Cal Test Electronics CT3686 User's Manual

CT3686 & CT3687
User Manual
Safety Summary
To avoid personal injury and/or product damage, review and comply with the following safety precautions. These precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this probe.
A CAUTION statement calls attention to an operating procedure, practice, or condition, which, if not followed correctly, could result in damage to or destruction of parts or the entire product.
Do Not Work Alone
Do not work alone when working with high voltages.
Inspect the Probe
Inspect the probe and accessories for cracks and frayed or broken leads before each use. If defects or damages are noted, DO NOT USE the probe.
Dry Conditions
Hands, shoes, oor and work bench must be dry. Avoid making
measurements under humidity, dampness or other environmental conditions that might affect safety.
Do Not Remove the Probe’s Casing
Removal of the probe’s casing may expose you to electric shock. If necessary, disconnect the inputs and outputs of the probe before opening the case.
Hazardous Contact
To avoid injury, remove jewelry such as rings, watches, and other metallic objects. Do not touch exposed connections and components when power is present.
Unexpected Charges
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit exists. Capacitors
800-572-1028 caltestelectronics.com 1
inside the instrument may retain a charge even if the instrument is disconnected from its source of supply.
Use Only in Ofce-Type Indoor Setting
The probe is designed to be used in ofce-type indoor environments.
Do not operate the probe:
z In the presence of noxious, corrosive, or flammable fumes,
gases, vapors, chemicals, or finely-divided particulates.
z In environments where there is a danger of any liquid being
spilled on the probe.
z In air temperatures exceeding the specified operating
temperatures.
z In atmospheric pressures outside the specified altitude limits
or where the surrounding gas is not air.
Not for Critical Applications
This probe is not authorized for use in contact with the human body or
for use as a component in a life-support device or system.
Do Not Substitute Parts
Do not install substitute parts or perform any unauthorized modication
to the instrument.
Only Qualied Personnel
Only qualied personnel should use this probe. This differential
voltage probe is designed to be used by personnel who are trained,
experienced, or otherwise qualied to recognize hazardous situations
and who are trained in the safety precautions necessary to avoid possible injury when using such a device.
Observe Maximum Working Voltage
Do not use the CT3686 or CT3687 probes above their maximum
working volatage ranges. See “Electrical Specications” on page 10.
Use Proper Power Source
To ensure this probe functions well, use four AA cells or 6 VDC/200 mA or regulated 9 VDC/120 mA mains adapter or power leads. Do not
operate this probe from a power source that applies more than the
800-572-1028 caltestelectronics.com 2
voltage speci ed.
Must be Grounded
This probe is grounded by the shell of the BNC connector through the grounding conductor of the power cord of the measurement instrument. Before making connections to the input leads of this probe, ensure that the output BNC connector is attached to the BNC connector of the measurement instrument, and that the measurement instrument is properly grounded. Whenever it is likely that the ground protection is impaired, you must make the instrument inoperative and secure it against any unintended operation.
Use Fused Test Probes if Necessary
If this probe is intended to use for measurements in circuits of installation CAT III, it should incorporate the use of fused test probes.
Compliance Statements
Disposal of Old Electrical & Electronic Equipment
(Applicable in the European Union and other European countries with separate collection systems). This product
is subject to Directive 2012/19/EU of the European
Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being
put on the market after August 13, 2005, and should not
be disposed of as unsorted municipal waste. Please utilize your local WEEE collection facilities in the disposition of
this product and otherwise observe all applicable requirements.
This probe is in compliance with IEC-61010-031 CAT II, Pollution
Degree 2.
800-572-1028 caltestelectronics.com 3
1 Introduction
1.1 Overview
Differential probes allow safe, accurate measurement between two voltage points where neither point is referenced to ground. The
CT3686 and CT3687 both offer 100 MHz bandwidth and can test up to ±700 V (DC + AC peak) and ±1400 V (DC + AC peak), respectively.
Compatible with oscilloscopes from all major manufacturers, the probes can be battery operated, powered by a universal adapter (optional), or powered by USB power lead (optional) if the oscilloscope is so equipped.
Features:
z Meets IEC 61010-031 safety standard z 100 MHz bandwidth (-3 dB) z Up to ±700 V (DC + AC peak) common mode (CT3686) z Up to ±1400 V (DC + AC peak) common mode (CT3687) z Selectable attenuation settings of 10x/100x (CT3686) z Selectable attenuation settings of 100x/1000x (CT3687) z Power and over-range indicators z High accuracy (±2%) z High CMRR z Powered by 4 AA batteries (included) z Power adapter, CT3723 (optional) z USB power lead, CT4122 (optional)
800-572-1028 caltestelectronics.com 4
1.2 Initial Inspection
This probe is tested prior to shipment. It is therefore ready for immediate use upon receipt. An initial physical inspection should be made to ensure that no damage has been sustained during shipment. After the inspection, verify the contents of the shipment. The kit contains:
z Differential probe z (2) Hook probes, black & red z (4) AA batteries z Offset adjustment tool z User manual
2 Product Description
2.1 Front Panel
Output BNC
Power On/O Switch
Over-range Indicator
cable, 90 cm, with 4 mm Banana Plug Ground Lead
Oset Adjustment
Attenuation Function Switch
LED Power Indicator
Figure 1 Front Panel Diagram
800-572-1028 caltestelectronics.com 5
Hook Probes
Leads, 60 cm
3 Using the Probe
3.1 Replacing the Batteries
Before using the differential probe for the rst time, the batteries
supplied with the device must be inserted in the battery compartment (unless you are using the power adapter or USB power lead).
WARNING
At the time of inserting or replacing the batteries, the input leads must not be connected to an item to be tested. Never operate the probe with the case open.
Slide back the battery cover. If necessary, the old AA batteries can then be removed and the new ones inserted into the compartment. Always ensure the batteries are positioned for proper polarity. After inserting the batteries, close the case. When the batteries are low, the power
indicator will start to icker and dim.
3.2 Inspection Procedure
1. Connect the BNC output connector to the vertical input the
oscilloscope.
2. Power on the probe.
3. Set the oscilloscope input to DC coupling and 1V/div. Center
the trace on the display.
4. Set the attenuation of the oscilloscope to match the low
setting on the probe.
5. Connect the hook probes to the leads.
6. Connect the black hook probe to the ground connection on the oscilloscope and the red hook probe to the test signal on
the oscilloscope (1 kHz for example).
7. A wave matching the test signal should display on the screen of the oscilloscope and this means this probe is working properly.
800-572-1028 caltestelectronics.com 6
3.3 Getting Started
1. Connect the hook probes to the leads.
2. Connect the probe to the oscilloscope with the BNC cable. When using a portable or ungrounded oscilloscope, connect the output ground lead to ground.
3. Switch the probe “ON.”
4. Switch to the desired attenuation ratio.
5. Use the hook probes to contact the circuit to be tested.
CAUTION
This probe is used to carry out differential measurements between two points on the circuit under test. This probe is not for electrically insulating the circuit under test and the measuring instrument.
3.4 Overrange Indicator
The overrange indicator lights when the voltage of the input signal exceeds the linear operating range of the probe. When this happens, the signal on the probe output may not accurately represent the signal on the probe input.
800-572-1028 caltestelectronics.com 7
3.5 Vertical Scale on Oscilloscope
The actual vertical scale of the oscilloscope is equal to the attenuation factor multiplied by the range of vertical scale selected on the
oscilloscope. For example, with the CT3686 set on attenuation 10x, the oscilloscope on 0.5 V/div, the real vertical scale is 10 x 0.5 = 5 V/div. With the probe on 100x, the real vertical scale is 100 x 0.5 = 50 V/div. These values apply when the oscilloscope is set to the typical 1 MΩ impedance input. When the oscilloscope is set to 50 Ω input, the actual vertical scale will be doubled: 10 V/div for the 10x setting and 100 V/div for the 100x setting. See Table 1.
Vertical Scale on Oscilloscope
Scope Input
Impedance
1 MΩ 10x 10x 0.5 V/div 5 V/div
1 MΩ 100x 100x 0.5 V/div 50 V/div
50 Ω 10x 20x 0.5 V/div 10 V/div
50 Ω 100x 200x 0.5 V/div 100 V/div
Table 1 Oscilloscope Readings
Probe
Attenuation
Setting
Actual
Attenuation
Setting
Vertical Scale
Reading on the
Oscilloscope
Actual Vertical
Scale of the
Oscilloscope
Offset Zero Procedure
The CT3686 and CT3687 can be adjusted to zero the probe’s offset voltage using the offset adjustment tool supplied with the probe. Follow this procedure to perform the offset adjustment.
1. Connect the probe to Channel 1 of the oscilloscope. Turn
on the probe power. You may use the USB power cable or batteries to power the probe. Set the probe attenuation ratio
to the low setting (10x for CT3686 and 100x for CT3687).
2. Short the + and - probe inputs together with the hook tips.
3. Turn on power to the oscilloscope. Leave both the
800-572-1028 caltestelectronics.com 8
instrument and the probe turned on for 30 minutes to
stabilize.
4. Press [Default Setup] and [Auto] on the oscilloscope.
5. Press the Channel 1 button, then press the Probe softkey
and set the attenuation to match that of the probe.
6. Set the oscilloscpe to DC coupled mode and the scope
offset to 0 volts.
7. Set the oscilloscope to average mode (x16) or high- resolution mode to reduce oscilloscope noise.
8. Using the offset adjustment tool (included), adjust the probe
offset voltage to 0 volts
4 Cleaning
This probe does not require any particular cleaning. If necessary, clean
the case with a cloth slightly moistened with soapy water. Make sure
the probe is completely dry before reconnecting it to an oscilloscope.
WARNING
Dry the probe thoroughly before attempting to make voltage measurements.
CAUTION
Do not subject the probe to solvents or solvent fumes as these can cause deterioration of the probe body and cables.
800-572-1028 caltestelectronics.com 9
Specications
All specications apply to the unit after a temperature stabilization time of 20 minutes over an ambient temperature range of 25 °C ± 5 °C.
Electrical Specications
CT3686 CT3687
Bandwidth (-3dB) 100 MHz (driving 1 MΩ oscilloscope input)
Rise Time (10%-90%) 3.5 ns
Attenuation ratio 10x/100x 100x/1000x
Accuracy ±2%
CMRR (typical)
Maximum Differential Input
Voltage
(DC + AC peak)
Maximum Common Mode Input
Voltage
(DC + AC peak)
Absolute Maximum Rated Input
Voltage (each side to ground)
Input Impedance 4 MΩ // 7 pF (each side to ground)
Output Voltage Swing ±7 V (driving 1 MΩ oscilloscope input)
Offset (typical) ±5 mV
Noise (typical) 0.9 mVrms
Source Impedance 50 Ω
Power Supply
-85 dB @ 50 Hz
-55 dB @ 1 MHz
±70 V @ 10x
±700 @ 100x
±700 V ±1400 V
700 Vrms CAT II 1000 Vrms CAT II
4 AA batteries (included) or
CT3723 power adapter (optional)
CT4122 USB power lead (optional)
-80 dB @ 60 Hz
-50 dB @ 1 MHz
±140 V @ 100x
±1400 @ 1000x
Mechanical Characteristics
Weight 500 g
Dimensions 207 x 83 x 38 mm
BNC Cable Length 90 cm
Input Leads Length 30 cm each
800-572-1028 caltestelectronics.com 10
Environmental Characteristics
Operating Temp/Humidity -10°C to 40°C / Up to 85% RH
Storage Temp/Humidity -30°C to 70°C / Up to 85% RH
Pollution Degree Pollution Degree 2
Altitude
Operating: 3,000 m
Nonoperating: 15,300 m
Safety Specications
IEC 61010-031 CAT II
Specications are subject to change without notice. To ensure the most current version of this
manual, please download the current version from our website: caltestelectronics.com
5 Voltage Derating Curve
The derating curve of the absolute maximum input voltage in common mode is show as follows:
CT3686 & CT3687 Maximum Input Voltage Derating Curve
10,000
1,000
1000 1000
100
20
10
Voltage (Vrms)
10
5
10
6
10
7
10
8
Frequency (Hz)
Figure 2 Derating Curve
800-572-1028 caltestelectronics.com 11
6 Service & Warranty Information
6.1 Limited One-Year Warranty
Cal Test Electronics warrants this product to be free from defective
material or workmanship for a period of 1 year from the date of original
purchase. Under this warranty, Cal Test Electronics is limited to repairing the defective device when returned to the factory, shipping charges prepaid, within the warranty period.
Units returned to Cal Test Electronics that have been subject to abuse, misuse, damage or accident, or have been connected, installed or adjusted contrary to the instructions furnished by Cal Test Electronics, or that have been repaired by unauthorized persons, will not be covered by this warranty.
Cal Test Electronics reserves the right to discontinue models, change
specications, price, or design of this device at any time without notice
and without incurring any obligation whatsoever.
The purchaser agrees to assume all liabilities for any damages and/or bodily injury which may result from the use or misuse of this device by the purchaser, his employees, or agents.
This warranty is in lieu of all other representations or warranties expressed or implied and no agent or representative of Cal Test Electronics is authorized to assume any other obligation in connection with the sale and purchase of this device.
6.2 Service
If you have a need for calibration or repair services, technical or sales support, please contact us:
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
800-572-1028 or 714-221-9330
caltestelectronics.com
800-572-1028 caltestelectronics.com 12
800-572-1028 caltestelectronics.com 13
800-572-1028 caltestelectronics.com 14
© 2017 Cal Test Electronics450847-001
Loading...