The107-LC TEMPERATURE PROBE FOR METDATA1 is warranted by
CAMPBELL SCIENTIFIC, INC. to be free from defects in materials and
workmanship under nor mal use and service for twelve (12) months from date of
shipment unless specifi ed otherwise. Batteries have no warranty. CAMPBELL
SCIENTIFIC, INC.'s obligation under this warranty is limited to repairing or
replacing (at CAMPBELL SCIENTIFIC, INC.'s option) defective products.
The customer shall assume all costs of removing, reinstalling, and shipping
defective products to CAMPBELL SCIENTIFIC, INC. CAMPBELL
SCIENTIFIC, INC. will return such products by surface carrier prepaid. This
warranty shall not apply to any CAMPBELL SCIENTIFIC, INC. products
which have been subjected to modification, misuse, neglect, accidents of
nature, or shipping damage. This warranty is in lieu of all other warranties,
expressed or implied, including warranties of merchantability or fitness for a
particular purpose. CAMPBELL SCIENTIFIC, INC. is not liable for special,
indirect, incidental, or consequential damages.
Products may not be returned without prior authorization. The following
contact information is for US and International customers residing in countries
served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs
for customers wi thin their territories. Please visi t www.campbellsci.com to
determine which Campbell Scientific company serves your country. To obtain
a Returned Materials Authorization (RMA), contact CAMPBELL
SCIENTIFIC, INC., phone (435) 753-2342. After an applications engineer
determines the nature of the problem, an RMA number will be issued. Please
write this number clearly on the outside of the shipping container.
CAMPBELL SCIENTIFIC's shipping address is:
CAMPBELL SCIENTIFIC, INC.
RMA#_____
815 West 1800 North
Logan, Utah 84321-1784
CAMPBELL SCIENTIFIC, INC. does not accept collect calls.
107-LC Table of Contents
PDF viewers note: These page numbers refer to the printed version of this document. Use
the Adobe Acrobat® bookmarks tab for links to specific sections.
6. Maintenance and Calibration .....................................5
i
This is a blank page.
Model 107-LC Temperature Probe
for MetData1
1. General
The 107 Air/Soil/Water Temperature Probe uses a thermistor to measure
temperature. Custom lead lengths are available up to 1000 ft.
For air temperature measurement, a 41303 radiation shield is used to mount
and limit solar radiation loading of the 107. The 107 probe is designed to be
buried or submerged in water up to 100 feet.
1.1 Specifications
Temperature
Measurement Range: -35° to +50°C
Thermistor InterChangeability Error: Typically <±0.2°C over 0°C to 60°C; ±0.4 @ -35°C.
Polynomial
Linearization Error:<±0.5°C over -35°C to +50°C.
NOTE
2. Accuracy
The black outer jacket of the cable is Santoprene
compound was chosen for its resistance to temperature extremes,
moisture, and UV degradation. However, this jacket will support
combustion in air. It is rated as slow burning when tested
according to U.L. 94 H.B. and will pass FMVSS302. Local fire
codes may preclude its use inside buildings.
The overall probe accuracy is a combination of the thermistor's
interchangeability specification, the precision of the bridge resistors, and the
polynomial error. In a "worst case" all errors add to an accuracy of ±0.4°C
over the range of -24° to 48°C and ±0.9°C over the range of -38°C to 53°C.
The major error component is the interchangeability specification of the
thermistor, tabulated in Table 2-1. For the range of 0° to 50°C the
interchangeability error is predominantly offset and can be determined with a
single point calibration. Compensation can then be done with an offset entered
in the measurement instruction. The bridge resistors are 0.1% tolerance with a
10 ppm temperature coefficient. Polynomial errors are tabulated in Table 2-2
and plotted in Figure 2-1.
TABLE 2-3. Temperature, Re sistance, and Datalogger Output
TemperatureResistanceOutput
-40.004067212-39.18
-38.003543286-37.55
-36.003092416-35.83
-34.002703671-34.02
-32.002367900-32.13
-30.002077394-30.18
-28.001825568-28.19
-26.001606911-26.15
-24.001416745-24.11
-22.001251079-22.05
-20.001106485-20.00
-18.00980100-17.97
-16.00869458-15.95
-14.00772463-13.96
-12.00687276-11.97
-10.00612366-10.00
10.0020780710.06
12.0018780312.07
14.0016992414.06
16.0015392316.05
18.0013958818.02
20.0012672919.99
22.0011517921.97
24.0010479623.95
Temperature
Temperature (°C)Tolerance (
−400.40
−300.40
−200.32
−100.25
0 to +500.20
TABLE 2-2. Polynomial Error
-40 to +56<±1.0°C
-38 to +53<±0.5°C
-24 to +48<±0.1°C
°COHMS°C
-8.00546376-8.02
-6.00488178 -6.05
-4.00436773-4.06
-2.00391294-2.07
0.00351017-0.06
2.003152881.96
4.002835583.99
6.002553376.02
8.002302108.04
±°C)
2
Model 107-LC Temperature Probe for MetData1
26.009544925.94
28.008702627.93
30.007942829.95
32.007256731.97
34.006636533.99
36.006075236.02
38.005566838.05
40.005105840.07
42.004687342.07
44.004307144.05
46.003961346.00
48.003646547.91
50.003359849.77
52.003098351.59
54.002859553.35
56.002641355.05
58.002441956.70
60.002259358.28
3. Installation
FIGURE 2-1. Error Produced by Polynomial Fit to Published Values
The 107 must be housed inside a radiation shield when used in the field. The
41303 Radiation Shield (see Figure 3-1) mounts to a CM6 or CM10 tripod. The
UT018 mounting arm and UT6P Radiation Shield mount to a UT30 tower.
The standard lead length of 6 feet and 9 feet allow the 107 to be mounted at a 2
meter height on the CM6/CM10 tripod or the UT30 tower respectively.
3
Model 107-LC Temperature Probe for MetData1
4. Connection
5. Programming
NOTE
FIGURE 3-1. 107 and 41303 Radiation Shield
on a CM6/CM10 Tripod Mast
The 107 cable is attached to the MetData1 connector #4 or #7. Connector #4
uses single-ended input 4L (channel 8) and connector #7 uses single-ended
input 5H (channel 9).
Information in this section is not necessary when programming
the MetData1 with the Short Cut Program Builder software.
Instruction 11 is used to measure temperature. Instruction 11 provides AC
excitation, makes a single-ended voltage measurement, and calculates
temperature with a fifth order polynomial. A multiplier of 1.0 and an offset of
0.0 yields temperature in Celsius. For Fahrenheit, use a multiplier of 1.8 and an
offset of 32.
4
Model 107-LC Temperature Probe for MetData1
MetData1 Datalogger Program Example
01: Temp 107 Probe (P11)
1: 1Rep
2: 8 or 9**IN Chan
3: 3Excite all reps w/EXchan 3
4: 11*Loc [:Air_Temp]
5: 1Mult
6: 0Offset
* Proper entries will vary with input location assignments.
** Specify input channel #8 for connector #4. Specify input channel #9 for connector #7.
6. Maintenance and Calibration
The 107 Probe requires minimal maintenance. Check monthly to make sure the
radiation shield is free from debris.
For most applications it is unnecessary to calibrate the 107 to eliminate the
thermistor offset. However, for those users that are interested, the following
briefly describes calibrating the 107 probes.
A single point calibration can be performed to determine the 107 temperature
offset (thermistor interchangeability). This calibration will not remove the
polynomial error. The value of the offset must be chosen so that the probe
outputs the temperature calculated by the polynomial, not the actual calibration
temperature. For example, a 107 is placed in a calibration chamber that is at
0°C and the probe outputs 0.1°C. The offset is -0.16, because at 0°C the
polynomial calculates a temperature of -0.06°C (Table 7-1).
5
Model 107-LC Temperature Probe for MetData1
This is a blank page.
6
This is a bla nk page.
Campbell Scientific Companies
Campbell Scientific, Inc. (CSI)
815 West 1800 North
Logan, Utah 84321
UNITED STATES
www.campbellsci.com
info@campbellsci.com
Campbell Scientific Africa Pty. Ltd. (CSAf)
PO Box 2450
Somerset West 7129
SOUTH AFRICA
www.csafrica.co.za
sales@csafrica.co.za
Campbell Scientific Australia Pty. Ltd. (CSA)
PO Box 444
Thuringo wa Cent ra l
QLD 4812 AUSTRALIA
www.campbellsci.com.au
info@campbellsci.com.au
Campbell Scientific do Brazil Ltda . (CSB)
Rua Luisa Crapsi Orsi, 15 Butantã
CEP: 005543-000 São Paulo SP BRAZIL
www.campbellsci.com.br
suporte@campbellsci.com.br
Campbell Scientific Canada Corp. (CSC)
11564 - 149th Street NW
Edmonton, Alberta T5M 1W7
CANADA
www.campbellsci.ca
dataloggers@campbellsci.ca
Campbell Scientific Ltd. (CSL)
Campbell Park
80 Hathern Road
Shepshed, Loughborough LE12 9GX
UNITED KINGDOM
www.campbellsci.co.uk
sales@campbellsci.co.uk
Campbell Scientific Ltd. (France)
Miniparc du Verger - Bat. H
1, rue de Terre Neuve - Les Ulis
91967 COURTABOEUF CEDEX
FRANCE
www.campbellsci.fr
campbell.scientific@wanadoo.fr
Campbell Scientific Spain, S. L.
Psg. Font 14, local 8
08013 Barcelona
SPAIN
www.campbellsci.es
info@campbellsci.es
Please visit www.campbellsci.com to obtain contact information for your local US or International representative.