
C
A
I
R
E,
Breeze
Concentrator
CE
AL
Breeze
Breeze
Breeze
OCSI
Messenger
Oxygen
Oxygen
Oxygen
Concentrator
Concentrator
Concentrator

|
Preface
on
--.
ma
=
mn
SERVICE
Breeze
Breeze
Breeze
CAIRE,
3505
Burnsville,
Customer/Technical
Toll
Toll
Phone:
Fax:
This
manua!
Oxygen
experienced
MANUAL
Oxygen
OCSI
Messenger
Inc.
County
Free
Free
Concentrator,
Road:42
MN
Phone
Fax
(U.S.A.):
|
covers
personnel
Concentrator
Oxygen
West
55306-3803
Service:
(U.S.A.):
use
and
or
the
only.
1-800-48
1-888-WE
1-952-882-5179
1-952-882-5178
maintenance
Breeze
Concentrator
Oxygen
CAIRE — (1-800-482-2473)
CAIRE
of
the
Messenger
-
Concentrator
(1-888-932-2473,
Breeze
Oxygen
Oxygen
Concentrator.
to
place
Concentrator,
It
is
intended
an
the
order)
Breeze
for
OCS!
use
by
No
attempt
Operating
should
Instruction
be
made
booklet
to
maintain
have
been
this
equipment
read
and
until
fully
understood.
both
this
manual
and
Patient
Page
2

III
Preventative
Mai
ntenance
Schedule
——
=_=
Clean
Cabinet
(Located
panel.)
Replace
Check
Prescribed
Check
Adequate
+
+
O2
1-5
5-6
Alarm
Intake
right
side
of
Cabinet
9v
Intake
Concentration
Flowrate
LPM
min.
LPM
min.
Function
Battery
Voltage
Filter
control
Filter
at
Max.
90%
87%
.
to
Verify
Breeze
Weekly
(by
patient)
CAIRE
10848321
Model:
OCSI
Oxygen
Ref. No.:
Concentrator
As
Needed | (approx.) | (approx.)
120
Days | 365
Days
Indicated
Messenger
Reporter
By
When
following
the
maintenance
+
+ +
+
unit
has
Replace
Replace
Replace
If
necessary,
As
the
the
needed,
an
oxygen
items
compressor
final
filter.
the 9 volt
rebuild
vacuum
concentration
should
battery.
intake
or
replace
dust
be
reviewed:
the
and
dirt
level
below
Hepa
filter.
compressor.
from
the
the
interior
of
minimum
cabinet.
specification,
the
Page
5

ll!
Preventative
Maintenance
Schedule
Clean
Cabinet
(Located
panel.)
Replace
Check
Prescribed
Check
Adequate
O2
+
+
Alarm
Intake
right
side
Cabinet
1-5
5-6
9v
Intake
Concentration
Flowrate
LPM
min.
LPM
min.
Function
Battery
Filter
of
control
Filter
90%
87%
to
Voltage
Breeze
at
Max.
Verify
CAIRE
Messenger
Weekly
(by
patient) | Needed
Ref.
10961833
Model:
Oxygen
As
No.:
Concentrator
120
Days | 365
(approx.) | (approx.)
Days
Indicated
Messenger
Reporter
By
When
following
the
maintenance
+
+
+
+
+
*
unit
has
Replace
Replace
Replace
If
necessary,
As
Verify
the
the
the 9 voit
needed,
accuracy
an
oxygen
items
compressor
final
rebuild
vacuum
of
concentration
should
filter.
battery.
Messenger.
intake
or
replace
dust
be
and
leve!
reviewed:
Hepa
the
dirt
filter.
compressor.
from
the
below
interior
the
of
minimum
cabinet.
specification,
the
,
Page
6

IV
Troubleshooting
--
—
—
Audible
Power
Loose/disconnected
capacitor
Faulty
Seized
Defective
_Defective
Faulty
Audible
Power
No
Faulty
Loose
Line
Faulty
*
Audible
Thermal
because
alarm
outage,
wire.
motor-run
or
partially
or
circuit
power
alarm
cord
power
at
unit
or
disconnected
voltage
or
disconnected
alarm
of
not
power
switch
high
activated,
PROBABLE
tripped
or
cold
seized
shorted
switch.
activated,
wall
is
activated,
wiring.
breaker.
PROBABLE
plugged
power
switch.
low.
PROBABLE
has
internal
Concentrator
CAUSE
circuit
breaker.
motor-run
start
capacitor.
compressor.
Concentrator
CAUSE
in.
outlet.
wiring.
OCSI.
Concentrator
CAUSE
shut
down
temperature.
is
or
cold
is
not
is
not
compressor
not
running,
start
runnin
running,
circuit
Reset
panel).
Repair
Replace
Replace
Repair.
Replace
Replace.
,
Plug
Check
Replace.
Repair.
Unit
voltage (115
Replace
push
or
reconnect.
capacitor.
or
circuit
circuit
circuit
breaker
in
power.cord.
house
restart
or
breaker
Replace/repair
Remove
Locate
Possible
breaker
repair
is
repair.
unit
tripped,
button
compressor.
breaker.
has
circuit
automatic
volts + 11.5
has
fan.
any
obstructions
in
well
defective
no
REMEDY
circuit
not
tripped,
REMEDY
breaker
not
REMEDY
or
upon
volts).
tripped,
ventilated
thermal
alarm
breaker
fuse.
restoration
to
switch.
(top
no
alarm
Repair.
red
alarm
intake
area.
lights.
control
lights.
of
proper
.
is
on.
or
exhaust.
No
oxygen
Humidifier
Tubing
blocked
Finai
Control
Low
Leak
following
Restriction
Restriction
Final
Page
between
or
filter
valve
oxygen
or
ATF
filter
7
flow,
unit
PROBABLE
or
patient
kinked.
blocked.
flow,
restriction
in
internal
restricted.
tubing
ATF
is
closed.
flow
PROBABLE
module.
humidifier
to
running
CAUSE
blocked
module
adjusting
CAUSE
in
tubing
or
patient
ATF
module.
no
alarm
or
and
oxygen
valve
or
tubing.
indications.
kinked.
outlet
full
open.
component
REMEDY
Clean
tubing.
Remove
components
necessary.
Replace.
Turn
flow.
.
Repair
Repair
Using
from
greater
Replace.
or
replace
unit
valve
or
replace
or
replace
portable
ATF
module
flow,
cabinet
for
blockage.
counter
as
as
flowmeter
outlet,
replace
humidifier,
and
Repair
clockwise
REMEDY
required.
required.
check
if
unable
module.
repair
check
tubing
or
to
open/increase
-
maximum
to
get 6 Ipm
or
replace
replace
delivery
and
as
or

IV.
Low
Output
Small
low
concentration.
of 4 Ipm
Slightly
filters
compressor.
that
ATF
contamination.
discharge
Leak
flowmeter.
Troubleshooting
concentration,
flow
compressed
pressure
or
higher.
low
or
minor
do
not
cause
module
pressure
in
oxygen
70%
PROBABLE
set
above 6 ipm.
air
alarm
This
will
compressor
valve
Behavior
low
is
damaged
Check
to
circuit
to.85%
CAUSE
leak
that
but
occur
output
or
is
similar
pressure
or
has
for
confirm
between
O,,
does
does
only
at
due
seal
problems
to
alarm.
been
high
problem.
ATF
yellow
oxygen
alarm
not
cause
cause
flows
to
clogged
small
leaks
exposed
compressor
module
low
with
to
and
OCSI
Model).
Adjust
flow
Repair
module.
Repair
Replace ATF
Repair
any
or
leak.
to a lower
leaks
replace
module.
REMEDY
setting.
between.
components.
compressor
and
ATF
Concentration
OCSI
Audible
Relief
Low
out
alarm
Compressor
pressure.
alarm.
A
leak
exists
the
low
Faulty
pressure
Flow
rate
low
pressure
valve
ATF
modute
pressure
Fauity
Output
Compressed
Clogged
relief
concentration,
flow
Low
compressor
Faulty
ATF
Leak
in
above
PROBABLE
of
calibration.
activated,
PROBABLE
is
Low
in
pressure
is
set
in
sensor
audio
PROBABLE
drive
to
vent
valve.
PROBABLE
set
well
air
filters.
module.
oxygen
circuit.
85%
worn
pressure
the
compressed
alarm
to
switch.
excess
activating.
alarm
motor
relief
valve.
below
above 6 LPM.
leak.
output.
O,,
CAUSE
Concentrator
CAUSE
or
sensor
activate.
of 7 LPM
activated,
CAUSE
has
70%
CAUSE
yellow alarm
failed
causing
has
air
lines
resulting
concentrator
failed,
O2,
causing
red
(OCSI
is
running,
low | Repair
activated
causing
in
the
high
alarm
(OCSI
Model).
REMEDY
Recalibrate
red
alarm
Repair
Replace.
Adjust
running,
Check
as
required.
Replace.
model).
Adjust
Repair
Repair
Replace.
Replace.
Repair.
light
or
replace
or
replace
flow
rate
no
alarm
for
faulty
flow
to a lower
or
replace
or
replace.
or
replace.
is
compressor.
to
indications.
component
on.
REMEDY
leaking
prescribed
components.
components.
REMEDY
REMEDY
setting.
setting,
and
repair
or
replace
Page
8

V
Electrical
System
Components
πας
TEMPERATURE
A
temperature
cabinet
event
AIR
A
twin
motor
RED
A
this
are
GREEN
A
When
provided
temperature.
of
cabinet
COMPRESSOR
cylinder
is
provided,
LIGHT
red
LED
light.
provided
LIGHT
neon
AC
units
by a circuit
SWITCH
switch
ventilating
air
compressor
supplying
(only
is
When
green
for
provided
units
by a circuit
(only
are
for
lamp
equipped
(Thermal
is
provided
This
protection
fan
(Reference
compressed
Base
to
indicate
are
board
Base
is
board
Models)
equipped
mounted
Models)
provided
with
the
mounted
Switch)
to
shut
off
is
provided
failure
or a blocked
#20)
with
an
integral
air
(Reference # 50)
an
alarm
with
the
red
(Reference
to
indicate
OCS!
option
green
(Reference
the
air
compressor
mainly
ventilating
split
for
the
oxygen
condition.
OCSI
indicating
indicating
option
#49)
that the
the
green
#32)
to
phase,
The
9-volt
lamp.
unit
lamp
lamp.
in
the
protect
capacitor
concentration
the
is
the
air
intake/exhaust.
DC
battery
red
lamp
properly
functions
event
unit
in
run,
electric
process.
powers
functions
powered.
are
of
the
high
OCS]
The
sensing
Included
functions
light).
70%
70%
the
MESSENGER
The
system
CIRCUIT
OCSI
technique
on
If
the
and
the
red
BOARD
system
this
(red
light),
oxygen
85%
lamp
the
OCSI.
CONTROL
Messenger
provides a means
is
circuit
Control
telephone/computer
(optional)
an
on
board
to
determine
board
low
oxygen
concentration
yellow
will
lamp
light.
BOARD
Board
of
interface.
(Reference
oxygen
the
oxygen
are
the
indicating
concentration
is
85%
is
lit,
and
See
section
」
(optional)
interfaces
remotely
This
circuit
#51)
sensing
concentration
lights
(yellow
or
higher
if
the
oxygen
VIII
Service
(not
shown)
with
the
monitoring
board
system
OCSI
the
carries
that
in
the
for:
low
light},
the
green
concentration
Instructions,
*
board.
Breeze operation
various
utilizes
battery
and
lamp
The
an
product
power
is lit,
for
Messenger
power
ultrasonic
stream.
and
alarm
on
(green
if
between
falls
below
calibration
via
a
and
signals.
~
of
Page
11

IX__Spare
Ref.
No.
4
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
Part
Number
10844493
10848321
10740408
10740444
10786490
14165
10848582
10740424
10844549
10848339
10844670
10861058
10739933
10740446
10801389
10855871
10739992
14709
10740420
10786481
CA110060
10855900
10877375
10850421
10879320
10893666
888-006-103
10844506
13320
15058
Parts
List
Breeze
Breeze
Cabinet
Cabinet
O2
02
Circuit
Power
LED
LED
Control
Flow
Control
Access
Cover
Compressor
Caster
Patient
Silicone
Hour
Battery 9 Volt
Compressor
Relief
HEPA
Black
Silicone
Hose
Heat
Cooling
Fan
OCSÍ
Filter
Filter
Outlet
Fitting
Outiet
Nut
Breaker
Switch
Control
Control
Panel
Meter
Vaive
Panel
Bolts
Bacteria
Tubing
Meter
Valve
Intake
Muffler
Tubing
Clamp
Exchanger
Fan
Mount
Insert
——
Frame
(panel
(panel
(8
amp)
Button
Panel
Panel
{no
plumbing
0-6
LPM
Knob
Door:
Cover
Filter
3/16"
Assembly
30
PSI
Filter
3/8"
Description
mount)
mount)
Label
(base
Label
(OCSI
included)
(includes
ID
Clear
ID
Blue
model)
model)
foam)
a
-
Spares
Per
100
Units
3
3
1 1
1
1
1 1
1 1
1 1
1 1
1 1
1
1 1
6 6
1
1
3
1
1
1 1
1 1
1 1
3
1 1
1
1
1 1
1 4
4
Quantity
Used
AIR
AJR
1
1
1
1
1
1
4
1
5
ft.
1
1
13
in.
4
4
Quantity
Ordered

IX
Spare
Ref.
No.
29
30
31
-
32
33
34
35
36
37
38
39
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
57
Part
Number
14156
10829453
10780881
10785518
10875951
10800546
10747773
10784451
10747829
10747845
10855889
10914754
10822631
10844602
10844531
10871686
10786609
10740432
701-007-401
.
10739925
10740264
10914762
10846616
10879311
10880013
10786570
CA403730
10847424
10915870
10740430
10888496
Parts
ㆍ
List
Mounting
Fan
Capacitor
Capacitor
Thermal
Compressor
Compressor
Compressor
Compressor
Compressor
Compressor
Power
Power
ATF
Terminal
Fan
Cold
ATF
ATF
Check
Control
Power
Green
Red
OCSI
Pressure
Wire
Cable
Heat
Base
Strain Relief
Strain
15
Bracket
Switch
Cord
Cord
Module
Block
Cord
Start
Capacitor
Mounting
Mount
Light
Clip
Valve
Valve
On/Off
Light
Harness
Harness
Board
Switch
Clip
Tie
Exchanger
Model
Control
Relief (0.30"
Bolt
UF
(150F)
Mount
Mount
Mount
Mount
Mount
Mounting
(0.25"
(0.30"
Kit
(only
(0.25"
Spring
Footpad
Shoulder
Shoulder
Grommet
OD)
OD)
Bolt
Switch
(base
unit
Mounting
Panel
OD
OD
Description
Bracket
(base
model
model
SN
4419704060
Foam
w/Lights
Cord)
Cord)
Bolt
Washer
only)
only)
ous
`
and
newer)
men
|
Spares
100
Per
Units
1
1 1
1
1
1
1
1
1
1
1
1 1
1
1 1
1 1
1
1 1
4
1
1 1
1 1
1 1
1 1
1 1
1
1 1
1
2
1
1 1
1
1 1
Quantity
Used
4
1
1
4
4
4
4
4
2
4
1
3
3
1
1
2
4in.
1
παρω
Quantity
Ordered
NÍ
Page
1

IX
Ordering
The
Spare
Information
following
replacement
1.
Compile
ordered.
2.
Fill
out a purchase
a.
Purchase
b.
Name
c.
Name
d.
Quantity,
ordered.
steps
parts
a
list
order
and
and
part
Parts
should
for
an
existing
of
all
order
number,
address
address
number,
List
be
used
unit:
equipment
containing
of
billing
of
shipping
description,
when
and
the
location.
location.
and
ordering
replacement
following
unit
cost
new
equipment
parts
information:
for
each
item
to
or
be
nm
|
3.
Telephone
begin
Toll
Toll
Phone:
Fax:
4.
Mail
to:
CAIRE,
3505
Burnsville,
All
new
collect,
UPS
“prepaid”,
added
invoice
particular
order.
immediate
Free
Phone
Free
Fax
the
completed
Inc.
County
MN
equipment
via
your
to
the
final
date.
All
carrier
or
fax
processing
(U.S.A.):
(U.S.A.):
Road
42
55306-3803
will
specified
and
the
invoice.
shipments
or
method
CAIRE
purchase
West
be
shipped
carrier.
shipping
Payment
will
of
shipment
us
at
one
of
the
of
the
order:
1-800-48
1-888-WE
CAIRE
CAIRE
1-952-882-5179
1-952-882-5178
order
for
U.S.A.
either
“prepaid”,
All
replacement
charges
originate
for
equipment
for
replacement
from
is
desired,
numbers
listed
(1-800-482-2473)
(1-888-932-2473)
confirmation
F.O.B.
parts
will
and
parts
Burnsville,
Minnesota.
specify
below
of
the
order
Burnsville,
be
sent
by
parts
will
be
is
CAIRE
If
when
placing
to
or
a
-
Page?
`

XI
Return
Policy
때때
vat
prata
==
When a Breeze
should
If a problem
encountered,
be
inspected
with
reference
Troubleshooting
lf
these
procedures
the
problem,
the
taken:
1.
Call
CAIRE,
numbers,
Service
with
problem
“Return
assigned
Purchase
and
Representative.
the
unit.
cannot
Authorization
to
Order
referenced,
Technical
Service
time.
2.
Carefully
the
container,
3.
Write
the
Note:
4..
Return
carrier
Otherwise
package
unit
in
its
precisely
the
Return Authorization
top
of
the
the
unit
to:
Concentrator
is
immediately.
the
unit
should
should
Chart
in
Section
do
not
provide a solution
following
using
one
steps
of
the
request a Medical
State
If
it
is
determined
be
solved
by
Number"
the
unit
or
part(s).
Number
please
give
is
this
Representative
the
parts,
original
shipping
unit
or
shipping
as
container.
will
be
parts
by
shipped.
returned.
received,
it
be
be
made
to
the
IV,
page
7-8.
should
be
toll-free
Technical
the
problem
that
the
the
TSR,
a
will
be
Ifa
to
be
number
to
at
or
repack
Number
the
that
on
professional
for
Restocking
If
it
becomes
CAIRE
use
the
1.
Notify
CAIRE
When
personnel,
Policy
after
the
following
the
using
contacting
following
a.
State
equipment
b.
Give
be
returned.
c.
State
2. A Return
issued
CAIRE
When
factory,
must
Note:
3.
All
to:
CAIRE,
3505
Burnsville,
in
for
the
the
appear
Otherwise
equipment
Inc.
County
MN
necessary
shipment
“Restock
Customer
one
to
has
Service
of
the
cancel
Policy”
customer
it
will
be
necessary
information:
the
the
the
Serial
quantity
to
equipment
and
be
returned.
Number
Authorization
the
name
of
the
the
equipment
equipment
Return
on
box
must
Road
is
Authorization
the
packing
will
be
be
returned
42
West
55306-3803
an
been
received,
procedure:
Department
toll-free
numbers.
service
to
description
of
each
purchase
Number
will
distributor
to
be
returned.
shipped
to
slip
and
returned.
“prepaid”
-
U.S.A.
order
with
relay
the
of
unit
date.
be
by
the
Number
boxes.
at
to
CAIRE,
3505
Inc.
County
Burnsville,
All
equipment
shipped
When
CAIRE,
the
of
“prepaid:
the
it
distributor
the
“Repair
giving a detailed
performed.
Page
23
Road
MN
55306-3803
returned
defective
will
be
serviced
as
soon
Cost
listing
42
West
to
CAIRE
item(s)
Sheet”
is
and
as
possible. A copy
will
of
any
U.S.A.
must
be
received
returned
be
at
to
enclosed
maintenance
4.
Finally, a “Credit
the
distributor
received,
CAIRE.
Memo”
when
all
inspected,
will
equipment
and
restocked
be
issued
has
by
to
been