Caiman Lektrafuse Instructions For Use Manual

Aesculap®
Aesculap Endoscopic Technology Valid for units with software version 2.0 and higher
Instructions for use/Technical description
Lektrafuse HF generator GN200
Lektrafuse HF-Generator GN200
Mode d’emploi/Description technique
Générateur HF Lektrafuse GN200
Instrucciones de manejo/Descripción técnica
Generador AF Lektrafuse GN200
Istruzioni per l’uso/Descrizione tecnica
Generatore HF Lektrafuse GN200
Instruções de utilização/Descrição técnica
Gerador de AF Lektrafuse GN200
Gebruiksaanwijzing/Technische beschrijving
Lektrafuse HF-generator GN200
Bruksanvisning/Teknisk beskrivning
Lektrafuse HF-generator GN200
Инструкция по примению/Техническое описание
ВЧ-генератор Lektrafuse GN200
Návod k použití/Technický popis
VF generátor Lektrafuse GN200
Instrukcja użytkowania/Opis techniczny
Generator HF Lektrafuse GN200
Návod na použivanie/Technický opis
VF generátor Lektrafuse GN200
Kullanım Kılavuzu/Teknik açiklama
Lektrafuse HF jeneratörü GN200
2
3
1
9
4678 5
101112131415
2
Aesculap®
Lektrafuse HF generator GN200
Aesculap
®
Lektrafuse HF generator GN200
Legend
1 Signal lamp HF-ON (surrounding ring) 2 Operating mode selection button (MODE) (middle) 3 Connection socket (foot switch) 4 Signal lamp regrasp 5 Display 6 Signal lamp error 7 Signal ring instrument 8 Connection socket (instrument) 9 Indicator lamp power ON 10 Equipotential connector 11 Fuse holder (with 2 fuses) 12 OFF/ON switch 13 Mains power input socket 14 Type plate 15 Ventilation slots
Symbols on product and packages
Caution Observe important safety information such as warn­ings and precautions in the instructions for use.
Follow the instructions for use
Marking of electric and electronic devices according to directive 2002/96/EC (WEEE), see Disposal
Foot switch
Regrasp
Type CF applied part, defibrillation-protected
Equipotentialization
Fuse
Non-ionizing radiation
Alternating current
Batch designation
Serial number
Order number
Date of manufacture
LOT
3
Contents
1. Safe handling 3
1.1 Safe for use in conformance with IEC regulations 4
2. Product description 5
2.1 Scope of supply 5
2.2 Components required for operation 5
2.3 Intended use 5
2.4 Operating principle 5
2.5 Acoustic warning signal 6
2.6 Monitoring functions 6
2.7 Output power diagrams 7
2.8 Maximum peak output voltage (Up) 7
3. Preparation and setup 7
3.1 First use 7
4. Working with the Lektrafuse HF generator GN200 8
4.1 System set-up 8
4.2 Function checks 10
4.3 Safe operation 10
5. Validated reprocessing procedure 11
5.1 General safety instructions 11
5.2 General information 11
5.3 Preparations at the place of use 11
5.4 Preparation before cleaning 11
5.5 Cleaning/disinfection 12
5.6 Wipe disinfection for electrical devices without sterilization 13
5.7 Inspection, maintenance and checks 13
5.8 Storage and transport 13
6. Maintenance 13
6.1 Safety inspection 13
7. Troubleshooting list 15
7.1 Regrasp error 15
7.2 Warnings 16
7.3 Error reports 16
7.4 Failure messages 17
7.5 Problem solving 17
7.6 User maintenance 18
7.7 Fuse replacement 18
8. Technical Service 18
9. Accessories/Spare parts 18
10. Technical data 18
10.1 Ambient conditions 19
11. Disposal 19
12. Distributor in the US/Contact in Canada for product information and complaints 19
1. Safe handling
CAUTION Federal law restricts this device to sale by or on order of a physician!
Note
These instructions for use only describe the set-up, function and operation of the Lektrafuse generator GN200 and is not suitable for initiating a beginner in high-frequency surgery.
Ŷ General risk factors associated with surgical procedures are not
described in this documentation.
Ŷ It is the operating surgeon's responsibility to ensure that the surgical
procedure is performed correctly.
Ŷ The operating surgeon must have a thorough understanding of both
the hands-on and conceptual aspects of the established operating techniques.
Ź Transport the product only in its original box. Ź Prior to use, check that the product is in good working order. Ź Observe “Notes on Electromagnetic Compatibility (EMC)“, see
TA022130.
Ź The Lektrafuse HF generator GN200 fulfills the requirements of
CISPR 11, Class A.
Ź To prevent damage caused by improper setup or operation, and in order
not to compromise warranty and manufacturer liability: – Use the product only according to these instructions for use. – Follow the safety and maintenance instructions. – Only combine Aesculap products with each other. – Adhere to application instructions according to relevant norms, see
Safe for use in conformance with IEC regulations.
DANGER
Risk of injury to patients due to inappropriate application!
Ź The product and accessories should only be
operated by qualified or trained and experi­enced personnel.
WARNING
Risk of injury and material damage due to inappro­priate use of the product!
Ź Use the product only in accordance with the
intended use.
WARNING
Risk of injury and material damage due to improper handling of the product!
The Lektrafuse HF generator together with the accessories forms a system.
Ź Follow the instructions for use of the Lektrafuse
HF generator accessories.
Ź Follow the instructions for use of all products
used.
4
Aesculap®
Lektrafuse HF generator GN200
Ź Inspect the accessories regularly: Electrode cables and endoscopic
accessories, in particular, must be checked for possible damage to the insulation.
Ź Keep the instructions for use accessible for the user. Ź Always adhere to applicable standards.
1.1 Safe for use in conformance with IEC regulations
The operating environment
Ź Ensure that the device does not come in direct contact with the patient
or in the sterile area respectively.
Ź Ensure that the user does not come into direct contact with the patient
and HF generator at the same time.
Patient safety
DANGER
Risk of death by electric shock!
Ź Do not open the product. Ź Only ever connect the product to power mains
with equipment grounding conductor.
WARNING
Risk of injury from ignition or explosion of flam­mable gases! Sparks may occur when using the HF generator as directed.
Ź Do not use the device in explosion-hazard zones. Ź When operating in the head or thoracic region,
avoid using combustible anesthetics and accel­erating gases (e.g. nitrous oxide or oxygen) or, when using such substances, ensure they are extracted from the region of operation.
Ź If possible, use incombustible cleaning and dis-
infecting agents.
Ź If combustible cleaning and disinfecting agents
or solvents have to be used: Verify that such agents have evaporated prior to commencing HF surgery.
Ź Be sure that no inflammable liquids accumulate
under the patient’s body or in body cavities (e.g. the vagina). Before using the HF generator, wipe up all fluids.
Ź Ensure the absence of any endogenous, combus-
tible gases.
Ź Check that oxygen-soaked materials (e.g.
absorbent cotton or mull) are kept at a safe dis­tance from the HF field, so that they cannot ignite.
CAUTION
Risk of interference with other devices! HF generators create potentially harmful magnetic
fields during normal use.
Ź Be sure that no electronic devices that could be
damaged by an electromagnetic field are set up in the vicinity of the HF generator.
CAUTION
Restriction to view and/or side-effects due to the development of steam/smoke during HF surgery!
Ź If necessary, use smoke suction.
DANGER
Danger to life from inadequate preparation or operational errors in the HF generator!
Ź Be sure that the HF generator is in perfect
working order.
Ź Ensure that neither the foot switch nor the hand
switch has been penetrated by conductive fluids (e.g. blood, amniotic fluid).
Ź Ensure there is no short circuit in the foot or
hand switch cables.
DANGER
Risk of burns suffered by the patient due to inad­vertent activation of the HF generator!
Ź Switch off the HF generator immediately using
the power OFF/ON switch in the event that it is activated accidentally.
Ź Always exercise particular care when operating
the foot switch.
DANGER
Risk of injury to the patient due to an unintended rise of the HF output voltage due to a fault in the HF generator!
Ź Stop using the HF surgical device as soon as it
shows even the slightest anomaly.
WARNING
Risk of injury to patients/users due to defective power cord or missing protective ground connec­tions!
Ź Check the mains power cord/protective ground
connections.
WARNING
Danger of injuries due to muscle contraction, caused by stimulation of the nerves and muscles!
Ź Work with particular care on sensitive struc-
tures.
5
Always be sure to do the following:
Ź Position the patient in such a way that s/he is not in contact with any
metal parts that are grounded or have a significant electric capacity against ground (e.g. operating table, fixtures). If necessary, interpose antistatic drape.
Ź Ensure that the patient will not be in contact with any damp cloths,
drapes or bedding.
Ź Safeguard areas prone to strong perspiration against skin contact with
the trunk of the patient’s body by inserting antistatic tissue between such areas and the trunk.
Ź Siphon off urine with a catheter. Ź For heart surgery, ground the HF generator via the potential equaliza-
tion connection.
Ź For patients with cardiac pacemaker or other active implants, consult
with the relevant medical specialist prior to applying HF surgery, so that irreparable damage to the pacemaker or implant can be avoided.
Ź If possible, remove from the vicinity of HF electrodes any electrodes
from physiological monitoring devices that are not equipped with pro­tective resistors or HF dampers.
Ź Do not use needle electrodes for intraoperative monitoring. Ź Arrange the wires and cables of monitoring devices in such a way that
they do not come in contact with the patient’s skin.
Ź Keep the leads to the HF electrodes as short as possible, and arrange
them in such a way that they do not touch the patient or any other wires or cables.
Ź Should the output power appear insufficient with the usual settings,
check that: – the working electrodes are clean, – all plug connections are properly in place.
Ź Never lay active electrodes on or next to the patient. Ź Put down active electrodes that are not needed at any particular
moment, in such a way that they will not touch the patient.
Ź For operations involving unavoidable, continuous contact between the
electrodes and the patient (e.g. endoscopic procedures), press the power OFF/ON switch 12 to switch off the HF generator immediately after any inadvertent activation of the electrode.
Ź Do not remove hot electrodes from the patient’s body immediately
after performing cutting or coagulation procedures.
Ź During operation of the generator, do not touch the electrodes on the
instrument (on the upper or lower side) as the electrodes can heat up and this may lead to burns or injuries.
2. Product description
2.1 Scope of supply
2.2 Components required for operation
Ŷ Power cord Ŷ Foot switch (optional) Ŷ Caiman Instrument
2.3 Intended use
The Lektrafuse HF generator GN200 is used for vessel sealing and vessel division in open and minimally invasive surgery. The instruments can seal vessels of up to and including 7 mm.
The Lektrafuse HF generator is not suitable for use in tube steriliza­tion/tube coagulation for sterilization.
With respect to the electric shock hazard, the Lektrafuse HF generator meets the classification and safety requirements of a type CF device.
The Lektrafuse HF generator is intended only for indoor operation and storage.
2.4 Operating principle
The Lektrafuse HF generator GN200 works on a micro-processor con­trolled basis and converts the voltage of the mains supply into a high-fre­quency alternating current for bipolar vessel sealing.
The sealing process is carried out via a closed control circuit. The sealing can be started and stopped via a button on the instrument or a foot switch.
The Lektrafuse HF generator GN200 has two operating modes:
Ŷ Standard mode: Pre-set operating mode Ŷ Plus mode: Operating mode with increased energy output
Designation Art. no.
High frequency surgical device GN200
Instructions for use TA022414
Notes on Electromagnetic Compatibility TA022130
6
Aesculap®
Lektrafuse HF generator GN200
2.5 Acoustic warning signal
2.6 Monitoring functions
Self-test
As soon as it is switched on, the unit performs a self-test of the control elements, the acoustic warning signal, the microprocessor and the hard­ware function. During this phase, the message "Performing Self-Test" appears in the display 5.
Fig.1
Continuous test cycle during operation
During operation, safety-relevant functions and signals are monitored through a continuous test cycle. As soon as a critical error is detected, the HF generator terminates the HF activation. In display 5, the error number of the corresponding error is displayed and an acoustic warning signal is emitted, see Troubleshooting list.
Status Acoustic warning signal Notes
System failure 3 repetitions of: 1 long beep, 3 KHz, 1 beep, 2 KHz Following each system failure, if identified (F001,...)
REGRASP error 3 repetitions of: 1 beep, 2 KHz,
1 beep, 3 KHz
Following each regrasp error, if identified
System error 3 beeps, 3 KHz Following each failure report, if identified (E001,...)
System warning 3 beeps, 3 KHz Following each warning, if identified
Start HF discharge 2 beeps, 480 Hz When starting the HF discharge
HF discharge active 1 beep, 480 Hz
Ŷ In Standard mode: Repetition with 1.8 Hz Ŷ In Plus mode: Repetition with 2.8 Hz
Continually during HF discharge
End of HF discharge 3 beeps, 530 Hz When a sealing process has been successfully completed (no
failure reports or errors)
Power up self-test 1 beep, 2 KHz During the self-test, in order to check that the signal tone is
functioning
7
2.7 Output power diagrams
Representation of the output power as a function of the load resistance.
Fig.2 Output power diagram Caiman 12 mm Seal & Cut instrument
family
Fig.3 Output power diagram Caiman 5 mm Seal & Cut instrument
family
2.8 Maximum peak output voltage (Up)
Note
The following data allow the user to judge whether the HF generator GN200 is suitable for a given accessory (insulation rating).
Maximum peak output voltage (Up): 200 V
3. Preparation and setup
Non-compliance with the following instructions will preclude all respon­sibility and liability in this respect on the part of Aesculap.
Ź When setting up and operating the product, adhere to
– national regulations for installation and operation, – national regulations on fire and explosion protection.
Note
For the safety of patients and users it is essential that the mains power cord and, especially, the protective earth connection are intact. In many cases defective or missing protective earth connections are not registered imme­diately.
Ź Connect the device via the potential equalization terminal at the rear
panel of the device to the potential equalization system of the room used for medical purposes.
Note
The potential equalization lead can be ordered from the manufacturer as article no. GK535 (4 m length) or TA008205 (0.8 m length).
Ź In order to ensure sufficient ventilation and cooling of the HF genera-
tor, leave at least 10 cm to 15 cm of space around the HF generator.
3.1 First use
WARNING
Risk of injury to patient or user caused by inade­quate accessories (insulation rating)!
Ź Make certain that the accessory voltage rating
specified in the product documentation is higher than maximum peak output voltage.
WARNING
Risk of injury and/or product malfunction due to incorrect operation of the electromedical system!
Ź Adhere to the instructions for use of any medi-
cal device.
8
Aesculap®
Lektrafuse HF generator GN200
4. Working with the Lektrafuse HF gener-
ator GN200
4.1 System set-up
Connecting the accessories
Combinations of accessories that are not mentioned in the present instructions for use may only be employed if they are specifically intended for the respective application, and if they do not compromise the perfor­mance and safety characteristics of the products.
Only CAIMAN instruments can be used in conjunction with the Lektrafuse HF generator GN200, see Accessories/Spare parts.
Connecting the power supply
The voltage of the mains power supply must correspond to the voltage indicated on the type plate of the device.
The device is fitted with a universal power supply unit, which means that it automatically adapts to mains voltages ranging from 100 V to 240 V without having to switch between voltage ranges.
Ź Ensure that the device is switched off. If necessary, switch off with the
power OFF/ON switch 12.
Ź Plug in the power cord in mains power input socket 13 in the rear
panel of the HF surgical device.
Ź Plug in the mains plug at the building mains power socket.
Switching on the device
Ź Switch on the device with the power OFF/ON switch 12.
Indicator lamp power ON 9 is illuminated. The device performs a self­test, see Continuous test cycle during operation:
If no instrument is connected, the message "Attach Instrument" appears in the display 5 after the self-test. If the self-test fails, an error message is shown in the display 5.
Fig.4
Connecting the foot switch
Connecting the foot switch GN201 is optional. Both the foot switch plug and connection socket (foot switch) 3 have a
dot marking. To connect the plug to the connection socket correctly, these markings must be aligned.
Fig.5
DANGER
Risk of injury due to unapproved configuration using additional components!
Ź For all applied components, ensure that their
classification matches that of the application component (e.g. Type BF or Type CF) of the respective device.
DANGER
Risk of death from electric shock!
Ź Only connect the product to power mains with
equipment grounding conductor.
Ź Set up the device in a way that switching it off
using the power OFF/ON switch is possible and straightforward.
Ź Set up the device in a way that disconnecting
the power cord is straightforwardly possible.
9
Ź Align the foot switch plug and insert it into the foot switch connection
socket 3 until it engages. When the device and the foot switch have been successfully con-
nected, the message "Footswitch attached" appears for 3 seconds in the display 5.
Fig.6
Note
Do not activate the foot switch continually during HF discharge. HF dis­charge is started and stopped by pressing once on the foot switch and then releasing.
Remove the foot switch
Ź Turn the locking ring counter-clockwise and withdraw the connecting
plug at the same time The message "Footswitch removed" appears in the display 5 for 3 sec-
onds.
Fig.7
Connecting the instrument
Only instruments of the CAIMAN Seal & Cut product family can be used in conjunction with the Lektrafuse HF generator GN200.
The instrument connector has an arrow marking and the connection socket 8 has a dot marking. To connect the plug to the connection socket correctly, these markings must be aligned.
Fig.8
Ź Connect the instrument with the connection socket 8.
The green signal ring 7 around the instrument connector lights up. The message "Ready to Seal" appears in the display 5.
Fig.9
10
Aesculap®
Lektrafuse HF generator GN200
Selecting the operating mode
The Lektrafuse HF generator has two operating modes for the sealing pro­cess. The selected operating mode will be retained until the HF generator is switched off. The operating mode can be changed during the procedure.
Ź Changing the operating mode: Press operating mode selection button
(MODE) 2.
Ŷ Standard mode: Pre-set operating mode Ŷ Plus mode: Operating mode with increased energy output
The operating mode is independent from the connected instrument. Changing or reconnecting the instrument does not change the operating mode.
The active operating mode is indicated in the display as follows:
Ŷ Border Ŷ Bold font
Fig.10
Fig.11
Note
In case of a regrasp error, the active operating mode is not displayed. The active operating mode can be displayed by pressing the operating mode selection button (MODE) 2.
Note
The Plus mode can be distinguished from Standard mode by the higher fre­quency of the beep during HF discharge.
Shutting down
Ź Switch off the device with the power OFF/ON switch 12.
All the poles of the device are disconnected from the power supply.
Ź Withdraw mains cable.
The device is completely disconnected from the power supply.
4.2 Function checks
Prior to each use, carry out a functional test of the device:
Ź Check that the accessories have no visible damage. Ź Prepare and set up the unit, see Preparation and setup. Ź Check the functionality of the following elements, one after the other,
in the following sequence: – Switch on the power OFF/ON switch 12; the power on signal
lamp 9 lights up
– Automatic self-test after each time the device is switched on: brief
acoustic signal, all display elements light up briefly – Operating mode selection, see Selecting the operating mode – foot switch, see Connecting the foot switch – Instrument, see Connecting the instrument – Activation with instrument button and foot switch – Switch off the power OFF/ON switch 12
4.3 Safe operation
Ź Start the sealing process (HF discharge) by pressing the button on the
instrument or foot switch once. After completing the sealing process, the HF discharge is automatically
stopped. During sealing process, the message "Sealing in Progress" is shown on
the display 5.
Fig.12
To activate/deactivate device
Ź Switch on/switch off the device with the power OFF/ON switch 12.
Operating mode
The Lektrafuse HF generator GN200 automatically recognizes the con­nected instrument. The internal device settings are loaded. The user can choose between two operating modes.
WARNING
Risk of injury and/or malfunction!
Ź Always carry out a function check prior to using
the product.
WARNING
Risk of injury when using the product beyond the field of view!
Ź Apply the product only under visual control.
11
Activating the HF current
Ź Be sure that the patient is prepared in such a way that the HF genera-
tor can be applied without any risk to the patient.
Ź Make sure that the required operating mode is selected. Ź Ensure that the instrument and foot switch, if used, are correctly con-
nected.
Ź Activate the HF generator via the hand switch or the foot switch. Ź Do not activate the hand switch or foot switch continually during HF
discharge.
Note
The vessel sealing process is started and ended by pressing the foot switch or hand switch on the instrument once. After completing the sealing pro­cess, the device stops the HF discharge automatically. The sealing process can be interrupted by pressing the button once more.
5. Validated reprocessing procedure
5.1 General safety instructions
Note
Adhere to national statutory regulations, national and international stan­dards and directives, and local, clinical hygiene instructions for sterile pro­cessing.
Note
For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possi­ble variants of CJD, observe the relevant national regulations concerning the reprocessing of products.
Note
Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results.
Note
Successful processing of this medical device can only be ensured if the pro­cessing method is first validated. The operator/sterile processing techni­cian is responsible for this.
The recommended chemistry was used for validation.
Note
If there is no final sterilization, then a virucidal disinfectant must be used.
Note
For up-to-date information about reprocessing and material compatibility, see also the Aesculap Extranet at https://extranet.bbraun.com.
The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
5.2 General information
Dried or affixed surgical residues can make cleaning more difficult or inef­fective and lead to corrosion. Therefore the time interval between appli­cation and processing should not exceed 6 h; also, neither fixating pre­cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used.
Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel.
Residues containing chlorine or chlorides e.g. in surgical residues, medi­cines, saline solutions and in the service water used for cleaning, disinfec­tion and sterilization will cause corrosion damage (pitting, stress corro­sion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then dry­ing.
Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or
FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems:
Ŷ Optical changes of materials, e.g. fading or discoloration of titanium or
aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes.
Ŷ Material damage such as corrosion, cracks, fracturing, premature aging
or swelling.
Ź Do not use metal cleaning brushes or other abrasives that would dam-
age the product surfaces and could cause corrosion.
Ź Further detailed advice on hygienically safe and material-/value-pre-
serving reprocessing can be found at www.a-k-i.org, link to Publica­tions, Red Brochure – Proper maintenance of instruments.
5.3 Preparations at the place of use
Ź Remove any visible surgical residues to the extent possible with a
damp, lint-free cloth.
Ź Transport the dry product in a sealed waste container for cleaning and
disinfection within 6 hours.
5.4 Preparation before cleaning
Ź Disconnect the unit from mains power. Ź Remove accessories.
12
Aesculap®
Lektrafuse HF generator GN200
5.5 Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure
Validated cleaning and disinfection procedure
DANGER
Risk of electric shock and fire hazard!
Ź Unplug the device before cleaning. Ź Do not use flammable or explosive cleaning or
disinfecting solutions.
Ź Ensure that no fluids will penetrate the product.
CAUTION
Damage to, or destruction of the product caused by mechanical cleaning/disinfection!
Ź Only clean and disinfect the product manually. Ź Do not sterilize the product under any circum-
stances.
CAUTION
Damage to the product due to inappropriate clean­ing/disinfecting agents!
Ź Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer’s instructions for the respective cleaning/disin­fecting agent.
CAUTION
Damage to the product due to inappropriate clean­ing/disinfecting agents and/or excessive tempera­tures!
Ź Use cleaning agents and disinfectants according
to the manufacturer's instructions.
Ź Observe specifications regarding concentration,
temperature and exposure time.
Validated procedure Special features Reference
Wipe disinfection for electrical devices with­out sterilization
Chapter Wipe disinfec-
tion for electrical devices without sterilization
13
5.6 Wipe disinfection for electrical devices without sterilization
RT: Room temperature
Phase I
Ź Remove any visible residues with a disposable disinfectant wipe. Ź Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Ź Observe the specified application time (1 min minimum).
5.7 Inspection, maintenance and checks
Ź Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
Ź Set aside the product if it is damaged. Ź Perform a safety inspection annually, see Maintenance.
5.8 Storage and transport
Ź Transport the product only in its original box. Ź For storage and transport conditions, see Ambient conditions.
6. Maintenance
No servicing is required for the Lektrafuse HF generator GN200.
Note
Maintenance must only be carried out by authorized personnel. If neces­sary, the circuit diagrams and the service manual, which contains all the necessary documents, are provided for this.
6.1 Safety inspection
A safety inspection must be carried out annually. The person in charge of the inspection will document the inspection find-
ings and measurement values according to the printed inspection proto­col.
Ź Only have the product and its accessories inspected by persons who
possess the requisite training, knowledge and experience and who are authorized to do so.
Ź If there are significant deviations from the inspection protocol values,
or if the prescribed maximum values are exceeded: Send in the device.
For technical service, please contact your national B. Braun/Aesculap agency, see Technical Service.
Phase Step T
[°C/°F]t[min]
Conc. [%]
Water quality Chemical
IWipe disinfectionRT 1 - - Meliseptol HBV wipes 50 % Propan-1-ol
CAUTION
Damage to the product caused by switch-on too early after storage/transport at temperature below +10 °C (50 °F)!
Ź Allow the HF generator to acclimate at room
temperature for about 1 hour.
14
Aesculap®
Lektrafuse HF generator GN200
Fig.13 Inspection protocol form Safety inspection
Inspection protocol – safety inspection
TEST INTERVAL: 1 Year TYPE OF DEVICE: MANUFACTURED BY:
Bipolar Electrosurgical Unit GN200
Aesculap AG, Am Aesculap-Platz 78532 Tuttlingen/Germany
SN: ...............................
INVENTORY NO.:
...............................
RESPONSIBLE
ORGANIZATION: ..........................................................................
Measurement
equipment: ...........................................................................................................................................................................................
tcidreV NOITCEPSNI FO EPOCS
1.) VISUAL INSPECTION
PASS FAIL
1.1 Inspection of power cable for visually detectable damage
1.2 Inspection of foot switch for visually detectable damage elbadaer dna tneserp llits etalp epyT 3.1
1.4 Check fuse links for prescribed values
1.5 Overall condition of device: dirt, damage
2.) ELECTRICAL INSPECTION
Measured value
Verdict
Pass / Fail
2.1
Protective earth resistance incl. mains cable
according IEC62353:2007
R
max
=0,3 Ω (at ≥0,2 A)
2.2
Earth leakage current
according IEC60601-1*
N.C. S.F.C.
I
max
= 0,50 mA
I
max
= 1,00 mA
2.3
Enclosure leakage current/touch current
according IEC60601-1*
N.C. S.F.C.
I
max
= 0,10 mA
I
max
= 0,50 mA
2.4
Patient leakage current
according IEC60601-1*
N.C. S.F.C.
I
max
= 0,01 mA
I
max
= 0,05 mA
2.5
Equipment leakage current - alternative method
according IEC62353:2007
I
max
= 1,00 mA
2.6
Applied part leakage current - alternative method
according IEC62353:2007
I
max
= 0,05 mA
2.7 HF-leakage current bipolar
according IEC60601-2-2:2009
I
max
= 122mA
2.8 Insulation resistance
according IEC62353:2007 (Preparation according Service Manual)
test voltage = 500 V DC
tuptuo FH tsniaga sniaM 1.8.2
R
min
= 7 MΩ
gnisuoh tsniaga sniaM 2.8.2
R
min
= 2 MΩ
gnisuoh tsniaga tuptuo-FH 3.8.2
R
min
= 2 MΩ
2.9 HF-output power measurement with induction-free resistor
Instrument Channel R
Load
Rated value
05 1 5C
Ω
145 Watt
±
20 %
C24/44 1 50
Ω
150 Watt
±
20 %
05 2
Ω
150 Watt
±
20 %
2.10 DC resistance between the two HF output poles
according IEC60601-2-2:2009 section 201.8.4.102
R
min
= 2 MΩ
2.11
Function test according to instructions for use
Performed:
TLUSER TSET
Deficiencies were determined, by which patient, user or other individuals can be endangered
No Yes
Repair …………………………………………………….. Next date of inspection
erutangiS/rotcepsnI etaD/ecalP Responsible Organization
* The listed allowable values comply with the requirements of the IEC60601-1:1988+A1:1991+A2:1995 and IEC60601-1:2005
15
7. Troubleshooting list
7.1 Regrasp error
A regrasp error, i.e. potentially insufficient sealing of the vessels, is dis­played as follows:
Ŷ Unambiguous sound sequence and interruption of the sound sequence
of HF operation, also stops HF discharge
Ŷ The yellow regrasp error light 4 on the front side of the HF generator
lights up
Ŷ Warning message in the display 5 on the front side of the HF generator
Procedure if a regrasp error occurs:
Ź Do not cut the vessel. Ź Open the instrument. Ź Check the sealed vessel.
When correct hemostasis has occurred:
Ź Cut the vessel manually with surgical scissors.
When hemostasis has not occurred in the correct way:
Ź Place the instrument on the vessel again and lock it in place.
Note
The regrasp error must not be confirmed. Activating the hand switch or foot switch will start the next sealing process immediately.
Note
In case of a regrasp error, the active operating mode is not displayed. The active operating mode can be displayed by pressing the operating mode selection button (MODE) 2.
Ź Start additional HF cycle. Ź If the additional HF cycle leads to a normal seal, the vessel can be cut,
and the instrument can be opened and removed.
Ź If a regrasp error occurs again, do not cut the vessel, open the instru-
ment and manually cut the sealed vessel with surgical scissors. When so doing, ensure that the hemostasis is correct.
Regrasp error message Significance Remedy
REGRASP INDICATOR – OPEN Visually check seal – Manually cut
- Clean jaws OR Regrasp thicker tissue
Check the vessel sealing. Contact areas of the jaw part must be covered with dried vessel residue or bodily fluids.
- or ­The gripped bundle of tissue is too thin
Clean contact areas with sterile water and a sponge Grip a thicker tissue bundle
REGRASP INDICATOR – SHORT Visually check seal – Manually cut Check for metal or occlusion in jaw Remove excess fluids
There is metal (e.g. staples) or other foreign material between the electrodes in the jaw part
Open and check jaw part Remove any foreign material Remove excess suction fluids in the operat­ing area
REGRASP INDICATOR – TIME Visually check seal – Reseal as needed Seal cycle interrupted Seal endpoint not reached
HF generator has reached its maximum sealing time and has not ascertained any complete seal
- or ­The user has ended HF discharge during sealing
Check the quality of the seal
16
Aesculap®
Lektrafuse HF generator GN200
7.2 Warnings
A warning is indicated as follows:
Ŷ Individual tone (different from a failure, an error or normal HF opera-
tion)
Ŷ Warning message in the display 5 on the front side of the HF generator
7.3 Error reports
An error is indicated as follows:
Ŷ Unambiguous tone sequence (different from a failure or normal HF
operation)
Ŷ Failure message in the display 5 on the front side of the HF generator
Warning message Significance Remedy
Generator Cooling Please Wait
generator temperature too high Check the airflow around the generator
Generator Too Cold Please Wait
Generator must first warm up Do not switch off the generator whilst it is warming up
Error code Error report Significance Remedy
E001 ERROR E001
Instrument HF activation button error Release Instrument RF-ON Button Remove and reattach instrument
Instrument "HF activation" button is pressed while the instrument is con­nected to the HF generator
Release the “HF activation" button Remove instrument and reconnect it
E002 ERROR E002
Instrument Error Remove and reattach OR Remove and replace instrument
HF generator cannot find instrument
- or ­Instrument is defective and must be replaced
Remove instrument and reconnect it
- or ­Replace instrument
E003 ERROR E003
Footswitch Error Release footswitch Remove and reattach footswitch
Foot switch is pressed while the foot switch is connected to the HF generator
Release foot switch Remove foot switch and reconnect it
17
7.4 Failure messages
A failure is indicated as follows:
Ŷ Clear sound sequence (as necessary, may interrupt the sound of the HF
discharge)
Ŷ The red error light 6 on the front side of the HF generator lights up Ŷ Error message in the display 5 on the front side of the HF generator
Note
If the failure occurs during HF discharge, the vessel sealing may be incom­plete or insufficient.
In order to ensure that sufficient hemostasis has been achieved, follow the procedure for regrasp errors, see Regrasp error.
Note
If the failure continues after turning the device off and on again twice, con­tact your national B. Braun/Aesculap agency, see Technical Service.
7.5 Problem solving
Failure code Cause Remedy
F001 Software error on the HF generator Switch off HF generator, wait 5 s, switch HF generator on
F002 Dosage error of HF generator Switch off HF generator, wait 5 s, switch HF generator on
F003 Hardware error Switch off HF generator, wait 5 s, switch HF generator on
F004 Internal communication error of the HF generator Switch off HF generator, wait 5 s, switch HF generator on
F005 Overheating of HF generator Switch off HF generator and leave to cool down
Check the air stream around the generator
F006 Error in the operating mode selection button (MODE) 2 (activated
when switching on)
Switch off HF generator, wait 5 s, release operating mode selection button (MODE) 2, switch HF generator on
F007 Error with HF energy output Switch off HF generator, wait 5 s, switch HF generator on
F008 Fault: HF outside the calibration Switch off HF generator, wait 5 s, switch HF generator on
F009 Fault: Internal power supply Switch off HF generator, wait 5 s, switch HF generator on
Finding Remedy
HF generator does not switch on Ensure that the power cord has been correctly plugged into the power supply 13 of the HF
generator. Ensure that the power cord is connected to a functioning power source. Ensure that the power OFF/ON switch 12 of the HF generator is in the power ON position. Check the fuse in the fuse box and replace if necessary, see Fuse replacement. If the error continues, replace power cord.
The device is switched on, but no displays are illumi­nated and the self-test is not being carried out
Switch off HF generator, wait 5 s, switch HF generator on again. If the error continues, contact your national B. Braun/Aesculap agency, see Technical Ser­vice.
Self-test fails Switch off HF generator, wait 5 s, switch HF generator on again.
If the error continues, contact your national B. Braun/Aesculap agency, see Technical Ser­vice.
HF generator is active (power ON signal lamp 9 lights up), and accessories are connected, but there is no HF discharge
Verify that the required foot switch (if used ) is properly connected and plugged in. Ensure that the Caiman Seal & Cut Instrument is correctly connected and plugged in (green ring 1 is lit up). In order to start the HF discharge, activate the button on the instrument or the foot switch. If the error continues, replace Caiman instrument.
18
Aesculap®
Lektrafuse HF generator GN200
7.6 User maintenance
Ź In case of error, follow the instructions in the display and note the error
code.
Ź If the error continues after following the instructions in the display,
contact your national B. Braun/Aesculap agency, see Technical Service.
7.7 Fuse replacement
Specified fuses: 2 units Aesculap art. no. TA021404: G fuse link, time lag (T) 8 A, breaking capacity H (1 500 A)
Ź Use a small screwdriver to release the clip on the fuse holder 11 and
tilt up cover.
Ź Use a small screwdriver to release the fuse holder 11 and pull it out. Ź Replace both fuses. Ź Reinsert the fuse holder 11. Ź Close the cover.
Fig.14
Note
If the fuses burn out frequently, the device is faulty and should be repaired, see Technical Service.
8. Technical Service
Ź For service and repairs, please contact your national B. Braun/Aesculap
agency.
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses.
Service addresses
Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95 -1601 Fax: +49 (7461) 14 -939 E-Mail: ats@aesculap.de Or in the US: Aesculap Inc. Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 USA Aesculap Repair Hotline Phone: +1 (800) 214 -3392 Fax: +1 (314) 895 -4420 Other service addresses can be obtained from the address indicated above.
9. Accessories/Spare parts
Note
Other accessories, applied parts and replacement parts are described in Aesculap brochure C-304-81 and C-902-02
10. Technical data
Classification acc. to Directive 93/42/EEC
DANGER
Risk of death by electric shock!
Ź Unplug the device before changing the fuses!
DANGER
Danger to life of patients and users if the product malfunctions and/or protective measures fail or are not used!
Ź Do not perform any servicing or maintenance
work under any circumstances while the product is being used on a patient.
Ź Do not modify the product.
Art. no. Designation
TA021404 Fuse
Caiman 5 instrument family
Caiman 12 instrument family
GN201 Foot switch
Art. no. Designation Class
GN200 High frequency surgical device IIb
19
10.1 Ambient conditions
Note
The atmospheric pressure of 700 hPa corresponds to a maximum operating altitude of 3 000 m.
Note
After transporting or storing at temperature outside the operating temper­ature range, leave the HF generator for one hour before use so that it can reach room temperature.
11. Disposal
Note
The user institution is obliged to process the product before its disposal, see Validated reprocessing procedure.
Ź Detailed information concerning the disposal of the product is avail-
able through your national B. Braun/Aesculap agency, see Technical Service.
12. Distributor in the US/Contact in Can­ada for product information and com­plaints
Aesculap Inc. 3773 Corporate Parkway Center Valley, PA, 18034, USA
Mains voltage ranges (Current intake)
100V~–240 V~ (6.7 A–2.8 A)
Frequency 50–60 Hz
Protection class (acc. to IEC / DIN EN 60601-1)
I
Foot switch circuit ignition-safe acc. to IEC/DIN EN 60601,
approved for use in “medical environment”
Output power 2 x 150 W on 50 Ohm
Device protection Time-Current characteris­tic Breaking capacity Construction
T8 AH/250 V T (Time lag) H (1 500 A) 5 mm x 20 mm
HF operating frequency 460 kHz ± 1 %, Quasi-Sinus
Operating mode Int 10 s/30 s
Weight 8.1 kg
Dimensions (L x W x H) 400 mm x 355 mm x 100 mm
Application part type acc. to IEC/DIN 60601-1
CF
Recovery time of the defi­brillation-protected out­put
0 seconds
CISPR11 Class A
EMC IEC/DIN EN 60601-1-2
Conforming to standard IEC/DIN EN 60601-1
IEC/DIN EN 60601-2-2
Operation Storage and transport
Temperature
Relative humidity
non-condensing non-condensing
Atmospheric pressure
-10-10
50
10
90
Adhere to national regulations when disposing of or recy­cling the product, its components and its packaging! The recycling pass can be downloaded from the Extranet as a PDF document under the respective article number. (The recycling pass includes disassembling instructions for the product, as well as information for proper disposal of components harmful to the environment.) Products carrying this symbol are subject to separate col­lection of electrical and electronic devices. Within the European Union, disposal is taken care of by the manufac­turer as a free-of-charge service.
20
Aesculap®
Lektrafuse HF-Generator GN200
Aesculap
®
Lektrafus e HF-Gen erator G N200
Legende
1 Signallampe HF-ON (umliegender Ring) 2 Taster Betriebsartenwahl (MODE) (Mitte) 3 Anschlussbuchse (Fußpedal) 4 Signallampe Regrasp 5 Anzeige 6 Signallampe Fehler 7 Signalring Instrument 8 Anschlussbuchse (Instrument) 9 Signallampe Netz-EIN 10 Anschluss (Potenzialausgleich) 11 Sicherungshalter (mit 2 Sicherungen) 12 Schalter Netz-AUS/EIN 13 Netzeingangsbuchse 14 Typenschild 15 Lüftungsschlitze
Symbole an Produkt und Verpackung
Vorsicht Wichtige sicherheitsbezogene Angaben wie Warnhin­weise und Vorsichtsmaßnahmen in Gebrauchsanwei­sung beachten.
Gebrauchsanweisung befolgen
Kennzeichnung von Elektro- und Elektronikgeräten entsprechend Richtlinie 2002/96/EG (WEEE), siehe Entsorgung
Fußpedal
Regrasp
Defibrillationsgeschütztes Anwendungsteil des Typs CF
Potenzialausgleich
Sicherung
Nichtionisierende Strahlung
Wechselstrom
Chargenbezeichnung
Seriennummer
Bestellnummer
Herstelldatum
LOT
21
Inhaltsverzeichnis
1. Sichere Handhabung 21
1.1 Sichere Handhabung gemäß IEC-/VDE-Bestimmungen 22
2. Gerätebeschreibung 23
2.1 Lieferumfang 23
2.2 Zum Betrieb erforderliche Komponenten 23
2.3 Verwendungszweck 23
2.4 Funktionsweise 23
2.5 Signaltöne 24
2.6 Überwachungsfunktionen 24
2.7 Diagramme Ausgangsleistung 25
2.8 Maximale Ausgangsspitzenspannung (Up) 25
3. Vorbereiten und Aufstellen 25
3.1 Erstinbetriebnahme 25
4. Arbeiten mit dem Lektrafuse HF-Generator GN200 26
4.1 Bereitstellen 26
4.2 Funktionsprüfung 28
4.3 Bedienung 28
5. Validiertes Aufbereitungsverfahren 29
5.1 Allgemeine Sicherheitshinweise 29
5.2 Allgemeine Hinweise 29
5.3 Vorbereitung am Gebrauchsort 29
5.4 Vorbereitung vor der Reinigung 29
5.5 Reinigung/Desinfektion 30
5.6 Wischdesinfektion bei elektrischen Geräten ohne Sterilisation 31
5.7 Kontrolle, Wartung und Prüfung 31
5.8 Lagerung und Transport 31
6. Instandhaltung 31
6.1 Sicherheitstechnische Kontrolle 31
7. Fehler erkennen und beheben 33
7.1 Regrasp-Fehler 33
7.2 Warnungen 34
7.3 Störungsmeldungen 34
7.4 Fehlermeldungen 35
7.5 Problembehebung 36
7.6 Fehlerbehebung durch den Anwender 36
7.7 Sicherungswechsel 36
8. Technischer Service 37
9. Zubehör/Ersatzteile 37
10. Technische Daten 37
10.1 Umgebungsbedingungen 38
11. Entsorgung 38
1. Sichere Handhabung
Hinweis
Diese Gebrauchsanweisung beschreibt nur Aufbau, Funktion und Bedie­nung des Lektrafuse Generators GN200 und ist nicht dazu geeignet, einen Anfänger in die Hochfrequenz-Chirurgie einzuführen.
Ŷ Allgemeine Risiken eines chirurgischen Eingriffs sind in dieser
Gebrauchsanweisung nicht beschrieben.
Ŷ Der Operateur trägt die Verantwortung für die sachgemäße Durchfüh-
rung des operativen Eingriffs.
Ŷ Der Operateur muss sowohl theoretisch als auch praktisch die aner-
kannten Operationstechniken beherrschen.
Ź Produkt nur im Originalkarton transportieren. Ź Vor der Anwendung des Produkts Funktionsfähigkeit und ordnungsge-
mäßen Zustand prüfen.
Ź „Hinweise zur elektromagnetischen Verträglichkeit (EMV)“ beachten,
siehe TA022130.
Ź Der Lektrafuse HF-Generator GN200 erfüllt die Anforderungen nach
CISPR 11 Klasse A.
Ź Um Schäden durch unsachgemäßen Aufbau oder Betrieb zu vermeiden
und die Gewährleistung und Haftung nicht zu gefährden: – Produkt nur gemäß dieser Gebrauchsanweisung verwenden. – Sicherheitsinformationen und Instandhaltungshinweise einhalten. – Nur Aesculap-Produkte miteinander kombinieren. – Anwendungshinweise gemäß Norm einhalten, siehe Sichere Hand-
habung gemäß IEC-/VDE-Bestimmungen.
Ź Zubehör regelmäßig prüfen: Insbesondere sollten Elektrodenkabel und
endoskopisches Zubehör auf mögliche Beschädigungen an der Isola­tion geprüft werden.
Ź Gebrauchsanweisung für den Anwender zugänglich aufbewahren. Ź Gültige Normen einhalten.
GEFAHR
Verletzungsgefahr für den Patienten durch unsach­gemäße Anwendung!
Ź Produkt und Zubehör nur von Personen betrei-
ben lassen, die die erforderliche Ausbildung, Fachkenntnis oder Erfahrung haben.
WARNUNG
Verletzungsgefahr und Sachschäden bei Benutzung des Produkts entgegen seinem Verwendungszweck!
Ź Produkt nur gemäß Verwendungszweck verwen-
den.
WARNUNG
Verletzungsgefahr und Sachschäden durch falsche Handhabung des Produkts!
Der Lektrafuse HF-Generator bildet mit dem Zube­hör ein System.
Ź Gebrauchsanweisung des Zubehörs des
Lektrafuse HF-Generators einhalten.
Ź Gebrauchsanweisung aller verwendeten Pro-
dukte einhalten.
22
Aesculap®
Lektrafuse HF-Generator GN200
1.1 Sichere Handhabung gemäß IEC-/VDE-Bestim-
mungen
Operationsumfeld
Ź Sicherstellen, dass das Gerät nicht in direkten Kontakt mit dem Patien-
ten bzw. in den Sterilbereich gelangt.
Ź Sicherstellen, dass der Anwender nicht zeitgleich in direkten Kontakt
mit dem Patienten und dem HF-Generator gelangt.
Sicherheit für den Patienten
GEFAHR
Lebensgefahr durch elektrischen Stromschlag!
Ź Produkt nicht öffnen. Ź Produkt nur an ein Versorgungsnetz mit Schutz-
leiter anschließen.
WARNUNG
Verletzungsgefahr durch Zündung oder Explosion von brennbaren Gasen! Beim bestimmungsgemäßen Gebrauch des HF-Generators können Funken ent­stehen.
Ź Nicht im explosionsgefährdeten Bereich einset-
zen.
Ź Bei Operationen im Bereich von Kopf und Tho-
rax den Einsatz von zündfähigen Anästhetika und verbrennungsfördernden Gasen (z. B. Lach­gas, Sauerstoff) vermeiden oder diese Stoffe absaugen.
Ź Wenn möglich, nicht brennbare Stoffe zur Rei-
nigung und Desinfektion verwenden.
Ź Wenn brennbare Reinigungs-, Desinfektions-
und Lösungsmittel verwendet werden: Sicher­stellen, dass diese Stoffe vor dem Einsatz der HF-Chirurgie verdunstet sind.
Ź Sicherstellen, dass sich keine brennbaren Flüs-
sigkeiten unter dem Patienten oder in Körper­höhlen (z. B. Vagina) sammeln. Bevor der HF­Generator eingesetzt wird, alle Flüssigkeiten abwischen.
Ź Sicherstellen, dass keine endogenen Gase vor-
handen sind, die sich entzünden können.
Ź Sicherstellen, dass mit Sauerstoff getränkte
Materialien (z. B. Watte, Mull) so weit vom HF­Umfeld entfernt sind, dass sie sich nicht entzün­den können.
VORSICHT
Gefahr von Beeinträchtigungen anderer Geräte! Beim bestimmungsgemäßen Gebrauch des HF-
Generators entstehen elektromagnetische Störfel­der.
Ź Sicherstellen, dass keine elektronischen Geräte
im Umfeld des HF-Generators aufgestellt sind, die durch elektromagnetische Störfelder beein­trächtigt werden können.
VORSICHT
Sichtbehinderung und/oder Nebenwirkungen durch Dampf-/Rauchentwicklung bei HF-Chirurgie!
Ź Ggf. Rauchabsaugung verwenden.
GEFAHR
Lebensgefahr durch mangelnde Vorbereitung oder Fehler am HF-Generator!
Ź Sicherstellen, dass der HF-Generator fehlerfrei
arbeitet.
Ź Sicherstellen, dass keine leitfähigen Flüssigkei-
ten (z. B. Blut, Fruchtwasser) in Fußpedal oder Handschalter eingedrungen sind.
Ź Sicherstellen, dass im Fußpedal- oder Hand-
schalter-Kabel kein Kurzschluss vorliegt.
GEFAHR
Verbrennungsgefahr für den Patienten durch unbe­absichtigtes Aktivieren des HF-Generators!
Ź HF-Generator nach unbeabsichtigtem Aktivie-
ren sofort am Schalter Netz-AUS/EIN ausschal­ten.
Ź Fußpedal oder Handschalter immer mit beson-
ders hoher Aufmerksamkeit bedienen.
GEFAHR
Verletzungsgefahr für den Patienten durch uner­wünschten Anstieg der HF-Ausgangsleistung bei Versagen des HF-Generators!
Ź Gerät bei kleinsten Unregelmäßigkeiten nicht
mehr verwenden.
WARNUNG
Verletzungsgefahr für den Patienten/Anwender durch defekte Netzzuleitung/nicht vorhandene Schutzleiterverbindung!
Ź Netzzuleitung/Schutzleiterverbindungen prüfen.
WARNUNG
Gefahr von Verletzungen durch Muskelkontraktion verursacht durch Stimulation der Nerven und Mus­keln!
Ź An empfindlichen Strukturen mit besonderer
Aufmerksamkeit arbeiten.
23
Folgendes immer sicherstellen:
Ź Patienten so lagern, dass er keine Metallteile berührt, die geerdet sind
oder beträchtliche Kapazität gegen Erde haben (z. B. Operationstisch, Halterungen). Bei Bedarf antistatische Tücher dazwischen legen.
Ź Sicherstellen, dass der Patient keine feuchten Tücher oder Unterlagen
berührt.
Ź Bereiche mit starker Schweißabsonderung und Haut-an-Haut-Berüh-
rungen am Körperstamm durch Zwischenlegen von antistatischen Tüchern schützen.
Ź Urin über Katheter ableiten. Ź Bei Operationen am Herzen HF-Generator über Potenzialausgleich
erden.
Ź Bei Patienten mit Herzschrittmachern oder anderen aktiven Implanta-
ten vor der Anwendung der HF-Chirurgie den entsprechenden Facharzt konsultieren, um irreparable Schäden oder Fehlfunktionen am Schritt­macher oder dem Implantat zu vermeiden.
Ź Elektroden von physiologischen Überwachungsgeräten ohne Schutz-
widerstände oder HF-Drosseln möglichst weit entfernt von den HF­Elektroden anbringen.
Ź Keine Nadelelektroden für die Überwachung verwenden. Ź Leitungen von Überwachungsgeräten so anbringen, dass sie nicht auf
der Haut liegen.
Ź Leitungen zu den HF-Elektroden möglichst kurz halten und so führen,
dass sie weder den Patienten noch andere Leitungen berühren.
Ź Bei unzureichender Leistung mit den gewohnten Einstellungen sicher-
stellen, dass: – die Arbeitselektroden sauber sind, – die Steckverbindungen korrekt sind.
Ź Aktive Elektrode nie auf oder neben dem Patienten ablegen. Ź Zeitweise nicht benötigte aktive Elektroden so ablegen, dass sie den
Patienten nicht berühren.
Ź Bei Operationen, bei denen sich ein ständiger Kontakt der Elektroden
zum Patienten nicht vermeiden lässt (z. B. bei endoskopischen Opera­tionen), HF-Generator bei unbeabsichtigtem Aktivieren der Elektrode sofort am Schalter Netz-AUS/EIN 12 ausschalten.
Ź Heiße Elektrode nicht unmittelbar nach dem Schneiden oder Koagulie-
ren aus dem Körper entfernen.
Ź Während des Betriebs des Generators die Elektroden am Instrument
(auf der oberen oder unteren Seite) nicht berühren, da die Elektroden sich erwärmen können und dies zu Verbrennungen oder Verletzungen führen kann.
2. Gerätebeschreibung
2.1 Lieferumfang
2.2 Zum Betrieb erforderliche Komponenten
Ŷ Netzkabel Ŷ Fußpedal (optional) Ŷ Caiman Instrument
2.3 Verwendungszweck
Der Lektrafuse HF-Generator GN200 wird zur Gefäßversiegelung und Gefäßtrennung in der offenen und minimal-invasiven Chirurgie einge­setzt. Die Instrumente können Gefäße bis einschließlich 7 mm versiegeln.
Der Lektrafuse HF-Generator ist nicht für den Einsatz bei Tubensterilisa­tion/Tubenkoagulation zur Sterilisation geeignet.
Der Lektrafuse HF-Generator ist zur Anwendung am Herzen zugelassen (Typ CF).
Der Lektrafuse HF-Generator ist für den Betrieb und die Lagerung in geschlossenen Räumen bestimmt.
2.4 Funktionsweise
Der Lektrafuse HF-Generator GN200 arbeitet mikroprozessorgesteuert und wandelt die Netzspannung in einen hochfrequenten Wechselstrom für die bipolare Gefäßversiegelung um.
Der Versiegelungsprozess erfolgt über einen geschlossenen Regelkreis. Die Versiegelung kann über die Taste am Instrument oder ein Fußpedal gestar­tet und gestoppt werden.
Der Lektrafuse HF-Generator GN200 besitzt zwei Betriebsarten:
Ŷ Standard-Modus: Voreingestellte Betriebsart Ŷ Plus-Modus: Betriebsart mit erhöhter Energieabgabe
Bezeichnung Art.-Nr.
Hochfrequenz-Chirurgiegerät GN200
Gebrauchsanweisung TA022414
Hinweise zur elektromagnetischen Verträglichkeit
TA022130
24
Aesculap®
Lektrafuse HF-Generator GN200
2.5 Signaltöne
2.6 Überwachungsfunktionen
Selbsttest
Beim Einschalten führt das Gerät einen Selbsttest durch, der Bedienele­mente, Signalton, Mikroprozessor und Hardwarefunktion prüft. Während dieser Phase erscheint in Anzeige 5 die Meldung “Performing Self-Test”.
Abb.1
Permanenter Test während des Betriebs
Während des Betriebs werden sicherheitsrelevante Funktionen und Sig­nale zyklisch getestet. Wenn kritische Fehler erkannt werden, beendet der HF-Generator die HF-Aktivierung. In Anzeige 5 wird die Fehlernummer des entsprechenden Fehlers angezeigt und ein Signalton ausgegeben, siehe Fehler erkennen und beheben.
Zustand/Status Signalton Bemerkung
System Fehlermeldung 3 Wiederholungen von: 1 langer Piepton, 3 KHz,
1 Piepton, 2 KHz
Nach jedem System-Fehler, wenn erkannt (F001,...)
REGRASP-Fehler 3 Wiederholungen von: 1 Piepton, 2 KHz,
1 Piepton, 3 KHz
Nach jedem Regrasp-Fehler, wenn erkannt
System Störungsmeldung 3 Pieptöne, 3 KHz Nach jeder Störungsmeldung, wenn erkannt (E001,..)
System Warnung 3 Pieptöne, 3 KHz Nach jeder Warnung, wenn erkannt
HF-Abgabe Start 2 Pieptöne 480 Hz Beim Start der HF-Abgabe
HF-Abgabe Aktiv Piepton, 480 Hz
Ŷ Im Standard-Modus: Wiederholung mit 1,8 Hz Ŷ Im Plus-Modus: Wiederholung mit 2,8 Hz
Kontinuierlich während der HF-Abgabe
HF-Abgabe Ende 3 Pieptöne, 530 Hz Bei erfolgreicher Beendigung eines Versiegelungsprozesses
(keine Störungsmeldung oder Fehler)
Power Up Selbsttest 1 Piepton, 2 KHz Beim Selbsttest, um die Funktionalität des Signaltons zu über-
prüfen
25
2.7 Diagramme Ausgangsleistung
Darstellung der Ausgangsleistung als Funktion des Lastwiderstands.
Abb.2 Diagramm Ausgangsleistung Caiman 12 mm Seal & Cut Instru-
mentenfamilie
Abb.3 Diagramm Ausgangsleistung Caiman 5 mm Seal & Cut Instru-
mentenfamilie
2.8 Maximale Ausgangsspitzenspannung (Up)
Hinweis
Die folgenden Angaben ermöglichen dem Anwender zu beurteilen, ob der HF-Generator GN200 für bestimmtes Zubehör (Qualität der Isolierung) geeignet ist.
Maximale Ausgangsspitzenspannung (U
p
): 200 V
3. Vorbereiten und Aufstellen
Wenn die folgenden Vorschriften nicht beachtet werden, übernimmt Aesculap insoweit keinerlei Verantwortung.
Ź Beim Aufstellen und Betrieb des Produkts einhalten:
– die nationalen Installations- und Betreiber-Vorschriften, – die nationalen Vorschriften über Brand- und Explosionsschutz.
Hinweis
Die Sicherheit des Anwenders und des Patienten hängt u. a. von einer intak­ten Netzzuleitung, insbesondere von einer intakten Schutzleiterverbindung ab. Defekte oder nicht vorhandene Schutzleiterverbindungen werden häu­fig nicht sofort erkannt.
Ź Gerät über den an der Geräterückwand montierten Anschluss für
Potenzialausgleichsleiter mit dem Potenzialausgleich des medizinisch genutzten Raums verbinden.
Hinweis
Die Potenzialausgleichsleitung ist unter der Artikel-Nr. GK535 (4 m lang) bzw. TA008205 (0,8 m lang) beim Hersteller erhältlich.
Ź Um eine ausreichende Belüftung und Kühlung des HF-Generators
sicherzustellen, rund um der HF-Generator mindestens 10 cm bis 15 cm Raum lassen.
3.1 Erstinbetriebnahme
WARNUNG
Verletzungsgefahr für den Patienten/Anwender durch Verwendung von unzureichendem Zubehör (Qualität der Isolierung)!
Ź Sicherstellen, dass die in den Begleitpapieren
des Zubehörs angegebene Bemessungszubehör­spannung größer ist als die maximale Ausgangs­spitzenspannung.
WARNUNG
Verletzungsgefahr und/oder Fehlfunktion des Pro­dukts durch Fehlbedienung des medizinisch-elekt­rischen Systems!
Ź Gebrauchsanweisungen aller medizinischen
Geräte einhalten.
26
Aesculap®
Lektrafuse HF-Generator GN200
4. Arbeiten mit dem Lektrafuse HF-Gene-
rator GN200
4.1 Bereitstellen
Zubehör anschließen
Zubehörkombinationen, die nicht in der Gebrauchsanweisung erwähnt sind, dürfen nur verwendet werden, wenn sie ausdrücklich für die vorge­sehene Anwendung bestimmt sind. Leistungsmerkmale sowie Sicherheits­anforderungen dürfen nicht nachteilig beeinflusst werden.
Nur CAIMAN Instrumente können in Verbindung mit dem Lektrafuse HF­Generator GN200 eingesetzt werden, siehe Zubehör/Ersatzteile.
Spannungsversorgung anschließen
Die Netzspannung muss mit der Spannung übereinstimmen, die auf dem Typschild des Geräts angegeben ist.
Das Gerät ist mit einem Weitbereichsnetzteil ausgerüstet, so dass die Netzspannung zwischen 100 V und 240 V liegen kann, ohne dass der Netzbereich umgeschaltet werden muss.
Ź Sicherstellen, dass das Gerät ausgeschaltet ist. Gegebenenfalls mit
Schalter Netz-AUS/EIN 12 ausschalten.
Ź Netzkabel an der Rückseite des Geräts in Netzeingangsbuchse 13 ein-
stecken.
Ź Netzstecker in die Steckdose der Hausinstallation stecken.
Gerät einschalten
Ź Gerät mit Schalter Netz-AUS/EIN 12 einschalten.
Die Signallampe Netz-EIN 9 leuchtet. Das Gerät führt einen Selbsttest durch, siehe Permanenter Test während des Betriebs:
Ist kein Instrument angeschlossen, erscheint nach dem Selbsttest in Anzeige 5 die Meldung “Attach Instrument”. Scheitert der Selbsttest, wird eine Fehlermeldung auf Anzeige 5 ausgegeben.
Abb.4
Fußpedal anschließen
Der Anschluss des Fußpedals GN201 ist optional. Am Fußpedalstecker und an der Anschlussbuchse Fußpedal 3 ist jeweils
ein Punkt als Markierung angezeichnet. Um den Stecker korrekt in die Anschlussbuchse einstecken zu können, müssen diese Markierungen über­einander stehen.
Abb.5
GEFAHR
Verletzungsgefahr durch unzulässige Konfiguration bei Verwendung weiterer Komponenten!
Ź Sicherstellen, dass bei allen verwendeten Kom-
ponenten die Klassifikation mit der Klassifika­tion des Anwendungsteils (z. B. Typ BF oder Typ CF) des eingesetzten Geräts übereinstimmt.
GEFAHR
Lebensgefahr durch elektrischen Stromschlag!
Ź Produkt nur an ein Versorgungsnetz mit Schutz-
leiter anschließen.
Ź Gerät so aufstellen, dass das Ausschalten über
Schalter Netz-AUS/EIN problemlos möglich ist.
Ź Gerät so aufstellen, dass eine Trennung vom
Netzkabel problemlos möglich ist.
27
Ź Fußpedalstecker ausrichten und bis zum Einrasten in Anschlussbuchse
Fußpedal 3 einstecken. Bei erfolgreicher Verbindung zwischen Gerät und Fußpedal erscheint in
Anzeige 5 für 3 Sekunden die Meldung "Footswitch attached".
Abb.6
Hinweis
Fußpedal während der HF-Abgabe nicht kontinuierlich betätigen. Die HF­Abgabe wird gestartet und beendet durch einmaliges Drücken und Loslas­sen des Fußpedals.
Fußpedal entfernen
Ź Rast-Ring gegen Uhrzeigersinn drehen und gleichzeitig Anschlussste-
cker abziehen In Anzeige 5 erscheint für 3 Sekunden die Meldung "Footswitch remo-
ved".
Abb.7
Instrument anschließen
Ausschließlich Instrumente der CAIMAN Seal & Cut Produktfamilie kön­nen in Verbindung mit dem Lektrafuse HF-Generator GN200 eingesetzt werden.
Am Instrumentenstecker ist ein Pfeil und an der Anschlussbuchse 8 ein Punkt als Markierung angezeichnet. Um den Stecker korrekt in die Anschlussbuchse stecken zu können, müssen diese Markierungen überei­nander stehen.
Abb.8
Ź Instrument in Anschlussbuchse 8 stecken.
Der grüne Signalring 7 um den Instrumentenanschluss leuchtet auf. In Anzeige 5 erscheint die Meldung "Ready to Seal".
Abb.9
28
Aesculap®
Lektrafuse HF-Generator GN200
Betriebsart wählen
Der Lektrafuse HF-Generator besitzt zwei Betriebsarten für den Versiege­lungsprozess. Die gewählte Betriebsart bleibt bis zum Ausschalten des HF­Generators erhalten. Die Betriebsart kann während des Eingriffs geändert werden.
Ź Betriebsart wechseln: Taster Betriebsartenwahl (MODE) 2 betätigen.
Ŷ Standard-Modus: Voreingestellte Betriebsart Ŷ Plus-Modus: Betriebsart mit erhöhter Energieabgabe
Die eingestellte Betriebsart ist unabhängig vom angeschlossenen Instru­ment. Das Wechseln oder Wiederverbinden des Instruments ändert nicht die eingestellte Betriebsart.
Die aktive Betriebsart ist im Display folgendermaßen gekennzeichnet:
Ŷ Rahmen Ŷ Fette Schrift
Abb.10
Abb.11
Hinweis
Bei einem Regrasp-Fehler wird die aktive Betriebsart nicht angezeigt. Die aktive Betriebsart kann durch Betätigen des Tasters Betriebsartenwahl (MODE) 2 angezeigt werden.
Hinweis
Der Plus-Modus kann durch die höhere Frequenz des Pieptons während der HF-Abgabe vom Standard-Modus unterschieden werden.
Außerbetriebnahme
Ź Gerät mit Schalter Netz-AUS/EIN 12 ausschalten.
Alle Pole des Geräts werden vom Versorgungsnetz getrennt.
Ź Netzkabel abziehen.
Das Gerät ist vollständig vom Versorgungsnetz getrennt.
4.2 Funktionsprüfung
Korrekte Funktion des Geräts vor jedem Einsatz prüfen:
Ź Sicherstellen, dass das Zubehör keine sichtbaren Schäden hat. Ź Gerät vorbereiten und aufstellen, siehe Vorbereiten und Aufstellen. Ź Funktion folgender Elemente in der vorgegebenen Reihenfolge prüfen:
–Schalter Netz-AUS/EIN12 einschalten, Signallampe Netz-Ein 9
leuchtet
– Automatischer Selbsttest nach jedem Einschalten: kurzes akusti-
sches Signal, alle Anzeige-Elemente leuchten kurz auf – Betriebsartenwahl, siehe Betriebsart wählen – Fußpedal, siehe Fußpedal anschließen – Instrument, siehe Instrument anschließen – Aktivierung mit Taster an Instrument und Fußpedal –Schalter Netz-AUS/EIN12 ausschalten
4.3 Bedienung
Ź Start des Versiegelungsprozesses (HF-Abgabe) durch einmaliges Drü-
cken des Tasters am Instrument oder des Fußpedals. Nach Beendigung des Versiegelungsprozesses wird die HF-Abgabe
automatisch gestoppt. Während des Versiegelungsprozesses erscheint in Anzeige 5 die Mel-
dung "Sealing in Progress".
Abb.12
Gerät ein-/ausschalten
Ź Gerät mit Schalter Netz-AUS/EIN 12 einschalten/ausschalten.
Betriebsart
Der Lektrafuse HF-Generator GN200 erkennt automatisch das angeschlos­sene Instrument. Interne Geräteeinstellungen werden geladen. Der Nutzer kann zwischen zwei Betriebsarten wählen.
WARNUNG
Verletzungsgefahr und/oder Fehlfunktion!
Ź Vor jedem Gebrauch Funktionsprüfung durch-
führen.
WARNUNG
Verletzungsgefahr durch Verwendung des Produkts außerhalb des Sichtbereichs!
Ź Produkt nur unter visueller Kontrolle anwenden.
Loading...
+ 232 hidden pages