CAEN will repair or replace any product within the guarantee period if the Guarantor declares
that the product is defective due to workmanship or materials and has not been caused by
mishandling, negligence on behalf of the User, accident or any abnormal conditions or
operations.
CAEN declines all responsibility for damages or
injuries caused by an improper use of the Modules due
to negligence on behalf of the User. It is strongly
recommended to read thoroughly the CAEN User's
Manual before any kind of operation.
CAEN reserves the right to change partially or entirely the contents of this Manual at any time
and without giving any notice.
Disposal of the Product
The product must never be dumped in the Municipal Waste. Please check your local
regulations for disposal of electronics products.
3.2.LOCAL CONTROL ...................................................................................................................................17
3.4.1. Serial Links....................................................................................................................................27
3.4.1.1.USB communication ................................................................................................................................. 27
3.4.1.2.RS232 communication ..............................................................................................................................28
3.4.1.3.RS485 communication ..............................................................................................................................28
3.4.2. Communication Control................................................................................................................30
3.4.2.1.Remote Control: Main Menu.....................................................................................................................30
3.4.2.2.Remote Control: General Menu ................................................................................................................ 30
FIG.2.1:MOD.N1471 SERIES FRONT PANEL ............................................................................................................9
FIG.2.2:MOD.N1471 SERIES BACK PANEL............................................................................................................10
FIG.2.3:LOCAL CONTROL PANEL...........................................................................................................................11
FIG.2.4:CHANNEL CONTROL PANEL AND KILL SCHEME.........................................................................................11
FIG.2.5:N1471HVSTATUS CONTROL PANEL .......................................................................................................12
FIG.3.34:BOARD STATUS MENU...........................................................................................................................30
FIG.3.35:CHANNELS MENU ..................................................................................................................................31
Fig. 1.1: Mod. N1471 4 Channel Programmable HV Power Supply
The Mod. N1471 provides 4 independent High Voltage channels in a single width NIM
mechanics. Two and one channel versions (N1471A and N1471B) are also available.
Each channel can provide a ±5.5kV / 300 µA max output.
Channels have common floating return (common return insulated from the crate ground);
HV outputs are delivered through SHV connectors.
The HV output RAMP-UP and RAMP-DOWN rates may be selected independently for
each channel in the range 1÷500 V/s in 1 V/s steps.
Safety features include:
• OVERVOLTAGE and UNDERVOLTAGE warning when the output voltage differs from
the programmed value by more than 2% of set value (minimum 10V).
• Programmable VMAX protection limit
• OVERCURRENT detection: if a channel tries to draw a current larger than its
programmed limit, it enters TRIP status, keeping the maximum allowed value for a
programmable time (TRIP), before being switched off
• Channels can be enabled or disabled individually through the Interlock logic.
Module control can take place either locally, assisted by a Graphic color display or
remotely, via USB, RS232 or RS485, the latter allowing to build a N1471s’ daisy chain
network.
It is also controllable via TCP/IP by the Smart Fan Unit of CAEN NIM 8301 crate.
Control software is available (Mod SW1470), able to manage up to 32 modules (up to
128 channels).
Table 1.1: Available items
Code Model Description
WN1471X05AAC N1471 4 Channel NIM Programmable HV Power Supply (± 5.5 kV, 300 µA, 5 nA res.)
WN1471A05AAC N1471A 2 Channel NIM Programmable HV Power Supply (± 5.5 kV, 300 µA, 5 nA res.)
WN1471B05AAC N1471B 1 Channel NIM Programmable HV Power Supply (± 5.5 kV, 300 µA, 5 nA res.)
Local settings monitoring
Parameter and Mode setting
2.4.2. Channel control section
Fig. 2.4: Channel control panel and Kill scheme
NAME: TYPE: FUNCTION:
HV_EN/OFF/KILL 3 POS. SWITCH
ON RED LED
REMOTE KILL
+ GREEN LED
- YELLOW LED
AMP 280370-2
Channel Enable and turning OFF/KILL
HV On enabled
The channel is KILLED as no current flows across the
1Kohm resistor; this is achieved either as the 1-2 contacts
are open or as a +4÷6Vdc voltage is fed to pin 1 (see note)
Positive polarity
Negative polarity
1
NPO: Filename: Number of pages: Page:
00112/07:N1471.MUTx/01 N1471_REV1.DOC 38 11
1
OFF: Channel turned off according to RAMP DOWN setting; KILL: Channel turned off at fastest available rate
ON RED LED
ALARM RED LED/LEMO CONN. Out
INTERLOCK RED LED/LEMO CONN. In
2.4.3.1. Alarm signal
HV On enabled (at least one channel ON)
Alarm status signaled (active LOW)
Interlock signal
Fig. 2.6: N1471 ALARM electrical scheme
Alarm signal output signal electrical scheme is reported in the figure above. The
maximum output is 400mA@12V on ± Pins.
2.4.3.2. Interlock signal
Fig. 2.7: N1471 INTERLOCK electrical scheme
A schematic diagram of the Interlock input is shown in the figure above, where the diode
is part of optocoupler stage.
Interlock means that channels are hardware disabled. The interlock operation is
explained by the following table:
NPO: Filename: Number of pages: Page:
00112/07:N1471.MUTx/01 N1471_REV1.DOC 38 12
Loading...