Cal Test Electronics CT4076-NA User Manual

CT4076-NA
User Manual
Safety Summary
To avoid personal injury and/or product damage, review and comply with the following safety precautions. These precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument.
voltage probe is designed to be used by personnel who are trained,
experienced, or otherwise qualifi ed to recognize hazardous situations
and who are trained in the safety precautions necessary to avoid possible injury when using such a device.
This instrument is intended for usage within Measurement Category I (CAT I) only.
Do not work alone when working with high voltages.
For your own safety, inspect the probe and accessories for cracks and frayed or broken leads before each use. If defects are noted, DO NOT USE the probe.
Hands, shoes, fl oor and work bench must be dry.
Avoid making measurements under humidity, dampness or other environmental conditions that might affect safety.
The probe should be kept clean and free of any conductive contamination.
Do not remove the probe casing. Removal of the probe’s casing may expose you to electric shock.
Do not use the probe with its case open.
Disconnect the inputs and outputs of the probe before opening the case.
800-572-1028 caltestelectronics.com 1
Use only in offi ce-type indoor setting
The instrument is designed to be used in offi ce-type indoor
environments. Do not operate the instrument:
l In the presence of noxious, corrosive, or flammable fumes,
gases, vapors, chemicals, or finely-divided particulates.
l In environments where there is a danger of any liquid being
spilled on the instrument.
l In air temperatures exceeding the specified operating
temperatures.
l In atmospheric pressures outside the specified altitude limits or
where the surrounding gas is not air.
Not for critical applications.
This instrument is not authorized for use in contact with the human body or for use as a component in a life-support device or system.
Hazardous voltages may be present in unexpected locations in circuitry
being tested when a fault condition in the circuit exists.
Do not substitute parts that are not approved by Cal Test Electronics or modify this instrument. Return the instrument to Cal Test Electronics for service and repair to ensure that safety and performance features are maintained.
WARNING statement calls attention to an operating procedure,
A practice, or condition, which, if not followed correctly, could result in injury or death to personnel.
CAUTION statement calls attention to an operating procedure,
A practice, or condition, which, if not followed correctly, could result in damage to or destruction of parts or the entire product.
800-572-1028 caltestelectronics.com 2
Compliance Statements
Disposal of Old Electrical& Electronic Equipment
(Applicable in the European Union and other European countries with separate collection systems). This product is subject to Directive 2012/19/EU of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as
unsorted municipal waste. Please utilize your local
WEEE collection facilities in the disposition of this
product and otherwise observe all applicable requirements.
800-572-1028 caltestelectronics.com 3
1 Introduction
Overview
Use the CT4076-NA 35-MHz differential probe to make safe and accurate fl oating measurements with an oscilloscope. The CT4076-NA differential probe allows conventional earth-grounded oscilloscopes to be used for fl oating signal measurements of up to ±7.5 kV for both
differential and common mode voltage.
Features
l 35 MHz bandwidth l Selectable attenuation settings of 100x or 1000x l Up to ±7.5 kV differential and common mode voltage l Compatible with most oscilloscopes
l Powered by included 9 VDC mains adapter
Initial Inspection
This unit is tested prior to shipment. It is therefore ready for immediate use upon receipt. An initial physical inspection should be made to ensure that no damage has been sustained during shipment. After the inspection, verify the contents of the shipment. The included
accessories for this product are:
l Two fully insulated pincer style clips (black & red) l Two fully insulated alligator clips (black & red) l Two 60 cm input leads (black & red) l One 100 cm insulated BNC output cable l One 9 V mains adapter l User manual
800-572-1028 caltestelectronics.com 4
2 Product Description
Front Panel
4 mm Input Safety Banana Jacks
Power On/Off Indicator (Green)
Output BNC Connection
Figure 1 Front Panel Features
3 Using the Probe
Getting Started
Overload Indicator (Red)
Attenuation Function Switch
External Power
Connection
1. Connect the leads to the inputs on the probe.
2. Connect the clips to the leads.
3. Connect the probe to the oscilloscope with the BNC cable.
4. Connect the probe to the mains adapter.
5. Switch the probe from “OFF” to the desired attenuation ratio. The power light should come on.
6. Connect the clips to the circuit to be tested.
800-572-1028 caltestelectronics.com 5
Vertical Scale on Oscilloscope
The actual vertical scale of the oscilloscope is equal to the attenuation factor multiplied by the range of vertical scale selected on the oscilloscope. For example, with the probe on factor 100x, the
oscilloscope on 0.5 V/div, the real vertical scale is 100 x 0.5 = 50 V/ div. With the probe on 1000x, the real vertical scale is 1000 x 0.5 = 500 V/div. These values apply when the oscilloscope is set to 50 Ω input. When the oscilloscope is set to 1 MΩ input, the actual vertical scale will be halved: 25 V/div for the 100x probe setting and 250 V/div for the
1000x probe setting. See the chart below.
Vertical Scale on Oscilloscope
Probe
Output
Impedance
50 Ω 50 Ω 10x 0.5 V/div 50 V/div
50 Ω 50 Ω 1000x 0.5 V/div 500 V/div
50 Ω 1 MΩ 10x 0.5 V/div 25 V/div
50 Ω 1 MΩ 1000x 0.5 V/div 250 V/div
Scope Input
Impedance
Probe
Attenuation
Setting
Vertical Scale
Reading on the
Oscilloscope
Actual Vertical Scale
of the Oscilloscope
External Power Source
Power consumption of the probe is about 200 mA, thus it is not suited for battery usage. Use the supplied adapter only. CAUTION: The probe
may be damaged when the input power exceeds 12 VDC.
Using a DMM
Using a DMM rather than an oscilloscope can signifi cantly increase your accuracy. Use the CT2944-50 plus the CT3956 converter to use
the probe with a DMM.
800-572-1028 caltestelectronics.com 6
4 Cleaning
This probe does not require any particular cleaning. If necessary, clean the case with a cloth slightly moistened with soapy water
WARNING
Dry the probe thoroughly before attempting to make voltage measurements.
CAUTION
Do not subject the probe to solvents or solvent fumes as these can cause deterioration of the probe body and cables.
800-572-1028 caltestelectronics.com 7
5 Specifi cations
All specifi cations apply to the unit after a temperature stabilization time of 15 minutes over an
ambient temperature range of 25 °C ± 5 °C.
CT4076-NA Specifi cations
Operating Parameters
Bandwidth 35 MHz
Rise Time 10 ns
Attenuation 100x / 1000x
Accuracy ±2%
AC CMRR -80 dB @ 60 Hz
Input Impedance Between inputs: 50 MΩ // 1.3 pF
Input Voltage
Differential Voltage (DC+ACpk) ±750 V / ±7.5 kV
Common-Mode Voltage (DC+ACpk)
Absolute Max Voltage (DC+ACpk)
Output Voltage
Swing ±8 V (±4 V into 50 Ω load)
Source Impedance 50 Ω
General
Power Supply External 9 VDC power supply
Power Consumption 200 mA about (9 VDC)
Operating Temperature/ Humidity
Storage Temperature/Humidity -30°C to 70°C / 10% to 90% RH
Cable Length 100 cm
Input Leads Length 60 cm each
Weight 350 g
Dimensions 220 x 85 x 30 mm
-60 dB @ 100 Hz
-50 dB @ 1 MHz
Each input to ground: 25 MΩ // 2.6 pF
±7.5 kV or 5.5 kVrms
±7.5 kV or 5.5 kVrms
0°C to 50°C / 10% to 85% RH
Specifi cations are subject to change without notice. To ensure the most current version of
this manual, please download the current version from our website: caltestelectronics.com.
800-572-1028 caltestelectronics.com 8
6 Voltage Derating Curve
The derating curve of the absolute maximum input voltage in common mode is show as follows:
Figure 1. CT4076-NA Derating Curve
800-572-1028 caltestelectronics.com 9
7 Service & Warranty Information
Limited One-Year Warranty
Cal Test Electronics warrants this product to be free from defective material or workmanship for a period of 1 year from the date of original purchase. Under this warranty, Cal Test Electronics is limited to repairing the defective device when returned to the factory, shipping charges prepaid, within the warranty period.
Units returned to Cal Test Electronics that have been subject to abuse, misuse, damage or accident, or have been connected, installed or adjusted contrary to the instructions furnished by Cal Test Electronics,
or that have been repaired by unauthorized persons, will not be
covered by this warranty.
Cal Test Electronics reserves the right to discontinue models, change
specifi cations, price, or design of this device at any time without notice
and without incurring any obligation whatsoever.
The purchaser agrees to assume all liabilities for any damages and/or bodily injury which may result from the use or misuse of this device by the purchaser, his employees, or agents.
This warranty is in lieu of all other representations or warranties expressed or implied and no agent or representative of Cal Test
Electronics is authorized to assume any other obligation in connection
with the sale and purchase of this device.
Service
If you have a need for calibration or repair services, technical or sales support, please contact us:
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
800-572-1028 or 714-221-9330
caltestelectronics.com
800-572-1028 caltestelectronics.com 10
© 2017 Cal Test Electronics450860-001 Rev. A
Loading...