Caliber IV User Manual

Caliber IVDigital
Hygrometer/Thermometer
User Manual
Thank you for purch asin g th e Calibe r IV di gita l hygrometer and thermometer. We are certain that you will find
this is the most dependable and accurate unit available in the consumer market . To maintain your hygrometer in optimal working order, we recommend that you review this manual thoroughly to understand how to use the advanced features of the Caliber IV™.
1. Humidity Reading
2. Temperature Reading
3. Set
4. Max/Min/Clear
5. Fahrenheit/Celsius
6. Battery Tray
7. Magnetic Mount
Getting Started
1. Slide the battery tray out to insert the battery plus side up. This will activate the display and begin the collection of readings.
2. Mount the unit onto the desired location. The best location is in a central area where the airflow to the unit will not be blocked. To attach the magnet mount to a surface, first ensure that the surface is clean and free of any dust or other debris. For best results, use an alcohol swab to clean the area where the adhesive will be applied. Remove the backing paper from the adhesive disc on the back of the magnet and apply firm pressure to hold the magnet in place for about 30 seconds. Allow the adhesive to dry for at least 24 hours before attaching the hygrometer.
Switching Between Celsius & Fahrenheit
To switch the display between Fahrenheit and Celsius temperature readings, move the Fahrenheit/Celsius Switch back and forth.
MAX/MIN Recording & Display
The Caliber IV™ will record the minimum and maximum values for
both humidity and temperature. To view the maximum values, press the MAX/MIN/Clear button one time. The maximum value will be displayed with the word MAX at the top of the temperature reading. To view the minimum values, press the MAX/MIN/Clear button two times. The minimum value will be displayed with the word MIN at the top of the temperature reading. To reset the values, while viewing either the MAX or MIN value, press and hold the MAX/MIN/Clear button for about 5 seconds. When the display shows empty values, that MAX or MIN value has been reset. Both values reset independently. Removing or replacing the battery will also reset both the minimum and maximum values.
1
2
6
3 4
5
7
79 Heisser Court
Farmingdale, NY 11735
www.cigaroasis.com
Calibrating the Humidity Display
Your hygrometer is factory calibrated and should not need any further adjustments. The following instructions should be performed only to correct an error in calibration.
1. Test the humidity display settings by placing the Caliber IV
next to another unit that is known to be accurate.
2. Determine the amount o f calibration needed by finding the difference bet ween the di splay value and the ex pected va lue. For example, if you are expecting 70% but your unit displays 71%, the difference in calibration is 71-70 = 1. If you need the display lower as in this example, change the calibration by -1%. If you need the display higher, add (+) the value.
3. Press the Set button for 3 seconds to p lace the unit into calibration mode.
4. In calibration mode, you will see the current adjustment setting flashing on the di splay. If this is th e first time y ou are calibrating your unit, it will show 0%. If you have already calibra ted the unit previously , you will see the cur rent adjustment.
5. Press the Min/Max/Clear button to adjust the display based on your calibration. Pressing the button once will change the adjustment by 1%. The Caliber IV adjustment range is +/-6% in the sequence: 0, +1, +2, +3, +4, +5, +6, -6, -5, -4, -3, -2, -1, 0.
6. When you have completed your adjustment, press the Set button o ne more t ime and it wil l take the unit o ut o f calibration mode. If you leave the unit in calibration mode for more than one minute without pressing any buttons, the unit will automatically will change back to operations mode.
Restoring Factory Calibration
To restore your Caliber IV back to factory calibration, simply follow the calibration instructions above and adjust the display back to 0%. This will restore the unit back to factory calibration.