VisionAire Family
User Controls & System Status Indicators
ISO 7000; Graphical symbols for use on equipment—Index and synopsis
Keep away from rain, keep dry. Reg. # 0626
Name and address of manufacturer. Reg. # 3082
Caution, consult accompanying docu­ments. Reg. # 0434A
Catalog Number. Reg. # 2493
Serial Number. Reg. # 2498
This way up. Reg. # 0623
Fragile, handle with care. Reg. # 0621
ISO 7010: Graphical symbols—Safety colors and safety signs—Registered safety signs
The instruction manual must be read. Reg. # M002
Keep away from open ame, re,
sparks. Open ignition source and smoking prohibited. Reg. # P003
Do not smoke near unit or while operating unit. Reg. # P002
Type BF applied part (degree of protection against electric shock). Reg. # 5333
Warning. Reg. # W001
Council Directive 93/42/EEC; concerning medical devices
Authorized representative in the European Community This device complies with the requirements of Directive 93/42/EEC concerning medical devices. It bears the CE marking as shown.
Internal Symbols
Keep away from ammable materials,
oil and grease.
Safety agency for CAN/CSA C22.2 No. 60601-1-14 for medical
electrical equipment. Certied for
both the U.S. and Canadian markets, to the applicable U.S. and Canadian standards.
Do not disassemble.
When present on the device alarm panel indicates external power inter­ruption has been detected.
2018 Labeling: When present on the device alarm panel indicates low ox­ygen concentration in device output.
ON (power switch on)
OFF (power switch off)
Date of Manufacture
Class II equipment
21 CFR 801.15: Code of Federal Regulations Title 21
Federal law restricts this device to sale by or on the order of a physician.
IEC 60601-1: Medical electrical equipment Part 1 General requirements for basic safety and essential performance
Drip Proof Equipment - IP21
2 - ENG PN MN238-C4 A | User Manual
Page 3
VisionAire Family
Council Directive 2012/19/EU: waste electrical and electronic equipment (WEEE)
WEEE
This symbol is to remind the equip­ment owners to return it to a recycling facility at the end of its life, per Waste Electrical and Electronic Equipment (WEEE) Directive. Our products will comply with the restriction of Hazardous Substances (RoHS) directive. They will not con­tain more than trace amounts of lead or other hazardous material content.
This product may be covered by one or more pat­ents, US and international. Please visit our website below for the listing of applicable patents. Pat.: patents.gtls.io.
AirSep® VisionAire® Oxygen Concentrator
This User Manual will acquaint you with CAIRE’s VisionAire Oxygen Concentrator. Make sure you read and understand all of the information contained in this guide before operating your concentrator. Should you have any questions, your Equipment Provider will be happy to answer them for you.
What is the Oxygen Concentrator
The air we breathe contains approximately 21% oxygen, 78% nitrogen, and 1% other gasses. In the VisionAire oxygen concentrator, room air is drawn into the machine through the air intakes. It then passes through an adsorbent material called molec­ular sieve. This material separates the oxygen from the nitrogen and allows only the oxygen to pass
through. The result is a ow of high-concentration
oxygen delivered to the user.
Note: There is never a danger of depleting the oxygen in a room when you use your Oxygen Concentrator unit.
PN MN238-C4 A | User Manual ENG - 3
Page 4
VisionAire Family
Why Your Physician Prescribed Oxygen
Many people suffer from a variety of heart, lung,
and other respiratory diseases. A signicant number of these people can benet from supplemental
oxygen therapy at home, when traveling, or while participating in daily activities away from home.
Oxygen is a gas that makes up 21% of the room air we breathe. Our bodies depend on a steady supply to function properly. Your physician prescribed a
ow or setting to address your particular respiratory
condition.
Although oxygen is a non-addictive drug, unautho­rized oxygen therapy can be dangerous. You must seek medical advice before you use this oxygen concentrator. The Equipment Provider who supplies your oxygen equipment will demonstrate how to set
the prescribed ow rate.
WARNING: “NO SMOKING – OXYGEN IN USE” SIGNS MUST BE PROMI­NENTLY DISPLAYED IN THE HOME, OR WHERE OXYGEN IS IN USE. USERS AND THEIR CAREGIVERS MUST BE INFORMED ABOUT THE DANGERS OF SMOKING IN THE PRESENCE OF, OR WHILE USING, MEDICAL OXYGEN.
Operator Profile
Concentrators are intended to supply supplemental oxygen to users suffering from discomfort due to
ailments which effect the efciency of one’s lungs
to transfer oxygen in the air to their bloodstream. Stationary oxygen concentrators (SOCs) do not store or contain oxygen. They do not need to be re-
lled, and can recharge anywhere AC or DC power.
Oxygen concentrator use requires a physician’s prescription and is not intended for life support use.
Although oxygen therapy can be prescribed for users of all ages, the typical oxygen therapy user is older than 65 years of age and suffers from a variety of respiratory diseases, including Chronic Obstructive Pulmonary Disorder (COPD). Users typically have good cognitive abilities and must be able to communicate discomfort. If the user is unable to communicate discomfort, or unable to read and understand the concentrator labeling and instructions for use, then use is recommended only under the supervision of one who can. If any dis­comfort is felt while using the concentrator, users are advised to contact their healthcare provider. Users are also advised to have back-up oxygen available (i.e. cylinder oxygen) in the event of a power outage or concentrator failure. There are no other unique skills or user abilities required for concentrator use.
CAUTION: The Manufacturer rec­ommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition, or mechanical failure. Consult your physician or Equipment Provider for the type of reserve system required.
It is very important to select only the pre­scribed level of oxygen. Do not change the flow selection unless you have been directed to do so by a licensed clinician.
The Oxygen Concentrator may be used during sleep under the recommendation of a licensed clinician.
4 - ENG PN MN238-C4 A | User Manual
Page 5
Safety Features
The following information will acquaint you with safety features of the VisionAire Oxygen Concen­trator. Make sure you read and understand all the information contained in this manual before you operate your unit. Should you have any questions, your Equipment Provider will be happy to answer them for you.
Device warning label and alarm display.
• Compressor Motor: A pressure relief valve is
tted to the compressor outlet and is calibrated to
280 kPa (40 psig). Thermal safety is ensured by a thermostat situated in the stator winding of the compressor (135°C / 275 °F).
• General Malfunction: If any of the conditions listed below occurs, the general malfunction light
) will illuminate and an audible intermittent
( alarm will activate.
This includes:
• Obstruction to the ow of oxygen such as a
pinch or kink in the delivery cannula, trig­gered by high product tank pressure
• High device product tank pressure condition of greater than 33psig (±1)
• Low device product tank pressure condition of less than 5psig (±1)
• High device temperature of greater than 135°C (275 °F), triggered by low product tank pressure if the thermal switch located within the compressor trips (shutting down the compressor)
VisionAire Family
• Oxygen Monitor: The oxygen monitor detects any drop in concentration below 82%. If this oc­curs the low oxygen concentration warning light ( ) will illuminate. If the low O2 condition persists, an audible intermittent alarm will also activate in addition to the alarm light.
• Power Failure: In the event the unit is operating and a loss of power occurs, the power warning light ( intermittent alarm will activate. See label image.
• Product Filter: ≥ 10 μm lter
) will illuminate and an audible
PN MN238-C4 A | User Manual ENG - 5
Page 6
VisionAire Family
Unpacking Your VisionAire
Verify that all of the components listed are included in the package. If any items are missing, contact your oxygen provider immediately.
• Stationary Oxygen Concentrator
Getting to Know Your VisionAire Oxygen Concentrator
First, become familiar with the important parts of your VisionAire Oxygen Concentrator.
A. On/Off (I/0) Power Switch: Starts and stops the
operation of the unit.
B. Circuit Breaker Reset Button: Resets the unit
after electrical overload shutdown
C. Hour Meter: Records the unit’s total hours of
operation.
D. Flowmeter/Adjustment Knob: Controls and
indicates the oxygen ow rate in liters per minute
(lpm).
E. Oxygen Outlet: Provides connection for a
humidier (if required) or cannula.
F. Top and Side Handles: Enables convenience in
carrying the unit. G. Warning and Alarms Label H. Specication Label: Displays electrical speci-
cations and serial number. I. Power Cord: Allows connection of unit into
electrical outlet. J. Casters: Four casters enable unit to be easily
moved, as needed.
K. Humidier Bottle Recess: Area to place the
optional humidier bottle. L. Humidier Bottle (optional) M. Humidier Bottle Oxygen Outlet: Connection
for oxygen tubing/cannula.
N. Humidier Bottle Oxygen Outlet O. Humidier Bottle Tubing P. Humidier Bottle Fitting
Q. Oxygen Tubing / Cannula
WARNING: DO NOT USE EXTENSION CORDS WITH THIS UNIT OR CONNECT TOO MANY PLUGS INTO THE SAME ELECTRICAL OUTLET. THE USE OF EXTENSION CORDS COULD AD­VERSELY AFFECT THE PERFORMANCE OF THE DEVICE. TOO MANY PLUGS INTO ONE OUTLET CAN RESULT IN AN OVERLOAD TO THE ELECTRICAL PANEL CAUSING THE BREAKER/FUSE TO ACTIVATE OR FIRE IF THE BREAKER OR FUSE FAILS TO OPERATE.
6 - ENG PN MN238-C4 A | User Manual
Page 7
VisionAire Family
F
E
C
L
F
D
B
K
A
G
G
J
Figure 1a
I
N
M
H
O
Figure 1b
P
Q
PN MN238-C4 A | User Manual ENG - 7
Page 8
VisionAire Family
Important!
Safety Instructions are defined as follows:
WARNING: IMPORTANT SAFETY INFORMATION FOR HAZARDS THAT MIGHT CAUSE SERIOUS INJURY.
CAUTION: Important information for preventing damage to the VisionAire Family.
Note: Information needing special attention.
Indications for Use
WARNING: IT IS VERY IMPORTANT TO SELECT ONLY THE PRESCRIBED LEVEL OF OXYGEN. DO NOT CHANGE THE FLOW SELECTION UNLESS YOU HAVE BEEN DIRECTED TO DO SO BY A LICENSED CLINICIAN.
WARNING: FEDERAL (USA) LAW RESTRICTS THIS DEVICE TO SALE OR RENTAL BY ORDER OF A PHYSICIAN OR OTHER LICENSED HEALTH CARE PROVIDER.
WARNING: THIS UNIT IS NOT TO BE USED FOR LIFE SUPPORT. GERIATRIC, PEDIATRIC, OR ANY OTHER USER UNABLE TO COMMUNICATE DISCOMFORT WHILE USING THIS DEVICE MAY REQUIRE ADDITIONAL MONITORING. USERS WITH HEARING AND/OR SIGHT IMPAIRMENT(S) MAY NEED ASSISTANCE WITH MONITORING ALARMS. IF YOU FEEL DISCOMFORT OR ARE EXPERIENCING A MEDICAL EMERGENCY, SEEK MEDICAL ASSISTANCE IMMEDIATELY.
Contraindications for Use
WARNING: IN CERTAIN CIRCUM­STANCES, THE USE OF NON-PRE­SCRIBED OXYGEN CAN BE HAZARD­OUS. THIS DEVICE SHOULD ONLY BE USED WHEN PRESCRIBED BY A PHYSICIAN.
WARNING: NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS.
WARNING: AS WITH ANY ELECTRICALLY POWERED DEVICE, THE USER MAY EXPERIENCE PERIODS OF NON-OPERATION AS A RESULT OF ELECTRICAL POWER INTERRUPTION, OR THE NEED TO HAVE THE OXYGEN CONCENTRATOR SERVICED BY A QUALIFIED TECHNICIAN. THE OXYGEN CONCENTRATOR IS NOT APPROPRIATE FOR ANY USER WHO WOULD EXPERIENCE AD­VERSE HEALTH CONSEQUENCES AS THE RESULT OF SUCH TEMPORARY INTERRUPTION.
Safety Guidelines
WARNING: CAREFULLY REVIEW AND FAMILIARIZE YOURSELF WITH THE FOLLOWING IMPORTANT SAFETY IN­FORMATION ABOUT THE VISIONAIRE INTENSITY OXYGEN CONCENTRATOR.
WARNING: DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTAND­ING THIS MANUAL. IF YOU ARE UNABLE TO UN­DERSTAND THE WARNINGS AND INSTRUCTIONS, CONTACT YOUR EQUIPMENT PROVIDER BEFORE ATTEMPTING TO USE THIS EQUIPMENT; OTHER­WISE INJURY OR DAMAGE COULD OCCUR.
WARNING: SMOKING WHILE USING OXYGEN IS THE NUMBER ONE CAUSE OF FIRE INJURIES AND RELATED DEATHS. YOU MUST FOLLOW THESE SAFETY WARNINGS:
WARNING: DO NOT ALLOW SMOKING, CANDLES, OR OPEN FLAMES IN THE SAME ROOM WITH THE DEVICE OR THE OXYGEN-CARRYING ACCES­SORIES.
WARNING: SMOKING WHILE WEARING AN OXY­GEN CANNULA CAN CAUSE FACIAL BURNS AND POSSIBLY RESULT IN DEATH.
WARNING: REMOVING THE CANNULA AND PLACING IT ON CLOTHING, BEDDING, SOFAS, OR OTHER CUSHION MATERIAL WILL CAUSE A FLASH FIRE WHEN EXPOSED TO A CIGARETTE, HEAT SOURCE, SPARK OR OPEN FLAME.
8 - ENG PN MN238-C4 A | User Manual
Page 9
VisionAire Family
WARNING: IF YOU SMOKE, YOU MUST ALWAYS FOLLOW THESE THREE (3) IM­PORTANT STEPS FIRST: TURN OFF THE OXYGEN CONCENTRATOR, TAKE OFF THE CANNULA, AND LEAVE THE ROOM WHERE THIS DEVICE IS LOCATED.
WARNING: “NO SMOKING – OXYGEN IN USE” SIGNS MUST BE PROMINENTLY DISPLAYED IN THE HOME, OR WHERE OXYGEN IS IN USE. USERS AND THEIR CAREGIVERS MUST BE IN­FORMED ABOUT THE DANGERS OF SMOKING IN THE PRESENCE OF, OR WHILE USING, MEDICAL OXYGEN.
WARNING: FEDERAL (USA) LAW RESTRICTS THIS DEVICE TO SALE OR RENTAL BY ORDER OF A PHYSICIAN OR OTHER LICENSED HEALTH CARE PROVIDER.
WARNING: THIS DEVICE SUPPLIES HIGH-CON­CENTRATION OXYGEN THAT PROMOTES RAPID BURNING. DO NOT ALLOW SMOKING OR OPEN FLAMES WITHIN THE SAME ROOM OF (1) THIS DEVICE, OR (2) ANY OXYGEN-CARRYING ACCES­SORY. FAILURE TO OBSERVE THIS WARNING CAN RESULT IN SEVERE FIRE, PROPERTY DAMAGE AND / OR CAUSE PHYSICAL INJURY OR DEATH.
WARNING: DO NOT USE YOUR OXYGEN CON­CENTRATOR IN THE PRESENCE OF FLAMMABLE GASES. THIS CAN RESULT IN RAPID BURNING CAUSING PROPERTY DAMAGE, BODILY INJURIES OR DEATH.
WARNING: DO NOT LEAVE A NASAL CANNULA ON CLOTHING, BED COVERINGS OR CHAIR CUSHIONS. IF THE UNIT IS TURNED ON BUT NOT IN USE, THE OXYGEN WILL MAKE THE MATERI­AL FLAMMABLE. SET THE I/0 POWER SWITCH TO THE 0 (OFF) POSITION WHEN THE OXYGEN CONCENTRATOR IS NOT IN USE.
WARNING: USE NO OIL, GREASE, OR PETRO­LEUM-BASED OR OTHER FLAMMABLE PROD­UCTS WITH THE OXYGEN-CARRYING ACCESSO­RIES OR THE OXYGEN CONCENTRATOR. OXYGEN ACCELERATES THE COMBUSTION OF FLAMMA­BLE SUBSTANCES. ONLY WATER BASED, OXYGEN COMPATIBLE LOTIONS OR SALVES SHOULD BE USED.
WARNING: DO NOT LUBRICATE FITTINGS, CON­NECTIONS, TUBING, OR OTHER ACCESSORIES OF THE OXYGEN CONCENTRATOR TO AVOID THE RISK OF FIRE AND BURNS.
WARNING: ELECTRICAL SHOCK HAZARD. TURN OFF THE UNIT AND DISCONNECT THE POWER CORD FROM THE ELECTRIC OUTLET BEFORE YOU CLEAN THE UNIT TO PREVENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD. ONLY YOUR EQUIPMENT PROVIDER OR A QUALIFIED SERVICE TECHNICIAN SHOULD REMOVE THE COVERS OR SERVICE THE UNIT.
WARNING: CARE SHOULD BE TAKEN TO PREVENT THE OXYGEN CONCENTRATOR FROM GETTING WET OR ALLOWING FLUIDS TO ENTER THE UNIT. THIS CAN CAUSE THE UNIT TO MAL­FUNCTION OR SHUT DOWN, AND CAUSE AN INCREASED RISK FOR ELECTRICAL SHOCK OR BURNS.
WARNING: DO NOT USE LIQUID DIRECTLY ON THE UNIT. A LIST OF UNDESIRABLE CHEMICAL AGENTS INCLUDES BUT IS NOT LIMITED TO THE FOLLOWING: ALCOHOL AND ALCOHOL-BASED PRODUCTS, CONCENTRATED CHLORINE-BASED PRODUCTS (ETHYLENE CHLORIDE), AND OIL­BASED PRODUCTS (PINE-SOL®, LESTOIL®). THESE ARE NOT TO BE USED TO CLEAN THE PLASTIC HOUSING ON THE OXYGEN CONCEN­TRATOR, AS THEY CAN DAMAGE THE UNIT’S PLASTIC.
WARNING: CLEAN THE CABINET, CONTROL PANEL, AND POWER CORD ONLY WITH A MILD HOUSEHOLD CLEANER APPLIED WITH A DAMP CLOTH (NOT WET) OR SPONGE, AND THEN WIPE ALL SURFACES DRY. DO NOT ALLOW ANY LIQUID TO GET INSIDE THE DEVICE.
WARNING: THE OXYGEN CONCENTRATOR SHOULD NOT BE USED ADJACENT TO OR STACKED WITH OTHER EQUIPMENT. IF ADJACENT OR STACKED USE IS UNAVOIDABLE, THE DEVICE SHOULD BE OBSERVED TO VERIFY NORMAL OPERATION.
WARNING: ALWAYS PLACE OXYGEN SUPPLY TUBING AND POWER CORDS IN A MANNER THAT PREVENTS TRIP HAZARD OR POSSIBLE ACCIDENTAL STRANGULATION.
WARNING: NO MODIFICATION OF THIS EQUIP­MENT IS PERMITTED.
WARNING: USE OF CABLES AND ADAPTERS OTHER THAN THOSE SPECIFIED, WITH THE EXCEPTION OF CABLES AND ADAPTERS SOLD BY THE MANUFACTURER OF THE MEDICAL ELEC­TRICAL EQUIPMENT AS REPLACEMENT PARTS FOR INTERNAL COMPONENTS, MAY RESULT IN INCREASED EMISSIONS OF DECREASED IMMU­NITY OF THE OXYGEN CONCENTRATOR.
PN MN238-C4 A | User Manual ENG - 9
Page 10
VisionAire Family
WARNING: ENVIRONMENTAL CONDI­TIONS CAN AFFECT PERFORMANCE OF DEVICE. LOCATE IN CLEAN, PEST-FREE ENVIRONMENT.
WARNING: DEVICE SHOULD ONLY BE OPERAT­ED BY END USERS, TRAINED CARE-GIVERS OR TRAINED TECHNICIANS. CHILDREN SHOULD NOT OPERATE DEVICE.
WARNING: TO ENSURE RECEIVING THE THERAPEUTIC AMOUNT OF OXYGEN DELIVERY ACCORDING TO YOUR MEDICAL CONDITION THE VISIONAIRE MUST BE USED WITH THE SPECIFIC COMBINATION OF PARTS AND ACCESSORIES THAT ARE IN LINE WITH THE SPECIFICATION OF THE CONCENTRATOR MANUFACTURER AND THAT WERE USED WHILE YOUR SETTINGS WERE DETERMINED.
WARNING: USE OF THIS DEVICE AT AN ALTITUDE, TEMPERATURE, OR RELATIVE HUMIDITY OUT­SIDE OF THE SPECIFIED VALUES LISTED IN THIS MANUAL CAN ADVERSELY AFFECT THE FLOW­RATE AND THE OXYGEN CONCENTRATION, AND CONSEQUENTLY THE QUALITY OF THE THERAPY.
WARNING: THE USE OF SOME OXYGEN ADMIN­ISTRATION ACCESSORIES NOT SPECIFIED FOR USE WITH THIS OXYGEN CONCENTRATOR MAY IMPAIR ITS PERFORMANCE. RECOMMENDED ACCESSORIES ARE REFERENCED WITHIN THIS MANUAL.
WARNING: USE ONLY ELECTRICAL VOLTAGE AS SPECIFIED ON THE SPECIFICATION LABEL AFFIXED TO THE DEVICE.
WARNING: DO NOT USE EXTENSION CORDS WITH THIS UNIT OR CONNECT TOO MANY PLUGS INTO THE SAME ELECTRICAL OUTLET. THE USE OF EXTENSION CORDS COULD ADVERSELY AFFECT THE PERFORMANCE OF THE DEVICE. TOO MANY PLUGS INTO ONE OUTLET CAN RESULT IN AN OVERLOAD TO THE ELECTRICAL PANEL CAUS­ING THE BREAKER/FUSE TO ACTIVATE OR FIRE IF THE BREAKER OR FUSE FAILS TO OPERATE.
CAUTION: Federal (USA) law restricts this device to sale or rental by order of a physician or other licensed health care provider.
CAUTION: Do not position the unit so that it is difficult to access the power cord.
CAUTION: The concentrator should be located as to avoid smoke, pollutants or fumes.
CAUTION: Ensure concentrator is operated in an upright position.
CAUTION: Always place oxygen supply tubing and power cords in a manner that prevents a trip hazard.
CAUTION: Position the unit away from curtains or drapes, hot air registers or heaters. Be certain to place the unit on a flat surface and make sure all sides are at least 1 foot (30 cm) away from a wall or other obstruction. Do not place the unit in a confined area. Choose a dust and smoke free-location away from direct sun­light. Do not operate the unit outdoors unless the unit is plugged into a Ground Fault Circuit Interrupter (GFCI) protected outlet.
CAUTION: Do not operate this unit in a restrict­ed or confined space where ventilation can be limited. This can cause the device to overheat and affect performance.
CAUTION: Do not allow either the air intake or the air outlet vents to be blocked. DO NOT drop or insert any object into any openings on the device. This can cause the Oxygen Concentrator to overheat and impair performance.
10 - ENG PN MN238-C4 A | User Manual
Page 11
VisionAire Family
CAUTION: The Manufacturer rec­ommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition, or mechanical failure. Consult your physician or Equipment Provider for the type of reserve system required.
CAUTION: It is very important to select only the prescribed level of oxygen. Do not change the flow selection unless you have been directed to do so by a licensed clinician.
CAUTION: The Oxygen Concentrator may be used during sleep under the recommendation of a licensed clinician.
CAUTION: Operating or storing the Oxygen Concentrator outside of its normal operating temperature range can impair the performance of the unit. Refer to the specification section of this manual for storage and operating tempera­ture limits.
CAUTION: In the event of an alarm or you ob­serve the Oxygen Concentrator is not working properly; consult the troubleshooting section of this manual. If you cannot resolve the problem, consult your Equipment Provider.
CAUTION: If the audio alarm is weak or does not sound at all, consult your Equipment Pro­vider immediately.
CAUTION: If the humidifier bottle tubing is not properly connected to the humidifier bottle fitting or to the oxygen outlet, an oxygen leak can occur.
CAUTION: Normally, you should not need to ad­just the flowmeter on your unit. If you turn the flowmeter adjustment knob clockwise, you will decrease and can shut off the flow of oxygen from your unit. For your convenience, the flow­meter is marked in ½ LPM increments. For units equipped with the 2 LPM flowmeter option, the flowmeter is marked in 1/8 LPM increments for flow settings up to 2 LPM.
Note: Cannula must be non-kinking, which can be used for a total length of 25 ft. (7.6 m) max.
Ensure the cannula is fully inserted and secure. You should hear or feel oxygen flow to the prongs of the nasal cannula. If oxygen does not seem to flow, first verify that the flow meter ball is registering a flow. Then, place the tip of the cannula into a glass of water; if bubbles come out of the cannula, oxygen is flowing. If bubbles do not appear, refer to the troubleshooting section of this manual.
Note: Always follow the cannula manufacturer’s instructions for proper use. Replace the disposable cannula as recommended by the cannula manufac­turer or your Equipment Provider. Additional sup­plies are available from your Equipment Provider.
Note: The VisionAire Oxygen Concentrator must be operated for at least five minutes at 2 LPM before using the unit.
The VisionAire is appropriate for usage by two users, provided the combined flow is a minimum of 2 LPM and does not exceed the maximum capacity of the concentrator.
To Equipment Provider: The following oxygen ad­ministration accessories are recommended for use with the VisionAire Oxygen Concentrator:
• Humidifier Bottle: Part No. HU003-1
• Nasal Cannula with 7 feet (2.1 m) of tubing (6 LPM max): Part No. CU002-1
• OxySafe Kit: Part No. 20629671 Note: The Manufacturer does not recommend the
sterilization of this equipment. Note: If the unit has not been used for an extended
period of time, it needs to operate for several minutes before power failure alarm can become activated.
Note: The concentrator releases warm air out the bottom of the unit which can permanently discolor temperature sensitive flooring surfaces such as vinyl. The concentrator should not be used over flooring that is sensitive to heat staining. The Manufacturer is not responsible for flooring that becomes discolored.
Note: To prevent a void warranty, follow all manu­facturers’ instructions.
Note: Do not attempt any maintenance other than the possible solutions listed within the manual.
PN MN238-C4 A | User Manual ENG - 11
Page 12
VisionAire Family
Note: Portable and mobile radio frequency (RF) communications equipment can effect medical electrical equipment.
Note: There is never a danger of depleting the oxygen in a room when you use your Oxygen Concentrator unit.
Note: To Equipment Provider: The following oxygen administration accessories are recommended for use with the Oxygen Concentrator:
• Nasal Cannula with 7 feet (2.1 m) of tubing: CU002-1
• For any additional recommended accessories, please see the Accessories Catalog (PN ML­LOX0010) available on www.CAIREmedical.com.
A firebreak is required for use with any cannula.
• CAIRE offers the OxySafe™. This is intended to be used in conjunction with the oxygen concentrator. The OxySafe™ is a thermal fuse to stop the flow of gas in the event that the downstream cannula or oxygen tubing is ignited and burns to the OxySafe. It is placed in-line with the nasal cannula or oxygen tubing between the patient and the oxygen outlet of the VisionAire. For proper use of the OxySafe, always refer to the manufacturer’s instructions (included with each OxySafe kit). CAIRE offers an OxySafe™ kit that includes a cannula headpiece, OxySafe™, and 7’ tubing: PN
20629671.
WARNING: CET APPAREIL PRODUIT DE L’OXYGÈNE À CONCENTRATION ÉLEVÉE, FAVORISANT UNE COMBUS­TION RAPIDE. NE PAS PERMETTRE DE FUMER OU DES FLAMMES NUES DANS LA MÊME CHAMBRE: (1) CET APPAREIL OU (2) TOUT ACCESSOIRE CONTENANT DE L’OXYGÈNE. NE PAS UTILISER DE PRODUITS À BASE D’HUILE, DE GRAISSE OU DE PÉTROLE SUR OU À PROXIMITÉ DE L’UNITÉ. DÉCONNECTER LE CORDON D’ALI­MENTATION DE LA PRISE ÉLECTRIQUE AVANT DE NETTOYER OU DE FAIRE L’ENTRETIEN DE L’UNITÉ.
RISQUE DE CHOC ÉLECTRIQUE. NE PAS ENLEVER LES COUVERCLES LORSQUE L’UNITÉ EST BRANCHÉE. SEULS VOTRE FOURNISSEUR D’ÉQUIPEMENT OU UN TECHNICIEN DE SERVICE QUALIFIÉ DEVRAIT ENLEVER LES COUVERCLES OU FAIRE L’ENTRETIEN DE L’UNITÉ.
12 - ENG PN MN238-C4 A | User Manual
Page 13
VisionAire Family
Specifications
VisionAire
Flow Rates* 1 LPM – 5 LPM
±10% of indicated setting, or 200 mL, whichever is greater*
Dimensions 14.1 in. W x 11.5 in. D x 20.8 in. H
(35.8 cm W x 29.2 cm D x 52.8 cm H)
Weight 30 lbs (13.6 kg) Sound Pressure Level 45 dBa (1 to 5 LPM) Power Consumption 290 Watts O2 Concentration 90% (+ 5.5% / - 3%) Electrical Requirements 115 VAC / 60 Hz, 3.0A
230VAC /50 Hz, 1.5A 230VAC / 60 Hz, 1.5A
Operating Environment* 5° C to 40° C (41° F to 104° F ) at altitudes from up to 10,000 ft (3048
Altitude -1250 to 10,000 ft (-381 to 3048 m) (tested to 700 – 1060 hPa) Storage Environment -25° C - 70° C (-13°F - 158°F)
* Based on an atmospheric pressure range of 700 hPa to 1060 hPa at 70°F (21°C)
m) above sea level. 15 - 95% relative humidity (non-condensing)
The expected service life of the equipment is a minimum of ve years.
See Technical Manual (PN MN138-1) for Sound Power Level.
WARNING: USE OF THIS DEVICE AT AN ALTITUDE, TEMPERATURE, OR RELATIVE HUMIDITY OUTSIDE OF THE SPECIFIED VALUES LISTED IN THIS MANUAL CAN ADVERSELY AFFECT THE FLOWRATE AND THE OXYGEN CONCENTRATION, AND CONSEQUENTLY THE QUALITY OF THE THERAPY.
WARNING: USE OF THIS DEVICE AT AN ALTITUDE ABOVE 10,000 FT (3048 M) OR OUTSIDE A TEM­PERATURE OF 5° C TO 40° C (41° F TO 104° F ) OR A RELATIVE HUMIDITY ABOVE 95% IS EXPECTED TO ADVERSELY AFFECT THE FLOWRATE AND THE PERCENTAGE OF OXYGEN AND CONSEQUENTLY THE QUALITY OF THE THERAPY .
PN MN238-C4 A | User Manual ENG - 13
Page 14
VisionAire Family
Operating Instructions
Review the following information before you oper­ate your oxygen concentrator.
Note: The concentrator releases warm air out the bottom of the unit which can permanently discolor temperature sensitive flooring surfaces such as vinyl. The concentrator should not be used over flooring that is sensitive to heat staining. The Manufacturer is not responsible for flooring that becomes discolored.
Humidifier Bottle (Optional)
If additional humidication is required with your
oxygen therapy, perform the following steps each
time you ll or clean the humidier, which may
have been initially set up for your use.
1. Remove the humidier bottle from the humidi­er bottle recess.
2. Open the humidier bottle. If you have a pre­lled bottle, do not perform this step. Proceed to
step 5.
3. Fill the humidier bottle with cool or cold water (distilled water is preferred) to the ll line indi-
cated on the bottle. DO NOT OVERFILL.
4. Re-connect the top cover to the humidier bot­tle.
5. Place the humidier bottle in the humidier
bottle recess on the back of the concentrator
and connect the humidier bottle tubing to the oxygen outlet and the humidier bottle tting.
CAUTION: If the humidifier bottle tubing is not properly connected to the humidifier bottle fitting or to the oxygen outlet, an oxygen leak can occur.
Cannula Connection
Connect the tubing and cannula to the unit’s oxygen
outlet, or to the optional humidier’s oxygen outlet.
Note: Cannula must be non-kinking, which can be used for a total length of 25 ft. (7.6 m) max.
Ensure the cannula is fully inserted and secure. You should hear or feel oxygen flow to the prongs of the nasal cannula. If oxygen does not seem to flow, first verify that the flowmeter ball is registering a flow. Then, place the tip of the cannula into a glass of water; if bubbles come out of the cannula, oxygen is flowing. If bubbles do not appear, refer to the troubleshooting section of this manual.
Always follow the cannula manufacturer’s instruc­tions for proper use. Replace the disposable cannula as recommended by the cannula manufacturer or your Equipment Provider. Additional supplies are available from your Equipment Provider.
Starting the Concentrator
WARNING: “NO SMOKING – OXYGEN IN USE” SIGNS MUST BE PROMI­NENTLY DISPLAYED IN THE HOME, OR WHERE OXYGEN IS IN USE. USERS AND THEIR CAREGIVERS MUST BE INFORMED ABOUT THE DANGERS OF SMOKING IN THE PRESENCE OF, OR WHILE USING, MEDICAL OXYGEN.
WARNING: DO NOT USE EXTENSION CORDS WITH THIS UNIT OR CONNECT TOO MANY PLUGS INTO THE SAME ELECTRICAL OUTLET. THE USE OF EXTENSION CORDS COULD ADVERSELY AFFECT THE PERFORMANCE OF THE DEVICE. TOO MANY PLUGS INTO ONE OUTLET CAN RESULT IN AN OVERLOAD TO THE ELECTRICAL PANEL CAUS­ING THE BREAKER/FUSE TO ACTIVATE OR FIRE IF THE BREAKER OR FUSE FAILS TO OPERATE.
14 - ENG PN MN238-C4 A | User Manual
Page 15
VisionAire Family
CAUTION: Position the unit away from curtains or drapes, hot air registers or heaters. Be certain to place the unit on a flat surface and make sure all sides are at least 1 foot (30 cm) away from a wall or other obstruction. Do not place the unit in a confined area. Choose a dust and smoke free-loca­tion away from direct sunlight. Do not operate the unit outdoors unless the unit is plugged into a Ground Fault Circuit Interrupter (GFCI) pro­tected outlet.
CAUTION: Do not operate this unit in a restrict­ed or confined space where ventilation can be limited. This can cause the device to overheat and affect performance.
CAUTION: Do not allow either the air intake or the air outlet vents to be blocked. DO NOT drop or insert any object into any openings on the device. This can cause the
CAUTION: Oxygen Concentrator to overheat and impair performance.
1. Locate the unit near the electrical outlet in the room where you spend most of your time.
2. Insert the power cord plug into the electrical outlet.
3. Set the I/0 power switch to the “I” position to turn the unit on. An audible alarm will loudly sound for approximately 5 seconds.
4. The low oxygen concentration indicator remains on for a few minutes and until the oxygen concentration reaches minimum concentration requirements, (only pertains to unit equipped with an Oxygen Monitor)
5. To set the ow of supplemental oxygen, turn the owmeter adjustment knob left or right until the ball inside the owmeter centers on the ow line
number prescribed by your physician. To view
the owmeter at the proper angle, note that the
back line and the front number line must give appearance of one line.
6. The concentrator is now ready for use.
CAUTION: The Manufacturer rec­ommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition, or mechanical failure. Consult your physician or Equipment Provider for the type of reserve system required.
CAUTION: It is very important to select only the prescribed level of oxygen. Do not change the flow selection unless you have been directed to do so by a licensed clinician.
CAUTION: The Oxygen Concentrator may be used during sleep under the recommendation of a licensed clinician
CAUTION: Ensure concentrator is operated in an upright position.
Note: Optimal oxygen concentration is obtained within 10 minutes after the device is switched on (90% of the concentration is obtained after approximately 5 minutes).
Normally, you should not need to adjust the
owmeter on your unit. If you turn the owmeter
adjustment knob clockwise, you will decrease and
can shut off the ow of oxygen from your unit.
Turning the Concentrator Off
Set the I/0 power switch to the “0” position to turn off the unit.
PN MN238-C4 A | User Manual ENG - 15
Page 16
VisionAire Family
Cleaning, Care, and Proper Maintenance
Cabinet
WARNING: ELECTRICAL SHOCK HAZARD. TURN OFF THE UNIT AND DISCONNECT THE POWER CORD FROM THE ELECTRIC OUTLET BEFORE YOU CLEAN THE UNIT TO PREVENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD. ONLY YOUR EQUIP­MENT PROVIDER OR A QUALIFIED SERVICE TECHNICIAN SHOULD REMOVE THE COVERS OR SERVICE THE UNIT.
WARNING: CARE SHOULD BE TAKEN TO PREVENT THE OXYGEN CONCENTRATOR FROM GETTING WET OR ALLOWING FLUIDS TO ENTER THE UNIT. THIS CAN CAUSE THE UNIT TO MAL­FUNCTION OR SHUT DOWN, AND CAUSE AN INCREASED RISK FOR ELECTRICAL SHOCK OR BURNS.
WARNING: DO NOT USE OIL, GREASE, OR PETROLEUM-BASED OR OTHER FLAMMABLE PRODUCTS WITH THE OXYGEN-CARRYING ACCESSORIES OR THE OXYGEN CONCENTRATOR. OXYGEN ACCELERATES THE COMBUSTION OF FLAMMABLE SUBSTANCES.
WARNING: USE ONLY WATER-BASED LOTIONS OR SALVES THAT ARE OXYGEN COMPATIBLE PRIOR TO AND DURING THERAPY. NEVER USE PETROLEUM OR OIL-BASED LOTIONS OR SALVES TO AVOID THE RISK OF FIRE AND BURNS.
WARNING: DO NOT USE LIQUID DIRECTLY ON THE UNIT. A LIST OF UNDESIRABLE CHEMICAL AGENTS INCLUDES BUT IS NOT LIMITED TO THE FOLLOWING: ALCOHOL AND ALCOHOL-BASED PRODUCTS, CONCENTRATED CHLORINE-BASED PRODUCTS (ETHYLENE CHLORIDE), AND OIL­BASED PRODUCTS (PINE-SOL®, LESTOIL®). THESE ARE NOT TO BE USED TO CLEAN THE PLASTIC HOUSING ON THE OXYGEN CONCEN­TRATOR, AS THEY CAN DAMAGE THE UNIT’S PLASTIC. CLEAN THE CABINET, CONTROL PANEL, AND POWER CORD ONLY WITH A MILD HOUSE­HOLD CLEANER APPLIED WITH A DAMP CLOTH (NOT WET) OR SPONGE, AND THEN WIPE ALL SURFACES DRY. DO NOT ALLOW ANY LIQUID TO GET INSIDE THE DEVICE.
WARNING: CLEAN THE CABINET, CONTROL PANEL, AND POWER CORD ONLY WITH A MILD HOUSEHOLD CLEANER APPLIED WITH A DAMP (NOT WET) CLOTH OR SPONGE, AND THEN WIPE ALL SURFACES DRY. DO NOT ALLOW ANY LIQUID TO GET INSIDE THE CONCENTRATOR. PAY SPECIAL ATTENTION TO THE OXYGEN OUTLET FOR THE CANNULA CONNECTION TO MAKE SURE IT REMAINS FREE OF DUST, WATER, AND PARTICLES.
Note: Always follow the cannula manufacturer’s instructions for proper use. Replace the disposable cannula as recommended by the cannula manufac­turer or your Equipment Provider. Additional sup­plies are available from your Equipment Provider.
Note: The Manufacturer does not recommend the sterilization of this equipment.
Use a mild household cleaner applied with a damp cloth or sponge to clean the exterior of the concen­trator, and then wipe all surfaces dry. Do not allow any liquid to get inside the device. Device cabinet should be cleaned at a minimum between users.
Humidifier Bottle (optional)
• Check water level daily and add water as needed
• To clean and disinfect the humidier, follow your
Equipment Provider’s instructions, or the instruc-
tions included with the humidier bottle.
16 - ENG PN MN238-C4 A | User Manual
Page 17
Alarm Conditions
All alarms are low priority alarms.
Alarm Indicates Action
General malfunction
yellow light and intermittent audible alarm
Oxygen monitor yellow light
and intermittent audible alarm
Power failure yellow
light and intermittent audible alarm
high product tank pressure
OR low product tank pressure OR high device temperature
low oxygen concentration Contact equipment provider for service.
power failure Ensure device is plugged into a known, working outlet.
Ensure flowmeter is open to minimum flow rate or high­er. Ensure cannula is not kinked or obstructed. Remove any devices connected downstream of the outlet of the device. Ensure device has at least 12" of clearance on all sides and intakes are not obstructed. Ensure external gross particle intake filter is clean and not clogged. En­sure unit is within operating temperature range. If issue persists, contact equipment provider for service.
Ensure breaker switch is pushed in. If issue persists contact equipment provider for service.
VisionAire Family
PN MN238-C4 A | User Manual ENG - 17
Page 18
VisionAire Family
Troubleshooting
If your VisionAire Oxygen Concentrator fails to operate properly, refer to the chart on the following pages for possible causes and solutions and, if needed, consult your Equipment Provider.
If you cannot get the unit to operate, connect your nasal cannula, face mask, or other accessories to a reserve supplemental oxygen supply.
Note: Do not attempt any maintenance other than the possible solutions listed within this manual. Mainte­nance is the responsibility of the provider and will be tracked by the provider.
Note: To prevent a void warranty, follow all manufacturers’ instructions. Note: If the unit has not been used for an extended time period, it needs to operate for a minimum of 15
minutes before power failure alarm can become activated.
Problem Probable Cause Solution
Unit does not operate. Power failure condition causes an alarm to sound.
Limited oxygen flow. Dirty or obstructed humidifi-
Condensation collects in the oxygen tubing when you use the humidifier bottle.
Intermittent alarm sounds. See ‘Safety Features’ section
Power cord not connected into electrical outlet.
No power at electrical outlet. Check power source, wall switch, fuse, or
Oxygen concentrator circuit breaker is activated.
er bottle.
Defective nasal cannula, face mask, catheter, and/ or oxygen delivery tube, or other accessory.
Cannula tubing loose. Check cannula tubing connection at control
Unit not properly venti­lated. Elevated operating temperature.
for a description of auditory indicators.
Check power cord plug at the electrical outlet for a proper connection.
circuit breaker in-house. Press (do not hold in) the circuit breaker reset
button in the front of the unit. If the circuit breaker trips again or the alarm
continues to sound after the unit is turned on, contact your Equipment Provider.
Remove the humidifier bottle from the oxygen outlet. If flow is restored, clean or replace with a new humidifier bottle.
Remove and check accessories for kinks or obstructions. Replace if needed.
panel. Make sure unit is positioned away from
curtains or drapes, hot air registers, heaters, and fireplaces. Be certain to place the unit so all sides are at least 12 inches (30.5 cm) away from a wall or other obstruction. Do not place the unit in a confined area.
Refill humidifier bottle with COLD water. DO NOT OVERFILL. Allow oxygen tubing to dry
out, or replace with new tubing. Set I/0 power switch to 0 position, use your
reserve oxygen supply and consult your Equip­ment Provider immediately.
18 - ENG PN MN238-C4 A | User Manual
Page 19
VisionAire Family
Problem Probable Cause Solution
Oxygen monitor light remains lit (yellow).
All other problems. Set I/0 power switch to the 0 position, use
Accessories
Oxygen concentration is ≤ 82%.
Set I/0 power switch to the 0 position, use your reserve oxygen supply (if provided), and consult your Equipment Provider immediately.
your reserve oxygen supply and consult your Equipment Provider immediately.
For proper performance and safety, use only these listed accessories supplied by CAIRE through your oxygen provider. Use of accessories not listed below could adversely affect the performance and/ or safety of the concentrator.
VisionAire Family Standard Accessories
Humidifier Bottle (6–15 LPM)
Cannula, 25 feet (7.6 m) (6 LPM max)
Humidifier Bottle Tubing
Humidifier Bottle Fitting
Part Number – HU003-1
Part Number – CU002-4
Part Number – TU255-1
Part Number – F0655-1
Oxygen Tubing, 25 feet
Tubing/Cannula Connector
OxySafe Kit (Flow 25 LPM max. Pressure 87 PSIG max.)
Note: Additional options may be available for coun­try-specific power cords where noted above. Contact CAIRE or your oxygen provider if alternate options are needed for order.
Part Number – CU004-3
Part Number – CU009-1
Part Number – 20629671
PN MN238-C4 A | User Manual ENG - 19
Page 20
VisionAire Family
Pd 2.1
Pd 2.1
Pd 2.1
EMC Testing
Medical equipment needs special precautions regarding EMC and needs to be installed and put into ser­vice according to the EMC information provided in this section.
Guidance and Manufacturer’s Declaration —Electromagnetic Emissions
The VisionAire is intended for use in the electromagnetic environment specified below. The customer or the user of the VisionAire should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions CISPR 11
RF emissions CISPR 11 Harmonic emissions
IEC 61000-3-2 Voltage fluctuations/
flicker emissions IEC 61000-3-3
Group 1
Class B
Complies
Complies
Recommended separation distances between portable and mobile RF communications equipment and the VisionAire Units
The VisionAire is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the VisionAire can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the VisionAire as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
from 150 kHz to 80 MHz
Separation distance according to frequency of transmitter
The VisionAire uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The VisionAire is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies build­ings used for domestic purposes.
m
from 80 MHz to 800 MHz
from 800 MHz to 2,5 GHz
W
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73 1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
20 - ENG PN MN238-C4 A | User Manual
Page 21
VisionAire Family
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The VisionAire is intended for use in the electromagnetic environment specified below. The customer or the user of the VisionAire should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment –
Electromagnetic environment – guidance
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
<5 % U
T
(>95 % dip in U
T
for 0,5 cycle
±2 kV for power supply lines
N/A
±1 kV line(s) to line(s)
±2 kV line(s) to earth <5 % U
(>95 % dip in U
)
for 0,5 cycle
T
guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
)
T
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
40 % U
T
(60 % dip in U for 5 cycles
70 % U
T
(30 % dip in U for 25 cycles
<5 % U
T
(>95 % dip in U for 5 sec
)
T
)
T
T
)
40 % U (60 % dip in U for 5 cycles
70 % U (30 % dip in U for 25 cycles
<5 % U (>95 % dip in U for 5 sec
Power frequency magnetic field IEC
3 A / m 3 A / m
61000-4-8
NOTE U
is the A.C. mains voltage prior to application of the test level.
T
T
Mains power quality should be that of
)
a typical commercial or hospital envi-
T
ronment. If the user of the VisionAire requires continued operation during
T
T
power mains interruptions, it is recom­mended that the VisionAire is powered from an uninterruptible power supply
)
T
(UPS).
)
T
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
PN MN238-C4 A | User Manual ENG - 21
Page 22
VisionAire Family
Pd 2.1
Pd 2.1
Pd 2.1
Guidance and Manufacturer’s Declaration–Electromagnetic Immunity
The VisionAire is intended for use in the electromagnetic environment specified below. The customer or the user of the VisionAire should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment –
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 Vrms 150 kHz to 80 MHz
3 V/m 80 MHz to 2,5 GHz
3 Vrms
3 V/m
guidance
Portable and mobile RF communica­tions equipment should be used no closer to any part of the VisionAire, in­cluding cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
= 80 MHz to 800 MHz
= 800 MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufac­turer and d is the recommended sepa­ration distance in metres (m).
Field strengths from fixed RF transmit­ters, as determined by an electromag­netic site survey,
a
should be less than the compliance level in each frequency range. b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the VisionAire is used ex­ceeds the applicable RF compliance level above, the VisionAire should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the VisionAire.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
22 - ENG PN MN238-C4 A | User Manual
Page 23
VisionAire Family
Method of Disposal
Disposal of Waste
All waste from CAIRE’s VisionAire Oxygen Con­centrator must be disposed using the appropriate
methods specied by local authorities
Disposal of Device
In order to preserve the environment, the concentra­tor must be disposed using the appropriate methods
specied by local authorities.
Classification
Type of protection against electric shock: Class II Protection from electric shock is achieved by double insulation. Protective earthing or reliance upon installation conditions are not required.
Degree of protection against electric shock: Type BF Equipment providing a particular degree of protection against electric shock regarding
1) allowable leakage current;
2) reliability of protective earth connection (if present). Not intended for direct cardiac application.
Degree of protection against harmful ingress of water: Drip-proof equipment – IP21 Protection against ingress of solid foreign objects greater than 12.5 mm diameter. Equipment provided with an enclosure preventing of such an amount of falling liquid as might interfere with the satisfactory and safe operation of the equipment.
Method of cleaning and infection control allowed: Please refer to Maintenance section in the VisionAire Service Manuals.
Degree of safety of application in the presence of
ammable anesthetic gases:
Equipment not suited for such application.
Mode of operation: Continuous duty.
Note: Always follow the cannula manufacturer’s instructions for proper use. Replace the disposable cannula as recommended by the cannula manufac­turer or your Equipment Provider. Additional sup­plies are available from your Equipment Provider.
Note: The manufacturer does not recommend the sterilization of this equipment.
PN MN238-C4 A | User Manual ENG - 23
Page 24
VisionAire Family
Commandes d’utilisation et indicateurs d’état du système
ISO 7000 : symboles graphiques à utiliser sur l’équipement – index et sommaire
Lire le manuel d’utilisation avant la mise en marche. Reg. # 1641
Maintenir à l’abri de l’humidité et au sec. Reg. # 0626
Nom et adresse du fabricant. Reg. # 3082
Attention, consulter la documentation
ISO 7010 : symboles graphiques – couleurs de sécurité et étiquettes de sécurité – étiquettes de sécurité enregistrées
Directive du Conseil européen 93/42/CEE relative aux dispositifs médicaux
fournie. Reg. # 0434A
Référence catalogue. Reg. # 2493
Numéro de série. Reg. # 2498
Haut. Reg. # 0623
Fragile, manipuler avec soin. Reg. # 0621
Le manuel d’instruction doit être lu. Reg. # M002
Tenir à l’écart des ammes nues, du feu,
des étincelles. Source d’incendie ouverte interdite et interdiction de fumer. Reg. # P003
Ne pas fumer à proximité de l’appareil ou pendant son fonctionnement. Reg. # P002
Pièce appliquée de type BF (degré de protection contre les chocs électriques). Reg. # 5333
Avertissement. Reg. # W001
Représentant agréé au sein de l’Union européenne Cet appareil est conforme aux exigences de la Directive 93/42/CEE relative aux dispositifs médicaux. Il porte le marquage CE, comme illustré.
Symboles internes
Tenir éloigné des matériaux, huiles et
graisses inammables.
Agence de sécurité pour CAN/ CSA C22.2 no 60601-1-14 pour les
appareils électromédicaux. Certié
pour les marchés des États-Unis et du Canada, conformément aux normes des États-Unis et du Canada en vigueur.
Ne pas démonter.
Si présent sur le panneau d’alarme de l’appareil, indique qu’une interruption de l’alimentation externe a été détectée.
Étiquetage 2018 : si présent sur le panneau d’alarme de l’appareil, indique une faible concentration d’oxygène dans la sortie de l’appareil.
ON (mise en marche de l’interrupteur)
ON (mise en arrêt de l’interrupteur)
Date de fabrication
Équipement de Classe II
21 CFR 801.15 : Code de la réglementation fédérale, chapitre 21
La loi fédérale restreint cet appareil à la vente par, ou sur ordonnance d’un médecin.
CEI 60601-1 : Appareil électro-médical – Partie 1 – Règles générales de sécurité de base et performances essentielles.
Équipement protégé contre les gouttes d’eau - IP21
24 - FRE PN MN238-C4 A | Manuel d’utilisation
Page 25
VisionAire Family
Directive du Conseil européen 2012/19/UE : déchets d’équipements électriques et électroniques (DEEE)
DEEE
Ce symbole a pour objet d’inciter le propriétaire de l’équipement à le retourner à un centre de recyclage à la
n de son cycle de vie, conformément
à la directive relative aux déchets d’équipements électriques et électroniques (DEEE). Nos produits sont conformes à la Directive sur la limitation des substances dangereuses (RoHS). Ils ne contiendront pas plus qu’une quantité
inme de plomb ou d’autres substances
dangereuses.
Ce produit peut être protégé par un ou plusieurs brevets, américains ou internationaux. Consultez notre site Web ci-dessous pour voir la liste des brevets. Brev. : patents.gtls.io.
Concentrateur d’oxygène AirSep® VisionAire
Ce manuel d’utilisation a pour objet de vous familiariser avec votre concentrateur d’oxygène VisionAire de CAIRE. Assurez-vous d’avoir lu et compris toutes les informations incluses dans ce guide avant d’utiliser votre concentrateur. Pour toute question, contactez le fournisseur de l’appareil.
®
Qu’est-ce que le concentrateur d’oxygène ?
L’air que nous respirons contient environ 21 % d’oxygène, 78 % d’azote et 1 % d’autres gaz. Avec le concentrateur d’oxygène VisionAire, l’air de la pièce est aspiré dans la machine par les admissions d’air. Il passe ensuite au travers d’un matériau absorbant appelé tamis moléculaire, qui sépare l’oxygène de l’azote et capture ce dernier pour laisser uniquement passer l’oxygène. Le résultat ? Un débit d’une forte concentration d’oxygène transmis à l’utilisateur.
Remarque : l’utilisation du concentrateur d’oxygène ne risque en aucun cas d’épuiser tout l’oxygène de la pièce.
PN MN238-C4 A | Manuel d’utilisation 25 - FRE
Page 26
VisionAire Family
Pourquoi votre médecin vous prescrit-il de l’oxygène ?
Nombre d’individus souffrent de différentes maladies cardiaques, pulmonaires ou encore respiratoires.
La majorité d’entre eux peuvent bénécier d’une
oxygénothérapie complémentaire à la maison, lorsqu’ils voyagent, ou même lorsqu’ils participent à des activités à l’extérieur.
L’oxygène est un gaz qui constitue 21 % de l’air que nous respirons. Pour fonctionner correctement, notre corps a besoin d’un apport régulier. Votre médecin
vous a recommandé un débit ou un réglage à dénir an de traiter votre affection respiratoire.
Bien que l’oxygène ne constitue en aucun cas une drogue provoquant un état de dépendance, une oxygénothérapie non autorisée peut s’avérer dangereuse. N’oubliez pas de demander conseil à votre médecin avant d’utiliser ce concentrateur d’oxygène. Le fournisseur de cet appareil vous montrera comment régler le débit prescrit.
AVERTISSEMENT : DES INDICATIONS « INTERDICTION DE FUMER ­UTILISATION D’OXYGÈNE » DOIVENT ÊTRE AFFICHÉES DE FAÇON BIEN VISIBLE À VOTRE DOMICILE OU DANS TOUT LIEU OÙ DE L’OXYGÈNE EST UTILISÉ. LES UTILISATEURS ET LE PERSONNEL SOIGNANT DOIVENT ÊTRE INFORMÉS DES DANGERS LIÉS AU FAIT DE FUMER EN PRÉSENCE OU PENDANT L’UTILISATION D’OXYGÈNE MÉDICAL.
Profil de l’utilisateur
Les concentrateurs visent à fournir un apport d’oxygène supplémentaire aux utilisateurs pris d’un malaise découlant d’une maladie qui réduit la capacité des poumons à distribuer l’oxygène contenu dans l’air dans la circulation sanguine. Les concentrateurs d’oxygène stationnaires (COS) ne contiennent pas d’oxygène. Ils ne doivent donc pas être remplis et peuvent être rechargés partout où une alimentation CA ou CC est disponible. L’usage d’un concentrateur requiert une prescription médicale et n’est pas destiné
à des ns de réanimation.
Bien qu’une oxygénothérapie puisse être prescrite aux utilisateurs de tout âge, les patients sont généralement âgés de plus de 65 ans et souffrent de différentes pathologies respiratoires, notamment la maladie pulmonaire obstructive chronique (MPOC). En règle générale, les utilisateurs sont dotés de capacités
intellectuelles sufsantes pour exprimer leur malaise.
Dans le cas contraire, ou lorsque l’utilisateur est incapable de lire et de comprendre les étiquettes apposées sur le concentrateur ou encore le mode d’emploi, le concentrateur peut uniquement être utilisé sous la supervision d’une personne compétente. Si une gêne est ressentie pendant l’utilisation du concentrateur, il est conseillé aux utilisateurs de contacter leur prestataire de soins. Il est également recommandé aux utilisateurs d’être équipés d’un apport d’oxygène de secours, tel qu’une bouteille d’oxygène, dans le cas d’une coupure électrique ou d’une panne du concentrateur. L’utilisation du concentrateur n’exige aucune compétence ni connaissance particulière.
MISE EN GARDE : le fabricant recommande de disposer d’une autre source d’alimentation en oxygène en cas de coupure de courant, de déclenchement d’alarme de l’appareil ou de panne mécanique. Consultez votre médecin ou votre fournisseur agréé au sujet du type de système de réserve à utiliser.
Il est très important de sélectionner le débit d’oxygène qui vous a été prescrit. Ne modifiez pas le réglage de débit sans sans avis médical.
Le concentrateur d’oxygène peut être utilisé pendant le sommeil du patient sous la recommandation d’un clinicien agréé.
26 - FRE PN MN238-C4 A | Manuel d’utilisation
Page 27
Dispositifs de sécurité
Les informations suivantes ont pour objet de vous familiariser avec les dispositifs de sécurité du concentrateur d’oxygène VisionAire. Assurez-vous d’avoir lu et compris toutes les informations incluses dans ce manuel avant d’utiliser votre appareil. Pour toute question, contactez le fournisseur de l’appareil.
Étiquette d’avertissement de l’appareil et voyant d’alarme.
• Moteur du compresseur : la sortie du compresseur dispose d’une soupape de décompression calibrée à 280 kPa (40 psig). La sécurité thermique est assurée par un thermostat situé dans le bobinage du stator du compresseur (135 °C / 275 °F).
• Dysfonctionnement général : si l’une des conditions indiquées ci-dessous se produit, le témoin de dysfonctionnement général ( alarme sonore intermittente s’activera.
Cela comprend :
• une obstruction du débit d’oxygène comme un pincement ou un pli dans la canule de distribution, déclenchée par une pression élevée dans le réservoir de produit ;
• condition de pression élevée dans le réservoir de produit de l’appareil supérieure à 227 Kpa (33 psig) (±1) ;
• condition de pression faible dans le réservoir de produit de l’appareil inférieure à 34 Kpa (5 psig) (±1) ;
• température élevée de l’appareil supérieure à 135° C (275° F), déclenchée par une pression faible dans le réservoir de produit si l’interrupteur thermique situé dans le compresseur se déclenche (coupant ainsi le compresseur).
) s’allumera et une
VisionAire Family
• Moniteur d’oxygène : le moniteur d’oxygène détecte toute chute de concentration en dessous de 82 %. Si cela se produit le témoin d’avertissement de faible concentration d’oxygène ( condition de faible concentration en O2 persiste, une alarme sonore intermittente s’activera également, en plus du témoin.
• Panne d’alimentation : dans le cas où une panne de courant se produit alors que l’appareil est en marche, le témoin d’avertissement d’alimentation ( s’allumera et une alarme sonore intermittente s’activera. Voir l’illustration de l’étiquette.
• Filtre de produit : ltre de ≥ 10 μm
) s’allumera. Si la
)
PN MN238-C4 A | Manuel d’utilisation 27 - FRE
Page 28
VisionAire Family
Déballage de votre VisionAire
Vériez que tous les composants indiqués sont inclus
dans l’emballage. En l’absence d’un composant, contactez immédiatement votre fournisseur d’oxygène.
• Concentrateur d’oxygène stationnaire
Premier contact avec votre concentrateur d’oxygène VisionAire
Tout d’abord, familiarisez-vous avec les parties importantes de votre concentrateur d’oxygène VisionAire.
A. Interrupteur marche/arrêt (I/0) : démarre et arrête
l’appareil.
B. Bouton de réinitialisation du disjoncteur :
réinitialise l’appareil après un arrêt en cas de surcharge électrique
C. Compteur horaire : enregistre le nombre total
d’heures de fonctionnement de l’appareil.
Débitmètre/Bouton de réglage : contrôle et indique le
débit d’oxygène en litres par minute (l/m).
E. Sortie d’oxygène : permet un raccordement pour un
humidicateur (si nécessaire) ou une canule.
F. Poignées supérieures et latérales : permet un
transport pratique de l’appareil. G. Étiquette d’avertissement et d’alarmes H. Étiquette des caractéristiques techniques : indique
les caractéristiques électriques et le numéro de série. I. Cordon d’alimentation : permet le raccordement de
l’appareil à une prise secteur. J. Roulettes : quatre roulettes permettent de déplacer
facilement l’appareil, au besoin.
K. Logement du acon de l’humidicateur : zone où
poser le acon de l’humidicateur en option. L. Flacon de l’humidicateur (en option) M. Sortie d’oxygène du acon de l’humidicateur :
raccordement pour la tubulure/canule d’oxygène.
N. Sortie d’oxygène du acon de l’humidicateur O. Tubulure du acon de l’humidicateur P. Raccord du acon de l’humidicateur
Q. Tubulure / canule d’oxygène
AVERTISSEMENT : N’UTILISEZ PAS DE RALLONGE ÉLECTRIQUE AVEC CET APPAREIL ET VEILLEZ À NE PAS BRANCHER UN TROP GRAND NOMBRE D’APPAREILS SUR LA MÊME PRISE SECTEUR. L’UTILISATION D’UNE RALLONGE RISQUE D’AFFECTER LES PERFORMANCES DE L’APPAREIL. DE PLUS, SI VOUS EFFECTUEZ TROP DE BRANCHEMENTS SUR UNE MÊME PRISE, VOUS RISQUEZ DE PROVOQUER UNE SURCHARGE DU TABLEAU ÉLECTRIQUE ET DE DÉCLENCHER LE DISJONCTEUR OU DE PROVOQUER UN INCENDIE SI CELUI-CI NE FONCTIONNE PAS.
28 - FRE PN MN238-C4 A | Manuel d’utilisation
Page 29
VisionAire Family
F
E
C
L
F
D
B
K
A
G
G
J
Illustration 1a
I
N
M
H
O
Illustration 1 b
P
Q
PN MN238-C4 A | Manuel d’utilisation 29 - FRE
Page 30
VisionAire Family
Important !
Les instructions de sécurité se définissent comme indiqué ci-dessous :
AVERTISSEMENT : INFORMATIONS IMPORTANTES DE SÉCURITÉ RELATIVES AUX DANGERS POUVANT CAUSER DES BLESSURES GRAVES.
MISE EN GARDE : informations importantes pour éviter tout endommagement de la famille VisionAire.
Remarque : informations exigeant une attention particulière.
Indications d’utilisation
Contre-indications d’utilisation
AVERTISSEMENT : DANS CERTAINS CAS, L’UTILISATION D’OXYGÈNE SANS AVIS MÉDICAL PEUT S’AVÉRER DANGEREUSE. CET APPAREIL DOIT ÊTRE UTILISÉ UNIQUEMENT SUR PRESCRIPTION D’UN MÉDECIN.
AVERTISSEMENT : NE PAS UTILISER CET APPAREIL EN PRÉSENCE D’ANESTHÉSIQUES INFLAMMABLES.
AVERTISSEMENT : COMME TOUT DISPOSITIF ÉLECTRIQUE, CET APPAREIL PEUT NE PAS FONCTIONNER EN RAISON D’UNE COUPURE DE COURANT OU NÉCESSITER D’ÊTRE RÉPARÉ PAR UN TECHNICIEN QUALIFIÉ. LE CONCENTRATEUR D’OXYGÈNE NE PEUT ÊTRE UTILISÉ POUR LES UTILISATEURS QUI PRÉSENTERAIENT DES RÉACTIONS INDÉSIRABLES DÉCOULANT D’UNE INTERRUPTION TEMPORAIRE DE L’APPAREIL.
AVERTISSEMENT : IL EST TRÈS IMPORTANT DE SÉLECTIONNER LE DÉBIT D’OXYGÈNE QUI VOUS A ÉTÉ PRESCRIT. NE MODIFIEZ PAS LE RÉGLAGE DE DÉBIT SANS AVIS MÉDICAL.
AVERTISSEMENT : CONFORMÉMENT À LA LOI FÉDÉRALE DES ÉTATS-UNIS, LA VENTE ET LA LOCATION DE CET APPAREIL DOIVENT SE FAIRE SUR ORDONNANCE D’UN MÉDECIN OU D’UN PRESTATAIRE DE SOINS MUNI D’UNE LICENCE.
AVERTISSEMENT : CET APPAREIL NE DOIT PAS ÊTRE UTILISÉ À DES FINS DE RÉANIMATION. IL PEUT ÊTRE NÉCESSAIRE DE SURVEILLER LES UTILISATEURS DES SERVICES GÉRIATRIQUES, PÉDIATRIQUES OU TOUT AUTRE TYPE D’UTILISATEUR N’ÉTANT PAS EN MESURE DE COMMUNIQUER UN MALAISE LORS DE L’UTILISATION DE L’APPAREIL. LES UTILISATEURS SOUFFRANT DE DÉFICIENCE AUDITIVE ET/OU VISUELLE PEUVENT AVOIR BESOIN D’UNE ASSISTANCE POUR SURVEILLER LES ALARMES. SI VOUS VOUS SENTEZ MAL À L’AISE OU EN CAS D’URGENCE MÉDICALE, SOLLICITEZ IMMÉDIATEMENT UNE ASSISTANCE MÉDICALE.
Consignes de sécurité
AVERTISSEMENT : LISEZ ATTENTIVEMENT ET FAMILIARISEZ-VOUS AVEC LES INFORMATIONS DE SÉCURITÉ IMPORTANTES SUIVANTES AU SUJET DU CONCENTRATEUR D’OXYGÈNE VISIONAIRE INTENSITY.
AVERTISSEMENT : LISEZ ET ASSUREZ-VOUS D’AVOIR COMPRIS LES INFORMATIONS DE CE MANUEL AVANT D’UTILISER CET APPAREIL. SI VOUS NE COMPRENEZ PAS LES AVERTISSEMENTS ET LES INSTRUCTIONS, CONTACTEZ VOTRE FOURNISSEUR AGRÉÉ AVANT D’ESSAYER D’UTILISER L’APPAREIL AFIN D’ÉVITER TOUT RISQUE DE BLESSURE OU DE DÉGÂTS MATÉRIELS.
AVERTISSEMENT : FUMER EN UTILISANT DE L’OXYGÈNE CONSTITUE LA PREMIÈRE CAUSE DE BLESSURES ET DE DÉCÈS LIÉS AUX INCENDIES. VOUS DEVEZ IMPÉRATIVEMENT RESPECTER LES AVERTISSEMENTS DE SÉCURITÉ SUIVANTS :
AVERTISSEMENT : NE LAISSEZ PERSONNE FUMER ET N’UTILISEZ PAS DE BOUGIES OU DE FLAMMES NUES DANS LA MÊME PIÈCE QUE L’APPAREIL OU QUE LES ACCESSOIRES DE TRANSPORT D’OXYGÈNE.
AVERTISSEMENT : FUMER LORS DU PORT D’UNE CANULE À OXYGÈNE PEUT PROVOQUER DES BRÛLURES AU VISAGE OU MÊME LA MORT.
AVERTISSEMENT : LE FAIT DE RETIRER LA CANULE ET DE LA PLACER SUR UN VÊTEMENT, UN DRAP, UN SOFA OU TOUTE AUTRE MATIÈRE TEXTILE PEUT PROVOQUER UN EMBRASEMENT INSTANTANÉ EN CAS D’EXPOSITION À UNE CIGARETTE, À UNE SOURCE DE CHALEUR, À UNE ÉTINCELLE OU À UNE FLAMME NUE.
30 - FRE PN MN238-C4 A | Manuel d’utilisation
Page 31
VisionAire Family
AVERTISSEMENT : SI VOUS FUMEZ, VOUS DEVEZ TOUJOURS RESPECTER CES TROIS (3) ÉTAPES PRIMORDIALES APRÈS AVOIR UTILISÉ L’APPAREIL : ARRÊTEZ LE CONCENTRATEUR D’OXYGÈNE, RETIREZ LA CANULE ET QUITTEZ LA PIÈCE DANS LAQUELLE SE TROUVE L’APPAREIL.
AVERTISSEMENT : DES INDICATIONS « INTERDICTION DE FUMER - UTILISATION D’OXYGÈNE » DOIVENT ÊTRE AFFICHÉES DE FAÇON BIEN VISIBLE À VOTRE DOMICILE OU DANS TOUT LIEU OÙ DE L’OXYGÈNE EST UTILISÉ. LES UTILISATEURS ET LE PERSONNEL SOIGNANT DOIVENT ÊTRE INFORMÉS DES DANGERS LIÉS AU FAIT DE FUMER EN PRÉSENCE OU PENDANT L’UTILISATION D’OXYGÈNE MÉDICAL.
AVERTISSEMENT : CONFORMÉMENT À LA LOI FÉDÉRALE DES ÉTATS-UNIS, LA VENTE ET LA LOCATION DE CET APPAREIL DOIVENT SE FAIRE SUR ORDONNANCE D’UN MÉDECIN OU D’UN PRESTATAIRE DE SOINS MUNI D’UNE LICENCE.
AVERTISSEMENT : CET APPAREIL DISPENSE UN OXYGÈNE EXTRÊMEMENT PUR, SUSCEPTIBLE DE S’ENFLAMMER TRÈS VITE. NE FUMEZ PAS À PROXIMITÉ ET ÉVITEZ TOUTE FLAMME NUE DANS LA MÊME PIÈCE QUE (1) CET APPAREIL, OU QUE (2) TOUT AUTRE ACCESSOIRE DE TRANSPORT D’OXYGÈNE. LE NON-RESPECT DE CET AVERTISSEMENT PEUT AVOIR POUR CONSÉQUENCES UN INCENDIE, DES DÉGÂTS MATÉRIELS ET/OU DES BLESSURES GRAVES VOIRE MORTELLES.
AVERTISSEMENT : N’UTILISEZ PAS VOTRE CONCENTRATEUR D’OXYGÈNE EN PRÉSENCE DE GAZ INFLAMMABLES, AFIN D’ÉVITER TOUT RISQUE D’EMBRASEMENT RAPIDE SUSCEPTIBLE D’ENTRAÎNER DES DÉGÂTS MATÉRIELS, DES BLESSURES OU LA MORT.
AVERTISSEMENT : NE LAISSEZ JAMAIS UNE CANULE NASALE SUR UN VÊTEMENT, UN DRAP OU UN COUSSIN. SI L’APPAREIL EST ALLUMÉ SANS ÊTRE UTILISÉ, CES MATIÈRES RISQUENT DE S’ENFLAMMER AU CONTACT DE L’OXYGÈNE. ÉTEIGNEZ SYSTÉMATIQUEMENT LE CONCENTRATEUR D’OXYGÈNE LORSQU’IL N’EST PAS UTILISÉ.
AVERTISSEMENT : VEILLEZ À NE PAS UTILISER D’HUILE, DE GRAISSE, DE PRODUITS À BASE DE PÉTROLE OU D’AUTRES SUBSTANCES INFLAMMABLES AVEC LES ACCESSOIRES DE TRANSPORT D’OXYGÈNE OU AVEC LE CONCENTRATEUR D’OXYGÈNE. L’OXYGÈNE ACCÉLÈRE LA COMBUSTION DES SUBSTANCES INFLAMMABLES. SEULES DES LOTIONS ET DES CRÈMES À BASE D’EAU ET COMPATIBLES AVEC LA PRÉSENCE D’OXYGÈNE DOIVENT ÊTRE UTILISÉES.
AVERTISSEMENT : NE LUBRIFIEZ PAS LES RACCORDS, LES RACCORDEMENTS, LES TUBULURES, OU TOUT AUTRE ACCESSOIRE DU CONCENTRATEUR D’OXYGÈNE POUR ÉVITER TOUT RISQUE D’INCENDIE ET DE BRÛLURES.
AVERTISSEMENT : RISQUE DE CHOC ÉLECTRIQUE. ARRÊTEZ L’APPAREIL ET DÉBRANCHEZ LE CORDON D’ALIMENTATION DE LA PRISE ÉLECTRIQUE AVANT DE NETTOYER L’APPAREIL AFIN D’ÉVITER TOUT CHOC ÉLECTRIQUE ACCIDENTEL OU RISQUE DE BRÛLURE. SEUL LE FOURNISSEUR DE L’APPAREIL OU UN TECHNICIEN QUALIFIÉ EST HABILITÉ À RETIRER LES CAPOTS OU RÉPARER L’APPAREIL.
AVERTISSEMENT : PRENEZ TOUTES LES PRÉCAUTIONS NÉCESSAIRES POUR ÉVITER DE MOUILLER LE CONCENTRATEUR D’OXYGÈNE ET ASSUREZ-VOUS QU’AUCUN LIQUIDE NE PUISSE PÉNÉTRER À L’INTÉRIEUR DE L’APPAREIL, AFIN D’ÉVITER TOUT DYSFONCTIONNEMENT OU ARRÊT DE CELUI-CI, AINSI QU’UN RISQUE ACCRU DE CHOC ÉLECTRIQUE OU DE BRÛLURE.
AVERTISSEMENT : VEILLEZ À NE PAS UTILISER DE LIQUIDE DIRECTEMENT SUR L’APPAREIL. LES AGENTS CHIMIQUES À ÉVITER INCLUENT DE FAÇON NON EXHAUSTIVE : LES PRODUITS ALCOOLISÉS ET À BASE D’ALCOOL, LES PRODUITS À BASE DE CHLORE CONCENTRÉ (CHLORURE D’ÉTHYLÈNE) ET LES PRODUITS À BASE D’HUILE (PINE-SOL®, LESTOIL®). CES PRODUITS NE DOIVENT PAS ÊTRE UTILISÉS POUR NETTOYER LES PARTIES EN PLASTIQUE DU CONCENTRATEUR D’OXYGÈNE AU RISQUE DE LES ENDOMMAGER.
AVERTISSEMENT : NETTOYEZ LE BOÎTIER, LE PANNEAU DE COMMANDE ET LE CORDON D’ALIMENTATION UNIQUEMENT À L’AIDE D’UN DÉTERGENT DOUX APPLIQUÉ AU MOYEN D’UN CHIFFON OU D’UNE ÉPONGE LÉGÈREMENT HUMIDE, PUIS ESSUYEZ TOUTES LES SURFACES POUR LES SÉCHER. NE LAISSEZ AUCUN LIQUIDE PÉNÉTRER À L’INTÉRIEUR DE L’APPAREIL.
AVERTISSEMENT : LE CONCENTRATEUR D’OXYGÈNE NE DOIT PAS ÊTRE UTILISÉ LORSQU’IL EST ACCOLÉ À, OU EMPILÉ AVEC D’AUTRES ÉQUIPEMENTS. SI UN TEL POSITIONNEMENT NE PEUT PAS ÊTRE ÉVITÉ, IL CONVIENT DE SURVEILLER L’APPAREIL POUR S’ASSURER QU’IL FONCTIONNE CORRECTEMENT.
AVERTISSEMENT : PLACEZ TOUJOURS LES CORDONS D’ALIMENTATION ET LA TUBULURE D’ALIMENTATION EN OXYGÈNE DE SORTE À EMPÊCHER TOUTE PERSONNE DE TRÉBUCHER OU DE S’ÉTRANGLER ACCIDENTELLEMENT.
AVERTISSEMENT : AUCUNE MODIFICATION DE CET ÉQUIPEMENT N’EST AUTORISÉE.
AVERTISSEMENT : L’UTILISATION DE CÂBLES ET D’ADAPTATEURS AUTRES QUE CEUX SPÉCIFIÉS, À L’EXCEPTION DES CÂBLES ET ADAPTATEURS COMMERCIALISÉS PAR LE FABRICANT DE L’ÉQUIPEMENT ÉLECTRO-MÉDICAL POUR LE REMPLACEMENT DE COMPOSANTS INTERNES, RISQUE D’ENTRAÎNER UNE AUGMENTATION DES ÉMISSIONS ÉLECTROMAGNÉTIQUES OU UNE DIMINUTION DE L’IMMUNITÉ ÉLECTROMAGNÉTIQUE DU CONCENTRATEUR D’OXYGÈNE.
PN MN238-C4 A | Manuel d’utilisation 31 - FRE
Page 32
VisionAire Family
AVERTISSEMENT : LES CONDITIONS AMBIANTES PEUVENT AFFECTER LES PERFORMANCES DE L’APPAREIL. À UTILISER DANS UN ENVIRONNEMENT PROPRE ET EXEMPT D’ORGANISMES NUISIBLES.
AVERTISSEMENT : CET APPAREIL DOIT ÊTRE UTILISÉ UNIQUEMENT PAR LES UTILISATEURS FINAUX, LE PERSONNEL SOIGNANT OU DES TECHNICIENS QUALIFIÉS. NE PAS LAISSER À LA PORTÉE DES ENFANTS.
AVERTISSEMENT : POUR GARANTIR UNE QUANTITÉ D’OXYGÈNE THÉRAPEUTIQUE CORRESPONDANT À VOS BESOINS MÉDICAUX, LE VISIONAIRE DOIT ÊTRE UTILISÉ AVEC L’ASSOCIATION SPÉCIFIQUE DE PIÈCES ET D’ACCESSOIRES CORRESPONDANT AUX SPÉCIFICATIONS DU FABRICANT DU CONCENTRATEUR ET UTILISÉES LORSQUE VOS PARAMÈTRES ONT ÉTÉ DÉTERMINÉS.
AVERTISSEMENT : L’UTILISATION DE CET APPAREIL À UNE ALTITUDE, UNE TEMPÉRATURE OU UNE HUMIDITÉ RELATIVE DÉPASSANT LES VALEURS INDIQUÉES DANS CE MANUEL PEUT NUIRE AU DÉBIT ET À LA CONCENTRATION D’OXYGÈNE, ET DONC À LA QUALITÉ DE LA THÉRAPIE.
AVERTISSEMENT : LE RECOURS À DES ACCESSOIRES D’ADMINISTRATION D’OXYGÈNE NON PRÉVUS POUR UNE UTILISATION AVEC CE CONCENTRATEUR D’OXYGÈNE EST SUSCEPTIBLE DE NUIRE AUX PERFORMANCES DE L’APPAREIL. LES ACCESSOIRES RECOMMANDÉS SONT RÉFÉRENCÉS DANS CE MANUEL.
AVERTISSEMENT : UTILISEZ L’APPAREIL UNIQUEMENT À LA TENSION ÉLECTRIQUE INDIQUÉE SUR L’ÉTIQUETTE DES CARACTÉRISTIQUES TECHNIQUES APPOSÉE SUR L’APPAREIL.
AVERTISSEMENT : N’UTILISEZ PAS DE RALLONGE ÉLECTRIQUE AVEC CET APPAREIL ET VEILLEZ À NE PAS BRANCHER UN TROP GRAND NOMBRE D’APPAREILS SUR LA MÊME PRISE SECTEUR. L’UTILISATION D’UNE RALLONGE RISQUE D’AFFECTER LES PERFORMANCES DE L’APPAREIL. DE PLUS, SI VOUS EFFECTUEZ TROP DE BRANCHEMENTS SUR UNE MÊME PRISE, VOUS RISQUEZ DE PROVOQUER UNE SURCHARGE DU TABLEAU ÉLECTRIQUE ET DE DÉCLENCHER LE DISJONCTEUR OU DE PROVOQUER UN INCENDIE SI CELUI-CI NE FONCTIONNE PAS.
MISE EN GARDE : conformément à la loi fédérale des États-Unis, la vente et la location de cet appareil doivent se faire sur ordonnance d’un médecin ou d’un prestataire de soins muni d’une licence.
MISE EN GARDE : positionnez toujours le concentrateur de façon à ce que le cordon d’alimentation soit facilement accessible.
MISE EN GARDE : le concentrateur d’oxygène doit être placé à l’écart de toute source de fumée, de pollution ou de vapeurs.
MISE EN GARDE : assurez-vous d’utiliser le concentrateur en position verticale.
MISE EN GARDE : disposez toujours la tubulure d’alimentation en oxygène et les cordons d’alimentation de façon à prévenir tout risque de chute.
MISE EN GARDE : placez l’appareil à l’écart des rideaux, des draperies, des registres d’air chaud et des chauffages. Assurez-vous que l’appareil est posé sur une surface plane et que tous les côtés sont à au moins 30 cm (1 pied) d’un mur ou de tout autre obstacle. Ne mettez pas l’appareil dans un espace confiné. Choisissez un emplacement sans poussière ni fumée et à l’abri de la lumière du soleil directe. N’utilisez pas l’appareil à l’extérieur à moins que l’appareil soit branché à une prise protégée par un disjoncteur différentiel de fuite à la terre (DDFT).
MISE EN GARDE : n’utilisez pas cet appareil dans un espace fermé ou confiné offrant une ventilation potentiellement insuffisante. Ceci pourrait provoquer une surchauffe de l’appareil et une dégradation de ses performances.
MISE EN GARDE : ne laissez pas les orifices d’admission et d’évacuation de l’air s’obstruer. VEILLEZ À NE PAS faire tomber ou insérer des objets dans les ouvertures de l’appareil. Ceci pourrait provoquer une surchauffe du concentrateur d’oxygène et une dégradation de ses performances.
32 - FRE PN MN238-C4 A | Manuel d’utilisation
Page 33
VisionAire Family
MISE EN GARDE : le fabricant recommande de disposer d’une autre source d’alimentation en oxygène en cas de coupure de courant, de déclenchement d’alarme de l’appareil ou de panne mécanique. Consultez votre médecin ou votre fournisseur agréé au sujet du type de système de réserve à utiliser.
MISE EN GARDE : il est très important de sélectionner le débit d’oxygène qui vous a été prescrit. Ne modifiez pas le réglage de débit sans y avoir été invité par un médecin.
MISE EN GARDE : le concentrateur d’oxygène peut être utilisé pendant le sommeil du patient sous la recommandation d’un clinicien agréé.
MISE EN GARDE : l’utilisation ou le stockage du concentrateur d’oxygène dans des températures autres que celles recommandées peut affecter les performances de l’appareil. Reportez-vous à la section Caractéristiques techniques de ce manuel pour consulter les limites de température relatives au stockage et au fonctionnement de l’appareil.
MISE EN GARDE : si une alarme se déclenche ou si vous constatez que le concentrateur d’oxygène ne fonctionne pas correctement, reportez-vous à la section dépannage figurant dans ce manuel. Si vous ne parvenez pas à résoudre le problème, consultez votre fournisseur agréé.
MISE EN GARDE : si l’alarme sonore est faible ou inaudible, consultez immédiatement le fournisseur de l’appareil.
MISE EN GARDE : si la tubulure du flacon de l’humidificateur n’est pas correctement raccordée au raccord du flacon de l’humidificateur ou à la sortie d’oxygène, une fuite d’oxygène peut se produire.
MISE EN GARDE : normalement, il n’est pas nécessaire de régler le débitmètre sur votre appareil. Si vous tournez le bouton de réglage du débitmètre dans le sens des aiguilles d’une montre, vous réduirez et risquez de couper de débit d’oxygène produit par votre appareil. Pour votre confort, le débitmètre est marqué par incréments de ½ l/min. Pour les appareils équipés de l’option de débitmètre de 2 l/min, le débitmètre est marqué par incréments de 1/8 l/min pour les réglages de débit jusqu’à 2 l/min.
Remarque : la canule ne doit pas être pliée et peut être utilisée pour une longueur totale de 7,6 m (25 pieds) max.
Vérifiez que la canule est entièrement insérée et convenablement positionnée. Vous devriez entendre ou sentir le flux d’oxygène vers les branches de la canule nasale. Si l’oxygène semble ne pas circuler, vérifiez d’abord que la bille du débitmètre indique un débit. Puis, mettez le bout de la canule dans un verre d’eau ; si des bulles sortent de la canule, de l’oxygène circule. En l’absence de bulles, consultez la section dépannage de ce manuel.
Remarque : pour une utilisation correcte de la canule, reportez-vous toujours aux instructions du fabricant. Remplacez la canule jetable conformément aux recommandations du fabricant de la canule ou du fournisseur de l’équipement. D’autres fournitures sont disponibles auprès de votre fournisseur.
Remarque : le concentrateur d’oxygène VisionAire doit fonctionner pendant au moins cinq minutes à 2 l/min avant d’utiliser l’appareil.
Le VisionAire peut être utilisé par deux utilisateurs, tant que le débit combiné est d’au moins 2 l/min et ne dépasse pas le débit maximal du concentrateur.
À l’attention du fournisseur de l’équipement : les accessoires d’administration d’oxygène recommandés avec le concentrateur d’oxygène VisionAire sont les suivants :
• Flacon de l’humidificateur : Réf. HU003-1
• Canule nasale et tubulure de 2,1 mètres (7 pieds) de long (6 l/min max) : Réf. CU002-1
• Kit OxySafe : Réf. 20629671 Remarque : le fabricant ne recommande pas de
stériliser cet équipement. Remarque : en cas de longue période d’inutilisation,
l’appareil doit fonctionner plusieurs minutes avant que l’alarme de panne d’alimentation ne puisse s’activer.
Remarque : le concentrateur libère de l’air chaud par le bas de l’appareil, ce qui peut décolorer de façon permanente les planchers sensibles à la température comme le vinyle. Le concentrateur ne doit pas être utilisé sur des planchers sujets aux taches de chaleur. Le fabricant décline toute responsabilité en cas de décoloration des planchers.
Remarque : suivez toutes les instructions du fabricant afin d’éviter l’annulation de la garantie.
Remarque : ne tentez pas de réaliser des opérations de maintenance autres que les solutions de dépannage détaillées dans ce manuel.
PN MN238-C4 A | Manuel d’utilisation 33 - FRE
Page 34
VisionAire Family
Remarque : les équipements portables et mobiles de communication par radiofréquences (RF) peuvent affecter le fonctionnement des équipements médicaux électriques.
Remarque : l’utilisation du concentrateur d’oxygène ne risque en aucun cas d’épuiser tout l’oxygène de la pièce.
Remarque : à l’attention du fournisseur de l’équipement : les accessoires d’administration d’oxygène recommandés avec cet appareil sont les suivants :
• canule nasale et tube de 2,1 mètres (7 pieds) de long : référence 5408-SEQ
• pour consulter des accessoires recommandés supplémentaires, reportez-vous au catalogue d’accessoires (référence ML-LOX0010) disponible à l’adresse www.CAIREmedical.com.
Un coupe-feu est requis pour toute utilisation avec une canule.
• CAIRE offre l’OxySafe™. Cet accessoire est prévu pour être utilisé en association avec le concentrateur d’oxygène. L’OxySafe™ est un fusible thermique conçu pour stopper l’écoulement de gaz en cas de combustion remontant depuis la canule ou la tubulure d’oxygène. Il se place sur le circuit de la canule nasale/ de la tubulure d’oxygène, entre le patient et la sortie d’oxygène du système VisionAire.
Pour une utilisation correcte de l’OxySafe, reportez­vous toujours aux instructions du fabricant (fournies avec chaque kit OxySafe).
CAIRE dispose d’un kit OxySafe™ comportant un ensemble canule, l’OxySafe™ et une tubulure de 2,1 m (7 po) : PN 20629671.
AVERTISSEMENT : CET APPAREIL PRODUIT DE L’OXYGÈNE À CONCENTRATION ÉLEVÉE, FAVORISANT UNE COMBUSTION RAPIDE. NE PAS PERMETTRE DE FUMER OU DES FLAMMES NUES DANS LA MÊME CHAMBRE : (1) CET APPAREIL OU (2) TOUT ACCESSOIRE CONTENANT DE L’OXYGÈNE. NE PAS UTILISER DE PRODUITS À BASE D’HUILE, DE GRAISSE OU DE PÉTROLE SUR OU À PROXIMITÉ DE L’UNITÉ. DÉCONNECTER LE CORDON D’ALIMENTATION DE LA PRISE ÉLECTRIQUE AVANT DE NETTOYER OU DE FAIRE L’ENTRETIEN DE L’UNITÉ.
RISQUE DE CHOC ÉLECTRIQUE. NE PAS ENLEVER LES COUVERCLES LORSQUE L’UNITÉ EST BRANCHÉE. SEULS VOTRE FOURNISSEUR D’ÉQUIPEMENT OU UN TECHNICIEN DE SERVICE QUALIFIÉ DEVRAIT ENLEVER LES COUVERCLES OU FAIRE L’ENTRETIEN DE L’UNITÉ.
34 - FRE PN MN238-C4 A | Manuel d’utilisation
Page 35
VisionAire Family
Caractéristiques techniques
VisionAire
Débits* 1 l/min – 5 l/min
±10 % du réglage indiqué, ou 200 mL, selon la valeur la plus élevée*
Dimensions 35,8 cm L x 29,2 cm D x 52,8 cm H
(14,1 po L x 11,5 po D x 20,8 cm H)
Poids 13,6 kg (30 livres) Niveau de pression acoustiqu 45 dBa (1 à 5 l/min) Consommation 290 W Concentration en O2 90 % (+ 5,5 % / - 3 %) Exigences électriques 115 V CA / 60 Hz, 3,0 A
230 V CA / 50 Hz, 1,5 A 230 V CA / 60 Hz, 1,5 A
Conditions de fonctionnement* 5 °C à 40 °C (41 °F à 104 °F) à des altitudes jusqu’à 3 048 m (10 000 pieds)
Altitude -381 à 3048 m (-1250 à 10,000 ft) (testé à 700 - 1060 hPa) Conditions de stockage -25 °C – 70 °C (-13 °F – 158 °F)
* À une plage de pression atmosphérique de 700 hPa à 1060 hPa et une température de 21 °C (70 °F)
La durée de vie estimée de l’appareil est d’au moins cinq ans.
au-dessus du niveau de la mer. 15–95 % d’humidité (sans condensation)
La maintenance est la responsabilité du fournisseur et sera suivie par le fournisseur.
AVERTISSEMENT : L’UTILISATION DE CET APPAREIL À UNE ALTITUDE, UNE TEMPÉRATURE OU UNE HUMIDITÉ RELATIVE DÉPASSANT LES VALEURS INDIQUÉES DANS CE MANUEL PEUT NUIRE AU DÉBIT ET À LA CONCENTRATION D’OXYGÈNE, ET DONC À LA QUALITÉ DE LA THÉRAPIE.
AVERTISSEMENT : L’UTILISATION DE CET APPAREIL À UNE ALTITUDE SUPÉRIEURE À 3 048 M (10 000 PIEDS) OU À UNE TEMPÉRATURE SORTANT DE LA PLAGE DE 5 °C À 40 °C (41 °F À 104 °F) OU UNE HUMIDITÉ RELATIVE AU-DESSUS DE 95 % PEUT NUIRE AU DÉBIT ET AU POURCENTAGE D’OXYGÈNE, ET DONC À LA QUALITÉ DE LA THÉRAPIE.
PN MN238-C4 A | Manuel d’utilisation 35 - FRE
Page 36
VisionAire Family
Instructions d’utilisation
Lisez les informations suivantes avant d’utiliser votre concentrateur d’oxygène.
Remarque : le concentrateur libère de l’air chaud par le bas de l’appareil, ce qui peut décolorer de façon permanente les planchers sensibles à la température comme le vinyle. Le concentrateur ne doit pas être utilisé sur des planchers sujets aux taches de chaleur. Le fabricant décline toute responsabilité en cas de décoloration des planchers.
Flacon de l’humidificateur (en option)
Si une humidication est nécessaire pour votre
oxygénothérapie, suivez les étapes suivantes chaque
fois que vous remplissez ou nettoyez l’humidicateur,
qui peut avoir été initialement installé pour votre utilisation.
1. Retirez le acon de l’humidicateur de son
logement.
2. Ouvrez le acon de l’humidicateur. Si vous avez un acon pré-rempli, passez cette étape. Passez à
l’étape 5.
3. Remplissez le acon de l’humidicateur avec
de l’eau tiède ou froide (de l’eau distillée de préférence) jusqu’à la ligne de remplissage indiquée sur la bouteille. NE PAS TROP REMPLIR.
4. Rebranchez le capot supérieur au acon de l’humidicateur.
5. Placez le acon de l’humidicateur dans son
logement à l’arrière du concentrateur et raccordez
la tubulure du acon de l’humidicateur à la sortie d’oxygène et au raccord du acon de l’humidicateur.
MISE EN GARDE : si la tubulure du flacon de l’humidificateur n’est pas correctement raccordée au raccord du flacon de l’humidificateur ou à la sortie d’oxygène, une fuite d’oxygène peut se produire.
Raccord de la canule
Raccordez la tubulure et la canule à la sortie d’oxygène de l’appareil, ou à la sortie d’oxygène de
l’humidicateur en option.
Remarque : La canule ne doit pas être pliée et peut être utilisée pour une longueur totale de 7,6 m (25 pieds) max.
Vérifiez que la canule est entièrement insérée et convenablement positionnée. Vous devriez entendre ou sentir le flux d’oxygène vers les branches de la canule nasale. Si l’oxygène semble ne pas circuler, vérifiez d’abord que la bille du débitmètre indique un débit. Puis, mettez le bout de la canule dans un verre d’eau ; si des bulles sortent de la canule, de l’oxygène circule. En l’absence de bulles, consultez la section dépannage de ce manuel.
Pour une utilisation correcte de la canule, reportez­vous toujours aux instructions du fabricant. Remplacez la canule jetable conformément aux recommandations du fabricant de la canule ou du fournisseur de l’équipement. D’autres fournitures sont disponibles auprès de votre fournisseur.
Démarrage du concentrateur
AVERTISSEMENT : DES INDICATIONS « INTERDICTION DE FUMER ­UTILISATION D’OXYGÈNE » DOIVENT ÊTRE AFFICHÉES DE FAÇON BIEN VISIBLE À VOTRE DOMICILE OU DANS TOUT LIEU OÙ DE L’OXYGÈNE EST UTILISÉ. LES UTILISATEURS ET LE PERSONNEL SOIGNANT DOIVENT ÊTRE INFORMÉS DES DANGERS LIÉS AU FAIT DE FUMER EN PRÉSENCE OU PENDANT L’UTILISATION D’OXYGÈNE MÉDICAL.
AVERTISSEMENT : N’UTILISEZ PAS DE RALLONGE ÉLECTRIQUE AVEC CET APPAREIL ET VEILLEZ À NE PAS BRANCHER UN TROP GRAND NOMBRE D’APPAREILS SUR LA MÊME PRISE SECTEUR. L’UTILISATION D’UNE RALLONGE RISQUE D’AFFECTER LES PERFORMANCES DE L’APPAREIL. DE PLUS, SI VOUS EFFECTUEZ TROP DE BRANCHEMENTS SUR UNE MÊME PRISE, VOUS RISQUEZ DE PROVOQUER UNE SURCHARGE DU TABLEAU ÉLECTRIQUE ET DE DÉCLENCHER LE DISJONCTEUR OU DE PROVOQUER UN INCENDIE SI CELUI-CI NE FONCTIONNE PAS.
36 - FRE PN MN238-C4 A | Manuel d’utilisation
Page 37
VisionAire Family
MISE EN GARDE : Placez l’appareil à l’écart des rideaux, des draperies, des bouches de chaleur chaud et des chauffages. Assurez-vous que l’appareil est posé sur une surface plane et que tous les côtés sont à au moins 30 cm (1 pied) d’un mur ou de tout autre obstacle. Ne mettez pas l’appareil dans un espace confiné. Choisissez un emplacement sans poussière ni fumée et à l’abri de la lumière du soleil directe. N’utilisez pas l’appareil à l’extérieur à moins que l’appareil soit branché à une prise protégée par un disjoncteur différentiel de fuite à la terre (DDFT).
MISE EN GARDE : n’utilisez pas cet appareil dans un espace fermé ou confiné offrant une ventilation potentiellement insuffisante. Ceci pourrait provoquer une surchauffe de l’appareil et une dégradation de ses performances.
MISE EN GARDE : ne laissez pas les orifices d’admission et d’évacuation de l’air s’obstruer. VEILLEZ À NE PAS faire tomber ou insérer des objets dans les ouvertures de l’appareil. Cela peut provoquer une
MISE EN GARDE : surchauffe du concentrateur d’oxygène et une diminution des performances.
1. Placez l’appareil près de la prise secteur dans la pièce où vous passez le plus de temps.
2. Insérez la che du cordon d’alimentation dans la
prise secteur.
3. Mettez l’interrupteur d’alimentation I/0 sur « I » pour allumer l’appareil. Une alarme sonore se fera entendre très clairement pendant environ 5 secondes.
4. Le témoin de faible concentration en oxygène reste allumé pendant quelques minutes et jusqu’à ce que la concentration en oxygène atteigne les exigences minimales de concentration, (uniquement pour les appareils équipés d’un moniteur d’oxygène)
5. Pour régler le débit d’oxygène d’appoint, tournez le bouton de réglage du débitmètre vers la gauche ou la droite jusqu’à ce que la bille à l’intérieur du débitmètre se centre sur le numéro de ligne de débit prescrit par votre médecin. Pour voir le débitmètre sous le bon angle, notez que la ligne arrière et la ligne avant avec le numéro doivent sembler ne faire qu’une ligne.
6. Le concentrateur est maintenant prêt à être utilisé.
MISE EN GARDE : le fabricant recommande de disposer d’une autre source d’alimentation en oxygène en cas de coupure de courant, de déclenchement d’alarme de l’appareil ou de panne mécanique. Consultez votre médecin ou votre fournisseur agréé au sujet du type de système de réserve à utiliser.
MISE EN GARDE : il est très important de sélectionner le débit d’oxygène qui vous a été prescrit. Ne modifiez pas le réglage de débit sans y avoir été invité par un médecin.
MISE EN GARDE : le concentrateur d’oxygène peut être utilisé pendant le sommeil du patient sous la recommandation d’un clinicien qualifié
MISE EN GARDE : assurez-vous d’utiliser le concentrateur en position verticale.
Remarque : vous obtiendrez une concentration optimale en oxygène dans les 10 minutes après allumage de l’appareil (90 % de la concentration est obtenue après environ 5 minutes).
Normalement, il n’est pas nécessaire de régler le débitmètre sur votre appareil. Si vous tournez le bouton de réglage du débitmètre dans le sens des aiguilles d’une montre, vous réduirez et risquez de couper de débit d’oxygène produit par votre appareil.
Arrêt du concentrateur
Mettez l’interrupteur d’alimentation I/0 sur « 0 » pour arrêter l’appareil.
PN MN238-C4 A | Manuel d’utilisation 37 - FRE
Page 38
VisionAire Family
Nettoyage, entretien et maintenance correcte
Boîtier
AVERTISSEMENT : RISQUE DE CHOC ÉLECTRIQUE. ARRÊTEZ L’APPAREIL ET DÉBRANCHEZ LE CORDON D’ALIMENTATION DE LA PRISE ÉLECTRIQUE AVANT DE NETTOYER L’APPAREIL AFIN D’ÉVITER TOUT CHOC ÉLECTRIQUE ACCIDENTEL OU RISQUE DE BRÛLURE. SEUL LE FOURNISSEUR DE L’APPAREIL OU UN TECHNICIEN QUALIFIÉ EST HABILITÉ À RETIRER LES CAPOTS OU RÉPARER L’APPAREIL.
AVERTISSEMENT : PRENEZ TOUTES LES PRÉCAUTIONS NÉCESSAIRES POUR ÉVITER DE MOUILLER LE CONCENTRATEUR D’OXYGÈNE ET ASSUREZ-VOUS QU’AUCUN LIQUIDE NE PUISSE PÉNÉTRER À L’INTÉRIEUR DE L’APPAREIL, AFIN D’ÉVITER TOUT DYSFONCTIONNEMENT OU ARRÊT DE CELUI-CI, AINSI QU’UN RISQUE ACCRU DE CHOC ÉLECTRIQUE OU DE BRÛLURE.
AVERTISSEMENT : VEILLEZ À NE PAS UTILISER D’HUILE, DE GRAISSE, DE PRODUITS À BASE DE PÉTROLE OU D’AUTRES SUBSTANCES INFLAMMABLES AVEC LES ACCESSOIRES DE TRANSPORT D’OXYGÈNE OU AVEC LE CONCENTRATEUR D’OXYGÈNE. L’OXYGÈNE ACCÉLÈRE LA COMBUSTION DES SUBSTANCES INFLAMMABLES.
AVERTISSEMENT : SEULES DES LOTIONS ET DES CRÈMES À BASE D’EAU ET COMPATIBLES AVEC LA PRÉSENCE D’OXYGÈNE DOIVENT ÊTRE UTILISÉES AVANT ET PENDANT LA THÉRAPIE. N’UTILISEZ JAMAIS DE LOTIONS OU CRÈMES À BASE DE PÉTROLE OU D’HUILE POUR ÉVITER TOUT RISQUE D’INCENDIE OU DE BRÛLURES.
AVERTISSEMENT : VEILLEZ À NE PAS UTILISER DE LIQUIDE DIRECTEMENT SUR L’APPAREIL. LES AGENTS CHIMIQUES À ÉVITER INCLUENT DE FAÇON NON EXHAUSTIVE : LES PRODUITS ALCOOLISÉS ET À BASE D’ALCOOL, LES PRODUITS À BASE DE CHLORE CONCENTRÉ (CHLORURE D’ÉTHYLÈNE) ET LES PRODUITS À BASE D’HUILE (PINE-SOL®, LESTOIL®). CES PRODUITS NE DOIVENT PAS ÊTRE UTILISÉS POUR NETTOYER LES PARTIES EN PLASTIQUE DU CONCENTRATEUR D’OXYGÈNE AU RISQUE DE LES ENDOMMAGER. NETTOYEZ LE BOÎTIER, LE PANNEAU DE COMMANDE ET LE CORDON D’ALIMENTATION UNIQUEMENT À L’AIDE D’UN DÉTERGENT DOUX APPLIQUÉ AU MOYEN D’UN CHIFFON OU D’UNE ÉPONGE LÉGÈREMENT HUMIDE, PUIS ESSUYEZ TOUTES LES SURFACES POUR LES SÉCHER. NE LAISSEZ AUCUN LIQUIDE PÉNÉTRER À L’INTÉRIEUR DE L’APPAREIL.
AVERTISSEMENT : NETTOYEZ LE BOÎTIER, LE PANNEAU DE COMMANDE ET LE CORDON D’ALIMENTATION UNIQUEMENT AVEC UN CHIFFON OU UNE ÉPONGE HUMIDE (PAS MOUILLÉE) IMBIBÉE D’UN PRODUIT DE NETTOYAGE DOUX, PUIS ESSUYEZ TOUTES LES SURFACES. NE LAISSEZ AUCUN LIQUIDE PÉNÉTRER À L’INTÉRIEUR DU CONCENTRATEUR. FAITES PARTICULIÈREMENT ATTENTION À LA SORTIE D’OXYGÈNE DESTINÉE AU RACCORD DE LA CANULE, AFIN QU’ELLE NE CONTIENNE NI POUSSIÈRE, NI EAU, NI PARTICULES.
Remarque : pour une utilisation correcte de la canule, reportez-vous toujours aux instructions du fabricant. Remplacez la canule jetable conformément aux recommandations du fabricant de la canule ou du fournisseur de l’équipement. D’autres fournitures sont disponibles auprès de votre fournisseur.
Remarque : le fabricant ne recommande pas de stériliser cet équipement.
Utilisez un détergent doux avec un chiffon ou une éponge humide pour nettoyer l’extérieur du concentrateur, puis essuyez toutes les surfaces pour les sécher. NE laissez AUCUN liquide pénétrer à l’intérieur de l’appareil. Le boîtier de l’appareil doit être nettoyé au minimum entre chaque utilisateur.
Flacon de l’humidificateur (en option) :
• Vériez quotidiennement le niveau d’eau et ajoutez
de l’eau au besoin.
• Pour nettoyer et désinfecter l’humidicateur, suivez
les instructions du fournisseur de l’appareil, ou les
instructions fournies avec le acon de l’humidicateur.
38 - FRE PN MN238-C4 A | Manuel d’utilisation
Page 39
Conditions d’alarme
Toutes les alarmes sont des alarmes de faible priorité.
Alarme Message Action
Dysfonctionnement général : témoin jaune
et alarme sonore intermittente
Moniteur d’oxygène : témoin jaune et alarme sonore intermittente
Panne d’alimentation : témoin jaune et alarme sonore intermittente
pression élevée dans le réservoir de produit
OU faible pression du
réservoir de produit OU température élevée de
l’appareil Concentration faible en
oxygène
panne d’alimentation Assurez-vous que l’appareil est branché sur une prise
Assurez-vous que le débitmètre est ouvert au débit minimum ou au-dessus. Assurez-vous que la canule n’est pas pliée ou obstruée. Retirez tous les appareils connectés en aval de la sortie de l’appareil. Assurez-vous d’avoir au moins 300 mm (12 po) de dégagement de tous les côtés et que les admissions ne sont pas obstruées. Assurez-vous que le filtre à particules brutes d’admission est propre et dégagé. Assurez-vous que l’appareil est dans la plage de température de fonctionnement. Si le problème persiste, contactez le fournisseur de l’appareil pour réparation.
Contactez le fournisseur de l’appareil pour réparation.
secteur connue pour être en bon état de marche. Assurez­vous que l’interrupteur du disjoncteur est enfoncé. Si le problème persiste, contactez le fournisseur de l’appareil pour réparation.
VisionAire Family
PN MN238-C4 A | Manuel d’utilisation 39 - FRE
Page 40
VisionAire Family
Dépannage
Si votre concentrateur en oxygène VisionAire ne fonctionne pas correctement, consultez le tableau dans les pages suivantes pour les causes possibles et les solutions, et si besoin, consultez le fournisseur de l’appareil.
Si vous ne parvenez pas à faire fonctionner l’appareil, raccordez votre canule nasale, votre masque, ou tout autre accessoire à une réserve d’oxygène d’appoint.
Remarque : ne tentez pas de réaliser des opérations de maintenance autres que les solutions de dépannage détaillées dans ce manuel. La maintenance est la responsabilité du fournisseur et sera suivie par le fournisseur.
Remarque : suivez toutes les instructions du fabricant afin d’éviter l’annulation de la garantie. Remarque : en cas de longue période d’inutilisation, l’appareil doit fonctionner pendant au moins 15 minutes avant
que l’alarme de panne d’alimentation ne puisse s’activer.
Problème Cause probable Solution
L’appareil ne fonctionne pas. Une panne d’alimentation déclenche une alarme sonore.
Débit d’oxygène limité. Flacon de l’humidificateur sale
De la condensation s’accumule dans la tubulure d’oxygène lorsque vous utilisez le flacon de l’humidificateur.
Une alarme intermittente se fait entendre.
Le cordon d’alimentation n’est pas raccordé à la prise secteur.
Prise secteur non alimentée. Vérifiez la source d’alimentation, l’interrupteur
Le disjoncteur du concentrateur d’oxygène est activé.
ou obstrué.
Canule nasale, masque, cathéter, et/ou tube de distribution d’oxygène ou autre accessoire défectueux.
Tubulure de la canule desserrée.
Appareil mal ventilé. Température de fonctionnement élevée.
Consultez la section « Dispositifs de sécurité » pour une description des témoins sonores.
Vérifiez que le cordon d’alimentation est correctement branché sur la prise secteur.
mural, le fusible, ou le disjoncteur de la maison. Appuyez sur le bouton de réinitialisation du
disjoncteur (sans le maintenir) à l’avant de l’appareil.
Si le disjoncteur se déclenche de nouveau ou l’alarme continue de se faire entendre après l’allumage de l’appareil, contactez le fournisseur de l’appareil.
Retirez le flacon de l’humidificateur de la sortie d’oxygène. Si le débit revient, nettoyez ou remplacez le flacon de l’humidificateur.
Retirez et vérifiez les accessoires à la recherche de plis ou d’obstructions. Réparez si nécessaire.
Vérifiez le raccordement de la tubulure de la canule au niveau du panneau de commande.
Assurez-vous que l’appareil est à l’écart des rideaux, des draperies, des registres d’air chaud, des chauffages et des cheminées. Assurez-vous que tous les côtés de l’appareil sont à au moins 30,5 cm (12 pouces) d’un mur ou de tout autre obstacle. Ne mettez pas l’appareil dans un espace confiné.
Remplissez le flacon de l’humidificateur avec de l’eau FROIDE.
NE PAS TROP REMPLIR. Laissez la tubulure d’oxygène sécher, ou remplacez-la.
Mettez l’interrupteur d’alimentation I/0 sur 0, utilisez votre réserve d’oxygène et consultez immédiatement le fournisseur de l’appareil.
40 - FRE PN MN238-C4 A | Manuel d’utilisation
Page 41
VisionAire Family
Problème Cause probable Solution
Le témoin du moniteur d’oxygène reste allumé (jaune).
Tous les autres problèmes. Mettez l’interrupteur d’alimentation I/0 sur 0,
La concentration en oxygène est ≤ 82 %.
Accessoires
Mettez l’interrupteur d’alimentation I/0 sur 0, utilisez votre réserve d’oxygène (le cas échéant) et consultez immédiatement le fournisseur de l’appareil.
utilisez votre réserve d’oxygène et consultez immédiatement le fournisseur de l’appareil.
An de garantir des performances et une sécurité
optimales, utilisez uniquement les accessoires ci-dessous fournis par CAIRE par le biais de votre fournisseur d’oxygène. L’utilisation de produits différents pourrait nuire aux performances et/ou à la sécurité du concentrateur.
Accessoires standard de la famille VisionAire
Flacon de l’humidificateur (6–15 LPM)
Canule, 7,6 m (25 pieds) (6 l/min max)
Tubulure du flacon de l’humidificateur
Référence – HU003-1
Référence – CU002-4
Référence – TU255-1
Raccord du flacon de l’humidificateur
Tubulure d’oxygène, 7,6 m (25 pieds)
Connecteur de la tubulure/canule
Kit OxySafe (débit 25 l/ min max. Pression 6 bars/87 PSIG max.)
Remarque : si elles sont reprises ci-dessus, des options supplémentaires peuvent être disponibles pour les cordons d’alimentation spécifiques aux pays. Contactez CAIRE ou votre fournisseur d’oxygène si vous souhaitez commander d’autres accessoires.
Référence – F0655-1
Référence – CU004-3
Référence – CU009-1
Référence – 20629671
PN MN238-C4 A | Manuel d’utilisation 41 - FRE
Page 42
VisionAire Family
Test CEM
L’appareil médical doit faire l’objet de précautions particulières concernant la compatibilité électromagnétique (CEM) et doit être installé et mis en service conformément aux informations CEM fournies dans cette section.
Directives et déclaration du fabricant - Émissions électromagnétiques
Le système VisionAire est prévu pour une utilisation dans les environnements électromagnétiques spécifiés ci-dessous. Le client ou l’utilisateur du système VisionAire doit s’assurer que celui-ci est utilisé dans un environnement de ce type.
Test d’émissions Conformité Environnement électromagnétique - Directives
Émissions RF CISPR 11
Émissions RF CISPR 11 Émissions de courant
harmonique CEI 61000-3-2
Fluctuations de tension/ papillotement CEI 61000-3-3
Groupe 1
Classe B
Conforme
Conforme
Distances de séparation recommandées entre les équipements de communication RF mobiles et portables et les appareils VisionAire
Le VisionAire est prévu pour une utilisation dans un environnement dans lequel les perturbations dues aux radiofréquences rayonnées sont contrôlées. Le client ou l’utilisateur du système VisionAire peut prévenir les perturbations électromagnétiques en respectant les distances minimum recommandées ci-dessous pour la séparation entre les équipements de communication RF portables et mobiles (émetteurs) et l’appareil VisionAire, en fonction de la puissance maximum de sortie de l’équipement de communication.
Puissance
nominale de sortie
maximum de
l’émetteur
W
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
Pour les émetteurs avec une puissance nominale de sortie maximum ne figurant pas dans le tableau ci-dessus, la distance de séparation recommandée d en mètres (m) peut être déterminée à l’aide de l’équation applicable à la fréquence de l’émetteur, où P correspond à la puissance nominale de sortie maximum de l’émetteur en watts (W), indiquée par le fabricant de l’émetteur.
REMARQUE 1 : à 80 MHz et 800 MHz, la distance de séparation correspondant à la plage de fréquences supérieure s’applique.
REMARQUE 2 : il est possible que ces directives ne s’appliquent pas à toutes les situations. La propagation électromagnétique dépend des phénomènes d’absorption et de réflexion induits par les structures, les objets et les personnes.
Distance de séparation en fonction de la fréquence de l’émetteur
de 150 kHz à 80 MHz
d = 1,2√P
Le système VisionAire utilise de l’énergie RF uniquement pour son fonctionnement interne. Par conséquent, ses émissions RF sont très faibles et ne sont pas susceptibles de provoquer des interférences avec des équipements électroniques situés à proximité.
Le système VisionAire convient à une utilisation dans tous les établissements, y compris les habitations et les établissements directement connectés au réseau d’alimentation public basse tension qui alimente les bâtiments à usage d’habitation.
m
de 80 MHz à 800 MHz
d = 1,2√P
de 800 MHz à 2,5 GHz
d = 1,2√P
42 - FRE PN MN238-C4 A | Manuel d’utilisation
Page 43
VisionAire Family
Directives et déclaration du fabricant - Immunité électromagnétique
Le système VisionAire est prévu pour une utilisation dans les environnements électromagnétiques spécifiés ci-dessous. Le client ou l’utilisateur du système VisionAire doit s’assurer que celui-ci est utilisé dans un environnement de ce type.
Test d’immunité Niveau de test
CEI 60601
Environnement électromagnétique ­Directives
Contact ±6 kV Air ±8 kV
CEI 61000-4-2
Transitoires électriques rapides/
±2 kV pour les lignes d’alimentation
Salves CEI 61000-4-4
±1 kV pour les lignes d’entrée/sortie
±1 kV entre fils de ligne
Surtension CEI 61000-4-5
±2 kV entre un fil et la terre
<5 % U
T
(baisse >95 % en U pendant 0,5 cycle
Niveau de conformité
Contact ±6 kV Air ±8 kV
±2 kV pour les lignes d’alimentation
S/O
±1 kV entre fils de ligne
±2 kV entre un fil et la terre
<5 % U (baisse >95 % en U
)
T
pendant 0,5 cycle
Environnement électromagnétique ­Directives
Les sols doivent être en bois, en ciment ou carrelés. Si les sols sont revêtus d’un matériau synthétique, le taux d’humidité relative doit être d’au moins 30 %.
La qualité de l’alimentation secteur doit être celle d’un environnement commercial ou médical standard.
La qualité de l’alimentation secteur doit être celle d’un environnement commercial ou médical standard.
T
)
T
Creux de tension, coupures brèves et variations de tension sur les circuits d’alimentation électrique CEI 61000-4-11
Champ magnétique à la fréquence du réseau CEI 61000-4-8
REMARQUE : U
correspond à la tension secteur CA avant l’application du niveau de test.
T
40 % UT (baisse de 60 % en U pendant 5 cycles
70 % UT (baisse de 30 % en U pendant 25 cycles
<5 % UT (baisse >95 % en U pendant 5 s
40 % UT (baisse de 60 % en U
)
T
pendant 5 cycles
70 % UT (baisse de 30 % en U
)
T
pendant 25 cycles
<5 % UT (baisse >95 % en U
)
T
pendant 5 s
3 A/m 3 A/m
La qualité de l’alimentation secteur doit être celle d’un environnement
)
T
commercial ou médical standard. Si l’utilisateur du système VisionAire nécessite un fonctionnement continu de l’appareil en cas de coupure de courant, il est recommandé d’utiliser le système
)
T
VisionAire avec une alimentation sans interruption (UPS).
)
T
Les champs magnétiques à la fréquence du réseau doivent correspondre à ceux d’un environnement commercial ou médical standard.
PN MN238-C4 A | Manuel d’utilisation 43 - FRE
Page 44
VisionAire Family
Directives et déclaration du fabricant - Immunité électromagnétique
Le système VisionAire est prévu pour une utilisation dans les environnements électromagnétiques spécifiés ci-dessous. Le client ou l’utilisateur du système VisionAire doit s’assurer que celui-ci est utilisé dans un environnement de ce type.
Test d’immunité Niveau de test
CEI 60601
RF conduites CEI 61000-4-6
RF rayonnées CEI 61000-4-3
3 Vrms 150 kHz à 80 MHz
3 V/m 80 MHz à 2,5 GHz
Niveau de conformité Environnement électromagnétique -
Directives
3 Vrms
Les équipements de communication RF portables et mobiles ne doivent pas être utilisés à proximité des composants de l’équipement VisionAire, y compris les câbles, à une distance inférieure à la
3 V/m
distance de séparation recommandée calculée sur la base de l’équation applicable à la fréquence de l’émetteur.
Distance de séparation recommandée
d = 1,2√P
d = 1,2√P
= 80 MHz à 800 MHz
d = 1,2√P = 800 MHz à 2,5 GHz
où P correspond à la puissance nominale de sortie maximum de l’émetteur en watts (W), indiquée par le fabricant de l’émetteur, et d correspond à la distance de séparation recommandée en mètres (m).
L’intensité des champs émis par des émetteurs RF fixes, déterminée par l’étude électromagnétique d’un site inférieure au niveau de conformité pour chaque plage de fréquences
a
, doit être
b.
Des interférences peuvent se produire à proximité des équipements portant le symbole suivant :
REMARQUE 1 : à 80 MHz et 800 MHz, la plage de fréquences supérieure s’applique. REMARQUE 2 : il est possible que ces directives ne s’appliquent pas à toutes les situations. La propagation
électromagnétique dépend des phénomènes d’absorption et de réflexion induits par les structures, les objets et les personnes.
a
L’intensité des champs produits par des émetteurs fixes, tels que les stations de base pour les radiotéléphones (cellulaires/sans fil) et les radios mobiles, les radios amateurs, les radios AM et FM et la télévision ne peut pas être prédite théoriquement avec précision. Pour évaluer l’environnement électromagnétique dû aux émetteurs RF fixes, une étude électromagnétique du site doit être envisagée. Si l’intensité du champ, mesurée à l’emplacement où le système VisionAire est utilisé, dépasse le niveau de conformité RF applicable indiqué ci-dessus, il conviendra de s’assurer du bon fonctionnement du système VisionAire dans un tel environnement. En cas d’anomalie, il pourra être nécessaire de réaliser des mesures supplémentaires, par exemple en modifiant l’orientation ou la position du système VisionAire.
b
Sur la plage de fréquences de 150 kHz à 80 MHz, l’intensité des champs doit être inférieure à 3 V/m.
44 - FRE PN MN238-C4 A | Manuel d’utilisation
Page 45
VisionAire Family
Mode d’élimination
Élimination des déchets
Tous les déchets relatifs au concentrateur d’oxygène VisionAire de CAIRE doivent être éliminés selon les manières appropriées indiquées par les autorités locales
Élimination de l’appareil
An de préserver l’environnement, le concentrateur
doit être éliminé selon les manières appropriées indiquées par les autorités locales
Classification
Type de protection contre les chocs électriques : Classe II La protection contre les chocs électriques
est assurée par une double isolation.
Il n’est pas nécessaire d’effectuer une
mise à la terre de protection ou de se er
aux conditions d’installation.
Degré de protection contre les chocs électriques : Type BF Équipement assurant un degré particulier
de protection contre les chocs électriques concernant :
1) le courant de fuite autorisé ;
2) la abilité de la mise à la terre de
protection (le cas échéant).
Non destiné aux utilisations cardiaques
directes.
Degré de protection contre la pénétration nocive de l’eau : Équipement protégé contre les gouttes
d’eau - IP21
Protection contre la pénétration de corps
solides étrangers mesurant au moins 12,5 mm de diamètre et protection contre la pénétration de gouttes d’eau tombant à la verticale.
Mode autorisé de nettoyage et de prévention des infections : Veuillez consulter la section
Maintenance dans les manuels de réparation VisionAire.
Degré de sécurité d’utilisation en présence de gaz
anesthésiques inammables :
Appareil non adapté à ce type
Mode de fonctionnement : Fonctionnement en continu.
Remarque : pour une utilisation correcte de la canule, reportez-vous toujours aux instructions du fabricant. Remplacez la canule jetable conformément aux recommandations du fabricant de la canule ou du fournisseur de l’équipement. D’autres fournitures sont disponibles auprès de votre fournisseur.
Remarque : le fabricant ne recommande pas de stériliser cet équipement.
Copyright © 2018 Chart Industries. CAIRE Inc. se réserve le droit d’interrompre la commercialisation de ses produits ou de modifier les tarifs, les matériaux, les équipements, les niveaux de qualité, les descriptions, les caractéristiques et/ou les processus de ses produits, et ce à tout moment, sans avertissement préalable et sans aucune autre obligation ni conséquence. Nous nous réservons tous les droits non expressément stipulés dans le présent document, selon les conditions applicables.
d’application.
PN MN238-C4 A | Manuel d’utilisation 45 - FRE
Page 46
VisionAire Family
Steuerungselemente und Systemstatusanzeigen
ISO 7000; graphische Symbole für die Verwendung
auf dem Gerät — Index und Übersicht
Lesen Sie vor der Verwendung diese Bedienungsanleitung. Reg.-Nr. 1641
Vor Regen schützen, trocken aufbewahren. Reg.-Nr. 0626
Name und Adresse des Herstellers. Reg.-Nr. 3082
Vorsicht, Begleitdokumente beachten.
ISO 7010: Graphische Symbole—
Sicherheitsfarben und Sicherheitszeichen—
Registrierte Sicherheitszeichen
Richtlinie 93/42/EWG des Rates für
Medizinprodukte
Reg.-Nr. 0434A
Katalognummer. Reg.-Nr. 2493
Seriennummer. Reg.-Nr. 2498
Diese Seite nach oben. Reg.-Nr. 0623
Zerbrechlich, Vorsicht bei der Handhabung. Reg.-Nr. 0621
Die Betriebsanleitung ist aufmerksam durchzulesen. Reg.-Nr. M002
Von Flammen, offenem Feuer und Funken fernhalten. Offene Zündquellen und Rauchen verboten. Reg.-Nr. P003
Nicht in der Nähe des Geräts oder während dessen Betrieb rauchen. Reg.-Nr. P002
Anwendungsteil des Typs BF (Schutzgrad gegen Stromschlag). Reg.-Nr. 5333
Warnung: Reg.-Nr. W001
Autorisierter Vertreter in der Europäischen Gemeinschaft Dieses Gerät entspricht den Vorgaben der Richtlinie 93/42/EWG für Medizinprodukte. Es trägt die CE-Kennzeichnung wie dargestellt.
Interne Symbole
Von entzündlichen Materialien, Öl und Fett fernhalten.
Sicherheitsrichtlinie CAN/CSA C22.2 Nr. 60601-1-14 zu elektrischen Medizinprodukten. Sowohl für den US- als auch für den kanadischen Markt zugelassen; entspricht den US-Normen und kanadischen Normen.
Nicht zerlegen.
Wenn das Gerät über eine Alarmeinheit verfügt, meldet sie, dass eine Unterbrechung der externen Stromversorgung erkannt wurde.
Aufkleber 2018: Wenn das Gerät über eine Alarmeinheit verfügt, meldet sie eine niedrige Sauerstoffkonzentration am Geräteausgang.
EIN (Strom eingeschaltet)
AUS (Strom ausgeschaltet)
Herstellungsdatum
Gerät der Klasse II
21 CFR 801.15: Code of Federal Regulations, Titel 21
Nach den in den USA geltenden Gesetzen darf dieses Gerät nur von einem Arzt oder im Auftrag eines Arztes verkauft werden.
EN 60601-1: Medizinische elektrische Geräte – Teil 1: Allgemeine Festlegungen für die grundlegende Sicherheit und die wesentlichen Leistungsmerkmale
Tropfwassergeschütztes Gerät – Schutzart IP21
Richtlinie 2012/19/EU des Rates für Elektro- und Elektronik-Altgeräte (WEEE-Richtlinie)
46 - GER PN MN238-C4 A | Bedienungsanleitung
Page 47
WEEE-Richtlinie
Das Symbol soll den Besitzer des Geräts darauf hinweisen, dass das Gerät gemäß der Richtlinie über Elektro- und Elektronik-Altgeräte am Ende seiner Lebensdauer zu einer Recyclingstelle gebracht werden muss. Unsere Produkte entsprechen der Richtlinie zur Beschränkung der Verwendung bestimmter gefährlicher Stoffe (RoHS). Sie enthalten Blei oder andere gefährliche Materialien höchstens in Spuren.
Dieses Produkt kann von einem oder mehreren US-amerikanischen oder internationalen Patenten
abgedeckt sein. Auf unserer Webseite nden Sie eine Auistung der geltenden Patente. Pat.: patents.gtls.io.
VisionAire Family
AirSep® VisionAire® Sauerstoffkonzentrator
Diese Bedienungsanleitung macht Sie mit der Bedienung des CAIRE VisionAire­Sauerstoffkonzentrators vertraut. Lesen Sie die in dieser Anleitung enthaltenen Informationen sorgfältig durch, bevor Sie den Konzentrator in Betrieb nehmen. Wenden Sie sich bei Fragen an Ihren Geräteanbieter.
Was genau ist der Sauerstoffkonzentrator?
Die Atemluft enthält ca. 21 % Sauerstoff, 78 % Stickstoff und 1 % andere Gase. Bei dem VisionAire­Sauerstoffkonzentrator wird Raumluft durch die
Lufteinlässe in die Maschine eingesaugt. Sie ießt
dann durch ein Absorbermaterial, das sogenannte Molekularsieb. Dieses Material trennt den Sauerstoff vom Stickstoff und lässt nur den Sauerstoff hindurch. Das Ergebnis ist hochreiner Sauerstoff, der dem Anwender zugeführt wird.
Hinweis: Der Sauerstoff innerhalb eines Raumes kann nicht aufgebraucht werden, wenn Sie den Sauerstoffkonzentrator verwenden.
PN MN238-C4 A | Bedienungsanleitung 47 - GER
Page 48
VisionAire Family
Gründe für die Verschreibung von Sauerstoff durch den Arzt
Viele Menschen leiden unter verschiedenartigen Herz-, Lungen- und anderen Atemwegserkrankungen. Einem bedeutenden Teil dieser Menschen kann durch eine zusätzliche Sauerstoffzufuhr zu Hause, auf Reisen oder bei Tagesaktivitäten geholfen werden.
Das vorgeschriebene Sauerstoffniveau muss unbedingt eingehalten werden. Einstellungen der Flussrate dürfen nur geändert werden, wenn dies auf Anweisung eines approbierten Arztes erfolgt.
Der Gebrauch des Sauerstoffkonzentrators während des Schlafs darf nur auf Empfehlung durch qualifiziertes klinisches Personal erfolgen.
Die Raumluft, die wir einatmen, besteht zu 21 % aus dem Gas Sauerstoff. Unser Körper benötigt einen
stetigen Zuuss von Sauerstoff, damit er richtig
funktioniert. Ihr Arzt hat Ihnen eine Flussrate oder eine Einstellung gemäß Ihrem persönlichen respiratorischen Zustand verschrieben.
Obgleich Sauerstoff als Medikament nicht süchtig macht, kann eine unbefugte Sauerstofftherapie gefährlich sein. Vor der Verwendung dieses Sauerstoffkonzentrators müssen Sie einen Arzt konsultieren. Der Geräteanbieter Ihres Sauerstoffgeräts wird Ihnen zeigen, wie Sie die verschriebene Flussrate einstellen.
WARNUNG: SCHILDER MIT DER AUFSCHRIFT „RAUCHEN VERBOTEN – SAUERSTOFF IN GEBRAUCH“ MÜSSEN IM HAUS ODER AN DEM ORT, AN DEM DER SAUERSTOFF VERWENDET WIRD, DEUTLICH ERKENNBAR ANGEBRACHT SEIN. ANWENDER UND PFLEGEPERSONAL MÜSSEN ÜBER DIE GEFAHREN DES RAUCHENS IN DER NÄHE ODER WÄHREND DES GEBRAUCHS VON MEDIZINISCHEM SAUERSTOFF AUFGEKLÄRT WERDEN.
VORSICHT: Für den Fall eines Stromausfalls, Alarms oder mechanischen Versagens empfiehlt der Hersteller eine alternative Sauerstoffzufuhr. Den Arzt oder Geräteanbieter fragen, welche Art von Reservesystem benötigt wird.
Profil des Anwenders
Konzentratoren sind für die zusätzliche Sauerstoffzufuhr an den Anwender bestimmt, die unter Beschwerden aufgrund eines beeinträchtigten Übergangs des Sauerstoffs von der Lunge in den Blutkreislauf leiden. Stationäre Sauerstoffkonzentratoren (POCs) speichern und enthalten keinen Sauerstoff. Sie müssen nicht wieder aufgefüllt werden und können überall aufgeladen
werden, wo Wechsel- oder Gleichstrom ießt. Die
Verwendung eines Sauerstoffkonzentrators muss von einem Arzt verschrieben werden. Das Gerät ist nicht für lebenserhaltende Maßnahmen geeignet.
Obgleich Sauerstofftherapien Menschen jeden Alters verordnet werden können, handelt es sich in der Regel um Personen über 65 Jahre, die an einer Atemwegserkrankung wie z. B. der chronischen obstruktiven Lungenerkrankung (COPD) leiden. Üblicherweise verfügen die Anwender noch über gute kognitive Fähigkeiten und müssen dazu in der Lage sein, Unwohlsein mitzuteilen. Ist der Anwender nicht (mehr) in der Lage, Unwohlsein mitzuteilen oder die Aufkleber auf dem Konzentrator bzw. die Gebrauchsanweisung zu lesen und zu verstehen, darf das Gerät nur unter Aufsicht einer Person verwendet werden, die hierzu in der Lage ist. Treten während der Verwendung eines Konzentrators Beschwerden auf, sollten sich die betroffenen Anwender mit ihrem Arzt in Verbindung setzen. Es wird Anwendern empfohlen, Reservesauerstoff
(z. B. in Form einer Sauerstoffasche) bereitzuhalten,
der im Falle eines Stromausfalls oder eines sonstigen Ausfalls des Konzentrators verwendet werden kann. Weitere besondere Fertigkeiten oder Fähigkeiten des Anwenders sind für den Gebrauch des Sauerstoffkonzentrators nicht erforderlich.
48 - GER PN MN238-C4 A | Bedienungsanleitung
Page 49
Sicherheitsmerkmale
Die folgenden Informationen betreffen die Sicherheit des VisionAire-Sauerstoffkonzentrators. Lesen Sie die in dieser Anleitung enthaltenen Informationen aufmerksam durch, bevor Sie das Gerät in Betrieb nehmen. Wenden Sie sich bei Fragen an Ihren Geräteanbieter.
Warnaufkleber und Warnanzeige des Produkts.
• Kompressormotor: Am Kompressorauslass
bendet sich ein Überdruckventil, das auf 280 KPa
(40 psig) eingestellt ist. Die thermische Sicherheit wird durch ein Thermostat gewährleistet, das sich
in der Statorwicklung des Kompressors bendet
(135° C/275° F).
• Allgemeine Funktionsstörung: Tritt eine der unten aufgelisteten Bedingungen auf, leuchtet die Warnleuchte für allgemeine Funktionsstörungen (
) auf und ein Piepton ist zu hören.
Dies schließt Folgendes ein:
• Hindernisse, die den Sauerstoffuss
beeinträchtigen, z. B. Verengungen oder Knickstellen in der Kanüle; verursacht durch hohen Produktspeicherdruck
• Produktspeicherdruck höher als 33 psig (± 1)
• Produktspeicherdruck niedriger als 5 psig (± 1)
• Produktspeichertemperatur höher als 135° C (275° F); verursacht durch niedrigen Produktspeicherdruck, wenn der thermische Schalter im Kompressor ausgelöst wird (Ausschalten des Kompressors)
VisionAire Family
• Sauerstoffmonitor: Der Sauerstoffmonitor erkennt jedes Absinken der Konzentration unter 82 %. Tritt dieser Fall ein, leuchtet die Sauerstoffkonzentrations­Warnleuchte ( weiterhin niedrig, wird zusätzlich zur Warnleuchte ein Piepton aktiviert.
• Stromausfall: Wird das Gerät gerade betrieben und ein Stromausfall tritt auf, leuchtet die Stromwarnleuchte ( hören. Siehe Aufkleberbild.
• Produktlter: ≥ 10-μm-Filter
) auf. Bleibt die O2-Konzentration
) auf und ein Piepton ist zu
PN MN238-C4 A | Bedienungsanleitung 49 - GER
Page 50
VisionAire Family
Auspacken Ihres VisionAire
Überprüfen Sie, ob alle aufgeführten Komponenten im Paket enthalten sind. Falls etwas fehlen sollte, setzen Sie sich sofort mit Ihrem Sauerstoffanbieter in Verbindung.
• Stationärer Sauerstoffkonzentrator
Erste Schritte mit Ihrem VisionAire-Sauerstoffkonzentrator
Machen Sie sich zunächst mit den wichtigsten Teilen Ihres VisionAire-Sauerstoffkonzentrators vertraut.
A. Ein/Aus-Netzschalter (E/A): Startet und stoppt den
Betrieb des Geräts.
B. Reset-Taste des Schutzschalters: Setzt das Gerät
nach einer Abschaltung durch elektrische Überlast zurück
C. Stundenzähler: Erfasst die Gesamtbetriebszeit.
D. Durchussmesser/Einstellknopf: Regelt den
Sauerstoffvolumenstrom in Litern pro Minute (l/min) und zeigt ihn an.
E. Sauerstoffauslass: Anschluss für einen Befeuchter
(falls benötigt) oder eine Kanüle.
F. Griffe oben und an der Seite: Damit lässt sich das
Gerät bequem tragen.
G. Warn- und Alarmaufkleber
H. Spezikationsaufkleber: Listet elektrische Daten
und die Seriennummer auf.
I. Netzkabel: Darüber kann das Gerät an eine
Netzsteckdose angeschlossen werden.
J. Rollen: Mit den vier Rollen kann das Gerät (falls
nötig) leicht bewegt werden.
K. Einbuchtung für Befeuchterasche: Position der
optionalen Befeuchterasche. L. Befeuchterasche (optional) M. Sauerstoffauslass der Befeuchterasche: Darüber
können ein Sauerstoffschlauch oder eine Kanüle
angeschlossen werden.
N. Sauerstoffauslass der Befeuchterasche O. Schlauch der Befeuchterasche P. Befestigung der Befeuchterasche
Q. Sauerstoffschlauch/-kanüle
WARNUNG: KEINE VERLÄNGERUNGSKABEL FÜR DIESES GERÄT VERWENDEN UND NICHT ZU VIELE STECKER MIT DERSELBEN STECKDOSE VERBINDEN. DIE VERWENDUNG EINES VERLÄNGERUNGSKABELS KANN DIE LEISTUNG DES GERÄTES BEEINTRÄCHTIGEN. DIE VERBINDUNG ZU VIELER STECKER MIT EINER STECKDOSE KANN DEN SCHALTSCHRANK ÜBERLASTEN UND ZUM AUSLÖSEN VON TRENNSCHALTER/ SICHERUNG ODER BEI FEHLFUNKTION VON TRENNSCHALTER/SICHERUNG ZUM AUSBRUCH EINES FEUERS FÜHREN.
50 - GER PN MN238-C4 A | Bedienungsanleitung
Page 51
VisionAire Family
F
E
C
L
F
T
B
K
A
G
G
J
Abbildung 1a
I
N
M
H
O
Abbildung 1b
P
Q
PN MN238-C4 A | Bedienungsanleitung 51 - GER
Page 52
VisionAire Family
Wichtig!
Warn- und Sicherheitshinweise sind folgendermaßen gekennzeichnet:
WARNUNG: WEIST AUF GEFAHREN HIN, DIE SCHWERE VERLETZUNGEN ZUR FOLGE HABEN KÖNNEN.
VORSICHT: Weist auf Gefahren hin, die Sachschäden an der VisionAire­Produktserie verursachen können.
Hinweis: Besonders zu beachtende Informationen.
Gebrauchshinweise
WARNUNG: DAS VORGESCHRIEBENE SAUERSTOFFNIVEAU MUSS UNBEDINGT EINGEHALTEN WERDEN. EINSTELLUNGEN DER FLUSSRATE DÜRFEN NUR GEÄNDERT WERDEN, WENN DIES AUF ANWEISUNG EINES APPROBIERTEN ARZTES ERFOLGT.
WARNUNG: LAUT US-GESETZEN DARF DIESES GERÄT NUR AUF ANWEISUNG EINES ARZTES ODER EINES ANDEREN LIZENZIERTEN GESUNDHEITSDIENSTLEISTERS VERKAUFT ODER VERLIEHEN WERDEN.
WARNUNG: DIESES GERÄT DARF NICHT FÜR LEBENSERHALTENDE MASSNAHMEN EINGESETZT WERDEN. ÄLTERE, PÄDIATRISCHE ODER ANDERE ANWENDER, DIE IHR UNWOHLSEIN WÄHREND DER VERWENDUNG DES GERÄTS NICHT KOMMUNIZIEREN KÖNNEN, MÜSSEN MÖGLICHERWEISE ZUSÄTZLICH ÜBERWACHT WERDEN. ANWENDER MIT HÖR- UND/ODER SEHBEHINDERUNG BENÖTIGEN MÖGLICHERWEISE HILFE BEIM ÜBERWACHEN DER ALARME. BEI UNWOHLSEIN ODER EINEM MEDIZINISCHEN NOTFALL SOFORT ÄRZTLICHE HILFE ZURATE ZIEHEN.
Gegenanzeigen
WARNUNG: DIE VERWENDUNG VON SAUERSTOFF OHNE ÄRZTLICHE VERSCHREIBUNG IST UNTER UMSTÄNDEN GEFÄHRLICH. DAS GERÄT DARF NUR AUF VERSCHREIBUNG EINES ARZTES VERWENDET WERDEN.
WARNUNG: DAS GERÄT DARF NICHT VERWENDET WERDEN, WENN BRENNBARE ANÄSTHETIKA VORHANDEN SIND.
WARNUNG: WIE ALLE ELEKTRISCH BETRIEBENEN GERÄTE KANN AUCH DER SAUERSTOFFKONZENTRATOR, BEISPIELSWEISE BEI EINEM STROMAUSFALL ODER WENN EINE WARTUNG DURCH EINEN QUALIFIZIERTEN TECHNIKER ERFORDERLICH IST, AUSFALLEN. DER SAUERSTOFFKONZENTRATOR IST NICHT FÜR ANWENDER GEEIGNET, DIE DURCH EINE SOLCHE TEMPORÄRE UNTERBRECHUNG GESUNDHEITLICHEN SCHADEN NEHMEN WÜRDEN.
Sicherheitsrichtlinien
WARNUNG: DIE FOLGENDEN WICHTIGEN SICHERHEITSINFORMATIONEN ZUM VISIONAIRE INTENSITY­SAUERSTOFFKONZENTRATOR SORGFÄLTIG DURCHLESEN UND SICH MIT DEM INHALT VERTRAUT MACHEN.
WARNUNG: DAS GERÄT NUR NACH EINGEHENDEM STUDIUM UND VERSTÄNDNIS DIESER ANLEITUNG VERWENDEN. SOLLTEN DIE WARNUNGEN UND ANWEISUNGEN NICHT VERSTANDEN WERDEN, VOR DEM VERSUCH, DAS GERÄT ZU VERWENDEN, AN DEN GERÄTEANBIETER WENDEN; ANDERNFALLS KANN ES ZU VERLETZUNGEN ODER SCHÄDEN KOMMEN.
WARNUNG: DAS RAUCHEN WÄHREND DER SAUERSTOFFVERABREICHUNG IST DER HAUPTGRUND FÜR BRANDVERLETZUNGEN UND -TODE. DIESE SICHERHEITSWARNUNGEN MÜSSEN UNBEDINGT BEFOLGT WERDEN:
WARNUNG: RAUCHEN, KERZEN ODER OFFENE FLAMMEN SIND IN DEM RAUM, IN DEM SICH DAS GERÄT ODER EIN SAUERSTOFF BEFÖRDERNDES ZUBEHÖRTEIL BEFINDEN, VERBOTEN.
WARNUNG: RAUCHEN WÄHREND DES TRAGENS EINER SAUERSTOFFKANÜLE KANN ZU GESICHTSVERBRENNUNGEN UND ZUM TODE FÜHREN.
WARNUNG: EIN ABNEHMEN DER KANÜLE UND EINE LAGERUNG AUF KLEIDUNG, BETTWÄSCHE, SOFAS ODER ANDEREM POLSTERMATERIAL FÜHREN BEI VORHANDENSEIN EINER ZIGARETTE, WÄRMEQUELLE ODER OFFENEN FLAMME ZU EINER VERPUFFUNG.
52 - GER PN MN238-C4 A | Bedienungsanleitung
Page 53
VisionAire Family
WARNUNG: VOR DEM RAUCHEN MÜSSEN STETS DIE FOLGENDEN DREI (3) WICHTIGEN SCHRITTE BEFOLGT WERDEN: DEN SAUERSTOFFKONZENTRATOR AUSSCHALTEN, DIE KANÜLE ABNEHMEN UND DEN RAUM, IN DEM DAS GERÄT SICH BEFINDET, VERLASSEN.
WARNUNG: SCHILDER MIT DER AUFSCHRIFT „RAUCHEN VERBOTEN – SAUERSTOFF IN GEBRAUCH“ MÜSSEN IM HAUS ODER AN DEM ORT, AN DEM DER SAUERSTOFF VERWENDET WIRD, DEUTLICH ERKENNBAR ANGEBRACHT SEIN. ANWENDER UND PFLEGEPERSONAL MÜSSEN ÜBER DIE GEFAHREN DES RAUCHENS IN DER NÄHE ODER WÄHREND DES GEBRAUCHS VON MEDIZINISCHEM SAUERSTOFF AUFGEKLÄRT WERDEN.
WARNUNG: LAUT US-GESETZEN DARF DIESES GERÄT NUR AUF ANWEISUNG EINES ARZTES ODER EINES ANDEREN LIZENZIERTEN GESUNDHEITSDIENSTLEISTERS VERKAUFT ODER VERLIEHEN WERDEN.
WARNUNG: DIESES GERÄT LIEFERT HOCHKONZENTRIERTEN SAUERSTOFF, DER SCHNELL ZU BRÄNDEN FÜHREN KANN. RAUCHEN ODER OFFENE FLAMMEN SIND IM RAUM, IN DEM SICH (1) DAS GERÄT ODER (2) EIN SAUERSTOFF TRANSPORTIERENDES ZUBEHÖRTEIL BEFINDEN, NICHT ERLAUBT. DIE MISSACHTUNG DIESER WARNUNG KANN SCHWERE BRÄNDE, SACH- BZW. PERSONENSCHÄDEN ZUR FOLGE HABEN ODER ZUM TOD FÜHREN.
WARNUNG: DEN SAUERSTOFFKONZENTRATOR NICHT BEI VORHANDENSEIN ENTZÜNDLICHER GASE VERWENDEN. DIES KANN SCHNELL ZU BRÄNDEN FÜHREN, DIE SACHSCHÄDEN, VERLETZUNGEN ODER DEN TOD NACH SICH ZIEHEN KÖNNEN.
WARNUNG: EINE KANÜLE NICHT AUF KLEIDUNG, BETTWÄSCHE ODER STUHLPOLSTERN LIEGEN LASSEN. WENN DAS GERÄT EINGESCHALTET IST, ABER NICHT BETRIEBEN WIRD, KANN DER SAUERSTOFF ENTZÜNDLICH WIRKEN. DAS GERÄT AUSSCHALTEN, WENN ES NICHT IN GEBRAUCH IST.
WARNUNG: KEINE ÖL-, FETT-, ODER PETROLEUMBASIERTEN ODER ANDEREN ENTZÜNDLICHEN PRODUKTE MIT DEM SAUERSTOFFKONZENTRATOR ODER DESSEN SAUERSTOFF BEINHALTENDEM ZUBEHÖR VERWENDEN. SAUERSTOFF BESCHLEUNIGT DIE VERBRENNUNG ENTZÜNDLICHER SUBSTANZEN. ES SOLLTEN AUSSCHLIESSLICH WASSERBASIERTE, SAUERSTOFFVERTRÄGLICHE LOTIONEN ODER SALBEN VERWENDET WERDEN.
WARNUNG: BESCHLÄGE, ANSCHLÜSSE, SCHLÄUCHE ODER ANDERE ZUBEHÖRTEILE DES SAUERSTOFFKONZENTRATORS NICHT SCHMIEREN, UM EINE GEFÄHRDUNG DURCH BRÄNDE ODER VERBRENNUNGEN ZU VERMEIDEN.
WARNUNG: STROMSCHLAGGEFAHR. DAS GERÄT AUSSCHALTEN UND DAS NETZKABEL AUS DER NETZSTECKDOSE ZIEHEN, BEVOR DAS GERÄT GEREINIGT ODER GEWARTET WIRD, UM EINE STROMSCHLAGGEFAHR UND VERBRENNUNGSGEFAHR ZU VERMEIDEN. NUR DER GERÄTEANBIETER ODER EIN QUALIFIZIERTER WARTUNGSTECHNIKER DÜRFEN DIE ABDECKUNGEN ENTFERNEN ODER DAS GERÄT WARTEN.
WARNUNG: ES MUSS DARAUF GEACHTET WERDEN, DASS DER SAUERSTOFFKONZENTRATOR STETS TROCKEN BLEIBT UND KEINERLEI FLÜSSIGKEIT IN DAS GERÄT EINTRITT. ANDERNFALLS KANN ES ZU EINER FEHLFUNKTION ODER ZUR ABSCHALTUNG KOMMEN ODER ES BESTEHT EIN ERHÖHTES RISIKO EINES ELEKTRISCHEN SCHLAGS ODER VON VERBRENNUNGEN.
WARNUNG: KEINE FLÜSSIGKEITEN DIREKT AUF DEM GERÄT VERWENDEN. ZU DEN NICHT VERTRÄGLICHEN CHEMISCHEN MITTELN ZÄHLEN UNTER ANDEREM: ALKOHOL UND PRODUKTE AUF ALKOHOLBASIS, KONZENTRIERTE PRODUKTE AUF CHLORBASIS (ETHYLCHLORID) UND PRODUKTE AUF ÖLBASIS (PINE-SOL®, LESTOIL®). DIESE DÜRFEN NICHT VERWENDET WERDEN, UM DAS KUNSTSTOFFGEHÄUSE DES SAUERSTOFFKONZENTRATORS ZU REINIGEN, DA SIE DEN KUNSTSTOFF BESCHÄDIGEN KÖNNTEN.
WARNUNG: GEHÄUSE, BEDIENFELD UND STROMKABEL NUR MIT EINEM SANFTEN HAUSHALTSREINIGER, DER MIT EINEM FEUCHTEN (NICHT NASSEN) TUCH ODER SCHWAMM AUFGETRAGEN WIRD, REINIGEN. ANSCHLIESSEND ALLE OBERFLÄCHEN TROCKEN WISCHEN. KEINE FLÜSSIGKEITEN IN DAS INNERE DES GERÄTES GELANGEN LASSEN.
WARNUNG: DER SAUERSTOFFKONZENTRATOR SOLLTE NICHT DIREKT NEBEN ANDEREN GERÄTEN ODER MIT ANDEREN GERÄTEN GESTAPELT VERWENDET WERDEN. IST EINE SOLCHE SITUATION NICHT ZU VERMEIDEN, MUSS DAS GERÄT AUF ORDNUNGSGEMÄSSEN BETRIEB HIN ÜBERWACHT WERDEN.
WARNUNG: DIE SAUERSTOFFVERSORGUNGSLEITUNG UND DAS STROMKABEL IMMER SO PLATZIEREN, DASS NIEMAND DARÜBER STOLPERN ODER SICH DAMIT STRANGULIEREN KANN.
WARNUNG: AN DIESEM GERÄT DÜRFEN KEINE MODIFIKATIONEN VORGENOMMEN WERDEN.
WARNUNG: DIE VERWENDUNG VON ANDEREN KABELN UND ADAPTERN ALS DEN ANGEGEBENEN, MIT AUSNAHME DER KABEL UND ADAPTER, DIE VOM HERSTELLER DES MEDIZINISCHEN ELEKTRISCHEN GERÄTS ALS ERSATZTEILE FÜR INTERNE KOMPONENTEN VERKAUFT WERDEN, KANN ZU ERHÖHTEN EMISSIONEN ODER EINER REDUZIERTEN STÖRFESTIGKEIT DES SAUERSTOFFKONZENTRATORS FÜHREN.
PN MN238-C4 A | Bedienungsanleitung 53 - GER
Page 54
VisionAire Family
WARNUNG: DIE LEISTUNG DES GERÄTS KANN DURCH UNGÜNSTIGE UMWELTEINFLÜSSE VERMINDERT WERDEN. STELLEN SIE ES IN EINER SAUBEREN UND SCHADORGANISMUSFREIEN UMGEBUNG AUF.
WARNUNG: DAS GERÄT SOLLTE NUR VON ENDANWENDERN, GESCHULTEM PFLEGEPERSONAL ODER GESCHULTEN TECHNIKERN EINGESETZT WERDEN. KINDER SOLLTEN DAS GERÄT NICHT BEDIENEN.
WARNUNG: DAMIT GEWÄHRLEISTET IST, DASS SIE IHREM GESUNDHEITSZUSTAND GEMÄSS MIT DER ENTSPRECHENDEN THERAPIEDOSIS SAUERSTOFF VERSORGT WERDEN, MUSS DAS VISIONAIRE­GERÄT MIT GENAU DER KOMBINATION AN TEILEN UND ZUBEHÖR EINGESETZT WERDEN, DIE DEN HERSTELLERANGABEN DES KONZENTRATORS ENTSPRICHT UND DIE AUCH BEIM FESTLEGEN DER EINSTELLUNGEN VERWENDET WURDE.
WARNUNG: DIE VERWENDUNG DES GERÄTES AUSSERHALB DER IN DIESER ANLEITUNG ANGEGEBENEN BEREICHE FÜR HÖHE, TEMPERATUR ODER LUFTFEUCHTIGKEIT KANN DIE DURCHFLUSSRATE UND DEN SAUERSTOFFANTEIL UND SOMIT DIE QUALITÄT DER THERAPIE BEEINTRÄCHTIGEN.
WARNUNG: DER EINSATZ EINIGER SAUERSTOFFFÜHRENDER ZUBEHÖRTEILE, DIE NICHT FÜR DIE VERWENDUNG MIT DIESEM SAUERSTOFFKONZENTRATOR ANGEGEBEN SIND, KANN DESSEN LEISTUNG BEEINTRÄCHTIGEN. EMPFOHLENES ZUBEHÖR WIRD IN DIESER ANLEITUNG ERWÄHNT.
WARNUNG: NUR DIE ELEKTRISCHE SPANNUNG ANLEGEN, DIE AUF DEM GERÄTEETIKETT MIT DEN TECHNISCHEN DATEN ANGEGEBEN IST.
WARNUNG: KEINE VERLÄNGERUNGSKABEL FÜR DIESES GERÄT VERWENDEN UND NICHT ZU VIELE STECKER MIT DERSELBEN STECKDOSE VERBINDEN. DIE VERWENDUNG EINES VERLÄNGERUNGSKABELS KANN DIE LEISTUNG DES GERÄTES BEEINTRÄCHTIGEN. DIE VERBINDUNG ZU VIELER STECKER MIT EINER STECKDOSE KANN DEN SCHALTSCHRANK ÜBERLASTEN UND ZUM AUSLÖSEN VON TRENNSCHALTER/SICHERUNG ODER BEI FEHLFUNKTION VON TRENNSCHALTER/ SICHERUNG ZUM AUSBRUCH EINES FEUERS FÜHREN.
VORSICHT: Laut US-Gesetzen darf dieses Gerät nur auf Anweisung eines Arztes oder eines anderen lizenzierten Gesundheitsdienstleisters verkauft oder verliehen werden.
VORSICHT: Das Gerät immer so platzieren, dass das Stromkabel gut erreichbar ist.
VORSICHT: Der Konzentrator darf nicht mit Rauch, Schadstoffen oder Dämpfen in Kontakt kommen.
VORSICHT: Sicherstellen, dass der Konzentrator in aufrechter Position betrieben wird.
VORSICHT: Stromkabel und Sauerstoffschläuche so platzieren, dass eine Stolpergefahr vermieden wird.
VORSICHT: Das Gerät entfernt von Vorhängen oder Gardinen, Heißluftregistern oder Heizungen aufstellen. Sicherstellen, dass das Gerät auf einer flachen Oberfläche aufgestellt wird, darauf achten, dass an allen Seiten ein Mindestabstand von 30 cm (1 Fuß) zu Wänden oder anderen Hindernissen besteht. Das Gerät nicht auf engem Raum aufstellen. Einen staub- und rauchfreien Aufstellort wählen, der vor direkter Sonneneinstrahlung geschützt ist. Das Gerät nicht in Außenbereichen betreiben, sofern die Netzsteckdose, an der das Gerät angeschlossen ist, nicht über eine Fehlerstrom-Schutzeinrichtung (RCD) verfügt.
VORSICHT: Das Gerät nicht in einem eingeschränkten oder engen Bereich betreiben, in dem die Belüftung unzureichend sein könnte. Dies könnte zu einer Überhitzung des Geräts und einer Beeinträchtigung der Leistung führen.
VORSICHT: Sicherstellen, dass Lufteinlass und -auslass nicht blockiert werden. Keine Gegenstände in die Geräteöffnungen einführen und nichts in diese hineinfallen lassen. Hierdurch kann sich der Sauerstoffkonzentrator überhitzen und die Leistung beeinträchtigt werden.
54 - GER PN MN238-C4 A | Bedienungsanleitung
Page 55
VisionAire Family
VORSICHT: Für den Fall eines Stromausfalls, Alarms oder mechanischen Versagens empfiehlt der Hersteller eine alternative Sauerstoffzufuhr. Den Arzt oder Geräteanbieter fragen, welche Art von Reservesystem benötigt wird.
VORSICHT: Das vorgeschriebene Sauerstoffniveau muss unbedingt eingehalten werden. Einstellungen der Flussrate dürfen nur geändert werden, wenn dies auf Anweisung eines approbierten Arztes erfolgt.
VORSICHT: Der Gebrauch des Sauerstoffkonzentrators während des Schlafs darf nur auf Empfehlung durch qualifiziertes klinisches Personal erfolgen.
VORSICHT: Der Betrieb oder die Aufbewahrung des Sauerstoffkonzentrators außerhalb des normalen Betriebstemperaturbereichs kann die Leistung des Geräts beeinträchtigen. Lagerungs­und Temperaturgrenzwerte sind dem Abschnitt „Technische Daten“ in dieser Anleitung zu entnehmen.
VORSICHT: Sollte ein Alarm ausgegeben werden oder der Sauerstoffkonzentrator nicht ordnungsgemäß funktionieren, den Abschnitt „Fehlerbehebung“ dieser Anleitung zurate ziehen. An den Geräteanbieter wenden, wenn das Problem nicht gelöst werden kann.
VORSICHT: Wenn der akustische Warnton schwach ist oder gar nicht ertönt, schnellstmöglich an den Geräteanbieter wenden.
VORSICHT: Wenn die Schläuche der Befeuchterflasche nicht ordnungsgemäß am Anschluss der Befeuchterflasche oder am Sauerstoffauslass angeschlossen sind, kann Sauerstoff austreten.
VORSICHT: Im Normalfall muss der Durchflussmesser an Ihrem Gerät nicht eingestellt werden. Durch Drehen des Einstellknopfs des Durchflussmessers im Uhrzeigersinn wird der Sauerstofffluss zum Gerät verringert oder ausgeschaltet. Zur vereinfachten Einstellung verfügt der Durchflussmesser über Markierungen in Schritten von ½ l/min. Bei Geräten, die mit der Durchflussmesser-Option von 2 l/min ausgestattet sind, ist der Durchflussmesser für Durchflusseinstellungen von bis zu 2 l/min mit Markierungen in Schritten von 1/8 l/min versehen.
Hinweis: Die Kanüle darf kein Knicken zulassen und kann in einer Gesamtlänge bis max. 7,6 m (25 Fuß) verwendet werden.
Sicherstellen, dass die Kanüle vollständig eingeführt ist und fest sitzt. Der Sauerstofffluss zu den Zinken der Nasenkanüle sollte hörbar oder fühlbar sein. Wenn es scheint, als würde kein Sauerstoff strömen, muss zunächst sichergestellt werden, dass die Kugel im Durchflussmesser einen Durchfluss erfasst. Die Spitze der Kanüle anschließend in ein Glas mit Wasser stellen; wenn Luftblasen aus der Kanüle austreten, fließt Sauerstoff. Falls keine Luftblasen austreten, den Abschnitt „Fehlerbehebung“ in dieser Anleitung zurate ziehen.
Hinweis: Für einen ordnungsgemäßen Gebrauch stets den Anleitungen des Kanülenherstellers folgen. Die Einweg-Kanüle gemäß den Empfehlungen des Kanülenherstellers oder des Geräteanbieters ersetzen. Weiteres Zubehör erhalten Sie von Ihrem Geräteanbieter.
Hinweis: Der VisionAire-Sauerstoffkonzentrator muss vor der Verwendung des Geräts mindestens fünf Minuten bei 2 l/min oder einem höheren Durchfluss laufen gelassen werden.
Der VisionAire ist geeignet zur Verwendung durch zwei Anwender, vorausgesetzt, der Sauerstofffluss beträgt insgesamt mindestens 2 l/min und überschreitet nicht die maximale Kapazität des Konzentrators.
An den Geräteanbieter: Folgendes Sauerstoffanwendungszubehör wird für die Verwendung mit dem VisionAire Intensity­Sauerstoffkonzentrator empfohlen:
• Luftbefeuchterflasche: Teile-Nr. HU003-1
• Nasenkanüle mit 2,1 m (7 Fuß) langem Schlauch (6 l/m max.): Teile-Nr. CU002-1
• OxySafe-Kit: Teile-Nr. 20629671 Hinweis: Der Hersteller empfiehlt, dieses Gerät nicht zu
sterilisieren. Hinweis: Wenn das Gerät für einen längeren Zeitraum
nicht benutzt wurde, muss es einige Minuten laufen gelassen werden, bevor das Stromausfall-Warnsignal aktiviert werden kann.
Hinweis: Der Konzentrator gibt an der Unterseite des Geräts warme Luft ab, durch die sich temperaturempfindliche Bodenoberflächen (z. B. Vinyl) permanent verfärben können. Der Konzentrator sollte nicht über Bodenoberflächen eingesetzt werden, die gegenüber Verfärbungen durch Wärme anfällig sind. Der Hersteller übernimmt keine Haftung für sich verfärbende Bodenoberflächen.
Hinweis: Alle Anweisungen des Herstellers befolgen, damit die Garantie ihre Gültigkeit behält.
Hinweis: Nicht versuchen, das Gerät – abgesehen von den in dieser Anleitung aufgeführten Störungsbehebungen – zu warten.
PN MN238-C4 A | Bedienungsanleitung 55 - GER
Page 56
VisionAire Family
Hinweis: Tragbare und mobile Hochfrequenz (HF)­Kommunikationsausrüstung kann medizinische elektrische Geräte beeinträchtigen.
Hinweis: Der Sauerstoff innerhalb eines Raumes kann nicht aufgebraucht werden, wenn Sie den Sauerstoffkonzentrator verwenden.
Hinweis: An den Bereitsteller des Gerätes: Folgendes Sauerstoffanwendungszubehör wird für die Verwendung mit dem Sauerstoffkonzentrator empfohlen:
• Nasenkanüle mit 2,1 m langem Schlauch: PN 5408­SEQ
• Weiteres empfohlenes Zubehör finden Sie im Zubehör-katalog (PN ML-LOX0010) auf www. CAIREmedical.com.
Bei der Verwendung mit einer Kanüle ist eine Brandsperre erforderlich.
• Zu diesem Zweck bietet CAIRE die OxySafe™ an. Dieses Zubehör ist für die Verwendung zusammen mit dem Sauerstoffkonzentrator vorgesehen. Die OxySafe™ ist eine thermische Sicherung, die den Gasfluss stoppt, wenn die flussabwärts gerichtete Kanüle oder der Sauerstoffschlauch Feuer fängt und in Richtung OxySafe abbrennt. Die Sicherung wird linear zur Nasenkanüle oder dem Sauerstoffschlauch zwischen dem Patienten und dem Sauerstoffauslass des VisionAire eingesetzt.
Angaben zur ordnungsgemäßen Verwendung der OxySafe sind den Anweisungen des Herstellers zu entnehmen (im Lieferumfang des OxySafe-Kits enthalten).
CAIRE bietet ein OxySafe™-Kit an, das ein Kanülenkopfstück, OxySafe™ und einen Schlauch mit einer Länge von 7 Fuß (ca. 2 m) enthält: PN 20629671.
WARNUNG: CET APPAREIL PRODUIT DE L’OXYGÈNE À CONCENTRATION ÉLEVÉE, FAVORISANT UNE COMBUSTION RAPIDE. NE PAS PERMETTRE DE FUMER OU DES FLAMMES NUES DANS LA MÊME CHAMBRE: (1) CET APPAREIL OU (2) TOUT ACCESSOIRE CONTENANT DE L’OXYGÈNE. NE PAS UTILISER DE PRODUITS À BASE DHUILE, DE GRAISSE OU DE PÉTROLE SUR OU À PROXIMITÉ DE L’UNITÉ. DÉCONNECTER LE CORDON D’ALIMENTATION DE LA PRISE ÉLECTRIQUE AVANT DE NETTOYER OU DE FAIRE L’ENTRETIEN DE L’UNITÉ.
RISQUE DE CHOC ÉLECTRIQUE. NE PAS ENLEVER LES COUVERCLES LORSQUE L’UNITÉ EST BRANCHÉE. SEULS VOTRE FOURNISSEUR D’ÉQUIPEMENT OU UN TECHNICIEN DE SERVICE QUALIFIÉ DEVRAIT ENLEVER LES COUVERCLES OU FAIRE L’ENTRETIEN DE L’UNITÉ.
56 - GER PN MN238-C4 A | Bedienungsanleitung
Page 57
VisionAire Family
Technische Daten
VisionAire
Durchflussmengen* 1 l/min–5 l/min
± 10 % der angegebenen Einstellung oder 200 ml, je nachdem, welcher Wert höher ist*
Abmessungen 14,1 Zoll B x 11,5 Zoll B x 20,8 Zoll H
(35,8 cm W x 29,2 cm D x 52,8 cm H)
Gewicht 30 lbs (13,6 kg) Schalldruckpegel 45 dBa (1 bis 5 l/min) Stromverbrauch 290 Watt O2-Konzentration 90 % (+ 5,5 %/- 3 %) Elektrische Anforderungen 115 VAC/60 Hz, 3,0 A
230 VAC/50 Hz, 1,5 A 230 VAC/60 Hz, 1,5 A
Betriebsumgebung* 5° C bis 40° C (41° F bis 104° F) bei Höhen von bis 10.000 Fuß (3048 m)
Höhe über N.N. -381 bis 3048 m (-1250 bis 10,000 ft) (getestet auf 700 - 1060 hPa) Lagerumgebung -25° C bis 70° C (-13° F bis 158° F)
* Basierend auf einem Luftdruckbereich von 700–1060 hPa bei 21° C (70° F)
über dem Meeresspiegel. 15–95 % Feuchtigkeit (nicht kondensierend)
Die erwartete Lebensspanne des Geräts beträgt mindestens 5 Jahre.
Siehe Technisches Handbuch (PN MN138-1) für Schallleistungspegel.
WARNUNG: DIE VERWENDUNG DES GERÄTES AUSSERHALB DER IN DIESER ANLEITUNG ANGEGEBENEN BEREICHE FÜR HÖHE, TEMPERATUR ODER LUFTFEUCHTIGKEIT KANN DIE DURCHFLUSSRATE UND DEN SAUERSTOFFANTEIL UND SOMIT DIE QUALITÄT DER THERAPIE BEEINTRÄCHTIGEN.
WARNUNG: ES IST ZU ERWARTEN, DASS DIE VERWENDUNG DIESES GERÄTS IN HÖHENLAGEN ÜBER
3.048 METERN (10.000 FUSS) BEI AUSSENTEMPERATUREN VON 5° C BIS 40° C (41° F BIS 104° F ) ODER BEI EINER RELATIVEN FEUCHTIGKEIT ÜBER 95 % DIE DURCHFLUSSRATE UND DEN SAUERSTOFFANTEIL UND SOMIT DIE QUALITÄT DER THERAPIE BEEINTRÄCHTIGT.
PN MN238-C4 A | Bedienungsanleitung 57 - GER
Page 58
VisionAire Family
Gebrauchsanweisungen
Lesen Sie die folgenden Informationen aufmerksam durch, bevor Sie den Sauerstoffkonzentrator verwenden.
Hinweis: Der Konzentrator gibt an der Unterseite des Geräts warme Luft ab, durch die sich temperaturempfindliche Bodenoberflächen (z. B. Vinyl) permanent verfärben können. Der Konzentrator sollte nicht über Bodenoberflächen eingesetzt werden, die gegenüber Verfärbungen durch Wärme anfällig sind. Der Hersteller übernimmt keine Haftung für sich verfärbende Bodenoberflächen.
Befeuchterflasche (optional)
Wenn zusätzlich zur Sauerstofftherapie eine Befeuchtung benötigt wird, führen Sie die folgenden Schritte jedes Mal aus, bevor Sie den Befeuchter füllen oder reinigen, der ursprünglich für Ihre Benutzung eingerichtet wurde.
1. Entfernen Sie die Befeuchterasche aus der Einbuchtung für die Befeuchterasche.
2. Öffnen Sie die Befeuchterasche. Führen Sie diesen
Schritt nicht durch, wenn die Flasche bereits gefüllt ist. Fahren Sie mit Schritt 5 fort.
3. Füllen Sie die Befeuchterasche bis zur
Markierung auf der Flasche mit kaltem Wasser (nach Möglichkeit destilliert) auf. NICHT ÜBERFÜLLEN.
4. Verbinden Sie die obere Abdeckung wieder mit der
Befeuchterasche.
5. Stellen Sie die Befeuchterasche wieder in die Einbuchtung für die Befeuchterasche an der
Rückseite des Konzentrators und schließen
Sie den Schlauch der Befeuchterasche an
den Sauerstoffauslass und den Anschluss der
Befeuchterasche an.
VORSICHT: Wenn die Schläuche der Befeuchterflasche nicht ordnungsgemäß am Anschluss der Befeuchterflasche oder am Sauerstoffauslass angeschlossen sind, kann Sauerstoff austreten.
Anschluss der Kanüle
Verbinden Sie die Schläuche und die Kanüle mit dem Sauerstoffauslass oder mit dem Ausgang des optionalen Befeuchters.
Hinweis: Die Kanüle darf kein Knicken zulassen und kann in einer Gesamtlänge bis max. 7,6 m (25 Fuß) verwendet werden.
Sicherstellen, dass die Kanüle vollständig eingeführt ist und fest sitzt. Der Sauerstofffluss zu den Zinken der Nasenkanüle sollte hörbar oder fühlbar sein. Wenn es scheint, als würde kein Sauerstoff strömen, muss zunächst sichergestellt werden, dass die Kugel im Durchflussmesser einen Durchfluss erfasst. Die Spitze der Kanüle anschließend in ein Glas mit Wasser stellen; wenn Luftblasen aus der Kanüle austreten, fließt Sauerstoff. Falls keine Luftblasen austreten, den Abschnitt „Fehlerbehebung“ in dieser Anleitung zurate ziehen.
Für einen ordnungsgemäßen Gebrauch stets den Anleitungen des Kanülenherstellers folgen. Die Einweg-Kanüle gemäß den Empfehlungen des Kanülenherstellers oder des Geräteanbieters ersetzen. Weiteres Zubehör erhalten Sie von Ihrem Geräteanbieter.
Start des Konzentrators
WARNUNG: SCHILDER MIT DER AUFSCHRIFT „RAUCHEN VERBOTEN – SAUERSTOFF IN GEBRAUCH“ MÜSSEN IM HAUS ODER AN DEM ORT, AN DEM DER SAUERSTOFF VERWENDET WIRD, DEUTLICH ERKENNBAR ANGEBRACHT SEIN. ANWENDER UND PFLEGEPERSONAL MÜSSEN ÜBER DIE GEFAHREN DES RAUCHENS IN DER NÄHE ODER WÄHREND DES GEBRAUCHS VON MEDIZINISCHEM SAUERSTOFF AUFGEKLÄRT WERDEN.
58 - GER PN MN238-C4 A | Bedienungsanleitung
Page 59
WARNUNG: KEINE VERLÄNGERUNGSKABEL FÜR DIESES GERÄT VERWENDEN UND NICHT ZU VIELE STECKER MIT DERSELBEN STECKDOSE VERBINDEN. DIE VERWENDUNG EINES VERLÄNGERUNGSKABELS KANN DIE LEISTUNG DES GERÄTES BEEINTRÄCHTIGEN. DIE VERBINDUNG ZU VIELER STECKER MIT EINER STECKDOSE KANN DEN SCHALTSCHRANK ÜBERLASTEN UND ZUM AUSLÖSEN VON TRENNSCHALTER/SICHERUNG ODER BEI FEHLFUNKTION VON TRENNSCHALTER/ SICHERUNG ZUM AUSBRUCH EINES FEUERS FÜHREN.
VORSICHT: Das Gerät entfernt von Vorhängen oder Gardinen, Heißluftregistern oder Heizungen aufstellen. Sicherstellen, dass das Gerät auf einer flachen Oberfläche aufgestellt wird, darauf achten, dass an allen Seiten ein Mindestabstand von 30 cm (1 Fuß) zu Wänden oder anderen Hindernissen besteht. Das Gerät nicht auf engem Raum aufstellen. Einen staub- und rauchfreien Aufstellort wählen, der vor direkter Sonneneinstrahlung geschützt ist. Das Gerät nicht in Außenbereichen betreiben, sofern die Netzsteckdose, an der das Gerät angeschlossen ist, nicht über eine Fehlerstrom-Schutzeinrichtung (RCD) verfügt.
VORSICHT: Das Gerät nicht in einem eingeschränkten oder engen Bereich betreiben, in dem die Belüftung unzureichend sein könnte. Dies könnte zu einer Überhitzung des Geräts und einer Beeinträchtigung der Leistung führen.
VORSICHT: Sicherstellen, dass Lufteinlass und -auslass nicht blockiert werden. Keine Gegenstände in die Geräteöffnungen einführen und nichts in diese hineinfallen lassen. Dies könnte zu einer Überhitzung des Geräts und einer Beeinträchtigung der Leistung führen.
VisionAire Family
der Mitte der Durchuss-Liniennummer bendet,
die von Ihrem Arzt verordnet wurde. Damit sich der
Durchussmesser im korrekten Winkel bendet,
muss an der Markierungslinie auf der Rück- und Vorderseite die gleiche Zahl sichtbar sein.
6. Der Konzentrator ist nun einsatzbereit.
VORSICHT: Für den Fall eines Stromausfalls, Alarms oder mechanischen Versagens empfiehlt der Hersteller eine alternative Sauerstoffzufuhr. Den Arzt oder Geräteanbieter fragen, welche Art von Reservesystem benötigt wird.
VORSICHT: Das vorgeschriebene Sauerstoffniveau muss unbedingt eingehalten werden. Einstellungen der Flussrate dürfen nur geändert werden, wenn dies auf Anweisung eines approbierten Arztes erfolgt.
VORSICHT: Der Gebrauch des Sauerstoffkonzentrators während des Schlafs darf nur auf Empfehlung durch qualifiziertes klinisches Personal erfolgen.
VORSICHT: Sicherstellen, dass der Konzentrator in aufrechter Position betrieben wird.
Hinweis: Die optimale Sauerstoffkonzentration wird innerhalb von 10 Minuten nach dem Einschalten des Geräts erreicht (90 % der Konzentration werden nach ca. 5 Minuten erreicht).
Im Normalfall muss der Durchussmesser an Ihrem
Gerät nicht eingestellt werden. Durch Drehen
des Einstellknopfs des Durchussmessers im Uhrzeigersinn wird der Sauerstoffuss zum Gerät
verringert oder ausgeschaltet.
Ausschalten des Konzentrators
Stellen Sie den E/A-Netzschalter auf die Stellung „A“, um das Gerät auszuschalten.
1. Das Gerät in dem Raum, in dem Sie sich hauptsächlich aufhalten, in der Nähe einer Netzsteckdose aufstellen.
2. Stecken Sie den Stromkabelstecker in die Netzsteckdose.
3. Stellen Sie den E/A-Netzschalter auf die Stellung „E“, um das Gerät einzuschalten. Ein lautes akustisches Warnsignal ertönt für ca. 5 Sekunden.
4. Die Anzeige für eine geringe Sauerstoffkonzentration bleibt für einige Minuten aktiv, bis die Sauerstoffkonzentration die Mindestkonzentration erreicht (nur bei Geräten mit Sauerstoffmonitor).
5. Drehen Sie den Einstellknopf des
Durchussmessers zum Einstellen des Flusses
der zusätzlichen Sauerstoffzufuhr nach links oder
rechts, bis sich die Kugel im Durchussmesser in
PN MN238-C4 A | Bedienungsanleitung 59 - GER
Page 60
VisionAire Family
Reinigung, Pflege und ordnungsgemäße Wartung
Schrank
WARNUNG: STROMSCHLAGGEFAHR. DAS GERÄT AUSSCHALTEN UND DAS NETZKABEL AUS DER NETZSTECKDOSE ZIEHEN, BEVOR DAS GERÄT GEREINIGT ODER GEWARTET WIRD, UM EINE STROMSCHLAGGEFAHR UND VERBRENNUNGSGEFAHR ZU VERMEIDEN. NUR DER GERÄTEANBIETER ODER EIN QUALIFIZIERTER WARTUNGSTECHNIKER DÜRFEN DIE ABDECKUNGEN ENTFERNEN ODER DAS GERÄT WARTEN.
WARNUNG: ES MUSS DARAUF GEACHTET WERDEN, DASS DER SAUERSTOFFKONZENTRATOR STETS TROCKEN BLEIBT UND KEINERLEI FLÜSSIGKEIT IN DAS GERÄT EINTRITT. ANDERNFALLS KANN ES ZU EINER FEHLFUNKTION ODER ZUR ABSCHALTUNG KOMMEN ODER ES BESTEHT EIN ERHÖHTES RISIKO EINES ELEKTRISCHEN SCHLAGS ODER VON VERBRENNUNGEN.
WARNUNG: KEINE ÖL-, FETT- ODER ERDÖLBASIERTEN ODER ANDEREN ENTZÜNDLICHEN PRODUKTE MIT DEM SAUERSTOFFKONZENTRATOR ODER DESSEN SAUERSTOFF BEINHALTENDEM ZUBEHÖR VERWENDEN. SAUERSTOFF BESCHLEUNIGT DIE VERBRENNUNG ENTZÜNDLICHER SUBSTANZEN.
WARNUNG: NUR SAUERSTOFFVERTRÄGLICHE LOTIONEN ODER SALBEN AUF WASSERBASIS VOR ODER WÄHREND DER SAUERSTOFFTHERAPIE VERWENDEN. NIEMALS LOTIONEN ODER SALBEN AUF ERDÖL- ODER ÖLBASIS VERWENDEN, UM EINE GEFÄHRDUNG DURCH BRÄNDE ODER VERBRENNUNGEN ZU VERMEIDEN.
WARNUNG: KEINE FLÜSSIGKEITEN DIREKT AUF DEM GERÄT VERWENDEN. ZU DEN NICHT VERTRÄGLICHEN CHEMISCHEN MITTELN ZÄHLEN UNTER ANDEREM: ALKOHOL UND PRODUKTE AUF ALKOHOLBASIS, KONZENTRIERTE PRODUKTE AUF CHLORBASIS (ETHYLCHLORID) UND PRODUKTE AUF ÖLBASIS (PINE-SOL®, LESTOIL®). DIESE DÜRFEN NICHT VERWENDET WERDEN, UM DAS KUNSTSTOFFGEHÄUSE DES SAUERSTOFFKONZENTRATORS ZU REINIGEN, DA SIE DEN KUNSTSTOFF BESCHÄDIGEN KÖNNTEN. GEHÄUSE, BEDIENFELD UND STROMKABEL NUR MIT EINEM SANFTEN HAUSHALTSREINIGER, DER MIT EINEM FEUCHTEN (NICHT NASSEN) TUCH ODER SCHWAMM AUFGETRAGEN WIRD, REINIGEN. ANSCHLIESSEND ALLE OBERFLÄCHEN TROCKEN WISCHEN. KEINE FLÜSSIGKEITEN IN DAS INNERE DES GERÄTES GELANGEN LASSEN.
WARNUNG: DAS GEHÄUSE, BEDIENFELD UND NETZKABEL MIT EINEM SCHONENDEN HAUSHALTSREINIGER REINIGEN. DAZU EIN FEUCHTES (ABER KEIN NASSES) TUCH ODER EINEN FEUCHTEN SCHWAMM VERWENDEN. ALLE BESTANDTEILE NACH DER REINIGUNG ABTROCKNEN. KEINE FLÜSSIGKEIT IN DAS GERÄT EINDRINGEN LASSEN. BESONDERS DARAUF ACHTEN, DASS DER SAUERSTOFFAUSLASS FÜR DEN ANSCHLUSS DER KANÜLE FREI VON STAUB, WASSER UND ANDEREN PARTIKELN BLEIBT.
Hinweis: Für einen ordnungsgemäßen Gebrauch stets den Anleitungen des Kanülenherstellers folgen. Die Einweg-Kanüle gemäß den Empfehlungen des Kanülenherstellers oder des Geräteanbieters ersetzen. Weiteres Zubehör erhalten Sie von Ihrem Geräteanbieter.
Hinweis: Der Hersteller empfiehlt, dieses Gerät nicht zu sterilisieren.
Das Gehäuse des Konzentrators nur mit einem sanften Haushaltsreiniger reinigen, der mit einem feuchten Tuch oder Schwamm aufgetragen wird.
Anschließend alle Oberächen trocken wischen. Keine
Flüssigkeiten in das Innere des Gerätes gelangen lassen. Der Geräteschrank sollte auf jeden Fall vor dem
Benutzerwechsel oder häuger gereinigt werden.
Befeuchterflasche (optional)
• Täglich den Wasserstand prüfen und falls nötig neues Wasser nachfüllen.
• Zur Reinigung und Desinfektion des Befeuchters den Anweisungen des Geräteanbieters oder den
Anweisungen, die mit der Befeuchterasche geliefert
wurden, folgen.
60 - GER PN MN238-C4 A | Bedienungsanleitung
Page 61
Warnzustände
Alle Warnsignale sind Warnsignale mit niedriger Priorität.
Warnsignal Bedeutung Maßnahme
Gelbe Warnleuchte für allgemeine Funktionsstörungen und Piepton
Gelbe Warnleuchte des Sauerstoffmonitors und Piepton
Gelbe Warnleuchte
für Stromausfall
und Piepton
Hoher Produktspeicherdruck
ODER niedriger
Produktspeicherdruck ODER Hohe Gerätetemperatur
Sauerstoffkonzentration gering
Stromausfall Sicherstellen, dass das Gerät mit einer funktionierenden
Sicherstellen, dass der Durchflussmesser die Mindestflussrate oder mehr durchlässt. Sicherstellen, dass die Kanüle weder umgeknickt noch verstopft ist. Alle Geräte entfernen, die dem Geräteausgang nachgeschaltet sind. Sicherstellen, dass rund um das Gerät ein Abstand von mindestens 12 Zoll besteht und alle Einlässe frei sind. Sicherstellen, dass der grobe Lufteinlass-Partikelfilter sauber und nicht verstopft ist. Sicherstellen, dass das Gerät sich im Betriebstemperaturbereich befindet. Sollte das Problem weiterhin bestehen, für die Wartung mit dem Geräteanbieter in Verbindung setzen.
Für die Wartung mit dem Geräteanbieter in Verbindung setzen.
Steckdose verbunden ist. Sicherstellen, dass der Schutzschalter nicht ausgelöst ist. Sollte das Problem weiterhin bestehen, für die Wartung mit dem Geräteanbieter in Verbindung setzen.
VisionAire Family
PN MN238-C4 A | Bedienungsanleitung 61 - GER
Page 62
VisionAire Family
Fehlerbehebung
Wenn Ihr VisionAire-Sauerstoffkonzentrator nicht ordnungsgemäß funktioniert, ziehen Sie die Tabelle auf den
folgenden Seiten zurate, um die möglichen Ursachen und Lösungen herauszunden, und wenden Sie sich bei
Bedarf an den Geräteanbieter.
Wenn das Gerät nicht funktioniert, schließen Sie die Nasenkanüle, Gesichtsmaske oder andere Zubehörteile an ein Reservesauerstoffgerät an.
Hinweis: Nicht versuchen, das Gerät – abgesehen von den in dieser Anleitung aufgeführten Störungsbehebungen – zu warten. Die Wartung liegt in der Verantwortung des Anbieters und wird vom Anbieter verfolgt.
Hinweis: Alle Anweisungen des Herstellers befolgen, damit die Garantie ihre Gültigkeit behält. Hinweis: Wenn das Gerät für einen längeren Zeitraum nicht benutzt wurde, muss es mindestens 15 Minuten laufen
gelassen werden, bevor das Stromausfall-Warnsignal aktiviert werden kann.
Problem Mögliche Ursache Lösung
Das Gerät funktioniert nicht. Durch einen Stromausfall wird ein akustisches Warnsignal aktiviert.
Begrenzter Sauerstofffluss Verunreinigte oder verstopfte
Beim Einsatz der Befeuchterflasche bildet sich im Sauerstoffschlauch Kondenswasser.
Pieptöne Unter „Sicherheitsfunktionen“
Das Stromkabel ist nicht mit einer Steckdose verbunden.
Es fließt kein Strom an der Steckdose.
Der Schutzschalter des Sauerstoffkonzentrators ist aktiviert.
Befeuchterflasche
Nasenkanüle, Gesichtsmaske, Katheter und/oder Sauerstoffzufuhrschlauch oder ein anderes Zubehörteil defekt.
Lockerer Kanülenschlauch Überprüfen Sie die Verbindung von
Das Gerät wird nicht ausreichend belüftet. Betriebstemperatur erhöht
finden Sie eine Beschreibung der akustischen Anzeigen.
Überprüfen Sie, ob der Stecker des Stromkabels ordnungsgemäß mit der Steckdose verbunden ist.
Überprüfen Sie die Stromquelle, den Stromschalter, die Sicherung oder den internen Schutzschalter.
Drücken Sie kurz auf den Schutzschalter an der Vorderseite des Gerätes.
Wenn der Schutzschalter ausgelöst wird oder das akustische Warnsignal weiterhin aktiv ist, nachdem das Gerät eingeschaltet wurde, wenden Sie sich an den Geräteanbieter.
Trennen Sie die Befeuchterflasche vom Sauerstoffauslass. Wenn der Fluss wiederhergestellt wurde, reinigen Sie die Flasche oder ersetzen Sie sie durch eine neue Befeuchterflasche.
Entfernen und prüfen Sie die Zubehörteile auf Knickstellen oder Verstopfungen. Bei Bedarf reparieren.
Kanülenschlauch und Bedienfeld. Das Gerät entfernt von Vorhängen oder Gardinen,
Heißluftregistern, Heizungen und Feuerstellen aufstellen. Darauf achten, dass an allen Seiten des Geräts ein Mindestabstand von 30,5 cm (12 Zoll) zu Wänden oder anderen Hindernissen besteht. Das Gerät nicht auf engem Raum aufstellen.
Die Befeuchterflasche mit KALTEM Wasser auffüllen.
NICHT ÜBERFÜLLEN. Den Sauerstoffschlauch trocknen lassen oder durch einen neuen Schlauch austauschen.
Bringen Sie den E/A-Netzschalter in die Stellung „A“, verwenden Sie die Ersatz­Sauerstoffzufuhr und wenden Sie sich sofort an den Geräteanbieter.
62 - GER PN MN238-C4 A | Bedienungsanleitung
Page 63
VisionAire Family
Problem Mögliche Ursache Lösung
Das gelbe Licht des Sauerstoffmonitors leuchtet weiterhin.
Alle übrigen Probleme. Bringen Sie den E/A-Netzschalter in die
Die Sauerstoffkonzentration beträgt ≤ 82 %.
Zubehör
Bringen Sie den E/A-Netzschalter in die Stellung „A“, verwenden Sie die Ersatz-Sauerstoffzufuhr (falls vorhanden) und wenden Sie sich sofort an den Geräteanbieter.
Stellung „A“, verwenden Sie die Ersatz­Sauerstoffzufuhr und wenden Sie sich sofort an den Geräteanbieter.
Um einen ordnungsgemäßen und sicheren Betrieb zu gewährleisten, bitte ausschließlich folgende Zubehörteile von CAIRE verwenden, die Ihnen von Ihrem Sauerstoff-Versorger bereitgestellt werden. Die Verwendung von Zubehörteilen, die hier nicht aufgeführt sind, kann die Funktion und/oder die Sicherheit des Geräts nachhaltig beeinträchtigen.
Standardzubehör der VisionAire-Serie
Luftbefeuchterflasche (6–15 LPM)
Kanüle, 7,6 m (25 Fuß) (max. 6 l/min)
Schlauch der Befeuchterflasche
Befestigung der Befeuchterflasche
Teilenummer – HU003-1
Teilenummer – CU002-4
Teilenummer – TU255-1
Teilenummer – F0655-1
Sauerstoffschlauch, 7,6 m Teilenummer – CU004-3 Schlauch-/
Kanülenanschluss OxySafe-Kit
(max. Flussrate 25 l/min, max. Druck 5,9 bar)
Hinweis: Landesspezifische Stecker sind, sofern oben angegeben, ebenfalls erhältlich. Falls weitere Ersatzteile bestellt werden müssen, wenden Sie sich bitte an CAIRE oder an Ihren Geräteanbieter.
Teilenummer – CU009-1
Teilenummer – 20629671
PN MN238-C4 A | Bedienungsanleitung 63 - GER
Page 64
VisionAire Family
EMV-Prüfung
Medizinische Geräte unterliegen hinsichtlich der elektromagnetischen Verträglichkeit (EMV) besonderen Vorsichtsmaßnahmen und müssen gemäß den in diesem Abschnitt bereitgestellten Informationen zur EMV installiert und in Betrieb genommen werden.
Leitlinien und Herstellererklärung – elektromagnetische Aussendungen
Der VisionAire ist für den Gebrauch in einer wie im Folgenden beschriebenen elektromagnetischen Umgebung bestimmt. Der Kunde oder Anwender des VisionAire muss sicherstellen, dass das Gerät in einer entsprechenden Umgebung verwendet wird.
Störaussendungsmessungen Übereinstimmung Elektromagnetische Umgebung – Leitlinien
HF-Aussendung CISPR 11
HF-Aussendung CISPR 11 Oberwellenaussendungen
IEC 61000-3-2
Spannungsschwankungen/ Flicker IEC 61000-3-3
Gruppe 1
Klasse B
stimmt überein
stimmt überein
Empfohlene Schutzabstände zwischen tragbaren und mobilen HF­Telekommunikationsgeräten und den VisionAire-Geräten
Der VisionAire ist für den Betrieb in einer elektromagnetischen Umgebung bestimmt, in der die HF-Störgrößen kontrolliert sind. Der Kunde oder der Anwender des VisionAire kann dazu beitragen, elektromagnetische Störungen zu vermeiden, indem er den Mindestabstand zwischen tragbaren und mobilen HF-Telekommunikationsgeräten (Sendern) und dem VisionAire – abhängig von der Ausgangsleistung des Kommunikationsgeräts – wie unten angegeben einhält.
Nennleistung des
Senders
W
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
Für Sender, deren maximale Nennleistung in obiger Tabelle nicht angegeben ist, kann der empfohlene Schutzabstand d in Metern (m) unter Verwendung der Gleichung ermittelt werden, die für die jeweilige Sendefrequenz gilt. Dabei ist P die maximale Nennleistung des Senders in Watt (W) gemäß Angabe des Senderherstellers.
ANMERKUNG 1 Bei 80 MHz und 800 MHz gilt der Schutzabstand für den höheren Frequenzbereich. ANMERKUNG 2 Diese Leitlinien sind möglicherweise nicht in allen Fällen anwendbar. Die Ausbreitung
elektromagnetischer Wellen wird durch Absorptionen und Reflexionen von Gebäuden, Gegenständen und Menschen beeinflusst.
von 150 kHz bis 80 MHz
d = 1,2√P
Trennungsabstand je nach Frequenz des Senders
Der VisionAire verwendet HF-Energie ausschließlich für den internen Betrieb. Daher ist die HF-Aussendung sehr gering, und es ist unwahrscheinlich, dass benachbarte elektronische Geräte gestört werden.
Der VisionAire ist für den Gebrauch in allen Einrichtungen einschließlich denen im Wohnbereich und solchen geeignet, die unmittelbar an ein öffentliches Versorgungsnetz angeschlossen sind, das auch Gebäude versorgt, die zu Wohnzwecken benutzt werden.
m
80 MHz bis 800 MHz
d = 1,2√P
von 800 MHz bis 2,5 GHz
d = 1,2√P
64 - GER PN MN238-C4 A | Bedienungsanleitung
Page 65
VisionAire Family
Leitlinien und Herstellererklärung – elektromagnetische Störfestigkeit
Der VisionAire ist für den Gebrauch in einer wie im Folgenden beschriebenen elektromagnetischen Umgebung bestimmt. Der Kunde oder Anwender des VisionAire muss sicherstellen, dass das Gerät in einer entsprechenden Umgebung verwendet wird.
Störfestigkeitsprüfung IEC 60601 Prüfpegel Übereinstimmungspegel Elektromagnetische
Elektromagnetische Umgebung – Leitlinien
IEC 61000-4-2
± 6 kV Kontakt ± 8 kV Luft
± 6 kV Kontakt ± 8 kV Luft
± 2 kV für
Schnelle transiente elektrische Störgrößen/ Bursts
IEC 61000-4-4
Netzleitungen
± 1 kV für Eingangs- und
± 2 kV für Netzleitungen
n. z.
Ausgangsleitungen
Stoßspannungen IEC 61000-4-5
± 1 kV Gegentaktspannung
± 2 kV Gleichtaktspannung
< 5 % U
T
(> 95 % Einbruch der U
)
T
für 0,5 Zyklen
± 1 kV Gegentaktspannung
± 2 kV Gleichtaktspannung
< 5 % UT (> 95 % Einbruch der U für 0,5 Zyklen
40 % UT
40 % UT (60 % Einbruch der U für 5 Zyklen
Spannungseinbrüche, Kurzzeitunterbrechungen
(60 % Einbruch der U
)
T
für 5 Zyklen
und Schwankungen der Versorgungsspannung gemäß IEC 61000-4-11
70 % UT (30 % Einbruch
der U
)
T
70 % UT (30 % Einbruch der U für 25 Zyklen
für 25 Zyklen
< 5 % UT
< 5 % UT (> 95 % Einbruch
der U
)
T
(> 95 % Einbruch der U für 5 Sekunden
für 5 Sekunden
Magnetfeld bei der Netzfrequenz gemäß
3 A/m 3 A/m
IEC 61000-4-8
HINWEIS: U
ist die Netzwechselspannung vor der Anwendung des Prüfpegels.
T
Umgebung – Leitlinien
Fußböden sollten aus Holz oder Beton bestehen oder mit Keramikfliesen versehen sein. Wenn der Fußboden mit synthetischem Material versehen ist, muss die relative Luftfeuchte mindestens 30 % betragen.
Die Qualität der Versorgungsspannung sollte der einer typischen Geschäfts­oder Krankenhausumgebung entsprechen.
Die Qualität der Versorgungsspannung sollte der einer typischen Geschäfts­oder Krankenhausumgebung entsprechen.
)
T
Die Qualität der Versorgungsspannung sollte der einer typischen Geschäfts-
)
oder Krankenhausumgebung
T
entsprechen. Wenn der Anwender des VisionAire auch bei Unterbrechungen der Netzstromversorgung auf einen fortgesetzten Betrieb angewiesen ist, wird empfohlen, den VisionAire
)
T
über eine unterbrechungsfreie Stromversorgung zu speisen.
)
T
Magnetfelder bei der Netzfrequenz sollten den typischen Werten, wie sie in der Geschäfts- und Krankenhausumgebung vorzufinden sind, entsprechen.
PN MN238-C4 A | Bedienungsanleitung 65 - GER
Page 66
VisionAire Family
Leitlinien und Herstellererklärung – elektromagnetische Störfestigkeit
Der VisionAire ist für den Gebrauch in einer wie im Folgenden beschriebenen elektromagnetischen Umgebung bestimmt. Der Kunde oder Anwender des VisionAire muss sicherstellen, dass das Gerät in einer entsprechenden Umgebung verwendet wird.
Störfestigkeitsprüfung IEC 60601 Prüfpegel Übereinstimmungspegel Elektromagnetische
Geleitete HF IEC 61000-4-6
Gestrahlte HF IEC 61000-4-3
3 Veff 150 kHz bis 80 MHz
3 V/m 80 MHz bis 2,5 GHz
3 Veff
3 V/m
Umgebung – Leitlinien
Tragbare und mobile Funkgeräte sollten in keinem geringeren Abstand zum VisionAire einschließlich der Leitungen verwendet werden als dem empfohlenen Schutzabstand, der nach der für die Sendefrequenz zutreffenden Gleichung berechnet wird.
Empfohlener Schutzabstand
d = 1,2√P
d = 1,2√P
= 80 MHz bis 800 MHz
d = 1,2√P = 800 MHz bis 2,5 GHz
Mit P als der Nennleistung des Senders in Watt (W) gemäß Angaben des Senderherstellers und d als empfohlenem Schutzabstand in Metern (m).
Die Feldstärke stationärer Funksender sollte bei allen Frequenzen gemäß einer Untersuchung vor Ort Übereinstimmungspegel sein.
a
geringer als der
b
In der Umgebung von Geräten, die das folgende Bildzeichen tragen, sind Störungen möglich:
ANMERKUNG 1 Bei 80 MHz und 800 MHz gilt der höhere Frequenzbereich. ANMERKUNG 2 Diese Leitlinien sind möglicherweise nicht in allen Fällen anwendbar. Die Ausbreitung
elektromagnetischer Wellen wird durch Absorptionen und Reflexionen von Gebäuden, Gegenständen und Menschen beeinflusst.
a
Die Feldstärke stationärer Sender, z. B. Basisstationen von Funktelefonen und mobilen Landfunkgeräten, Amateurfunkstationen, AM- und FM-Rundfunk- und Fernsehsendern, können theoretisch nicht genau vorherbestimmt werden. Um die elektromagnetische Umgebung hinsichtlich der stationären Sender zu ermitteln, sollte eine Standortanalyse erwogen werden. Wenn die gemessene Feldstärke an dem Standort, an dem der VisionAire benutzt wird, die obigen Übereinstimmungspegel überschreitet, sollte der VisionAire beobachtet werden, um die bestimmungsgemäße Funktion nachzuweisen. Wenn eine anormale Leistung festgestellt wird, können zusätzliche Maßnahmen erforderlich sein, z. B. eine veränderte Ausrichtung oder ein anderer Standort des VisionAire.
b
Über den Frequenzbereich von 150 kHz bis 80 MHz muss die Feldstärke unter 3 V/m liegen.
66 - GER PN MN238-C4 A | Bedienungsanleitung
Page 67
VisionAire Family
Entsorgungsmethode
Entsorgung gebrauchter und defekter Teile
Alle gebrauchten oder defekten CAIRE VisionAire­Sauerstoffkonzentratoren müssen ordnungsgemäß und in Übereinstimmung mit den örtlichen Gesetzen entsorgt werden.
Entsorgung des Gerätes
Zum Schutz der Umwelt muss der Konzentrator ordnungsgemäß und in Übereinstimmung mit den örtlichen Gesetzen entsorgt werden.
Klassifikation
Art des Schutzes vor elektrischem Schlag: Klasse II Der Schutz vor Stromschlag wird durch
eine doppelte Isolierung erreicht.
EIne Schutzerdung oder Sicherung
der Installationsbedingungen ist nicht erforderlich.
Grad des Schutzes vor elektrischem Schlag: Typ BF Das Gerät bietet einen bestimmten
Grad von Schutz gegen Stromschlag hinsichtlich
1) zulässigem Verluststrom;
2) Zuverlässigkeit der Schutzerdung (sofern vorhanden).
Nicht für die Direktanwendung am
Herzen geeignet.
Schutzgrad gegenüber schädlichem Eindringen von Wasser: Tropfwassergeschütztes Gerät –
Schutzart IP21
Schutz gegenüber dem Eindringen
von festen Fremdkörpern mit einem Durchmesser von mehr als 12,5 mm und Schutz gegenüber senkrecht fallendem Tropfwasser.
Zulässige Reinigungs- und Infektionskontrollmethoden: Siehe Abschnitt „Wartung“ in den
VisionAire-Wartungshandbüchern.
Grad der Anwendungssicherheit bei Vorhandensein entzündlicher anästhetischer Gase: Die Geräte sind für solche Anwendungen
nicht geeignet.
Betriebsmodus: Dauerbetrieb.
Hinweis: Für einen ordnungsgemäßen Gebrauch stets den Anleitungen des Kanülenherstellers folgen. Die Einweg-Kanüle gemäß den Empfehlungen des Kanülenherstellers oder des Geräteanbieters ersetzen. Weiteres Zubehör erhalten Sie von Ihrem Geräteanbieter.
Hinweis: Der Hersteller empfiehlt, dieses Gerät nicht zu sterilisieren.
Copyright © 2018 Chart Industries. CAIRE Inc. behält sich das Recht vor, die Vermarktung seiner Produkte einzustellen bzw. Preise, Werkstoffe, Ausrüstungsteile, Qualität, Beschreibung, Spezifikationen bzw. Prozesse ohne Vorankündigung zu einem beliebigen Zeitpunkt zu ändern, ohne dass hieraus irgendwelche Verpflichtungen oder Rechtsfolgen entstehen. Alle hier nicht ausdrücklich genannten Rechte bleiben im gesetzlich zulässigen Rahmen Chart Industries vorbehalten.
PN MN238-C4 A | Bedienungsanleitung 67 - GER
Page 68
VisionAire Family
Controles del usuario e indicadores de estado del sistema
Norma ISO 7000: símbolos gráficos que deben usarse en los equipos (índice y sinopsis)
Lea el manual del usuario antes de usar el equipo. Reg. n.º 1641
Mantenerlo alejado de la lluvia. Mantenerlo seco. Reg. n.º 0626
Nombre y domicilio del fabricante. Reg. n.º 3082
Precaución: consulte los documentos
ISO 7010: Símbolos gráficos: colores y símbolos de seguridad; símbolos de seguridad registrados
Directiva del Consejo 93/42/CEE sobre dispositivos médicos
que acompañan el producto. Reg. n.º 0434A
Número de catálogo. Reg. n.º 2493
Número de serie. Reg. n.º 2498
Este lado hacia arriba. Reg. n.º 0623
Frágil. Manipular con cuidado. Reg. n.º 0621
Debe leer el manual de instrucciones Reg. n.º M002
Mantener alejado de las llamas, del fuego y de las chispas. No acercar a fuentes de ignición abiertas. Se prohíbe fumar. Reg. n.º P003
No fumar cerca de la unidad ni mientras la esté utilizando. Reg. n.º P002
Parte aplicada tipo BF (grado de protección frente a choques eléctricos). Reg. n.º 5333
Advertencia. Reg. n.º W001
Representante autorizado en la Comunidad Europea Este dispositivo cumple con los requerimientos de la Directiva 93/42/ CEE para dispositivos médicos. Lleva la marca CE, tal como se muestra.
Símbolos internos
Mantener la unidad lejos de materiales
inamables, aceite y grasa.
Agencia de seguridad para la norma CAN/CSA C22.2 n.º 60601-1-14 para
equipos eléctricos médicos. Certicado
para los mercados de EE. UU. y Canadá, según las normas vigentes en ambos países.
No desmontar.
Cuando este símbolo aparece en el panel de alarma del dispositivo, indica que se ha detectado una interrupción de la fuente de alimentación externa.
Etiquetado 2018: Cuando este símbolo aparece en el panel de alarma del dispositivo, indica una baja concentración de oxígeno en la salida del dispositivo.
ON (interruptor de encendido)
OFF (interruptor de apagado)
Fecha de fabricación
Equipo de Clase II
21 CFR 801.15: Código de Normas Federales, Título 21
La legislación federal limita la venta de este dispositivo a médicos o por orden médica.
IEC 60601-1: Equipos médicos eléctricos, Parte 1: requisitos generales para la seguridad básica y el desempeño esencial.
Equipo IP21 a prueba de goteo
68 - SPA PN MN238-C4 A | Manual del usuario
Page 69
VisionAire Family
Directiva del Consejo 2012/19/UE: Directiva sobre residuos de aparatos eléctricos y electrónicos (RAEE)
RAEE
Este símbolo pretende recordar a los propietarios de los equipos que deben llevarlos a un centro de reciclado
al nal de su vida útil, de conformidad
con la Directiva sobre residuos de aparatos eléctricos y electrónicos (RAEE). Nuestros productos cumplirán con la Restricción sobre Sustancias Peligrosas (RoHS). No contendrán más que trazas de plomo o de otros materiales peligrosos.
Este producto puede estar cubierto por una o dos patentes, de EE. UU. o internacionales. Visite nuestro sitio web para obtener el listado de patentes aplicables. Pat.: patents.gtls.io.
Concentrador de oxígeno AirSep® VisionAire
Este manual le ayudará a familiarizarse con el concentrador de oxígeno VisionAire de CAIRE. Asegúrese de que lee y comprende toda la información contenida en esta guía antes de utilizar la unidad. En caso de que le surjan preguntas al respecto, el proveedor del equipo las contestará con gusto.
®
Qué es el concentrador de oxígeno
El aire que respiramos contiene aproximadamente un 21 % de oxígeno, un 78 % de nitrógeno y un 1 % de otros gases. En el concentrador de oxígeno VisionAire, el aire ambiente se aspira al interior de la máquina a través de las entradas de aire. A continuación, pasa a través de un material adsorbente llamado tamiz molecular. Este material separa el oxígeno del nitrógeno y permite que solo el oxígeno pase a través de él. El resultado es el suministro al usuario de un
ujo de oxígeno de alta concentración.
Aclaración: en ningún caso existe el peligro de reducir el oxígeno presente en una habitación al usar el concentrador de oxígeno.
PN MN238-C4 A | Manual del usuario 69 - SPA
Page 70
VisionAire Family
Por qué su médico le ha prescrito oxígeno
Muchas personas padecen multitud de enfermedades cardíacas, pulmonares y otras enfermedades respiratorias. Un número considerable de estas personas
pueden beneciarse del tratamiento con oxígeno
suplementario en casa, al viajar, o mientras realiza actividades cotidianas fuera de casa.
El oxígeno es un gas que constituye el 21 % del aire ambiente que respiramos. El cuerpo humano depende del suministro constante de este para funcionar
correctamente. Su médico le ha prescrito un ujo o una conguración para tratar su trastorno respiratorio
particular.
Aunque el oxígeno no se trata de una droga adictiva, la oxigenoterapia sin autorización puede ser peligrosa. Debe consultar con un médico antes de utilizar el concentrador de oxígeno. El proveedor de equipos que suministra el equipo de oxígeno le mostrará cómo ajustar el caudal que se le ha prescrito.
ADVERTENCIA: DEBEN EXHIBIRSE LETREROS DE "NO FUMAR: OXÍGENO EN USO" EN EL HOGAR O EN EL SITIO DONDE SE USE EL OXÍGENO. LOS USUARIOS Y LOS CUIDADORES SANITARIOS DEBEN RECIBIR INFORMACIÓN ACERCA DE LOS PELIGROS DE FUMAR AL USAR OXÍGENO MEDICINAL O EN PRESENCIA DE ESTE.
Perfil del operador
Los concentradores están destinados a suministrar oxígeno suplementario a los usuarios con molestias
debidas a alteraciones que afectan a la eciencia de
sus pulmones para transferir el oxígeno en el aire a su torrente sanguíneo. Los concentradores de oxígeno
jos (SOC) no almacenan o contienen oxígeno.
No necesitan ser rellenados y pueden recargarse en cualquier fuente de alimentación de CA o CC. El uso del concentrador de oxígeno requiere una prescripción médica y no está destinado a soporte vital.
Aunque puede prescribirse un tratamiento con oxígeno a usuarios de todas las edades, el usuario habitual de la oxigenoterapia es mayor de 65 años y padece diversas enfermedades respiratorias, incluida la enfermedad pulmonar obstructiva crónica (EPOC). Por lo general, los usuarios presentan buenas capacidades cognitivas, y deben ser capaces de comunicar su incomodidad. Si el usuario es incapaz de comunicar su incomodidad, o es incapaz de leer y comprender el etiquetado y las instrucciones de uso del concentrador, el uso se recomienda solo bajo la supervisión de alguien que pueda. Si se siente alguna molestia durante el uso del concentrador, se recomienda a los usuarios que se pongan en contacto con su proveedor de atención médica. Asimismo, se recomienda a los usuarios que dispongan de una fuente de oxígeno de respaldo (por ejemplo, un tanque de oxígeno) para contar con un suministro de oxígeno en caso de que se produzca un corte eléctrico o una avería del concentrador. El uso del concentrador de oxígeno no requiere de otras habilidades o capacidades singulares del usuario.
PRECAUCIÓN: el fabricante recomienda disponer de una fuente alternativa de oxígeno complementario en caso de corte eléctrico, condición de alarma o avería mecánica. Consulte a su médico o al proveedor del equipo para determinar el tipo de sistema de reserva que necesita.
Es muy importante seleccionar solo el nivel de oxígeno prescrito. No cambie la selección de flujo a menos que un profesional médico acreditado así se lo haya indicado.
El concentrador de oxígeno puede usarse cuando duerme si así se lo ha recomendado un profesional autorizado.
70 - SPA PN MN238-C4 A | Manual del usuario
Page 71
Características de seguridad
La siguiente información le ayudará a familiarizarse con las características de seguridad del concentrador de oxígeno VisionAire. Asegúrese de que lee y comprende toda la información contenida en este manual antes de manipular la unidad. En caso de que le surjan preguntas al respecto, el proveedor del equipo estará encantado de contestarlas.
Etiqueta de advertencia del dispositivo e indicadores de alarma.
• Motor del compresor: la salida del compresor está equipada con una válvula de alivio de presión que está calibrada a 280 kPa (40 psig). La seguridad térmica se garantiza mediante un termostato situado en el bobinado del estátor del compresor (135 °C/275 °F).
• Avería general: si se presentan algunas de las condiciones enumeradas a continuación, la luz de
avería general ( alarma sonora intermitente.
Esto incluye:
• Obstrucción del ujo de oxígeno, como un
pellizco o una torcedura en la cánula de suministro, provocada por la alta presión del tanque de producto
• Presión alta del tanque de producto del dispositivo de más de 33 psig (± 1)
• Presión baja del tanque de producto del dispositivo de menos de 5 psig (± 1)
• Temperatura alta del dispositivo de más de 135 °C (275 °F), provocada por la baja presión del tanque de producto si el interruptor térmico situado dentro del compresor se activa (lo que apaga el compresor)
) se encenderá y se activará una
VisionAire Family
• Control de oxígeno: El control de oxígeno detecta cualquier caída a una concentración de menos del 82 %. Si esto ocurre, la luz de advertencia (
) del concentrador de oxígeno se iluminará. Si la condición de O2 bajo persiste, se activará una alarma sonora intermitente, además de la luz de alarma.
• Corte de suministro eléctrico: En caso de que la unidad esté en funcionamiento y se produzca una pérdida de energía, la luz de advertencia de alimentación ( ) se iluminará y se activará una alarma sonora intermitente. Véase la imagen de la etiqueta.
• Filtro de producto: ltro de ≥ 10 μm
PN MN238-C4 A | Manual del usuario 71 - SPA
Page 72
VisionAire Family
Desembalaje del VisionAire
Compruebe que todos los componentes enumerados están incluidos en el paquete. Si falta algún elemento, póngase en contacto con el proveedor de su concentrador de oxígeno inmediatamente.
• Concentrador de oxígeno jo
Conocer su concentrador de oxígeno VisionAire
En primer lugar, familiarícese con las partes importantes del concentrador de oxígeno VisionAire.
A. Interruptor de encendido/apagado (I/0): inicia
y detiene el funcionamiento de la unidad.
B. Botón de reinicio del disyuntor: reinicia la unidad
después de una parada por sobrecarga eléctrica.
C. Horómetro: registra las horas totales de
funcionamiento de la unidad.
D. Medidor de ujo/perilla de ajuste: controla e indica
el caudal de oxígeno en litros por minuto (lpm).
E. Salida de oxígeno: ofrece la conexión para un
humidicador (si es necesario) o una cánula.
F. Asas superior y laterales: facilitan el transporte de
la unidad.
G. Etiqueta de advertencia y alarma
H. Etiqueta de especicaciones: presenta las
especicaciones eléctricas y el número de serie.
I. Cable de alimentación: permite la conexión de la
unidad a la toma eléctrica.
J. Ruedas: las cuatro ruedas permiten mover
fácilmente la unidad cuando sea necesario.
K. Hueco para la botella humidicadora: espacio
donde colocar la botella humidicadora opcional. L. Botella humidicadora (opcional) M. Salida de oxígeno para la botella humidicadora:
conexión para el tubo de oxígeno/cánula.
N. Salida de oxígeno para la botella humidicadora O. Tubo de la botella humidicadora P. Conector de la botella humidicadora
Q. Tubo de oxígeno/cánula
ADVERTENCIA: NO UTILICE CABLES DE EXTENSIÓN EN ESTA UNIDAD NI CONECTE DEMASIADOS ENCHUFES A LA MISMA TOMA ELÉCTRICA. EL USO DE CABLES DE EXTENSIÓN PODRÍA AFECTAR DE FORMA ADVERSA EL RENDIMIENTO DEL DISPOSITIVO. DEMASIADOS ENCHUFES EN UNA TOMA PUEDEN OCASIONAR UNA SOBRECARGA DEL PANEL ELÉCTRICO Y HACER QUE EL DISYUNTOR/FUSIBLE SE ACTIVE O DISPARE SI EL DISYUNTOR O EL FUSIBLE NO FUNCIONAN.
72 - SPA PN MN238-C4 A | Manual del usuario
Page 73
VisionAire Family
F
E
D
G
Ilustración 1a
C
L
F
B
K
A
G
J
I
N
M
H
O
Ilustración 1b
Q
PN MN238-C4 A | Manual del usuario 73 - SPA
P
Page 74
VisionAire Family
¡Importante!
Las instrucciones de seguridad se definen de la siguiente manera:
ADVERTENCIA: INFORMACIÓN DE SEGURIDAD IMPORTANTE SOBRE PELIGROS QUE PUEDEN CAUSAR LESIONES GRAVES.
PRECAUCIÓN: Información importante sobre la prevención de daños en la familia VisionAire.
Aclaración: información que requiere atención especial.
Indicaciones de uso
ADVERTENCIA: ES MUY IMPORTANTE SELECCIONAR SOLO EL NIVEL DE OXÍGENO PRESCRITO. NO CAMBIE LA SELECCIÓN DE FLUJO A MENOS QUE UN PROFESIONAL MÉDICO ACREDITADO ASÍ SE LO HAYA INDICADO.
ADVERTENCIA: LAS LEYES FEDERALES (DE LOS ESTADOS UNIDOS) LIMITAN LA VENTA O ALQUILER DE ESTE DISPOSITIVO POR PARTE DE UN MÉDICO U OTRO PROVEEDOR DE ATENCIÓN MÉDICA ACREDITADO.
ADVERTENCIA: LA UNIDAD NO DEBE USARSE COMO SOPORTE VITAL. EN LOS PACIENTES GERIÁTRICOS, PEDIÁTRICOS O CUALQUIER OTRO USUARIO INCAPAZ DE COMUNICAR SU INCOMODIDAD MIENTRAS USA ESTE DISPOSITIVO, PODRÍA SER NECESARIO UTILIZAR UN SISTEMA DE SUPERVISIÓN O VIGILANCIA ADICIONAL. LOS USUARIOS CON TRASTORNOS AUDITIVOS O VISUALES PUEDEN NECESITAR ASISTENCIA PARA CONTROLAR LAS ALARMAS. SI SIENTE MALESTAR O TIENE UNA EMERGENCIA MÉDICA, BUSQUE ASISTENCIA MÉDICA DE INMEDIATO.
Contraindicaciones de uso
ADVERTENCIA: EN ALGUNAS CIRCUNSTANCIAS, EL USO DE OXÍGENO NO PRESCRITO PUEDE SER PELIGROSO. ESTE DISPOSITIVO SOLO DEBERÁ UTILIZARSE CUANDO LO PRESCRIBA UN MÉDICO.
ADVERTENCIA: NO UTILIZAR EN PRESENCIA DE ANESTÉSICOS INFLAMABLES.
ADVERTENCIA: AL IGUAL QUE CON CUALQUIER DISPOSITIVO ELÉCTRICO, EL USUARIO PUEDE EXPERIMENTAR PERÍODOS DE FALTA DE FUNCIONAMIENTO COMO RESULTADO DE UN CORTE EN LA ENERGÍA ELÉCTRICA, O LA NECESIDAD DE TAREAS DE INSPECCIÓN EN EL CONCENTRADOR DE OXÍGENO POR PARTE DE UN TÉCNICO CUALIFICADO. EL CONCENTRADOR DE OXÍGENO NO ES APROPIADO PARA USUARIOS QUE EXPERIMENTARÍAN CONSECUENCIAS ADVERSAS EN LA SALUD COMO RESULTADO DE TAL INTERRUPCIÓN TEMPORAL.
Instrucciones de seguridad
ADVERTENCIA: REVISE CON ATENCIÓN Y FAMILIARÍCESE CON LA SIGUIENTE INFORMACIÓN IMPORTANTE DE SEGURIDAD ACERCA DEL CONCENTRADOR DE OXÍGENO VISIONAIRE INTENSITY.
ADVERTENCIA: NO UTILICE ESTE EQUIPO SIN ANTES LEER Y COMPRENDER ESTE MANUAL. SI NO COMPRENDE LAS ADVERTENCIAS E INSTRUCCIONES, CONTACTE CON EL PROVEEDOR DE SU EQUIPO ANTES DE INTENTAR USARLO; DE LO CONTRARIO, PODRÍAN PRODUCIRSE LESIONES O DAÑOS.
ADVERTENCIA: FUMAR MIENTRAS SE USA OXÍGENO ES LA CAUSA PRINCIPAL DE LESIONES POR INCENDIOS Y MUERTES RELACIONADAS. DEBE SEGUIR LAS ADVERTENCIAS DE SEGURIDAD:
ADVERTENCIA: NO PERMITA QUE HAYA CIGARRILLOS ENCENDIDOS, VELAS O LLAMAS EXPUESTAS EN LA MISMA HABITACIÓN EN QUE SE ENCUENTREN EL DISPOSITIVO O LOS ACCESORIOS QUE TRANSPORTAN OXÍGENO.
ADVERTENCIA: FUMAR MIENTRAS SE USA UNA CÁNULA DE OXÍGENO PUEDE CAUSAR QUEMADURAS FACIALES Y POSIBLEMENTE LA MUERTE.
ADVERTENCIA: RETIRAR LA CÁNULA Y COLOCARLA SOBRE LA ROPA, LA CAMA, EL SOFÁ U OTRO MATERIAL ACOLCHADO CAUSARÁ UNA LLAMARADA SI SE LA EXPONE A UN CIGARRILLO, UNA FUENTE DE CALOR O UNA LLAMA VIVA.
74 - SPA PN MN238-C4 A | Manual del usuario
Page 75
VisionAire Family
ADVERTENCIA: SI FUMA, DEBE SEGUIR SIEMPRE ESTOS TRES (3) IMPORTANTES PASOS ANTES DE HACERLO: APAGUE EL CONCENTRADOR DE OXÍGENO, RETIRE LA CÁNULA Y SALGA DE LA HABITACIÓN DONDE SE ENCUENTRE EL DISPOSITIVO.
ADVERTENCIA: DEBEN EXHIBIRSE LETREROS DE "NO FUMAR: OXÍGENO EN USO" EN EL HOGAR O EN EL SITIO DONDE SE USE EL OXÍGENO. LOS USUARIOS Y LOS CUIDADORES SANITARIOS DEBEN RECIBIR INFORMACIÓN ACERCA DE LOS PELIGROS DE FUMAR AL USAR OXÍGENO MEDICINAL O EN PRESENCIA DE ESTE.
ADVERTENCIA: LAS LEYES FEDERALES (DE LOS ESTADOS UNIDOS) RESTRINGEN ESTE DISPOSITIVO A SU VENTA O ALQUILER POR PARTE DE UN MÉDICO U OTRO PROVEEDOR DE ATENCIÓN MÉDICA ACREDITADO.
ADVERTENCIA: ESTE DISPOSITIVO SUMINISTRA OXÍGENO DE ALTA CONCENTRACIÓN, LO QUE LO HACE MUY INFLAMABLE. NO PERMITA QUE SE FUME NI LA PRESENCIA DE LLAMAS EXPUESTAS DENTRO DE LA MISMA HABITACIÓN EN LA QUE SE ENCUENTRE (1) ESTE DISPOSITIVO O (2) ALGÚN ACCESORIO QUE CONTENGA OXÍGENO. SI NO SE TIENE EN CUENTA ESTA ADVERTENCIA, PUEDEN PRODUCIRSE UN INCENDIO GRAVE, DAÑOS A LOS BIENES O HASTA LESIONES FÍSICAS O LA MUERTE.
ADVERTENCIA: NO USE SU CONCENTRADOR DE OXÍGENO EN PRESENCIA DE GASES INFLAMABLES. ESTO PUEDE GENERAR UN FUEGO RÁPIDO QUE PUEDE PROVOCAR DAÑOS A LOS BIENES, LESIONES CORPORALES O LA MUERTE.
ADVERTENCIA: NO DEJE UNA CÁNULA NASAL SOBRE O DEBAJO DE LA ROPA, LA CAMA O LOS ALMOHADONES DE UNA SILLA. SI LA UNIDAD ESTÁ ENCENDIDA PERO NO EN USO, EL OXÍGENO PODRÍA VOLVER INFLAMABLES ESOS MATERIALES. COLOQUE EL INTERRUPTOR DE ENCENDIDO I/0 EN LA POSICIÓN 0 (APAGADO) CUANDO EL CONCENTRADOR DE OXÍGENO NO ESTÉ EN USO.
ADVERTENCIA: NO USE ACEITE, GRASA, PRODUCTOS A BASE DE PETRÓLEO NI OTROS PRODUCTOS INFLAMABLES CON LOS ACCESORIOS DE TRANSPORTE DE OXÍGENO O CON EL CONCENTRADOR DE OXÍGENO. EL OXÍGENO ACELERA LA COMBUSTIÓN DE LAS SUSTANCIAS INFLAMABLES. SOLO DEBEN USARSE LOCIONES O BÁLSAMOS COMPATIBLES CON EL OXÍGENO, A BASE DE AGUA.
ADVERTENCIA: NO LUBRIQUE LOS CONECTORES, LAS CONEXIONES NI OTROS ACCESORIOS DEL CONCENTRADOR DE OXÍGENO PARA EVITAR EL RIESGO DE INCENDIO Y QUEMADURAS.
ADVERTENCIA: RIESGO DE DESCARGA ELÉCTRICA. APAGUE LA UNIDAD Y DESCONECTE EL CABLE DE ALIMENTACIÓN DE LA TOMA ELÉCTRICA ANTES DE LIMPIAR LA UNIDAD PARA PREVENIR UNA DESCARGA ELÉCTRICA ACCIDENTAL O EL RIESGO DE QUEMADURAS. ÚNICAMENTE EL PROVEEDOR DE SU EQUIPO O UN TÉCNICO DE SERVICIO CUALIFICADO DEBEN QUITAR LAS TAPAS O REALIZAR LA INSPECCIÓN DE LA UNIDAD.
ADVERTENCIA: HAY QUE TENER CUIDADO DE EVITAR QUE EL CONCENTRADOR DE OXÍGENO SE MOJE Y NO HAY QUE DEJAR QUE NINGÚN LÍQUIDO INGRESE A LA UNIDAD. ESTO PUEDE CAUSAR UNA AVERÍA O DESACTIVACIÓN DE LA UNIDAD Y PRODUCIR UN MAYOR RIESGO DE DESCARGA ELÉCTRICA O QUEMADURAS.
ADVERTENCIA: NO USE LÍQUIDO EN CONTACTO DIRECTO CON LA UNIDAD. A CONTINUACIÓN, SE INCLUYE UNA LISTA DE ALGUNOS AGENTES QUÍMICOS NO DESEABLES, ENTRE OTROS: ALCOHOL Y PRODUCTOS A BASE DE ALCOHOL, PRODUCTOS A BASE DE CLORO CONCENTRADO (CLORURO DE ETILENO) Y PRODUCTOS A BASE DE ACEITE (PINE-SOL®, LESTOIL®). ESTOS NO DEBEN USARSE PARA LIMPIAR LA CARCASA DE PLÁSTICO DEL CONCENTRADOR DE OXÍGENO, YA QUE PUEDEN DAÑAR EL PLÁSTICO DE LA UNIDAD.
ADVERTENCIA: LIMPIE EL GABINETE, EL PANEL DE CONTROL Y EL CABLE DE ALIMENTACIÓN SOLO CON UN LIMPIADOR DOMÉSTICO APLICADO CON UN PAÑO HÚMEDO (NO MOJADO) O UNA ESPONJA, Y LUEGO SEQUE TODAS LAS SUPERFICIES. NO PERMITA QUE NINGÚN LÍQUIDO SE INTRODUZCA EN EL DISPOSITIVO.
ADVERTENCIA: EL CONCENTRADOR DE OXÍGENO NO DEBE USARSE ADYACENTE A O APILADO CON OTROS EQUIPOS. SI EL USO ADYACENTE O APILADO ES INEVITABLE, EL DISPOSITIVO DEBE OBSERVARSE PARA VERIFICAR SU FUNCIONAMIENTO NORMAL.
ADVERTENCIA: SIEMPRE COLOQUE EL TUBO DE SUMINISTRO DE OXÍGENO Y LOS CABLES DE ALIMENTACIÓN DE FORMA TAL QUE EVITE EL RIESGO DE TROPIEZOS O DE ESTRANGULACIÓN ACCIDENTAL.
ADVERTENCIA: NO SE PERMITE REALIZAR MODIFICACIONES A ESTE EQUIPO.
ADVERTENCIA: EL USO DE CABLES Y ADAPTADORES DIFERENTES A LOS ESPECIFICADOS, CON EXCEPCIÓN DE LOS CABLES Y DE LOS ADAPTADORES VENDIDOS POR EL FABRICANTE DEL EQUIPO MÉDICO ELÉCTRICO COMO PIEZAS DE REPUESTO PARA COMPONENTES INTERNOS, PUEDE DERIVAR EN UNA MAYOR EMISIÓN DE MENOR INMUNIDAD DEL CONCENTRADOR DE OXÍGENO.
PN MN238-C4 A | Manual del usuario 75 - SPA
Page 76
VisionAire Family
ADVERTENCIA: LAS CONDICIONES DEL ENTORNO PUEDEN AFECTAR AL RENDIMIENTO DEL DISPOSITIVO. SITÚELO EN UN ENTORNO LIMPIO Y LIBRE DE ORGANISMOS NOCIVOS.
ADVERTENCIA: SOLO LOS USUARIOS FINALES, LOS CUIDADORES SANITARIOS CUALIFICADOS O TÉCNICOS CUALIFICADOS DEBEN MANIPULAR EL DISPOSITIVO. LOS NIÑOS NO DEBEN MANIPULAR EL DISPOSITIVO.
ADVERTENCIA: PARA GARANTIZAR QUE RECIBE LA CANTIDAD TERAPÉUTICA DE OXÍGENO ACORDE A SU PROBLEMA DE SALUD, LA UNIDAD VISIONAIRE DEBE UTILIZARSE CON LA COMBINACIÓN ESPECÍFICA DE PIEZAS Y ACCESORIOS QUE RESPONDAN A LAS ESPECIFICACIONES DEL FABRICANTE DEL CONCENTRADOR Y QUE SE UTILIZARON AL ESTABLECER LOS AJUSTES PARA USTED.
ADVERTENCIA: EL USO DE ESTE DISPOSITIVO A UNA ALTITUD, TEMPERATURA O HUMEDAD RELATIVA FUERA DE LOS VALORES ESPECIFICADOS EN ESTE MANUAL PUEDE AFECTAR NEGATIVAMENTE AL CAUDAL Y LA CONCENTRACIÓN DE OXÍGENO Y, EN CONSECUENCIA, A LA CALIDAD DEL TRATAMIENTO.
ADVERTENCIA: EL USO DE CIERTOS ACCESORIOS DE LA ADMINISTRACIÓN DE OXÍGENO NO ESPECIFICADOS PARA ESTE CONCENTRADOR DE OXÍGENO PUEDE AFECTAR DE FORMA ADVERSA A SU RENDIMIENTO. LOS ACCESORIOS RECOMENDADOS SE MENCIONAN DENTRO DE ESTE MANUAL.
ADVERTENCIA: USE ÚNICAMENTE EL VOLTAJE ELÉCTRICO INDICADO EN LA ETIQUETA DE ESPECIFICACIONES ADHERIDA AL DISPOSITIVO.
ADVERTENCIA: NO UTILICE CABLES DE EXTENSIÓN EN ESTA UNIDAD NI CONECTE DEMASIADOS ENCHUFES A LA MISMA TOMA ELÉCTRICA. EL USO DE CABLES DE EXTENSIÓN PODRÍA AFECTAR DE FORMA NEGATIVA AL RENDIMIENTO DEL DISPOSITIVO. DEMASIADOS ENCHUFES EN UNA TOMA PUEDEN OCASIONAR UNA SOBRECARGA DEL PANEL ELÉCTRICO Y HACER QUE EL DISYUNTOR/FUSIBLE SE ACTIVE O DISPARE SI EL DISYUNTOR O EL FUSIBLE NO FUNCIONAN.
PRECAUCIÓN: las leyes federales (de los Estados Unidos) limitan la venta o alquiler de este dispositivo por parte de un médico u otro proveedor de atención médica acreditado.
PRECAUCIÓN: no coloque la unidad de manera que el acceso al cable de alimentación resulte complicado.
PRECAUCIÓN: el concentrador debe ubicarse de forma tal que se puedan evitar el humo, los contaminantes o los gases tóxicos.
PRECAUCIÓN: asegúrese de que el concentrador se utiliza en posición vertical.
PRECAUCIÓN: siempre coloque el tubo de suministro de oxígeno y los cables de alimentación de forma tal que evite el riesgo de tropiezos.
PRECAUCIÓN: coloque la unidad alejada de cortinas, rejillas de conductos de ventilación o calefactores. Asegúrese de colocar la unidad en una superficie plana y de que todos los lados se encuentran al menos a 30 cm (1 pie) de una pared u otro elemento de obstrucción. No coloque la unidad en un área reducida. Elija una ubicación donde no haya polvo ni humo y apartada de la luz directa del sol. No utilice la unidad en el exterior a menos que se encuentre conectada a una toma eléctrica protegida por un interruptor accionado por corriente de pérdida a tierra (GFCI).
PRECAUCIÓN: no utilice esta unidad en un espacio restringido o reducido donde la ventilación pueda ser limitada. Esto puede ocasionar que el dispositivo se sobrecaliente y que el rendimiento se vea afectado.
PRECAUCIÓN: no permita que se bloqueen las aberturas de entrada y de salida de aire. NO deje caer ni inserte objetos en ninguna abertura del dispositivo. Esto puede ocasionar que el concentrador de oxígeno se sobrecaliente y que su rendimiento se vea perjudicado.
76 - SPA PN MN238-C4 A | Manual del usuario
Page 77
VisionAire Family
PRECAUCIÓN: el fabricante recomienda disponer de una fuente alternativa de oxígeno complementario en caso de corte eléctrico, condición de alarma o avería mecánica. Consulte a su médico o al proveedor del equipo para determinar el tipo de sistema de reserva que necesita.
PRECAUCIÓN: es muy importante seleccionar solo el nivel de oxígeno prescrito. No cambie la selección de flujo a menos que un profesional médico acreditado así se lo haya indicado.
PRECAUCIÓN: el concentrador de oxígeno puede usarse cuando duerme si así se lo ha recomendado un profesional autorizado.
PRECAUCIÓN: operar o guardar el concentrador de oxígeno fuera de su intervalo de temperatura normal de funcionamiento puede afectar de forma negativa al rendimiento de la unidad. Consulte la sección de especificaciones de este manual para conocer los límites de temperatura de funcionamiento y almacenamiento.
PRECAUCIÓN: en caso de alarma o si observa que el concentrador de oxígeno no funciona correctamente, consulte la sección de resolución de problemas de este manual. Si no puede resolver el problema, consulte con el proveedor de su equipo.
PRECAUCIÓN: si la alarma sonora es débil o no suena, consulte con el proveedor de su equipo inmediatamente.
PRECAUCIÓN: si el tubo de la botella humidificadora no está conectado correctamente al conector de la botella humidificadora o a la salida de oxígeno, puede producirse una fuga de oxígeno.
PRECAUCIÓN: por lo general, no debería necesitar ajustar el medidor de flujo de la unidad. Si gira la perilla de ajuste del medidor de flujo en el sentido de las agujas del reloj, reducirá y podrá cortar el flujo de oxígeno desde la unidad. Para su comodidad, el medidor de flujo está marcado con incrementos de ½ lpm. En las unidades equipadas con la opción de medidor de flujo de 2 lpm, el medidor de flujo está marcado con incrementos de 1/8 lpm para ajustar el flujo hasta a 2 lpm.
Aclaración: la cánula no debe estar torcida y puede usarse en una longitud total máxima de 7,6 m (25 pies).
Asegúrese de que la cánula esté bien insertada y quede firme. Debe escuchar o sentir el flujo de oxígeno en las puntas de la cánula nasal. Si parece que el oxígeno no fluye, compruebe primero que la esfera indicadora del medidor de flujo registra flujo. A continuación, coloque la punta de la cánula en un vaso de agua; si salen burbujas de la cánula, hay flujo de oxígeno. Si no aparecen burbujas, consulte la sección de resolución de problemas de este manual.
Aclaración: siempre siga las instrucciones del fabricante de la cánula para un uso correcto. Reemplace la cánula desechable de acuerdo con las recomendaciones del fabricante de la cánula o del proveedor de su equipo. El proveedor de su equipo dispone de accesorios de repuesto.
Aclaración: el concentrador de oxígeno VisionAire debe dejarse funcionar durante al menos cinco minutos a 2 lpm antes de usarlo.
El VisionAire es adecuado para su utilización por dos usuarios, siempre que el flujo combinado sea de un mínimo de 2 lpm y no exceda la capacidad máxima del concentrador.
Al proveedor del equipo: se recomienda usar los siguientes accesorios para la administración de oxígeno con el concentrador de oxígeno VisionAire Intensity:
• Botella humidificadora: Pieza n.º HU003-1
• Cánula nasal con un tubo de 2,1 m (7 pies; 6 lpm máx.): Pieza n.º CU002-1
• Kit OxySafe: Pieza n.º 20629671 Aclaración: el fabricante no recomienda la
esterilización de este equipo. Aclaración: si no se ha usado la unidad durante un
periodo de tiempo prolongado, necesita funcionar durante varios minutos antes de que se active la alarma de fallo de corriente.
Aclaración: el concentrador expulsa aire caliente desde la parte inferior de la unidad que puede decolorar permanentemente las superficies de revestimiento de suelos sensibles a la temperatura, como el vinilo. No debería utilizarse el concentrador sobre un suelo que sea sensible a la decoloración térmica. El fabricante no se hace responsable de la decoloración de los suelos.
Aclaración: para evitar la anulación de la garantía, siga todas las instrucciones del fabricante.
Aclaración: no intente realizar tareas de mantenimiento aparte de las soluciones posibles enumeradas en el manual.
PN MN238-C4 A | Manual del usuario 77 - SPA
Page 78
VisionAire Family
Aclaración: los equipos de comunicaciones portátiles y móviles de radiofrecuencia (RF) pueden afectar los equipos eléctricos médicos.
Aclaración: en ningún caso existe el peligro de reducir el oxígeno presente en una habitación al usar el concentrador de oxígeno.
Aclaración: al proveedor del equipo: se recomiendan los siguientes accesorios para la administración de oxígeno para usar con el concentrador de oxígeno:
• Cánula nasal con un tubo de 2,1 m (7 pies): PN 5408-SEQ
• Para ver accesorios recomendados adicionales, consulte el catálogo de accesorios (PN ML-LOX0010), disponible en www.CAIREmedical.com.
Se necesita un cortafuegos para usarlo con cualquier cánula.
• CAIRE ofrece el OxySafe™. Está destinado a emplearse conjuntamente con el concentrador de oxígeno. El OxySafe™ es un fusible térmico que detiene el flujo de gas en caso de que la cánula descendente o el tubo de oxígeno se enciendan y ardan hacia el OxySafe. Se coloca en línea con la cánula nasal o el tubo de oxígeno entre el paciente y la salida de oxígeno del sistema VisionAire.
Para un uso correcto del OxySafe, siempre consulte las instrucciones del fabricante (que vienen con cada conjunto de OxySafe).
CAIRE ofrece un conjunto de OxySafe™ que incluye un accesorio de la cánula para la cabeza, OxySafe ™ y un tubo de 2,1 m (7 pies): PN 20629671.
ADVERTENCIA: CET APPAREIL PRODUIT DE L’OXYGÈNE À CONCENTRATION ÉLEVÉE, FAVORISANT UNE COMBUSTION RAPIDE. NE PAS PERMETTRE DE FUMER OU DES FLAMMES NUES DANS LA MÊME CHAMBRE: (1) CET APPAREIL OU (2) TOUT ACCESSOIRE CONTENANT DE L’OXYGÈNE. NE PAS UTILISER DE PRODUITS À BASE D’HUILE, DE GRAISSE OU DE PÉTROLE SUR OU À PROXIMITÉ DE L’UNITÉ. DÉCONNECTER LE CORDON D’ALIMENTATION DE LA PRISE ÉLECTRIQUE AVANT DE NETTOYER OU DE FAIRE L’ENTRETIEN DE L’UNITÉ.
RISQUE DE CHOC ÉLECTRIQUE. NE PAS ENLEVER LES COUVERCLES LORSQUE L’UNITÉ EST BRANCHÉE. SEULS VOTRE FOURNISSEUR D’ÉQUIPEMENT OU UN TECHNICIEN DE SERVICE QUALIFIÉ DEVRAIT ENLEVER LES COUVERCLES OU FAIRE L’ENTRETIEN DE L’UNITÉ.
78 - SPA PN MN238-C4 A | Manual del usuario
Page 79
VisionAire Family
Especificaciones
VisionAire
Velocidades de flujo* 1-5 lpm
± 10 % del ajuste indicado, o 200 ml, el que sea mayor*
Dimensiones 35,8 cm ancho x 29,2 cm profundidad x 52,8 cm alto
(14,1 pulgadas de ancho x 11,5 pulgadas de profundidad x 20,8 pulgadas de alto)
Peso 13,6 kg (30 lb) Nivel de presión de sonido 45 dBa (1 a 5 lpm) Consumo de energía 290 vatios Concentración de O2 90 % (+5,5 %/-3 %) Requisitos eléctricos 115 VAC/60 Hz, 3,0 A
230 VAC/50 Hz, 1,5 A 230 VAC/60 Hz, 1,5 A
Ambiente de funcionamiento* 5 °C a 40 °C (41 °F a 104 °F) a altitudes de hasta 3048 m (10 000 pies) sobre
Altitud Entorno de almacenamiento -25 °C a 70 °C (-13 °F a 158 °F)
* Con base en el intervalo de presión atmosférica de 700 hPa a 1060 hPa a 21 °C (70 °F).
el nivel del mar. Humedad relativa de 15-95 % (sin condensación)
-381 a 3048 m (-1250 a 10,000 pies) (probado a 700 - 1060 hPa)
La vida útil esperada del dispositivo es de un mínimo de cinco años.
Consulte el Manual técnico (PN MN138-1) para conocer el nivel de potencia acústica.
ADVERTENCIA: EL USO DE ESTE DISPOSITIVO A UNA ALTITUD, TEMPERATURA O HUMEDAD RELATIVA FUERA DE LOS VALORES ESPECIFICADOS EN ESTE MANUAL PUEDE AFECTAR NEGATIVAMENTE AL CAUDAL Y LA CONCENTRACIÓN DE OXÍGENO Y, EN CONSECUENCIA, A LA CALIDAD DEL TRATAMIENTO.
ADVERTENCIA: EL USO DE ESTE DISPOSITIVO A UNA ALTITUD SUPERIOR A 3048 METROS (10 000 PIES), A UNA TEMPERATURA QUE NO SE ENCUENTRE ENTRE 5 °C Y 40 °C (41 °F A 104 °F) O A UNA HUMEDAD RELATIVA SUPERIOR AL 95 % AFECTARÁN NEGATIVAMENTE AL CAUDAL Y EL PORCENTAJE DE OXÍGENO Y, EN CONSECUENCIA, A LA CALIDAD DEL TRATAMIENTO.
PN MN238-C4 A | Manual del usuario 79 - SPA
Page 80
VisionAire Family
Instrucciones de funcionamiento
Revise la siguiente información antes de manipular el concentrador de oxígeno.
Aclaración: el concentrador expulsa aire caliente desde la parte inferior de la unidad que puede decolorar permanentemente las superficies de revestimiento de suelos sensibles a la temperatura, como el vinilo. No debería utilizarse el concentrador sobre un suelo que sea sensible a la decoloración térmica. El fabricante no se hace responsable de la decoloración de los suelos.
Botella humidificadora (opcional)
Si se necesita humidicación adicional para su
oxigenoterapia, siga los pasos siguientes cada vez que
llene o limpie el humidicador, que puede haber sido
instalado inicialmente para su uso.
1. Retire la botella humidicadora del hueco para la botella humidicadora.
2. Abra la botella humidicadora. Si la botella viene
llena, no realice este paso. Continúe con el paso 5.
3. Llene la botella humidicadora con agua fresca o
fría (es preferible el agua destilada) hasta la línea de llenado indicada en la botella. NO LLENAR POR ENCIMA DE LA LÍNEA.
4. Vuelva a conectar la cubierta superior a la botella
humidicadora.
5. Coloque la botella humidicadora en el hueco para la botella humidicadora en la parte posterior
del concentrador y conecte el tubo de la botella
humidicadora a la salida de oxígeno y al conector de la botella humidicadora.
PRECAUCIÓN: si el tubo de la botella humidificadora no está conectado correctamente al conector de la botella humidificadora o a la salida de oxígeno, puede producirse una fuga de oxígeno.
Conexión de la cánula
Conecte el tubo y la cánula a la salida de oxígeno de
la unidad, o a la salida de oxígeno del humidicador
opcional.
Aclaración: la cánula no debe estar torcida y puede usarse en una longitud total máxima de 7,6 m (25 pies).
Asegúrese de que la cánula esté bien insertada y quede firme. Debe escuchar o sentir el flujo de oxígeno en las puntas de la cánula nasal. Si parece que el oxígeno no fluye, compruebe primero que la esfera indicadora del medidor de flujo registra flujo. A continuación, coloque la punta de la cánula en un vaso de agua; si salen burbujas de la cánula, hay flujo de oxígeno. Si no aparecen burbujas, consulte la sección de resolución de problemas de este manual.
Siempre siga las instrucciones del fabricante de la cánula para un uso correcto. Reemplace la cánula desechable de acuerdo con las recomendaciones del fabricante de la cánula o del proveedor de su equipo. El proveedor de su equipo dispone de accesorios de repuesto.
Encendido del concentrador
ADVERTENCIA: DEBEN EXHIBIRSE LETREROS DE "NO FUMAR: OXÍGENO EN USO" EN EL HOGAR O EN EL SITIO DONDE SE USE EL OXÍGENO. LOS USUARIOS Y LOS CUIDADORES SANITARIOS DEBEN RECIBIR INFORMACIÓN ACERCA DE LOS PELIGROS DE FUMAR AL USAR OXÍGENO MEDICINAL O EN PRESENCIA DE ESTE.
ADVERTENCIA: NO UTILICE CABLES DE EXTENSIÓN EN ESTA UNIDAD NI CONECTE DEMASIADOS ENCHUFES A LA MISMA TOMA ELÉCTRICA. EL USO DE CABLES DE EXTENSIÓN PODRÍA AFECTAR DE FORMA ADVERSA EL RENDIMIENTO DEL DISPOSITIVO. DEMASIADOS ENCHUFES EN UNA TOMA PUEDEN OCASIONAR UNA SOBRECARGA DEL PANEL ELÉCTRICO Y HACER QUE EL DISYUNTOR/FUSIBLE SE ACTIVE O DISPARE SI EL DISYUNTOR O EL FUSIBLE NO FUNCIONAN.
80 - SPA PN MN238-C4 A | Manual del usuario
Page 81
VisionAire Family
PRECAUCIÓN: coloque la unidad alejada de cortinas, rejillas de conductos de ventilación o calefactores. Asegúrese de colocar la unidad en una superficie plana y de que todos los lados se encuentran al menos a 30 cm (1 pie) de una pared u otro elemento de obstrucción. No coloque la unidad en un área reducida. Elija una ubicación donde no haya polvo ni humo y apartada de la luz directa del sol. No utilice la unidad en el exterior a menos que se encuentre conectada a una toma eléctrica protegida por un interruptor accionado por corriente de pérdida a tierra (GFCI).
PRECAUCIÓN: no utilice esta unidad en un espacio restringido o reducido donde la ventilación pueda ser limitada. Esto puede ocasionar que el dispositivo se sobrecaliente y que el rendimiento se vea afectado.
PRECAUCIÓN: no permita que se bloqueen las aberturas de entrada y de salida de aire. NO deje caer ni inserte objetos en ninguna abertura del dispositivo. Esto puede provocar que
PRECAUCIÓN: el concentrador de oxígeno se sobrecaliente y que su rendimiento se vea perjudicado.
1. Coloque la unidad cerca de la toma eléctrica en la habitación en la que pasa la mayor parte del tiempo.
2. Inserte el enchufe del cable de alimentación en la toma eléctrica.
3. Coloque el interruptor de encendido I/0 en la posición "I" para encender la unidad. Sonará una alarma fuerte durante aproximadamente 5 segundos.
4. El indicador de la concentración de oxígeno permanecerá encendido durante unos minutos hasta que la concentración de oxígeno alcance
las especicaciones de concentración mínimas
(solo para las unidades equipadas con control de oxígeno).
5. Para ajustar el ujo de oxígeno suplementario, gire la perilla de ajuste del medidor de ujo hacia
la izquierda o la derecha hasta que la esfera que se
encuentra dentro del medidor de ujo se centre en la línea numerada con el ujo que le ha prescrito su médico. Para ver el medidor de ujo desde el
ángulo adecuado, tenga en cuenta que la línea posterior y la línea numerada deben verse como una sola línea.
6. El concentrador ya está preparado para su uso.
PRECAUCIÓN: el fabricante recomienda disponer de una fuente alternativa de oxígeno complementario en caso de corte eléctrico, condición de alarma o avería mecánica. Consulte a su médico o al proveedor del equipo para determinar el tipo de sistema de reserva que necesita.
PRECAUCIÓN: es muy importante seleccionar solo el nivel de oxígeno prescrito. No cambie la selección de flujo a menos que un profesional médico acreditado así se lo haya indicado.
PRECAUCIÓN: el concentrador de oxígeno puede usarse cuando duerme si así se lo ha recomendado un profesional autorizado.
PRECAUCIÓN: asegúrese de que el concentrador se utiliza en posición vertical.
Aclaración: la concentración de oxígeno óptima se alcanza en los 10 minutos siguientes al encendido del dispositivo (el 90 % de la concentración se obtiene después de aproximadamente 5 minutos).
Por lo general, no debería necesitar ajustar el medidor
de ujo de la unidad. Si gira la perilla de ajuste del medidor de ujo en el sentido de las agujas del reloj, reducirá y podrá cortar el ujo de oxígeno desde
la unidad.
Apagado del concentrador
Coloque el interruptor de encendido I/0 en la posición "0" para apagar la unidad.
PN MN238-C4 A | Manual del usuario 81 - SPA
Page 82
VisionAire Family
Limpieza, cuidados y mantenimiento adecuado
Gabinete
ADVERTENCIA: RIESGO DE DESCARGA ELÉCTRICA. APAGUE LA UNIDAD Y DESCONECTE EL CABLE DE ALIMENTACIÓN DE LA TOMA ELÉCTRICA ANTES DE LIMPIAR LA UNIDAD PARA PREVENIR UNA DESCARGA ELÉCTRICA ACCIDENTAL O EL RIESGO DE QUEMADURAS. ÚNICAMENTE EL PROVEEDOR DE SU EQUIPO O UN TÉCNICO DE SERVICIO CUALIFICADO DEBEN QUITAR LAS TAPAS O REALIZAR LA INSPECCIÓN DE LA UNIDAD.
ADVERTENCIA: HAY QUE TENER CUIDADO DE EVITAR QUE EL CONCENTRADOR DE OXÍGENO SE MOJE Y NO HAY QUE DEJAR QUE NINGÚN LÍQUIDO INGRESE A LA UNIDAD. ESTO PUEDE CAUSAR UNA AVERÍA O DESACTIVACIÓN DE LA UNIDAD Y PRODUCIR UN MAYOR RIESGO DE DESCARGA ELÉCTRICA O QUEMADURAS.
ADVERTENCIA: NO USE ACEITE, GRASA, PRODUCTOS A BASE DE PETRÓLEO NI OTROS PRODUCTOS INFLAMABLES CON LOS ACCESORIOS DE TRANSPORTE DE OXÍGENO O CON EL CONCENTRADOR DE OXÍGENO. EL OXÍGENO ACELERA LA COMBUSTIÓN DE LAS SUSTANCIAS INFLAMABLES.
ADVERTENCIA: USE SOLO LOCIONES O BÁLSAMOS A BASE DE AGUA COMPATIBLES CON EL OXÍGENO ANTES Y DURANTE LA OXIGENOTERAPIA. NUNCA USE LOCIONES NI BÁLSAMOS A BASE DE PETRÓLEO O ACEITES PARA EVITAR EL RIESGO DE INCENDIO Y QUEMADURAS.
ADVERTENCIA: NO USE LÍQUIDO EN CONTACTO DIRECTO CON LA UNIDAD. A CONTINUACIÓN, SE INCLUYE UNA LISTA DE ALGUNOS AGENTES QUÍMICOS NO DESEABLES, ENTRE OTROS: ALCOHOL Y PRODUCTOS A BASE DE ALCOHOL, PRODUCTOS A BASE DE CLORO CONCENTRADO (CLORURO DE ETILENO) Y PRODUCTOS A BASE DE ACEITE (PINE-SOL®, LESTOIL®). ESTOS NO DEBEN USARSE PARA LIMPIAR LA CARCASA DE PLÁSTICO DEL CONCENTRADOR DE OXÍGENO, YA QUE PUEDEN DAÑAR EL PLÁSTICO DE LA UNIDAD. LIMPIE EL GABINETE, EL PANEL DE CONTROL Y EL CABLE DE ALIMENTACIÓN SOLO CON UN LIMPIADOR DOMÉSTICO APLICADO CON UN PAÑO HÚMEDO (NO MOJADO) O UNA ESPONJA, Y LUEGO SEQUE TODAS LAS SUPERFICIES. NO PERMITA QUE NINGÚN LÍQUIDO SE INTRODUZCA EN EL DISPOSITIVO.
ADVERTENCIA: LIMPIE EL GABINETE, EL PANEL DE CONTROL Y EL CABLE DE ALIMENTACIÓN SOLO CON UN LIMPIADOR DOMÉSTICO APLICADO CON UN PAÑO HÚMEDO (NO MOJADO) O UNA ESPONJA, Y LUEGO SEQUE TODAS LAS SUPERFICIES. NO PERMITA QUE NINGÚN LÍQUIDO SE INTRODUZCA EN EL CONCENTRADOR. PRESTE ESPECIAL ATENCIÓN A LA SALIDA DE OXÍGENO PARA LA CONEXIÓN DE LA CÁNULA PARA ASEGURARSE DE QUE SE MANTENGA LIBRE DE POLVO, AGUA Y PARTÍCULAS.
Aclaración: siempre siga las instrucciones del fabricante de la cánula para un uso correcto. Reemplace la cánula desechable de acuerdo con las recomendaciones del fabricante de la cánula o del proveedor de su equipo. El proveedor de su equipo dispone de accesorios de repuesto.
Aclaración: el fabricante no recomienda la esterilización de este equipo.
Utilice un limpiador doméstico aplicado con un paño húmedo o una esponja para limpiar el exterior del
concentrador, y luego seque todas las supercies.
No permita que ningún líquido ingrese al dispositivo. El gabinete del dispositivo debe limpiarse, como mínimo, entre usuarios.
Botella humidificadora (opcional)
• Compruebe el nivel de agua diariamente y añada el agua necesaria.
• Para limpiar y desinfectar el humidicador,
siga las instrucciones del proveedor de su equipo o las instrucciones suministradas con la botella
humidicadora.
82 - SPA PN MN238-C4 A | Manual del usuario
Page 83
Condiciones de alarma
Todas las alarmas son alarmas de baja prioridad.
Alarma Indica Acción
Luz amarilla de
avería general y alarma sonora intermitente
Luz amarilla del control de oxígeno
y alarma
sonora intermitente Luz amarilla de corte
de suministro eléctrico
y alarma
sonora intermitente
presión alta del tanque de producto
O presión baja del tanque
de producto O temperatura alta del
dispositivo
concentración de oxígeno baja
corte de suministro eléctrico Asegúrese de que el dispositivo está enchufado en una
Asegúrese de que el medidor de flujo está abierto con el caudal mínimo o superior. Asegúrese de que la cánula no esté retorcida ni obstruida. Quite cualquier dispositivo conectado más allá de la salida del dispositivo. Asegúrese de que el dispositivo tiene al menos 30,5 cm (12 pulgadas) de espacio libre a todo su alrededor y que las entradas no están obstruidas. Asegúrese que el filtro externo de entrada para partículas gruesa está limpio y no está obstruido. Asegúrese de que la unidad se encuentra dentro del intervalo de la temperatura de funcionamiento. Si el problema persiste, comuníquese con el proveedor del equipo para inspección.
Comuníquese con el proveedor del equipo para inspección.
toma eléctrica que se sabe que funciona. Asegúrese que el interruptor está pulsado. Si el problema persiste, comuníquese con el proveedor del equipo para inspección.
VisionAire Family
PN MN238-C4 A | Manual del usuario 83 - SPA
Page 84
VisionAire Family
Resolución de problemas
Si su concentrador de oxígeno VisionAire no funciona correctamente, consulte la tabla en las páginas siguientes para conocer las posibles causas y soluciones y, si lo necesita, consulte al proveedor de su equipo.
Si no consigue que la unidad funcione, conecte la cánula nasal, la mascarilla o el accesorio correspondiente a un suministro de oxígeno suplementario de reserva.
Aclaración: no intente realizar tareas de mantenimiento aparte de las soluciones posibles enumeradas en este manual. El mantenimiento es responsabilidad del proveedor y será rastreado por el proveedor.
Aclaración: para evitar la anulación de la garantía, siga todas las instrucciones del fabricante. Aclaración: si no se ha usado la unidad durante un periodo de tiempo prolongado, necesita funcionar durante un
mínimo de 15 minutos antes de que se active la alarma de fallo de corriente.
Problema Causa posible Solución
La unidad no funciona. El estado de corte de suministro eléctrico hace que suene la alarma.
Flujo de oxígeno limitado. Botella humidificadora sucia u
Se forma condensación en el tubo de oxígeno cuando utiliza la botella humidificadora.
Cable de alimentación no conectado a la toma eléctrica.
No hay tensión en la toma eléctrica.
El disyuntor del concentrador de oxígeno está activado.
obstruida.
Cánula nasal, mascarilla, catéter o tubo de suministro de oxígeno, u otro accesorio, defectuosos.
Tubo de la cánula suelto. Compruebe la conexión del tubo de la cánula en
Ventilación deficiente de la unidad. Temperatura elevada de funcionamiento.
Compruebe que el enchufe del cable de alimentación está correctamente enchufado a la toma eléctrica.
Compruebe la fuente de alimentación, el interruptor de pared, los fusibles o los disyuntores dentro de la vivienda.
Presione (no mantenga presionado) el botón de reinicio del disyuntor en la parte frontal de la unidad.
Si el disyuntor se activa de nuevo o la alarma continúa sonando después de que la unidad se encienda, comuníquese inmediatamente con el proveedor de su equipo.
Quite la botella humidificadora de la salida de oxígeno. Si se restaura el flujo, límpiela o reemplácela por una botella humidificadora nueva.
Retire los accesorios y compruebe si presentan torceduras u obstrucciones. Reemplácelo si es necesario.
el panel de control. Asegúrese de que la unidad está alejada de
cortinas, rejillas de conductos de ventilación, calefactores y chimeneas. Asegúrese de colocar la unidad de manera de que todos los lados se encuentren al menos a 30,5 cm (12 pulgadas) de una pared u otro elemento de obstrucción. No coloque la unidad en un área reducida.
Vuelva a llenar la botella humidificadora con agua FRÍA.
NO LLENAR POR ENCIMA DE LA LÍNEA. Deje que el tubo de oxígeno se seque, o reemplácelo por un tubo nuevo.
84 - SPA PN MN238-C4 A | Manual del usuario
Page 85
VisionAire Family
Suena una alarma intermitente.
Problema Causa posible Solución
La luz del control de oxígeno permanece encendida (amarilla).
Todos los demás problemas. Coloque el interruptor I/0 en la posición 0, utilice
En la sección "Características de seguridad" puede consultar una descripción de los indicadores sonoros.
La concentración de oxígeno es ≤ 82 %.
Coloque el interruptor I/0 en la posición 0, utilice su suministro de oxígeno de reserva y consulte al proveedor de su equipo inmediatamente.
Coloque el interruptor I/0 en la posición 0, utilice su suministro de oxígeno de reserva (si se ha suministrado) y consulte con el proveedor de su equipo inmediatamente.
su suministro de oxígeno de reserva y consulte con el proveedor de su equipo inmediatamente.
Accesorios
Para obtener un rendimiento y seguridad adecuados, utilice solo los accesorios contenidos en la tabla suministrados por CAIRE a través del proveedor de su concentrador de oxígeno. El uso de accesorios que no se mencionan a continuación podría afectar de manera negativa al rendimiento o a la seguridad del concentrador.
Accesorios estándar de la familia VisionAire
Botella humidificadora (6–15 LPM)
Cánula, 7,6 m (25 pies) (6 lpm máx.)
Tubo de la botella humidificadora
Conector de la botella humidificadora
Pieza n.º HU003-1
Pieza n.º CU002-4
Pieza n.º TU255-1
Pieza n.º F0655-1
Tubo de oxígeno, 7,6 m
Conector del tubo/ cánula
Kit OxySafe (flujo máx. 25 lpm; presión máx. 87 psig)
Aclaración: puede haber opciones adicionales disponibles para cables de alimentación específicos del país que no se mencionan arriba. Póngase en contacto con CAIRE o con el proveedor de su concentrador de oxígeno si necesita encargar opciones alternativas.
Pieza n.º CU004-3
Pieza n.º CU009-1
Pieza n.º 20629671
PN MN238-C4 A | Manual del usuario 85 - SPA
Page 86
VisionAire Family
Prueba de EMC
Los equipos médicos requieren precauciones especiales con respecto a la compatibilidad electromagnética (EMC) y deben instalarse y ponerse en servicio según la información sobre compatibilidad electromagnética provista en esta sección.
Orientación y declaración del fabricante: emisiones electromagnéticas
El sistema VisionAire está destinado para su uso en el ambiente electromagnético que se especifica a continuación. El cliente o el usuario del sistema VisionAire deben verificar que se use en tal ambiente.
Prueba de emisiones Cumplimiento Ambiente electromagnético: orientación
Emisiones de radiofrecuencia
CISPR 11
Emisiones de radiofrecuencia
CISPR 11 Emisiones armónicas
IEC 61000-3-2 Fluctuaciones de voltaje/
emisiones de flicker IEC 61000-3-3
Grupo 1
Clase B
Cumple
Cumple
Distancias recomendadas de separación entre equipos portátiles y móviles de comunicaciones de radiofrecuencia y las unidades VisionAire
El VisionAire está destinado para su uso en un ambiente electromagnético en el que se controlan las interferencias de radiofrecuencia irradiada. El cliente o el usuario del sistema VisionAire pueden ayudar a prevenir la interferencia electromagnética al mantener una distancia mínima entre los equipos portátiles y móviles de comunicación por radiofrecuencia (transmisores) y el VisionAire, tal como se recomienda a continuación, de acuerdo con la potencia máxima de salida de los equipos de comunicaciones.
Potencia de salida
nominal máxima
del transmisor
W
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
Para transmisores con una potencia nominal máxima de salida no mencionada anteriormente, la distancia de separación recomendada d en metros (m) puede calcularse usando la ecuación aplicable a la frecuencia del transmisor, donde P es la potencia nominal máxima de salida del transmisor en vatios (W) de acuerdo con el fabricante del transmisor.
NOTA 1: A 80 MHz y 800 MHz, se aplica la distancia de separación para el rango de mayor frecuencia. NOTA 2: Estas pautas pueden no ser de aplicación en todas las situaciones. La propagación electromagnética se ve
afectada por la absorción y la reflexión de estructuras, objetos y personas.
Distancia de separación de acuerdo con la frecuencia del transmisor
de 150 kHz a 80 MHz
d =1.2√P
El sistema VisionAire usa energía de radiofrecuencia solo para su función interna. En consecuencia, sus emisiones de radiofrecuencia son muy bajas y no es probable que causen interferencias en equipos electrónicos cercanos.
El sistema VisionAire es adecuado para usar en todos los establecimientos, incluidos los establecimientos domésticos y aquellos conectados directamente a la red pública de energía eléctrica de bajo voltaje que alimenta edificios usados para fines domésticos.
m
de 80 MHz a 800 MHz
d =1.2√P
de 800 MHz a 2,5 GHz
d =1.2√P
86 - SPA PN MN238-C4 A | Manual del usuario
Page 87
VisionAire Family
Orientación y declaración del fabricante: inmunidad electromagnética
El sistema VisionAire está destinado para su uso en el ambiente electromagnético que se especifica a continuación. El cliente o el usuario del sistema VisionAire deben verificar que se use en tal ambiente.
Prueba de inmunidad
Ambiente electromagnético: orientación
IEC 61000-4-2
Ráfagas eléctricas de transitorios muy rápidos
IEC 61000-4-4
Sobrecarga IEC 61000-4-5
Nivel de prueba IEC 60601
Contacto de ± 6 kV Aire de ± 8 kV
± 2 kV para las líneas de alimentación
± 1 kV para líneas de entrada/salida
± 1 kV línea(s) a línea(s)
± 2 kV línea(s) a tierra
< 5 % U
T
(Caída > 95 % en UT) para 0,5 ciclos
Nivel de cumplimiento
Contacto de ± 6 kV Aire de ± 8 kV
± 2 kV para las líneas de alimentación
N/A
± 1 kV línea(s) a línea(s)
± 2 kV línea(s) a tierra < 5 % UT
(Caída > 95 % en UT) para 0,5 ciclos
Ambiente electromagnético: orientación
Los suelos deben ser de madera, hormigón o cerámica. Si los suelos están cubiertos con material sintético, la humedad relativa debe ser de al menos un 30 %.
La calidad de la red eléctrica debe ser la de un ambiente comercial u hospitalario típico.
La calidad de la red eléctrica debe ser la de un ambiente comercial u hospitalario típico.
Caídas de voltaje, interrupciones por cortocircuitos y variaciones de
40 % UT (Caída del 60 % en UT) para 5 ciclos
40 % UT (Caída del 60 % en UT) para 5 ciclos
voltaje en las líneas de entrada de las fuentes de alimentación IEC 61000-4-11
70 % UT (Caída del 30 % en UT) para 25 ciclos
< 5 % UT (Caída > 95 % en UT) para 5 segundos
70 % UT (Caída del 30 % en UT) para 25 ciclos
< 5 % UT (Caída > 95 % en UT) para 5 segundos
Campo magnético a frecuencia industrial
3 A/m 3 A/m
IEC 61000-4-8
NOTA: U
es el voltaje de CA de la red eléctrica antes de la aplicación del nivel de prueba.
T
La calidad de la red eléctrica debe ser la de un entorno comercial u hospitalario típico. Si el usuario del sistema VisionAire requiere un funcionamiento continuo durante cortes de la red eléctrica, se recomienda que el VisionAire se alimente desde una fuente de alimentación sin interrupciones (UPS).
Los campos magnéticos a frecuencia industrial deben tener un nivel característico de una ubicación típica en un entorno comercial u hospitalario típico.
PN MN238-C4 A | Manual del usuario 87 - SPA
Page 88
VisionAire Family
Orientación y declaración del fabricante: inmunidad electromagnética
El sistema VisionAire está destinado para su uso en el ambiente electromagnético que se especifica a continuación. El cliente o el usuario del sistema VisionAire deben verificar que se use en tal ambiente.
Prueba de inmunidad
Radiofrecuencia conducida
IEC 61000-4-6
RF irradiada IEC 61000-4-3
Nivel de prueba IEC 60601
3 Vrms de 150 kHz a 80 MHz
3 V/m de 80 MHz a 2,5 GHz
Nivel de cumplimiento Ambiente electromagnético:
orientación
3 Vrms
Los equipos portátiles y móviles de comunicaciones por radiofrecuencia no deben usarse más cerca de las piezas del VisionAire, incluidos los cables, que la distancia de separación recomendada
3 V/m
calculada a partir de la ecuación aplicable a la frecuencia del transmisor.
Distancia recomendada de separación
d =1.2√P
=1.2√P = 80 MHz a 800 MHz
d
d =1.2√P = 800 MHz a 2,5 GHz
donde P es la clasificación máxima de potencia de salida del transmisor en vatios (W) de acuerdo con el fabricante del transmisor y d es la distancia de separación recomendada en metros (m).
Las concentraciones de campo de transmisores fijos de RF, determinadas por un estudio electromagnético del
a
sitio,
deberían ser inferiores al nivel de cumplimiento en cada rango de frecuencia. b
Puede producirse interferencia en la cercanía de los equipos marcados con el símbolo siguiente:
NOTA 1: A 80 MHz y 800 MHz, se aplica el rango de mayor frecuencia. NOTA 2: Estas pautas pueden no ser de aplicación en todas las situaciones. La propagación electromagnética se ve
afectada por la absorción y la reflexión de estructuras, objetos y personas.
a
Las concentraciones de campo de los transmisores fijos, tales como las estaciones de base para teléfonos de radio (móviles/inalámbricos) y radios móviles terrestres, radios de aficionados y las transmisiones de radio de AM y FM y la transmisión de TV, no pueden predecirse teóricamente con precisión. Para evaluar el ambiente electromagnético debido a transmisores fijos de RF, debe considerarse un estudio electromagnético del sitio. Si la concentración de campo medida en la ubicación donde se usa el VisionAire supera el nivel correspondiente de cumplimiento de RF anterior, el VisionAire debe observarse para verificar su funcionamiento normal. Si se observa un rendimiento anormal, pueden necesitarse medidas adicionales, tales como la reorientación o la reubicación del sistema VisionAire.
b
En el rango de frecuencia de 150 kHz a 80 MHz, las concentraciones de campo deben ser inferiores a 3 V/m.
88 - SPA PN MN238-C4 A | Manual del usuario
Page 89
VisionAire Family
Método de eliminación
Eliminación de residuos
Todos los residuos del concentrador de oxígeno VisionAire de CAIRE deben eliminarse según los
métodos adecuados especicados por las autoridades
locales.
Eliminación del dispositivo
Con el n de preservar el medio ambiente, el
concentrador de oxígeno VisionAire de CAIRE debe
eliminarse según los métodos adecuados especicados
por las autoridades locales.
Clasificación
Tipo de protección frente a descargas eléctricas: Clase II La protección frente a descargas
eléctricas se consigue mediante un doble aislamiento.
No son necesarias la toma a tierra de
seguridad ni la dependencia de las condiciones de la instalación.
Grado de protección frente a descargas eléctricas: Tipo BF Equipo que proporciona un grado particular de protección frente a descargas eléctricas en lo que respecta a
1) corriente de fuga permitida;
2) abilidad de la puesta a tierra de seguridad
(si está presente).
No está destinado para aplicación cardíaca
directa.
Grado de protección frente a la penetración de agua con efectos perjudiciales: Equipo IP21 a prueba de goteo Protección frente a la penetración de objetos
sólidos extraños con más de 12,5 mm de diámetro y protección frente a goteos de agua de caída vertical.
Método permitido de limpieza y control de infección: Consulte la sección de mantenimiento en
los manuales de mantenimiento del sistema VisionAire.
Grado de seguridad de la aplicación en presencia de
gases anestésicos inamables:
El equipo no está preparado para tal aplicación.
Modo de funcionamiento: Funcionamiento continuo.
Aclaración: siempre siga las instrucciones del fabricante de la cánula para un uso correcto. Reemplace la cánula desechable de acuerdo con las recomendaciones del fabricante de la cánula o del proveedor de su equipo. El proveedor de su equipo dispone de accesorios de repuesto.
Aclaración: el fabricante no recomienda la esterilización de este equipo.
Copyright © 2018 Chart Industries. CAIRE Inc. se reserva el derecho de suspender sus productos o modificar sus precios, los materiales, el equipo, la calidad, las descripciones, las especificaciones o los procesos de sus productos en cualquier momento, sin previo aviso, y sin ningún compromiso o consecuencia. Queda reservado cualquier otro derecho que no se encuentre indicado explícitamente en este documento, según corresponda.
PN MN238-C4 A | Manual del usuario 89 - SPA
Page 90
VisionAire Family
Controlli utente e indicatori di stato del sistema
ISO 7000; Simboli grafici per l’utilizzo
sull’apparecchiatura - Indice e sinossi
Leggere il manuale dell’utente prima dell’uso. Reg. n. 1641
Tenere lontano dalla pioggia e conservare in luogo asciutto. Reg. n. 0626
Nome e indirizzo del produttore. Reg. n. 3082
Attenzione, consultare i documenti di
ISO 7010: Simboli grafici - Colori di sicurezza
e simboli di sicurezza - Simboli di sicurezza
registrati
Direttiva del Consiglio 93/42/CEE relativa ai
dispositivi medici
accompagnamento. Reg. n. 0434A
Numero di catalogo. Reg. n. 2493
Numero di serie. Reg. n. 2498
Lato alto. Reg. n. 0623
Fragile, maneggiare con cura. Reg. n. 0621
Leggere il manuale di istruzioni. Reg. n. M002
Conservare lontano da amme libere,
fuoco e scintille. È vietato fumare e utilizzare fonti di accensione aperte. Reg. n. P003
Non fumare nei pressi dell’unità o durante il suo utilizzo. Reg. n. P002
Parte applicata di tipo BF (grado di protezione da scossa elettrica). Reg. n. 5333
Avvertenza. Reg. n. W001
Rappresentante autorizzato per la Comunità Europea Il presente dispositivo è conforme ai requisiti della Direttiva 93/42/CEE relativa ai dispositivi medici e reca il marchio CE come illustrato.
Simboli interni
Tenere lontano da materiali
inammabili, olio e grasso.
Agenzia per la sicurezza per CAN/ CSA C22.2 N. 60601-1-14 per apparecchiature elettromedicali
Certicato per i mercati statunitense
e canadese in conformità ai relativi standard applicabili.
Non smontare.
Quando è presente sul pannello allarmi del dispositivo, indica che è stata rilevata un’interruzione dell’alimentazione esterna.
Etichettatura 2018: quando è presente sul pannello allarmi del dispositivo, indica una concentrazione di ossigeno ridotta nell’uscita del dispositivo.
ON (interruttore acceso)
OFF (interruttore spento)
Data di produzione
Apparecchiatura di Classe II
21 CFR 801.15: Code of Federal Regulations, Title 21 (Codice delle normative federali USA, Titolo 21)
La legge federale degli Stati Uniti limita la vendita del presente dispositivo ai soli medici o dietro prescrizione medica.
IEC 60601-1: Apparecchi elettromedicali Parte 1 Prescrizioni generali per la sicurezza di base e le prestazioni essenziali
Apparecchiatura a tenuta stagna - IP21
90 - ITL PN MN238-C4 A | Manuale dell’utente
Page 91
VisionAire Family
Direttiva del Consiglio 2012/19/UE: rifiuti di apparecchiature elettriche ed elettroniche (RAEE)
RAEE
Questo simbolo ha lo scopo di ricordare ai proprietari dell’apparecchio di consegnarlo presso un centro per il riciclaggio al termine della durata utile, in conformità alla Direttiva sui
riuti di apparecchiature elettriche ed
elettroniche (RAEE). I nostri prodotti sono conformi alle restrizioni dettate dalla Direttiva sulle sostanze pericolose (RoHS). È possibile che essi contengano tracce, ma non quantità rilevanti, di piombo o altri materiali pericolosi.
Questo prodotto può essere coperto da uno o più brevetti, validi negli Stati Uniti e all’estero. Visitare il nostro sito Web che segue per un elenco dei brevetti
Concentratore di ossigeno AirSep® VisionAire
Il presente manuale dell’utente consente di acquisire dimestichezza con il concentratore di ossigeno VisionAire prodotto da CAIRE. Assicurarsi di aver letto e compreso tutte le informazioni contenute in questa guida prima di utilizzare il concentratore. In caso di domande, rivolgersi al fornitore dell’apparecchiatura.
®
Cosa si intende per Concentratore di ossigeno
L’aria che respiriamo è composta all’incirca dal 21% di ossigeno, 78% di azoto e 1% di altri gas. Nel concentratore di ossigeno VisionAire, l’aria dell’ambiente viene aspirata nella macchina attraverso gli ingressi dell’aria, passando poi attraverso un materiale assorbente denominato setaccio molecolare. Questo materiale separa ossigeno e azoto, consentendo il passaggio solo dell’ossigeno. Il risultato è
l’erogazione all’utente di un usso di ossigeno ad alta
concentrazione.
Nota: non sussiste mai il rischio di esaurire l’ossigeno nella stanza in cui si usa il concentratore di ossigeno.
PN MN238-C4 A | Manuale dell’utente 91 - ITL
Page 92
VisionAire Family
Perché il medico prescrive l’ossigenoterapia?
Molte persone soffrono di svariate patologie cardiache, polmonari e respiratorie di altro tipo. Un numero
signicativo di tali persone può trarre benecio da
un’ossigenoterapia supplementare a casa, in viaggio o durante la partecipazione ad attività quotidiane lontano da casa.
L’ossigeno è un gas che compone il 21% dell’aria che respiriamo. Per funzionare correttamente, l’organismo ne richiede una fornitura costante. Il medico prescrive
un usso o un’impostazione per fronteggiare una
condizione respiratoria particolare.
Benché l’ossigeno sia un farmaco che non produce dipendenza, l’ossigenoterapia non autorizzata può essere pericolosa. È necessario consultare un medico prima di utilizzare questo concentratore di ossigeno. Il fornitore dell’apparecchio per l’ossigeno dimostrerà come impostare la portata prescritta.
AVVERTENZA: IN CASA O NEL LUOGO IN CUI È UTILIZZATO L’OSSIGENO OCCORRE AFFIGGERE CARTELLI "NON FUMARE – OSSIGENO IN USO" IN MODO CHE SIANO BEN VISIBILI. GLI UTENTI E COLORO CHE PRESTANO ASSISTENZA DEVONO ESSERE INFORMATI DEI PERICOLI DEL FUMO IN PRESENZA O DURANTE L’UTILIZZO DI OSSIGENO MEDICO.
Profilo operatore
I concentratori sono destinati a fornire ossigeno supplementare agli utenti che soffrono di disagi
causati da disturbi che incidono sull’efcienza dei
polmoni nel trasferimento dell’ossigeno nell’aria al sangue. I concentratori di ossigeno stazionari (SOC) non conservano né contengono ossigeno. Non devono essere riforniti e possono essere ricaricati ovunque sia disponibile alimentazione CA o CC. L’utilizzo del concentratore di ossigeno richiede una prescrizione del medico e non è destinato all’uso come salvavita.
Sebbene l’ossigenoterapia possa essere prescritta a utenti di tutte le età, il tipico utente dell’ossigenoterapia ha superato i 65 anni e soffre di svariate malattie respiratorie, tra cui la broncopneumopatia cronica ostruttiva (BPCO). Gli utenti in genere dispongono di buone capacità cognitive e devono essere in grado di comunicare il disagio. Se l’utente non è in grado di comunicare il disagio o di leggere e comprendere le etichette e le istruzioni per l’uso del concentratore, l’utilizzo è consigliato solo sotto supervisione. In caso di disagio durante l’uso del concentratore, gli utenti sono invitati a rivolgersi al proprio fornitore di assistenza sanitaria. Gli utenti sono inoltre invitati a disporre di una fonte di ossigeno di riserva (ovvero, una bombola di ossigeno) in caso di interruzione dell’alimentazione o guasto al concentratore. Non sono previste altre abilità peculiari o necessarie dell’utente per l’uso del concentratore.
ATTENZIONE: il produttore consiglia di tenere a disposizione un’altra sorgente di ossigeno supplementare in caso di interruzioni di corrente, condizioni di allarme o guasti meccanici. Per conoscere il tipo di sistema di riserva necessario, rivolgersi al proprio medico o al fornitore dell’apparecchiatura.
È molto importante selezionare solo il livello di
ossigeno prescritto. Non modificare la selezione
del flusso se non in caso di precise istruzioni da
parte di un medico autorizzato.
È possibile utilizzare il concentratore di ossigeno
durante il sonno solo su raccomandazione di un
medico qualificato.
92 - ITL PN MN238-C4 A | Manuale dell’utente
Page 93
Funzionalità di sicurezza
Le seguenti informazioni consentono di acquisire dimestichezza con le funzionalità di sicurezza del concentratore di ossigeno VisionAire. Assicurarsi di aver letto e compreso tutte le informazioni in esso contenute prima di utilizzare l’unità. In caso di domande, rivolgersi al fornitore dell’apparecchiatura.
Etichetta di avvertenza del dispositivo e display allarmi.
• Motore compressore: nell’uscita del compressore vi è una valvola di massima pressione calibrata a 280 kPa (40 psig). La sicurezza termica è garantita da un termostato situato nell’avvolgimento dello statore del compressore (135° C / 275° F).
• Malfunzionamento generale: se si verica una delle
condizioni elencate di seguito, si accenderà la spia di malfunzionamento generale (
allarme acustico intermittente.
Le condizioni includono:
• ostruzione del usso di ossigeno come una
cannula di erogazione piegata o schiacciata, attivata da una pressione elevata del serbatoio prodotto
• condizione di pressione elevata del serbatoio prodotto del dispositivo superiore a 33 psig (±1)
• condizione di pressione ridotta del serbatoio prodotto del dispositivo inferiore a 5 psig (±1)
• temperatura elevata del dispositivo superiore a 135° C (275° F), attivata da una pressione ridotta del serbatoio prodotto se l’interruttore termico collocato all’interno del compressore scatta (spegnendo il compressore)
) e si attiverà un
VisionAire Family
• Monitor dell’ossigeno: rileva qualsiasi calo della concentrazione al di sotto dell’82%. In tal caso, si accenderà la spia di avvertenza concentrazione di ossigeno bassa ( persiste, oltre alla spia di allarme si attiverà anche un allarme acustico intermittente.
• Guasto di alimentazione: nel caso in cui l’unità sia in
funzione e si verichi una perdita di alimentazione,
si accenderà la spia di avvertenza alimentazione (
) e si attiverà un allarme acustico
intermittente. Vedere l’immagine dell’etichetta.
• Filtro prodotto: ltro ≥ 10 μm
). Se la condizione di O2 basso
PN MN238-C4 A | Manuale dell’utente 93 - ITL
Page 94
VisionAire Family
Disimballaggio del VisionAire
Vericare che la confezione contenga tutti i
componenti elencati di seguito. In caso di elementi mancanti, contattare immediatamente il fornitore di ossigeno.
• Concentratore di ossigeno sso
Cenni preliminari sul concentratore di ossigeno VisionAire
Anzitutto, acquisire familiarità con le parti importanti del concentratore di ossigeno VisionAire.
A. Interruttore di accensione/spegnimento (On/Off,
I/0): avvia e interrompe il funzionamento dell’unità.
B. Pulsante di ripristino interruttore automatico:
ripristina l’unità dopo un arresto per sovraccarico elettrico.
C. Contatore: registra le ore di funzionamento totali
dell’unità.
D. Manopola di regolazione/Flussometro: controlla e
indica la portata di ossigeno in litri al minuto (l/min).
E. Uscita dell’ossigeno: fornisce il collegamento per
un umidicatore (se richiesto) o cannula.
F. Maniglie laterali e superiori: consentono di
trasportare comodamente l’unità.
G. Etichetta di avvertenza e allarmi.
H. Etichetta delle speciche: visualizza le speciche
elettriche e numero di serie.
I. Cavo di alimentazione: consente il collegamento
dell’unità a una presa elettrica.
J. Ruote: quattro ruote che consentono di spostare
facilmente l’unità, ove necessario.
K. Rientranza per acone umidicatore: are in cui
collocare il acone umidicatore opzionale. L. Flacone umidicatore (opzionale). M. Presa dell’ossigeno acone umidicatore:
collegamento per il cannula/tubo dell’ossigeno.
N. Presa dell’ossigeno acone umidicatore. O. Tubo acone umidicatore. P. Raccordo acone umidicatore.
Q. Cannula/tubo dell’ossigeno.
AVVERTENZA: NON UTILIZZARE PROLUNGHE CON QUESTA UNITÀ O EVITARE DI COLLEGARE TROPPE SPINE ALLA STESSA PRESA ELETTRICA. L’USO DI PROLUNGHE PUÒ AVERE EFFETTI AVVERSI SULLE PRESTAZIONI DEL DISPOSITIVO. IL COLLEGAMENTO DI TROPPE SPINE A UN’UNICA PRESA PUÒ PROVOCARE UN SOVRACCARICO DEL QUADRO ELETTRICO, CAUSANDO L’ATTIVAZIONE DELL’INTERRUTTORE/ FUSIBILE O UN INCENDIO IN CASO DI MANCATO FUNZIONAMENTO DELL’INTERRUTTORE/FUSIBILE.
94 - ITL PN MN238-C4 A | Manuale dell’utente
Page 95
VisionAire Family
F
E
C
L
F
P
B
K
A
G
G
J
Figura 1a
I
N
M
H
O
Figura 1b
P
Q
PN MN238-C4 A | Manuale dell’utente 95 - ITL
Page 96
VisionAire Family
Importante!
Le istruzioni per la sicurezza sono definite come indicato di seguito:
AVVERTENZA: IMPORTANTI INFORMAZIONI PER LA SICUREZZA RELATIVE AI PERICOLI CHE POSSONO CAUSARE GRAVI LESIONI.
ATTENZIONE: informazioni importanti per evitare danni a VisionAire Family.
Nota: informazioni che richiedono particolare attenzione.
Indicazioni per l’uso
AVVERTENZA: È MOLTO IMPORTANTE SELEZIONARE SOLO IL LIVELLO DI OSSIGENO PRESCRITTO. NON MODIFICARE LA SELEZIONE DEL FLUSSO SE NON IN CASO DI PRECISE ISTRUZIONI DA PARTE DI UN MEDICO AUTORIZZATO.
AVVERTENZA: LA LEGGE FEDERALE (USA) LIMITA LA VENDITA O IL NOLEGGIO DI QUESTO DISPOSITIVO DIETRO PRESCRIZIONE DI UN MEDICO O DI UN ALTRO PROFESSIONISTA SANITARIO QUALIFICATO.
AVVERTENZA: QUESTA UNITÀ NON DEVE ESSERE UTILIZZATA COME SUPPORTO SALVAVITA. I PAZIENTI GERIATRICI, PEDIATRICI O ALTRI UTENTI INCAPACI DI COMUNICARE I PROPRI DISTURBI DURANTE L’USO DI QUESTO DISPOSITIVO POTREBBERO NECESSITARE DI UN MONITORAGGIO AGGIUNTIVO. GLI UTENTI CON DANNI ALL’UDITO E/O ALLA VISTA POTREBBERO NECESSITARE DI ASSISTENZA PER MONITORARE GLI ALLARMI. SE SI MANIFESTANO DISTURBI O IN CASO DI EMERGENZA MEDICA, RIVOLGERSI IMMEDIATAMENTE A UN MEDICO.
96 - ITL PN MN238-C4 A | Manuale dell’utente
Controindicazioni all’uso
AVVERTENZA: IN ALCUNE CIRCOSTANZE, L’USO DI OSSIGENO NON PRESCRITTO PUÒ ESSERE PERICOLOSO. QUESTO DISPOSITIVO VA UTILIZZATO SOLO DIETRO PRESCRIZIONE MEDICA.
AVVERTENZA: NON UTILIZZARE IN PRESENZA DI ANESTETICI INFIAMMABILI.
AVVERTENZA: ANALOGAMENTE A QUALSIASI ALTRO DISPOSITIVO ELETTRICO, POTREBBERO VERIFICARSI PERIODI DI INATTIVITÀ DOVUTI ALL’INTERRUZIONE DELL’ALIMENTAZIONE ELETTRICA, OPPURE ALLA NECESSITÀ DI FAR RIPARARE IL CONCENTRATORE DI OSSIGENO DA UN TECNICO QUALIFICATO. IL CONCENTRATORE DI OSSIGENO NON È INDICATO PER UTENTI CHE A SEGUITO DELL’INTERRUZIONE TEMPORANEA POTREBBERO SUBIRE CONSEGUENZE NEGATIVE SULLA PROPRIA SALUTE.
Linee guida per la sicurezza
AVVERTENZA: ESAMINARE ATTENTAMENTE E ACQUISIRE DIMESTICHEZZA CON LE SEGUENTI INFORMAZIONI DI SICUREZZA IMPORTANTI CHE RIGUARDANO IL CONCENTRATORE DI OSSIGENO VISIONAIRE INTENSITY.
AVVERTENZA: NON UTILIZZARE QUESTO DISPOSITIVO SENZA PRIMA AVER LETTO E COMPRESO IL PRESENTE MANUALE. PER EVITARE POSSIBILI LESIONI O DANNI, SE NON SI È IN GRADO DI COMPRENDERE LE AVVERTENZE E LE ISTRUZIONI, CONTATTARE IL FORNITORE DELL’APPARECCHIATURA PRIMA DI PROCEDERE ALL’UTILIZZO.
AVVERTENZA: FUMARE DURANTE L’USO DELL’OSSIGENO COSTITUISCE LA CAUSA NUMERO UNO DI USTIONI E MORTI CORRELATE AGLI INCENDI. È NECESSARIO ATTENERSI ALLE SEGUENTI AVVERTENZE DI SICUREZZA.
AVVERTENZA: NON CONSENTIRE AL PAZIENTE DI FUMARE O DI USARE CANDELE O FIAMME LIBERE ALL’INTERNO DELLA STANZA IN CUI SI TROVA IL DISPOSITIVO OPPURE QUALSIASI ALTRO ACCESSORIO PER L’EROGAZIONE DELL’OSSIGENO.
AVVERTENZA: NON FUMARE MENTRE SI INDOSSA UNA CANNULA PER L’OSSIGENO PER EVITARE IL RISCHIO DI USTIONI AL VISO E POSSIBILI CONSEGUENZE LETALI.
AVVERTENZA: RIMUOVENDO LA CANNULA E POSIZIONANDOLA SU INDUMENTI, BIANCHERIA, DIVANI O ALTRI MATERIALI IMBOTTITI SI POSSONO VERIFICARE LAMPI DI FUOCO IN CASO DI ESPOSIZIONE A SIGARETTE, FONTI DI CALORE, SCINTILLE O FIAMME.
Page 97
VisionAire Family
AVVERTENZA: SE SI FUMA, È SEMPRE NECESSARIO SEGUIRE PRIMA QUESTI TRE (3) PASSI IMPORTANTI: SPEGNERE IL CONCENTRATORE DI OSSIGENO, RIMUOVERE LA CANNULA E LASCIARE LA STANZA IN CUI SI TROVA IL DISPOSITIVO.
AVVERTENZA: IN CASA O NEL LUOGO IN CUI È UTILIZZATO L’OSSIGENO OCCORRE AFFIGGERE CARTELLI "NON FUMARE – OSSIGENO IN USO" IN MODO CHE SIANO BEN VISIBILI. GLI UTENTI E COLORO CHE PRESTANO ASSISTENZA DEVONO ESSERE INFORMATI DEI PERICOLI DEL FUMO IN PRESENZA O DURANTE L’UTILIZZO DI OSSIGENO MEDICO.
AVVERTENZA: LA LEGGE FEDERALE (USA) LIMITA LA VENDITA O IL NOLEGGIO DI QUESTO DISPOSITIVO DIETRO PRESCRIZIONE DI UN MEDICO O DI UN ALTRO PROFESSIONISTA SANITARIO QUALIFICATO.
AVVERTENZA: QUESTO DISPOSITIVO EROGA OSSIGENO A ELEVATA CONCENTRAZIONE, CHE FAVORISCE IL RAPIDO SVILUPPO DI INCENDI. NON CONSENTIRE AL PAZIENTE DI FUMARE O DI USARE FIAMME LIBERE ALL’INTERNO DELLA STANZA IN CUI SI TROVA (1) QUESTO DISPOSITIVO O (2) QUALSIASI ALTRO ACCESSORIO PER L’EROGAZIONE DELL’OSSIGENO. IL MANCATO RISPETTO DI QUESTA AVVERTENZA PUÒ COMPORTARE GRAVI INCENDI, DANNI MATERIALI E/O PROVOCARE LESIONI FISICHE O IL DECESSO.
AVVERTENZA: NON UTILIZZARE IL CONCENTRATORE DI OSSIGENO IN PRESENZA DI GAS INFIAMMABILI, POICHÉ POSSONO VERIFICARSI FENOMENI DI COMBUSTIONE RAPIDA, CHE POSSONO COMPORTARE DANNI MATERIALI, LESIONI CORPOREE O IL DECESSO.
AVVERTENZA: NON LASCIARE LA CANNULA NASALE SU INDUMENTI, COPERTE O CUSCINI DI SEDIE. SE L’UNITÀ È ACCESA MA NON IN USO, L’OSSIGENO PUÒ RENDERE INFIAMMABILE IL MATERIALE. QUANDO IL CONCENTRATORE DI OSSIGENO NON È IN USO, IMPOSTARE L’INTERRUTTORE DI ALIMENTAZIONE I/0 SULLA POSIZIONE 0 (OFF).
AVVERTENZA: NON UTILIZZARE PRODOTTI A BASE DI OLIO, GRASSO O PETROLIO O ALTRI PRODOTTI INFIAMMABILI CON GLI ACCESSORI PER IL TRASPORTO DELL’OSSIGENO O IL CONCENTRATORE DI OSSIGENO. L’OSSIGENO ACCELERA LA COMBUSTIONE DELLE SOSTANZE INFIAMMABILI. UTILIZZARE ESCLUSIVAMENTE LOZIONI O UNGUENTI A BASE D’ACQUA COMPATIBILI CON L’OSSIGENO.
AVVERTENZA: PER EVITARE IL RISCHIO DI INCENDI O USTIONI, NON LUBRIFICARE RACCORDI, CONNESSIONI, TUBI O ALTRI ACCESSORI DEL CONCENTRATORE DI OSSIGENO.
AVVERTENZA: PERICOLO DI SCOSSA ELETTRICA. SPEGNERE L’UNITÀ E SCOLLEGARE IL CAVO DI ALIMENTAZIONE DALLA PRESA ELETTRICA PRIMA DI PULIRE L’UNITÀ PER PREVENIRE IL PERICOLO DI SCOSSE ELETTRICHE E DI USTIONI ACCIDENTALI. GLI INTERVENTI DI RIPARAZIONE E LA RIMOZIONE DELLE COPERTURE DEVONO ESSERE ESEGUITI ESCLUSIVAMENTE DAL FORNITORE DELL’APPARECCHIATURA O DA UN TECNICO DI ASSISTENZA QUALIFICATO.
AVVERTENZA: PRESTARE ATTENZIONE A EVITARE CHE IL CONCENTRATORE DI OSSIGENO SI BAGNI O CHE FLUIDI ENTRINO NELL’UNITÀ. CIÒ PUÒ PROVOCARE UN MALFUNZIONAMENTO O UN ARRESTO DELL’UNITÀ E UN RISCHIO SUPERIORE DI USTIONI O SCOSSE ELETTRICHE.
AVVERTENZA: NON UTILIZZARE LIQUIDI DIRETTAMENTE SULL’UNITÀ. L’ELENCO DELLE SOSTANZE CHIMICHE NON DESIDERATE COMPRENDE, IN VIA NON LIMITATIVA: ALCOL E PRODOTTI A BASE ALCOLICA, PRODOTTI CONCENTRATI A BASE DI CLORO (CLORURO DI ETILENE) E PRODOTTI A BASE OLEOSA (PINE­SOL®, LESTOIL®). QUESTE SOSTANZE NON VANNO UTILIZZATE PER PULIRE L’ALLOGGIAMENTO DI PLASTICA SUL CONCENTRATORE DI OSSIGENO, POICHÉ POSSONO DANNEGGIARE LE PARTI IN PLASTICA DELL’UNITÀ.
AVVERTENZA: PULIRE IL CONTENITORE, IL PANNELLO DI CONTROLLO E IL CAVO DI ALIMENTAZIONE SOLO CON UN DETERGENTE DELICATO PER USO DOMESTICO, APPLICANDOLO CON UNA SPUGNA O UN PANNO UMIDO (NON BAGNATO), QUINDI ASCIUGARE TUTTE LE SUPERFICI. NON FAR PENETRARE LIQUIDI NEL DISPOSITIVO.
AVVERTENZA: NON UTILIZZARE IL CONCENTRATORE DI OSSIGENO ACCANTO O IMPILATO SU ALTRE APPARECCHIATURE. SE NON È POSSIBILE EVITARE TALE MODALITÀ DI UTILIZZO, OSSERVARE IL DISPOSITIVO PER VERIFICANE IL NORMALE FUNZIONAMENTO.
AVVERTENZA: SISTEMARE SEMPRE I TUBI DI MANDATA DELL’OSSIGENO E I CAVI DI ALIMENTAZIONE IN MODO DA PREVENIRE PERICOLI DI INCIAMPO O LO STRANGOLAMENTO ACCIDENTALE.
AVVERTENZA: NON SONO AMMESSE MODIFICHE A QUESTO DISPOSITIVO.
AVVERTENZA: L’USO DI CAVI E ADATTATORI DIVERSI DA QUELLI SPECIFICATI, A ECCEZIONE DI QUELLI VENDUTI DAL PRODUTTORE DEL DISPOSITIVO ELETTROMEDICALE COME PARTI DI RICAMBIO PER I COMPONENTI INTERNI, PUÒ PROVOCARE UN AUMENTO DELLE EMISSIONI E UNA RIDUZIONE DELL’IMMUNITÀ DEL CONCENTRATORE DI OSSIGENO.
PN MN238-C4 A | Manuale dell’utente 97 - ITL
Page 98
VisionAire Family
AVVERTENZA: LE CONDIZIONI AMBIENTALI POSSONO INFLUIRE SULLE PRESTAZIONI DEL DISPOSITIVO. POSIZIONARLO IN UN AMBIENTE PULITO E DISINFESTATO.
AVVERTENZA: IL DISPOSITIVO DEVE ESSERE UTILIZZATO SOLO DAGLI UTENTI FINALI O DA ASSISTENTI SANITARI E TECNICI FORMATI. IL DISPOSITIVO NON DEVE ESSERE UTILIZZATO DAI BAMBINI.
AVVERTENZA: PER ASSICURARSI DI RICEVERE LA QUANTITÀ TERAPEUTICA DI OSSIGENO EROGATO ADATTA ALLE PROPRIE CONDIZIONI MEDICHE, L’UNITÀ VISIONAIRE DEVE ESSERE UTILIZZATA CON LA COMBINAZIONE SPECIFICA DI PARTI E ACCESSORI IN LINEA CON LE SPECIFICHE DEL PRODUTTORE DEL CONCENTRATORE E UTILIZZATI QUANDO SONO STATE STABILITE LE PROPRIE IMPOSTAZIONI.
AVVERTENZA: L’UTILIZZO DI QUESTO DISPOSITIVO A UN’ALTITUDINE, TEMPERATURA O UMIDITÀ RELATIVA AL DI FUORI DEI VALORI SPECIFICI ELENCATI NEL PRESENTE MANUALE PUÒ INFLUIRE IN MODO NEGATIVO SULLA PORTATA E LA CONCENTRAZIONE DI OSSIGENO E DI CONSEGUENZA SULLA QUALITÀ DELLA TERAPIA.
AVVERTENZA: L’USO DI ALCUNI ACCESSORI PER LA SOMMINISTRAZIONE DI OSSIGENO NON INDICATI PER L’USO CON QUESTO CONCENTRATORE DI OSSIGENO PUÒ PREGIUDICARNE LE PRESTAZIONI. GLI ACCESSORI RACCOMANDATI SONO INDICATI IN RIFERIMENTO NEL PRESENTE MANUALE.
AVVERTENZA: UTILIZZARE SOLO LA TENSIONE ELETTRICA SPECIFICATA SULL’ETICHETTA DELLE SPECIFICHE APPLICATA SUL DISPOSITIVO.
AVVERTENZA: NON UTILIZZARE PROLUNGHE CON QUESTA UNITÀ O EVITARE DI COLLEGARE TROPPE SPINE ALLA STESSA PRESA ELETTRICA. L’USO DI PROLUNGHE PUÒ AVERE EFFETTI AVVERSI SULLE PRESTAZIONI DEL DISPOSITIVO. IL COLLEGAMENTO DI TROPPE SPINE A UN’UNICA PRESA PUÒ PROVOCARE UN SOVRACCARICO DEL QUADRO ELETTRICO, CAUSANDO L’ATTIVAZIONE DELL’INTERRUTTORE/FUSIBILE O UN INCENDIO IN CASO DI MANCATO FUNZIONAMENTO DELL’INTERRUTTORE/FUSIBILE.
ATTENZIONE: la legge federale (USA) limita la vendita o il noleggio di questo dispositivo dietro prescrizione di un medico o di un altro professionista sanitario qualificato.
ATTENZIONE: non posizionare l’unità in modo che il cavo di alimentazione sia difficile da raggiungere.
ATTENZIONE: il concentratore va posizionato in modo da evitare fumo, agenti inquinanti o vapori.
ATTENZIONE: assicurarsi che il concentratore sia utilizzato in posizione verticale.
ATTENZIONE: posizionare sempre il tubo di erogazione dell’ossigeno e i cavi di alimentazione in modo che non vi sia pericolo di inciampare.
ATTENZIONE: posizionare l’unità lontano da tende o drappeggi, valvole di regolazione dell’aria calda o stufette elettriche. Assicurarsi che l’unità si trovi su una superficie piana e che tutti i lati si trovino ad almeno 30 cm (1 piede) dal muro o da altre ostruzioni. Non collocare l’unità in un’area confinata. Scegliere una posizione priva di fumo e polvere lontana dalla luce del sole diretta. Non utilizzare l’unità all’aperto, a meno che non sia collegata a una presa protetta con un interruttore automatico differenziale.
ATTENZIONE: non utilizzare questa unità in spazi confinati o angusti in cui la ventilazione possa essere limitata. In questo modo il dispositivo potrebbe surriscaldarsi pregiudicando le prestazioni.
ATTENZIONE: non far bloccare l’ingresso o gli sfiati di uscita dell’aria. NON far cadere o inserire alcun oggetto nelle aperture presenti sul dispositivo. In questo modo il concentratore di ossigeno potrebbe surriscaldarsi pregiudicando le prestazioni.
98 - ITL PN MN238-C4 A | Manuale dell’utente
Page 99
VisionAire Family
ATTENZIONE: il produttore consiglia di tenere a disposizione un’altra sorgente di ossigeno supplementare in caso di interruzioni di corrente, condizioni di allarme o guasti meccanici. Per conoscere il tipo di sistema di riserva necessario, rivolgersi al proprio medico o al fornitore dell’apparecchiatura.
ATTENZIONE: è molto importante selezionare solo il livello di ossigeno prescritto. Non modificare la selezione del flusso se non in caso di precise istruzioni da parte di un medico autorizzato.
ATTENZIONE: è possibile utilizzare il concentratore di ossigeno durante il sonno solo su raccomandazione di un medico qualificato.
ATTENZIONE: l’utilizzo o la conservazione del concentratore di ossigeno a condizioni che non rientrano nel normale intervallo di temperatura di funzionamento può influire negativamente sulle prestazioni dell’unità. Per i limiti di temperatura di funzionamento e conservazione, fare riferimento alla sezione delle specifiche riportata all’interno del presente manuale.
ATTENZIONE: in caso di allarme o qualora si osservi un funzionamento non corretto del concentratore di ossigeno, consultare la sezione Risoluzione dei problemi del presente manuale. Se non si riesce a risolvere il problema, rivolgersi al fornitore dell’apparecchiatura.
ATTENZIONE: se l’allarme acustico è debole o non suona, consultare immediatamente il fornitore dell’apparecchiatura.
ATTENZIONE: se il tubo del flacone umidificatore non è collegato correttamente al raccordo corrispondente o alla presa dell’ossigeno, può verificarsi una perdita di ossigeno.
ATTENZIONE: di norma, il flussometro incluso nell’unità non necessita di regolazioni. Ruotando la manopola di regolazione del flussometro in senso orario, è possibile ridurre e interrompere il flusso di ossigeno dall’unità. Per comodità, il flussometro è contrassegnato in incrementi di ½ l/min. Per le unità dotate dell’opzione di flussometro da 2 l/min, il flussometro è contrassegnato in incrementi di 1/8 l/min per le impostazioni di flusso fino a 2 l/min.
Nota: la cannula non deve essere piegata e può essere utilizzata per una lunghezza totale di 25 piedi (7,6 m) max.
Accertarsi che la cannula sia completamente inserita e salda. Si dovrebbe udire o avvertire il flusso di ossigeno attraverso la biforcazione della cannula nasale. Se sembra che l’ossigeno non fuoriesca, verificare anzitutto che la sfera del flussometro stia registrando un flusso. Quindi, posizionare la punta della cannula in un bicchiere di acqua, se dalla cannula escono delle bolle, c’è fuoriuscita di ossigeno. In caso contrario, fare riferimento alla sezione Risoluzione dei problemi del presente manuale.
Nota: per l’utilizzo adeguato, attenersi sempre alle istruzioni del produttore della cannula. Sostituire la cannula monouso come consigliato dal produttore della cannula o dal fornitore dell’apparecchiatura. Il fornitore dell’apparecchiatura è in grado di fornire ulteriori pezzi di ricambio.
Nota: il concentratore di ossigeno VisionAire deve funzionare per almeno cinque minuti a 2 l/min prima di poter essere utilizzato.
VisionAire è idoneo all’utilizzo doppio, sempre che il flusso combinato sia di almeno 2 l/min e non superi la capacità massima del concentratore.
Al fornitore dell’apparecchiatura: con il concentratore di ossigeno VisionAire sono raccomandati per l’uso i seguenti accessori per la somministrazione dell’ossigeno:
• flacone umidificatore: N. parte HU003-1
• cannula nasale con tubo da 2,1 m (7 piedi) (6 l/min max): N. parte CU002-1
• kit OxySafe: N. parte 20629671 Nota: il produttore non raccomanda la sterilizzazione
di questo dispositivo. Nota: se l’unità non è stata utilizzata per un periodo
di tempo prolungato, sono necessari alcuni minuti di funzionamento prima che si attivi l’allarme del guasto di alimentazione.
Nota: dalla parte inferiore del concentratore fuoriesce dell’aria calda che potrebbe scolorire in modo permanente superfici sensibili alla temperatura, come il vinile. Il concentratore non deve essere utilizzato su pavimenti sensibili alla termocolorazione. Il produttore non è responsabile dello scolorimento dei pavimenti.
Nota: per evitare l’annullamento della garanzia, attenersi a tutte le istruzioni del produttore.
Nota: non tentare di eseguire interventi di manutenzione, a eccezione delle soluzioni possibili elencate all’interno del manuale.
PN MN238-C4 A | Manuale dell’utente 99 - ITL
Page 100
VisionAire Family
Nota: le apparecchiature di comunicazione a radiofrequenza (RF) portatili e mobili possono avere effetti sui dispositivi elettromedicali.
Nota: non sussiste mai il rischio di esaurire l’ossigeno nella stanza in cui si usa il concentratore di ossigeno.
Nota: al fornitore dell’apparecchiatura: con il concentratore di ossigeno sono raccomandati per l’uso i seguenti accessori per la somministrazione dell’ossigeno:
• Cannula nasale con tubicino da 2,1 m (7 piedi): PN 5408-SEQ.
• Per gli altri accessori raccomandati, consultare l’apposito Catalogo degli accessori (PN ML-LOX0010) disponibile su www.CAIREmedical.com.
È richiesta una fascia tagliafuoco per l’utilizzo con qualsiasi cannula.
• CAIRE produce OxySafe™. Da utilizzare insieme al concentratore di ossigeno. OxySafe™ è un fusibile termico in grado di interrompere il flusso di gas nel caso in cui il tubo della cannula a valle o quello dell’ossigeno dovessero prendere fuoco e bruciare in direzione di OxySafe. È posizionato in linea con la cannula nasale o il tubo dell’ossigeno tra il paziente e l’uscita dell’ossigeno di VisionAire.
Per l’utilizzo corretto di OxySafe, fare sempre riferimento alle istruzioni del produttore (incluse in ciascun kit OxySafe).
CAIRE produce un kit OxySafe™ che comprende cuffia della cannula, OxySafe™ e tubi da 2,1 m (7 piedi): N.P.
20629671.
AVVERTENZA: CET APPAREIL PRODUIT DE L’OXYGÈNE À CONCENTRATION ÉLEVÉE, FAVORISANT UNE COMBUSTION RAPIDE. NE PAS PERMETTRE DE FUMER OU DES FLAMMES NUES DANS LA MÊME CHAMBRE: (1) CET APPAREIL OU (2) TOUT ACCESSOIRE CONTENANT DE L’OXYGÈNE. NE PAS UTILISER DE PRODUITS À BASE D’HUILE, DE GRAISSE OU DE PÉTROLE SUR OU À PROXIMITÉ DE L’UNITÉ. DÉCONNECTER LE CORDON D’ALIMENTATION DE LA PRISE ÉLECTRIQUE AVANT DE NETTOYER OU DE FAIRE L’ENTRETIEN DE L’UNITÉ.
RISQUE DE CHOC ÉLECTRIQUE. NE PAS ENLEVER LES COUVERCLES LORSQUE L’UNITÉ EST BRANCHÉE. SEULS VOTRE FOURNISSEUR D’ÉQUIPEMENT OU UN TECHNICIEN DE SERVICE QUALIFIÉ DEVRAIT ENLEVER LES COUVERCLES OU FAIRE L’ENTRETIEN DE L’UNITÉ.
100 - ITL PN MN238-C4 A | Manuale dell’utente