This equipment is guaranteed against defects in materials and workmanship.
This guarantee applies for 24 months from date of delivery. We will repair or
replace products which prove to be defective during the guarantee period
provided they are returned to us prepaid. The guarantee will not apply to:
Equipment which has been modified or altered in any way without the
written permission of Campbell Scientific
Batteries
Any product which has been subjected to misuse, neglect, acts of God or
damage in transit.
Campbell Scientific will return guaranteed equipment by surface carrier
prepaid. Campbell Scientific will not reimburse the claimant for costs incurred
in removing and/or reinstalling equipment. This guarantee and the Company’s
obligation thereunder is in lieu of all other guarantees, expressed or implied,
including those of suitability and fitness for a particular purpose. Campbell
Scientific is not liable for consequential damage.
Please inform us before returning equipment and obtain a Repair Reference
Number whether the repair is under guarantee or not. Please state the faults as
clearly as possible, and if the product is out of the guarantee period it should
be accompanied by a purchase order. Quotations for repairs can be given on
request. It is the policy of Campbell Scientific to protect the health of its
employees and provide a safe working environment, in support of this policy a
“Declaration of Hazardous Material and Decontamination” form will be
issued for completion.
When returning equipment, the Repair Reference Number must be clearly
marked on the outside of the package. Complete the “Declaration of
Hazardous Material and Decontamination” form and ensure a completed copy
is returned with your goods. Please note your Repair may not be processed if
you do not include a copy of this form and Campbell Scientific Ltd reserves
the right to return goods at the customers’ expense.
Note that goods sent air freight are subject to Customs clearance fees which
Campbell Scientific will charge to customers. In many cases, these charges are
greater than the cost of the repair.
Campbell Scientific Ltd,
80 Hathern Road,
Shepshed, Loughborough, LE12 9GX, UK
Tel: +44 (0) 1509 601141
Fax: +44 (0) 1509 601091
Email: support@campbellsci.co.uk
www.campbellsci.co.uk
PLEASE READ FIRST
About this manual
Please note that this manual was originally produced by Campbell Scientific Inc. primarily for the North
American market. Some spellings, weights and measures may reflect this origin.
Some useful conversion factors:
Area: 1 in2 (square inch) = 645 mm2
Length: 1 in. (inch) = 25.4 mm
1 ft (foot) = 304.8 mm
1 yard = 0.914 m
1 mile = 1.609 km
Mass: 1 oz. (ounce) = 28.35 g
In addition, while most of the information in the manual is correct for all countries, certain information
is specific to the North American market and so may not be applicable to European users.
Differences include the U.S standard external power supply details where some information (for
example the AC transformer input voltage) will not be applicable for British/European use. Please note,
however, that when a power supply adapter is ordered it will be suitable for use in your country.
Reference to some radio transmitters, digital cell phones and aerials may also not be applicable
according to your locality.
Some brackets, shields and enclosure options, including wiring, are not sold as standard items in the
European market; in some cases alternatives are offered. Details of the alternatives will be covered in
separate manuals.
Part numbers prefixed with a “#” symbol are special order parts for use with non-EU variants or for
special installations. Please quote the full part number with the # when ordering.
1 lb (pound weight) = 0.454 kg
Pressure: 1 psi (lb/in2) = 68.95 mb
Volume: 1 UK pint = 568.3 ml
1 UK gallon = 4.546 litres
1 US gallon = 3.785 litres
Recycling information
At the end of this product’s life it should not be put in commercial or domestic refuse but
sent for recycling. Any batteries contained within the product or used during the
products life should be removed from the product and also be sent to an appropriate
recycling facility.
Campbell Scientific Ltd can advise on the recycling of the equipment and in some cases
arrange collection and the correct disposal of it, although charges may apply for some
items or territories.
For further advice or support, please contact Campbell Scientific Ltd, or your local agent.
DANGER — MANY HAZARDS ARE ASSOCIATED WITH INSTALLING, USING, MAINTAINING, AND WORKING ON
OR AROUND TRIPODS, TOWERS, AND ANY ATTACHMENTS TO TRIPODS AND TOWERS SUCH AS SENSORS, CROSSARMS, ENCLOSURES, ANTENNAS, ETC. FAILURE TO PROPERLY AND COM P LE TE LY ASS E M BLE ,
INSTALL, OPERATE, USE, AND MAINTAIN TRIPODS, TOWERS, AND ATTACHMENTS, AND FAILURE TO HEED
WARNINGS, INCREASES THE RISK OF DEATH, ACCIDENT, SERIOUS INJURY, PROPERTY DAMAGE, AND
PRODUCT FAILURE. TAKE ALL REASONABLE PRECAUTIONS TO AVOID THESE HAZARDS. CHECK WITH YOUR
ORGANIZATION'S SAFETY COORDINATOR (OR POLICY) FOR PROCEDURES AND REQUIRED PROTECTIVE
EQUIPMENT PRIOR TO PERFORMING ANY WORK.
Use tripods, towers, and attachments to tripods and towers only for purposes for which they are designed. Do not
exceed design limits. Be familiar and comply with all instructions provided in product manuals. Manuals are
available at www.campbellsci.eu or by telephoning +44(0) 1509 828 888 (UK). You are responsible for conformance
with governing codes a nd regulations, including safety regulations , and the integrity and location o f structures o r land
to which towers, tripods, and any attachments are attached. Installation sites should be evaluated and approved by a
qualified engineer. If questions or co ncerns arise regarding installation, use, or maintenance of tripods, towers,
attachments, or electrical connections, consult with a licensed and qualified engineer or electrician.
General
•Prior to performing site or installation work, obtain required approvals and permits. Comply with all
governing structure-height regulations, such as those of the FAA in the USA.
•Use only qualified personnel for installation, use, and maintenance of tripods and towers, and any
attachments to tripods and towers. The use of licensed and qualified contractors is highly recommended.
• Read all applicable instructions carefully and understand procedures thoroughly before beginning work.
• Wear a hardhat and eye protection, and take other appropriate safety precautions while working on or
around tripods and towers.
•Do not climb tripods or towers at any time, and prohibit climbing by other persons. Take reasonable
precautions to secure tripod and tower sites from trespassers.
•Use only manufacturer recommended parts, materials, and tools.
Utility and Electrical
•You can be killed or sustain serious bodily injury if the tripod, tower, or attachments you are installing,
constructing, using, or maintaining, or a tool, stake, or anchor, come in contact with overhead or underground utility lines.
•Maintain a distance of at least one-and-one-half times structure height, or 20 feet, or the distance
required by applicable law, whichever is greater, between overhead utility lines and the structure (tripod,
tower, attachments, or tools).
•Prior to performing site or installation work, inform all utility companies and have all underground utilities
marked.
•Comply with all electrical codes. Electrical equipment and related grounding devices should be installed
by a licensed and qualified electrician.
Elevated Work and Weather
• Exercise extreme caution when performing elevated work.
• Use appropriate equipment and safety practices.
• During installation and maintenance, keep tower and tripod sites clear of un-trained or non-essential
personnel. Take precautions to prevent elevated tools and objects from dropping.
•Do not perform any work in inclement weather, including wind, rain, snow, lightning, etc.
Maintenance
•Periodically (at least yearly) check for wear and damage, including corrosion, stress cracks, frayed cables,
loose cable clamps, cable tightness, etc. and take necessary corrective actions.
•Periodically (at least yearly) check electrical ground connections.
WHILE EVERY ATTEMPT IS MADE TO EMBODY THE HIGHEST DEGREE OF SAFETY IN ALL CAMPBELL
SCIENTIFIC PRODUCTS, THE CUSTOMER ASSUMES ALL RISK FROM ANY INJURY RESULTING FROM IMPROPER
INSTALLATION, USE, OR MAINTENANCE OF TRIPODS, TOWERS, OR ATTACHMENTS TO TRIPODS AND TOWERS
SUCH AS SENSORS, CROSSARMS, ENCLOSURES, ANTENNAS, ETC.
Contents
PDF viewers: These page numbers refer to the printed version of this document. Use the
PDF reader bookmarks tab for links to specific sections.
9-1. Symptom, Cause, and Solutions ......................................................... 29
B-1. CR1000 Wiring for One Probe Example Program ........................... B-1
B-2. CR1000 Wiring for Two Probe Example Program .......................... B-1
B-3. CR1000 Wiring For Multiplexer Example Program ........................ B-2
B-4. CR200(X) Wiring for Example Program ......................................... B-3
D-1. CS650 SDI-12 Command and Response Set .................................. D-1
iii
iv
NOTE
CS650 and CS655 Water Content
Reflectometers
1. Introduction
The CS650 and CS655 are multiparameter smart sensors that use innovative
techniques to monitor soil volumetric water content, bulk electrical conductivity,
and temperature. They output an SDI-12 signal that many of our dataloggers can
measure.
The CS650 has 30 cm length rods, whereas the CS655 has 12 cm length rods.
This manual uses CS650 to reference model numbers CS650 and CS655. Unless
specifically stated otherwise, information in the manual applies equally to both
models.
This manual provides information only for CRBasic dataloggers. It
is also compatible with most of our retired Edlog dataloggers. For
Edlog datalogger support, see an older manual at
www.campbellsci.com/old-manuals or contact a Campbell Scientific
application engineer for assistance.
2. Cautionary Statements
READ AND UNDERSTAND the Precautions section at the front of this
manual.
Although the CS650 is rugged, it should be handled as precision scientific
instrument.
External RF sources can affect the probe’s operation. Therefore, the probe
should be located away from significant sources of RF such as ac power lines
and motors.
3. Initial Inspection
Upon receipt of the CS650, inspect the packaging and contents for damage.
File damage claims with the shipping company.
The model number and cable length are printed on a label at the connection
end of the cable. Check this information against the shipping documents to
ensure the expected product and cable length are received.
1
CS650 and CS655 Water Content Reflectometers
4. Quickstart
Short Cut is an easy way to program your datalogger to measure the CS650 and
assign datalogger wiring terminals. The following procedures show using Short Cut to program the CS650.
1. Install Short Cut by clicking on the install file icon. Get the install file from
either www.campbellsci.com, the ResourceDVD, or find it in installations of
LoggerNet, PC200W, PC400, or RTDAQ software.
2. The Short Cut installation should place a shortcut icon on the desktop of your
computer. To open Short Cut, click on this icon.
3. When Short Cut opens, select New Program.
2
User Manual
4. Select Datalogger Model and Scan Interval (default of 5 seconds is OK for
most applications). Click Next.
5. Under the Available Sensors and Devices list, select the Sensors |
Meteorological | Soil Moisture | CS650/CS655 Water Content
Reflectometer. Four options are available that monitor different parameters.
In this tutorial, we’ll select CS650/CS655 Water Content Reflectometer
(VWC, EC, T, P, PA, and VR). Click to move the selection to the
Selected device window. The soil temperature defaults to degree C, and the
sensor is measured hourly. These can be changed by clicking the
Temperature or Measure sensor box and selecting one of the other options.
3
CS650 and CS655 Water Content Reflectometers
6. After selecting the sensor, click at the left of the screen on Wiring Diagram
to see how the sensor is to be wired to the datalogger. The wiring diagram
can be printed out now or after more sensors are added.
5. Overview
7. Select any other sensors you have, then finish the remaining Short Cut steps
to complete the program. The remaining steps are outlined in Short CutHelp,
which is accessed by clicking on Help | Contents |Programming Steps.
8. If LoggerNet, PC400, RTDAQ, or PC200W is running on your PC, and the
PC to datalogger connection is active, you can click Finish in Short Cut and
you will be prompted to send the program just created to the datalogger.
9. If the sensor is connected to the datalogger, as shown in the wiring diagram in
step 6, check the output of the sensor in the datalogger support software data
display to make sure it is making reasonable measurements.
The CS650 measures volumetric water content, electrical conductivity, dielectric
permittivity, and temperature of soils or other porous media. These values are
reported through SDI-12 communication.
4
User Manual
Figure 5-1. CS650 Water Content Reflectometer
Volumetric water content information is derived from the probe’s sensitivity to the
dielectric permittivity of the medium surrounding the probe stainless-steel rods.
The CS650 is configured as a water content reflectometer, with the two parallel
rods forming an open-ended transmission line. A differential oscillator circuit is
connected to the rods, with an oscillator state change triggered by the return of a
reflected signal from one of the rods. The two-way travel time of the
electromagnetic waves that are induced by the oscillator on the rod varies with
changing dielectric permittivity. Water is the main contributor to the bulk
dielectric permittivity of the soil or porous media, so the travel time of the
reflected wave increases with increasing water content and decreases with
decreasing water content, hence the name water content reflectometer.
Electrical conductivity is determined by exciting the rods with a known nonpolarizing waveform and measuring the signal attenuation.
Temperature is measured with a thermistor in contact with one of the rods.
It is well known that transmission line oscillators used for water content
measurements suffer from unwanted increases in oscillation period as increasing
electrical conductivity causes transmission line signal attenuation. The CS650
handles this problem by making an electrical conductivity measurement and then
correcting the oscillator period accordingly. On-board processing within the
sensor head calculates electrical conductivity from the signal attenuation
measurement and combines the result with the oscillation period measurement to
calculate the dielectric permittivity of the media and finally applies the Topp
equation (Topp et al. 1980) to estimate volumetric water content.
Probe electronics are encapsulated in the rugged epoxy probe head.
A five conductor cable including the drain or shield wire is used to provide power
and ground as well as serial communication with the CS650. The CS650 is
intended to communicate with SDI-12 recorders, including Campbell Scientific
dataloggers. The orange Rx wire can be used to communicate by means of RS232 Tx/Rx. The A200 USB-to-Serial Module allows RS-232 serial
communication between a computer and the CS650 by means of Campbell
Scientific’s Device Configuration Utility (DevConfig) software.
5
CS650 and CS655 Water Content Reflectometers
Table 6-1. Size Specifications
CS650
CS655
Rods
300 mm long
3.2 mm diameter
32 mm spacing
120 mm long
3.2 mm diameter
32 mm spacing
Probe Head
L 85 mm
W 63 mm
D 18 mm
L 85 mm
W 63 mm
D 18 mm
Probe Weight
280 g
240 g
Cable Weight
35 g m-1
35 g m-1
The CS650’s cable can terminate in:
Connector that attaches to a prewired enclosure (option –PW). Refer to
www.campbellsci.com/prewired-enclosuresfor more information.
Campbell Scientific also offers the CS650-LC, CS655-LC, and CWS655. The
CS650-LC and CS655-LC include a connector for attaching the sensor to an
ET107 weather station. The CWS655 is a wireless version of our CS655; refer to
the Wireless Sensor Network manual for more information.
6. Specifications
Features:
Larger sample volume reduces error
Measurement corrected for effects of soil texture and electrical
conductivity
Estimates soil-water content for a wide range of mineral soils
Versatile sensor—measures dielectric permittivity, bulk electrical
conductivity (EC), and soil temperature
Compatible with Campbell Scientific CRBasic dataloggers: CR6,
CR200(X) series, CR800 series, CR1000, CR3000, and CR5000
Table 6-1 compares the size of the CS650 and CS655 reflectometers.
6
6.1 Electrical Specifications
Sensor Output:
SDI-12
Serial RS-232
Warmup Time:
3 s
Measurement Time:
3 ms to measure
600 ms to complete SDI-12 command
Power Supply
Requirements:
6 Vdc to 18 Vdc
Must be able to supply 45 mA @ 12 Vdc
Maximum Cable Length:
610 m (2000 ft) combined length for 1 – 10
sensors connected to the same datalogger
control port
Electromagnetic
Compatibility:
Œ compliant (EMC compliant performance
criteria available upon request)
Meets EN61326 requirements for protection
against electrostatic discharge and surge
External RF sources can affect CS650
measurements. CS650 circuitry should be
located away from radio transmitter aerials and
cables, or measurements ignored during RF
transmissions.
Active (3 ms):
45 mA typical @ 12 Vdc
(80 mA @ 6 Vdc, 35 mA @ 18 Vdc)
Quiescent:
135 A @ 12 Vdc
Average Current Drain:
I = 0.09n + [3.5 + 0.024(n-1)]n/s
I = average current in milliamps
n = number of CS650’s
s = number of seconds between measurements
(see Figure 6-1)
Figure 6-1 shows average current drain for different measurement rates and
quantities of CS650 probes. If the time between measurements is five minutes or
longer, average current drain may be approximated at 0.15 milliamps per sensor.
6.2 Operational Specifications
Table 6-2 provides the operational specifications.
8
User Manual
Volumetric Water Content using Topp Equation (m3/m3)
Range
5% to 50%
Accuracy†
3% VWC typical in
mineral soils where
solution EC 3 dS/m
3% VWC typical in
mineral soils where
solution EC 10 dS/m
Precision‡
<0.05%
Electrical Conductivity
Range Solution EC
0 to 3 dS/m
0 to 8 dS/m
Range Bulk EC
0 to 3 dS/m
0 to 8 dS/m
Accuracy†
(5% of reading + 0.05 dS/m)
Precision‡
0.5% of BEC
Temperature
Soil Measurement Range
–10 to 70 °C
Operational Range
0 to 70 °C
Accuracy†
0.5 C for probe body buried in soil
Precision‡
0.02 °C
Sensing Volume*
7800 cm3
3600 cm3
*Sensing Volume approximately 7.5 cm radius around each probe rod and 4.5 cm
beyond the end of the rods
†Accuracy specifications are based on laboratory measurements in a series of
solutions with dielectric permittivities ranging from 1 to 81 and solution electrical
conductivities ranging from 0 to 3 dS/m.
‡Precision describes the repeatability of a measurement. It is determined for the
CS650 by taking repeated measurements in the same material. The precision of the
CS650 is better than 0.05 % volumetric water content and 0.01 dS/m electrical
conductivity.
7. Installation
If you are programming your datalogger with Short Cut, skip Section 7.3,
Datalogger Wiring (p. 10), and Section 7.4, Programming (p. 11). Short Cut does this
work for you. See Section 4, Quickstart (p. 2), for a Short Cut tutorial.
7.1 Orientation and Placement
The CS650 measures the bulk dielectric permittivity, average volumetric water
content, and bulk EC along the length of the rods, which is 30 cm for the CS650
and 12 cm for the CS655. The probe rods may be inserted vertically into the soil
surface or buried at any orientation to the surface. The probe may be installed
horizontal to the surface to detect the passing of wetting fronts or other vertical
water fluxes.
The sensitive volume depends on the surrounding media. In soil, the sensitive
volume extends approximately 7.5 cm (3 in) from the rods along their length and
4.5 cm (1.8 in) beyond the end of the rods. Consequently, if the probe is buried
horizontally closer than 7.5 cm from the soil surface, it will include air above the
surface in its measurements and underestimate soil water content.
9
CS650 and CS655 Water Content Reflectometers
The thermistor used to measure temperature is in contact with one of the stainless
steel rods at the base of the epoxy probe body. Because of the low thermal
conductivity of stainless steel, the thermistor does not measure the average
temperature along the rod, but instead provides a point measurement of the
temperature within the epoxy. For a valid soil temperature reading, the probe
body must be in thermal equilibrium with the soil. If the probe is installed
vertically with the epoxy probe body above the surface, then the probe body must
be shielded from solar radiation and in direct contact with the soil or media of
interest.
7.2 Proper Insertion
The method used for probe installation can affect the accuracy of the
measurement. The probe rods should be kept as close to parallel as possible when
installed to maintain the design wave guide geometry. The probe is more
sensitive to permittivity close to the rods so probes inserted in a manner which
generates air voids around the rods will have reduced measurement accuracy. In
most soils, the soil structure will recover from the disturbance during probe
insertion.
In some applications, installation can be improved by using the CS650G insertion
guide tool (Figure 7-1). The CS650G is inserted into the soil and then removed.
This makes proper installation of the water content reflectometer easier in dense or
rocky soils.
10
Figure 7-1. CS650G Insertion Guide Tool
7.3 Datalogger Wiring
Our dataloggers typically use SDI-12 to measure the sensor because RS-232
communication requires more control ports per CS650 and RS-232 programming
is more complicated than SDI-12 programming. SDI-12 communication also
allows up to ten probes to be given different addresses and then share a single
control port.
User Manual
Table 7-1. CS650 Wiring Code for SDI-12
Colour
Function
Datalogger Connection
Green
SDI-12 Data
SDI-12 Input, Control Port,
or U* configured for SDI-12
measurement
Red
SDI-12 Power
12 V
Black
SDI-12 Reference
G
Clear
Shield
G
Orange
Not Used
G
NOTE
NOTE
Table 7-1 shows the SDI-12 wiring for the CS650 water content reflectometer.
SDI-12 data is transmitted to a CRBasic datalogger odd numbered control port or
U terminal. Wiring information for RS-232 communications is provided in
Section 8.1.1.2, RS-232 Wiring(p. 12).
The orange Rx wire is only used for RS-232 Tx/Rx communication,
and should be grounded when using SDI-12.
7.4 Programming
Short Cut is the best source for up-to-date datalogger programming code.
Programming code is needed,
If your data acquisition requirements are simple, you can probably create and
maintain a datalogger program exclusively with Short Cut. If your data
acquisition needs are more complex, the files that Short Cut creates are a great
source for programming code to start a new program or add to an existing custom
program.
Short Cut cannot edit programs after they are imported and edited in
CRBasic Editor.
A Short Cut tutorial is available in Section 4, Quickstart (p. 2). If you wish to
import Short Cut code into CRBasic Editor to create or add to a customized
program, follow the procedure in Appendix A, Importing Short Cut Code into a Program Editor(p. A-1). Programming basics for CRBasic dataloggers are
provided here. Complete program examples for select CRBasic dataloggers can
be found in Appendix B, Example Programs(p. B-1). Programming basics and
programming examples for Edlog dataloggers are provided at
www.campbellsci.com\old-manuals.
when creating a program for a new datalogger installation
when adding sensors to an existing datalogger program
The SDI12Recorder() measurement instruction programs CRBasic dataloggers
(CR6, CR200(X)-series, CR800-series, CR1000, CR3000, and CR5000) to
measure the sensor. This instruction sends a request to the sensor to make a
measurement and then retrieves the measurement from the sensor. See Section
8.2, SDI-12 Measurements (p. 19), for more information.
When using a CR200(X), the SDI12Recorder() instruction has the following
syntax:
The A200 Sensor-to-PC Interface allows communication between a CS650 and a
PC, allowing sensor settings to be changed through Device Configuration Utility
(DevConfig) software.
8.1.1 Using the A200
8.1.1.1 Driver Installation
If the A200 has not been previously plugged into your PC and your PC operating
system is not Windows 7, the A200 driver needs to be loaded onto your PC.
8.1.1.2 RS-232 Wiring
Drivers should be loaded before plugging the A200 into the PC.
The A200 drivers can be downloaded, at no charge, from:
www.campbellsci.com/downloads.
One end of the A200 has a terminal block while the other end has a type B female
USB port. The terminal block provides 12V, G, Tx, and Rx terminals for
connecting the sensor (see Figure 8-1 and Table 8-1).
12
Figure 8-1. A200 Sensor-to-PC Interface
A data cable, pn #17648, ships with the A200. This cable has a USB type-A male
connector that attaches to a PC’s USB port, and a type B male connector that
attaches to the A200’s USB port.
Table 8-1. CS650 Wiring Code
for RS-232 and A200
Colour
Function
A200 Terminal
Orange
RxD
Rx
Green
TxD
Tx
Red
Power
+12 Vdc
Black
Reference
G
Clear
Shield
G
8.1.1.3 Powering the Sensor
The A200 provides power to the sensor when it is connected to a PC’s USB port.
An internal DC/DC converter boosts the 5 Vdc supply from the USB connection
to a 12 Vdc output that is required to power the sensor.
8.1.1.4 Determining which COM Port the A200 has been Assigned
When the A200 driver is loaded, the A200 is assigned a COM port number. This
COM port number is needed when using DevConfig. Often, the assigned COM
port will be the next port number that is free. However, if other devices have been
installed in the past (some of which may no longer be plugged in), the A200 may
be assigned a higher COM port number.
User Manual
To check which COM port has been assigned to the A200, you can monitor the
appearance of a new COM port in the list of COM ports offered in your software
package such as LoggerNet before and after the installation, or look in the
Windows Device Manager list under the ports section (access via the control
panel).
8.1.2 Device Configuration Utility (DevConfig)
DevConfig may be downloaded from the Campbell Scientific website,
www.campbellsci.com/downloads.
Connect the CS650 to the A200 as shown in Table 8-1. Connect the PC to the
A200 USB port with the supplied USB cable.
Launch DevConfig and select CS650 Series from the Device Type menu on the
left. Select 9600 from the Baud Rate drop-down menu.
13
CS650 and CS655 Water Content Reflectometers
Select the appropriate PC serial port from the list of available COM ports shown
when the browse button on the lower left is selected (see Section 8.1.1.4,
Determining which COM Port the A200 has been Assigned(p. 13)).
Select Ok and then Connect to begin communication with the CS650.
14
8.1.2.1 Settings Editor Tab
The Settings Editor tab shows settings stored in the CS650 firmware. Settings
that may be modified include User Name, SDI-12 Address, and RS-232 Baud
Rate. Attempts to change any of the other settings will result in a “Commit failed.
Unrecognized error condition” error message. DevConfig polls the CS650 every
two seconds while connected and the results are displayed in the Real-Time
Measurements field (Table 8-2). This is useful for verifying probe performance.
User Manual
Default communication settings are 9600 baud, no parity, 1 stop bit, 8 data bits,
and no error checking. After any changes to CS650 settings, select Apply to write
the changes to the CS650 firmware. A configuration summary is then shown.
The summary may be printed or saved electronically for future reference.
15
CS650 and CS655 Water Content Reflectometers
Table 8-2. Real-Time Measurements
Measurement Field Name
Meaning
VWC
Volumetric Water Content
EC (dS/m)
Bulk Electrical Conductivity
TS (°C)
Soil Temperature
Ka
Bulk Dielectric Permittivity
PA (S)
Period Average
VR
Voltage Ratio
8.1.2.2 Send OS Tab
The Send OS tab is used to update the firmware in the CS650. The firmware is
available at www.campbellsci.com/downloads. The file to send will have a
filename extension of .a43, such as CS65X.Std.00.36.a43. Sending a new
operating system will not affect any of the user-modified settings or probe specific
multiplier and offset settings.
16
User Manual
To download a new operating system, follow the Operating System Download
Procedure listed on the Send OS tab.
8.1.2.3 Terminal Tab
The Terminal tab may be used to send serial commands directly to the CS650.
See Table 8-3 for a list of serial interface commands.
To send a command from the Terminal tab, left click in the field to get a flashing
black cursor, then press <Enter> several times until the CS650> prompt is shown.
At the prompt, type in the command then <Enter>.
17
CS650 and CS655 Water Content Reflectometers
Table 8-3. CS650 Terminal Commands
Command
Values Returned
Units
0
1) Volumetric Water Content,
2) Electrical Conductivity,
3) Temperature
m3/m3
dS/m
°C
1
1) Permittivity,
2) Electrical Conductivity,
3) Temperature
dS/m
°C
2
1) Period,
2) Voltage Ratio,
3) Temperature
Sec
°C
3
1) Volumetric Water Content,
2) Electrical Conductivity,
3) Temperature
4) Permittivity,
5) Period,
6) Voltage Ratio,
m3/m3
dS/m
°C
Sec
5
1) Copyright information
2) OS version and Date
3) Product Serial Number
4) Product User Name
5) SDI-12 Address
H or h
Help Menu
18
8.2 SDI-12 Measurements
Table 8-4. CS650 SDI-12 Commands
SDI-12 command
(“a” is the sensor
address)
Values Returned
Units
aM!
1) Volumetric Water Content,
2) Electrical Conductivity,
3) Temperature
m3/m3
dS/m
°C
aM1!
1) Permittivity,
2) Electrical Conductivity,
3) Temperature
dS/m
°C
aM2!
1) Period,
2) Voltage Ratio,
3) Temperature
Sec
°C
aM3!
1) Volumetric Water Content,
2) Electrical Conductivity,
3) Temperature
4) Permittivity,
5) Period,
6) Voltage Ratio,
m3/m3
dS/m
°C
Sec
aM4! .. aM9!
No Values Returned
?!
Returns the SDI-12 Address
aI!
CampbellSci, OS version, Product
Serial Number
NOTE
The CS650 responds to SDI-12 commands M!, M1!, M2!, M3!, ?!, and I!. Table
8-4 shows the values returned for each of these commands.
See Section 7.3, Datalogger Wiring(p. 10), for SDI-12 wiring details.
This section briefly describes using the SDI-12 commands.
Additional SDI-12 information is available at Appendix D, SDI-12
Sensor Support (p. D-1), www.sdi-12.org, or
www.youtube.com/user/CampbellScientific.
User Manual
8.2.1 Use of Multiplexers
Up to 10 CS650 probes may be connected to the same datalogger control port as
long as each one has a unique SDI-12 address. The CS650 ships with a default
SDI-12 address of 0 unless otherwise specified at the time of ordering. The SDI12 address may be changed through DevConfig software (see Section 8.1, A200 and Device Configuration Utility(p. 12)) or with a terminal emulator in SDI-12
transparent mode (see Appendix D, SDI-12 Sensor Support(p. D-1)).
SDI-12 communication is established using the SDI12Recorder() CRBasic
instruction. See Appendix D, SDI-12 Sensor Support(p. D-1), for more detail on
SDI-12 communication.
Multiplexers such as Campbell Scientific’s AM16/32B may be used to connect up
to 32 CS650 probes to a single control port. When using multiplexers, the
simplest configuration is for all probes to have the same SDI-12 address.
19
CS650 and CS655 Water Content Reflectometers
CAUTION
When multiplexing CS650 probes, the switched 12V channel should be used so
that power to the sensor may be turned off under program control before the
multiplexer switches to the next channel.
Failure to turn off the switched 12 volt channel before clocking
the multiplexer will result in damage to the multiplexer relays.
The proper sequence in the datalogger program for measuring CS650 probes on a
multiplexer is:
1. Set RES control port high to enable multiplexer
2. Pulse CLK control port to advance to next multiplexer channel
3. Set switched 12 volt channel high to supply power to CS650
4. Send SDI-12 command(s) to CS650
5. Set switched 12 volt channel low to remove power from CS650
6. Repeat steps 2 through 5 for each CS650 connected to the multiplexer
7. Set RES control port low to disable multiplexer
Program examples in Appendix B, Example Programs(p. B-1), show the commands
used in CRBasic.
8.3 Water Content Reflectometer Method for Measuring
Volumetric Water Content
8.3.1 Description of Measurement Method
For the water content measurement, a differential emitter-coupled logic (ECL)
oscillator on the circuit board is connected to the two parallel stainless steel rods.
The differentially driven rods form an open-ended transmission line in which the
wave propagation velocity is dependent upon the dielectric permittivity of the
media surrounding the rods. An ECL oscillator state change is triggered by the
return of a reflected signal from the end of one of the rods.
The fundamental principle for CS650 water content measurement is that the
velocity of electromagnetic wave propagation along the probe rods is dependent
on the dielectric permittivity of the material surrounding the rods. As water
content increases, the propagation velocity decreases because of increasing
dielectric permittivity. Therefore, the two-way travel time of the rod signal is
dependent upon water content, hence the name water content reflectometer.
Digital circuitry scales the high-speed oscillator output to an appropriate
frequency for measurement by an on-board microprocessor. Increases in
oscillation period resulting from signal attenuation are corrected using an
electrical conductivity measurement. A calibration equation converts period and
electrical conductivity to bulk dielectric permittivity. The Topp equation is used
to convert from permittivity to volumetric water content.
20
8.3.2 The Topp Equation
The relationship between dielectric permittivity and volumetric water content in
mineral soils has been described by Topp et al. (1980) in an empirical fashion
using a 3rd degree polynomial. With v the volumetric water content and Ka the
bulk dielectric permittivity of the soil, the equation presented by Topp et al. is
User Manual
bulksolution
vsolid
v = -5.3*10-2 + 2.92*10-2Ka – 5.5*10-4K
It has been shown in numerous research efforts that this equation works well in
most mineral soils, so a soil specific calibration of the CS650 probe is usually not
necessary. If a soil specific calibration is desired, the user can generate an
equation relating Ka to v following the methods described in Section 8.4, Water Content Reflectometer User-Calibration(p. 23).
8.3.3 Electrical Conductivity
8.3.3.1 Soil Electrical Conductivity
The quality of soil water measurements which apply electromagnetic fields to
wave guides is affected by soil electrical conductivity. The propagation of
electromagnetic fields in the configuration of the CS650 is predominantly affected
by changing dielectric permittivity due to changing water content, but it is also
affected by electrical conductivity. Free ions in soil solution provide electrical
conduction paths which result in attenuation of the signal applied to the
waveguides. This attenuation both reduces the amplitude of the high-frequency
signal on the probe rods and reduces the bandwidth. The attenuation reduces
oscillation frequency at a given water content because it takes a longer time to
reach the oscillator trip threshold.
It is important to distinguish between soil bulk electrical conductivity and soil
solution electrical conductivity. Soil solution electrical conductivity refers to the
conductivity of the solution phase of soil. Soil solution electrical conductivity,
the solution from the solid and then measuring the electrical conductivity of the
extracted solution.
can be determined in the laboratory using extraction methods to separate
solution
2
+ 4.3*10-6K
a
3
a
The relationship between solution and bulk electrical conductivity can be
described by (Rhoades et al., 1976)
with
solution;
being the electrical conductivity of the bulk soil;
bulk
, the solid constituents; v, the volumetric water content; and, a
solid
solution
, the soil
soil-specific transmission coefficient intended to account for the tortuosity of the
flow path as water content changes. See Rhoades et al., 1989 for a form of this
equation which accounts for mobile and immobile water. This publication also
discusses soil properties related to CS650 operation such as clay content and
compaction. The above equation is presented here to show the relationship
between soil solution electrical conductivity and soil bulk electrical conductivity.
Most expressions of soil electrical conductivity are given in terms of solution
conductivity or electrical conductivity from extract since it is constant for a soil.
Bulk electrical conductivity increases with water content so comparison of the
electrical conductivity of different soils must be at the same water content.
The calibration equation in the CS650 firmware corrects the oscillation frequency
for the effects of
10 dS m-1 for the CS655. This is equivalent to
dS m-1 and 2.7 dS m-1 respectively. If
up to 3 dS m-1 for the CS650 and up to
solution
bulk
exceeds these limits, the CS650 probe
bulk
values of approximately 0.8
will return 99999 for dielectric permittivity and volumetric water content. The
measured period average and voltage ratio values will continue to be reported
even if the bulk EC is outside the operational range of the probe.
21
CS650 and CS655 Water Content Reflectometers
8.3.3.2 Temperature Correction of Soil Electrical Conductivity
The EC value reported by the CS650 is bulk electrical conductivity. This value is
temperature dependent, changing by 2% per degree Celsius. To compensate for
the effect of temperature, EC readings may be converted to a standard
temperature, such as 25 °C using the following equation:
where EC25 is the
temperature T
(°C).
soil
EC25 = ECT / (1 + 0.02*(T
value at 25 °C and ECT is the
bulk
soil
-25)
bulk
value at soil
8.3.4 Error Sources in Water Content Reflectometer Measurement
8.3.4.1 Probe-to-Probe Variability Error
All manufactured CS650s/CS655s are checked in standard media to develop a
probe specific span and offset value for electrical conductivity and dielectric
permittivity measurements. These probe specific values are written to the probe’s
firmware and minimize probe-to-probe variability.
8.3.4.2 Insertion Error
The method used for probe insertion can affect the accuracy of the measurement.
The probe rods should be kept as close to parallel as possible when inserted to
maintain the design wave guide geometry. The sensitivity of this measurement is
greater in the regions closest to the rod surface than at distances away from the
surface. Probes inserted in a manner that generates air voids around the rods will
indicate lower water content than actual. In some applications, installation can be
improved by using insertion guides or a pilot tool. Campbell Scientific offers the
CS650G insertion tool.
8.3.5 Temperature Dependence and Correction
The two temperature dependent sources of error in CS650 water content
measurements are the effect of temperature on the operation of the probe
electronics and the effect of temperature on the dielectric permittivity of the soil.
The effect of temperature on probe electronics is minimal with period average
readings varying by less than 0.5% of the 20 °C reading over the range of 10 to 30
°C and less than 2% of the 20 °C reading over the range of –10 to 70 °C.
The larger error is caused by the change in dielectric permittivity of soil with
temperature. This is mostly due to the high temperature dependence of the
permittivity of water, which varies from a value of 88 at 0 °C to 64 at 70 °C.
Since water is the major contributor to bulk dielectric permittivity of soil,
temperature related changes to the permittivity of water will lead to
overestimation of volumetric water content at temperatures below 20 °C and
underestimation of volumetric water content at temperatures above 20 °C.
The Topp equation does not account for soil temperature. The effect of
temperature on the soil permittivity is related to soil specific properties such as
porosity and the permittivity of the soil solid phase with temperature.
Consequently, a general equation that corrects volumetric water content for
temperature for all soils is not available.
A temperature correction equation that works well in quartz sand is given by:
is the temperature corrected volumetric water content, T is soil
Corr
temperature in °C, and is the volumetric water content value at soil temperature
T.
8.3.5.1 Accurate Soil Temperature Measurement
The thermistor used for measuring soil temperature is located in the probe head
and is in contact with one of the stainless steel rods. In order to make an accurate
soil temperature measurement, the probe head should be buried in the soil so that
it is insulated from diurnal temperature fluctuations.
8.4 Water Content Reflectometer User-Calibration
8.4.1 Need for Soil Specific Calibration Equation
While the Topp equation has been determined to work well in a wide range of
mineral soils, there are soils for which a user-derived calibration will optimize
accuracy of the volumetric water content measurement. The Topp equation
underestimates the water content of some organic, volcanic, and fine textured
soils. Additionally, porous media with porosity greater than 0.5 or bulk density
greater than 1.55 g cm–3 may require a media-specific calibration equation.
In these cases, the user may develop a calibration equation to convert CS650
permittivity to volumetric water content over the range of water contents the probe
is expected to measure.
8.4.2 User-Derived Calibration Equation
The relationship between soil permittivity and volumetric water content may be
described by a quadratic equation or a 3rd order polynomial. In many applications,
a linear equation similar to Ledieu et al (1986) gives required accuracy.
Quadratic form:
v(Ka) = C0 + C1*Ka + C2*K
with v the volumetric water content, Ka the bulk dielectric permittivity of the soil,
and Cn , the calibration coefficient.
3rd degree polynomial form:
v(Ka) = C0 + C1*Ka + C2*K
with v the volumetric water content, Ka the bulk dielectric permittivity of the soil,
and Cn , the calibration coefficient.
Linear form:
with v the volumetric water content, Ka the bulk dielectric permittivity of the soil,
and Cn , the calibration coefficient.
v(Ka) = C0 + C1*K
2
+ C3*K
a
0.5
a
2
a
3
a
Two data points from careful measurements can be enough to derive a linear
calibration. A minimum of three data points are needed for a quadratic
calibration. With three evenly spaced water contents covering the expected range,
the middle water content data point will indicate whether a linear or polynomial
calibration equation is needed.
23
CS650 and CS655 Water Content Reflectometers
A minimum of four data points are required for derivation of a 3rd degree
polynomial. Data points should be spaced as evenly as practical over the expected
range of water content and include the wettest and driest expected values.
8.4.3 Collecting Laboratory Data for Calibration
Water content reflectometer data needed for CS650 calibration are the CS650
permittivity reading and an independently determined volumetric water content.
From this data, the probe response to changing water content can be described by
a linear or polynomial function as described in Section 8.4.2, User-Derived Calibration Equation(p. 23).
Required equipment:
CS650 connected to datalogger programmed to measure permittivity
Cylindrical sampling devices to determine sample volume for bulk
density, such as copper tubing with diameter of 2.5 cm (1 in) and
length at about 5 cm (2 ins).
Containers and scale to measure soil sample mass
Oven to dry samples (microwave oven can also be used)
The calibration coefficients are derived from a curve fit of known water content
and probe permittivity output. The number of data sets needed to derive a
calibration depends on the form of the calibration equation. At least three data
sets should be generated to determine whether the linear form is valid. If a
polynomial is to be used, four data sets will determine whether the function is a
quadratic or third order polynomial. Accuracy requirements may require
additional data sets. Consider the expected range of soil water content and include
data sets from the highest and lowest expected water contents.
The measurement sensitive volume around the probe rods must be completely
occupied by the calibration soil. Only soil should be in the region within
10 cm (4 in) of the rod surface. The probe rods can be buried in a tray of soil that
is dry or nearly dry. The soil will be homogeneous around the probe rods if it is
poured around the rods while dry. Also, a 20 cm diameter PVC pipe with length
about 35 cm can be closed at one end and used as the container.
It is important that the bulk density of the soil used for calibration be similar to the
bulk density of the undisturbed soil. Using dry soil without compaction will give
a typical bulk density, 1.1 to 1.4 g cm–3. This is especially important when bulk
density is greater than 1.55 g cm–3. Compaction of the calibration soil to similar
bulk density at the field site is necessary for an accurate calibration.
The typically used method for packing a container of soil to uniform bulk density
is to roughly separate the soil into three or more equal portions and add one
portion to the container with compaction. Evenly place the first loose soil layer in
the bottom of the container. Compact by tamping the surface to a level in the
container that is correct for the target bulk density. Repeat for the remaining
layers. Prior to placing successive layers, scarify (loosen) the top of the existing
compacted layer.
24
The container to hold the soil during calibration should be non-metal and large
enough that the rods of the probe are no closer than about 10 cm from any
container surface.
Pack the container as uniformly as possible in bulk density with relatively dry soil
(volumetric water content <10%).
User Manual
g
wetdry
dry
mm
m
bulk
dry
cylinder
m
volume
Probe rods can be buried in a tray or inserted into a column. When using a
column, insert the rods carefully through surface until rods are completely
surrounded by soil. Movement of rods from side-to-side during insertion can form
air voids around rod surface and lead to measurement error.
Collect the probe permittivity output. Repeat previous step and this step three or
four times.
Determine volumetric water content by subsampling soil column after removing
probe or using mass of column. If subsampling is used, remove soil from column
and remix with samples used for water content measurement. Repack column.
Water can then be added to the top of the container. It must be allowed to
equilibrate. Cover the container during equilibration to prevent evaporation. The
time required for equilibration depends on the amount of water added and the
hydraulic properties of the soil. Equilibration can be verified by frequently
observing the CS650 permittivity output. When permittivity is constant,
equilibration is achieved. Collect a set of calibration data values and repeat the
water addition procedure again if needed.
With soil at equilibrium, record the CS650 permittivity.
Take subsamples of the soil using containers of known volume. This is necessary
for measurement of bulk density. Copper tubing of diameter 2.5 cm (1 in) and
length about 5 cm (2 ins) works well. The tubes can be pressed into the soil
surface.
It is good to take replicate samples. Three carefully handled samples will provide
good results.
The sample tubes should be pushed evenly into the soil. Remove the tube and
sample and gently trim the ends of excess soil. Remove excess soil from outside
of tube.
Remove all the soil from tube to a tray or container of known mass that can be put
in oven or microwave. Weigh and record the wet soil mass.
Water is removed from the sample by heating with oven or microwave. Oven
drying requires 24 hours at 105 °C. Microwave drying typically takes 20 minutes
depending on microwave power and sample water content. ASTM Method
D4643-93 requires heating in microwave for 3 minutes, cooling in desiccator then
weighing and repeating this process until measured mass is constant.
Gravimetric water content is calculated after the container mass is accounted for.
For the bulk density
the dry mass of the sample is divided by the sample tube volume.
The volumetric water content is the product of the gravimetric water content and
the bulk density
25
CS650 and CS655 Water Content Reflectometers
vgbulk
*
The average water content for the replicates and the recorded CS650 permittivity
are one datum pair to be used for the calibration curve fit.
8.4.4 Collecting Field Data for Calibration
Required equipment:
CS650 connected to datalogger programmed to measure probe
permittivity
Cylindrical sampling devices to determine sample volume for bulk
density, such as copper tubing with diameter of 2.5 cm (1 in) and
length about 5 cm (2 ins).
Containers and scale to measure soil sample mass
Oven to dry samples (microwave oven can also be used)
Data needed for CS650 calibration are the CS650 permittivity output and an
independently determined volumetric water content. From this data, the probe
response to changing water content can be described by a function as described in
Section 8.4.2, User-Derived Calibration Equation(p. 23).
The calibration coefficients are derived from a curve fit of known water content
and probe permittivity output. The number of data sets needed to derive a
calibration depends on the form of the calibration equation. At least three data
sets should be generated to determine whether the linear form is valid. If a
polynomial is to be used, four data sets will determine whether the function is a
quadratic or third order polynomial. Accuracy requirements may require
additional data sets. Consider the expected range of soil water content and include
data sets from the highest and lowest expected water contents.
Collecting measurements of CS650 permittivity and core samples from the
location where the probe is to be used will provide the best on-site soil-specific
calibration. However, intentionally changing water content in soil profiles can be
difficult.
A vertical face of soil can be formed with a shovel. If the CS650 is to be used
within about 0.5 metres of the surface, the probe can be inserted into the face and
water added to the surface with percolation. After adding water, monitor the
CS650 permittivity to determine if the soil around the rods is at equilibrium.
With soil at equilibrium, record the CS650 permittivity.
Soil hydraulic properties are spatially variable. Obtaining measurements that are
representative of the soil on a large scale requires multiple readings and sampling.
The average of several core samples should be used to calculate volumetric water
content. Likewise, the CS650 should be inserted at least 3 times into the soil
recording the permittivitys following each insertion and using the average.
Remove the CS650 and take core samples of the soil where the probe rods were
inserted. This is necessary for measurement of bulk density. Copper tubing of
diameter 2.5 cm (1 in) and length about 5 cm (2 ins) works well. The tubes can
be pressed into the soil surface.
It is good to take replicate samples at locations around the soil surface. Three
carefully handled samples will provide good results.
26
User Manual
g
wetdry
dry
mm
m
bulk
dry
cylinder
m
volume
vgbulk
*
volume
d
h
**
2
2
The sample tubes should be pushed evenly into the soil surface. Remove the tube
and sample and gently trim the ends of excess soil. Remove excess soil from
outside of tube.
Remove all the soil from tube to a tray or container of known mass that can be put
in oven or microwave. Weigh and record the wet soil mass. If samples must be
stored prior to weighing, seal the container with tape or inside a plastic bag to
prevent water loss and store away from direct sunlight.
Water is removed from the sample by heating with oven or microwave. Oven
drying requires 24 hours at 105 °C. Microwave drying typically takes 20 minutes
depending on microwave power and sample water content. ASTM Method
D4643-93 requires heating in microwave for 3 minutes, cooling in desiccator then
weighing and repeating this process until mass is constant.
Gravimetric water content is calculated after the container mass is accounted for.
For the bulk density,
8.4.5 Calculations
the dry mass of the sample is divided by the sample tube volume.
The volumetric water content is the product of the gravimetric water content and
the bulk density
The average water content for the replicates and the recorded CS650 period are
one datum pair to be used for the calibration curve fit.
The empty cylinders used for core sampling should be clean and both empty mass
and volume are measured and recorded. For a cylinder, the volume is
where d is the inside diameter of the cylinder and h is the height of the cylinder.
During soil sampling it is important that the cores be completely filled with soil
but not extend beyond the ends of the cylinder.
Once soil core samples are obtained, place the soil-filled cylinder in a small tray
of known empty mass. This tray will hold the core sample during drying in an
oven.
To obtain m
, subtract the cylinder empty mass and the container empty mass
wet
from the mass of the soil filled cylinder in the tray. Remove all the soil from the
cylinder and place this soil in the tray. Dry the samples using oven or microwave
methods as described above.
27
CS650 and CS655 Water Content Reflectometers
g
wetdry
dry
mm
m
bulk
dry
cylinder
m
volume
vgbulk
*
To obtain m
for m
dry
, weigh the tray containing the soil after drying. Subtract tray mass
dry
. Calculate gravimetric water content, g, using
To obtain soil bulk density, use
Volumetric water content is calculated using
28
9. Maintenance and Troubleshooting
Table 9-1. Symptom, Cause, and Solutions
Symptom
Possible Cause
Solution
All CS650 output
values read 0
No SDI12Recorder
instruction in datalogger
program
Conditional statement
that triggers reading is
not evaluating as true
Add SDI12Recorder
instruction to datalogger
program
Check logic of conditional
statement that triggers
readings
First value reads NAN
and all other values read
0* or never change
from one measurement
to another
CS650 SDI-12 address
does not match address
specified in datalogger
program
Change probe address or
modify program so that
they match
(*or all values read
NAN if the program
examples in this manual
are followed)
CS650 green wire not
attached to SDI port
specified in datalogger
program
Connect wire to correct
control port or modify
program to match wiring
CS650 not being
powered
Make sure red wire is
connected to 12V or
SW12V and black wire to
G.
If using SW12 to power
sensor, make sure red wire
is connected and datalogger
program switches SW12 on.
VWC reading is
9999999
Soil bulk permittivity is
outside probe’s
operational range
Modify program to collect
permittivity value and try
soil specific calibration
EC reading is 9999999
Soil bulk electrical
conductivity is outside
probe’s operational
range
If using CS650, try CS655
Readings erratic,
including NAN’s and
9999999’s
Multiple probes with
same SDI-12 address
sharing same control
port
Give probes unique
addresses or put on separate
control ports
The CS650 does not require periodic maintenance. Table 9-1 provides
troubleshooting information.
User Manual
29
CS650 and CS655 Water Content Reflectometers
10. References
Ledieu, J., P. De Ridder, P. De Clerck, and S. Dautrebande. 1986. “A method of
measuring soil moisture by time-domain reflectometry,” J. Hydrol. 88:319-
328.
Rhoades, J.D., P.A.C. Raats, and R.J. Prather. 1976. Effects of liquid-phase
electrical conductivity, water content and surface conductivity on bulk soil
electrical conductivity. Soil Sci. Soc. Am. J., 40: 651-653.
conductivity and soil salinity: New formulations and calibrations. Soil Sci.
Soc. Am. J., 53:433-439.
Topp, G.C., J.L. Davis & A.P. Annan. 1980. “Electromagnetic determination of
soil water content: measurements in coaxial transmission lines,” Water
Resources Research, v. 16, No. 3:574-582.
30
NOTE
Appendix A. Importing Short Cut Code
into a Program Editor
This tutorial shows:
How to import a Short Cut program into a program editor for additional
refinement
How to import a wiring diagram from Short Cut into the comments of a
custom program
Short Cut creates files that can be imported into either CRBasic Editor program
editor. These files normally reside in the C:\campbellsci\SCWin folder and have
the following extensions:
Use the following procedure to import Short Cut code into CRBasic Editor
(CR200(X), CR1000, CR800, CR3000, CR5000 dataloggers).
1. Create the Short Cut program following the procedure in Section 4,
Quickstart(p. 2). Finish the program and exit Short Cut. Make note of the file
name used when saving the Short Cut program.
2. Open CRBasic Editor.
3. Click File | Open. Assuming the default paths were used when Short Cut was
installed, navigate to C:\CampbellSci\SCWin folder. The file of interest has a
“.CR6”, “.CR2”, “.CR1”, “.CR8”, “.CR3, or “.CR5” extension, for CR6,
CR200(X), CR1000, CR800, CR3000, or CR5000 dataloggers, respectively.
Select the file and click Open.
4. Immediately save the file in a folder different from \Campbellsci\SCWin, or
save the file with a different file name.
Once the file is edited with CRBasic Editor, Short Cut can no longer
be used to edit the datalogger program. Change the name of the
program file or move it, or Short Cut may overwrite it next time it is
used.
5. The program can now be edited, saved, and sent to the datalogger.
6. Import wiring information to the program by opening the associated .DEF
file. Copy and paste the section beginning with heading “-Wiring for
CRXXX–” into the CRBasic program, usually at the head of the file. After
pasting, edit the information such that a ' character (single quotation mark)
begins each line. This character instructs the datalogger compiler to ignore
the line when compiling the datalogger code.
A-1
Appendix A. Importing Short Cut Code
A-2
Table B-1. CR1000 Wiring for One Probe Example Program
CR1000
CS650
12V
Red
C1
Green
G
Black, Orange, Clear
Table B-2. CR1000 Wiring for Two Probe Example Program
CR1000
CS650’s (wiring same for both)
12V
Red
C1
Green
G
Black, Orange, Clear
Appendix B. Example Programs
B.1 CR1000 Programs
B.1.1 CR1000 with One CS650 Probe
This CRBasic example program measures one CS650 probe on a CR1000 every
15 minutes, storing hourly averages of volumetric water content, electrical
conductivity, and soil temperature and samples of permittivity, period average and
voltage ratio. The CS650 has an SDI-12 address of 0. Wiring for the example is
shown in Table B-1.
Public CS650(6)
'Assign aliases to the public array
Alias CS650(1)=VWC: Alias CS650(2)=EC: Alias CS650(3)=TSoil
Alias CS650(4)=Perm: Alias CS650(5)=PerAvg: Alias CS650(6)=VoltR
Units VWC = m^3/m^3: Units EC = dS/m: Units TSoil = deg C
DataTable (DatoutCS650,1,-1)
DataInterval (0,60,Min,2)
Average (3,CS650(1),FP2,False)
Sample(3,CS650(4),IEEE4)
EndTable
BeginProg
Scan (15,Min,0,0)
SDI12Recorder (CS650(1),1,0,"M3!",1.0,0)
CallTable DatoutCS650 'Call Data Table
NextScan
EndProg
B.1.2 CR1000 with Two CS650 Probes on Same Control Port
This CRBasic example program measures two CS650 probes on a CR1000 every
15 minutes, storing hourly averages of volumetric water content, electrical
conductivity, and soil temperature and samples of permittivity, period average and
voltage ratio. The first CS650 has an SDI-12 address of 0 and the second has an
address of 1. Wiring for the example is shown in Table B-2. Assignment of
aliases and units is not shown in this example.
B-1
Appendix B. Example Programs
Table B-3. CR1000 Wiring For Multiplexer Example Program
CR1000
AM16/32B (2x32 mode)
CS650
12V
12V
G GND
C2
RES
C3
CLK
SW12
COM ODD H
C1
COM ODD L
G
COM Ground
High Channels 1H – 12H
Red
Low Channels 1L – 12L
Green
Ground Channels to Left of
Low Channels
Black, Orange, Clear
Public CS650(6)
Public CS650_2(6)
DataTable (DatoutCS650,1,-1)
DataInterval (0,60,Min,2)
Average (3,CS650(1),FP2,False)
Sample(3,CS650(4),IEEE4)
Average (3,CS650_2(1),FP2,False)
Sample(3,CS650_2(4),IEEE4)
EndTable
BeginProg
Scan (15,Min,0,0)
SDI12Recorder (CS650(1),1,0,"M3!",1.0,0)
SDI12Recorder (CS650_2(1),1,1,"M3!",1.0,0)
CallTable DatoutCS650 'Call Data Table
NextScan
EndProg
B.1.3 CR1000 with 12 CS650 Probes on Multiplexer
This CRBasic example program measures 12 CS650 probes on a AM16/32B
multiplexer every 15 minutes, storing hourly averages of volumetric water
content, electrical conductivity, soil temperature, permittivity, period average, and
voltage ratio. All probes are addressed with SDI-12 address of 0. In this example,
the probes are powered through the switched 12V channel and require 3 seconds
warm-up time per probe. Total time to measure all 12 probes is more than 36
seconds. Alternately, all of the red wires for the probes could be connected to a
bus separate from the multiplexer with the bus connected to 12V for continuous
power. This would decrease measurement time. Wiring for the example is shown
in Table B-3. Assignment of aliases and units is not shown in this example.
B-2
Appendix B. Example Programs
Table B-4. CR200(X) Wiring for Example Program
CR200(X)
CS650’s (Wiring same for all)
SW Battery
Red
C1/SDI-12
Green
G channels
Black, Orange, Clear
Dim LCount
Public CS650(12,6)
DataTable (DatoutCS650,1,-1)
DataInterval (0,60,Min,2)
Average (72,CS650(),IEEE4,False)
EndTable
BeginProg
Scan (15,Min,0,0)
PortSet(2,1) 'Turn AM16/32 Multiplexer On
Delay(0,150,mSec)
LCount=1
SubScan(0,uSec,12)
PulsePort(3,10000) 'Switch to next AM16/32 channel
SW12 (1 ) 'Apply power to CS650
Delay (0,3,Sec) 'Wait three seconds for probe to warm up
SDI12Recorder (CS650(LCount,1),1,0,"M!",1.0,0)
LCount=LCount+1
SW12 (0) 'Remove power from CS650
NextSubScan
PortSet(2,0) 'Turn AM16/32 Multiplexer Off
Delay(0,150,mSec)
CallTable DatoutCS650 'Call Data Table
NextScan
EndProg
B.2 CR200X with Three CS650 Probes
This CRBasic example program measures three CS650 probe on a CR200X every
15 minutes, storing hourly averages of volumetric water content, electrical
conductivity, soil temperature, permittivity, period average, and voltage ratio. The
CS650s have SDI-12 addresses of 0, 1, and 2, respectively. Sensors are powered
with the SWBatt channel, which requires a 3 second warm-up time. Alternately,
the red wires may be connected to Battery + for continuous power which would
reduce measurement time. Wiring for the example is shown in Table B-4.
Assignment of aliases and units is not shown in this example.
Public CS650(18)
DataTable (CS650,1,-1)
DataInterval (0,60,Min)
Average (18,CS650(),False)
EndTable
BeginProg
Scan (15,Min)
SWBatt (1 ) 'Apply power to CS650's
Delay (3,sec) 'Warm-up time of 3 seconds
B-3
Appendix B. Example Programs
'CS650 #1
SDI12Recorder (CS650(1),"0M3!",1,0)
'CS650 #2
SDI12Recorder (CS650(7),"1M3!",1,0)
'CS650 #3
SDI12Recorder (CS650(13),"2M3!",1,0)
SWBatt (0 ) 'Remove power from CS650's
CallTable CS650 'Call Data Table
NextScan
EndProg
B-4
g
water
soil
wetdry
dry
m
m
mm
m
v
water
soil
water
water
soil
soil
gsoil
water
volume
volume
m
m
*
bulk
dry
sample
m
volume
1
bulk
solid
Appendix C. Discussion of Soil Water
Content
The water content reflectometer measures volumetric water content. Soil water
content is expressed on a gravimetric and a volumetric basis. To obtain the
independently determined volumetric water content, gravimetric water content
must first be measured. Gravimetric water content (g) is the mass of water per
mass of dry soil. It is measured by weighing a soil sample (m
sample to remove the water, then weighing the dried soil (m
Volumetric water content (v) is the volume of liquid water per volume of soil.
Volume is the ratio of mass to density (b) which gives:
), drying the
wet
).
dry
The density of water is close to 1 and often ignored.
Soil bulk density (
) is used for
bulk
and is the ratio of soil dry mass to sample
soil
volume.
Another useful property, soil porosity (), is related to soil bulk density as shown
by the following expression.
The term
is the density of the soil solid fraction and is approximately 2.65 g
solid
cm–3.
C-1
Discussion of Soil Water Content
C-2
Table D-1. CS650 SDI-12 Command and Response Set
Name
Command
Response
Acknowledge
Active
a!
a<CR><LF>
Send Identification
aI!
allccccccccmmmmmmvvvxxx...xx<CR>
<LF>
Change Address
aAb!
b<CR><LF>
Address Query
?!
a<CR><LF>
Start Measurement
aM!
atttn<CR><LF>
Send Data
aD0!
aD1!
a<values><CR><LF>
a<values><CR><LF>
Additional
Measurements
aM1!
aM2!
aM3!
atttn<CR><LF>
atttn<CR><LF>
atttn<CR><LF>
Appendix D. SDI-12 Sensor Support
D.1 SDI-12 Command Basics
SDI-12 commands have three components:
Sensor address (a) – a single character, and is the first character of the command.
CS650 sensors are usually assigned a default address of zero unless option –VS is
selected at the time of ordering. Sensors with the –VS option are addressed with
the last digit of the probe’s serial number. This allows for multiple CS650’s to be
connected to a single control port without requiring the user to change the SDI-12
addresses from zero.
Command body (e.g., M1) – an upper case letter (the “command”) followed by
alphanumeric qualifiers.
Command termination (!) – an exclamation mark.
An active sensor responds to each command. Responses have several standard
forms and terminate with <CR><LF> (carriage return – line feed).
SDI-12 commands supported by the CS650 are listed in Table D-1. Continuous
and concurrent measurements are not supported.
Address Query Command (?!)
Change Address Command (aAb!)
Command ?! requests the address of the connected sensor. The sensor replies to
the query with the address, a.
Sensor address is changed with command aAb!, where a is the current address and
b is the new address. For example, to change an address from 0 to 2, the
command is 0A2!. The sensor responds with the new address b, which in this case
is 2.
D-1
Appendix D. SDI-12 Sensor Support
Send Identification Command (aI!)
Sensor identifiers are requested by issuing command aI!. The reply is defined by
the sensor manufacturer, but usually includes the sensor address, SDI-12 version,
manufacturer’s name, and sensor model information. Serial number or other
sensor specific information may also be included.
An example of a response from the aI! command is:
313CampbellCS65X 000Std.00.35=2196405 <CR><LF>
where:
Address = 3
SDI-12 version =1.3
Manufacturer = Campbell
Sensor model = CS65X
OS version = 000Std.00.35
Sensor serial number = 2196405
Start Measurement Commands (aM!)
A measurement is initiated with M! commands. The response to each command
has the form atttnn, where
a = sensor address
ttt = time, in seconds, until measurement data are available
nn = the number of values to be returned when one or more subsequent D!
commands are issued.
Start Measurement Command (aMv!)
Qualifier v is a variable between 1 and 3 that requests variant data. Variants
include different subsets of the CS650 probe output:
M0! Volumetric Water Content (Bulk Electrical Conductivity
M2! Period (Voltage Ratio (Temperature (°C)
M3! Volumetric Water Content (), Bulk Electrical Conductivity
(Temperature (°C), Permittivity (Period (Voltage Ratio (
D-2
Aborting a Measurement Command
A measurement command (M!) is aborted when any other valid command is sent
to the sensor.
Send Data Command (aDv!)
This command requests data from the sensor. It is normally issued automatically
by the datalogger after measurement commands aMv!. In transparent mode, the
user asserts this command to obtain data. If the expected number of data values
are not returned in response to an aD0! command, the data logger issues aD1!.
The limiting constraint is that the total number of characters that can be returned
Appendix D. SDI-12 Sensor Support
to an aD0! command is 35 characters. If the number of characters exceed the
limit, the remainder of the response are obtained with the subsequent aD1!
command.
D.2 Changing the SDI-12 Address Using Terminal
Emulator and a Datalogger
Up to ten CS650’s or other SDI-12 sensors can be connected to a single datalogger
control port. Each SDI-12 device on the same control port must have a unique
SDI-12 address. The CS650 supports addresses of 0-9, a-z, and A-Z.
The factory-set SDI-12 address for the CS650 is 0 when the probe is ordered with
the –DS option or the last digit of its serial number when ordered with the –VS
option. The CS650 SDI-12 address is changed by issuing the aAb! command
where a is the current address and b is the new address. The current address can
be found by issuing the ?! command.
The easiest way to change the address on a CS650 sensor is with DevConfig and
an A200 Sensor to PC Interface as described in Section 8.1, A200 and Device Configuration Utility(p. 12). However if an A200 is not available, it is possible to
change the address by connecting a single CS650 to an SDI-12 compatible control
port on a datalogger and utilizing SDI-12 transparent mode to send commands
directly to the sensor.
D.2.1 SDI-12 Transparent Mode
System operators can manually interrogate and enter settings in probes using
transparent mode. Transparent mode is useful in troubleshooting SDI-12 systems
because it allows direct communication with probes. Datalogger security may
need to be unlocked before transparent mode can be activated.
Transparent mode is entered while the PC is in telecommunications with the
datalogger through a terminal emulator program. It is easily accessed through
Campbell Scientific datalogger support software, but is also accessible with
terminal emulator programs such as Windows HyperTerminal. Datalogger
keyboards and displays cannot be used.
The terminal emulator is accessed by navigating to the Datalogger menu in
PC200W, the Tools menu in PC400, or the Datalogger menu in the Connect
screen of LoggerNet.
The following examples show how to use LoggerNet software to enter transparent
mode and change the SDI-12 address of a CS650 sensor. The same steps are used
to enter transparent mode with PC200W and PC400 software after accessing the
terminal emulator as previously described.
D-3
Appendix D. SDI-12 Sensor Support
D.2.2 CR200(X) Series Datalogger Example
1. Connect a single CS650 to the datalogger as follows:
2. In the LoggerNet Connect screen navigate to the Datalogger menu and select
3. Click on the Open Terminal button.
4. Press the <enter> key until the datalogger responds with the “CR2XX>”
5. To query the CS650 for its current SDI-12 address, key in ?! <enter> and the
Green to Control Port C1/SDI12
Black, Orange, Clear to G
Red to Battery +
Terminal Emulator. The “Terminal Emulator” window will open. In the
Select Device menu, located in the lower left-hand side of the window, select
the CR200Series station.
prompt. At the “CR2XX>” prompt, make sure the All Caps Mode box is
checked and enter the command SDI12 <enter>. The response “SDI12>”
indicates that the CS650 is ready to accept SDI-12 commands.
CS650 will respond with its SDI-12 address. If no characters are typed within
60 seconds, then the mode is exited. In that case, simply enter the command
SDI12 again and press <enter>.
D-4
Figure D-1. SDI-12 transparent mode on CR200(X)-series datalogger using
control port C1/SDI12 and changing SDI-12 address from 0 to 1
6. To change the SDI-12 address, key in aAb!<enter> where a is the current
address from the above step and b is the new address. The CS650 will change
its address and the datalogger will respond with the new address. To exit
SDI-12 transparent mode select the Close Terminal button.
D.2.3 CR1000 Datalogger Example
1. Connect a single CS650 to the datalogger as follows:
Green to Control Port C1
Black, Orange, Clear to G
Red to 12V
2. In the LoggerNet Connect screen navigate to the Datalogger menu and select
Terminal Emulator. The “Terminal Emulator” window will open. In the
Select Device menu, located in the lower left-hand side of the window, select
the CR1000 station.
3. Click on the Open Terminal button.
4. Press the <enter> key until the datalogger responds with the “CR1000>”
prompt. At the “CR1000>” prompt, make sure the All Caps Mode box is
checked and enter the command SDI12 <enter>. At the “Enter Cx Port 1, 3,
5, or 7” prompt, key in the control port number where the CS650 green lead is
connected and <enter>. The response “Entering SDI12 Terminal” indicates
that the CS650 is ready to accept SDI-12 commands.
Appendix D. SDI-12 Sensor Support
5. To query the CS650 for its current SDI-12 address, key in ?! <enter> and the
CS650 will respond with its SDI-12 address. If no characters are typed within
60 seconds, then the mode is exited. In that case, simply enter the command
SDI12 again, press <enter>, and key in the correct control port number when
prompted.
Figure D-2. SDI-12 transparent mode on CR1000 datalogger using control
port 1 and changing SD1-12 address from 3 to 1
6. To change the SDI-12 address, key in aAb!<enter> where a is the current
address from the above step and b is the new address. The CS650 will change
its address and the datalogger will respond with the new address. To exit
SDI-12 transparent mode, press the Esc key or wait for the 60 second timeout,
then select the Close Terminal button.
D-5
CAMPBELL SCIENTIFIC COMPANIES
Campbell Scientific, Inc. (CSI)
815 West 1800 North
Logan, Utah 84321
UNITED STATES
www.campbellsci.cominfo@campbellsci.com
Campbell Scientific Africa Pty. Ltd. (CSAf)
PO Box 2450
Somerset West 7129
SOUTH AFRICA
www.csafrica.co.zasales@csafrica.co.za
Campbell Scientific Australia Pty. Ltd. (CSA)
PO Box 8108
Garbutt Post Shop
QLD 4814 AUSTRALIA
www.campbellsci.com.auinfo@campbellsci.com.au
Campbell Scientific do Brazil Ltda. (CSB)
Rua Apinagés, nbr. 2018 - Perdizes
CEP: 01258-00 São Paulo SP BRAZIL
www.campbellsci.com.brvendas@campbellsci.com.br
Campbell Scientific Canada Corp. (CSC)
14532 – 131 Avenue NW
Edmonton, Alberta T5L 4X4
CANADA
www.campbellsci.cadataloggers@campbellsci.ca
Campbell Scientific Centro Caribe S.A. (CSCC)
300N Cementerio, Edificio Breller
Santo Domingo, Heredia 40305
COSTA RICA
www.campbellsci.cc info@campbellsci.cc
Campbell Scientific Ltd. (CSL)
80 Hathern Road, Shepshed, Loughborough LE12 9GX
UNITED KINGDOM
www.campbellsci.co.uk sales@campbellsci.co.uk
Campbell Scientific Ltd. (France)
3 Avenue de la Division Leclerc
92160 ANTONY
FRANCE
www.campbellsci.frinfo@campbellsci.fr
Campbell Scientific Spain, S. L.
Avda. Pompeu Fabra 7-9
Local 1 - 08024 BARCELONA
SPAIN
www.campbellsci.esinfo@campbellsci.es
Campbell Scientific Ltd. (Germany)
Fahrenheitstrasse13, D-28359 Bremen
GERMANY
www.campbellsci.deinfo@campbellsci.de
Campbell Scientific (Beijing) Co., Ltd.
8B16, Floor 8 Tower B, Hanwei Plaza
7 Guanghua Road, Chaoyang, Beijing 100004
P.R. CHINA
www.campbellsci.cominfo@campbellsci.com.cn
Please visit www.campbellsci.com to obtain contact information for your local US or International representative.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.