Quick RefeRence Guide
®
®
Battery Life
SeQual® Eclipse 5
Transportable Oxygen Concentrator with autoSAT
Continuous Flow Pulse Flow
0.5 LPM – 4.4 Hours 2.0 LPM – 2.0 Hours 1.0 (16 mL) – 5.4 Hours 4.0 (64 mL) – 4.0 Hours 7.0 (128 mL) – 2.5 Hours
1.0 LPM – 3.7 Hours 3.0 LPM – 1.3 Hours 2.0 (32 mL) – 5.1 Hours 5.0 (80 mL) – 3.7 Hours 8.0 (160 mL) – 2.0 Hours
3.0 (48 mL) – 4.9 Hours 6.0 (96 mL) – 3.5 Hours 9.0 (192 mL) – 1.7 Hours
NOTE: Pulse flow times are calculated based on 12BPM. Higher breath rates will decrease battery times on pulse mode only.
Recharge Time
1. The typical time to recharge the battery inside the Eclipse
on AC or DC power is between 2–5 hours, dependent
upon the flow setting.
The battery may be re-charged while the Eclipse is either
ON or OFF.
2. The battery will recharge using the optional desktop
charger in approximately 3 hours.
3. It is acceptable to leave the Eclipse plugged into AC
power at all times while it is OFF. Note that while the
unit remains plugged in and OFF, the internal fan may
continue to run while the battery is charging and produce
a faint humming noise.
Battery Gauge
The Eclipse 5 has a battery status gauge on the right side
of the display screen. Each of the 5 horizontal gray bars
represents approximately 20% of the total battery charge.
When a battery is being charged, the indicator bars will blink
in a waterfall-type fashion. If a battery is not installed, or is
improperly installed, the battery status gauge will not appear
on the screen.
On/Off
Press and hold the ON/OFF button (top left Blue Button) for
two seconds to power on the Eclipse 5. Upon power up, the
audible alarm should sound for 3 seconds, and all three lights
on above the display screen should illuminate. It is normal for
the yellow and/or red lights to remain on for up to 5 minutes
during warm-up. Once warm-up is complete, only the green
light should be on. Press and hold the ON/OFF button for two
seconds to power off the Eclipse 5.
Changing Flow Mode
Press the flow mode button (bottom left blue button) to toggle
between pulse and continuous flow. The light adjacent to
the button will illuminate when in pulse mode and will blink
with each breath detection and delivery. If the unit is in pulse
mode but does not detect a breath, it will switch to continuous
mode.
Call Us Toll-Free Today!
1.800.482.2473
WWW.CAIREMEDICAL.COM

AC Power Supply
1. The Eclipse 5 will run at all flow rates and recharge a
battery while connected to AC power.
2. Each AC Power supply has a removable cord. If the AC
power supply is plugged into and receiving power from a
wall outlet, a green LED light on the power supply brick
will illuminate.
3. If the AC Power Supply is properly connected to the
Eclipse 5, the external power light on the far right side of
the control panel will illuminate green.
4. The Eclipse AC Power supply is rated for 100-240VAC
and 50-60Hz. When traveling, the power supply will
automatically adjust to the specific voltage for each
country. If needed, you may use outlet adapters that can
be purchased from CAIRE or from a local electronics store.
DC Power Supply
1. Each Eclipse 5 comes with a DC power supply that will
allow the unit to be powered from an automotive-style
power outlet.
2. The Eclipse can operate at all flow settings while using DC
power.
3. The Eclipse battery will recharge while in use and
connected to DC power at 2.0 LPM or lower, and at all
pulse flow settings. There is not sufficient power to run
the Eclipse and recharge the battery on DC power at 2.5
LPM and 3.0 LPM settings.
4. If the DC power supply is plugged into and receiving
power from a DC outlet, and the power supply is properly
plugged into the side of the Eclipse, a green LED on the
power supply brick will illuminate.
5. If the DC Power Supply is properly connected to the
Eclipse 5, the external power light on the far right side
of the control panel will illuminate green. The Eclipse DC
power supply is rated between 11.5-16 VDC.
6. It is suggested to always have a motor vehicle cranked
and running before inserting the DC power supply into the
accessory outlet.
Maintenance
The Air Intake filter must be removed and cleaned by the
patient at least once weekly. The air intake filter is located
on the upper rear of the Eclipse 5 and can be removed and
cleaned with soap and water. Allow to dry thoroughly before
re-installing the filter.
Nasal Cannula and Tubing
The Eclipse 5 uses a standard nasal cannula. Tubing length
should not exceed 50 ft (15.2 m) on a continuous flow
setting, or 7 ft (2.1 m) on a pulse flow setting or when using a
humidifier bottle on continuous flow.
Humidifier Bottle
A humidifier bottle may be used with the Eclipse on
CONTINUOUS FLOW SETTINGS ONLY, and ONLY WHEN THE
UNIT IS STATIONARY. If the unit is to be transported or moved,
the humidifier bottle MUST be removed.
This Quick Reference Guide is not intended to replace the detailed information available in the User Manual, so please refer to the User Manual for
important critical information regarding use and operation of the product.
CAIRE Inc. reserves the right to discontinue its products, or change the prices, materials, equipment, quality, descriptions, specifications and/or
processes to its products at any time without prior notice and with no further obligation or consequence. All rights not expressly stated herein are
reserved by us, as applicable.