CAIRE LIBERATOR 20, LIBERATOR 30, LIBERATOR 37, LIBERATOR 45, LIBERATOR 60 Service manual

L IBERATOR 20
30 37
45 60
Liberator
L IBERATOR
S E R V I C E
M A N U A L
I Preface
Page 2
LIBERATOR 20 LIBERATOR 30 LIBERATOR 37
LIBERATOR 45 LIBERATOR 60
CAIRE, Inc. 3505 County Road 42 West Burnsville, MN 55306-3803
Customer/Technical Service:
Toll Free Phone (U.S.A.): 1-800-48 CAIRE (1-800-482-2473) Toll Free Fax (U.S.A.): 1-888-WE CAIRE (1-888-932-2473) (To place an order) Phone: 1-952-882-5000 Fax: 1-952-882-5178
This manual covers use and maintenance of the Liberator 20, Liberator 30, Liberator 37, Liberator 45, and Liberator 60 units (see below). It is intended for use by experienced personnel only.
No attempt should be made to fill or maintain this equipment until both this manual and Patient Operating Instruction booklet have been read and fully understood.
The following abbreviations are used throughout this manual:
FCV — Flow Control Valve PRV — Primary Relief Valve LED — Light Emitting Diode QDV — Quick Disconnect Valve LO2 — Liquid Oxygen RA — Return Authorization LPM — Liters Per Minute RP — Repair Procedure NER — Normal Evaporation Rate R/R — Removal and Replacement POI — Patient Operating Instructions SRV — Secondary Relief Valve
Definition of Terms
WARNING Description of a condition that can result in personal injury or death.
CAUTION Description of a condition that can result in equipment or component damage.
NOTE A statement containing information important enough to emphasize or repeat.
(ITEM) Item numbers used throughout this manual are shown on the illustrations beginning on page 35.
Model Side Fill Top Fill Dual Fill (0-6 LPM)
L20 10562007 10564150 10660379 L30 10561995 10564133 10564176 L37 10562023 10564168 10660328 L45 10562015 10564141 10564184 L60 11068983 11069020 11208641
Model Side Fill Top Fill Dual Fill (0-10 LPM)
L20 10879389 10879434 10879477 L30 10879400 10879442 10879493 L37 10879418 10879451 10879514 L45 10879426 10879469 10879522 L60 11075905 11075921 11374892
Model Side Fill Top Fill Dual Fill (0-15 LPM)
L45 11177805 11018951 TBD L60 11075948 11075972 11374876
IITable of Contents
Page 3
I. Preface 2
II. Table of Contents 3
III. Equipment Description 4
IV. Specifications 5
V. Safety 6
VI. Theory of Operation 7-9
VII. Unpacking/Setup Instructions 10
VIII. Operation 11-12
IX. Routine Maintenance 13-14
X. Troubleshooting/Repair 15
Troubleshooting Chart 16-17 Repair Procedures 18-31 Service Tools/Equipment/Supplies 32
XI. Parts Illustration (Top Fill G3.0 Units) 35
Parts Illustration (Side Fill G3.0 Units) 38 Parts Illustration (Dual Fill G3.0 Units) 42 Parts Illustration (Dual Fill G3.1 Units) 45 Parts Listing (Top Fill G3.0 Units) 33-34 Parts Listing (Side Fill G3.0 Units) 36-37 Parts Listing (Dual Fill G3.0 Units) 39-41 Parts Listing (Dual Fill G3.1 Units) 43-44
XII.Ordering Information/Return Policy 46-47
III Equipment Description
Page 4
The CAIRELiberator is the stationary component of the Liberator/Stroller supplementary oxygen system. The Liberator incorporates a stainless steel cryogenic container with the valves, plumbing, and associated hardware required to deliver gaseous oxygen to the patient at near ambient temperature.
The Liberator is comprised of four major assemblies. Grouped according to function, they are:
1. Cryogenic Container – This assembly is a double walled, vacuum insulated dewar for storing liquid oxygen at ap­proximately -300 degrees Fahrenheit. The inner pres­sure vessel is designed to safely hold liquid oxygen and is protected from over pressurization with the primary re­lief valve. The vacuum insulation between the inner and outer vessel keeps the outside heat from evaporating the cold liquid
2. Breathing Circuit – This circuit consists of the manifold assembly, fixed orifice rotary flow control valve (FCV), a breathing coil, and a warming coil. It withdraws liquid oxygen from the cryogenic container, warms it to near ambient temperature, and meters the oxygen gas to the patient. The water that is formed when the cold liquid is converted to gas is collected in the condensate bottle. An economizer regulator is used to reduce any exces­sive head pressure while oxygen is being provided to the patient.
3. Shroud Assembly – The shroud assembly houses and protects the breathing circuit and liquid level meter. Labels listing safety information and patient operating instructions are affixed to the side of the shroud.
4. Meter – This system uses a capacitance probe and an elec­tronic level meter (LED) readout to measure and display the product contents at the touch of the meter switch.
An optional roller base can be provided to help move the Liberator.
* For Top Fill or Dual Models Only
Meter Switch
Liquid Level Meter
Shroud
Portable Release
Button*
Top Fill Fitting*
(QDV)
Flow Control Valve
(FCV)
Vent Valve
Economizer Regulator
Breathing/Warming
Coils
Side Fill Fitting
(QDV)
Cryogenic Container
(Dewar)
Condensate Bottle
Inner Vessel
Vacuum Insulation
IVSpecifications (Nominal Values)
Page 5
LIBERATOR 20 LIBERATOR 30 LIBERATOR 37 LIBERATOR 45 LIBERATOR 60
Capacity
Pounds (L02): 52 75 92 112 144 Liquid Liters: 21.6 31.2 38.3 46.6 60 Gaseous Liters: 17,780 25,650 31,460 38,300 49,200
Selectable Flow Rates
Liters per minute (0-6 LPM): Off, .25, .5, .75, 1, 1.5, 2, 2.5, 3, 4, 5, 6 Liters per minute (0-10 LPM): Off, .5, .75, 1, 1.5, 2, 3, 4, 5, 6, 8, 10 Liters per minute (0-15 LPM) Off, .5, 1, 2, 3, 4, 5, 6, 8, 10, 12, 15
Flow Rate Accuracy ± .1 liter per minute or ±10% of flow setting, whichever is greater
Hours of Operation
@ 2 liters/minute: 148 213 262 319 410
Standard Fill Connections Side Side Side Side Side
Top Top Top Top Top
Operating Pressure
(Economizer Regulator) PSIG: 20 20 20 20 20
Primary Relief Valve Setting
PSIG: 23 23 23 23 23
Secondary Relief Valve Setting
PSIG: 30 30 30 30 30
Normal Evaporation Rate
Pounds per day: 1.5 1.5 1.5 1.5 1.6
Filling Time (Fast Fill Technique)
Minutes: Warm 2.5 3.0 3.5 4.0 5.5
Cold 1.5 2.0 2.5 3.0 4.0
Height
Inches: 24.5 29.5 32.75 37 39
Diameter
Inches: 14 14 14 14 16
Weight, Empty
Empty (Lbs.): 39 45 50 55 66 Full (Lbs.): 91 120 142 167 210
Fill Connector Type Side Mounted Rotary Coupling (Standard)
Top Mounted Push Coupling (Optional)
V Safety
Page 6
Oxygen, as it exists at standard atmospheric pressure and temperature, is a colorless, odorless, and tasteless gas. Oxygen constitutes 21% of the atmosphere, by volume. Aside from its well-documented ability to sustain life, oxygen also supports combustion, even though it is nonflammable. Many substances which will burn in air, burn at a faster rate and at a higher temperature in an oxygen enriched atmosphere. A few materials that do not burn in air will burn as the oxygen concentration increases. Some greases and many liquid sol­vents become extremely hazardous materials when placed in an oxygen-enriched environment. In its liquid form, oxygen is still odorless and tasteless, but is now pale blue in color. At an operating pressure of 20 psig, the temperature of liq­uid oxygen is about -280° Fahrenheit. Skin exposed to such a low temperature can become severely frostbitten.
These hazards require that certain safety precautions be taken when working with or around gaseous and/or liquid oxygen:
1. Never permit combustible substances such as greases, oils, solvents, or other compounds not oxygen compati­ble to contact any component of the unit exposed to higher-than-atmospheric concentrations of gaseous or liquid oxygen. This especially applies to tubing, fittings, and valves.
2. Keep the oxygen equipment away from open flames or electrical appliances such as heaters, stoves, toasters, and other devices with heating elements.
3. Never permit smoking in an area where oxygen equip­ment is repaired, filled, or used.
4. Always wear goggles, a face shield, and insulated gloves when working with or around liquid oxygen.
While CAIRE equipment is designed and built to the most rigid standards, no piece of mechanical equipment can ever be made 100% foolproof. Strict compliance with proper safety practices is necessary when using a Liberator or Stroller. We recommend that our distributors emphasize safety and safe handling practices to their employees and customers. While safety features have been designed into the unit and safe operations are anticipated, it is necessary that all distributor personnel carefully read and fully understand WARNINGS, CAUTIONS, and NOTES throughout the man- ual. Periodic review of this information is recommended.
WARNING: Excess accumulation of oxygen creates an oxy­gen-enriched atmosphere (defined by the Compressed Gas Association as an oxygen concentration above 23%). In an oxygen-enriched atmosphere, flammable items may burn vig­orously and may explode. Certain items considered non-com­bustible in air may burn rapidly in such an environment. Keep all organic materials and other flammable substances away from possible contact with oxygen; particularly oil, grease, kerosene, cloth, wood, paint, tar, coal dust, and dirt which may contain oil or grease. DO NOT permit smoking or open flame in any area where oxygen is stored, handled, or used. Failure to comply with this warning may result in serious personal injury.
WARNING: In the event a unit is dropped, tipped over, or unreasonably abused, immediately, but cautiously, raise the container to its normal vertical position. If substantial con­tainer damage has occurred, remove the liquid oxygen from the vessel in a safe manner (RP 24). Purge the unit with an in­ert gas (nitrogen) and promptly return it to CAIRE for in­spection. The container should be prominently marked “CONTAINER DROPPED, INSPECT FOR DAMAGE.” Failure to comply with these procedures may result in per­sonal injury and can seriously damage the container.
WARNING: Personnel must remove liquid oxygen and de­pressurize the Liberator or Stroller before removing parts or loosening fittings from a unit. Failure to do so may result in personal injury because of the extreme cold of the liquid oxy­gen and the pressure in the vessel. External valves and fittings can become extremely cold during liquid transfer.
WARNING: During transfer of liquid oxygen, components will become extremely cold. Care should be used to avoid any contact with these components, as serious burns may result.
WARNING: Keep filled unit upright at all times. Tip over of filled unit may result in liquid oxygen leakage and/or oxy­gen enriched atmosphere.
WARNING: Only use replacement equipment which is com­patible with liquid oxygen and has been cleaned for oxygen use. Do not use regulators, fittings, hoses, etc. which have been previously used in a non-oxygen service.
CAUTION: The Liberator should be moved by utilizing the roller base or hand truck. The Liberator must be used, stored, and transported in a vertical position. Do not lay, store, or ship on its side.
IVTheory of Operation
Page 7
Filling
The Liberator is filled by connecting a pre-purged transfer line with a fill adapter from a larger liquid oxygen source to the Liberator female (side fill) or male (top fill) quick dis­connect valve. The Liberator vent valve is opened. The pres­sure of the oxygen gas above the liquid in the source con­tainer forces liquid oxygen through the transfer line and into the Liberator inner vessel.
There will be some oxygen vaporized during filling. This gas is discharged through the vent valve. When the Liberator is full, liquid oxygen is expelled. Closing the Liberator vent valve and disconnecting the fill adapter from the Liberator QDV terminates the fill process.
Saturation Pressure
The saturation point of a liquid is a steady-state condition where the liquid has absorbed the maximum amount of heat it can. Such a liquid is defined as being at its saturation pres­sure. For each saturation pressure, there is a corresponding temperature; the higher the saturation pressure, the higher the liquid temperature.
There are two conditions which can seriously affect the overall efficiency and operation of the system:
1. Saturation pressure of the liquid oxygen in the fill source is substantially higher than the Liberator operating pres­sure (oversaturated).
2. Saturation pressure of the liquid oxygen in the fill source is substantially lower than the Liberator operating pres­sure (undersaturated).
For example, when a Liberator is filled from a liquid source saturated at 100 psig, larger transfer losses will occur. This is because the Liberator is designed to operate at 20 psig, and the liquid it is filled with is saturated at a much higher pressure and at its correspondingly higher temperature. It is necessary for this liquid to desaturate to a lower pressure and temperature before the relief valve will close and the Liberator will operate properly.
In order to become saturated at 20 psig, the liquid oxygen must give up enough heat for its temperature to be lowered to that temperature corresponding to a pressure of 20 psig, as shown in the graph. It accomplishes this by a vigorous boil­ing action. All of the gas generated by this boiling is vented through the primary relief valve, and is lost.
If the saturation pressure of the liquid oxygen in the filling vessel is lower than the normal operating pressure of the Liberator, e.g., 10 psig, then the pressure building rod of the Liberator must function to raise the system pressure to the operating pressure of 20 psig. It may require as much as several hours for the undersaturated liquid to become satu­rated at 20 psig. The time required for saturation to 20 psig depends on the initial liquid saturation pressure.
FILLING
SATURATION CURVE
Vent Valve
Liquid
Gas
Fill Connection
(QDV)
Gas
-240
-250
-260
-270
TEMPERATURE (˚F )
2
-280
LIQUID O
-290
-300
0 255075100
(-278)
(-297)
SATURATION PRESSURE (PSIG)
(-251)
(-258)
VI Theory of Operation
Page 8
WARNING: Low oxygen flow rates to the patient may result if the Liberator is filled with under-saturated liq­uid oxygen.
To minimize the effect of undersaturated liquid on the Liberator, a fixed orifice has been installed in the outlet of the vent valve. This orifice regulates the back pressure in the unit during the fill process, resulting in more correct satura­tion pressures in the Liberator.
Operation
With liquid oxygen in the unit and the flow control valve and vent valve closed, the pressure in the inner vessel will remain near the primary relief valve setting of 23 psig.
In the Liberator, as in all vacuum-insulated cryogenic con­tainers, some liquid (oxygen in this case) is always evapo­rating into a gas because no insulating system is perfect. The rate of generation of this gas, with the flow control valve closed, is called the normal evaporation rate (NER). This gas is lost through the primary relief valve.
When the flow control valve is at any setting other than off, and the economizer valve is open (pressure above 20 psig), gaseous oxygen is forced from the head space in the inner vessel, through the economizer valve, to the breathing coil.
This process conserves or “economizes” liquid oxygen by withdrawing the head gas first, instead of allowing it to es­cape through the relief valve.
In the breathing coil, the cold gaseous oxygen is warmed to near-ambient temperature while flowing to the flow control valve, where it is metered and dispensed.
Whenever gas is removed from the space above the liquid oxygen, the inner vessel internal pressure begins to drop slightly. When the pressure drops to 20 psig, the economizer valve closes, forcing liquid oxygen up the withdrawal tube and through the warming coil where it becomes gas.
The gas then flows through the bypass tee to the breathing coil, the flow control valve, and then the patient.
As the pressure in the container increases over 20 psig, the economizer valve opens, and the cycle repeats, maintaining constant oxygen flow, at the set flow rate, to the patient.
OPERATION PRESSURE
ABOVE 20 PSI
OPERATION PRESSURE
BELOW 20 PSI
Gas to Patient
Breathing
Coils
Warm-Up
Economizer
(Closed)
Coils
Breathing
Coils
Economizer
(Open)
Gas
Gas to Patient
VITheory of Operation
Page 9
Liquid Level Measurement
Liberators are equipped with a unique liquid level measure­ment system. This system measures the level of liquid oxygen inside the unit with a capacitance probe and displays that liq­uid level on an LED bar graph meter called "Sur-Cal 3".
The liquid level probe consists of two concentric stainless steel cylinders, which are inside the inner vessel. As the liq­uid oxygen level rises, the capacitance of this assembly goes up. The meter measures this capacitance and displays it on the LED bar graph.
Electrical connection between the Sur-Cal 3meter and the probe is made via a single conductor plug connector. The male plug carries the coaxial cable from the meter. The fe­male receptacle carries the same wire from the probe. A sin­gle ground wire is connected from the meter to a male spade terminal on the corresponding meter ground wire. The Sur­Cal 3is equipped with a short-circuit indicator which turns all LED’s “off” when capacitance above approximately 375pF is reached.
The meter is powered by a nine volt battery. The meter has a low battery (LOW BATT) indicator which signals the need for battery replacement.
LIQUID LEVEL
METER
Probe Outer
Cylinder
Liquid Fill Tube
(Probe Inner Cylinder)
Plug Connector
Female Quick Connect Valve
Meter Package
LOW BATT.
+
+
Meter
Switch
Battery
3/4
1/4
+
+
9 Volt
F
1/2
E
_
_
VII Unpacking and Setup
Page 10
Unpacking
1. Always inspect carton for shipping damage. Report any damage to freight company before signing bill of lad­ing.
2. Check description on carton against your order.
3. Unpack unit, including condensation bottle, POI, and hu­midifier elbow kit.
4. Set aside several sets of packing materials in case a unit must be returned to factory.
Setup
1. Install condensation bottle on bracket located below shroud. Remove cap from bottle. Put condensation hose inside bottle.
2. Install humidifier elbow following instructions in carton.
3. Visually inspect the Liberator for damage from improper handling. Note any container dents, cracks in shroud, missing or loose screws or other hardware, bent quick disconnect valve or humidifier adapter.
4. Check for smooth operation of the flow control selector, making sure that a positive detent is felt at all settings. The flow control knob should be secure and properly aligned.
5. Check the vent valve (Item 41) for smooth operation.
6. If possible, connect a Stroller to the Liberator to check for smooth coupling, and to make sure the Stroller is in proper alignment with the Liberator when mated.
7. Verify operation of the Sur-Cal 3meter by depressing the push button. LED segments will light, displaying the level of liquid oxygen. If the unit is empty, only the bottom segment should light. If it does not, or if other erroneous indications are given, refer to the Troubleshooting/Repair Section.
8. Check all labels for damage and wipe away any dust on unit with a clean, dry, lint-free cloth.
9. If desired, flow control knob (Item 1) can be adjusted so it will not exceed maximum prescribed flow rate.
a. Using hex key wrench, loosen setscrew if used (Item
17) in flow control knob and remove knob (Item 1). b. Remove flow rate decal number disc (Item 2). c. Remove two hex head screws (Item 4) from flow lock
plate (Item 3) and remove plate.
d. Remove locking pin (Item 65) from its storage position
on flow lock plate (Item 3) and place in underside of hole corresponding to maximum allowable flow rate.
e. Replace flow lock plate (Item 3), number disc
(Item 2) and knob. Tighten screws (Item 4) to 4-6 inch­pounds. Verify flow lock is at correct position.
ILLUSTRATION
FOR STEP 9
17
2
2
1/2
1/2
1
1
3/4
1/2
3
4
5
6
1/4
0
1
2
4
65
3
65
VIIIOperation
Page 11
Transport
CAUTION: Always ship, store, or transport a Liberator, empty or full, in an upright position, properly secured to pre­vent damage.
Specifically designed roller bases are available for moving the Liberators short distances on smooth surfaces. Hand trucks are also available for moving the Liberators.
The Liberator 20, 30, 37, 45 and 60 may be moved about or transported in a vehicle while full without damage. However, the equipment should not be dropped or handled roughly, or necktube damage may occur.
Filling
1. Fill Source
a. Fill source must contain a sufficient amount of liquid
oxygen to completely fill the Liberator (approximately 120% of Liberator volume).
b. The liquid oxygen in the fill source must be saturated
at 35-60 psig (50 psig is optimal).
c. The fill source must have correct fitting (5/8” * 45°
male flare) to connect to transfer line.
WARNING: Fill source must be in a well ventilated area to prevent formation of oxygen enriched atmosphere.
2. Fill Procedure
a. Required Equipment:
Fill source as outlined aboveLiquid oxygen transfer lineMale transfer line adapter for side fill LiberatorsFemale transfer line adapter for top fill LiberatorsLiberator vent valve wrenchEye protectionInsulated gloves
b. If refilling partially filled Liberator, verify flow rates are
within tolerance specifications before filling.
c. Verify that liquid level meter is operating properly. The
LED display should indicate approximate level in unit. The low battery LED should not be lit.
NOTE: If flow rates are out of specifications or liquid level meter operates improperly, refer to Troubleshooting/ Repair section.
d. Verify that fill source has enough properly saturated
(35-60 psig) liquid oxygen to fill Liberator.
e. Connect transfer line to fill source. Connect proper
transfer line adapter to transfer line.
f. Fully open liquid valve on fill source.
WARNING: Wear insulated gloves and eye protec­tion whenever working with liquid oxygen.
VIII Operation
Page 12
g. For male fill adapter:
Purge transfer line by placing adapter poppet (remove protective cover) against side of unit and pressing in un­til poppet opens. Hold the poppet open for approxi­mately five seconds, then release.
NOTE: Purge the transfer line any time fill source valve has been closed.
h. Wipe reservoir fill connector with lint free rag if moist. i. Weigh unit as required by local and federal standards. j. Open reservoir vent valve.
k. Connect transfer line (liquid hose) to reservoir to begin
fill.
l. Connect a pressure gauge to oxygen outlet and open
the flow control valve to 2 lpm or greater.
m. While filling throttle the vent valve as required to keep
pressure at 20 psi nominal.
An alternate method to throttle the vent valve is to at­tach a flow meter to the oxygen outlet, set the flow con­trol valve to 2 lpm, and then throttle the vent valve to maintain a flow of 2 lpm. This is equivalent to using a pressure gauge.
n. When liquid spurts from vent outlet, disconnect trans-
fer line.
o. Close reservoir vent valve immediately after removing
transfer line.
p. Disconnect pressure gauge (or flow meter) from oxygen
outlet.
CAUTION: Do not allow excessive venting of liquid oxygen through the vent valve. Prolonged exposure may freeze the valve in the open position.
q. Replace protective cover on adapter QDV and hang
adapter and transfer line using hook provided.
r. Verify that all flow rates are within tolerance specifi-
cations. Verify that liquid level meter indicates full.
NOTE: The liquid level indicating system is accurate only after the vent valve is closed, and the oxygen has been stabilized for five minutes.
IX
Routine Maintenance/Schedule A, Annual
There are two schedules for routine maintenance which the home health care distributor may follow. These schedules allow the distributor maximum flexibility while assuring that equipment is operating properly.
Schedule A – Annual
A. Introduction
Routine maintenance is a series of steps used to assure that equipment is functioning properly.
1. If a unit fails to pass a given test, one of two things may be done.
a. Refer to Troubleshooting/Repair (Section X) of this
manual.
b. Return to CAIRE for repair.
2. Schedule – Maximum of one year between routine main­tenance testing. Unit should be tested when a problem is suspected.
B. Procedure
Follow the steps in order listed. If the unit fails any step, refer to Troubleshooting/Repair (Section X) of this manual.
1. Visual Inspection:
a. Remove any LO2prior to maintenance (RP24). b. Look for damaged or missing parts. c. Verify that meter reads empty (one LED) and that the
low battery LED is not lit.
2. Component Test:
a. Remove shroud (RP2). b. Pressurize to 25 psig (RP12) and check that PRV is
open.
c. Pressurize to 34 psig (RP12) and check that SRV is
open. d. Recalibrate meter (RP5). e. Test pressure retention (RP15). f. Replace shroud (RP2).
3. Flow Test:
a. Fill with approximately 15 lbs. (2 LEDs) of properly
saturated LO
2
.
b.
Set FCV to maximum setting and run for one hour
minimum. c. Check all flow settings to the following chart and
check pressure to be at least 18 psi.
4. Check Efficiency of Unit:
a. Set FCV to zero and allow bottle to warm up (10-15
minutes).
b. Inspect bottle for cold or sweaty condition and for ex-
cessive venting from relief valve (some venting is normal).
c. If either condition is observed, conduct NER test
(RP32).
5. Prepare for Use:
a. Empty contents (RP24). b. Allow unit to sit until warm (2-4 hours). c. Clean outside of unit with household glass cleaner
and lint free cloth (do not get in any valves).
Page 13
Reservoir Flow Test
FCV Setting LPM
OFF -0-
0.25 0.15 to 0.35
0.50 0.40 to 0.60
0.75 0.65 to 0.85
1.00 0.90 to 1.10
1.50 1.35 to 1.65
2.00 1.80 to 2.20
2.50 2.25 to 2.75
3.00 2.70 to 3.30
4.00 3.60 to 4.40
5.00 4.50 to 5.50
6.00 5.40 to 6.60
8.00 7.20 to 8.80
10.00 9.00 to 11.00
12.00 10.80 to 13.20
15.00 13.50 to 16.50
IX
Routine Maintenance/Schedule B, Continuous
Page 14
Schedule B – Continuous
A. Introduction Continuous maintenance is a set of test or inspections done
consistently to assure equipment is functioning properly. It can be done with equipment in service, by drivers or other per­sonnel.
1. If a unit fails to pass a given test, it should be taken out of service and sent to the Repair Center/Department for fur­ther inspection.
2. Schedule – Checks are made when the driver sees patient and when moving equipment between patients.
B. Procedure These inspections/ tests are done by driver as part of the
Standard Fill Procedure every time the reservoir is filled.
1. Visual Inspection: a. Broken shroud.
b. Cold sweaty bottle (vacuum problem). c. QDV not deformed.
2. Check prescription flow rate(s). Erie liter meter (± 0.25 lpm) can be used.
3. Check meter. Push meter button before fill and verify that battery is not low and that meter is within one LED of table below. After filling, verify that meter reads full.
These inspections/tests are done between patients.
1. Visual Inspection: a. Broken shroud/flow control knob. b. Cold sweaty bottle or excessive venting from relief
valve (vacuum problem). Some venting from relief
valve is normal. c. QDV not deformed. d. Inspect under shroud (without removal) for any visible
dirt or contaminants. e. Inspect drain tube for visible dirt. Clean with a 6” cot-
ton swab to remove dirt.
2. Verify that meter battery is not low and meter is within one LED of table. If unit is empty, verify meter reads empty, then fill with approximately 15 pounds of liquid oxygen and verity with chart.
3. Set FCV to maximum flow rate for one hour. Check all flow settings to the following chart and check pressure to be at least 18 psi.
4. If unit fails one of the above test, return to Repair Center/Department or CAIRE. if units passes all tests, clean outside of unit with household glass cleaner and lint free cloth (do not get in any valves).
Contents vs. Meter Reading
Pounds of Oxygen
LED Lib 20 Lib 30 Lib 37 Lib 45 Lib 60
1 0.0-9.8 0.0-14.6 0.0-17.8 0.0-21.2 0.0-21.6 2 9.9-14.7 14.7-21.9 17.9-26.7 21.3-31.8 21.7-33.4 3 14.8-19.6 22.0-29.2 26.8-35.6 31.9-42.4 33.5-53.1 4 19.7-24.5 29.3-36.5 35.7-44.5 42.5-53.0 53.2-71.9 5 24.6-29.4 36.6-43.8 44.6-53.4 53.1-63.6 72.0-86.3 6 29.5-34.3 43.9-51.1 53.5-62.3 63.7-74.2 86.4-100.5 7 34.4-39.2 51.2-58.4 62.4-71.2 74.3-84.8 100.6-121.9 8 39.3-44.1 58.5-65.7 71.3-80.1 84.9-95.4 122.0-130.9 9 Above 44.1 Above 65.8 Above 80.2 Above 95.5 Above 131
Reservoir Flow Test
FCV Setting LPM
OFF -0-
0.25 0.15 to 0.35
0.50 0.40 to 0.60
0.75 0.65 to 0.85
1.00 0.90 to 1.10
1.50 1.35 to 1.65
2.00 1.80 to 2.20
2.50 2.25 to 2.75
3.00 2.70 to 3.30
4.00 3.60 to 4.40
5.00 4.50 to 5.50
6.00 5.40 to 6.60
8.00 7.20 to 8.80
10.00 9.00 to 11.00
12.00 10.80 to 13.20
15.00 13.50 to 16.50
Page 15
XTroubleshooting (Table of Contents)
Introduction ………………………………………………………15 Troubleshooting Charts…………………………………………16-17 Procedures
RP1 General ………………………………………………18 RP2 Shroud Assembly R/R ………………………………19 RP3 Condensation Bottle Bracket R/R ……………………19 RP4 Battery R/R …………………………………………19 RP5A Sur-Cal 3 Meter Calibration ……………………19-21 RP5B Sur-Cal 3.1™ Meter Calibration ……………………21 RP6 Manifold Capacitance Test …………………………22 RP7 Manifold Assembly R/R …………………………22-23 RP8 Resolder Feed-thru Wire ……………………………23 RP9 Manifold Harness Assembly R/R ……………………23 RP10 Clean/Dry Probe and Dewar …………………………23 RP11 Sur-Cal 3TMMeter R/R ………………………………24 RP12 Relief Valve Test ……………………………………24 RP13 Primary Relief Valve (PRV) R/R ……………………24 RP14 Secondary Relief Valve (SRV) R/R …………………25 RP15 Pressure Retention Test ………………………………25 RP16 Plumbing Leak Test …………………………………25 RP17 Warming and Breathing Coil Assembly R/R ………26 RP18 Vent Valve R/R ………………………………………26 RP19 Fill ……………………………………………………26 RP20 QDV Lip Seal R/R (Side Fill Only) ………………26-27 RP21 QDV Poppet Assembly R/R (Side Fill Only) ………27 RP22 QDV Assembly R/R (Top Fill and Dual Only) ………27 RP23 Stroller Pop-Off Assembly (Top Fill and Dual Only)…27 RP24 Empty Unit……………………………………………28 RP25 Warm Unit ………………………………………28-29 RP26 Flow Rate Test ………………………………………29 RP27 Operating Pressure Test………………………………29 RP28 Flow Meter Verification………………………………29 RP29 Economizer Regulator R/R …………………………30 RP30 Flow Control Valve (FCV) R/R ……………………30 RP31 Dewar R/R ………………………………………30-31 RP32 Normal Evaporation Rate (NER) Test ………………31
RP33 Cleaning Unit…………………………………………31 Tools ………………………………………………………………32 Fixtures/Equipment ………………………………………………32 Supplies ……………………………………………………………32
Introduction
1. These procedures are designed to be performed only by qualified personnel with proper equipment.
2. Any failure during routine maintenance checks will refer you to this section. See troubleshooting chart for appropriate procedure.
Loading...
+ 35 hidden pages