This equipment is guaranteed against defects in materials and workmanship.
This guarantee applies for twelve months from date of delivery. We will
repair or replace products which prove to be defective during the guarantee
period provided they are returned to us prepaid. The guarantee will not apply
to:
• Equipment which has been modified or altered in any way without the
written permission of Campbell Scientific
• Batteries
• Any product which has been subjected to misuse, neglect, acts of God or
damage in transit.
Campbell Scientific will return guaranteed equipment by surface carrier
prepaid. Campbell Scientific will not reimburse the claimant for costs incurred
in removing and/or reinstalling equipment. This guarantee and the Company’s
obligation thereunder is in lieu of all other guarantees, expressed or implied,
including those of suitability and fitness for a particular purpose. Campbell
Scientific is not liable for consequential damage.
Please inform us before returning equipment and obtain a Repair Reference
Number whether the repair is under guarantee or not. Please state the faults as
clearly as possible, and if the product is out of the guarantee period it should
be accompanied by a purchase order. Quotations for repairs can be given on
request. It is the policy of Campbell Scientific to protect the health of its
employees and provide a safe working environment, in support of this policy a
“Declaration of Hazardous Material and Decontamination” form will be
issued for completion.
When returning equipment, the Repair Reference Number must be clearly
marked on the outside of the package. Complete the “Declaration of
Hazardous Material and Decontamination” form and ensure a completed copy
is returned with your goods. Please note your Repair may not be processed if
you do not include a copy of this form and Campbell Scientific Ltd reserves
the right to return goods at the customers’ expense.
Note that goods sent air freight are subject to Customs clearance fees which
Campbell Scientific will charge to customers. In many cases, these charges are
greater than the cost of the repair.
Campbell Scientific Ltd,
Campbell Park, 80 Hathern Road,
Shepshed, Loughborough, LE12 9GX, UK
Tel: +44 (0) 1509 601141
Fax: +44 (0) 1509 601091
Email: support@campbellsci.co.uk
www.campbellsci.co.uk
PLEASE READ FIRST
About this manual
Please note that this manual was originally produced by Campbell Scientific Inc. primarily for the
North American market. Some spellings, weights and measures may reflect this origin.
Some useful conversion factors:
2
Area: 1 in
Length: 1 in. (inch) = 25.4 mm
1 ft (foot) = 304.8 mm
1 yard = 0.914 m
1 mile = 1.609 km
In addition, while most of the information in the manual is correct for all countries, certain information
is specific to the North American market and so may not be applicable to European users.
Differences include the U.S standard external power supply det ails where some information (for
example the AC transformer input voltage) will not be applicable for British/European use. Please
note, however, that when a power supply adapter is ordered it will be suitable for use in your country.
Reference to some radio transmitters, digital cell phones and aerials may also not be applicabl e
according to your locality.
(square inch) = 645 mm2
Mass: 1 oz. (ounce) = 28.35 g 1 lb (pound weight) = 0.454 kg
Pressure: 1 psi (lb/in
Volume: 1 UK pint = 568.3 ml
1 UK gallon = 4.546 litres
1 US gallon = 3.785 litres
2
) = 68.95 mb
Some brackets, shields and enclosure options, including wiring, are not sold as standard items in the
European market; in some cases alter n a tiv es are offered. Details of the alt ernatives will be covered in
separate manuals.
Part numbers prefixed with a “#” symbol are special order parts for use with non-EU variants or for
special installations. Please quote the full part number with the # when ordering.
Recycling information
At the end of this product’s life it should not be put in commercial or domestic refuse
but sent for recycling. Any batteries contai ned within the product or used during the
products life should be removed from the product and also be sent to an appropriate
recycling facility.
Campbell Scientific Ltd can advise on the recycling of the equipment and in some cases
arrange collection and the correct disposal of it, although charges may apply for some
items or territories.
For further advice or support, please contact Campbell Scientific Ltd, or your local agent.
Campbell Scientific Ltd, Campbell Park, 80 Hathern Road, Shepshed, Loughborough, LE12 9GX, UK
Tel: +44 (0) 1509 601141 Fax: +44 (0) 1509 601091
Email: support@campbellsci.co.uk
www.campbellsci.co.uk
Contents
PDF viewers note: These page numbers refer to the printed version of this document.
Use the Adobe Acrobat® bookmarks tab for links to specific sections.
8-1. USB and RS-485 Output Elements ....................................................... 23
8-2. Multipliers and Offsets for Analogue Outputs...................................... 24
10-1. Output Modes for EC100 Instruction ................................................. 34
10-2. Bits in the Sonic Diagnostic Flag ....................................... ................ 35
10-3. Bits in the Gas Diagnostic Flag .......................................................... 36
10-4. ConfigCmd Values for Setting and Retrieving Settings ..................... 38
A-1. Filter Time Delays for Various Bandwidths ...................................... A-3
iii
EC155 Packing Information
The EC155 components are placed in a foam cutout that helps protect them from damage during shipment. The
EC155 should look like one of the photographs below, depending on the sample cell option ordered. After unpacking,
it is recommended to save the foam cutout as the EC155 components should be placed in the foam cutout whenever
the EC155 is transported to another location.
Note: Another box containing the Sonic Head and its components will also be shipped with the EC155 analyzer if
Sensing Head Option –SH has been ordered.
EC100 Electronics
with Enclosure
Extra Pocket for End
User Added Parts
Gas Sensor Head Cable
Sample Cell Cable
#26563 USB Cable
EC100 Electronics
with Enclosure
#27007 CD-ROM
Certificate of Conformance
Sample Cell with Intake
and Gas Sensor Head
(if ordered –SI)
#25995 EC100
Enclosure Mounting
#26570 EC155
Mounting Platform
Figure 1. EC155 with Sample Cell Option -SI
Extra Pocket for End
User Added Parts
#27007 CD-ROM
Certificate of Conformance
#25995 EC100
Enclosure Mounting
#26559
Boom Adapter
#26559
Boom Adapter
Gas Sensor Head Cable
Sample Cell Cable
#26563 USB Cable
Figure 2. EC155 with Sample Cell Option -SN
Campbell Scientic, Inc.
|
USA
AUSTRALIA | BRAZIL | CANADA | COSTA RICA | ENGLAND | FRANCE | GERMANY | SOUTH AFRICA | SPAIN
|
815 W 1800 N
Sample Cell with
Gas Sensor Head
(if ordered –SN)
|
Logan, Utah 84321-1784
#26570 EC155
Mounting Platform
|
(435) 753-2342
|
www.campbellsci.com
Printed 6/1/2011
EC155 CO2 and H2O Closed-Path Gas
Analyzer and EC100 Electronics with
Optional CSAT3A 3D Sonic Anemometer
1. Introduction
The EC155 is an in-situ, closed-path, mid-infrared absorption gas analyzer that
measures molar mixing ratios of carbon dioxide and water vapour, along with
sample cell temperature and pressure. The EC155 may be used in conjunction
with the CSAT3 sonic anemometer, which measures orthogonal wind
components.
Before using the EC155, please study
• Section 2, Cautionary Statements
• Section 3, Initial Inspection
• Section 6, Installation
More details are available in the remaining sections.
2. Cautionary Statements
• DANGER:
oThe scrubber bottles (see Section 9.5, Replacing the EC155
Desiccant/CO
sodium hydroxide (caustic soda, NaOH) and anhydrous magnesium
perchlorate (Mg(ClO
Avoid direct contact with the chemicals.
Ensure your work area is well ventilated and free of
Store used chemical bottles in a sealed container until
Dispose of chemicals and bottles properly.
Materials Safety Data Sheets (MSDS) are provided in
• WARNING:
o Do not carry the EC155 or CSAT3A by the arms or carry the EC155
by the strut between the arms. Always hold them by the block,
where the upper and lower arms connect.
o Handle the EC155 carefully. The optical source may be damaged by
rough handling, especially while the analyzer is powered.
o Over-tightening bolts will damage or deform the mounting hardware.
Scrubber Bottles) contain the strong oxidizing agents
2
).
4)2
reactive compounds, including liquid water.
disposal.
Appendix C. MSDS are updated periodically by chemical
manufacturers. Obtain current MSDS at
www.campbellsci.com
.
1
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
• CAUTION:
o Grounding the EC100 measurement electronics is critical. Proper
grounding to earth (chassis) will ensure maximum ESD (electrostatic
discharge) protection and improve measurement accuracy.
o Do not connect or disconnect the gas analyzer or sonic connectors
while the EC100 is powered.
o The SDM, USB, and RS-485 output options include EC155
diagnostic data. Be aware that the absence of diagnostic data in the
analogue output option could make troubleshooting difficult and
may lead to the user not being aware of potential problems with the
instrumentation (see Section 8, EC100 Outp uts ).
o Resting the analyzer on its side during the zero-and-span procedure
may result in measurement inaccuracy.
o When cleaning the gas-analyzer window, make sure the alcohol and
any residual water completely evaporate before proceeding with the
zero-and-span procedure (see Section 9.3, Cleaning Analyzer Windows).
3. Initial Inspection
Upon receipt of your equipment, inspect the packaging and contents for damage.
File damage claims with the shipping company.
4. Overview
The EC155 is a closed-path, mid-infrared absorption analyzer that measures molar
mixing ratios of carbon dioxide and water vapour, along with sample cell
temperature and pressure. It has been designed specifically for eddy covariance
flux measurements and may be used in conjunction with the CSAT3A 3D sonic
anemometer head. The analyzer has a rugged, aerodynamic design with low
power requirements, making it suitable for field applications.
The EC155 gas analyzer connects directly to the EC100 electronics, which
computes real-time CO
cell of the analyzer. A CSAT3A sonic anemometer head may also be connected
to the EC100.
The EC155 has been designed specifically to address issues of aerodynamics,
power consumption, performance during precipitation events, ambient air density
fluctuations, temporal synchronicity, and system integration. Its unique design
enables it to operate with only 4.8 W power; it has minimal spatial displacement
from the sample volume of a CSAT3A sonic anemometer; the EC100 electronics
synchronize data from the EC155 and CSAT3A; and the analyzer is easily
integrated into the CPEC200 closed-path eddy covariance system, a turn-key
system containing data acquisition and control instrumentation, a sample pump,
and optional zero-and-span valve module.
and H2O molar mixing ratios of the air inside the sample
2
2
5. Specifications
5.1 Measurements
Features
• To compute carbon dioxide, water vapour, and sensible heat fluxes using the
These measurements are required to compute carbon dioxide and water vapour
fluxes using the:
• Standard outputs:
eddy-covariance method, the EC155 measures:
o absolute carbon dioxide
o water vapour mixing ratios
o three-dimensional wind speed (requires CSAT3A)
o sonic air temperature (requires CSAT3A)
o sample cell temperature
o barometric pressure.
o CO
mixing ratio, H2O mixing ratio
2
o gas analyzer diagnostic flags
o cell temperature, cell pressure
o CO
signal strength, H2O signal strength
2
o differential pressure
o air temperature and air pressure are auxiliary sensor inputs.
User Manual
• Additional outputs:
, uy, and uz orthogonal wind components
o u
x
o sonic temperature (based on the measurement of c, the speed of
sound)
o sonic diagnostic flags.
Compatibility: CR1000
CR3000
CR5000
Measurement
Rate: 100 Hz
Output bandwidth
Output rate
Operating temperature: -30
Gas analyzer
Measurement precision
CO
density: 0.2 mg·m
2
O density: 0.00350 g·m
H
2
Factory calibrated range
CO
: 0 to 1830 mg·m
2
H
O: 0 to 83 ppt (-60
2
Analyzer temp: -30
2
2
: 5, 10, 12.5, 20, or 25 Hz
: 5 to 50 Hz
o
to 50oC
1
o
to 50oC
-3
(0.15 µmol·mol
-3
(0.006 mmol·mol
-3
(0 to 1000 ppm)
o
to 37oC dew point)
-1
)
-1
)
Baro pressure:70 to 106 kPa
CO
performance
2
Zero max drift
Gain Drift: ±0.1% of reading·°C
Sensitivi t y t o H
3
: ±0.55 mg·m-3·°C-1 (±0.3 μmol·mol·°C
O: ±5.6 x 10
2
-5
µmol CO
-1
(maximum)
-1
·mol
H2O (max)
2
-1
)
3
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
H
O performance
2
Zero max drift
3
: ±0.037 g·m-3·°C-1 (±0.05 mmol·mol-1·°C
Gain Drif t: ±0.3% of reading·°C
Sensitivity to CO
CSAT3A sonic measurement precision
u
: 1 mm·s
x
u
: 1 mm·s
y
: 0.5 mm·s
u
z
Sonic temperature: 0.025
CSAT3A sonic accuracy
Offset error
, uy: < ±8 cm·s
u
x
u
: < 4 cm·s
z
: ±0.05 mol H2O·mol
2
4
-1
-1
-1
o
C
5
-1
-1
Gain error
Wind vector ±5° horizontal: < ±2% of reading
Wind vector ±10° horizontal: < ±3% of reading
Wind vector ±20° horizontal: < ±6% of reading
CSAT3A sonic reporting range
Full scale wind: ±65.553 m/s
Sonic temperature: -50° to +60°C
Sample cell sensors
6
Barometer
Internal basic barometer
Manufacturer: Freescale
Model: MPXAZ6115A
Accuracy: ±2.2 kPa (0
Linear rise: ±2.2 kPa at 0
Optional enhanced barometer:
Manufacturer: Campbell Scientific
Model: CS106 (Vaisala PTB110)
Accuracy: ±0.85 kPa (0
Linear rise: 0.85 kPa at 0°C to 1.90 kPa at -30°C
o 25°C
o 85 kPa
o 19 mmol/mol H
o 326 mmol/mol CO
o 25 Hz bandwidth.
2
user selectable
3
-30° to 50°C
4
noise rms
5
assumes:
o -30° to +50°C
o wind speed <30 m·s
o azimuth angles between ±170°
6
refer to manufacturer’s product brochure or manual for details
O concentration
2
concentration
2
-1
5.2 Output Signals
Features
• EC100 electronics outputs data using:
User Manual
o CS SDM
o RS-485
o USB
o Analogue out
Digital
SDM (Synchronous Device for Measurement)
1
Data type: FLOAT
RS-485
Data type: ASCII
Output Rate: 5 to 50 Hz (user selectable)
Baud rate: 1200 to 230400 bps (user selectable)
USB
Data type: ASCII
Output rate: 5 to 50 Hz (user selectable)
Analogue (two outputs for CO
and H2O molar mixing ratios)
2
Voltage range: 0 mV to 5000 mV
Resolution: 76 µV (16 bit)
Update rate: 150 Hz
Accuracy (at 25
o
C): ±3 mV
CO
mixing ratio equation: µmol/mol = 211.27 (V
2
) – 56.34
out
Full scale range: -56 to 1000 µmol/mol
H
O mixing ratio equation: mmol/mol = 11.31 (V
2
) – 3.04
out
Full scale range: -3 to 53 mmol/mol
1
Synchronous Device for Measurement. A Campbell Scientific, Inc. proprietary serial interface for
datalogger to peripheral and sensor communication. See Section 8.1, SDM Output for details.
5
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
5.3 Physical Description
Sample cell volume: 5.9 cm
Sample ce ll length: 12.0 cm (4.72 in)
Sample cell diameter: 7.94 mm (0.313 in)
Spatial separation between
EC155 optional intake and
CSAT3A sample volume: 15 cm (6.0 in)
Length of tubing from tip
of optional heated intake
to sample cell: 58.4 cm (23 in)
Inside diameter of intake tubing: 2.67 mm (0.105 in)
Dimensions
Analyzer: 42.7 cm x 7.4 cm x 10.1 cm (16.8 in x
9 in x 4.0)
Len gth of optional intake: 38.1 cm (15.0 in) EC1 00 electronics: 24.1 cm x 35.6 cm x 14 cm (9.5 in x 14 in
x 5.5 in)
Cable length:3 m (9.8 ft) from analyzer to EC100
electronics
3
(0.36 in3)
Weight
Analyzer: 3.9 kg (8.5 lbs) Mounting hardware:0.4 kg (0.9 lbs)
EC1 00 electronics and
enclosure: 3.2 kg (7 lbs)
Connections
Pump: 3/8 inch Swagelok
Zero/Span: 1/4 inch Swagelok Sample Intake:1/8 inch Swagelok or Optional Heated
Intake Assembly
6
User Manual
Figure 5-1. Dimensions of EC155 analyzer head with optional heated intake
Figure 5-2. Dimensions of EC155 analyzer head without optional heated intake
7
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
5.4 Power Requirements
Voltage supply: 10 to 16 Vdc
6. Installation
6.1 Mounting
Power at 25
Power at 25
Power at 25
mode (CSAT3A fully powered
and EC155 in stand-by): 3.0 W
Power for optional heated inta ke: set by user, 0 to 0.7 W.
The EC155 is supplied with mounting hardware to attach it to the end of a
horizontal pipe of 1.31 inch outer diameter, such as the CM202, CM204, or
CM206 crossarm (p/n #1790x). The EC155 mounting hardware also
accommodates an optional CSAT3A sonic anemometer, placing it at the proper
position when the EC155 is configured with the optional heated intake assembly.
The following steps describe the normal mounting procedure with the optional
heated intake assembly and optional CSAT3A sonic head. Other mounting
arrangements are acceptable as long as the analyzer is upright. The bottom of the
analyzer has two #6-32 UNC-thread mounting holes for applications that do not
use the EC155 mounting platform.
Refer to Figure 6-1 throughout this section.
o
C including CSAT3A: 4.8 W
o
C excluding CSAT3A: 4.0 W
o
C in power-down
WARNING
a. Mount a CM202, CM204, or CM206 crossarm (p/n #1790X) to a tripod or
other vertical structure using a CM210 crossarm-to-pole bracket (p/n
#17767). The crossarm should be within ± 7 degrees of horizontal to allow
the CSAT3A sonic anemometer to be levelled.
Do not carry the EC155 by the intake or the CSAT3A by t he
arms. Always hold the instrume nts by the body or base.
b. Mount the CM250 levelling mount (p/n #26559) on the end of the crossarm.
Tighten the set screws on the levelling mount.
c. Bolt the mounting platform (p/n #26570) to the CM250 levelling mount (p/n
#26559).
d. Place the EC155 gas analyzer on the mounting platform so the four rubber
feet fit into the platform holes, and tighten the captive screws located on the
bottom of the platform into the mounting holes on the bottom of the analyzer.
e. If a CSAT3A is being used, mount it on the end of the mounting platform
using the captive CSAT3A mounting bolt.
f. Level the assembly by slightly loosening the bolt in the CM250 levelling
mount. Adjust the assembly until the levelling bubble on top of the CSAT3A
is in the bulls eye. Retighten the bolt.
8
User Manual
WARNING
WARNING
NOTE
Over-tightening bolts will damage or deform the mounting
hardware.
Use caution when handling the EC155 gas analyzer. The
optical source may be damaged by rough handling, especially
when the EC155 is po we r ed .
If the assembly is to be mounted on a high tower, it can be hoisted
using the handle on the front of the analyzer and the holes in the
mounting platform.
CSAT3A Sonic
Anemometer Head
EC155 Gas
Analyzer
Mounting Platform
(p/n #26570)
CM250 Leveling
Mount (p/n #26559)
CM20X Crossarm
(p/n #1790X)
Figure 6-1. Exploded view of mounting the EC155 gas analyzer and the
CSAT3A sonic head
9
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
NOTE
The CSAT3A sonic anemometer is an updated version of the
CSAT3, designed to work with the EC100 electronics. An existing
CSAT3 may be upgraded to a CSAT3A. Contact Campbell
Scientific for details.
g. Attach the EC100 electronic enclosure to the mast, tripod leg, or other part of
the mounting structure. To do this, attach the EC100 enclosure mounting
bracket (p/n #26604) to the pipe by loosely tightening the u-bolts around the
pipe. The u-bolts are found in the mesh pocket inside the EC100 enclosure.
If the pipe does not run vertically up-and-down (e.g., if you are attaching the
enclosure to a leg of a tripod), rotate the bracket to the side of the pipe. As
the enclosure must hang up-right, angle adjustments may need to be made by
loosening the four nuts and rotating the bracket plates relative to one another.
If the necessary angle cannot be reached in the given orientation, the four nuts
may be removed and the top plate indexed by 90 degrees to allow the bracket
to travel in the other direction (see Figure 6-2). Once adjusted, tighten all the
nuts. Finally attach the EC100 enclosure to the bracket by loosening the bolts
on the back of the enclosure, hanging the enclosure on the mounting bracket
(it should slide into place and be able to securely hang from the bracket), and
tightening the bolts (see Figure 6-3).
10
Figure 6-2. EC100 enclosure mounting bracket mounted on a vertical
mast (left) and a tripod leg (right)
User Manual
6.2 Plumbing
6.2.1 Flow
6.2.2 Pressure
Figure 6-3. Exploded view of mounting the EC100 enclosure
h. Remove the EC100 enclosure desiccant from the plastic bag and put it back in
the mesh pocket of the enclosure. Adhere the humidity indicator card to the
inside of the enclosure.
The EC155 has a small sample cell volume (5.9 cm3) to give good frequency
response at a relatively low flow rate. The sample cell residence time is 50 ms for
a nominal 7 LPM flow. The CPEC200 pump module is designed to provide this
flow for the EC155, but other user-supplied pumps may be used. There is no
specific limitation to the flow rate that may be used with the EC155, but the
sample cell pressure must be considered.
The EC155 is designed to be used near ambient pressure, but it will not be
damaged by operation under vacuum. The EC155 includes a differential pressure
sensor to measure the sample cell pressure relative to ambient pressure, which has
a range of ± 7 kPa. If the EC155 is operated less than 7 kPa from ambient
pressure, the user must attach a separate, user-supplied pressure sensor.
The pressure drop in the optional heated intake assembly is approximately 2.5 kPa
at 7 LPM flow with no filter. The filter adds approximately 1 kPa pressure drop
when it is clean. This pressure drop will increase as the filter clogs. The filter
should be replaced before the differential pressure reaches -7 kPa (unless the user
has supplied a pressure sensor with a wider range). See Section 9.2, Intake Filter Replacement for details on replacing the filter.
11
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
If the EC155 is configured without the optional heated intake assembly, there is
tubing connecting the sample inlet fitting to the sample cell that will drop the
pressure approximately 4 kPa at 7 LPM.
6.2.3 Filtration
The EC155 will not be damaged by use without an inlet filter, although a coarse
screen (up to 1 mm hole size) is suggested to keep large debris out. Over time
particulates will collect on the optical windows, reducing the signal levels until the
windows must be cleaned (see Section 9.3, Cleaning Analyzer Windows). Using a
filter on the inlet will increase the time before the windows must be cleaned. A
finer pore size will keep the windows clean longer, but will need to be replaced
more frequently. The optional heated intake assembly includes a 20 micron filter
element, which gives a compromise between filter replacement and window
cleaning.
6.2.4 Plumbing Connections
There are three connections to the EC155 sample cell: the sample inlet, zero-andspan inlet, and pump outlet, as illustrated in Figure 6-4. In the normal mode, a
vacuum pump pulls an air sample from the sample inlet through the sample cell.
In zero-and-span mode the pump is turned off and a zero-and-span gas is pushed
backwards through the sample cell, exhausted out through the sample inlet.
Figure 6-4. End view s of the analyzer showing the sample intake ( optional
heated intake not shown), pump outlet, and zero-and-span intake
Analyzer Head Cable
Sample Cell Cable
6.2.4.1 Sample Intake
12
The EC155 can be ordered with a factory-installed intake assembly, or with a
Swagelok fitting to attach a user-supplied intake assembly. If the EC155 is
configured with the intake assembly, it is installed at the factory. No further
assembly is required.
6.2.4.2 Pump
6.2.4.3 Zero and Span
User Manual
If the EC155 is configured with no intake assembly, it has a 1/8 inch Swagelok
fitting at the front end for connection to the user-supplied intake assembly (see
Figure 6-4).
In normal mode the EC155 uses a vacuum pump to pull an air sample through the
sample cell. See the discussion on flow and pressure in the previous section for
pump requirements. The CPEC200 pump module (p/n #26399-x) is designed for
use with the EC155. Connect the CPEC200 pump module or user-supplied
sample pump to the 3/8 inch Swagelok fitting at the back end of the analyzer
labelled Pump.
The zero-and-span inlet connects to the pump connection passage near the outlet
of the sample cell. During normal operation the zero-and-span inlet should be
plugged, either with a Swagelok 1/4 inch plug, or with a tube connecting to a
closed valve or manifold system such as the CPEC200 valve module. During zero
and span the zero or span gas can be pushed into this fitting to flow backward
through the sample cell and exhausted through the intake assembly.
NOTE
The CPEC200 system includes a valve module controlled by a
CR3000 datalogger, which automates the zero gas and CO2 span gas
flows during the zero-and-span procedure.
6.3 Wiring and Connections
Figure 6-5 and Figure 6-6 below show EC100 electronics panel and the bottom of
the EC100 enclosure, respectively. Refer to these figures during wiring and
connecting.
Figure 6-5. EC100 electronics front panel. The picture on the left shows
the panel as it is shipped from the factory (enhanced barometer
shown). The picture on the right shows the panel after the user has
done all the wiring and made all connections (basic barometer used).
13
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
Figure 6-6. Bottom of EC100 enclosure
CAUTION
Do not connect or disconnect the EC155 gas analyzer head
or CSAT3 sonic head while the EC100 is powered.
a. Connect the EC155 gas analyzer head. Begin by removing the black rubber
cable entry plug (p/n #26224) on the bottom right of the EC100 enclosure.
(This plug can be stored in the mesh pocket of the enclosure). Now insert the
cable entry plug that is attached to the large cable of the EC155 gas analyzer
head into the vacant slot. Push the connector at the end of the cable onto its
mating connector (labelled Gas Analyzer) and tighten the thumbscrews (see
Figure 6-6). The EC155 gas analyzer cable is approximately 3 metres in
length.
b. Connect the EC155 sample cell cable. Unscrew the sample cell connector
cover, which is found on the bottom of the EC100 enclosure. Insert the 12prong sample cell cable connector into the female connector on the enclosure
and screw it firmly in place. The EC155 sample cell cable is approximately 3
metres in length.
c. Connect the CSAT3A sonic head (skip this step if not using the CSAT3A).
Similar to (a), begin by removing the black rubber cable entry plug found on
the bottom left of the EC100 enclosure. Insert the cable entry plug on the
CSAT3A cable into the slot and connect the male end to the female connector
labelled Sonic Anemometer on the EC100 electronics (see Figure 6-5).
Tighten the thumbscrews. The CSAT3A cable is approximately 3 metres in
length.
NOTE
14
Unlike previous models of the CSAT3 3D sonic anemometer, the
CSAT3A sonic head and the EC155 gas analyzer head have
embedded calibration information. This means that any CSAT3A
and any EC155 may be used with any EC100.
User Manual
d. Ground the EC100 by attaching a thick wire (e.g., 12 AWG) to the grounding
lug found on the bottom of the EC100 enclosure. The other end of the wire
should be connected to earth (chassis) ground (i.e., grounding rod). For more
details on grounding, see the CR3000 datalogger manual, grounding section.
CAUTION
Grounding the EC100 and other electrical components in the
measurement system is critical. Proper grounding to earth
(chassis) will ensure the maximum ESD (electrostatic
discharge) protection and higher measurement accuracy.
e. Connect a communications signal cable to the EC100. Loosen the nut on one
of the cable entry seals (Cable 1 or Cable 2) on the bottom of the EC100,
remove the plastic plug (the plug can be stored in the mesh pocket in the
enclosure), insert the cable, and retighten the nut by hand. Refer to the
sections below on SDM, USB, RS-485, and analogue communications for
information on required signal cable types and connections to the EC100
panel.
1. SDM Communications: Use cable CABLE4CBL-L (p/n #21972). “L”
denotes the length of the cable, which is customer-specified at time of
order. Table 6-1below details which colour of wire in the cable should
be connected to each terminal found on the SDM connector of the EC100
panel.
Table 6-1. EC100 SDM out put to a Campbell Scientif ic CR1000, CR3000,
or CR5000 Datalogger
EC100 Channel Description Colour
SDM-C1 SDM Data Green
SDM-C2 SDM Clock White
SDM-C3 SDM Enable Red (or brown)
G Digital Ground Black
G Shield Clear
2. USB Communications: Use the EC100 USB cable (p/n #26561) to
connect a PC to the on the bottom of the EC100 enclosure.
3. RS-485 Communications: Use cable CABLE3TP-L (p/n #26987) for
lengths less than 500 ft. The connector on the EC100 panel labelled RS-485 displays which terminals are for receiving and transmitting.
4. Analogue Output: Use CABLE4CBL-L (p/n #21972) or CABLE2TP-L
(#26986-L). Once again, the customer specifies the length of this cable
at time of order. The connector labelled Analogue Outputs on the EC100
panel indicates where each wire should be connected (CO
signal, H
O voltage signal, and two ground connections).
2
voltage
2
f. Wire power and ground (i.e., power reference) cable CABLEPCBL-L (p/n
#21969-L) to the EC100. Feed the cable through one of the cable port
openings in the bottom of the EC100 enclosure and attach the ends into the
green EC100 power connector (p/n #3768). Plug the connector into the
female power connector on the EC100 panel. Ensure that the power and
ground ends are going to the appropriate terminals labelled 12V and ground,
respectively.
g. Connect the power cable to a power source. The power and ground ends may
be wired to the 12V and G ports, respectively, of a Campbell Scientific
datalogger or to another 12 Vdc source.
15
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
h. Once power is applied to the EC100, three LED status lights on the EC100
panel will illuminate. The power LED will be green and the sonic and gas
LEDs will be red until the unit has warmed up and is ready to make
measurements at which time the LEDs will be green. If after several minutes
the LED lights do not turn green, a diagnostic flag has been detected. Check
the individual diagnostic bits to determine the specific fault. Diagnostics may
be monitored using the Status window of ECMon, the user interface software
included with the EC155 (see Section 7, Settings), or with a datalogger (see
Section 10, Datalogger Programming with CRBasic). The diagnostics may
reveal that the unit needs to be serviced (e.g., clean the optical windows of the
sample cell, clear the CSAT3A transducers of ice or debris, etc.).
7. Settings
Operation of the EC155 can be customized by changing the values of the settings.
Factory defaults will work well for most applications, but the user may adjust the
settings with a PC using either the ECMon software (see Section 7.3, ECMon) or
the Device Configuration Utility (see Section 7.4, Device Configur ation Utility),
or with a datalogger using the EC100Configure() instruction (see Section 10.2,
EC100Configure() Instruction).
7.1 Factory Defaults
Table 7-1 shows the default value for each of the settings.
EC100 Basic or EC100 Enhanced (depending
on order)
Auto-Detect (Enabled for EC155)
Disabled
16
7.2 Details
7.2.1 SDM Address
7.2.2 Bandwid t h
User Manual
This section gives an explanation for each setting.
This parameter must be set to use SDM output from the EC100. See Section 8.1,
SDM Output for details on using SDM output.
Each SDM device on the SDM bus must have a unique address. The EC155 has a
factory default SDM address of 1, but may be changed to any integer value
between 0 and 14. The value 15 is reserved as an SDM group trigger.
The EC100 has a user-selectable low-pass filter to select the bandwith (5, 10, 12.5,
20, or 25 Hz). Setting the bandwith to a lower value will reduce noise. However, it
must be set high enough to retain the high-frequency fluctuations in the CO
H
O, or the high frequency contributions to the flux will be lost. The factory
2
default bandwidth of the EC100 is 20 Hz, which is sufficient for most flux
applications. Lower bandwith settings may be used for higher measurement
heights, which inherently have lower frequency content. Refer to Appendix A for
more information on the digital filter options.
and
2
If a spectral analysis is being done to evaluate the experimental setup, the
bandwidth should be set to the Nyquist frequency, which is half the datalogger
sample rate (for SDM output) or half the unprompted output rate (for USB and
RS-485 output). This ensures that the data will not be under-sampled and that
higher frequency variations will not be aliased to lower frequencies. Note that if
too small a bandwidth is selected, high frequency fluxes may be under-measured.
7.2.3 Unprompted Output
If the EC100 is to output data in one of the unprompted modes (USB or RS-485,
see Section 8.2, USB or RS-485 Output), this setting must be set accordingly. The
factory default is to disable the unprompted output, assuming data will be logged
via SDM (see Section 8.1, SDM Output).
Only one unprompted output type (i.e., USB, RS-485) may be selected at a given
time. The rate at which the EC100 outputs these data is determined by the
Unprompted Output Rate setting.
7.2.4 Unprompted Output Rate
This setting determines the output rate for unprompted output (USB or RS-485,
see Section 8.2, USB or RS-485 Output). If the unprompted output is disabled,
this parameter is not used. The factory default output rate is 10 Hz, but it may be
set to 10, 25, or 50 Hz.
7.2.5 RS-485 Baud Rate
If the unprompted output mode is set to RS-485, this parameter determines the
baud rate. Otherwise this setting is not used. The RS-485 baud rate defaults to
115200 bps, although the user may enter another value.
7.2.6 Analogue Output
The EC100 has two analogue outputs for CO2 and H2O molar mixing ratios (see
Section 8.3, Analogue Outputs for more information). These outputs may be
enabled/disabled with this setting. The default is for analogue output to be
disabled.
17
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
7.2.7 ECMon Update Rate
This setting determines the rate at which data are sent over the USB connection to
the PC while running ECMon. The default setting of 10 Hz should be adequate in
most situations.
7.2.8 Temperature Sensor
The EC155 measures the temperature of the sample cell block with a
thermocouple embedded in the block. With the Auto-Select default setting, the
EC100 will automatically detect that an EC155 is connected to the electronics and
will report temperature measurements from the sample cell thermocouple.
To diagnose problems with the temperature measurement, a fixed temperature
value may be used, or the temperature sensor may be selected manually.
7.2.9 Fixed Temperature Value
If the Temperature Sensor setting is None, the EC155 will use the value of this
setting for the sample temperature. This mode is intended for troubleshooting
only. In normal operation, the Temperature Sensor is set to Auto-Select, and this
setting is not used.
7.2.10 Pressure Sensor
This setting determines which pressure sensor will be used to measure the
barometric pressure. The EC100 always includes the EC100 basic barometer, but
it may be ordered with the optional EC100 enhanced barometer. This setting is
factory defaulted to the enhanced barometer if it is ordered, and to the basic
barometer otherwise.
There are two other possible settings for the Pressure Sensor. First, the user may
provide his or her own pressure sensor. In this case the setting should be changed
to User Supplied, with the appropriate values for gain and offset entered (see
below). This option may be used if the EC155 sample cell is to be used outside
the range of the differential pressure sensor (see Section 7.2.11) For this mode the
user-supplied pressure sensor must be plumbed to the EC155 sample cell, and the
Differential Pressure sensor setting should be disabled.
The final option is to select None for the Pressure Sensor setting. The EC100
will use a fixed (see below) value for pressure. This mode is intended for
troubleshooting only.
7.2.10.1 Pressure Gain
If the Pressure Sensor is set to User Supplied, this setting gives the gain factor
(kPa/V) used to convert measured voltage to pressure. Normally the Pressure Sensor is set to EC100 Basic or EC100 Enhanced, and this setting is not used.
7.2.10.2 Pressure Offset
If the Pressure Sensor is set to User Supplied, this setting gives the offset (kPa)
used to convert measured voltage to pressure. Normally the Pressure Sensor is
set to EC100 Basic or EC100 Enhanced, and this setting is not used.
7.2.10.3 Fixed Pressure Value
If the Pressure Sensor setting is None, the EC155 will use the value of this
setting for the barometric pressure. This mode is intended for troubleshooting
only. In normal operation this setting is not used.
18
7.2.11 Differential Pressure
The EC155 includes a differential pressure sensor to measure the pressure
difference between the inside of the sample cell and barometric pressure. With the
Auto-Select default, the EC100 detects the presence of the EC155 and
automatically enables the differential pressure measurement. This pressure
difference is then added to the barometric pressure measurement to give the
pressure in the sample cell.
This setting is only used for closed-path analyzers such as the EC155. It is the
difference between ambient pressure and sample cell pressure. In the case of an
open path analyzer such as the IRGASON or the EC150, this setting should be
disabled.
The EC155 sample cell differential pressure sensor has a range of ± 7 kPa, which
will accommodate most applications. If the sample cell is to be used outside this
range, the user must disable the Differential Pressure sensor and connect a usersupplied pressure sensor (see Section 7.2.10, Pressure Sensor).
7.2.12 Heater Control
When enabled, this setting applies a voltage between 0 and 4.5 Volts to heaters
near the optical windows of the analyzer. Heated windows will inhibit the
formation of condensation, such as dew and frost, and help the analyzer recover
more quickly when precipitation has blocked the optical path. This setting,
however, should be used with caution; excessive heating may lead to density
fluctuations and subsequent flux measurement errors. More information about
heater control, and specifically using a datalogger for control, may be found at the
end of Section 10.2, EC100Configure() Instruction.
User Manual
7.3 ECMon
Settings for the EC155 are easily verified and/or changed by using the Windows
PC support software ECMon (ECMon is short for Eddy Covariance Monitor),
which is found on the EC150 & EC155 Supp or t CD (p/n #27007) or on the
company website in the Support|Downloads section
(www.campbellsci.com/downloads
Before installing ECMon, read the file titled Read.me.text found on the EC150 & EC155 Support CD. This will direct the user to install USB drivers (also found on
the Support CD), which are required for communications between the PC and the
EC100 via the EC100 USB cable (p/n #26561). Once the drivers are installed,
download and run the ECMon.exe install file. Launch ECMon, and connect the
EC100 electronics to the PC with the included EC100 USB cable (p/n #26561).
The USB connection for the EC100 electronics is found on the bottom of the
enclosure (see Figure 6-6). Once connected, select the appropriate
communications port in the ECMon Main Page and click Connect (see Figure
7-1). Next click on the Setup button. All of the above settings are now available
for the user to change (see Figure 7-2).
Besides changing settings, ECMon is also a useful tool for other common tasks
such as:
• Monitoring real-time data from the EC155 using the Display window
• Performing a manual zero and span of the instrument (see Section 9.4, Zero
and Span)
• Troubleshooting and monitoring diagnostics using the Status window (see
Figure 7-3).
).
19
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
Figure 7-1. The Main window of ECMon
20
Figure 7-2. The Setup window in ECMon
Figure 7-3. The ECMon Status window
User Manual
7.4 Device Configuration Utility
The Device Configuration Utility software may also be used to change settings,
although ECMon is generally preferred because of its more user-friendly interface.
Device Configuration may be downloaded from the EC150 & EC155 Support CD
(p/n #27007), or may be downloaded free of charge from the Campbell Scientific
website in the Support|Downloads section (www.campbellsci.com/downloads
Device Configuration requires a USB driver to communicate with the EC100,
similar to ECMon. See Section 7.3, ECMon for notes on installing a USB driver.
After launching the Device Configuration Utility, the user should select “EC100”
from the list of device types. The EC100 electronics should be connected to the
PC with the EC100 USB cable (p/n #26561) and the appropriate USB port
selected before connecting. Once connected, the settings tab displays all the
current settings. The Apply button must be clicked to save any changes.
The Device Configuration Utility is also used to send an updated operating system
to the EC100 electronics. The Send OS tab gives directions on this procedure.
8. EC100 Outputs
The EC100 outputs data in one of four types: SDM, USB, RS-485, or analogue.
In general Campbell Scientific recommends that SDM be used if a Campbell
Scientific datalogger is responsible for data collection. However, RS-485 output
is recommended over SDM if cable lengths exceed 100 metres. If a PC is being
used as the collection vehicle, USB and RS-485 are suitable outputs. Analogue
output may also be used, however only CO
will be output. More information regarding each output type is provided in the
sections below.
mixing ratio and H2O mixing ratio
2
).
21
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
NOTE
8.1 SDM Output
The EC100 synchronously samples the gas in the EC155 sample
cell and the CSAT3A sonic head. However, a delay induced by the
intake assembly must be accounted for to ensure maximum
covariance. The exact delay will depend on the length and size of
the intake tubing and the pump flow rate. See Appendix A or the
CPEC200 manual for details.
SDM (Synchronous Device for Measurement) is a Campbell Scientific
communication protocol that allows synchronized measurement and rapid
communication between a Campbell Scientific datalogger and multiple devices
including the EC155. Although nearly all of the Campbell Scientific dataloggers
support SDM, only the CR1000, CR3000, and CR5000 dataloggers support the
EC155.
To use SDM data output, connect an SDM cable from the EC100 (see Section 6.3,
Wiring and Connections) to a CR1000, CR3000, or CR5000 datalogger. On
CR1000 dataloggers, the SDM protocol uses ports C1, C2, and C3. These are
multipurpose control ports that are SDM-activated when an SDM instruction is
used in the datalogger’s program. On CR3000 and CR5000 dataloggers, the SDM
protocol uses SDM-dedicated ports SDM-C1, SDM-C2, and SDM-C3.
Each SDM device on the SDM bus must have a unique address. The EC155 has a
factory default SDM address of 1, but may be changed to any integer value
between 0 and 14 (see Section 7.2.1, SDM Address).
The sample rate for SDM output is determined by the inverse of the datalogger
scan interval, as set by the user in the datalogger program. Data are output from
the EC100 when a request is received from the logger, i.e. a prompted output
mode. The number of data values sent from the EC100 to the datalogger is also
set by the user in the datalogger program. CRBasic, the programming language
used by Campbell Scientific dataloggers, uses the EC100() instruction to get data
from an EC155. This instruction is explained in detail under Section 10,
Datalogger Programming with CRB asic of this manual.
8.2 USB or RS-485 Output
In contrast to the SDM output mode, which is prompted by a datalogger, data can
also be output from the EC100 via USB or RS485 in an upprompted mode. In this
case the EC100 sends out data without initiation from the receiving device, at a
rate determined by the EC100. Only one unprompted output type (i.e., USB, RS-
485) may be selected at a given time. RS-485 output is recommended over SDM
if cable lengths exceed 100 metres. If a Campbell Scientific datalogger is not
being used to collect the data from the EC155, either unprompted mode is
recommended.
To use USB or RS-485 output, connect a USB or RS-485 cable from the EC100 to
the receiving device (see Section 6.3, Wiring and Connections), and configure the
settings (see Section 7, Settings).
The Unprompted Output parameter must be set to USB or RS-485.
If RS-485 is selected, the RS-485 Baud Rate must be set.
22
The Unprompted Output Rate must be set to the desired output rate.
All output data will be in ASCII format, with each data element separated by a
comma. Each record will terminate with a carriage return and line feed. Table
User Manual
8-1 below lists the elements in each output array, and Figure 8-1 shows an
example USB data feed in terminal mode.
Table 8-1. USB and RS-485 Output Elements
Data Element Description Units/comments
1 Ux m/s
2 Uy m/s
3 Uz m/s
4 Sonic Temperature °C
5 Sonic Diagnostic Flag
6 CO2 Concentration µmol/mol
7 H2O Concentration mmol/mol
8 Gas Diagnostic Flag
9 Air Temperature °C
10 Air Pressure kPa
11 CO2 Signal Strength Nominally 0.0 to 1.0
12 H2O Signal Strength Nominally 0.0 to 1.0
13 Sample Cell Pressure Differential kPa
14 Counter Arbitrary
15 Signature Arbitrary in hexadecimal
Figure 8-1. An example of USB data output in terminal mode
The final data element in each row or output array is the signature, a four character
hexadecimal value that is a function of the specific sequence and number of bytes
in the output array. The recording device (i.e., PC or datalogger) calculates its
own signature using each transmitted byte until encountering the transmitted
signature. The computed signature and the transmitted signature are compared. If
they match, the data were received correctly. This is very similar to a CyclicRedundancy-Check (CRC).
In most situations, a PC begins by reading in the ASCII data and extracting the
last four ASCII characters, casting them as Long data type. The signature is then
calculated on the science data sent from the EC155, starting with CO
and ending
2
on the counter. All the characters after the counter are not part of the signature.
Once the signature is computed using the algorithm below, it is compared to the
transmitted signature. If signatures do not match, the data should be disregarded.
The following block of code is an example implementation of Campbell
Scientific’s signature algorithm in the programming language C. To generate the
signature of an output array of bytes, the seed needs to be initialized to 0xaaaa and
a pointer passed to the first byte of the output array. The number of bytes in the
23
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
output array should be entered in as the swath. The returned value is the
computed signature.
//signature(), signature algorithm.
// Standard signature is initialized with a seed of 0xaaaa.
// Returns signature.
unsigned short signature( unsigned char* buf, int swath,
unsigned short seed ) {
unsigned char msb, lsb;
unsigned char b;
int i;
msb = seed >> 8;
lsb = seed;
for( i = 0; i < swath; i++ ) {
b = (lsb << 1) + msb + *buf++;
if( lsb & 0x80 ) b++;
msb = lsb;
lsb = b;
}
return (unsigned short)((msb << 8) + lsb);
}
8.3 Analogue Outputs
9. Maintenance
9.1 Routine Maintenance
If analogue output is enabled, the EC100 will output two analogue signals that
correspond to CO
Volts. Table 8-2 below gives the multipliers and offsets for the analogue outputs.
Table 8-2. Multipliers and Offsets for Analogue Outputs
Mixing Ratio
(µmol mol
CO2 211.27 -56.34
H2O 11.31 -3.04
There are five basic types of maintenance for the EC155/EC100:
• intake filter replacement (if EC155 was ordered with an intake)
• analyzer window cleaning
• zero and spanning,
• replacing analyzer desiccant/scrubber bottles
• factory recalibration.
The following items should be examined periodically:
density and H2O density. These signals range from 0 to +5
2
-1
)
Voltage Output Multiplier
(µmol mol-1 V-1)
Offset
(µmol mol-1)
24
• Check the humidity indicator card in the EC100 enclosure. If the highest dot
has turned pink, replace or recharge the desiccant bags. Replacement
desiccant bags may be purchased (# 6714), or old ones may be recharged by
heating in an oven. See the manual ENC10/1 2, ENC12/14, ENC14/16, ENC16/18, available at www.campbellsci.com
desiccant bags.
• Make sure the Power and Gas LED status lights on the EC100 panel are
green. If not, verify that all sensors are connected securely and that the
, for more details on recharging
instruments are powered. Also check the individual diagnostic bits for the
specific fault. See Tables 10-2 and 10-3.
9.2 Intake Filter Replacement
This section only applies if your EC155 was ordered with the intake assembly.
The differential pressure between the sample cell and ambient pressure should be
monitored in the output data. (This can also be done using the display screen of
ECMon). If the differential pressure approaches the limit of the full scale range
(-7 kPa), it is likely that the intake filter is clogged and should be replaced. To
replace the filter, follow these steps:
a. Stop the air flow through the EC155.
b. Locate one of the EC155 intake filters (p/n #26072) in the mesh pocket of the
EC100 enclosure.
c. Remove the old filter by pulling on the small santoprene tab on the edge of
the filter (see Figure 9-1). Once removed, make sure the underlying
aluminium disk and intake hole are free from debris.
d. Place a new filter on the aluminium disk. Press along the santoprene edge to
make sure it is well-seated.
User Manual
The standard intake filter has a sintered disk with 20 micron pore size. For
NOTE
dusty sites, an intake filter with 40 micron pores may be ordered (p/n
#28698), which will increase the lifetime of the filter. Ideally, the
appropriate pore size will result in the filter needing replacement at the
same time the windows need cleaning (see Section 9.3). For extremely
dusty conditions, lowering the flow rate through the analyzer will further
increase the lifetime of the filter, although this will also result in a decrease
in frequency response.
Santoprene Tab
9.3 Cleaning Analyzer Windows
Figure 9-1. The underside of the optional heated intake
The windows of the analyzer should be cleaned if the signal strength of CO2 or
O drops below 80% of the original value. (These values may be monitored in
H
2
the output data, or they can be viewed with ECMon.) To clean the windows,
follow these steps:
25
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
a. Stop the air flow through the EC155.
b. Loosen the two captive thumbscrews (one on each end of the EC155), and lift
the top portion of the EC155 shell, leaning it back against the lower shell.
See Figure 9-2.
c. Loosen the thumbscrew on the cable clamp at the back of the analyzer to
release the cable, and loosen the two long thumbscrews found above the
sample cell. Rotate the latches so that the struts on the analyzer are free to
move upwards. See Figure 9-3.
d. Lift the analyzer head off the sample cell (see Figure 9-4), taking care not to
lose the O-rings (# 26212) surrounding the optical windows (see Figure 9-5).
If an O-ring is lost, two replacement O-rings may be found in the mesh pocket
of the EC100 enclosure, or new ones may be ordered.
e. Wash the windows with isopropyl alcohol using cotton swabs or a non-
scratching tissue or cloth.
f. Put the analyzer back in place, making sure the O-rings are still intact. The
analyzer’s label should face out to the side.
g. Rotate the latches back in place to hold the analyzer’s struts down, and
tighten the long thumbscrews by hand. Also make sure the analyzer cable is
seated properly in the cable clamp and tighten the thumbscrew by hand.
Thumbscrew
h. Put the top portion of the EC155 shell back in place, and tighten the
thumbscrews.
Top Shell
Thumbscrew
Cable Clamp
Thumbscrew
Figure 9-2. The EC155 analyzer with the top shell open
26
User Manual
Figure 9-3. By loosening the thumbscrews above the sample cell, the
latches may be spun from position A to position B, thus freeing the
struts of the analyzer.
Figure 9-4. The EC155 analyzer and sample cell with shell top open
27
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
Optical Window
9.4 Zero and Span
O-ring
Optical Window
O-ring
Figure 9-5. Analyzer removed from sample cell and shell
As is the case with optical instrumentation, the EC155 may drift slightly with
exposure to natural elements. Thus, a zero-and-span procedure should be
performed occasionally. The first part of the procedure listed below simply
measures the CO
allows the CO
and H2O span and zero, without making any adjustments. This
2
and H2O gain factors to be calculated. These gain factors quantify
2
the state of the analyzer before the zero-and-span procedure, and in theory could
be used to correct recent measurements for drift. The last part of the zero-and-span
procedure adjusts internal processing parameters to correct subsequent
measurements.
NOTE
28
If the EC155 was purchased as part of a CPEC200 closed-path eddy
covariance system, consult the CPEC200 manual. The CPEC200
system has an optional valve module to allow the datalogger to
automate the zero-and-span procedure.
It is imperative that the zero-and-span procedure be done correctly and not rushed;
allocate plenty of time for the procedure. During a normal zero-and-span
procedure a PC running the ECMon software is used to monitor and control the
EC155. However, the zero-and-span procedure can also be performed using either
the Device Configuration Utility software or a datalogger running the
EC100Configure() instruction (see Section 10.2, EC100Configure() Instruction).
User Manual
To check and then set the EC155 zero and span, follow the steps below:
a. Connect the EC100 to a PC with the EC100 USB cable (p/n #26561), and
launch ECMon on the PC. Select the appropriate USB port, and press
Connect. The main screen should now be reporting real-time CO
concentrations.
Increase to
~900 seconds
and H2O
2
Figure 9-6. ECMon Display window showing real-time CO
and H2O
2
b. Check the differential pressure and replace the intake filter as needed (see
Section 9.2, Intake Filter Replacement).
c. Check the signal strengths and clean the windows as needed (see Section 9.3,
Cleaning Analyzer Windows).
d. Verify that the Gas LED status light on the EC100 panel is green. Also make
sure the analyzer is resting right-side-up. If the zero-and-span procedure is
being performed on-site, this should already be the case.
e. If the EC155 is configured with the heated intake assembly, connect the zero-
and-span gas to the Zero/Span inlet at the back of the analyzer. The zero-and-
span gas will be pushed backwards through the EC155 sample cell and
exhausted through the heated intake assembly. If the sample pump is the
CPEC200 pump module, it may be left connected and simply shut off. A
fraction of the zero-and-span gas will be pushed through the pump, but not
enough to affect the equilibration time. Alternatively, the sample pump may
be disconnected and the Pump connection plugged.
If the EC155 is configured with a sample inlet fitting to connect to the user’s
own intake assembly, there are two options for connecting the zero-and-span
gas:
1. Connect the zero-and-span gas to the Sample inlet, and disconnect the
pump, leaving the Pump connection open. The zero-and-span gas will be
pushed forward through the EC155 sample cell and exhausted out the
Pump fitting. In this case the Zero/Span connection may be left plugged.
2. Connect the zero-and-span gas to the Zero/Span inlet, and disconnect the
intake tube from the Sample connection. Disconnect the sample pump
29
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
and plug the Pump connection. The zero-and-span gas will be pushed
backwards through the EC155 sample cell and exhausted through the
Sample fitting.
f. Click on the Zero/Span button on the main screen of ECMon. A real-time
graph at the bottom of the window will appear that displays concentrations of
and H2O (see Figure 9-6). Next, allow CO2 span gas to flow through the
CO
2
sample cell. The exact flow rate is not important since sample cell pressure is
being measured; however, a flow rate should be high enough to flush the
tubing and sample cell in a reasonable time. If the tubing from the CO
tank to the EC155 is kept short, the CO
even at relatively low flow rate (< 0.5 LPM). However if the tubing is long
(e.g. if the EC155 is left in place at the top of the tower) it may take a few
minutes to flush the tube, and a higher flow rate (> 1 LPM) may be useful to
reduce the equilibration time.
Once gas begins to flow through the sample cell, watch the graph on ECMon
for the measurement readings to stabilize. Once stable, write down the
reported CO
concentration.
2
will equilibrate in several seconds
2
2
span
NOTE
NOTE
Optimally the concentration of span CO
should be near the
2
concentration of CO2 being measured in the field. Also, the user is
advised to use CO
gas. The use of reference CO
mixtures in dry ambient air for the CO2 span
2
gas mixtures in pure nitrogen will
2
lead to errors due to a carrier gas effect on pressure-broadening of
the CO
absorption lines since oxygen gas has a smaller line-
2
broadening coefficient than nitrogen.
g. Stop the flow of CO
span gas, and provide H2O span gas to the analyzer. A
2
dew point generator is often used for this. Allow a relatively high flow rate
for the first several minutes to more quickly stabilize the system, and then
decrease the flow to 0.2 to 0.4 L/min before making the measurement.
Higher flow rates should not be used when taking the measurement because
back-pressure on the dew point generator will cause errors. Write down the
reported H
As H
O may adsorb to surfaces inside the tubing and sample cell,
2
O concentration.
2
allow plenty of time for the system to reach equilibrium.
h. Stop the flow of H
O span gas, and allow zero air (no CO2 or H2O) to flow
2
through the analyzer. Dry nitrogen is often used as zero air. The exact flow
rate is not important since sample cell pressure is being measured, however, a
flow rate should be high enough to flush the tubing and sample cell within a
reasonable time period. Wait for the measurement readings to stabilize and
write down the reported values for CO
and H2O concentrations.
2
30
i. Examine the measurements that were written down for span CO
and zero air. Compute the drift in instrument gain using the following
equation:
gain−=
actual
zerospan
measmeas
span
, span H2O,
2
where,
• span
• span
• zero
is the known concentration of the span gas
actual
is the measured concentration
meas
is the measured concentration with zero gas.
meas
User Manual
Note that in the zero-and-span window of ECMon, span
is reported to the
actual
right of the box where the user enters the span dew-point temperature. The
software calculates span
by taking into account the dew-point temperature
actual
and current ambient temperature and pressure. The equations used for this
calculation may be found in Appendix B. If drift (offset or gain) for CO
H
O is excessive, it may be time to replace the desiccant and CO2 scrubber
2
2
or
bottles (see Section 9.5, Replacing Desiccant and Scrubber Bottles).
j. With the zero air still flowing and measurements stabilized, click on the Zero
and H2O button in the Zero/Span window. This will cause the analyzer
CO
2
to adjust the values of its CO
CO
and H2O concentrations to read zero. Verify the CO2 and H2O
2
Zero and H2O Zero parameters, forcing the
2
concentrations now read zero.
k. Now, remove the zero air source and replace it with the CO
span gas. Allow
2
the gas to flow through the sample cell. Watch for readings to stabilize.
Figure 9-7. ECMon Ze r o/ Span window
l. Enter the known concentration of CO
(in ppm) in the Span Concentration
2
box and press Span. This will cause the analyzer to adjust the value of its
Span parameter, forcing the measured CO
CO
2
specified. Verify the CO
m. Replace the CO
concentration reads the correct value.
2
span gas with an H2O span gas of known dew point. Allow
2
concentration to the value
2
the gas to flow through the sample cell; as before, higher flows may be
desired for a couple minutes to more quickly establish equilibrium before
resuming a flow between 0.2 and 0.4 L/min. Wait for the readings to
stabilize.
31
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
n.Enter the known dew point (in °C) in the Span Dew Point box and press
Span. This will cause the analyzer to adjust the value of its H
parameter, forcing the measured dew point to the value specified. Verify the
dew point reads the correct value.
o. The zero-and-span procedure is now complete.
9.5 Replacing the EC155 Desiccant/CO2 Scrubber Bottles
If more than one year has passed since replacing the desiccant/scrubber, or if zeroand-span readings have drifted excessively (see Section 9.4, Zero and Span
above), the desiccant/scrubber bottles (p/n #26511) within the EC155 analyzer
head should be replaced as follows:
a. Remove the analyzer in the same way as explained in Section 9.3, Cleaning
Analyzer Windows of this manual.
b. Unscrew the large metal plug found at the base of the analyzer next to the
analyzer cable; it should only be hand-tight (see Figure 9-8). Once the plug is
removed, tip the analyzer up so the desiccant/scrubber bottle falls out. Insert
a new bottle lid-first into the analyzer. Firmly screw the plug back in place.
c. On the other end of the analyzer, remove the two seal-screws from the metal
cap (see Figure 9-9). Carefully pull the cap off. Tip the analyzer up so the
desiccant/scrubber bottle falls out. Insert a new bottle lid-first. Push the cap
back on, and use two new seal-screws (included with replacement
desiccant/scrubber bottles) to hold the cap in place.
O Span
2
CAUTION
CAUTION
While the metal cap is removed, avoid touching the detector
and its electroni c s .
d. Insert the analyzer back into place, making sure to latch the analyzer’s struts
down. Do not operate the analyzer for at least 24 hours (longer if in humid
environments) to give the chemicals time to purge the air inside the analyzer.
A zero-and-span procedure should then be performed before resuming
measurements.
The scrubber bottles contain strong oxidizing agents. Avoid
direct contact with the chemicals inside the bottles. Also
ensure your work area is well ventilated and free of any
reactive compounds, including liquid water. Store used
chemical bottles in a sealed container until disposal.
The chemical bottles should be disposed of according to local and federal
regulations. For more information, MSDS (Material Safety Data Sheet) forms for
the chemicals are included in Appendix C.
32
User Manual
Figure 9-8. Replacing the source housing desiccant/scrubber bottle
Figure 9-9. Replacing the detector housing desiccant/scrubber bottle
9.6 Factory Recalibration
When the EC155 is manufactured, it goes through an extensive calibration
process, covering a wide range of temperatures, pressures, and gas concentrations.
All CO
traceable to the WMO Mole Fraction Scale maintained by the Central Carbon
Dioxide Laboratory and the Carbon Cycle Greenhouse Gases Group of the Global
Monitoring Division/National Oceanographic and Atmospheric Administration in
Boulder, CO, USA.
After an extended period of time in the field, the EC155 may need to undergo this
factory calibration again in order to ensure valid measurements. When
recalibration is deemed necessary, contact Campbell Scientific.
For the CSAT3A, refer to the CSAT3A instruction manual for information on
recalibration.
calibration gases used in this process are mixtures of CO2 in ambient air
2
10. Datalogger Programming with CRBasic
CRBasic supports two instructions to communicate with the EC100 via SDM.
The first is the EC100() instruction, which reads measurement data from the
EC100. The second is the EC100Configure() instruction, which receives and
sends configuration settings.
33
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
10.1 EC100() Instruction
The EC100() instruction is used to retrieve data from the EC155 via SDM. The
instruction syntax is:
EC100(Dest,SDMAddress,EC100Cmd)
Dest is the input variable name in which to store the data from the EC155. The
length of the input variable array will depend on the selected value for the
EC100CMd. A value of -99999 will be loaded into Dest(1) if a signature error on
SDM data occurs.
Command Input Variable Length
0 8
1 12
213
SDMAddress defines the address of the EC155 with which to communicate.
Valid SDM addresses are 0 through 14. Address 15 is reserved for the
SDMTrigger() instruction.
EC100Cmd is a parameter that requests the data to be retrieved from the analyzer.
The results for the command will be returned in the array specified by the Dest
parameter. A numeric code is entered to request the data, as shown in Table 10-1.
Table 10-1. Output Modes for EC100 Instruction
Output
Mode
Data
Field
Description
Units
1 Ux m/s
2 Uy m/s
3 Uz m/s
4 Sonic Temperature ºC
0, 1, 2,
5 Sonic Diagnostic Flag
CO
µmol/mol
6
2
7 H2O mmol/mol
8 Gas Diagnostic Flag
9 Air Temperature ºC
10 Air Pressure kPa
1, 2
11
12
2 13
CO
Signal Strength
2
O Signal Strength
H
2
Sample Cell Pressure
Differential
nominally 0.0 ≤ strength
≤1.0
nominally 0.0 ≤ strength
≤1.0
kPa
As shown, all output modes give two diagnostic values, the Sonic Diagnostic Flag and the Gas Diagnostic Flag. The values contain a bit field, with each bit
representing a monitored condition. When a certain condition is detected, the
corresponding bit is set. The value remains set until the event that caused the
condition is no longer present. Table 10-2 and Table 10-3 below describe the bits
in the sonic diagnostic flag and the gas diagnostic flag, respectively.
34
User Manual
Table 10-2. Bits in the Sonic Diagnostic Flag
hex
bit
value decimal Name
Function
0 0x1 1 Low Amp Amplitude is too low
1 0x2 2 High Amp Amplitude is too high
2 0x4 4 Tracking Poor signal lock
3 0x8 8 Hi 3 Axis DC
Delta temperature exceeds
limits
4 0x10 16 Acquiring
Acquiring ultrasonic
signals
5 0x20 32 Cal Mem Err
Sonic head calibration
signature error
35
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
Table 10-3. Bits in the Gas Diagnostic Flag
bit hex value decimal Name Function
0 0x1 1 Bad Data Data are suspect (there is
1 0x2 2 Sys Fault General system fault
2 0x4 4 Sys Startup Gas analyzer is starting up
3 0x8 8 Motor Speed Motor speed outside of
4 0x10 16 TEC Temp TEC temperature exceeds
5 0x20 32 Light Power Source power exceeds
6 0x40 64 Light Temp Invalid source temperature
7 0x80 128 Light I Source current exceeds
8 0x100 256 Power Off Gas head not powered
9 0x200 512 Chan Err Gas input data out of sync
10 0x400 1024 Amb Temp Invalid ambient
11 0x800 2048 Amb Press Invalid ambient pressure
12 0x1000 4096 CO2 I CO2 I exceeds limits
13 0x2000 8192 CO2 Io CO2 Io exceeds limits
14 0x4000 16384 H2O I H2O I exceeds limits
15 0x8000 32768 H2O Io H2O Io exceeds limits
16 0x10000 65536 CO2 Io Var Moving variation in CO2 Io
17 0x20000 131072 H2O Io Var Moving variation in H2O Io
18 0x40000 262144 CO2 Io Ratio CO2 signal level too low
19 0x80000 524288 H2O Io Ratio H2O signal level too low
20 0x100000 1048576 Cal Mem Err Gas head calibration
21 0x200000 2097152 Heater Control Heater control error
This instruction is another way, besides ECMon and Device Configuration, to
retrieve and modify settings. ECMon and Device Configuration are user-
interactive, while the EC100Configure() instruction allows automated control
under CRBasic datalogger programming.
EC100Configure() is a processing instruction. Whether running in pipeline mode
or sequential mode the datalogger will execute the instruction from processing.
This functionality allows the instruction to be placed in conditional statements.
Running from processing also introduces ramifications when attempting to
execute the EC100Configure() instruction while other SDM instructions are
executing in pipeline mode. This instruction locks the SDM port during the
duration of its execution. If the pipelined SDM task sequencer needs to run while
the SDM is locked, it will be held off until the instruction completes. This locking
will likely result in skipped scans when reconfiguring an EC155.
For the EC155 to save settings, it must go through a lengthy write-read-verify
process. To avoid saving the settings after each set command, the result code can
be used to determine if any settings were modified from their original value.
When a change is detected the save settings command (command code 99) can
then be sent to the EC155. The DestSource parameter variable should be set to
2718 to save the settings. The reception of this command is acknowledged but
since it takes up to a second to complete, a successful return code does not mean
that all of the data was successfully written to the appropriate non-volatile
memory.
Result is a variable that contains a value indicating the success or failure of the
command. A result code of 0 means that the command was successfully executed.
If reading a setting, 0 in the result code means that the value in the DestSource
variable is the value the desired setting has in the EC155. When writing a setting,
if the result code is 0, the value and setting were compatible, but the value was not
changed because it contained the same value that was sent. A return code of 1
from the set operation means that the value was valid, different, set and
acknowledged. This allows CRBasic code to control whether or not to save the
settings. NAN (i.e., not a number) indicates that the setting was not changed or
acknowledged or a signature failure occurred.
SDMAddress defines the address of the EC155 to configure. Valid SDM
addresses are 0 through 14. Address 15 is reserved for the SDMTrigger()
instruction.
ConfigCmd is a variable that indicates whether to get or set a setting. The options
are listed in Table 10-4.
DestSource is a variable that will contain the value to read when getting a setting,
or that will contain the value to send when writing a setting to the EC155.
37
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
Table 10-4. ConfigCmd Values for Setting and Retrieving Settings
ConfigCmd
Variable
Set Retrieve
0 100
Setting Description
(some settings list possible values for the DestSource
variable)
21 121 PowerDown: 0 = Gas Head On, 1 = Gas Head Off
99 N/A Save Settings to EEPROM memory
,
2
10.2.1 ConfigCmd 11 Zero-and-Span Control
38
To perform zeroing of CO2 and H
O, ConfigCmd 11 is set to 1. After the EC155
2
completes the zero, it will write the value to -1. The datalogger can poll this value
or simply wait for a period of time to allow the zeroing to complete. To perform
span, the CO
CO
2
the proper value in ppm CO
(ConfigCmd 11) to 2. After the CO
Control setting will change to -2. H
Span Concentration setting (Config Cm d 12 ) must be written to
2
prior to setting the Span/Zero Control setting
2
span is completed, the value of the Span/Zero
2
O span is similar to CO2. First the H2O dew
2
point value (ConfigCmd 13) must be written to the desired value. Then the
Span/Zero Control setting is set to 3. After the EC155 completes the span, the
span control setting is written as -3. ConfigCmd’s 14 through 17 automatically
store the results of the zero-and-span procedure. Each result is a coefficient used
in the gas analyzer’s algorithms for calculating gas concentrations.
10.2.2 ConfigCmd 18 Heater Voltage
Normally the EC100Configure() instruction is run in the datalogger’s processing
task. Skipped scans can occur when the EC100Configure() instruction executes.
When changing operational parameters, these skipped scans are acceptable.
However, it may not be acceptable when changing the heater voltage. ConfigCmd
218 allows the EC100Configure() instruction to operate in the SDM task, thus
avoiding skipped scans. When using ConfigCmd 218, the command must be a
constant and the instruction cannot be placed in a conditional statement.
If the EC155 includes the optional heated intake assembly, this setting gives the
voltage applied to the heater. It can be set to -1 to disable the heater, or set it to
any voltage between 0 and 4.5375 V. The heater prevents condensation in the
intake tube.
The resistance of the heater in the intake assembly is 30 ohms, so the heater power
will be given by:
User Manual
2
V
P
H
The maximum power (at 4.5 V) is 0.7 (W). The heater may be operated
continuously at full power, over the full range of operating temperatures. If
ambient conditions are dry enough to prevent condensation without heating the
intake, the power may be turned down to conserve power. Note that the CPEC200
system automatically controls the intake heater power as needed to prevent
condensation.
(W)
=
30
39
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
10.3 Example CRBasic Program
'CR3000 Series Datalogger
'CR3000 Series Datalogger
Public sonic_irga(13)
Alias sonic_irga(1) = Ux
Alias sonic_irga(2) = Uy
Alias sonic_irga(3) = Uz
Alias sonic_irga(4) = Ts
Alias sonic_irga(5) = diag_sonic
Alias sonic_irga(6) = CO2
Alias sonic_irga(7) = H2O
Alias sonic_irga(8) = diag_irga
Alias sonic_irga(9) = cell_tmpr
Alias sonic_irga(10) = cell_press
Alias sonic_irga(11) = CO2_sig_strgth
Alias sonic_irga(12) = H2O_sig_strgth
Alias sonic_irga(13) = diff_press
Units Ux = m/s
Units Uy = m/s
Units Uz = m/s
Units Ts = C
Units diag_sonic = arb
Units CO2 = umol/mol
Units H2O = mmol/mol
Units diag_irga = arb
Units cell_tmpr = C
Units cell_press = kPa
Units CO2_sig_strgth = arb
Units H2O_sig_strgth = arb
Units diff_press = kPa
The EC155 is a non-dispersive mid-infrared absorption analyzer. Infrared
radiation is generated in the larger block of the analyzer before propagating
through a 12 cm sample cell. Chemical species located within the sample cell will
absorb radiation at characteristic frequencies. A mercury cadmium telluride
(MCT) detector in the smaller block of the gas analyzer measures the decrease in
radiation intensity due to absorption, which can then be related to analyte
concentration using the Beer-Lambert Law:
ε
−
ePP
=
o
User Manual
cl
where P is irradiance after passing through the optical path, P
is initial irradiance,
o
ε is molar absorptivity, c is analyte concentration, and l is pathlength.
In the EC155, radiation is generated by applying constant power to a tungsten
lamp, which acts as a 2200 K broadband radiation source. Specific wavelengths
are then selected using interference filters located on a spinning chopper wheel.
For CO
the molecule’s asymmetric stretching vibrational band. For H
, radiation with a wavelength of 4.3 µm is selected, as it corresponds to
2
O, radiation at 2.7
2
µm, corresponding to water’s symmetric stretching vibrational band, is used.
The EC155 is a dual wavelength single beam analyzer; thus, rather than using a
separate reference cell and detector, the initial intensity of the radiation is
calculated by measuring the intensity of nearby, non-absorbing wavelengths
(4 µm for CO
and 2.3 µm for H2O). These measurements account for any source
2
and detector aging and window contamination.
The chopper wheel spins at a rate of 150 revolutions per second, and the detector
is measured 512 times per revolution, resulting in a detector sampling rate of 76.8
kHz. The detector is maintained at -40
o
C using a 3-stage thermoelectric cooler
and is coupled to a low noise pre-amp module.
The EC100 electronics digitize and process the detector data (along with ancillary
data such as sample temperature and pressure) to give the CO
and H2O
2
concentration for each chopper wheel revolution (150 Hz), filtered to the userspecified bandwidth. The EC100 also synchronously measures and processes data
from an optional CSAT3A 3D sonic anemometer head.
41
EC155 CO2 and H2O Closed-Path Gas Analyzer and EC100 Electronics
with Optional CSAT3A 3D Sonic Anemometer
42
Appendix A. Filter Bandwidth and Time
Delay
The EC100 measures CO2 and H2O from the EC155 Gas Analyzer Head (as well
as wind velocity and sonic temperature from the optional CSAT3A Sonic H ea d) at
150 Hz and then applies a user-selectable low-pass filter. The available filter
bandwidths are 5, 10, 12.5, 20, and 25 Hz. Figure A-1 shows the amplitude
response of these filters. The EC100 filters provide a flat pass band, a steep
transition from pass band to stop ban d, and a well-attenuated stop band. Figure
A-2 compares the EC100 10-Hz filter to a 50-msec movi ng average filter with
approximately the same bandwidth.
The ideal eddy-covariance filte r is one that is wide enough to preserve the lowfrequency signal variations that transport flu x and narrow enou gh to at ten uate
high-frequency noise. In addition, to minimize a liasin g (the misinterpretation of
high-frequency variation as lower-frequency variation) the measurement
bandwidth must be less than half of the sample rate, or the datalogger scan rate.
Two factors complicate choosing the ideal eddy-covariance bandwidth. First, the
flux signal bandwidth varies from one ins tallation to another, and the flux signal
bandwidth varies with mean wind speed at a given installation. Second, the fast
sample rate required to anti-alias a desired signa l bandwidth may result in large,
unwieldy data sets.
Fortunately, the covariance calculation itself relaxe s the ne ed for the idea l
bandwidth. First, the time-averaged (typicall y thirty-minute) covariance
calculations inherently reduce noise, and second, aliasing does not degrade the
accuracy of covariance calculations. Therefore, the factory default for the EC100
bandwidth (20 Hz) is rather wide to preserve the signal variations that transport
flux, and that bandwidth is suitable for most flux applications. Additional
bandwidths are available for experimenters desiring to match the EC100 fi lter
bandwidth to their data acquisition sample rate to avoid aliasing. In this case, the
selected bandwidth should be one-half of the sample rate (or datalogger sca n rate),
and experimenters should be careful to avoid attenuati on of fl ux-car ryi n g signals.
The EC100 Electronics synchronously sample the gas in the EC155 sample cell
and the CSAT3A Sonic Head. However, delays induced by the intake asse m bly
must be accounted for. The exact delay will de pend on the length and size of the
intake tubing and the pump flow rate. This delay needs to be experimentally
determined by shifting the time delay until the covariance of the vertical wi nd and
the gas concentrations are m aximized.
Experimenters wishing to synchronize their EC100 data with other measurem ents
in the data acquisition system must account for the time delay of the EC100 filter.
Table A-1 shows the delay for each of the filter bandwidths. The EC100 provides
a constant time delay for all spectral components within each filter’s pass band.
The following examples show how to use Table A-1. To synchronize EC1 00 data
to other datalogger measurements when the datalogger scan rate is 25 Hz and the
EC100 bandwidth is set to 20 Hz (a 200-msec time delay from Table A-1), delay
the non-EC100 data by five datalogger scans . Similarly, for a 10-Hz datalogger
scan rate and the same 20-Hz EC100 bandwidth, delay the non-EC100 data by two
datalogger scans to match the EC100 data. For the best synchronicity, choose a
datalogger scan interval that is an integer multiple of the EC100 filter delay.
The EC100 measures the gas and wind data at 150 Hz, and the 150-Hz data are
down-sampled to the datalogger’s scan rate through SDM communications (see
Section 5). This process synchronizes the EC100 gas and wind data with other
A-1
Appendix A. Filter Bandwidth and Time Delay
signals measured by the datalogger to within ±3.333 ms (plus or minus one-half of
the inverse of 150 Hz). Alternatively, when sending data to a non-Campbell data
acquisition system, the EC100 down-samples its USB and RS-485 outputs to a
user-selectable rate of 10, 25, or 50 Hz. Although the gas and wind data from the
EC100 remain synchronized with one another, the user must consider the downsampled output interval when synchronizing the EC100 data with other
measurements in their system. These slowe r output intervals will increase the
asynchronicity of EC100 data with other system measurements.
EC100 Bandwidths (Amplitude Responses)
10
1
0.1
No Units
0.01
5 Hz
10 Hz
12.5 Hz
20 Hz
25 Hz
0.001
0.0001
11070
Hertz
Figure A-1. Frequency Response of the EC100 Filter at various bandwidths.
A-2
Appendix A. Filter Bandwidth and Time Delay
EC100 10-Hz Filter Compared to 20-msec Moving Average (Amplitude Responses)
10
1
0.1
EC100 10-Hz Bandwidth Filter
No Units
0.01
0.001
0.0001
10-Hz Bandwidth from a 50-m sec Moving Average
11070
Hertz
Figure A-2. Frequency response comparison of the EC100 10-Hz bandwidth and
a 50-msec moving average.
Table A-1. Filter Time Delays for Various Bandwidths
Bandwidth (Hz) Time Delay (ms)
5 800
10 400
12.5 320
20 200
25 160
A-3
Appendix A. Filter Bandwidth and Time Delay
A-4
Appendix B. Useful Equations
ρ
ρ
ρ
P
R
(
−
The following table lists all the variab les and constants used in the equations
below:
Table of Variables and Constants
Variable or Constant Description Units
c
v
d
X
c
X
v
M
M
M
T
e
f
T
d
c
d
v
tmpdT_
Mass Density from Molar Mixing Ratios
Mass Density mg m-3
CO
2
O Mass Density g m-3
H
2
-3
Mass Density of Dry Air g m
CO
Molar Mixing Ratio
2
(concentration relative to dry air)
H
O Molar Mixing Ratio
2
(concentration relative to dry air)
Molecular Weight of CO
Molecular Weight of dry air 0.029 g mmol
Molecular Weight of H
Ambient Pressure kPa
Universal Gas Constant
Ambient Temperature
Vapour Pressure kPa
Enhancement Factor Arbitrary
Dew Point Temperature
Temporary variable for dew point
calculation
44 mg mmol-1
2
O 0.018 g mmol-1
2
µmol mol
mmol mol
o
C
o
C
Arbitrary
-1
-1
-1
6
−
kPa m3 K-1 mmol-1
103143.8
×
⎛
MX
ρ
=
c
ρ
=
v
=
ρ
d
=
ρ
d
⎛
⎜
ρ
=
d
⎜
⎝
⎜
6
⎜
⎝
()()
()
TR
⎛
⎜
−
P
⎜
1000
⎝
TR
()
PM
()
TR
+
P
()
PMX
)
MeP
d
15.27310
+
vv
XTR
++
100015.273
(B-3)
15.273+
⎞
PX
v
⎟
M
d
⎟
+
X
v
⎠
15.273
+
15.273
(B-4)
⎛
⎞
⎜
⎟
1
−
⎟
⎜
1000
⎠
⎝
⎞
ρ
vcc
⎟
−
(B-1)
⎟
MTR
v
⎠
(B-2)
v
⎞
X
vd
⎟
(B-5)
⎟
X
+
v
⎠
B-1
Appendix B. Useful Equations
()(
)
P
−
+
ρ
Dew Point from Molar Mixing Ratio
T
97.240
tmpd
_
T
=
d
502.17
−
(B-6)
T
tmpd
_
⎛
⎜
lnT
=
tmpd
_
Water Vapour Molar Mixing Ratio from Dew Point
X
v
Water Vapour Mass Density from Dew Point
ρ
=
v
Vapour Pressure from Molar Mixing Ratio and Water Vapour Density
⎜
⎝
e
= (B-9)
1000
e
61121.0
()()
EXPfe
⋅⋅=
61121.0018.0
()
TR
PX
v
()
100061121.0
⎛
502.17
⎜
⎜
97.240
⎝
f
EXP
15.273
⎞
⎟
(B-7)
⎟
Xf
+⋅
v
⎠
295
−−
109.5102.300072.1PTPf
×+×+= (B-8)
⎞
T
d
⎟
(B-10)
⎟
T
+
d
⎠
⎛
502.17
⎜
⎜
97.240
⎝
⎞
T
d
⎟
(B-11)
⎟
T
++
d
⎠
PX
e+=
e
=
Equations (1) and (2) were derived from Leuni ng, 2004; Eq. 6.23.
Equations (6) - (8) and (10) - (11) were derived from Buck, 1981; Eq. 2a, 3a, and
6.
v
(B-12)
1000
X
v
()
TR
v
M
15.273
(B-13)
v
B-2
Appendix C. MSDS (Material Safety Data
Sheet) Forms
C.1 Magnesium Perchlorate MSDS
C-1
Appendix C. MSDS (Material Safety Data Sheet) Forms
C.2 Decarbite MSDS
C-2
Appendix C. MSDS (Material Safety Data Sheet) Forms
C-3
CAMPBELL SCIENTIFIC COMPANIES
Campbell Scientific, Inc. (CSI)
815 West 1800 North
Logan, Utah 84321
UNITED STATES
www.campbellsci.com
Campbell Scientific Africa Pty. Ltd. (CSAf)
Somerset West 7129
www.csafrica.co.za
Campbell Scientific Australia Pty. Ltd. (CSA)
QLD 4814 AUSTRALIA
www.campbellsci.com.au
Campbell Scientific do Brazil Ltda. (CSB)
Rua Luisa Crapsi Orsi, 15 Butantã
CEP: 005543-000 São Paulo SP BRAZIL
www.campbellsci.com.br
Campbell Scientific Canada Corp. (CSC)
11564 - 149th Street NW
Edmonton, Alberta T5M 1W7
www.campbellsci.ca
Campbell Scientific Centro Caribe S.A. (CSCC)
300N Cementerio, Edificio Breller
Santo Domingo, Heredia 40305
www.campbellsci.cc • info@campbellsci.cc
Campbell Scientific Ltd. (CSL)
Shepshed, Loughborough LE12 9GX
UNITED KINGDOM
www.campbellsci.co.uk
Campbell Scientific Ltd. (France)
3 Avenue de la Division Leclerc
www.campbellsci.fr
Campbell Scientific Spain, S. L.
Avda. Pompeu Fabra 7-9
Local 1 - 08024 BARCELONA
www.campbellsci.es
Campbell Scientific Ltd. (Germany)
Fahrenheitstrasse13, D-28359 Bremen
www.campbellsci.de
Please visit www.campbellsci.com to obtain contact information for your local US or International representative.
•info@campbellsci.com
PO Box 2450
SOUTH AFRICA
•sales@csafrica.co.za
PO Box 8108
Garbutt Post Shop
•info@campbellsci.com.au
•suporte@campbellsci.com.br
CANADA
•dataloggers@campbellsci.ca
COSTA RICA
Campbell Park
80 Hathern Road
•sales@campbellsci.co.uk
92160 ANTONY
FRANCE
•info@campbellsci.fr
SPAIN
•info@campbellsci.es
GERMANY
•info@campbellsci.de
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.