Campbell Scientific AVW200 User Manual

INSTRUCTION MANUAL
AVW200-series 2-Channel
Vibrating Wire Spectrum
Copyright © 2008- 2014
Campbell Scientific, Inc.
Analyzer Modules

Warranty

“PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless otherwise specified in the corresponding Campbell pricelist or product manual. Products not manufactured, but that are re-sold by Campbell, are warranted only to the limits extended by the original manufacturer. Batteries, fine-wire thermocouples, desiccant, and other consumables have no warranty. Campbell’s obligation under this warranty is limited to repairing or replacing (at Campbell’s option) defective products, which shall be the sole and exclusive remedy under this warranty. The customer shall assume all costs of removing, reinstalling, and shipping defective products to Campbell. Campbell will return such products by surface carrier prepaid within the continental United States of America. To all other locations, Campbell will return such products best way CIP (Port of Entry) INCOTERM® 2010, prepaid. This warranty shall not apply to any products which have been subjected to modification, misuse, neglect, improper service, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied. The warranty for installation services performed by Campbell such as programming to customer specifications, electrical connections to products manufactured by Campbell, and product specific training, is part of Campbell’s product warranty. CAMPBELL EXPRESSLY DISCLAIMS AND EXCLUDES ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Campbell is not liable for any special, indirect, incidental, and/or consequential damages.”

Assistance

Products may not be returned without prior authorization. The following contact information is for US and international customers residing in countries served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs for customers within their territories. Please visit
www.campbellsci.com to determine which Campbell Scientific company serves
your country.
To obtain a Returned Materials Authorization (RMA), contact CAMPBELL SCIENTIFIC, INC., phone (435) 227-9000. After an application engineer determines the nature of the problem, an RMA number will be issued. Please write this number clearly on the outside of the shipping container. Campbell Scientific’s shipping address is:
CAMPBELL SCIENTIFIC, INC. RMA#_____ 815 West 1800 North Logan, Utah 84321-1784
For all returns, the customer must fill out a “Statement of Product Cleanliness and Decontamination” form and comply with the requirements specified in it. The form is available from our web site at www.campbellsci.com/repair. A completed form must be either emailed to repair@campbellsci.com or faxed to (435) 227-9106. Campbell Scientific is unable to process any returns until we receive this form. If the form is not received within three days of product receipt or is incomplete, the product will be returned to the customer at the customer’s expense. Campbell Scientific reserves the right to refuse service on products that were exposed to contaminants that may cause health or safety concerns for our employees.

Table of Contents

PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to specific sections.
1. Overview ...................................................................... 1
1.1 Design Features .................................................................................... 1
1.2 Specifications ....................................................................................... 3
1.3 Communication .................................................................................... 3
1.3.1 Datalogger ..................................................................................... 3
1.3.1.1 PakBus Protocol/Direct RS-232 Connection ...................... 4
1.3.1.2 PakBus Protocol/Wireless Connection ............................... 4
1.3.1.3 PakBus Protocol/MD485 Communication ......................... 4
1.3.1.4 SDI-12 Communication Mode ........................................... 4
1.3.2 Computer ...................................................................................... 5
1.3.2.1 Device Configuration Utility .............................................. 5
1.3.2.2 LoggerNet .......................................................................... 5
1.3.2.3 Terminal Commands .......................................................... 5
1.4 Use with Multiplexers .......................................................................... 5
2. Measurements ............................................................. 7
2.1 Vibrating Wire ..................................................................................... 7
2.2 Temperature ....................................................................................... 10
3. Quick Start Guides ................................................... 11
3.1 One or Two Sensors (no multiplexers) ............................................... 11
3.1.1 Direct RS-232 Connection .......................................................... 11
3.1.2 Wireless Connection ................................................................... 12
3.2 Multiplexers Controlled by AVW200 ................................................ 14
3.2.1 Direct RS-232 Connection .......................................................... 14
3.2.2 Wireless Connection ................................................................... 15
3.3 Multiplexers Controlled by Datalogger .............................................. 17
3.3.1 SDI-12 Communication .............................................................. 17
4. Connections .............................................................. 18
4.1 Sensor Wiring (no multiplexers) ........................................................ 18
4.2 Power and Ground .............................................................................. 19
4.3 Datalogger Wiring (Direct Connection) ............................................. 20
4.4 Wireless Connections (AVW206, AVW211, AVW216) ................... 21
4.5 Multiplexer Wiring ............................................................................. 22
4.5.1 AVW200 Controlling the Multiplexer ........................................ 22
4.5.2 Datalogger Controlling the Multiplexer ...................................... 23
5. Device Configuration (DevConfig) Utility ............... 25
5.1 Connecting to DevConfig................................................................... 25
5.2 Deployment Tab ................................................................................. 26
5.2.1 Communications ......................................................................... 26
5.2.2 Measurement ............................................................................... 29
5.3 Data Monitor ...................................................................................... 30
i
Table of Contents
5.4 Send OS ...................................................................................... 32
5.5 Troubleshoot .............................................................................. 33
5.6 Settings Editor ............................................................................ 36
5.7 Terminal ..................................................................................... 36
6. Programming ............................................................. 37
6.1 AVW200 Instruction ......................................................................... 37
6.1.1 Pipeline Mode ............................................................................ 41
6.1.2 Sequential Mode ......................................................................... 41
6.2 SDI-12 Measurements ....................................................................... 42
6.2.1 SDI12 Recorder() Instruction ..................................................... 42
6.2.2 Extended SDI-12 Commands ..................................................... 44
6.2.3 Use with Multiplexers ................................................................ 44
7. Example Programs .................................................... 45
7.1 AVW200() Instruction (no multiplexers) .......................................... 45
7.1.1 Direct RS-232 Connection (two sensors) ................................... 45
7.1.2 Wireless/One Sensor/Resistance Converted to Temperature ..... 46
7.2 AVW200() Instruction Controlling Two Multiplexers ...................... 47
7.3 AVW200( ) Instruction Running in the Pipeline Mode ..................... 48
7.4 AVW200( ) Instruction Running in the Sequential Mode ................. 49
7.4.1 AVW200 Controlling Two Multiplexers in Sequential Mode ... 50
7.4.2 Datalogger Controlling Two Multiplexers in Sequential Mode . 50
7.5 SDI-12 Example ................................................................................ 51
8. Troubleshooting Communication Problems .......... 54
8.1 Unable to Communicate with DevConfig or Terminal Emulator ...... 54
8.2 Datalogger to AVW200 Communication .......................................... 54
8.3 Wireless Communications ................................................................. 54
Appendices
Conversion from Hertz ........................................... A-1
A.
A.1 Displacement Example .................................................................... A-1
B. Thermistor Information ........................................... B-1
B.1 Converting Resistance to Temperature ............................................ B-1
B.1.1 Resistance Conversion Example – Geokon Sensor .................. B-1
B.2 Accuracy and Resolution ................................................................. B-1
C. Antennas, Antenna Cables, and Surge
Protectors for the AVW206, AVW211, and
AVW216 ................................................................. C-1
C.1 Antennas .......................................................................................... C-1
C.2 Antenna Cables ................................................................................ C-5
C.3 Surge Protectors .............................................................................. C-5
C.3.1 Electrostatic Issues ................................................................... C-5
C.3.2 Antenna Surge Protector Kit .................................................... C-6
C.4 Part 15 FCC Compliance Warning .................................................. C-6
ii
Table of Contents
D. The Public Table ..................................................... D-1
D.1 Forced Measurement Program ........................................................ D-4
E. Status Table ............................................................ E-1
F. Time Series and Spectrum Graph Information .... F-1
F.1 Good Sensor Examples .................................................................... F-1
F.2 Good Sensors with Noise ................................................................. F-3
G. CR10X Programming Example .............................. G-1
H. Additional Programming Examples ...................... H-1
H.1 AVW200-Controlled Multiplexer ................................................... H-1
H.1.1 Direct RS-232 Connection ....................................................... H-1
H.1.2 Wireless/Sensors with Different Frequencies .......................... H-4
H.2 Datalogger-Controlled Multiplexer ................................................. H-7
I. Using MD485 Multidrop Modems with AVW200
Interfaces ................................................................. I-1
Figures
I.1 Required Settings .............................................................................. I-1
I.2 Connections ....................................................................................... I-2
I.2.1 Datalogger to MD485 ................................................................ I-3
I.2.2 MD485 to MD485 ...................................................................... I-3
I.2.3 MD485 to AVW200 ................................................................... I-4
I.2.4 Multiplexer Connections ............................................................ I-4
I.3 Programming ..................................................................................... I-4
I.3.1 Example Program ....................................................................... I-4
1-1. Network of AVW200s and AM16/32Bs (using a direct RS-232
connection) ....................................................................................... 6
1-2. Network of AVW206s and AM16/32Bs (wireless) ............................. 6
1-3. Network of AVW200 Interfaces (SDI-12) ........................................... 7
2-1. Cutaway of Vibrating Wire Sensor ...................................................... 8
2-2. DevConfig plots showing the AVW200 measurement approach. ...... 10
4-1. Wiring for Sensor Connections .......................................................... 19
4-2. Ground Lug and Power Connectors on the AVW200 ........................ 20
4-3. AVW206 with Whip Antenna ............................................................ 21
4-4. Example AM16/32-series to AVW200 Hookup (multiplexers
controlled by AVW200) ................................................................. 23
4-5. AM16/32B to AVW200 Hookup (AM16/32Bs controlled by
datalogger and using SDI-12) ......................................................... 24
4-6. Multiplexer to Datalogger Power/Control Hookup (multiplexer
controlled by datalogger) ................................................................ 24
5-1. Opening Page in DevConfig .............................................................. 26
5-2. Deployment Communications Editor in DevConfig .......................... 27
5-3. Deployment/Measurement Tab in DevConfig ................................... 29
5-4. Data Display/Public Table in DevConfig ........................................... 30
5-5. Data Display/Status Table in DevConfig ........................................... 31
5-6. Sending an Operating System using DevConfig ................................ 32
iii
Table of Contents
5-7. Opening Page of the Troubleshoot Tool ............................................ 33
5-8. Options Tab of the Troubleshoot Tool .............................................. 34
5-9. Graphs for Evaluating Spectral Analysis of a Sensor ........................ 35
A-1. Geokon Calibration Report of a Sensor without a Thermistor ........ A-2
B-1. Temperature Measurement Error at Three Temperatures as a
Function of Lead Length. Wire is 22 AWG with 16 ohms
per 1000 feet. ............................................................................... B-2
B-2. Temperature Measurement Error on a 1000 foot Lead. Wire is
22 AWG with 16 ohms per 1000 feet. ......................................... B-3
B-3. Temperature Measurement Error on a 3000 foot Lead. Wire is
22 AWG with 16 ohms per 1000 feet. ......................................... B-3
B-4. Temperature Measurement Error on a 5000 foot Lead. Wire is
22 AWG with 16 ohms per 1000 feet. ......................................... B-4
C-1. Some FCC Approved Antennas ...................................................... C-4
C-2. Example COAX RPSMA-L Cable for Yagi or Omni Colinear ....... C-5
C-3. Antenna Surge Protector .................................................................. C-5
F-1. Good Sensor with a Narrower Range (200 to 2200 Hz) ................... F-2
F-2. Good Sensor with a Wider Range (200 to 6500 Hz) ........................ F-2
F-3. Good Sensor with Noise (600 to 1800 Hz) ....................................... F-3
F-4. Good Sensor with Noise (450 to 6500 Hz) ....................................... F-4
I-1. Deployment tab in DevConfig with proper MD485 configuration.... I-1
I-2. Point-to-Point MD485-to-AVW200 Network ................................... I-2
I-3. Point-to-Multipoint MD485-to-AVW200 Network .......................... I-2
I-4. MD485 and its connectors. ................................................................ I-3
I-5. MD485-to-MD485 Connections and Grounding ............................... I-4
Tables
2-1. AVW200 Measurement Inputs ............................................................ 8
2-2. AVW200 Measurement Outputs ......................................................... 9
4-1. Datalogger to AVW200 Cable Options ............................................. 20
4-2. 17855 or SC110’s DTE Cable Wiring ............................................... 21
4-3. Datalogger to Spread Spectrum Radio Connections .......................... 22
5-1. AVW206 Power Modes and the Recommended Corresponding
RF401 Power Modes...................................................................... 28
5-2. Terminal Mode Commands ............................................................... 36
6-1. SDI-12 Command Codes ................................................................... 43
7-1. Wiring for Example 7.1.1 .................................................................. 45
7-2. Wiring for Example 7.2 ..................................................................... 47
7-3. Wiring for Example 7.3 ..................................................................... 48
7-4. Wiring for Sequential Mode Examples ............................................. 50
7-5. SDI-12 Command Codes ................................................................... 52
D-1. Description of the Public Table ....................................................... D-1
E-1. Status Fields and Descriptions .......................................................... E-1
iv
CAUTION
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
The AVW200 series consist of a base model (AVW200) and three wireless models (AVW206, AVW211, AVW216). The wireless models combine the AVW200 with a spread spectrum radio. The different model numbers of the wireless versions are for different spread spectrum frequency ranges.
Compatible Radios
AVW206—910 to 918 MHz (US/Canada) RF401
AVW211—920 to 928 MHz (Australia/Israel) RF411
AVW216—2.450 to 2.482 GHz (worldwide) RF416
No product using the 24XStream radio, including the AVW216, will be available for sale in Europe after 1/1/2015 due to changes in EU legislation. Consequently, purchase of the AVW216 is not recommended for use in Europe in new networks that may require future expansion.
Throughout this manual AVW200 will refer to all models unless specified otherwise. Likewise, AVW206 typically refers to all wireless models, and RF401 refers to the corresponding spread spectrum radio.

1. Overview

The AVW200 module allows the measurement of vibrating-wire strain gauges, pressure transducers, piezometers, tiltmeters, crackmeters, and load cells. These sensors are used in a wide variety of structural, hydrological, and geotechnical applications because of their stability, accuracy, and durability. Up to two vibrating wire or vibrating strip transducers can be connected to the AVW200. More sensors can be measured by using multiplexers (see Section
1.4, Use with Multiplexers).

1.1 Design Features

Historically, vibrating-wire sensors suffered from one major problem— external noise. The AVW200 significantly reduces and, in most cases, eliminates the problem of incorrect readings due to noise sources. The noise problems were overcome by advancement in technology and mathematical processing from previous time-domain based measurements.
Read more! Vibrating Wire Measurements are discussed in detail in Section 2.
i
, resulting in frequency based measurements—a complete departure
To simplify programming of the AVW200, engineers implemented firmware to eliminate several parameters that were necessary in programming the older Campbell Scientific interfaces (i.e., AVW1, AVW4, AVW100).
i
U.S. Patent No. 7,779,690
1
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
The eliminated parameters are:
Number of steps
Number of cycles
Time of Swept Frequency
These parameters are now part of the AVW200 internal operating system and require no user input. The user only needs to input the lower frequency range, upper frequency range, and excitation voltage of the sensor.
Read more! Detailed programming information is provided in Section 6
.
The AVW200 returns five or six values per measurement. The first value is the vibrating wire frequency in Hz. The sixth value is the optional thermistor measurement on Ohms. Values two through five are diagnostic information giving an indication or validation of the measurement.
(1) = Frequency in Hertz (2) = SignalStrength in mV_RMS (3) = Signal/Noise Ratio (unitless) (4) = Noise Frequency Hz (5) = DecayRatio (unitless) (6) = Thermistor output in Ohms of resistance (see Section 2.2, Temperature)
The on-board diagnostics should be monitored to determine faulty wiring, incorrect frequency range, sensor degradation over time, or to determine faulty sensors.
Read more! Information on how to use the on-board diagnostics is provided in Appendix F.
2
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules

1.2 Specifications

1.3 Communication

1.3.1 Datalogger

The AVW200 module is designed to work with and complement Campbell Scientific dataloggers, as well as data acquisition products from other manufacturers.
3
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
1.3.1.1 PakBus Protocol/Direct RS-232 Connection
When using the PakBus protocol, the AVW200() instruction in CRBasic supports an AVW200 connected to a datalogger via a cable attached to the AVW200’s RS-232 port. You can run the datalogger AVW200() instruction in either the pipeline or sequential mode. This instruction is contained in the following datalogger operating systems:
CR800std.06 or greater
CR1000std.15 or greater
CR3000std.08 or greater
Multiplexers can be controlled by either the AVW200 or the datalogger.
Read more! You can find Quick Start Guides in Section 3 instruction description in Section 6.1, and programming examples in Section 7.
1.3.1.2 PakBus Protocol/Wireless Connection
An AVW206, AVW211, or AVW216 interface can transmit data to a corresponding spread spectrum radio that is attached to the datalogger. The AVW200() instruction in CRBasic supports these wireless interfaces. You can run the AVW200() instruction in either the pipeline or sequential mode. This instruction is contained in the following datalogger operating systems:
CR800std.06 or greater
CR1000std.15 or greater
CR3000std.08 or greater
Multiplexers can be controlled by the wireless AVW200.
Read more! You can find Quick Start Guides in Section 3 instruction description in Section 6.1, and a programming example in Section 7.1.2.
1.3.1.3 PakBus Protocol/MD485 Communication
For situations where wireless communication is impractical, MD485 multidrop modems may extend the distance between the AVW200 interfaces. Multiplexers must be controlled by the AVW200.
, AVW200()
, AVW200()
4
Read more! Appendix I
1.3.1.4 SDI-12 Communication Mode
SDI-12 is the only option available for our CR5000, CR10X, and CR23X dataloggers and non-Campbell Scientific dataloggers. Our CRBasic dataloggers use the SDI12Recorder instruction and Edlog dataloggers (i.e., CR10X, CR23X) use Instruction 105. The SDI12Recorder instruction should only be run in the sequential mode. When using SDI-12, multiplexers must be controlled by the datalogger.
Read more! You can find a Quick Start Guide in Section 3.3.1, SDI12Recorder instruction description in Section 6.2, and a programming example in Section 7.5.
describes using MD485 modems with AVW200s.
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules

1.3.2 Computer

1.3.2.1 Device Configuration Utility
The Device Configuration (DevConfig) Utility supports AVW200 configuration, operating system download, and vibrating wire spectrum analysis troubleshooting. To use DevConfig, the AVW200 must be connected to a PC and a power source. DevConfig is bundled in Campbell Scientific’s datalogger support software and can also be acquired, at no cost, from Campbell Scientific’s website. DevConfig 1.10 or greater is required.
Read more! DevConfig is described in detail in Section 5
1.3.2.2 LoggerNet
LoggerNet supports datalogger programming, accesses the status and public tables, and displays network routing. Please ensure that the AVW200 CRBasic instruction is included. If using LoggerNet 3.4.1 or lower, the user needs to download the most recent OS for the datalogger. This installation installs the required CRBasic files on the user’s computer so that the AVW200 instruction shows up in the editor.
Read more! Programming information is provided in Section 6 refer to the LoggerNet manual and help.
1.3.2.3 Terminal Commands
You can monitor the AVW200 by entering terminal commands in the terminal emulator of DevConfig or LoggerNet. HyperTerminal or ProComm can also be used. The AVW200 needs to be connected to a computer with a standard RS-232 cable (CSI model 10873).
Read more! Terminal Mode Commands are discussed in Section 5.6

1.4 Use with Multiplexers

For the AVW200 interfaces, the AM16/32B multiplexer is recommended over its predecessors (i.e., AM16/32, AM16/32A). The AM16/32B has a clocking mode that can use relay addressing to go directly to a specific channel— reducing power consumption and wear on the relay switches.
.
. Also,
.
Up to 32 vibrating wire sensors without thermistors or 16 vibrating wire sensors with thermistors can be connected to one multiplexer. Two multiplexers can be connected to one AVW200. Using a direct RS-232 connection, up to four AVW200 interfaces can be connected to one CR1000 or CR3000; up to two AVW200 interfaces can be connected to one CR800 or CR850. This allows up to 256 vibrating wire sensors (128 with temperature) to be measured by one CR1000 or CR3000 (see FIGURE 1-1) or 128 vibrating wire sensors (64 with temperature) to be measured with one CR800 or CR850. FIGURE 1-2 and FIGURE 1-3 indicate the number of interfaces that can be connected to one datalogger when using wireless interfaces or the SDI-12 protocol.
Read more! You can find Quick Start Guides in Sections 3.2 and 3.3 wiring information in 4.5, programming information in Section 6, and programming examples in Section 7.
,
5
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
CR3000
CR1000
AVW200
COM3 (C5/C6)
COM4 (C7/C8)
AM16/32B
128 – Vibrating Wire Sensors in 4x16 configuration 256
COM1 (C1/C2)
COM2 (C3/C4)
CR3000 CR800, CR850 CR1000
128 – Vibrating Wire Sensors in 4x16 configuration 256
PakAddr = 200
PakAddr = 201
PakAddr = 202
PakAddr = 203
AVW206
AM16/32B
RF401
– Vibrating Wire Sensors in 2x32 configuration
FIGURE 1-1. Network of AVW200s and AM16/32Bs (using a direct RS-
232 connection)
6
– Vibrating Wire Sensors in 2x32 configuration
FIGURE 1-2. Network of AVW206s and AM16/32Bs (wireless)
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
CR10X CR5000 CR23X CR800, CR850 CR1000 CR3000
AVW200
SDI-12
AM16/32B
Datalogger MUST Control Multiplexers in SDI
0
1
2
3
-12 Mode
FIGURE 1-3. Network of AVW200 Interfaces (SDI-12)

2. Measurements

2.1 Vibrating Wire

The spectral approach implemented by the AVW200 offers significantly improved noise immunity when compared to older period-averaging techniques implemented by other vibrating-wire interfaces (AVW1, AVW4, and AVW100). Testing revealed more than two to three orders of magnitude better noise immunity with the AVW200. In addition, the spectral analysis gives improved frequency resolution (0.001 Hz rms) during quiet conditions.
The AVW200 measures the resonant frequency of the taut wire in a vibrating wire sensor (see FIGURE 2-1) with the following procedure. First, the AVW200 excites the wire with a swept-frequency excitation signal. Next, the AVW200 records the response from the vibrating wire. Finally, the AVW200 Fourier transforms the recorded response and analyzes the resulting spectrum to determine the wire’s resonant frequency. This analysis also provides diagnostic information indicating the quality of the resonant-frequency measurement.
7
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
TABLE 2-1. AVW200 Measurement Inputs
Plucking/ Pickup Coil
Vibrating Wire
Diaphragm
FIGURE 2-1. Cutaway of Vibrating Wire Sensor
There are three user-determined inputs to the AVW200 measurement process and five outputs from the measurement process. The input parameters control the excitation frequency range (BeginFreq and EndFreq) and the excitation amplitude (ExVolt); see TABLE 2-1. The frequency range supported spans from 100 Hz to 6500 Hz.
Input Units Description
BeginFreq Hz Minimum excitation and analysis frequency
EndFreq Hz Maximum excitation and analysis frequency
ExVolt Unitless Excitation voltage
1: 5 Volts peak to peak 2: 12 Volts peak to peak
Therm50_60Hz (see Section 2.2)
Unitless Thermistor measurement parameter
0: No thermistor measurement _60Hz Use 60 Hz noise rejection _50Hz: Use 50 Hz noise rejection
The measurement outputs are resonant frequency, response amplitude, signal­to-noise ratio, noise frequency, and decay ratio; see TABLE 2-2. The raw frequency measurement output of the AVW200 is in Hertz unlike our previous
2
interfaces, which output kHz
or 1/T2 (where T is the period in milliseconds). The Hertz output is converted to the appropriate units of measurement (e.g., pressure, strain, displacement) by using information provided on the sensor’s calibration report.
Read more! You can find an example of converting Hertz to displacement in Appendix A and an example program that converts Hertz to displacement in Section 7.1.2.
8
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
TABLE 2-2. AVW200 Measurement Outputs
Output Units Description
Resonant Frequency Hz Frequency of the peak response
Response Amplitude mV rms Amplitude of the peak response1
Signal-to-Noise Ratio unitless Response amplitude divided by
1
amplitude of largest noise candidate
Noise Frequency Hz Frequency of largest noise candidate1
Decay Ratio unitless Ending time-series amplitude divided
Thermistor Resistance
by the beginning time-series amplitude
Ohms On-gage thermistor resistance2
1
(see Section 2.2)
Notes:
1. Use for measurement diagnostics.
2. Optional output, not measured if Therm50_60Hz is set to 0.
When using firmware version Std.04 (or higher) and the Response Amplitude diagnostic is measured as less than 0.01 mV rms (10 microvolts), the Resonant Frequency reading will be modified to warn the user about the occurrence of low signal strength amplitudes. If SDI-12 is used to communicate with the AVW200-series device, the frequency will be given as -9,999,999 under those conditions. For all other communication methods, the frequency will be given as NAN (not-a-number) when experiencing this low signal strength condition.
If the user desires the frequency to be returned as NAN for a higher (i.e., more pessimistic) threshold than 0.01 mV, this can be done by using an optional parameter in the AVW200 CRBasic Instruction. See Section 6.1, AVW200 Instruction, for details about how this can be done.
The Resonant Frequency reading is also used to warn the user when there is an invalid voltage supply in the hardware of the device (firmware Std.04 and higher). If an internal calibration factor is outside of the expected range, then the value of -555,555 is returned for the frequency measurement. This indicates to the user that there is a hardware issue on the device which requires a factory examination and/or repair. Contact Campbell Scientific for instructions when this value is given as the Resonant Frequency reading.
FIGURE 2-2 (a representative output from the Device Configurator’s Troubleshooter) illustrates the AVW200 measurements. The bottom graph shows the raw time series data recorded from a vibrating sensor after the sensor has been excited with the frequency-swept voltage signal, and the top graph shows the spectrum after the AVW200 has applied the fast Fourier transform (FFT). In addition to the wire’s resonant frequency, the spectrum shows the response amplitude, noise amplitude, and noise frequency. The AVW200 computes the signal-to-noise ratio diagnostic by dividing the response amplitude by the noise amplitude. The AVW200 computes the decay ratio diagnostic from the time series ending amplitude divided by the beginning amplitude shown on the bottom graph in FIGURE 2-2.
9
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
Response
Amplitude
Noise Amplitude
Resonant Frequency
Noise Frequency
Beginning Amplitude
Ending Amplitude
FIGURE 2-2. DevConfig plots showing the AVW200 measurement
Please note that the use of the special FFT algorithm to achieve better noise immunity does require time for computation, which limits the maximum vibrating wire measurement rate to 2 seconds per sensor. Running a program at rates faster than this will result in compile/download errors.
Read more! You can find Troubleshoot tool information in Section 5.5 Appendix F; and detailed programming information in Section 6.

2.2 Temperature

The AVW200 contains a completion resistor for measuring the internal thermistor contained in many vibrating wire sensors. The thermistor’s resistance changes with the internal temperature of the sensor. This temperature can be used to correct errors in the vibrating wire measurement due to thermal expansion/contraction of the sensor body. The temperature correction is often used when the temperature of the medium that the sensor is measuring is changing (e.g. water temperature in a river or shallow lake). Temperature is calculated by applying the resistance to a known equation such as the Steinhart-Hart equation. The Steinhart-Hart coefficients for your sensor are found in the sensor’s user manual.
approach.
and
10
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
Cable that Comes with Sensor
17855 Pigtailed Cable or 18663 Null Modem Cable
19246 Power Cable
Sensors
AVW200
Sensors
Datalogger
Power
Supply
Read more! You can find an example program that converts resistance to temperature in Section 7.1.2, and detailed information about the thermistors in Appendix B.

3. Quick Start Guides

The AVW200 can be used in many types of systems—from simple to complex. The following quick start guides provide steps used to set up a system for some example configurations.

3.1 One or Two Sensors (no multiplexers)

3.1.1 Direct RS-232 Connection

For this simple configuration, the AVW200 can be used straight from the box (i.e., settings do not need to be changed in DevConfig). The sensor(s) are attached directly to the AVW200. The AVW200 is connected directly with the datalogger via the 17855 cable or 18663 cable. The 17855 cable terminates in pigtails for connection to datalogger control port pairs (C1/C2…C7/C8). The 18663 Null Modem Cable has a DB9 connector for attachment to the datalogger’s RS-232 port.
The following steps are used to measure the sensor(s):
1. Attach the vibrating wire sensor(s) to the AVW200 as shown in
FIGURE 4-1.
2. Use the 17855 cable to attach the AVW200 to a control port pair on the
datalogger (i.e., C1/C2, C3/C4, C5/C6, C7/C8), or use the 18663 Null Modem cable to attach the AVW200 to the RS-232 port on the datalogger.
3. Connect one end of the 19246 power cable to the 12V and G terminals on
the AVW200 and the other end to the 12V and G terminals on the datalogger or external power supply.
Read more! You can find power connection information in Section 4.2
,
and datalogger connection information in Section 4.3.
4. Create a CRBasic program that includes an AVW200() instruction for each
of the sensors.
11
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
Cable that Comes with Sensor
19246 Power Cable
SC12 CS I/O Cable
Wireless Connection
Sensors
AVW206
Sensors
Power Supply
Datalogger
RF401
Onsite
Office
NOTE
Check the manufacturer’s specification for the sensors frequency and excitation range before picking the begin/end frequencies and excitation voltage.
For example, the following AVW200() instructions can be used to measure two sensors:
AVW200(Result,Com1,200,200,Dst(1,1),1,1,1,1000,3500,2,_60HZ,1,0) AVW200(Result,Com1,200,200,Dst(2,1),2,1,1,1000,3500,2,_60HZ,1,0)
Where, AVW200 connects to datalogger control ports 1 & 2 via 17855 cable
(option Com1) Begin Frequency = 1000 End frequency = 3500 Excitation voltage = 12 V peak to peak (option 2)
Read more! A thorough description of the AVW200() instruction and its parameters is provided in Section 6.1. A complete example program that measures two vibrating wire sensors (no multiplexer) is included in Section 7.1.1.

3.1.2 Wireless Connection

12
For this example configuration, the sensor(s) are attached directly to the AVW206. The AVW206 interface transmits the data to an RF401 spread spectrum radio that is connected to the datalogger.
At the AVW206 site, do the following steps:
1. Configure the AVW206 for RF communications as described in Section
5.1, Connecting to DevConfig, and Section 5.2.1, Communications.
2. Attach the vibrating wire sensor(s) to the AVW206 as shown in FIGURE
4-1.
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
NOTE
NOTE
3. Connect an antenna (or antenna cable with Yagi or omnidirectional
antenna attached) to the Antenna Connector on the side of the AVW206.
Read more! Description of our antenna options is provided in Appendix C
4. Use the power cable to connect the 12V and G terminals on the AVW206
to the 12V and G terminals on the PS100 or another power supply.
At the datalogger/RF401 site, do the following steps:
1. Configure the RF401 radio so that its parameters match the AVW206.
The protocol setting for the RF401 must be PB Aware or PB Node.
2. Attach an antenna (or antenna cable with Yagi or omnidirectional antenna
attached) to the Antenna Connector on the RF401.
3. Use the SC12 serial cable to attach the datalogger’s CS I/O port to the
RF401’s CS I/O port. The datalogger’s CS I/O port applies power to the RF401.
Read more! More information about using spread spectrum radios is provided in the RF401/RF411/RF416 Spread Spectrum Data Radio/Modem manual.
4. Create a CRBasic program that includes an AVW200() instruction for each
of the sensors.
.
Check the manufacturer’s specification for the sensors frequency and excitation range before picking the begin/end frequencies and excitation voltage.
For example, the following AVW200() instructions can be used to measure two sensors:
AVW200(Result,ComSDC7,200,200,Dst(1,1),1,1,1,1000,3500,2,_60HZ,1,0) AVW200(Result,ComSDC7,200,200,Dst(2,1),2,1,1,1000,3500,2,_60HZ,1,0)
Where, RF401 = configured for SDC7 Begin Frequency = 1000 End frequency = 3500 Excitation voltage = 12 V peak to peak (option 2)
Read more! A thorough description of the AVW200() instruction and its parameters is provided in Section 6.1. A complete example program that measures one vibrating wire sensor (no multiplexer) is included in Section
7.1.2.
13
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
Cable that Comes with Sensor
CABLE4CBL-L Four Conductor Cable
17855 Pigtailed Cable or 18663 Null Modem Cable
19246 Power Cable
Multiplexer
AVW200
Multiplexer
Datalogger
Power
Supply
Sensors
Sensors

3.2 Multiplexers Controlled by AVW200

3.2.1 Direct RS-232 Connection

For this example configuration, vibrating wire sensors are attached to multiplexers, which are controlled by the AVW200. The AVW200 is connected directly with the datalogger.
14
The following steps are used:
1. If you are not using the default multiplexer, go to the
Deployment/Measurement tab in DevConfig and select the multiplexer you are using (Section 5.1, Connecting to DevConfig, and Section 5.2.1, Communications). The default multiplexer for the AVW200 is the AM16/32A.
2. Attach the multiplexers to the AVW200 as shown in FIGURE 4-4.
3. Connect the sensors to the multiplexers.
Read more! Information about connecting the vibrating wire sensors to the multiplexer is provided in the AM16/32B manual.
4. Use the 17855 cable to attach the AVW200 to control port pairs on the
datalogger, or use the 18663 Null Modem cable to attach the AVW200 to the RS-232 port on the datalogger.
5. Connect one end of the 19246 power cable to the 12V and G terminals on
the AVW200 and the other end to the 12V and G terminals on the datalogger or external power supply.
Read more! You can find power and ground connection information in Section 4.2 and datalogger connection information in Section 4.3.
6. Create a CRBasic program that includes the AVW200() instruction for
each of the multiplexers that the AVW200 will control.
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
Cable that Comes with Sensor
CABLE4CBL-L Four Conductor Cable
19246 Power Cable
SC12 CS I/O Cable
Wireless Connection
Multiplexer
Multiplexer
Sensors
Sensors
AVW206
Power
Supply
Datalogger
RF401
Onsite
Office
NOTE
Check the manufacturer’s specification for the sensors frequency and excitation range before picking the begin/end frequencies and excitation voltage.
For example, the following AVW200() instructions can be used to control two multiplexers:
AVW200(Data1(),Com1,200,200,mux1(1,1),1,1,16,450,3000,2,_60HZ,1,0) AVW200(Data2(),Com1,200,200,mux2(1,1),2,1,16,450,3000,2,_60HZ,1,0)
Where, AVW200 connects to datalogger control ports 1 & 2 via 17855 cable
(option Com1) Each multiplexer has 16 sensors connected to it. Begin Frequency = 450 End frequency = 3000 Excitation voltage = 12 V peak to peak (option 2)
Read more! A thorough description of the AVW200() instruction and its parameters is provided in Section 6.1. Complete example programs that control two multiplexers are included in Sections 7.2, 7.3, and 7.4.1.

3.2.2 Wireless Connection

For this example configuration, vibrating wire sensors are attached to multiplexers, which are controlled by an AVW206. The AVW206 interface transmits the data to an RF401 spread spectrum radio that is connected to the datalogger.
15
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
NOTE
NOTE
At the AVW206 site, do the following steps:
1. Use DevConfig to configure the AVW206 for RF communications
(Section 5.1, Connecting to DevConfig, and Section 5.2.1, Communications).
2. If you are not using the default multiplexer, go to the
Deployment/Measurement tab in DevConfig and select the multiplexer you are using (Section 5.1, Connecting to DevConfig, and Section 5.2.1, Communications). The default multiplexer for the AVW206 is the AM16/32A.
3. Attach the multiplexers to the AVW206 as shown in FIGURE 4-4.
4. Connect the sensors to the multiplexers.
Read more! Information about connecting the vibrating wire sensors to the multiplexer is provided in the AM16/32B manual.
5. Connect an antenna (or antenna cable with Yagi or omnidirectional
antenna attached) to the Antenna Connector on the side of the AVW206.
Read more! Description of our antenna options is provided in Appendix C
6. Use the power cable to connect the 12V and G terminals on the AVW206
to the 12V and G terminals on the PS100 or another power supply.
At the datalogger/RF401 site, do the following steps:
1. Configure the RF401 radio so that its parameters match the AVW206.
The protocol setting for the RF401 must be PB Aware or PB Node.
2. Attach an antenna (or antenna cable with Yagi or omnidirectional antenna
attached) to the Antenna Connector on the RF401.
3. Use the SC12 serial cable to attach the datalogger’s CS I/O port to the
RF401’s CS I/O port. The datalogger’s CS I/O port applies power to the RF401.
Read more! More information about using spread spectrum radios is provided in the RF401/RF411/RF416 Spread Spectrum Data Radio/Modem manual.
4. Create a CRBasic program that includes the AVW200() instruction for
each of the multiplexers that the AVW200 will control.
.
16
Check the manufacturer’s specification for the sensors frequency and excitation range before picking the begin/end frequencies and excitation voltage.
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
Cable that Comes with Sensor
CABLE4CBL Cable
CABLE3CBL Cable
19246 Power Cable
AVW200
Multiplexer
Power
Supply
Datalogger
Sensors
NOTES
For example, the following AVW200() instructions can be used to control the multiplexers:
AVW200(Data1(),ComSDC7,200,200,mux1(1,1),1,1,16,450,3000,2,_60HZ,1,0) AVW200(Data2(),ComSDC7,200,200,mux2(1,1),2,1,16,450,3000,2,_60HZ,1,0)
Where, RF401 = configured for SDC7 Each multiplexer has 16 sensors connected to it. Begin Frequency = 450 End frequency = 3000 Excitation voltage = 12 V peak to peak (option 2)
Read more! A thorough description of the AVW200() instruction and its parameters is provided in Section 6.1. A complete example program that controls two multiplexers is included in Appendix H.1.2.

3.3 Multiplexers Controlled by Datalogger

3.3.1 SDI-12 Communication

For this example configuration, SDI-12 is used to measure the vibrating wire sensors. The vibrating wire sensors are attached to multiplexers, which are controlled by the datalogger.
(1) When using SDI-12, multiplexers have to be controlled by the datalogger.
(2) SDI-12 is the only option available for our CR10X, CR23X, and CR5000 dataloggers.
(3) SDI-12 is the only option available for non-Campbell Scientific dataloggers.
17
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
NOTE
The following steps are used:
1. Access DevConfig to configure the AVW200 for SDI-12 communications.
Go to the Deployment/Measurement tab in DevConfig and enter the SDI-12 Address, multiplexer type, begin frequency, end frequency, and excitation (see Section 5.1, Connecting to DevConfig, and Section 5.2.2, Measurement).
2. Use a CABLE4CBL-L cable to connect the AVW200 to the multiplexers
(see FIGURE 4-5).
SDI-12 uses the CLK and RESET on the multiplexer instead of the CLK and RESET address on the AVW200.
3. Connect the sensors to the multiplexers.
Read more! Information about connecting the vibrating wire sensors to the multiplexer is provided in the AM16/32B manual.
4. Use a CABLE4CBL-L cable to connect the multiplexers to the datalogger
(see FIGURE 4-6).
5. Use a CABLE3CBL-L cable to connect the C1 terminal on the AVW200
6. Connect one end of the 19246 power cable to the 12V and G terminals on
Read more! You can find power and ground connection information in Section 4.2, and datalogger connection information in Section 4.3.
7. Program the datalogger. Our CRBasic dataloggers use the SDI12Recorder
Read more! The SDI12Recorder instruction is discussed in Section 6.2 An example CRBasic program is provided in Section 7.5, and an example Edlog program is provided in Appendix G.

4. Connections

4.1 Sensor Wiring (no multiplexers)

Up to two vibrating wire sensors can be directly connected to the AVW200 (see FIGURE 4-1). Sensor cabling is sold as a part of the sensor (refer to the sensor manual for wire colors). Cable options for connecting the AVW200 to the datalogger are listed in TABLE 4-1.
to a control port and ground on the datalogger.
the AVW200 and the other end to the 12V and G terminals on the datalogger or external power supply.
instruction and Edlog dataloggers (i.e., CR10X, CR23X) use Instruction
105. The SDI12Recorder instruction should only be run in the sequential mode.
.
18
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
NOTE
AVW200
FIGURE 4-1. Wiring for Sensor Connections

4.2 Power and Ground

Each AVW200 has a ground lug for connection to earth ground and a green connector for attachment to a power source (see FIGURE 4-2).
Only connect the AVW200 ground lug to earth ground when the AVW200 is not directly connected to the datalogger. When a datalogger is in the same enclosure, only connect the datalogger’s ground lug to earth ground.
The AVW200’s ground lug is connected to earth ground via an 8 AWG wire. This connection should be as short as possible.
The 19246 power cable is shipped with each AVW200 for connection to a power source. The cable terminates in pigtails that attach to the 12V and G terminals on the AVW200 and the power source. Often the AVW200 is powered by the datalogger, but another 12 Vdc power source may be used.
19
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
TABLE 4-1. Datalogger to AVW200 Cable Options
Connects to a power source via 19246 power cable
Connects to earth ground via an 8 AWG wire when
Indicates AVW200 is connected to a power source
not directly connected to a datalogger
FIGURE 4-2. Ground Lug and Power Connectors on the AVW200

4.3 Datalogger Wiring (Direct Connection)

There are three options for connecting the AVW200 directly to the datalogger (see TABLE 4-1); the cable is ordered as a common accessory.
Communication Protocol Cable Used
PakBus (direct RS-232 connection)
PakBus (direct RS-232 connection)
SDI-12 CABLE3CBL-L
18663 Null Modem Cable
17855 or SC110’s DTE Cable (pigtail to DB9; see TABLE 4-2)
Cable
Datalogger Port for Cable Attachment AVW200
Port for CR800, CR850
RS-232 RS-232 N/A N/A N/A RS-232
Control port pair (C1/C2 or C3/C4)
C1 or C3 and G
CR1000, CR3000 CR5000 CR10X CR23X
Control port pair (C1/C2, C3/C4, C5/C6, or C7/C8)
C1, C3, C5, or C7 and G
N/A N/A N/A RS-232
SDI12 and G
C1 through C8 and G
C5 through C8 and G
Cable
Attachment
C1 and G
20
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
TABLE 4-2. 17855 or SC110’s DTE Cable Wiring
Ground Lug (connect to earth ground via 8 AWG wire)
Lights up when radio is powered; blinks when radio is transmitting
Whip Antenna
Wire Color of 17855 or SC110’s DTE Cable CR800, CR850 CR1000, CR3000
Brown C1 or C3 C1, C3, C5, or C7
White C2 or C4 C2, C4, C6, or C8
Yellow G G

4.4 Wireless Connections (AVW206, AVW211, AVW216)

The connector on the side of the AVW206, AVW211, and AVW216 is for attaching a whip antenna or an antenna cable (see FIGURE 4-3).
FIGURE 4-3. AVW206 with Whip Antenna
21
AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
TABLE 4-3. Datalogger to Spread Spectrum Radio Connections
NOTES
NOTE
Wireless communication requires the appropriate spread spectrum radio to be connected to the datalogger (see TABLE 4-3). DevConfig is used to configure the AVW206, AVW211, or AVW216 for RF communications (Section 5.1,
Connecting to DevConfig, and Section 5.2.1, Communications).
Vibrating Wire Interface Model
AVW206 RF401 SC12 CS I/O CS I/O
AVW211 RF411 SC12 CS I/O CS I/O
AVW216 RF416 SC12 CS I/O CS I/O
Required Spread Spectrum Radio Model
(1) The AVW206, AVW211, and AVW216 are not compatible with the RF450, RF400, RF410, and RF415 spread spectrum radios.
(2) For communication between the vibrating wire interface and spread spectrum radio to be successful, the radio’s Protocol must be set to PakBus Aware or PakBus Node.
Read more! Description of our antenna options is provided in Appendix C. More information about using spread spectrum radios is provided in the RF401/RF411/RF416 Spread Spectrum Data Radio/Modem manual.
Cable used for datalogger to radio connection
Radio Port for Cable Attachment
Datalogger Port for Cable Attachment

4.5 Multiplexer Wiring

22
Wire the sensors to the multiplexer according to the multiplexer manual. Other multiplexer wiring depends on whether the AVW200 will control the multiplexer or the datalogger will control the multiplexer.

4.5.1 AVW200 Controlling the Multiplexer

For most applications, it is desirable to have the AVW200 control the multiplexer. A CABLE4CBL or MUXSIGNAL cable connects to the multiplexer’s COM terminals, and another CABLE4CBL cable or the MUXPOWER cable connects to the multiplexer’s CLK and RESET terminals (see FIGURE 4-4).
When two multiplexers are connected to the AVW200, they share the CLK and RESET lines. Because of this, while one multiplexer is making measurements, the other multiplexer will advance its channels but without making any measurements.
Loading...
+ 82 hidden pages