt. This Agreement
shall be governed and construed in accordance with the laws of the State of
Utah, USA. Any dispute resulting from this Agreement will be settled in
arbitration.
This Agreement sets forth the entire understanding of the parties and
supersedes all prior agreements, arrangements and communications, whether
oral or written pertaining to the subject matter hereof. This Agreement shall
not be modified or amended except by the mutual written agreement of the
parties. The failure of either party to enforce any of the provisions of this
Agreement shall not be construed as a waiver of such provisions or of the right
of such party thereafter to enforce each and every provision contained herein.
If any term, clause, or provision contained in this Agreement is declared or
held invalid by a court of competent jurisdiction, such declaration or holding
shall not affect the validity of any other term, clause, or provision herein
contained. Neither the rights nor the obligations arising under this Agreement
are assignable or transferable.
If within 30 days of receiving the BMP5 Direct SDK product developer does
not agree to the terms of license, developer shall return all materials without
retaining any copies of the product and shall remove any use of the BMP5
Direct Binaries in any applications developed or distributed by Developer. In
the absence of such return, CSI shall consider Developer in agreement with the
herein, stated license terms and conditions.
ii
Limited Warranty
The following warranties are in effect for ninety (90) days from the date of
shipment of the original purchase. These warranties are not extended by the
installation of upgrades or patches offered free of charge:
Campbell Scientific warrants that the installation media on which the software
is recorded and the documentation provided with it are free from physical
defects in materials and workmanship under normal use. The warranty does not
cover any installation media that has been damaged, lost, or abused. You are
urged to make a backup copy (as set forth above) to protect your investment.
Damaged or lost media is the sole responsibility of the licensee and will not be
replaced by Campbell Scientific.
Campbell Scientific warrants that the software itself will perform substantially
in accordance with the specifications set forth in the instruction manual when
properly installed and used in a manner consistent with the published
recommendations, including recommended system requirements. Campbell
Scientific does not warrant that the software will meet licensee’s requirements
for use, or that the software or documentation are error free, or that the
operation of the software will be uninterrupted.
Campbell Scientific will either replace or correct any software that does not
perform substantially according to the specifications set forth in the instruction
manual with a corrected copy of the software or corrective code. In the case of
significant error in the installation media or documentation, Campbell
Scientific will correct errors without charge by providing new media, addenda,
or substitute pages. If Campbell Scientific is unable to replace defective media
or documentation, or if it is unable to provide corrected software or corrected
documentation within a reasonable time, it will either replace the software with
a functionally similar program or refund the purchase price paid for the
software.
All warranties of merchantability and fitness for a particular purpose are
disclaimed and excluded. Campbell Scientific shall not in any case be liable for
special, incidental, consequential, indirect, or other similar damages even if
Campbell Scientific has been advised of the possibility of such damages.
Campbell Scientific is not responsible for any costs incurred as a result of lost
profits or revenue, loss of use of the software, loss of data, cost of re-creating
lost data, the cost of any substitute program, telecommunication access costs,
claims by any party other than licensee, or for other similar costs.
This warranty does not cover any software that has been altered or changed in
any way by anyone other than Campbell Scientific. Campbell Scientific is not
responsible for problems caused by computer hardware, computer operating
systems, or the use of Campbell Scientific’s software with non-Campbell
Scientific software.
Licensee’s sole and exclusive remedy is set forth in this limited warranty.
Campbell Scientific’s aggregate liability arising from or relating to this
agreement or the software or documentation (regardless of the form of action;
e.g., contract, tort, computer malpractice, fraud and/or otherwise) is limited to
the purchase price paid by the licensee.
Campbell Scientific may from time-to-time, and without notice, make changes
to one or more components of the SDK or make changes to one or more
components of other software on which the SDK relies. In no instance will
Campbell Scientific be responsible for any costs or liabilities incurred by you
or other third parties as a result of these changes. There is no written or implied
warranty provided with the SDK software other than as stated herein.
Developer agrees to bear all warranty responsibility of any derivative products
distributed by Developer.
BMP5 Direct SDK Table of Contents
PDF viewers note: These page numbers refer to the printed version of this document. Use
the Adobe Acrobat® bookmarks tab for links to specific sections.
1. BMP5 Direct SDK Overview .....................................1-1
1.1 General Notes on BMP5 Direct SDK Usage ........................................ 1-1
1.2 Datalogger Program Table Structure .................................................... 1-1
A. Sample Program Table Structure .......................... A-1
v
Section 1. BMP5 Direct SDK Overview
The BMP5 Direct Software Development Kit (SDK) is a programming
interface that facilitates simple and direct communication with a single
datalogger containing a PakBus operating system. This SDK uses a simple
call-level API (SimplePB.DLL) that does not need to be registered on the PC.
However, the SimplePB.DLL wrapper accesses and therefore requires the
included communications engine, CORALIB3D.DLL, to be installed in the
same folder.
The BMP5 Direct SDK allows the creation of basic BMP5 application packets
that are sent to PakBus dataloggers over the PakBus network. Examples and
controls are installed by default in C:\Campbellsci\BMP5DirectSDK.
1.1 General Notes on BMP5 Direct SDK Usage
In order to communicate with the datalogger, a direct connection must be
created using the SimplePB.DLL. The BMP5 Direct SDK allows a connection
to a single datalogger through either an IP port or a COM port using the
OpenIPPort() or OpenPort() command.
Opening the COM or IP port starts the CORALIB3D.DLL. Therefore, the
COM port or IP port opened by the application must be closed before exiting
the application or the CORALIB3D.DLL will crash. Use the functions
CloseIPPort() or ClosePort() before the application exits to stop the
CORALIB3D.DLL and to properly close the connection. Starting the
CORALIB3D.DLL creates a working directory including log files by default in
C:\Campbellsci\SimplePB.
Once a connection is established, additional commands can be used to access
and administer the datalogger. A datalogger program can be sent, the clock can
be checked or set, values can be set, and data can be collected. Use a
combination of the commands found in the SimplePB.DLL to accomplish the
desired task.
1.2 Datalogger Program Table Structure
One default table in a PakBus datalogger is the Status table. The datalogger
may also include either a Public table or an Inlocs table. The Status table
contains values describing the datalogger and datalogger program and the
Public or Inlocs table stores all public variables or input locations. The user
defines any additional tables in the program sent to the datalogger.
When using the SimplePB.DLL to access specific tables and fields in a
datalogger program, the table number and the field number are implied by their
respective positions. For example, table number one is the first table in the
datalogger program and field number one is the first field following the
mandatory record number and timestamp fields within the table. The user must
understand the table structure of the program running in the datalogger because
table and field names and numbers are used to identify and access specific
tables and fields in several of the SimplePB.DLL commands.
1-1
Section 1. BMP5 Direct SDK Overview
1-2
Section 2. SimplePB.DLL Reference
The SimplePB.DLL is a call-level API that does not need to be registered on
the PC. However, the SimplePB.DLL wrapper accesses and therefore requires
the included communications engine, CORALIB3D.DLL, to be installed in the
same folder. The SimplePB wrapper provides an easy interface through the
CORALIB3D.DLL communications engine to access a single datalogger. The
following commands are available in the SimplePB.DLL:
2.1 OpenPort()
Opens a COM port to allow a direct connection to the datalogger.
Syntax
OpenPort(com_port_no, baud)
Parameters
com_port_no: Integer – The COM port to open.
baud: Integer – The baud rate used by the COM port.
Return Codes
2.2 ClosePort()
Syntax
Return Codes
0: Integer – Successful.
-1: Integer – COM port failed to open or is already open.
Closes the previously opened COM port.
ClosePort()
0 = Successful.
-1 = COM port failed to close or was not open.
2-1
Section 2. SimplePB.DLL Reference
2.3 OpenIPPort()
Opens an IP port to allow a connection to a datalogger.
Syntax
OpenIPPort(ip_address, tcp_port)
Parameters
ip_address: String – A valid IP address for the datalogger.
tcp_port: Integer – The port that will be used when communicating with the
datalogger.
Return Codes
0 = Successful.
-1 = IP port failed to open or is already open.
2.4 CloseIPPort()
Closes the previously opened IP port.
Syntax
CloseIPPort()
Return Codes
0 = Successful.
-1 = IP port failed to close or was not open.
2.5 GetClock()
Query the datalogger for its current date and time.
pakbus_address: Integer – The PakBus address of the datalogger.
device_type: Integer –The type of datalogger:
• 1=CR200
• 2=CR10XPB, CR23XPB, CR510PB
• 3=CR1000
• 4=CR3000
• 5=CR800, CR850
swath: Integer – The range within an indexed variable array to collect starting
at the location described in the field_name parameter. The requested swath
must be within the bounds of the indexed array or an error will occur.
table_name: String – The name of the table in the datalogger to collect.
2-4
field_name: String – The name of the field in the table where collection of
values should start.
return_data: Char – The location in memory where the data returned from the
datalogger exists.
Return Codes
Example of data returned by function call
2.8 SetValue()
Syntax
Section 2. SimplePB.DLL Reference
return_data_len: Integer – Number of bytes in the data returned from the
datalogger.
0 = Successful.
-1 = Communication timed out.
-2 = Port is not open.
12.753,111.9,1.239 (Swath of 3 values from fields)
pakbus_address: Integer – The PakBus address of the datalogger.
device_type: Integer – The type of datalogger:
• 1=CR200
• 2=CR10XPB, CR23XPB, CR510PB
• 3=CR1000
Return Codes
• 4=CR3000
• 5=CR800, CR850
table_no: Integer – The number of the table from which to collect data.
record_no: Integer – The record number where data collection will start. All
records following this record number will be included in the collection.
Therefore, if the record number is set to 0, all records in the table will be
collected. In addition, if the record number specified does not exist in the
datalogger, all existing records from the oldest to the newest will be returned.
However, if the record number is set to a negative number, only the most
recent record in the table will be collected. There is not a way to specify and
collect a single record from a table using this command unless that record is the
most recent record in the table.
return_data: Char – The location in memory where the data returned from the
datalogger exists.
return_data_len: Integer – Number of bytes in the data returned from the
datalogger.
0 = Successful.
1 = Success but more data to collect.
pakbus_address: Integer – The PakBus address of the datalogger.
device_type: Integer – The type of datalogger:
• 1=CR200
• 2=CR10XPB, CR23XPB, CR510PB
• 3=CR1000
• 4=CR3000
• 5=CR800, CR850
table_no: Integer – The number for the table from which to collect data.
record_no: Integer – The record number where data collection will start. All
records following this record number will be included in the collection.
Therefore, if the record number is set to 0, all records in the table will be
collected. In addition, if the record number specified does not exist in the
datalogger, all existing records from the oldest to the newest will be returned.
However, if the record number is set to a negative number, only the most
recent record in the table will be collected. There is not a way to specify and
collect a single record from a table using this command unless that record is the
most recent record in the table.
2-8
Return Codes
Example of data returned by function call
2.12 File_Send()
Section 2. SimplePB.DLL Reference
return_data: Char – The location in memory where the data returned from the
datalogger exists.
return_data_len: Integer – Number of bytes in the data returned from the
datalogger.
0 = Successful.
1 = Success but more data to collect.
pakbus_address: Integer – The Pakbus address of the datalogger.
device_type: Integer – The type of datalogger:
• 1=CR200
• 2=CR10XPB, CR23XPB, CR510PB
• 3=CR1000
• 4=CR3000
• 5=CR800, CR850
file_name: String – The path and file name of the program to be sent.
return_data: Char – The location in memory where the data returned from the
datalogger exists.
return_data_len: Integer – Number of bytes in the data returned from the
datalogger.
2-9
Section 2. SimplePB.DLL Reference
Return Codes
0 = Successful.
1 = Success but more data to transfer.
-1 = Communication timed out.
-2 = Port is not open.
-3 = Cannot open source file.
-4 = File name is to long.
-5 = Datalogger timed out.
-6 = File offset does not match.
-7 = Datalogger reported an error.
-8 = File control error.
-9 = Cannot get program status.
Example of data returned from a CR1000
OS Version : CR1000.Std.05
OS Signature : 19128
Serial Number : 1031
PowerUp Progr : CPU:Program.cr1
Compile Status: Datalogger Program Running
Program Name : CPU:Program.cr1
Program Sig. : 32083
Compile Result: Compiled in SequentialMode.
pakbus_address: Integer – The PakBus address of the datalogger.
device_type: Integer – The type of datalogger:
• 1=CR200
• 2=CR10XPB, CR23XPB, CR510PB
• 3=CR1000
• 4=CR3000
• 5=CR800, CR850
return_data: Char – The location in memory where the data returned from the
datalogger exists.
return_data_len: Integer – Number of bytes in the data returned from the
datalogger.
Return Codes
0 = Successful.
-1 = Communication timed out.
-2 = Cannot read table definitions from the datalogger.
Example of data returned by function call
1 Status
2 DataTable1
3 DataTable2
4 Public
2-12
2.16 GetDLLVersion()
Get the version of the SimplePB.dll being used.
Syntax
GetDLLVersion(return_data, return_data_len)
Parameters
return_data: Char – The location in memory where the data returned from the
datalogger exists.
return_data_len: Integer – Length of the data returned from the DLL.
Return Codes
0 = Successful
Section 2. SimplePB.DLL Reference
Example of data returned by function call
SimplePB.DLL Version 2.0 / 2,2,3,0I
2.17 GetLastResults()
Retrieves the return_data results from memory for the previous function as a
String. This function is useful for developers that don’t want to manage
memory pointers.
Syntax
GetLastResults()
Return Codes
0 = Successful
2.18 FileControl()
Used to manage programs and files on the datalogger.
pakbus_address: Integer - The PakBus address of the datalogger.
device_type: Integer - The type of datalogger:
1=CR200
2=CR10XPB, CR23XPB, CR510PB
3=CR1000
4=CR3000
5=CR800, CR850
file_name: String - The path and file name to be controlled on the datalogger.
command: Integer - The command issued to control the file:
1 = Compile and run
2 = Run on power up
3 = Make hidden
4 = Delete file
5 = Format device
6 = Compile and run (Preseve data if no table changed)
7 = Stop program
8 = Stop program and delete
9 = Make the specified file the operating system
10 = Compile and run but do not change the "run on power up" program
11 = Pause
12 = Resume
13 = Stop the running program, delete associated files, and mark as run
14 = Stop the running program, delete associated files, and mark as run
now and on power up
now but not on power up
Return Codes
0 = Successful
-1 = Communication timed out
-2 = Port is not open
2-14
Appendix A. Sample Program Table
Structure
The table structure of a PakBus datalogger is given in the example below. This
example shows a datalogger with two user defined tables plus the Status table
and Public or Inlocs table. The second table in the example below contains
three records and the third table contains four records. Both the Status table
and Public or Inlocs table will always return the most recent records and will
not contain any historical data records.
The first table is the Status table, which shows the status of the datalogger.
The Public or Inlocs table contains all public variables or input locations. All
other tables found in the datalogger are created and defined by the user in the
datalogger program. The tables in a PakBus datalogger will always contain a
record number and timestamp followed by the data fields.
Table 1 – Status
Record No Time Stamp Data Field 1 Data Field 2 Data Field 3-19 Data Field 20
Table 2 – User Defined
RN 0 Time Stamp Data Field 1 Data Field 2
RN 1 Time Stamp Data Field 1 Data Field 2
RN 2 Time Stamp Data Field 1 Data Field 2
Table 3 – User Defined
RN 0 Time Stamp Data Field 1 Data Field 2 Data Field 3 Data Field 4 Data Field 5
RN 1 Time Stamp Data Field 1 Data Field 2 Data Field 3 Data Field 4 Data Field 5
RN 2 Time Stamp Data Field 1 Data Field 2 Data Field 3 Data Field 4 Data Field 5
RN 3 Time Stamp Data Field 1 Data Field 2 Data Field 3 Data Field 4 Data Field 5
Table 4 – Public or Inlocs
Record No Time Stamp Data 1 Data 2 Data 3 Data 4 Data 5 Data 6
A-1
Appendix A. Sample Program Table Structure
CR200 Datalogger Program Tables
The following tables show the table structure from a program installed in a
CR200 datalogger. This program measures and stores the minimum battery
voltage and the minimum and maximum temperature over a 60-minute
interval. When communicating with a datalogger using the BMP5 Direct SDK,
knowing the table structure of the running program is necessary for some
commands.
Table Number 1 – Status
Field Number Field Name Units Notes:
Field 1 OSVersion Operating System Version
Field 2 OSDate Date of Operating System
Field 3 ProgName The name of the program running in the datalogger
Field 4 ProgSig The signature of the running program
Field 5 CalOffset
Field 6 PakBusAddress The PakBus address of the datalogger (1-4094)
Field 7 RfInstalled Radio Detected
Field 8 RfNetAddr Valid Addresses are 0-63
Field 9 RfAddress Valid Addresses are 0-1023
Field 10 RfHopSeq Valid numbers are 0-6
Field 11 RfPwrMode RF1_Sec
Field 12 Rf_ForceOn
Field 13 RfSignalLevel RF signal strength should be above 40
Field 14 RfRxPakBusCnt
Field 15 VarOutOfBounds
Field 16 SkipScan Program didn’t complete before the next execution interval
Field 17 TrapCode
Field 18 WatchDogCnt Number of Watchdog Errors
Field 19 ResetTables Clears All Stored Data
Field 20 BattVoltage Volts Current Battery Voltage
A-2
Appendix A. Sample Program Table Structure
Table Number 2 – Hourly: The Hourly table contains the minimum battery voltage and the minimum and
maximum temperature over a 60-minute interval.
Field Number Field Name Units Notes:
Field 1 Battery_Min Volts
Field 2 Battery_Time Time
Field 2 Temp_Min Deg C
Field 3 Temp_Max Deg C
Table Number 3 – Public: The Public table contains only the most recent “real-time” record for the variable
described in the datalogger program.
'Default Datalogger Battery Voltage measurement Batt_Volt:
Battery(Batt_Volt)
'109 Temperature Probe measurement Temp:
Therm109(Temp,1,1,1,1.0,0.0)
'Call Data Tables and Store Data
CallTable(Hourly)
NextScan
EndProg
A-3
Appendix A. Sample Program Table Structure
WeatherHawk Weather Station Tables
The following tables show the table structure from a default WeatherHawk
weather station program installed in a CR200 datalogger. When
communicating with a datalogger using the BMP5 Direct SDK, knowing the
table structure of the running program is necessary for some commands
Table Number 1 – Status
Field Number Field Name Units Notes:
Field 1 OSVersion Operating System Version
Field 2 OSDate Date of Operating System
Field 3 ProgName The name of the program running in the datalogger
Field 4 ProgSig The signature of the running program
Field 5 CalOffset
Field 6 PakBusAddress The PakBus address of the datalogger (1-4094)
Field 7 RfInstalled Radio Detected
Field 8 RfNetAddr Valid Addresses are 0-63
Field 9 RfAddress Valid Addresses are 0-1023
Field 10 RfHopSeq Valid numbers are 0-6
Field 11 RfPwrMode RF1_Sec
Field 12 Rf_ForceOn
Field 13 RfSignalLevel RF signal strength should be above 40
Field 14 RfRxPakBusCnt
Field 15 VarOutOfBounds
Field 16 SkipScan Program didn’t complete before the next execution interval
Field 17 TrapCode
Field 18 WatchDogCnt Number of Watchdog Errors
Field 19 ResetTables Clears All Stored Data
Field 20 BattVoltage Volts Current Battery Voltage
A-4
Appendix A. Sample Program Table Structure
Table Number 2 – SiteVal: The SiteVal table contains values that are stored for calculations by the
WeatherHawk program. Data is only stored when field “SaveSite” in the Public table is set to one.
Field Number Field Name Units Notes:
Field 1 Altitude_m Meter
Field 2 Latitude Degrees
Field 3 Longitude Degrees
Field 4 BPoffset_KPa KPa
Field 5 Int_Timer Minutes
Table Number 3 – Data1: This table contains data ouput at the Int_timer rate from the Public table. For example,
if Int_timer = 15 min, this table contains 15 min data.
Field Number Field Name Units Notes:
Field 1 BatVolt_V Volts
Field 2 BatVolt_V_Min Volts
Field 3 AirTemp_C_Avg Celsius
Field 4 RH_Avg Percent
Field 5 WindSpeed_ms_Avg m/s
Field 6 Solar_Avg W/m^2
Field 7 ETo mm
Field 8 AirTemp_C_Min Celsius
Field 9 AirTemp_C_TMn Time Example: 2004-01-25 13:49:50
Field 10 Max_AirTemp Celsius
Field 11 AirTemp_C_C_Tmx Time Example: 2004-01-25 13:49:50
Field 12 WindSpeed_ms_WVc(1) m/s Average Wind Speed
Field 13 WindSpeed_ms_WVc(2) Degrees Unit Vector Wind Direction
Field 14 WindSpeed_ms_Max m/s
Field 15 Baromete_KPa KPa
Field 16 RainYearly_mm mm
A-5
Appendix A. Sample Program Table Structure
Table Number 4 – Data2: This table contain daily data values.
Field Number Field Name Units Notes:
Field 1 BatVolt_V_Min Volts
Field 2 AirTemp_C_Max Celsius
Field 3 AirTemp_C_Min Celsius
Field 4 WindSpeed_ms_Max m/s
Field 5 RainYearly_mm mm
Field 6 DailyETo_mm mm
Table Number 5 – Public: The public table contains only the most recent “real-time” record.
Field Number Field Name Units Notes:
Field 1 SaveSite Set to one to save values to SiteVal table
Field 2 Latitude Degrees Decimal format: 41 deg 45 min = 14.75
Field 3 Longitude Degrees Decimal format: 41 deg 45 min = 14.75
Field 4 Altitude_m Meter
Field 5 Bpoffset_KPa KPa
Field 6 Int_timer Minutes
Field 7 RainReset Set to 1 to reset RainYearly_mm variable
Field 8 BatVolt_V Volts
Field 9 Solar W/m^2
Field 10 AirTemp_C Celsius
Field 11 RH Percent
Field 12 Barometer_KPa KPa Sea level adjustment barometric pressure
Field 13 WindSpeed_ms m/s
Field 14 WindDirect_deg Degrees
Field 15 RainYearly_mm mm Running Sum of Rainfall
Field 16 DailyETo_mm mm Running Sum of Eto (Resets at midnight)