EMS 20
Symbols Glossary
ISO 7000; Graphical symbols for use on equipment—Index and synopsis
ISO 7010: Graphical symbols—Safety colors and
safety signs—Registered safety signs
Operating temperature limitation of
these units is 10°C to 40°C. Storage
temperature limitation range is -40°C
to 70°C. Reg. # 0632
Portable humidity range is 15 to 95%.
Base humidity range is 30 to 75%.
Reg. # 2620
Keep away from rain, keep dry. Reg.
# 0626
Name and address of manufacturer.
Reg. # 3082
Caution, consult accompanying documents. Reg. # 0434A
Catalog Number. Reg. # 2493
Serial Number. Reg. # 2498
This way up. Reg. # 0623
Fragile, handle with care. Reg. #
0621
Frostbite may occur on contact with
cold liquid or gaseous oxygen, or
frosted parts. Warning low temperature. To warn of low temperature or
freezing conditions. Reg. # W010
The instruction manual must be read.
Reg. # M002
Keep away from open ame, re,
sparks. Open ignition source and
smoking prohibited. Reg. # P003
Do not smoke near unit or while
operating unit. Reg. # P002
Internal Symbols
Keep unit well ventilated at all times
Keep away from ammable materials,
oil and grease.
Wipe connector with clean dry cloth
before lling.
IEC 60417: Graphical Symbols for use on
equipment
Do not cover unit. These units normally vent oxygen. No. 5641
21 CFR 801.15: Code of Federal Regulations
Title 21
Federal law restricts this device to
sale by or on the order of a physician.
Council Directive 2012/19/EU: waste electrical
and electronic equipment (WEEE)
WEEE
This product may be covered by one or more
patents, US and international. Please visit our
website, Pat.: patents.cairemedical.com for listing
of applicable patents.
Warning. Reg. # W001
ADR: European Agreement concerning the International Carriage of Dangerous Goods by Road
Non-toxic gas.
Hazard Oxidizing substances: re
intensifying risk.
Refrigerated Liquid, USP; Produced
by Air Liquefaction
2 - ENG PN 13402391 D | User Manual

Specifications
• Mode of Operation: Continuous Flow
• Type of Protection Against Electrical Shock: Internally Powered Equipment
•
Equipment not suitable for use in the presence of ammable mixtures
Product Specifications
EMS 20
LOX Capacity 21.0 L 50.7 lb (23 kg)
Gaseous Equivalent Capacity 17,337 L
Weight, Empty 39 lb (17,96 kg)
Weight, Filled 89.7 lb (40,69 kg)
Height 24.5 in. (622 mm)
Diameter 14 in. (356 mm)
Typical use time at 2 LPM 6 days 12 hrs
Operating Pressure 20 psi (137 kPa)
Normal Evaporation Rate 1.6 lb/ day (0,73 kg/day)
Standard Flow Control Range Off, .25, .5, .75, 1, 1.5, 2, 2.5, 3, 4, 5, 6, 8, 10, 12, 15 LPM
Flow Rate Accuracy* +/- 10%
* This accuracy is only at 70F and 14.7 psig and with a calibrated accurate mass ow meter.
EMS 20
PN 13402391 D | User Manual 3 - ENG

EMS 20
Warning Information
Important: Read this manual thoroughly before
operating the EMS 20.
RX Only.
WARNING: THIS DEVICE IS NOT INTENDED FOR LIFE SUSTAINING USE.
WARNING: PATIENT OR OTHERS MAY BE ENTANGLED IN CANNULA OR OTHER TUBING CAUSING
ASPHYXIATION.
WARNING: IF YOU FEEL THE EQUIPMENT IS NOT
OPERATING PROPERLY, CALL YOUR HEALTH CARE
PROVIDER. DO NOT ATTEMPT TO REPAIR OR
ADJUST THE UNIT YOURSELF.
WARNING: DO NOT MODIFY THIS EQUIPMENT
WITHOUT AUTHORIZATION FROM THE MANUFACTURER.
WARNING: IF CONTINUITY OF OXYGEN SUPPLY IS
REQUIRED, ENSURE THAT AN ADEQUATE SUPPLY
OF OXYGEN AND/OR A SECONDARY OXYGEN SUPPLY IS AVAILABLE AT ALL TIMES DURING THERAPY.
WARNING: DO NOT ALLOW SMOKING,
CANDLES, OR OPEN FLAMES WITHIN 10
FEET (3 METERS) OF THE DEVICE, OR
CLOSER THAN 8 INCHES (20 CM) FROM
ANY SOURCE OF IGNITION.
WARNING: KEEP YOUR UNIT IN A WELLVENTILATED AREA.
WARNING: DO NOT STORE LIQUID
OXYGEN EQUIPMENT IN A CLOSET, CAR
TRUNK, OR OTHER CONFINED AREA. DO
NOT PLACE BLANKETS, DRAPERIES, OR
OTHER FABRICS OVER EQUIPMENT.
WARNING: THIS PRODUCT CAN EXPOSE YOU
TO CHEMICALS INCLUDING NICKEL, WHICH IS
KNOWN TO THE STATE OF CALIFORNIA TO CAUSE
CANCER. FOR MORE INFORMATION, GO TO
WWW.P65WARNINGS.CA.GOV.
Caution: Use the EMS 20 only as directed
by your doctor.
Keep ammable materials away from
this equipment. Aerosol sprays, oils and
grease, including facial creams and petroleum jelly, ignite easily and may burn
rapidly in the presence of oxygen.
Smoking while wearing an oxygen cannula can cause facial burns and possibly
result in death.
Removing the cannula and placing it
on clothing, bedding, sofas, or other
cushion material will cause a ash re
when exposed to a cigarette, heat source,
spark or ame.
If you smoke please: (1) turn off the
portable, (2) take off the cannula, and
(3) leave the room where the device is
located.
In the event of an accidental tip-over,
immediately but cautiously return the
unit into an upright position if possible.
If any liquid oxygen is escaping, leave
the area immediately and call your
healthcare provider. Do not attempt
to move the unit or stop the liquid
oxygen from escaping.
Do not touch frosted parts of any unit.
Do not store or operate the portable
coupled to the EMS 20.
Do not allow untrained personnel to
handle or operate this device.
Use of this device is prohibited on com-
mercial passenger and cargo air ights
by the Federal Aviation Administration.
The unit contains liquid oxygen which is
extremely cold, almost 300°F (-184°C).
Exposure to such a low temperature can
cause severe frostbite.
Liquid and gaseous oxygen, though
nonammable, cause other materials to
burn faster than normal. This hazard,
along with the low temperature of liquid
oxygen, warrants certain safety precautions.
4 - ENG PN 13402391 D | User Manual

EMS 20
Introduction
The CAIRE EMS 20 unit is a liquid oxygen storage
and delivery system intended to be connected to an
ambulance’s current oxygen supply system in place
of high pressure cylinders.
This operating instruction (OI) contains the instructions for using the EMS system, as well as information regarding available accessories. For detailed
instructions regarding accessories, please see their
respective documentation.
This device is not meant to supply oxygen directly
to a patient. A regulation device must be in place
between the unit and the patient.
Controls
1. Liquid Level Gauge
2. Liquid Level Switch
3. Output Valve
4. Fill Connection
5. Vent Valve
5
4
CAIRE EMS 20 shown.
1
2
PN 13402391 D | User Manual 5 - ENG
3