Symbols Glossary
ISO 7000; Graphical symbols for use on equip­ment—Index and synopsis
ISO 7010: Graphical symbols—Safety colors and safety signs—Registered safety signs
Operating temperature limitation of these units is 10°C to 40°C. Storage temperature limitation range is -40°C to 70°C. Reg. # 0632
Portable humidity range is 15 to 95%. Base humidity range is 30 to 75%. Reg. # 2620
Keep away from rain, keep dry. Reg. # 0626
Name and address of manufacturer. Reg. # 3082
Caution, consult accompanying docu­ments. Reg. # 0434A
Catalog Number. Reg. # 2493
Serial Number. Reg. # 2498
This way up. Reg. # 0623
Fragile, handle with care. Reg. # 0621
Frostbite may occur on contact with cold liquid or gaseous oxygen, or frosted parts. Warning low tempera­ture. To warn of low temperature or freezing conditions. Reg. # W010
The instruction manual must be read. Reg. # M002
Keep away from open ame, re,
sparks. Open ignition source and smoking prohibited. Reg. # P003
Do not smoke near unit or while operating unit. Reg. # P002
Internal Symbols
Keep unit well ventilated at all times
Keep away from ammable materials,
oil and grease.
Wipe connector with clean dry cloth
before lling.
IEC 60417: Graphical Symbols for use on equipment
Do not cover unit. These units nor­mally vent oxygen. No. 5641
21 CFR 801.15: Code of Federal Regulations Title 21
Federal law restricts this device to sale by or on the order of a physician.
Council Directive 2012/19/EU: waste electrical and electronic equipment (WEEE)
WEEE
This product may be covered by one or more patents, US and international. Please visit our website, Pat.: patents.cairemedical.com for listing of applicable patents.
Warning. Reg. # W001
ADR: European Agreement concerning the Inter­national Carriage of Dangerous Goods by Road
Non-toxic gas.
Hazard Oxidizing substances: re
intensifying risk.
Refrigerated Liquid, USP; Produced by Air Liquefaction
2 - ENG PN 13402391 D | User Manual
Specifications
• Mode of Operation: Continuous Flow
• Type of Protection Against Electrical Shock: Internally Powered Equipment
Equipment not suitable for use in the presence of ammable mixtures
Product Specifications
EMS 20
LOX Capacity 21.0 L 50.7 lb (23 kg)
Gaseous Equivalent Capacity 17,337 L Weight, Empty 39 lb (17,96 kg)
Weight, Filled 89.7 lb (40,69 kg)
Height 24.5 in. (622 mm)
Diameter 14 in. (356 mm) Typical use time at 2 LPM 6 days 12 hrs Operating Pressure 20 psi (137 kPa)
Normal Evaporation Rate 1.6 lb/ day (0,73 kg/day)
Standard Flow Control Range Off, .25, .5, .75, 1, 1.5, 2, 2.5, 3, 4, 5, 6, 8, 10, 12, 15 LPM
Flow Rate Accuracy* +/- 10%
* This accuracy is only at 70F and 14.7 psig and with a calibrated accurate mass ow meter.
EMS 20
PN 13402391 D | User Manual 3 - ENG
EMS 20
Warning Information
Important: Read this manual thoroughly before operating the EMS 20. RX Only.
WARNING: THIS DEVICE IS NOT IN­TENDED FOR LIFE SUSTAINING USE.
WARNING: PATIENT OR OTHERS MAY BE ENTAN­GLED IN CANNULA OR OTHER TUBING CAUSING ASPHYXIATION.
WARNING: IF YOU FEEL THE EQUIPMENT IS NOT OPERATING PROPERLY, CALL YOUR HEALTH CARE PROVIDER. DO NOT ATTEMPT TO REPAIR OR ADJUST THE UNIT YOURSELF.
WARNING: DO NOT MODIFY THIS EQUIPMENT WITHOUT AUTHORIZATION FROM THE MANUFAC­TURER.
WARNING: IF CONTINUITY OF OXYGEN SUPPLY IS REQUIRED, ENSURE THAT AN ADEQUATE SUPPLY OF OXYGEN AND/OR A SECONDARY OXYGEN SUP­PLY IS AVAILABLE AT ALL TIMES DURING THERAPY.
WARNING: DO NOT ALLOW SMOKING, CANDLES, OR OPEN FLAMES WITHIN 10 FEET (3 METERS) OF THE DEVICE, OR CLOSER THAN 8 INCHES (20 CM) FROM ANY SOURCE OF IGNITION.
WARNING: KEEP YOUR UNIT IN A WELL­VENTILATED AREA.
WARNING: DO NOT STORE LIQUID OXYGEN EQUIPMENT IN A CLOSET, CAR TRUNK, OR OTHER CONFINED AREA. DO NOT PLACE BLANKETS, DRAPERIES, OR OTHER FABRICS OVER EQUIPMENT.
WARNING: THIS PRODUCT CAN EXPOSE YOU TO CHEMICALS INCLUDING NICKEL, WHICH IS KNOWN TO THE STATE OF CALIFORNIA TO CAUSE CANCER. FOR MORE INFORMATION, GO TO WWW.P65WARNINGS.CA.GOV.
Caution: Use the EMS 20 only as directed by your doctor.
Keep ammable materials away from
this equipment. Aerosol sprays, oils and grease, including facial creams and pe­troleum jelly, ignite easily and may burn rapidly in the presence of oxygen.
Smoking while wearing an oxygen can­nula can cause facial burns and possibly result in death. Removing the cannula and placing it on clothing, bedding, sofas, or other
cushion material will cause a ash re
when exposed to a cigarette, heat source,
spark or ame.
If you smoke please: (1) turn off the portable, (2) take off the cannula, and (3) leave the room where the device is located.
In the event of an accidental tip-over, immediately but cautiously return the unit into an upright position if possible.
If any liquid oxygen is escaping, leave the area immediately and call your healthcare provider. Do not attempt to move the unit or stop the liquid oxygen from escaping.
Do not touch frosted parts of any unit.
Do not store or operate the portable coupled to the EMS 20.
Do not allow untrained personnel to handle or operate this device.
Use of this device is prohibited on com-
mercial passenger and cargo air ights
by the Federal Aviation Administration.
The unit contains liquid oxygen which is extremely cold, almost 300°F (-184°C). Exposure to such a low temperature can cause severe frostbite.
Liquid and gaseous oxygen, though
nonammable, cause other materials to
burn faster than normal. This hazard, along with the low temperature of liquid oxygen, warrants certain safety precau­tions.
4 - ENG PN 13402391 D | User Manual
EMS 20
Introduction
The CAIRE EMS 20 unit is a liquid oxygen storage and delivery system intended to be connected to an ambulance’s current oxygen supply system in place of high pressure cylinders.
This operating instruction (OI) contains the instruc­tions for using the EMS system, as well as informa­tion regarding available accessories. For detailed instructions regarding accessories, please see their respective documentation.
This device is not meant to supply oxygen directly to a patient. A regulation device must be in place between the unit and the patient.
Controls
1. Liquid Level Gauge
2. Liquid Level Switch
3. Output Valve
4. Fill Connection
5. Vent Valve
5
4
CAIRE EMS 20 shown.
1
2
PN 13402391 D | User Manual 5 - ENG