Campbell CD100 Instruction Manual

INSTRUCTION MANUAL
CD100 Mountable Display
with Keypad
Revision: 1/17
Guarantee
This equipment is guaranteed against defects in materials and workmanship. We will repair or replace products which prove to be defective during the guarantee period as detailed on your invoice, provided they are returned to us prepaid. The guarantee will not apply to:
Equipment which has been modified or altered in any way without the written permission of Campbell Scientific
Batteries
Any product which has been subjected to misuse, neglect, acts of God or damage in transit.
Campbell Scientific will return guaranteed equipment by surface carrier prepaid. Campbell Scientific will not reimburse the claimant for costs incurred
in removing and/or reinstalling equipment. This guarantee and the Company’s
obligation thereunder is in lieu of all other guarantees, expressed or implied, including those of suitability and fitness for a particular purpose. Campbell Scientific is not liable for consequential damage.
Please inform us before returning equipment and obtain a Repair Reference Number whether the repair is under guarantee or not. Please state the faults as clearly as possible, and if the product is out of the guarantee period it should be accompanied by a purchase order. Quotations for repairs can be given on request. It is the policy of Campbell Scientific to protect the health of its employees and provide a safe working environment, in support of this policy a
“Declaration of Hazardous Material and Decontamination” form will be
issued for completion.
When returning equipment, the Repair Reference Number must be clearly marked on the outside of the package. Complete the “Declaration of Hazardous Material and Decontaminationform and ensure a completed copy is returned with your goods. Please note your Repair may not be processed if you do not include a copy of this form and Campbell Scientific Ltd reserves the right to return goods at the customers’ expense.
Note that goods sent air freight are subject to Customs clearance fees which Campbell Scientific will charge to customers. In many cases, these charges are greater than the cost of the repair.
Campbell Scientific Ltd,
80 Hathern Road,
Shepshed, Loughborough, LE12 9GX, UK
Tel: +44 (0) 1509 601141
Fax: +44 (0) 1509 601091
Email: support@campbellsci.co.uk
www.campbellsci.co.uk
PLEASE READ FIRST
About this manual
Please note that this manual was originally produced by Campbell Scientific Inc. primarily for the North American market. Some spellings, weights and measures may reflect this origin.
Some useful conversion factors:
Area: 1 in2 (square inch) = 645 mm2
Length: 1 in. (inch) = 25.4 mm
1 ft (foot) = 304.8 mm 1 yard = 0.914 m 1 mile = 1.609 km
Mass: 1 oz. (ounce) = 28.35 g
1 lb (pound weight) = 0.454 kg
Pressure: 1 psi (lb/in2) = 68.95 mb
Volume: 1 UK pint = 568.3 ml
1 UK gallon = 4.546 litres 1 US gallon = 3.785 litres
In addition, while most of the information in the manual is correct for all countries, certain information is specific to the North American market and so may not be applicable to European users.
Differences include the U.S standard external power supply details where some information (for example the AC transformer input voltage) will not be applicable for British/European use. Please note,
however, that when a power supply adapter is ordered it will be suitable for use in your country.
Reference to some radio transmitters, digital cell phones and aerials may also not be applicable according to your locality.
Some brackets, shields and enclosure options, including wiring, are not sold as standard items in the European market; in some cases alternatives are offered. Details of the alternatives will be covered in separate manuals.
Part numbers prefixed with a “#” symbol are special order parts for use with non-EU variants or for special installations. Please quote the full part number with the # when ordering.
Recycling information
At the end of this product’s life it should not be put in commercial or domestic refuse but sent for recycling. Any batteries contained within the product or used during the products life should be removed from the product and also be sent to an appropriate recycling facility.
Campbell Scientific Ltd can advise on the recycling of the equipment and in some cases arrange collection and the correct disposal of it, although charges may apply for some items or territories.
For further advice or support, please contact Campbell Scientific Ltd, or your local agent.
Campbell Scientific Ltd, 80 Hathern Road, Shepshed, Loughborough, LE12 9GX, UK
Tel: +44 (0) 1509 601141 Fax: +44 (0) 1509 601091
Email: support@campbellsci.co.uk
www.campbellsci.co.uk
Precautions
DANGER — MANY HAZARDS ARE ASSOCIATED WITH INSTALLING, USING, MAINTAINING, AND WORKING ON OR AROUND TRIPODS, TOWERS, AND ANY ATTACHMENTS T O TRIP ODS AND T OWERS SUCH AS SENSORS, CROSSARMS, ENCLOSURES, ANTENNAS, ETC. FAILURE TO PROPERLY AND COM P LE TE LY AS S E MB LE , INSTALL, OPERATE, USE, AND MAINTAIN TRIPODS, TOWERS, AND ATTACHMENTS, AND F AI LURE TO HEED WARNINGS, INCREASES THE RISK OF DEATH, ACCIDENT, SERIOUS INJURY, PROPERTY DAMAGE, AND PRODUCT FAILURE. TAKE ALL REASONABLE PRECAUTIONS TO AVOID THESE HAZARDS. CHECK WITH YOUR ORGANIZATION'S SAFETY COORDINATOR (OR POLICY) FOR PROCE DURES AND REQUIRED PROTECTIVE EQUIPMENT PRIOR TO PERFORMING ANY WORK.
Use tripods, towers, and attachments to tripods and towers only for purposes for which they are designed. Do not exceed design limits. Be familiar and comply with all instructions provided in product manuals. Manuals are available at www.campbellsci.eu or by telephoning +44(0) 1509 828 888 (UK). You are responsible for conformance with govern i ng codes and regulations, including safety regul ations, and the integrity and locatio n of structures or land to which towers, tripods, and any attachments are attached. Installation sites should be evaluated and approved by a qualified engineer. If questions or concerns arise regarding installation, use, or maintenance of tripods, towers, attachments, or electrical connections, consult with a licensed and qualified engineer or electrician.
General
Prior to performing site or installation work, obtain required approvals and permits. Comply with all governing structure-height regulations, such as those of the FAA in the USA.
Use only qualified personnel for installation, use, and maintenance of tripods and towers, and any attachments to tripods and towers. The use of licensed and qualified contractors is highly recommended.
Read all applicable instructions carefully and understand procedures thoroughly before beginning work.
Wear a hardhat and eye protection, and take other appropriate safety precautions while working on or
around tripods and towers.
Do not climb tripods or towers at any time, and prohibit climbing by other persons. Take reasonable precautions to secure tripod and tower sites from trespassers.
Use only manufacturer recommended parts, materials, and tools.
Utility and Electrical
You can be killed or sustain serious bodily injury if the tripod, tower, or attachments you are installing, constructing, using, or maintaining, or a tool, stake, or anchor, come in contact with overhead o
r
u
nderground utility lines.
Maintain a distance of at least one-and-one-half times structure height, or 20 feet, or the distance r
equired by applicable law, whichever is greater, between overhead utility lines and the structure (tripod,
tower, attachments, or tools).
Prior to performing site or installation work, inform all utility companies and have all underground utilities marked.
Comply with all electrical codes. Electrical equipment and related grounding devices should be installed by a licensed and qualified electrician.
Elevated Work and Weather
Exercise extreme caution when performing elevated work.
Use appropriate equipment and safety practices.
During installation and maintenance, keep tower and tripod sites clear of un-trained or non-essential
personnel. Take precautions to prevent elevated tools and objects from dropping.
Do not perform any work in inclement weather, including wind, rain, snow, lightning, etc.
Maintenance
Periodically (at least yearly) check for wear and damage, including corrosion, stress cracks, frayed cables, loose cable clamps, cable tightness, etc. and take necessary corrective actions.
Periodically (at least yearly) check electrical ground connections.
WHILE EVERY ATTEMPT IS MADE TO EMBODY THE HIGHEST DEGREE OF SAFETY IN ALL CAMPBELL SCIENTIFIC PRODUCTS, THE CUSTOMER ASSUMES ALL RISK FROM ANY INJURY RESULTING FROM IMPROPER INSTALLATION, USE, OR MAINTENANCE OF TRIPODS, TOWERS, OR ATTACHMENTS TO TRIPODS AND TOWERS SUCH AS SENSORS, CROSSARMS, ENCLOSURES, ANTENNAS, ETC.
Table of Contents
PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to specific sections.
1. Introduction ................................................................ 1
2. Cautionary Statements .............................................. 1
3. Quickstart ................................................................... 2
4. Overview ..................................................................... 2
5. Specifications ............................................................. 3
6. Installation .................................................................. 3
7. Operation .................................................................... 7
8. Maintenance ............................................................... 8
Figures
3-1. Connecting the CD100 to the CR1000 datalogger ............................... 2
6-1. Template measurements for serial numbers less than 1311 ................. 4
6-2. Template measurements for serial numbers 1311 and higher .............. 5
6-3. Profile of installed CD100 ................................................................... 6
7-1. Display menu overview ....................................................................... 8
Tables
7-1. Keyboard Display Key Functions ........................................................ 7
i
CD100 Mountable Display with Keypad
1. Introduction
The CD100 is an easy-to-read panel-mountable datalogger data display. It includes a large-format keypad for input. The CD100 is compatible with several Campbell Scientific dataloggers. Refer to Section 5, Specifications for a complete list. When properly installed, the CD100 is weather and dust resistant. As a purchased service, Campbell Scientific offers the option of installing the CD100 into an enclosure.
2. Cautionary Statements
The CD100 is not compatible with the CR850 datalogger, nor should it be
used in tandem with the CR1000KD keyboard and display on any datalogger.
Weather resistance of the CD100 is dependent on proper installation into
the bulkhead of a weather tight enclosure. See Section 6, Installation for details of proper installation.
The CD100 is easily damaged if disassembled. If disassembly is required,
refer to Section 8, Maintenance
(p. 3),
(p. 3),
(p. 8).
1
CD100 Mountable Display with Keypad
3. Quickstart
1. If the CD100 is to be mounted into an enclosure bulkhead, refer to Section
6, Installation
2. Using an SC12 cable as shown in FIGURE 3-1, connect the 9-pin CS I/O
port of a CR1000 or other compatible datalogger to the 9-pin CS I/O port on the back of the CD100.
(p. 3), for instructions.
4. Overview
FIGURE 3-1. Connecting the CD100 to the CR1000 datalogger
3. Power the datalogger; the display should power up and be ready for use.
Keyboard commands are summarized in Section 7, Operation the datalogger manual.
The CD100 is a panel-mountable display with keypad. It facilitates user input and the display of data over a wide temperature range. The vacuum fluorescent element gives the CD100 high contrast, so it is easy to read in daylight conditions.
The CD100 is compatible with the dataloggers listed in Section 5, Specifications tandem with the CR1000KD on any datalogger.
When properly installed, the polyvinyl face and rear gasket make the CD100 weather resistant and prevent ingress of water and dust. The back of the CD100 is not weather resistant. If used outdoors, or in a condensing environment, the CD100 should be bulkhead mounted as recommended in Section 6, Installation
(p. 3).
(p. 3). It will not operate properly with the CR850 datalogger or in
(p. 7), and in
2
5. Specifications
NOTE
Features:
Compatibility: CR800 (not CR850)
Interface: CS I/O, SDC address 2
Temperature
Power Requirement
CD100 Mountable Display with Keypad
Operable with gloved hand
16 key alpha numeric keypad
8 line x 21 character display (64 x 128 pixel)
CR1000 CR3000
Operating: –45° to 85°C Storage: –50° to 85°C Transit: 1 hour at –60°C
Source: 12 Vdc from datalogger CS I/O port Current Draw
Inactive: 0.02 mA Active: 140 mA
6. Installation
Rating
Water and dust ingress
Installed as recommended: IP66 Uninstalled: None (vulnerable to water and dust)
Weight:
Dimensions:
Compliance:
As a purchased service, Campbell Scientific offers the option of installing the CD100 into an enclosure.
Install only on a flat, non-porous surface. The supplied gasket must create a weather-tight seal when installed or permanent damage to the display and other enclosure contents may occur.
1. Use the template shipped with the CD100 or create a template using th easurements shown in FIGURE 6-1, Template measurements for serial
m numbers less than 1311.
6 kg (0.80 lb)
0.3
17.5 x 13.7 x 3.6 cm (6.9 x 5.4 x 1.4 in)
View the EU Declaration of Conformity at
www.campbellsci.eu/cd100
e
2. Make the required cuts in the bulkhead.
3. Remove cutting debris from bulkhead.
3
CD100 Mountable Display with Keypad
Edge of
Cut
Edge of
CD100
Image Not Actual Size
Measurement Units = Inches
4. Mount the CD100 as shown in FIGURE 6-3, Profile of installed CD100.
5. Before tightening screws, adjust the gasket to align it just inside the edges of the display. Uniformly tighten the screws, but do not over-tighten.
F
IGURE 6-1. Template measurements for serial numbers less than
4
1311
CD100 Mountable Display with Keypad
F
IGURE 6-2. Template measurements for serial numbers 1311 and
higher
5
CD100 Mountable Display with Keypad
F
IGURE 6-3. Profile of installed CD100
6
7. Operation
TABLE 7-1. Keyboard Display Key Functions
CD100 Mountable Display with Keypad
An overview of keyboard functions is presented in TABLE 7-1, Keyboard Display Key Functions, and FIGURE 7-1, Display Menu Overview. Detailed
descriptions are available in the datalogger manual.
Key Special Function
[2] and [8] To navigate up and down through the menu list one line at a time
[Enter] Selects the line or toggles the option of the line the cursor is on
[Esc] Back up one level in the menu
[Home] Move cursor to top of the list
[End] Move cursor to bottom of the list
[Pg Up] Move cursor up one screen
[Pg Dn] Move cursor down one screen
[BkSpc] Delete character to the left
[Shift] Change alpha character selected
[Num Lock] Change to numeric entry
[Del] Delete
[Ins] Insert/change graph setup
[Graph] Graph
7
CD100 Mountable Display with Keypad
F
IGURE 7-1. Display menu overview
8. Maintenance
Active desiccant must be maintained in the datalogger enclosure to avoid water damage to the display electronics, the datalogger, and other contents. No other maintenance is required.
Depending on the environment and weather tightness of the installation, dust may accumulate between the surface window and the underlying display elements. If this occurs, return the display to Campbell Scientific for cleaning.
Do not disassemble the CD100. The printed circuit ribbon connecting the keypad to the display can be easily torn when not removed properly. If the display must be disassembled, take care to grasp the header and not the printed circuit itself. Sheer or torsional stresses must not be put on the printed circuit. The warranty does not cover damages caused by disassembly.
8
Campbell Scientific Companies
Campbell Scientific, Inc.
815 West 1800 North
Logan, Utah 84321
UNITED STATES
www.campbellsci.com info@campbellsci.com
Campbell Scientific Africa Pty. Ltd.
PO Box 2450
Somerset West 7129
SOUTH AFRICA
www.campbellsci.co.za cleroux@csafrica.co.za
Campbell Scientific Southeast Asia Co., Ltd.
877/22 Nirvana@Work, Rama 9 Road
Suan Luang Subdistrict, Suan Luang District
Bangkok 10250
THAILAND
www.campbellsci.asia info@campbellsci.asia
Campbell Scientific Australia Pty. Ltd.
PO Box 8108
Garbutt Post Shop QLD 4814
AUSTRALIA
www.campbellsci.com.au info@campbellsci.com.au
Campbell Scientific (Beijing) Co., Ltd.
8B16, Floor 8 Tower B, Hanwei Plaza
7 Guanghua Road
Chaoyang, Beijing 100004
P.R. CHINA
www.campbellsci.com • info@campbellsci.com.cn
Campbell Scientific do Brasil Ltda.
Rua Apinagés, nbr. 2018 ─ Perdizes
CEP: 01258-00 ─ São Paulo ─ SP
BRASIL
www.campbellsci.com.br vendas@campbellsci.com.br
Please visit www.campbellsci.com to obtain contact information for your local US or international representative.
Campbell Scientific Canada Corp.
14532 – 131 Avenue NW
Edmonton AB T5L 4X4
CANADA
www.campbellsci.ca dataloggers@campbellsci.ca
Campbell Scientific Centro Caribe S.A.
300 N Cementerio, Edificio Breller
Santo Domingo, Heredia 40305
COSTA RICA
www.campbellsci.cc • info@campbellsci.cc
Campbell Scientific Ltd.
Campbell Park
80 Hathern Road
Shepshed, Loughborough LE12 9GX
UNITED KINGDOM
www.campbellsci.co.uk sales@campbellsci.co.uk
Campbell Scientific Ltd.
3 Avenue de la Division Leclerc
92160 ANTONY
FRANCE
www.campbellsci.fr info@campbellsci.fr
Campbell Scientific Ltd.
Fahrenheitstraße 13
28359 Bremen
GERMANY
www.campbellsci.de • info@campbellsci.de
Campbell Scientific Spain, S. L.
Avda. Pompeu Fabra 7-9, local 1
08024 Barcelona
SPAIN
www.campbellsci.es info@campbellsci.es
Loading...