C
A
I
R
E,
Breeze
Concentrator
CE
AL
Breeze
Breeze
Breeze
OCSI
Messenger
Oxygen
Oxygen
Oxygen
Concentrator
Concentrator
Concentrator
|
Preface
on
--.
ma
=
mn
SERVICE
Breeze
Breeze
Breeze
CAIRE,
3505
Burnsville,
Customer/Technical
Toll
Toll
Phone:
Fax:
This
manua!
Oxygen
experienced
MANUAL
Oxygen
OCSI
Messenger
Inc.
County
Free
Free
Concentrator,
Road:42
MN
Phone
Fax
(U.S.A.):
|
covers
personnel
Concentrator
Oxygen
West
55306-3803
Service:
(U.S.A.):
use
and
or
the
only.
1-800-48
1-888-WE
1-952-882-5179
1-952-882-5178
maintenance
Breeze
Concentrator
Oxygen
CAIRE — (1-800-482-2473)
CAIRE
of
the
Messenger
-
Concentrator
(1-888-932-2473,
Breeze
Oxygen
Oxygen
Concentrator.
to
place
Concentrator,
It
is
intended
an
the
order)
Breeze
for
OCS!
use
by
No
attempt
Operating
should
Instruction
be
made
booklet
to
maintain
have
been
this
equipment
read
and
until
fully
understood.
both
this
manual
and
Patient
Page
2
III
Preventative
Mai
ntenance
Schedule
——
=_=
Clean
Cabinet
(Located
panel.)
Replace
Check
Prescribed
Check
Adequate
+
+
O2
1-5
5-6
Alarm
Intake
right
side
of
Cabinet
9v
Intake
Concentration
Flowrate
LPM
min.
LPM
min.
Function
Battery
Voltage
Filter
control
Filter
at
Max.
90%
87%
.
to
Verify
Breeze
Weekly
(by
patient)
CAIRE
10848321
Model:
OCSI
Oxygen
Ref. No.:
Concentrator
As
Needed | (approx.) | (approx.)
120
Days | 365
Days
Indicated
Messenger
Reporter
By
When
following
the
maintenance
+
+ +
+
unit
has
Replace
Replace
Replace
If
necessary,
As
the
the
needed,
an
oxygen
items
compressor
final
filter.
the 9 volt
rebuild
vacuum
concentration
should
battery.
intake
or
replace
dust
be
reviewed:
the
and
dirt
level
below
Hepa
filter.
compressor.
from
the
the
interior
of
minimum
cabinet.
specification,
the
Page
5
ll!
Preventative
Maintenance
Schedule
Clean
Cabinet
(Located
panel.)
Replace
Check
Prescribed
Check
Adequate
O2
+
+
Alarm
Intake
right
side
Cabinet
1-5
5-6
9v
Intake
Concentration
Flowrate
LPM
min.
LPM
min.
Function
Battery
Filter
of
control
Filter
90%
87%
to
Voltage
Breeze
at
Max.
Verify
CAIRE
Messenger
Weekly
(by
patient) | Needed
Ref.
10961833
Model:
Oxygen
As
No.:
Concentrator
120
Days | 365
(approx.) | (approx.)
Days
Indicated
Messenger
Reporter
By
When
following
the
maintenance
+
+
+
+
+
*
unit
has
Replace
Replace
Replace
If
necessary,
As
Verify
the
the
the 9 voit
needed,
accuracy
an
oxygen
items
compressor
final
rebuild
vacuum
of
concentration
should
filter.
battery.
Messenger.
intake
or
replace
dust
be
and
leve!
reviewed:
Hepa
the
dirt
filter.
compressor.
from
the
below
interior
the
of
minimum
cabinet.
specification,
the
,
Page
6
IV
Troubleshooting
--
—
—
Audible
Power
Loose/disconnected
capacitor
Faulty
Seized
Defective
_Defective
Faulty
Audible
Power
No
Faulty
Loose
Line
Faulty
*
Audible
Thermal
because
alarm
outage,
wire.
motor-run
or
partially
or
circuit
power
alarm
cord
power
at
unit
or
disconnected
voltage
or
disconnected
alarm
of
not
power
switch
high
activated,
PROBABLE
tripped
or
cold
seized
shorted
switch.
activated,
wall
is
activated,
wiring.
breaker.
PROBABLE
plugged
power
switch.
low.
PROBABLE
has
internal
Concentrator
CAUSE
circuit
breaker.
motor-run
start
capacitor.
compressor.
Concentrator
CAUSE
in.
outlet.
wiring.
OCSI.
Concentrator
CAUSE
shut
down
temperature.
is
or
cold
is
not
is
not
compressor
not
running,
start
runnin
running,
circuit
Reset
panel).
Repair
Replace
Replace
Repair.
Replace
Replace.
,
Plug
Check
Replace.
Repair.
Unit
voltage (115
Replace
push
or
reconnect.
capacitor.
or
circuit
circuit
circuit
breaker
in
power.cord.
house
restart
or
breaker
Replace/repair
Remove
Locate
Possible
breaker
repair
is
repair.
unit
tripped,
button
compressor.
breaker.
has
circuit
automatic
volts + 11.5
has
fan.
any
obstructions
in
well
defective
no
REMEDY
circuit
not
tripped,
REMEDY
breaker
not
REMEDY
or
upon
volts).
tripped,
ventilated
thermal
alarm
breaker
fuse.
restoration
to
switch.
(top
no
alarm
Repair.
red
alarm
intake
area.
lights.
control
lights.
of
proper
.
is
on.
or
exhaust.
No
oxygen
Humidifier
Tubing
blocked
Finai
Control
Low
Leak
following
Restriction
Restriction
Final
Page
between
or
filter
valve
oxygen
or
ATF
filter
7
flow,
unit
PROBABLE
or
patient
kinked.
blocked.
flow,
restriction
in
internal
restricted.
tubing
ATF
is
closed.
flow
PROBABLE
module.
humidifier
to
running
CAUSE
blocked
module
adjusting
CAUSE
in
tubing
or
patient
ATF
module.
no
alarm
or
and
oxygen
valve
or
tubing.
indications.
kinked.
outlet
full
open.
component
REMEDY
Clean
tubing.
Remove
components
necessary.
Replace.
Turn
flow.
.
Repair
Repair
Using
from
greater
Replace.
or
replace
unit
valve
or
replace
or
replace
portable
ATF
module
flow,
cabinet
for
blockage.
counter
as
as
flowmeter
outlet,
replace
humidifier,
and
Repair
clockwise
REMEDY
required.
required.
check
if
unable
module.
repair
check
tubing
or
to
open/increase
-
maximum
to
get 6 Ipm
or
replace
replace
delivery
and
as
or
IV.
Low
Output
Small
low
concentration.
of 4 Ipm
Slightly
filters
compressor.
that
ATF
contamination.
discharge
Leak
flowmeter.
Troubleshooting
concentration,
flow
compressed
pressure
or
higher.
low
or
minor
do
not
cause
module
pressure
in
oxygen
70%
PROBABLE
set
above 6 ipm.
air
alarm
This
will
compressor
valve
Behavior
low
is
damaged
Check
to
circuit
to.85%
CAUSE
leak
that
but
occur
output
or
is
similar
pressure
or
has
for
confirm
between
O,,
does
does
only
at
due
seal
problems
to
alarm.
been
high
problem.
ATF
yellow
oxygen
alarm
not
cause
cause
flows
to
clogged
small
leaks
exposed
compressor
module
low
with
to
and
OCSI
Model).
Adjust
flow
Repair
module.
Repair
Replace ATF
Repair
any
or
leak.
to a lower
leaks
replace
module.
REMEDY
setting.
between.
components.
compressor
and
ATF
Concentration
OCSI
Audible
Relief
Low
out
alarm
Compressor
pressure.
alarm.
A
leak
exists
the
low
Faulty
pressure
Flow
rate
low
pressure
valve
ATF
modute
pressure
Fauity
Output
Compressed
Clogged
relief
concentration,
flow
Low
compressor
Faulty
ATF
Leak
in
above
PROBABLE
of
calibration.
activated,
PROBABLE
is
Low
in
pressure
is
set
in
sensor
audio
PROBABLE
drive
to
vent
valve.
PROBABLE
set
well
air
filters.
module.
oxygen
circuit.
85%
worn
pressure
the
compressed
alarm
to
switch.
excess
activating.
alarm
motor
relief
valve.
below
above 6 LPM.
leak.
output.
O,,
CAUSE
Concentrator
CAUSE
or
sensor
activate.
of 7 LPM
activated,
CAUSE
has
70%
CAUSE
yellow alarm
failed
causing
has
air
lines
resulting
concentrator
failed,
O2,
causing
red
(OCSI
is
running,
low | Repair
activated
causing
in
the
high
alarm
(OCSI
Model).
REMEDY
Recalibrate
red
alarm
Repair
Replace.
Adjust
running,
Check
as
required.
Replace.
model).
Adjust
Repair
Repair
Replace.
Replace.
Repair.
light
or
replace
or
replace
flow
rate
no
alarm
for
faulty
flow
to a lower
or
replace
or
replace.
or
replace.
is
compressor.
to
indications.
component
on.
REMEDY
leaking
prescribed
components.
components.
REMEDY
REMEDY
setting.
setting,
and
repair
or
replace
Page
8