Campbell Scientific AM16-32A User Manual

AM16/32A Relay Multiplexer
Revision: 11/07
Copyright © 1987-2007
Campbell Scientific, Inc.

Warranty and Assistance

The AM16/32A RELAY MULTIPLEXER is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless specified otherwise. Batteries have no warranty. CAMPBELL SCIENTIFIC, INC.'s obligation under this warranty is limited to repairing or replacing (at CAMPBELL SCIENTIFIC, INC.'s option) defective products. The customer shall assume all costs of removing, reinstalling, and shipping defective products to CAMPBELL SCIENTIFIC, INC. CAMPBELL SCIENTIFIC, INC. will return such products by surface carrier prepaid. This warranty shall not apply to any CAMPBELL SCIENTIFIC, INC. products which have been subjected to modification, misuse, neglect, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied, including warranties of merchantability or fitness for a particular purpose. CAMPBELL SCIENTIFIC, INC. is not liable for special, indirect, incidental, or consequential damages.
Products may not be returned without prior authorization. The following
act
information is for US and International customers residing in countries
cont served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs for customers within their territories. Please visit www.campbellsci.com to determine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization (RMA), contact CAMPBELL SCIENTIFIC, INC., phone (435) 753-2342. After an applications engineer determines the nature of the problem, an RMA number will be issued. Please write this number clearly on the outside of the shipping container. CAMPBELL SCIENTIFIC's shipping address is:
CAMPBELL SCIENTIFIC, INC.
RMA#_____
e
815 W Logan, Utah 84321-1784
CAMPBELL SCIENTIFIC, INC. does not accept collect calls.
st 1800 North
AM16/32A Relay Multiplexer Table of Contents
PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks tab for links to specific sections.
1. Function........................................................................1
1.1 Typical Applications.................................................................................1
1.2 Compatibility............................................................................................2
2. Physical Description ...................................................2
3. AM16/32A Specifications ............................................ 3
4. Operation......................................................................5
4.1 The Control Terminals..............................................................................5
4.1.1 Reset................................................................................................6
4.1.2 Clock...............................................................................................6
4.1.3 Ground............................................................................................7
4.1.4 Power Supply..................................................................................7
4.2 Measurement Terminals ...........................................................................8
4.2.1 COM Terminals..............................................................................9
4.2.2 Sensor Input Terminals...................................................................9
5. Datalogger Programming..........................................10
5.1 Single Loop Instruction Sequence..........................................................11
5.2 Multiple Loop Instruction Sequence.......................................................16
5.3 CRBasic Programming...........................................................................18
5.3.1 CR5000 and CR3000 Programming .............................................19
5.3.2 CR1000, CR800, and CR850 Programming.................................20
5.4 General Programming Considerations....................................................22
6. Sensor Hook-up and Measurement Examples........22
6.1 Single-Ended Analog Measurement without Sensor Excitation.............22
6.2 Differential Analog Measurement without Sensor Excitation................23
6.3 Half Bridge Measurements.....................................................................24
6.3.1 Half Bridge Measurement with Completion Resistor at
Datalogger ......................................................................................24
6.3.2 Potentiometer Measurement..........................................................25
6.3.3 Four Wire Half Bridge (Measured Excitation Current).................25
6.4 Full Bridge Measurements......................................................................26
6.5 Full Bridges with Excitation Compensation...........................................27
i
AM16/32A Relay Multiplexer Table of Contents
6.6 Thermocouple Measurement.................................................................. 28
6.6.1 Measurement Considerations....................................................... 28
6.6.2 Single-ended Thermocouple Measurement.................................. 30
6.6.3 Differential Thermocouple Measurement .................................... 31
6.7 Mixed Sensor Types............................................................................... 31
6.7.1 Mixed Sensor Example: Soil Moisture Blocks and
Thermocouples............................................................................... 31
7. General Measurement Considerations ....................35
8. Installation..................................................................35
8.1 Mounting Tabs....................................................................................... 36
8.2 Controlling Humidity............................................................................. 36
Appendices
A. AM16/32A Improvements........................................A-1
Figures
1. AM16/32A Relay Multiplexer................................................................... 3
2. AM16/32A Relay Actuation Time vs. Temperature and Battery Voltage.. 5
3. AM16/32A to Datalogger Power/Control Hookup.................................... 6
4. Power and Ground Connections for External Power Supply..................... 8
5. Typical AM16/32A to Datalogger Signal Hookup (4x16 Mode)...............9
6. SCWIN (Short Cut for Windows Program Builder) ................................ 10
7. Example “4x16” Mode Program Loops for CR23X, CR10(X),
21X, and CR7 Dataloggers............................................................ 14
8. Example “2x32” Mode Program Loops for CR23X, CR10(X),
21X, and CR7 Dataloggers............................................................ 15
9. Wiring Diagram for Strain Gages and Potentiometers............................. 16
10. Single-ended Measurement without Excitation......................................23
11. Differential Measurement without Excitation........................................ 23
12. Half Bridge (Modified 107 Temperature Probe) Hook-up and
Measurement.................................................................................. 24
13. Potentiometer Hook-up and Measurement............................................. 25
14. Four Wire Half Bridge Hook-up and Measurement............................... 26
15. Full Bridge Measurement....................................................................... 26
16. Full Bridge Measurement with Excitation Compensation..................... 27
17. Differential Thermocouple Measurement with Reference
Junction at the Datalogger ............................................................. 29
18. Differential Thermocouple Measurement with Reference
Junction at the AM16/32A............................................................. 29
19. AM16/32A Aluminum Cover Plate....................................................... 30
20. Thermocouple and Soil Block Measurement......................................... 32
21. Mounting Tab Hole Pattern.................................................................... 36
ii

Cautionary Notes

The AM16/32A is not designed to multiplex power. Its intended function is to switch low level analog signals. Switched currents in excess of 30 mA will degrade the relay contacts involved, rendering that channel unsuitable for further low level analog measurement. Customers who need to switch power are directed to Campbell Scientific’s SDM-CD16AC, A6REL-12, or A21REL-12 relays.
Changing the setting of the mode switch from “4x16” to “2x32” connects COM ODD H to COM EVEN H and also COM ODD L to COM EVEN L. After wiring AM16/32A, exercise due care to avoid inadvertently putting excess voltage on a line or short circuiting a power supply which might damage datalogger, wiring panel, sensor or multiplexer (not covered under warranty).
iii
This is a blank page.

AM16/32A Relay Analog Multiplexer

1. Function

The primary function of the AM16/32A Multiplexer is to increase th e number of sensors that can be measured by a CR1000, CR3000, CR800, CR850, CR23X, CR10(X), 21X, or CR7 datalogger. The AM16/32A is positioned between the sensors and the datalogger. The AM16/32A is a replacement for CSI’s AM16/32 model. Mechanical relays in the AM16/32A connect each of the sensor channels in turn to a common output destined for the datalogger. The user program advances the multiplexer through the sensor channels making measurements and storing data.
A slide switch located on the AM16/32A’s top panel selects one of two modes of operation. In “2x32” mode the multiplexer can scan 32 sensor input channels, each with two lines. In “4x16” mode it can scan 16 input channels with four lines a piece. The datalogger program is written according to the selected mode and the sensors to be measured.
The maximum number of sensors that can be multiplexed by an AM16/32A depends primarily on the type(s) of sensors to be scanned. The following guidelines assume identical sensors:
Up to 32 single-ended or differential analog sensors that do not require excitation. For example: pyranometers and thermocouples (see Sections 6.1,
6.2, and 6.6).
Up to 32 single-ended sensors that require excitation. Example: some half bridges (see Section 6.3.1).
Up to 16 single-ended or differential sensors that require excitation. Examples: full bridges and four-wire half bridge with measured excitation (see Section 6.3.3 and 6.4).
In conjunction with a second AM16/32A, up to 16 six-wire full bridges (Section 6.5).

1.1 Typical Applications

The AM16/32A is intended for use in applications where the number of required sensors exceeds the number of datalogger input channels. Most commonly, the AM16/32A is used to multiplex analog sensor signals, although it can also be used to multiplex switched excitations, continuous analog outputs, or even certain pulse counting measurements (i.e., those that require only intermittent sampling). It is also possible to multiplex sensors of different, but compatible, types (e.g., thermocouples and soil moisture blocks, see Section 6.7.1).
1
AM16/32A Relay Analog Multiplexer
NOTE
For a discussion of single-ended versus differential analog measurements, please consult the Measurement section of your datalogger manual.
As purchased, the AM16/32A is intended for use in indoor, non-condensing environments. An enclosure is required for field or high humidity use. In applications where one or two multiplexers are deployed, the ENC 10/12 (10” x 12”) enclosure is recommended.

1.2 Compatibility

The AM16/32A is compatible with Campbell’s CR5000, CR800, CR850, CR3000, CR1000, CR23X, CR10(X), 21X, and CR7 dataloggers.
The AM16/32A is compatible with a wide variety of commercially available sensors. As long as relay contact current maximums are not exceeded (see Cautionary Notes, page iii), and no more than four lines are switched at a time, system compatibility for a specific sensor is determined by sensor-datalogger compatibility.
In CR1000, CR800, CR850, CR3000, CR23X, and CR10(X) applications the AM16/32A may be used to multiplex up to 16 Geokon vibrating wire sensors through one AVW1 vibrating wire interface.

2. Physical Description

The AM16/32A is housed in a 10.2 cm x 23.9 cm x 4.6 cm (4.0” x 9.4” x 1.8”) anodized aluminum case (Figure 1). The aluminum case is intended to reduce temperature gradients across the AM16/32A’s terminal strips. An aluminum cover plate is also included to this end. This is extremely important if thermocouples are being multiplexed (Section 6.6).
The case can be opened for inspection/cleaning by removing two phillips-head screws located on the under-side of the case. Mounting tabs are provided so the AM16/32A can be fastened to a flat surface or an enclosure plate (Section
8).
All connections to the AM16/32A are made on the top panel terminal blocks. The island of four terminals located near the mode switch are dedicated to the connecting of datalogger power and control lines (Section 4.1). The four “ODD” and “EVEN” “COM” terminals on the other side of the mode switch carry shielded multiplexed sensor signals destined for datalogger analog inputs. The remaining terminals on the AM16/32A are for sensor and sensor shield connection (Section 4.2). All of the inputs of the AM16/32A are protected with gas tubes. The terminals accept stripped and tinned lead wires up to 16 AWG or 1.6 mm in diameter. Datalogger-to-AM16/32A cabling requires a minimum of six and as many as nine individually insulated wires with shields.
2
FIGURE 1. AM16/32A Relay Multiplexer

3. AM16/32A Specifications

AM16/32A Relay Analog Multiplexer
Power*:
Current Drain:
Reset*:
Clock*:
Operational Temperature:
Operational Humidity:
Unregulated 12 VDC Minimum Operating Voltage:
from –55C to +40C = 11.3 VDC;
from +40C to +85C = 11.8 VDC (See Figure 2 for relay actuation times vs. temperature and supply voltage.)
Quiescent: < 210 uA Active: 6 mA typical in “2 x 32” mode 11 mA typical in “4 x 16” mode
A continuous signal between 3.5 VDC and 16 VDC holds the AM16/32A in an active state (where a clock pulse can trigger a channel advance). A signal voltage < 0.9VDC deactivates the AM16/32A (clock pulse will not trigger a scan advance; AM16/32A is also reset).
On the transition from <1.5 V to >3.5 V, a scan advance is actuated on the leading edge of the clock signal; clock pulse should be a minimum of 1 ms wide.
Standard: -25oC to +50oC Extended: -55oC to +85oC
0 - 95%, non-condensing
3
AM16/32A Relay Analog Multiplexer
Dimensions:
Length – 23.9 cm (9.4") Width - 10.2 cm (4.0") Depth - 4.6 cm (1.8")
Weight: Mounting Tab
Hole Spacing: Expandability
**
(nominal):
Maximum Cable Length:
Maximum Switching Current
***
:
Contact Specifications:
1.5 lbs. (approx.), 693 g. 1 inch x 3 inches x 9 inches. Up to 1/8 inch or 3
mm diameter screws (see Figure 21). 2 AM16/32As per CR800/CR850
4 AM16/32As per CR3000 4 AM16/32As per CR5000 4 AM16/32As per CR1000 4 AM16/32As per CR23X 4 AM16/32As per CR10(X) 4 AM16/32As per 21X 8 AM16/32As per CR7 725 Card
Depends on sensor and scan rate. In general, longer lead lengths necessitate longer measurement delays. Refer to datalogger manual for details.
500 mA Initial contact resistance: <0.1 ohm max.
Initial contact bounce: <1 ms Contact material: Gold clad silver alloy Wiper to N.O. contact capacitance: 0.5 pF Typical low-current (<30 mA) life: 5 x 10
7
operations
Relay Switching Characteristics
(applying 11.3 – 14 VDC):
ESD:
Thermal emf: 0.3 µV typical; 0.5 µV maximum Operate time: <10 ms over temperature and supply
ranges Break before make guaranteed by design
Air Discharge: complies with IEC61000-4-2, test level 4 (±15 kV) Contact Discharge: complies with IEC61000-4-2, test level 4 (±8 kV)
Surge:
Complies with IEC61000-4-5, test level 3 (±2 kV, 2 ohms coupling impedance)
*
Reset, Clock, and +12V inputs are protected by +16V transzorbs.
**
Assumes sequential activation of multiplexers and that each datalogger channel is uniquely dedicated. If your application requires additional multiplexing capability, please consult CSI for application assistance.
***
Switching currents greater than 30 mA (occasional 50 mA current is acceptable) will degrade the contact surfaces of the mechanical relays (increase their resistance). This will adversely affect the suitability of these relays to multiplex low voltage signals. Although a relay used in this manner no longer qualifies for low voltage measurem ent, it continues to be useful for switching currents in excess of 30 mA.
4
AM16/32A Relay Analog Multiplexer
12.0
10.0
8.0
6.0
4.0
2.0
RELAY ACTUATION TIME (ms)
0.0
10
9.6
10.4
10.8
11.2
11.61212.4
POWER SUPPLY VOLTAGE
65C
12.8
13.2
13.61414.4
50C
14.8
25C
15.2
-25C
16
15.6
FIGURE 2. AM16/32A Relay Actuation Time vs.
Temperature and Battery Voltage.

4. Operation

4.1. The Control Terminals

Subsection 4.1 discusses the terminals that control operation of the multiplexer. These terminals are located at the left-hand side of the multiplexer as shown in Figure 1. Subsection 4.2 d iscusses the use of sensor measurement terminals.
The CR5000, CR3000, CR800, CR850, CR1000, CR23X, CR10(X), 21X, and CR7 dataloggers connect to the AM16/32A as shown in Figure 3 (“4x16” mode). Figure 3 depicts control connections. Measurement connections are discussed in Section 6. The power, ground, reset, and clock connections remain essentially the same regardless of datalogger used.
With the CR5000, CR3000, CR800, CR850, CR1000, CR23X and CR10(X) the datalogger 12 VDC supply and ground terminals are connected to the AM16/32A 12V and ground terminals. One control port is required for clocking and a second control port for reset. The MUXPOWER cable (or equivalent) shield is grounded on both ends as illustrated below.
5
AM16/32A Relay Analog Multiplexer
RES
CLK
GND
12V
O
N
CR23X, CR5000
21X
MUXPOWER
SHIELD
CR10X,
CR800,
CR850
G G
12 V 12 V 12 V +12 V 12 V
G G G
C1-C4 C1-C8 C1-C8 EXCIT 1-4 EXCITATION C1-C4 C1-C8 C1-C8 C1-C8 725 Card
CR3000,
CR1000
FIGURE 3. AM16/32A to Datalogger Power/Control Hookup
With the 21X or CR7 the AM16/32A connects to the 12 VDC and “
terminals for power. One control port is used for reset, and one switched excitation channel is used for clock (on 725 card with CR7). If a switched excitation port is not available, an additional control port can be used to provide clock pulses to the multiplexer.
CR7
Control
4.1.1 Reset
4.1.2 Clock
The reset (“RES”) line is used to activate the AM16/32A. A signal in the range of +3.5 to +16 VDC applied to the reset terminal activates the multiplexer. When this line drops lower than +0.9 VDC, the multiplexer enters a quiescent, low-current-drain state. In the quiescent state the common (“COM”) terminals are electrically disconnected from all of the sensor input channels. Reset should always connect to a datalogger control port. Instruction 86 (option code 41 - 48 to activate, and 51 - 58 to deactivate) is generally used to activate/deactivate the multiplexer, however, in the case of the 21X or CR7 with older PROMS, Instruction 20 is commonly used. The CR800, CR850, CR3000, CR5000, and CR1000 uses the PortSet instruction to control the reset line.
Pulsing the AM16/32A “CLK” line high (“RES” line already high) advances the channel. When reset first goes high, the common terminals ODD H, ODD L and EVEN H, EVEN L are disconnected from all sensor input terminals. With the panel switch in “4x16” mode, when the first clock pulse arrives the “COM” terminals are switched to connect with sensor input channel 1 (blue lettering) consisting of 1H, 1L, 2H, and 2L. When a second clock pulse arrives the common lines are switched to connect to channel 2 (3H, 3L, 4H, 4L). The multiplexer advances on the leading edge of the positive going clock pulse. The voltage level must fall below 1.5 VDC and then rise above
3.5 VDC to clock the multiplexer. The CLK pulse should be at least 1 ms long. A delay (typically 10 to 20 ms) is inserted between the beginning of the CLK pulse and the measurement instruction to ensure sufficient settling time for relay contacts.
6
4.1.3 Ground
AM16/32A Relay Analog Multiplexer
With the 21X and CR7 dataloggers, switched excitation is generally used to clock the multiplexer (Instruction 22 configured for 5000 mV excitation). If no switched excitation channel is available, it is possible to clock using control ports. See Section 5.1 for details.
In the case of the CR5000, CR3000, CR800, CR850, CR1000, CR23X, and CR10(X), a control port is generally used to clock the multiplexer. Instruction 86 with the pulse port option (command code 71 through 78) generates a 10 ms pulse which works well.
The CR5000, CR3000, CR800, CR850, and CR1000 uses a control port controlled by PortSet, Delay, and SubScan/NextSubScan to create the Clock pulses (see program example in Section 5.3).
If several multiplexers are required, a CR5000, CR3000, CR800, CR850, CR1000, CR10(X) or CR23X control port can source sufficient current to drive up to six AM16/32A CLK or RES inputs wired in parallel.
The AM16/32A has a ground lug that should be connected to earth ground via an 8 AWG wire. This connection should be as short as possible.
The AM16/32A “GND” terminal is connected to datalogger power ground. The AM16/32A “GND” terminal is also connected to the MUXPOWER cable (or equivalent) SHIELD and, via that, to datalogger power ground (see Figure
3). If a separate power supply is used, the AM16/32A ground should also connect to the separate supply’s ground (Figure 4). An AM16/32A “COM
“G”) via the MUXSIGNAL cable (or equivalent) also according to Figure 5 (see 4.2.1). The datalogger itself must connect to earth ground by one of the methods described in the Installation and Maintenance Section of your datalogger operator’s manual.
4.1.4 Power Supply
The AM16/32A requires a continuous 12 VDC power supply for operation. The multiplexer's current drain is less than 210 microamps in the quiescent state and is typically 6 to 11 milliamps at 12 VDC when active (see current drain spec). The power supply is connected to the multiplexer terminals labeled “12V” (+) and “GND”. Connect the “GND” wire first for safety.
In many applications it is convenient to power the AM16/32A from a datalogger battery. For more power-intensive applications, an external, rechargeable, 12 VDC, 60 Amp Hr source may be advisable. Lead-acid supplies are recommended where solar or AC charging sources are available because they handle well being “topped off” by constant charging. The BPALK alkaline supply (12 Amp Hr) can be used to power the AM16/32A in applications where the average system current is low, or where it is convenient to frequently replace batteries. It is advisable to calculate the total power requirements of a system and the expected longevity of the power supply based on average system current drains (e.g. logger, multiplexer, other peripherals and sensors) at the expected ambient temperatures.
” terminal should connect to a datalogger ground terminal (“
” or
7
AM16/32A Relay Analog Multiplexer
The average power required to operate an AM16/32A depends on the percentage of time it is active per time period. For example, if a CR10X makes differential measurements on 32 thermocouples every minute, the average current drain due to the AM16/32A would be about ((.030 Sec/chan x 32 chan)/60 Sec) x 6 mA = 0.1 mA. Under the same conditions, a 2 second execution interval rate increases the average system current drain to about ((.030 Sec/chan x 32 chan)/2 Sec) x 6 mA = 2.9 mA. At a minimum, the power supply must be able to sustain the system between site visits anticipating the worst environmental extremes.
If a 21X power supply is used to power the AM16/32A, all low-level analog measurements (thermocouples, pyranometers, thermopiles, etc.) must be made differentially. Differential measurements are required because slight ground potentials are created along the 21X analog terminal strip when the 12V supply is used to power peripherals. This limitation reduces the number of available analog input channels and may mandate the use of an external supply for the AM16/32A (Figure 4).
FIGURE 4. Power and Ground Connections for External Power Supply.
Low supply voltage and high ambient temperatures affect the actuation time of the multiplexer relays (Figure 2). If your program does not allow the relay contacts sufficient time to close before a measurement is started, the result will be inaccurate or overranged values.

4.2 Measurement Terminals

Most of the terminals on the AM16/32A are dedicated to the connection of sensors to the multiplexer (Figure 1). Depending on the panel switch selection (“4x16” or “2x32” mode), the sensor input terminals are organized into 16 groups (blue letters) of 4 sensor inputs or 32 groups (white letters) of 2 sensor inputs. The terminals accept solid or tinned, stripped sensor leads. The four “COM” terminals marked “ODD H, L” and “EVEN H, L” located by the mode switch provide for attachment of the common signal leads that carry multiplexed sensor signals to the datalogger.
8
Loading...
+ 32 hidden pages