CAMBRIDGE CSW-15AT Installation, Setup And Operation Instructions

1
2
TABLE OF CONTENTS
1.0 Operation ------------------------------------------------------------------------ 4
1.1 Key Functions ---------------------------------------------------------- 4
1.2 Error Messages --------------------------------------------------------- 4
1.3 Numeric Entries -------------------------------------------------------- 5
1.4 Setup Parameters ------------------------------------------------------- 5
2.0 Scale Procedure ----------------------------------------------------------------- 6
2.1 Software Navigation Flowchart -------------------------------------- 6 - 7
2.2 Scale Menu Definitions ----------------------------------------------- 8 - 9
3.0 Calibration Procedure --------------------------------------------------------- 10
3.1 Calibration Menu Definitions ---------------------------------------- 10
3.2 Calibration -------------------------------------------------------------- 11
3.3 Linearity Correction --------------------------------------------------- 11
4.0 Communications Setup -------------------------------------------------------- 12
4.1 Communications Menu Definitions --------------------------------- 12
5.0 Testing Procedure ------------------------------------------------------------- 13
5.1 Testing Menu Definitions ------------------------------------------- 13
6.0 Warranty ------------------------------------------------------------------------ 14
7.0 Assistance ---------------------------------------------------------------------- 14
3
METER CONNECTIONS
To connect power via AC wall adapter cable to the CSW-15AT meter. First make sure the On/Off switch on the rear of the meter is in the off position. Connect the wall adapter to your 110VAC outlet. Next, connect the other end of the cable to the connector on the rear of the indicator.
On/OFF Switch
COMMUNICATIONS SIGNAL PORT
Pin # Description Pin # Description 1 ------ Shield 1 ------ +EXC 2 ------ RXD 2 ------ -SIG 3 ------ TXD 5 ------ Shield 4 ------ +5V 6 ------ -EXC 5 ------ SGND 7 ------ +SIG
4
1.0 OPERATION
1.1 Key Functions:
ZERO Brings the scale to a zero balance reading. If the ZERO key
is pressed and held for 5 seconds the Calibration zero value will be displayed.
GRS/NET Toggles the display between Gross weight and Net weight. This key is also used to enter Setup & Calibration Mode. To enter, press and hold this key until the Parameter (P xxx) Event counter is displayed, then release. Immediately after CodE is displayed, enter in sequence (within 5 seconds) TARE, lb/kg, GRS/NET, and PRINT/ENTER. The display will indicate ScAlE and you are now in Setup & Calibration mode.
Note: P xxx and C xxx are event counters that will
increment each time one or more changes are made
to the Scale or Calibration Parameters.
TARE Enters the Gross weight value into the Tare display and switches to the net display mode. If the TARE key is pressed and held for 5 seconds the current Tare value will be displayed.
Lb/kg Toggles the display between pounds and kilograms.
1.2 Error Messages:
ScnEg When the weight is more than 10 divisions negative from the zero
calibration point.
OLD The scale is in an overload condition.
BAtLo Will flash when the battery voltage falls to 10.8VDC and will be
displayed constantly when the voltage falls to 10.2VDC.
Err d More than 5000 scale divisions have been selected in S1 Ntep or S1
Angle mode. More than 20,000 scale divisions have been selected in
S1 No Mode.
5
1.3 Numeric Entries:
Once you have entered Setup and Calibration mode, (as described on the previous
page) some entries may require entering a numeric value.
To enter a numeric value:
1. Press and release the lb/kg key to move into the desired menu field.
2. Press and release the ZERO key, the first digit will begin to flash.
3. Use the up and down arrow keys to increase or decrease the value of this digit until you reach the desired value.
4. Press the TARE key to move to the next digit and repeat step 3.
When you are satisfied with the value of each digit press the PRINT/ENTER key to
save your entry and return to the menu.
1.4 Setup Parameters:
To begin press and hold this key until the Parameter (P xxx) Event counter is displayed,
then release. Immediately after CodE is displayed, enter in sequence (within 5 seconds) TARE,
lb/kg, GRS/NET, and PRINT/ENTER. The display will indicate ScAlE and you are now in Setup
& Calibration mode.
Note: P xxx and C xxx are event counters that will increment each time one or more Changes are made to the scale Scale or Calibration Parameters.
With the display indicating ScAlE, press the GRS/NET key to move down to change a parameter. This allows the operator to change any one of the Scale Parameters S1 thru S18. For example, press the GRS/NET key to move down until S2 is displayed, S2 is used to set the capacity of the scale. Press the lb/kg key to move right into the S2 parameter. The current capacity will be displayed. Press the ZERO key; the first digit will begin flashing. Enter the new capacity using the steps described in section 1.4 shown at the top of the page. Once you have the capacity entered, press the PRINT/ENTER key to enter the value and return to S2.
The same steps can be followed to enter the count by. Press the GRS/NET key to move down until S3 is displayed. Press the lb/kg key to move right into the S3 parameter. The current Count By will be displayed. Use the up and down arrow keys to increase or decrease the value. Use the left and right arrow keys to adjust the decimal point. When the desired Count By is entered, press the PRINT/ENTER key to enter the value and return to S3.