Unauthorized duplication of Cal Test Electronics documentation
materials is strictly prohibited. Customers are permitted to duplicate
and distribute Cal Test Electronics documentation for internal
educational purposes.
1 caltestelectronics.com 450888-001 Rev. A
Page 3
Cal Test Electronics
Safety Summary
To avoid personal injury and/or product damage, review and comply with
the following safety precautions. These precautions apply to both operating
and maintenance personnel and must be followed during all phases of
operation, service, and repair of this probe.
A WARNING statement calls attention to an operating procedure,
practice, or condition, which, if not followed correctly, could result
in injury or death to personnel.
A CAUTION statement calls attention to an operating procedure,
practice, or condition, which, if not followed correctly, could result
in damage to or destruction of parts or the entire product.
Do Not Work Alone
Do not work alone when working with high voltages.
Inspect the Probe
Inspect the probe and accessories for cracks and frayed or broken leads
before each use. If defects or damages are noted, DO NOT USE the
probe.
Dry Conditions
Hands, shoes, oor, and work bench must be dry. Avoid making
measurements under humidity, dampness, or other environmental
conditions that might affect safety.
Do Not Remove the Probe’s Casing
Removal of the probe’s casing may expose you to electric shock. If
necessary, disconnect the inputs and outputs of the probe before opening
the case.
Hazardous Contact
To avoid injury, remove jewelry such as rings, watches, and other metallic
objects. Do not touch exposed connections and components when power
is present.
Unexpected Charges
Hazardous voltages may be present in unexpected locations in circuitry
being tested when a fault condition in the circuit exists.
450888-001 Rev. A caltestelectronics.com 2
Page 4
Cal Test Electronics
Capacitors inside the instrument may retain a charge even if the instrument
is disconnected from its source of supply.
Use Only in Ofce-Type Indoor Setting
The probe is designed to be used in ofce-type indoor environments. Do
not operate the probe:
zIn the presence of noxious, corrosive, flammable fumes, gases,
vapors, chemicals, or finely-divided particulates.
zIn environments where there is a danger of any liquid being
spilled on the probe.
zIn air temperatures exceeding the specified operating
temperatures.
zIn atmospheric pressures outside the specified altitude limits or
where the surrounding gas is not air.
Not for Critical Applications
This probe is not authorized for use in contact with the human body or for
use as a component in a life-support device or system.
Do Not Substitute Parts
Do not install substitute parts or perform any unauthorized modication to
the instrument.
Only Qualied Personnel
Only qualied personnel should use this probe. This differential voltage
probe is designed to be used by personnel who are trained, experienced,
or otherwise qualied to recognize hazardous situations and who are
trained in the safety precautions necessary to avoid possible injury when
using such a device.
Observe Maximum Working Voltage
Do not use any probe above its maximum working voltage ranges. See
“Electronic Characteristics” on page 11.
Use Proper Power Source
Do not operate this probe from a power source that applies more than the
voltage specied.
3 caltestelectronics.com 450888-001 Rev. A
Page 5
Cal Test Electronics
Must be Grounded
This probe is grounded by the shell of the BNC connector through the
grounding conductor of the power cord of the measurement instrument.
Before making connections to the input leads of this probe, ensure
that the output BNC connector is attached to the BNC connector of the
measurement instrument, and that the measurement instrument is properly
grounded. Whenever it is likely that the ground protection is impaired,
you must make the instrument inoperative and secure it against any
unintended operation.
Compliance and Certications
CE Compliance
This product meets the essential requirements of the applicable European
Directives as follows:
This product meets the follow standard(s) of safety for electrical equipment
for measurement, control and laboratory use:
zIEC/EN 61326-1zIEC/EN 55011
Safety
This product meets the follow standard(s) of safety for electrical equipment
for measurement, control and laboratory use:
zIEC/EN 61010-1
450888-001 Rev. A caltestelectronics.com 4
Page 6
Cal Test Electronics
Disposal of Old Electrical & Electronic Equipment
(Applicable in the European Union and other European
countries with separate collection systems). This product is
subject to Directive 2012/19/EU of the European Parliament
and the Council of the European Union on waste electrical
and electronic equipment (WEEE), and in jurisdictions
adopting that Directive, is marked as being put on the market
after August 13, 2005, and should not be disposed of as
unsorted municipal waste. Please utilize your local WEEE
collection facilities in the disposition of this product and
otherwise observe all applicable requirements.
Terms and Symbols
The following symbols may appear on the product or in its documentation:
Direct current
5 caltestelectronics.com 450888-001 Rev. A
Both direct and alternating current
Caution, possibility of electric shock
Caution, see documentation for details
Double or reinforced insulation
Earth ground
Page 7
Cal Test Electronics
Denitions
MEASUREMENT CATEGORY - classication of testing and measuring
circuits according to the types of mains circuits to which they are intended
to be connected.
MEASUREMENT CATEGORY II (CAT II) - measuring circuits
connected directly to utilization points of the low-voltage mains
installation
POLLUTION - addition of foreign matter, solid, liquid, or gaseous (ionized
gases) that may produce a reduction of dielectric strength or surface
resistivity.
POLLUTION DEGREE 2 - only non-conductive pollution occurs except
that occasionally a temporary conductivity caused by condensation is
expected.
Working CAT rating is equal to that of the lowest rated element
within the test set-up.
450888-001 Rev. A caltestelectronics.com 6
Page 8
Cal Test Electronics
Introduction
Overview
The CT4196 differential probe allows safe, accurate measurement
between two voltage points where neither point is referenced to
ground. Designed for high sensitivity measurements, the probe offers
up to 60 MHz bandwidth. Compatible with oscilloscopes from all major
manufacturers, the probe is exclusively powered by the included 9 V
power adapter.
Features
zMeets IEC 61010-1 safety standardzSelectable attenuation settings of 1x, 5x, and 10x z60 MHz bandwidth (at 10x attenuation setting)z50 MHz (at 5x attenuation setting)z15 MHz (at 1x attenuation setting)zUp to ±35 V (DC + AC peak) common mode
zHigh accuracy (±2%)zPower indicator LEDzOver range indicator LED
Initial Inspection
This unit is tested prior to shipment but an initial physical inspection
should be made to ensure that no damage has been sustained during
shipment. After the inspection, verify the contents of the shipment. The
included items for this product are:
zDifferential probez(2) Pincer clips, black & redz(2) Alligator clips, black & redz(2) Sheathed 4 mm banana plug test leads, silicone jacketed,
black and red
zInsulated BNC cable, 100 cmz9 V power adapterzUser manual
7 caltestelectronics.com 450888-001 Rev. A
Page 9
Product Overview
Probe Description
Test leads,
black and red
!
!
INPUT
65V
MAX
600V
CAT III
600V
CAT III
Over range
Indicator (Red)
Over
Range
1x
5x
10x
OFF
Dierential Mode Common Mode
±35 V @ 10x
POWER
±3.5 V @ 1x
±35 V±18 V @ 5x
DC+AC peak
CT4196
60 MHz Dierential Probe
Cal Test Electronics
Alligator Clips,
black and red
Attenuation
Selector Switch
External Power
OUTPUT
≤ ± 8V
INTO 1MΩ
EXT.SUPPLY
9V
200mA MAX
Connection
Power
Indicator (Red)
Pincer Clips,
black and red
Insulated BNC
Cable
Figure 1 CT4196: Front Panel Features
450888-001 Rev. A caltestelectronics.com 8
Page 10
Cal Test Electronics
Using the Probe
WARNING At the time of powering on the probe, the input
leads must not be connected to an item to be tested. Never
operate the probe with the case open.
CAUTION This probe is used to carry out differential mea-
surements between two points on the circuit under test. This
probe is not designed for electrically insulating the circuit
under test or the measuring instrument.
Getting Started
1. Connect the BNC output connector to the vertical input of a
general purposed oscilloscope. The oscilloscope must have a
ground referenced.
2. Connect power adapter.
3. Select the proper attenuation ratio. When measuring signals
below 3.5 V, switch the attenuation ratio to 1x. For voltages up
to 18 V, set the attenuation ratio to 5x and for voltages up to 35
V set the attenuation ratio to 10x.
4. The power indicator LED should turn on.
5. Plug the pincer or alligator clips onto the test leads and connect
the circuit under test.
Vertical Scale on Oscilloscope
The actual vertical scale of the oscilloscope is equal to the attenuation
factor multiplied by the range of vertical scale selected on the
oscilloscope. For example, with the probe on factor 10x and the
oscilloscope on 0.5 V/div, the real vertical scale is 10 x 0.5 = 5 V/
div. With the probe on 5x, the real vertical scale is 5 x 0.5 = 2.5 V/
div. These values apply when the oscilloscope is set to the typical 1
MΩ impedance input. When the scope is set to 50 Ω input, the actual
vertical scale will be doubled: 10 V/div for the 10x setting and 5 V/div
for the 5x setting. See the chart below.
9 caltestelectronics.com 450888-001 Rev. A
Page 11
Cal Test Electronics
Vertical Scale on Oscilloscope
Scope Input
Impedance
1 MΩ1x1x0.5 V/div0.5 V/div
1 MΩ5x5x0.5 V/div2.5 V/div
1 MΩ10x10x0.5 V/div5 V/div
50 Ω1x2x0.5 V/div1 V/div
50 Ω5x10x0.5 V/div5 V/div
50 Ω10x20x0.5 V/div10 V/div
Table 1 Oscilloscope Readings
Probe
Attenuation
Setting
Actual
Attenuation
Setting
Vertical Scale
Reading on the
Oscilloscope
Actual Vertical
Scale of the
Oscilloscope
Cleaning
This probe does not require any particular cleaning. If necessary, clean
the case with a soft cloth.
WARNING
Dry the probe thoroughly before attempting to make voltage
measurements.
CAUTION
Avoid immersing or using abrasive cleaners or solvents
containing Benzene (or similar solvents) on the probe as
these can cause deterioration of the probe body and cables.
450888-001 Rev. A caltestelectronics.com 10
Page 12
Cal Test Electronics
Specications
All specications apply to the unit after a temperature stabilization time
of 20 minutes over an ambient range of 25°C ± 5°C.
Electrical Characteristics
Bandwidth
Rise Time
Attenuation
Accuracy±2% *
AC CMRR
Maximum Input Voltage
Maximum Input Voltage
Maximum Input Voltage
Absolute Maximum Rated
Input Voltage (each side
Input Impedance (each
Output Voltage Swing
* Accuracy based on DMM with 10 MΩ input impedance.
(1x)
(DC + AC peak)
(5x)
(DC + AC peak)
(10x)
(DC + AC peak)
to ground)
Input Impedance
(Differential)
side to ground)
Offset (typical)
Noise (typical)
Source Impedance
Power Supply
±4 V (driving 1 MΩ oscilloscope input)
9 V power adapter (included)
60 MHz (10x attenuation)
50 MHz (5x attenuation)
15 MHz (1x attenuation)
5.8 ns for 10x
7 ns for 5x
23 ns for 1x
1x, 5x, 10x
80 dB @ 60 Hz
60 dB @ 100 Hz
50 dB @ 1 MHz
±3.5 V
±18 V
±35 V
600 Vrms
1 MΩ // 2.5 pF
0.5 MΩ // 5 pF
±5 mV
2 mVrms
50 Ω
11 caltestelectronics.com 450888-001 Rev. A
Page 13
Cal Test Electronics
Mechanical Characteristics
Weight250 g
Dimensions195 x 55 x 30 mm
BNC Cable Length100 cm
Input Leads Length55 cm each
Environmental Characteristics
Operating Temp/Humidity
Storage Temp/Humidity
Pollution DegreePollution Degree 2
0°C to 50°C / 10% to 85% RH
-30°C to 70°C / 10% to 90% RH
Safety Specications
IEC 61010-1 CAT II
Specications are subject to change without notice. To ensure the most current version of this
manual, please download the current version from our website: caltesetlectronics.com.
450888-001 Rev. A caltestelectronics.com 12
Page 14
Cal Test Electronics
Volgage (DC+ACpeak)
Voltage Derating Curve
The derating curve of the absolute maximum input voltage in common
mode is shown as follows:
CT4196 Maximum Input Voltage Derating Curve
40
35
30
25
20
10
1KHz1MHz10MHz10KHz100KHz100MHz
Frequency
Figure 2 CT4196: Derating Curve
13 caltestelectronics.com 450888-001 Rev. A
20
15
Page 15
Cal Test Electronics
Service & Warranty Information
Limited One-Year Warranty
Cal Test Electronics warrants these products to be free from defective
material or workmanship for a period of 1 year from the date of original
purchase. Under this warranty, Cal Test Electronics is limited to
repairing the defective device when returned to the factory, shipping
charges prepaid, within the warranty period.
Units returned to Cal Test Electronics that have been subject to abuse,
misuse, damage, or accident, or have been connected, installed, or
adjusted contrary to the instructions furnished by Cal Test Electronics,
or that have been repaired by unauthorized persons, will not be
covered by this warranty.
Cal Test Electronics reserves the right to discontinue models, change
specications, price, or design of this device at any time without notice
and without incurring any obligation whatsoever.
The purchaser agrees to assume all liabilities for any damages and/or
bodily injury which may result from the use or misuse of this device by
the purchaser, his employees, or agents.
This warranty is in lieu of all other representations or warranties
expressed or implied and no agent or representative of Cal Test
Electronics is authorized to assume any other obligation in connection
with the sale and purchase of this device.
Service
If you have a need for calibration or repair services, technical, or sales
support, please contact us:
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
800-572-1028 or 714-221-9330
caltestelectronics.com
450888-001 Rev. A caltestelectronics.com 14