The MODEL HFP01SC SELF-CALIBRATING SOIL HEAT FLUX
PLATE is warranted by CAMPBELL SCIENTIFIC, INC. to be free from
defects in materials and workmanship under normal use and service for twelve
(12) months from date of shipment unless specified otherwise. Batteries have
no warranty. CAMPBELL SCIENTIFIC, INC.'s obligation under this
warranty is limited to repairing or replacing (at CAMPBELL SCIENTIFIC,
INC.'s option) defective products. The customer shall assume all costs of
removing, reinstalling, and shipping defective products to CAMPBELL
SCIENTIFIC, INC. CAMPBELL SCIENTIFIC, INC. will return such
products by surface carrier prepaid. This warranty shall not apply to any
CAMPBELL SCIENTIFIC, INC. products which have been subjected to
modification, misuse, neglect, accidents of nature, or shipping damage. This
warranty is in lieu of all other warranties, expressed or implied, including
warranties of merchantability or fitness for a particular purpose. CAMPBELL
SCIENTIFIC, INC. is not liable for special, indirect, incidental, or
consequential damages.
Products may not be returned without prior authorization. To obtain a
Returned Materials Authorization (RMA), contact CAMPBELL SCIENTIFIC,
INC., phone (435) 753-2342. After an applications engineer determines the
nature of the problem, an RMA number will be issued. Please write this
number clearly on the outside of the shipping container. CAMPBELL
SCIENTIFIC's shipping address is:
CAMPBELL SCIENTIFIC, INC.
RMA#_____
815 West 1800 North
Logan, Utah 84321-1784
CAMPBELL SCIENTIFIC, INC. does not accept collect calls.
Non-warranty products returned for repair should be accompanied by a
purchase order to cover the repair.
815 W. 1800 N.
Logan, UT 84321-1784
USA
Phone (435) 753-2342
FAX (435) 750-9540
www.campbellsci.com
Campbell Scientific Canada Corp.
11564 -149th Street
Edmonton, Alberta T5M 1W7
CANADA
Phone (780) 454-2505
FAX (780) 454-2655
1. Datalogger Connections for a Single-Ended Measurement........................ 3
2. Datalogger Connections for a Differential Measurement ...........................4
3. Wiring for Example 1.................................................................................5
4. Wiring for Example 2.................................................................................9
5. Wiring for Example 3...............................................................................14
6A. Wiring for Example 4............................................................................ 17
6B. Wiring for Example 4 ............................................................................17
7. Hukseflux and Campbell Scientific Variable Names................................24
i
Model HFP01SC
Self-Calibrating Soil Heat Flux Plate
1. General Description
The HFP01SC Soil Heat Flux plate consists of a thermopile and a film heater.
The thermopile measures temperature gradients across the plate. During the insitu field cali bration, the film heater is used to generate a heat flux through the
plate. The amount of power used to generate the calibration heat flux is
measured by the datalogger. Each plate is individually calibrated, at the
factory, to output flux.
In order to measure soil heat flux at the surface, several HFP01SCs are used to
measure the soil heat flux at a depth of eight centimeters. A TCAV Averaging
Soil Thermocouple is used to measure the temporal change in temperature of
the soil layer above the HFP01SC. Finally, a CS616 Water Content
Reflectometer is used to measure the soil water content. The temporal change
in soil temperature and soil water content are used to compute the soil storage
term.
The -L option on the model HFP01SC Soil Heat Flux plate (HFP01SC-L)
indicates that the cable length is user specified. The HFP01SC-L has two
cables, the first cable is the signal output cable and the second is the heater
input cable. Two analog inputs are required to measure the HFP01SC-L This
manual refers to the sensor as the HFP01SC.
2. Specifications
Operating Temperature:-30°C to +70°C
Storage Temperature:-30°C to +70°C
Plate Thickness:5 mm (0.2 in.)
Plate Diameter:80 mm (3.15 in.)
Power Consumption
During Self Calibration:1.3 W (108 mA @ 12 vdc)
Sensor:thermopile and film heater
Measurement Range:±100 W m
Signal Range (nominal):±5 mV for the above range (sensor), 0 to 2 V
(while the film heater is powered)
Accuracy:±3% of reading
Sensitivity (nominal):67 µV W
Sensor Resistance (nominal): 2 Ω
Heater Resistance (nominal): 100 Ω
Current Sensing Resistor:10Ω 1% 0.25 W
Thermal Conductivity:0.8 W m
-2
-1 m-2
-1 K-1
1
Model HFP01SC Self-Calibrating Soil Heat Flux Plate
Up to 1 m
3. Installation
2.5 cm
6 cm
2 cm
Ground Surface
8 cm
Partial emplacement of the HFP01SC and the TCAV
sensors is shown for illustration purposes. All sensors
must be completely inserted into the soil face before
the hole is backfilled.
FIGURE 1. Placement of Heat Flux Plates
The HFP01SC Soil Heat Flux plates, the TCAV Averaging Soil Temperature
probes, and the CS616 Water Content Reflectometer are installed as shown in
Figure 1.
The location of the heat flux plates and thermocouples should be chosen to be
representative of the area under study. If the ground cover is extremely varied,
it may be necessary to have additional sensors to provide a valid spatial average
of soil heat flux.
Use a small shovel to make a vertical slice in the soil. Excavate the soil to one
side of the slice. Keep this soil intact so that is can be replaced with minimal
disruption.
The sensors are installed in the undisturbed face of the hole. Measure the
sensor depths from the top of the hole. With a small kni fe, make a horizo nt al
cut eight centimeters below the surface into the undisturbed face of the hole.
Insert the heat flux plate into the horizontal cut.
NOTE
Install the HFP01SC in the soil such that the side with the text
“this side up” is facing the sky.
2
Model HFP01SC Self-Calibrating Soil Heat Flux Plate
H
U
K
S
E
F
L
U
X
H
E
A
T
F
L
U
X
P
L
A
T
E
CAUTION
In order for the HFP01SC to make quality soil heat flux
measurements, the plate must be in full contact with the
soil.
Never run the sensors leads directly to the surface. Rather, bury the sensor
leads a short distance back from the hole to minimized thermal conduction on
the lead wire. Replace the excavated soil back into it original position after all
the sensors are installed.
Signal (White)
Signal Reference (Green)
Shield (Clear)
Heater Resistor Signal (Yellow)
Heater Resistor Signal Reference (Purple)
Shield (Clear)
Power (Red)
Power Reference (Black)
FIGURE 2. HFP01SC Plate
TABLE 1. Datalogger Connectio ns for a Single-Ended Measurement
Connections to Campbell Scientific dataloggers are given in Tables 1 and 2.
The output of the HFP01SC can be measured using a single-ended analog
measurement (Instruction 1 or VoltSE()), however, a differential analog
measurement (Instruction 2 or VoltDiff()) is recommended.
The wiring convention is that the white wire is positive with respect to the
green wire, when energy is flo wing thr ough the transducer from the side with
the text “this side up” to the other side.
NOTE
The switched 12 vdc port on the black CR10X wiring panel can
source enough current to calibrate five HFP01SC plates; the
CR23X and CR5000 switched 12 vdc port can be used to
calibrate four HFP01SC plates. If additional HFP01SC plates
are needed an external relay is required to power the additional
plates (see example 4).
For dataloggers without a SW12V output (CR7X, 21X and CR10), a relay
(A21REL-12) is required for the in-situ calibration (see Example 4).
5. Example Programs
This section is for users who write their own datalogger programs. A
datalogger program to measure the sensor can be created using Campbell
Scientific’s Short Cut Program Builder software. You do not need to read this
section use Short Cut.
The HFP01SC has a nominal calibration of 15 W m
accompanied by a calibration certificate. Each sensor also has a unique
calibration label on it. The label is located on the pigtail end of the sensor
leads.
4
-2
mV-1. Each sensor is
Model HFP01SC Self-Calibrating Soil Heat Flux Plate
TABLE 3. Wiring for Exa mple 1
DescriptionColorCR10(X)
Sensor SignalWhite1H
Sensor Signal Refere nceGreenAG
ShieldClearG
Heater Resistor SignalYellow1L
Heater Resistor Signal ReferencePurpleAG
ShieldClearG
PowerRedSW12
Power ReferenceBlackG
External Power Controljumper wire SW12-CTRL to C8
Example 1. Sample CR10(X) Program Using a Single-Ended Measurement Instruction
;{CR10X}
;
*Table 1 Program
01: 1Execution Interval (seconds)
;Measure HFP01SC on smaller range.
;
1: Volt (SE) (P1)
1:48Set Port 8 High
;Used to filter data during and after calibration.
;
5: Do (P86)
1:18Set Flag 8 High
6: End (P95)
;End site calibration three minutes after calibration started.
;
7: If time is (P92)
1:4Minutes (Seconds --) into a
2:180Interval (same units as above)
3:30Then Do
7
Model HFP01SC Self-Calibrating Soil Heat Flux Plate
;Measure volta ge across curre nt shunt resistor (10 ohm 1% 0.25 W 50
;ppm/deg C) during calibration. This measurement is used to
;compute power.
;
8: Volt (SE) (P1)
Model HFP01SC Self-Calibrating Soil Heat Flux Plate
TABLE 6A. Wiring for Example 4
DescriptionColorCR10XA21REL-12
Sensor Signal #1White1H
Sensor Signal #2White1L
Sensor Signal #3White2H
Sensor Signal #4White2L
Sensor Signal #5White3H
Sensor Signal #6White3L
All Signal ReferencesGreenAG
All ShieldsClearG
Heater Resistor Signal #1Yellow4H
Heater Resistor Signal #2Yellow4L
Heater Resistor Signal #3Yellow5H
Heater Resistor Signal #4Yellow5L
Heater Resistor Signal #5Yellow6H
Heater Resistor Signal #6Yellow6L
All Heater Resistor Signal ReferencesPurpleAG
All ShieldsClearG
Sensor Power #1RedREL 1 NO
Sensor Power #2RedREL 1 NO
Sensor Power #3RedREL 2 NO
Sensor Power #4RedREL 2 NO
Sensor Power #5RedREL 3 NO
Sensor Power #6RedREL 3 NO
All Power ReferenceBlackG
TABLE 6B. Wiring for Example 4
DescriptionCR10XA21REL-12
Power12V+ 12V
Power ReferenceGGROUND
ControlC8CTRL 1
Controljumper from CTRL 2 to CTRL 1
Controljumper from CTRL 3 to CTRL 2
Powerjumper from REL 1 COM to +12V
Powerjumper from REL 2 COM to REL 1 COM
Powerjumper for REL 3 COM to REL 2 COM
17
Model HFP01SC Self-Calibrating Soil Heat Flux Plate
Example 4. Sample CR10X Program Using External Power and Relay
;{CR10X}
;
*Table 1 Program
01: 1Execution Interval (seconds)
;Measure HFP01SC on smallest range.
;
1: Volt (SE) (P1)
Model HFP01SC Self-Calibrating Soil Heat Flux Plate
6. Soil Heat Flux and Storage
The soil heat flux at the surface is calculated by adding the measured flux at a
fixed depth, d, to the energy stored in the layer above the heat flux plates. The
specific heat of the soil and the change in soil temperature, ∆T
interval, t, are required to calculate the stored energy.
The heat capacity of the soil is calculated by adding the specific heat of the dry
soil to that of the soil water. The values used for specific heat of dry soil and
water are on a mass basis. The heat capacity of the moist is given by:
CCCC C
=+ =+ρθ ρθρ(1)
()
sbdmw bdvww
ρ
w
θ
m
where C
of water, C
on a mass basis,
capacity of water.
This calculation requires site specific inputs for bulk density, mass basis soil
water content or volume basis soil water content, and the specific heat of the
dry soil. Bulk density and mass basis soil water content can be found by
sampling (Klute, 1986). The volumetric soil water content is measured by the
CS616 water content reflectometer. A value of 840 J kg
capacity of dry soil is a reasonable value for most mineral soils (Hanks and
Ashcroft, 1980).
θ
=(2)
v
ρ
b
is the heat capacity of moist soil, ρb is bulk density, ρw is the density
S
is the heat capacity of a dry mineral soil, θm is soil water content
d
θ
is soil water content on a volume basis, and Cw is the heat
v
, over the output
s
-1 K-1
for the heat
23
Model HFP01SC Self-Calibrating Soil Heat Flux Plate
The storage term is then given by Eq. (3) and the soil heat flux at the surface is
given by Eq. (4).
TCd
∆
ss
S
=
t
GG S
=+
sfccm
8
7. Maintenance
The HFP01SC requires minimal maintenance. Check the sensor leads monthly
for rodent damage.
8. In-Situ Calibration Theory
For detailed information on the theory of the in-situ calibration, see the Theory
section of the Hukseflux manual or visit the application section of the
Hukseflux web site at http://www.hukseflux.com/heat%20flux/applic&spec.pdf.
Equation 6 i n t he Hukseflux manual is used to compute a new calib ration every
three hours. The heater is on for a total of 180 seconds. Table 7 lists the
variables used in the Hukseflux manual and those in the example datalogger
programs.
(3)
(4)
TABLE 7. Hukseflux and Campbell Scientific
Variable Names
DescriptionHuksefluxCampbell
Soil Heat Flux
Output of Sensor in mVV
1/Sensitivity1/E
Output of Sensor during
ϕ
sen
sen2
V (0)mV_0
calibration at t=0 seconds
Output of Sensor during
V (180)mV_180
calibration at t=180 seconds
Output of Sensor after
V (360)mV_end
calibration and just before
output
Voltage Across fixed
Ω resistor
10
V
cur
Scientific
shf
shf_mV
cal
V_Rf
24
9. References
Model HFP01SC Self-Calibrating Soil Heat Flux Plate
Hanks, R. J., and G. L. Ashcroft, 1980: Applied Soil Physics: Soil Water and
Temperature Application. Springer-Verlag, 159 pp.
Klute, A., 1986: Method of Soil Analysis. No. 9, Part 1, Sections 13 and 21,
American Society of Agronomy, Inc., Soil Science Society of America,
Inc.
25
Model HFP01SC Self-Calibrating Soil Heat Flux Plate
This is a blank page.
26
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.