CADD Prizm 6100, 6001 OPERATOR’S MANUAL

OPERATOR’S MANUAL
This online version differs from
Certain information that is not intended
for patients has been removed.
OPERATOR’S MANUAL
h
PCS
Model 6100 and 6101 Ambulatory Infusion Pump
/
Pain Control System
This manual concerns only the CADD-Prizm® PCS (Pain Control System) Model 6100 and 6101 ambulatory infusion pumps. This pump can be pro­grammed to deliver medication at a constant rate and/or allow delivery of a bolus dose at a specified time interval. This manual is intended for clinician use only. Do not permit patients to have access to this manual. The pump has three security levels designed to limit patient access. Do not disclose the pump’s security codes or any other information that would allow inappropriate access to programming and operating functions.
The issue date of this Operator’s Manual is included for the clinician’s informa­tion. In the event one year has elapsed between the issue date and product use, the clinician should contact Smiths Medical MD, Inc. to see if a later revision of this manual is available.
Technical Assistance
If you have comments or questions concerning the operation of the CADD­Prizm® PCS pump, please call the number given below. When calling, please specify the pump’s software module. This information is located in the pump’s start-up screen.
Our staff at Smiths Medical MD is available to help clinicians twenty-four
®
hours a day with the programming and operation of the CADD-Prizm
PCS
infusion system.
Smiths Medical MD, Inc. 1265 Grey Fox Road St. Paul, Minnesota 55112 U.S.A. 1 800.426.2448 +1 651.633.2556 www.smiths-medical.com
ii
Read this entire Operator’s Manual before operating the CADD-Prizm® PCS ambulatory infusion pump.
Failure to properly follow warnings, cautions, and instructions could result in death or serious injury to the patient.
WARNINGS
• This Operator’s Manual should be used by clinicians only. Do not permit patients to have access to this manual, as the information contained would allow the patient complete access to all programming and operating functions. Improper programming could result in death or serious injury to the patient.
• For those patients who are likely to be adversely affected by unintended operations and failures, including interrupted medication or fluid delivery from the device, close supervision and provision for immediate corrective action should be provided.
• If the pump is used to deliver life-sustaining medication, an additional pump must be available.
• The pump is not to be used for delivery of blood or cellular blood prod­ucts.
• If the pump is dropped or hit, inspect the pump for damage. Do not use a pump that is damaged or is not functioning properly. Contact Customer Service to return a pump for service.
• Use of a syringe with the CADD DER-DELIVERY of medication. Syringe function can be adversely affected by variations in plunger dimension and lubricity, which can result in greater force required to move the syringe plunger. A syringe plunger will lose lubrication as it ages and, as a result, the amount of under-delivery will increase which could on occasion, be significant. Therefore, the type of medication and delivery accuracy required must be considered when using a syringe with the CADD® pump.
Administration Set may result in UN-
Clinicians must regularly compare the volume remaining in the syringe to the pump’s displayed values such as RES VOL and GIVEN in order to determine whether under-delivery of medication is occurring and if neces­sary, take appropriate action.
• System delivery inaccuracies may occur as a result of back pressure or fluid resistance, which depends upon drug viscosity, catheter size, and extension set tubing (for example, microbore tubing).
iii
• Do not administer drugs to the epidural space or subarachnoid space unless the drug is indicated for those spaces.
• To prevent the infusion of drugs that are not indicated for epidural space or subarachnoid space infusion, DO NOT use administration sets that incorporate injection sites.
• If a Medication Cassette Reservoir, CADD
Extension Set or CADD
Administration Set is used for epidural space or subarachnoid space drug delivery, it is strongly recommended that it be clearly differentiated from those used for other routes of infusion, for example, by color coding, or other means of identification.
• When the Air Detector is not installed, or is installed but turned off, the pump will not detect air in the fluid path. It is recommended that you periodically inspect the fluid path and remove any air to prevent air embolism.
• Follow the Instructions for Use provided with the Medication Cassette Reservoir and CADD
Extension Set, or the CADD™ Administration Set, paying particular attention to all warnings and cautions associated with their use.
• When the Upstream Occlusion Sensor is turned Off, the pump will not detect occlusions upstream (between pump and fluid container). It is recommended that you periodically inspect the fluid path for kinks, a closed clamp, or other upstream obstructions. Upstream occlusions may result in under- or non-delivery of medications.
• Do not disclose to the patient the pump’s security codes or any other information that would allow the patient complete access to all program­ming and operating functions.
• Ensure that the ±6% System Delivery Accuracy specification is taken into account when programming the pump and/or filling the Medication Cassette Reservoir. Failure to do so may result in medication in the reser­voir becoming depleted sooner than expected.
• Do not use rechargeable NiCad or nickel metal hydride (NiMH) batteries. Do not use carbon zinc (“heavy duty”) batteries. They do not provide sufficient power for the pump to operate properly.
• Always have new batteries available for replacement. If power is lost, non­delivery of drug will occur.
iv
• There is no pump alarm to alert users that a battery has not been properly installed or has become dislodged. An improperly installed or dislodged battery could result in loss of power and non-delivery of drug.
• If the pump is dropped or hit, the battery door may become broken or damaged. Do not use the pump if the battery door is damaged because the battery will not be properly secured; this may result in loss of power or non-delivery of drug.
• When you enter a new Demand Dose Lockout time, any lockout time in effect will be cleared. A Demand Dose could be requested and delivered immediately upon starting the pump, resulting in over-delivery.
• When you enter a new Max Doses per Hour value, any lockout time in effect will be cleared. A Demand Dose could be requested and delivered immediately upon starting the pump, resulting in over-delivery.
• Exercise care when using the Clinician Bolus function. Since there are no limits on the frequency of delivering a bolus, and since the amount of the bolus can be set as high as 20 ml (or the mg or mcg equivalent), you should not permit the patient to become familiar with the procedure for giving a Clinician Bolus.
• To prevent the patient from accessing the Clinician Bolus function, do not let the patient know the Clinician Bolus code.
• Always close the fluid path tubing with the clamp before removing the cassette from the pump to prevent unregulated gravity infusion.
• Attach the cassette properly. A detached or improperly attached cassette could result in unregulated gravity infusion of medication from the fluid container or a reflux of blood. If you are using a Deltec administration set or medication cassette reservor that does not have the flow stop feature (reorder number does not start with 21-73xx): you must use a CADD
Extension Set with anti­siphon valve or a CADD™ Administration Set with either an integral or add-on anti-siphon valve to protect against unregulated gravity infusion that can result from an improperly attached cassette.
• Do not prime the fluid path with the tubing connected to a patient as this ­could result in over-delivery of medication or air embolism.
• Ensure that the entire fluid path is free of all air bubbles before connecting to the patient to prevent air embolism.
• If Demand Doses are currently locked out, changing the Date and/or Time
v
will cancel the lockout period. This will allow a Demand Dose to be requested and delivered as soon as you restart the pump, resulting in over­delivery.
CAUTIONS
• Do not operate the pump at temperatures below +2°C (36°F) or above 40°C (104°F).
• Do not store the pump at temperatures below -20°C (-4°F) or above 60°C (140°F). Do not store the pump with a Medication Cassette Reservoir or CADD
• Do not expose the pump to humidity levels below 10% or above 90% relative humidity.
• Do not store the pump for prolonged periods with the battery installed. Battery leakage could damage the pump.
• If you are using a Medication Cassette Reservoir in which the medication is frozen, thaw at room temperature only. Do not heat in a microwave oven as this may damage the product and cause leakage.
• Do not immerse the pump in cleaning fluid or water. Do not allow solu­tion to soak into the pump, accumulate on the keypad, or enter the battery compartment, Data In/Out jack, Power jack or Air Detector port area. Moisture build-up inside the pump may damage the pump.
Administration Set attached.
• Do not clean the pump with acetone, other plastic solvents, or abrasive cleaners, as damage to the pump may occur.
• Do not expose the pump to therapeutic levels of ionizing radiation as permanent damage to the pump’s electronic circuitry may occur. The best procedure to follow is to remove the pump from the patient during thera­peutic radiation sessions. If the pump must remain in the vicinity during a therapy session, it should be shielded, and its ability to function properly should be confirmed following treatment.
• Do not expose the pump directly to ultrasound, as permanent damage to the pump’s electronic circuitry may occur.
• Do not use the pump in the vicinity of magnetic resonance imaging (MRI) equipment as magnetic fields may adversely affect the operation of the pump. Remove the pump from the patient during MRI procedures and keep it at a safe distance from magnetic energy.
• Do not use the pump near ECG equipment as the pump may interfere with
vi
the operation of the equipment. Monitor ECG equipment carefully when using this pump.
• Do not sterilize the pump.
• Do not use the pump in the presence of flammable anesthetics or explosive gases.
• Use only Smiths Medical MD accessories as using other brands may adversely affect the operation of the pump.
• Check appropriate medication stability for time and temperature to assure stability with actual pump delivery conditions.
SYMBOLS
Collect Separately
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician.
vii
Table of Contents
Warnings iii
Cautions vi
Section 1: General Description 1
Introduction ......................................................................................... 1
Indications ........................................................................................... 1
Epidural/Subarachnoid Administration ................................................ 1
Pump Diagram .................................................................................... 3
Description of Keys, Display and Features .......................................... 4
The Main Screen ................................................................................. 8
Getting Help Using the
Lock Levels ......................................................................................... 10
AutoLock ...................................................................................... 10
Security Codes .............................................................................. 10
Customizing the Security Codes .................................................... 10
Lock Level Table ................................................................................. 11
Section 2: Pump Setup and Programming 13
Installing the Battery ........................................................................... 13
Watching Power Up ............................................................................ 15
Changing to Lock Level 0 (LL0) ......................................................... 16
Programming the Pump: General Instructions ..................................... 17
PCA Delivery Method .......................................................................... 18
PCA Programming Screens .................................................................. 19
PCA Programming Example ................................................................ 24
Removing a Cassette ........................................................................... 30
Attaching a Cassette ............................................................................ 31
Priming the Tubing and Connecting to the Patient ............................... 34
Inserting the Tubing into the Air Detector ......................................... 36
Setting the Lock Level for the Patient .................................................. 38
Starting the Pump ................................................................................ 39
Programming with Upper Limits, Adjusting Doses in Lock Level 1 ..... 40
Key ........................................................ 9
Å
Section 3: Operating the Pump 41
Stopping the Pump .............................................................................. 41
Starting the Pump ................................................................................ 41
Starting a Clinician Bolus .................................................................... 43
Starting a Demand Dose ..................................................................... 45
Stopping a Demand Dose or Clinician Bolus ........................................ 46
viii
Resetting the Reservoir Volume .......................................................... 47
Section 4: Options 49
Overview: Accessing Options .............................................................. 49
Prime .................................................................................................. 50
Extended History, Viewing .................................................................. 51
AutoLock ............................................................................................ 53
Time ................................................................................................... 54
Date .................................................................................................... 55
Air Detector On/Off ............................................................................ 56
Event Log, Viewing ............................................................................. 57
Section 5: Biomed Toolbox 59
Overview: Accessing the Biomed Toolbox ........................................... 59
Micrograms On/Off ............................................................................ 60
Custom Concentrations ...................................................................... 60
Extended History On/Off ................................................................... 62
Max Doses Per Hour On/Off ................................................................ 62
PM (Preventive Maintenance) Reminder ............................................. 63
Custom Lock Level Code .................................................................... 63
Date Format ........................................................................................ 64
Power Source Status Display ................................................................ 65
Upstream Sensor On/Off (Model 6101 only) ....................................... 65
Air Detector Requirement ................................................................... 66
Description
General
Programming
Setup &
Operating
the Pump
Options
Section 6: Reference & Troubleshooting 67
Troubleshooting .................................................................................. 67
Alarms and Messages, Alphabetical List ............................................. 69
Cleaning the Pump and Accessories .................................................... 77
Cleaning the Battery Contacts ........................................................... 78
Exposure to Radiation or Magnetic Resonance Imaging (MRI) .......... 79
Continuous Rate Scroll Ranges ........................................................... 80
Demand Dose, Clinician Bolus Scroll Ranges: Milliliters ..................... 80
Demand Dose, Clinician Bolus Scroll Ranges: Milligrams ................... 81
Demand Dose, Clinician Bolus Scroll Ranges: Micrograms ................. 82
Military Time Conversion Chart ......................................................... 83
Specifications (Nominal) ..................................................................... 84
Printing Reports ................................................................................... 87
Index 88
Limited Warranty 91
ix
Toolbox
Biomed
Troubleshooting
Reference &
x
Section 1: General Description
Section 1: General Description
Introduction
The CADD-Prizm® PCS (Pain Control System) ambulatory infusion system provides measured drug therapy to patients in hospital or outpatient settings. Therapy should always be overseen by a physician or a certified, licensed healthcare professional. As appropriate, the patient should be instructed in using the pump.
Indications
The CADD-Prizm® PCS pump is indicated for intravenous, subcutaneous, epidu­ral space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both (such as patient-controlled analgesia).
Description
General
Epidural/Subarachnoid Administration
The selected drug must be used in accordance with the indications included in the package insert accompanying the drug. Administration of any drug by this pump is limited by any warnings, precautions, or contraindications in the drug labeling.
Analgesics
Administration of analgesics to the epidural space is limited to use with ind­welling catheters specifically indicated for either short- or long-term drug delivery.
Administration of analgesics to the subarachnoid space is limited to use with indwelling catheters specifically indicated for short-term drug delivery.
Anesthetics
Administration of anesthetics to the epidural space is limited to use with indwelling catheters specifically indicated for short-term drug delivery.
1
General
Description
Section 1: General Description
WARNING:
• Do not administer drugs to the epidural space or subarachnoid space unless the drug is indicated for administration to those spaces. Drugs not intended for epidural or subarachnoid space infusion could result in death or serious injury to the patient.
• To prevent the infusion of drugs that are not indicated for epidural space or subarachnoid space infusion, DO NOT use administration sets that incorpo­rate injection sites. The inadvertent use of injection sites for infusion of such drugs may cause death or serious injury to the patient.
• If a Medication Cassette Reservoir, CADD
Extension Set or CADD
Administration Set is used for epidural space or subarachnoid space drug delivery, it is strongly recommended that it be clearly differentiated from those used for other routes of infusion, for example, by color coding, or other means of identification. Drugs not intended for epidural or subarach­noid space infusion could result in death or serious injury to the patient.
2
Pump Diagram
Keypad
Power jack
Data In/Out
Air Detector
Port Cover
Display
jack
Indicator Lights
Amber Green
Section 1: General Description
Front View
Description
General
Air Detector
(optional)
Polemount
Bracket
Recess
Battery
compartment
Cassette
lock
Cassette
latch
Cassette
®
2000-03-07 D. Zurn
«Prizm Rear 3/4 (dark BW)»
Rear View
®
3
Section 1: General Description
Description of Keys, Display and Features
General
Description
Indicator Lights
Green: The green light blinks approximately every 3 seconds when the pump is running and delivering fluid as programmed.
Amber: The amber light flashes when the pump is stopped or an alarm condi­tion exists. It stays on continuously when the pump is inoperable. The display briefly describes the condition.
If both lights blink, delivery is still occurring but a condition exists of which you should be aware (for example, a low battery). Look at the display for a brief description of the condition.
Display with backlighting
The liquid crystal display (LCD) shows programming information and mes­sages. Backlighting helps keep the display visible in low light.
After a period of no key presses, backlighting turns off and the display blanks to save battery power (except during an alarm or when an external power source is in use). Press any key to turn the display back on.
NOTE: If you press you wish to start or stop the pump; press ´or Î. Do not use the display back on; this may deliver an inadvertent dose.
, the display will reappear with a message asking if
Í
to turn
Keypad
The keys on the keypad are described below. A key beeps when pressed if it is operable in the current lock level.
⁄ Œ
Å ¤
4
starts and stops pump delivery.
is used to view or change the pump’s current lock level. Lock levels are used to limit patient access to certain programming and operat­ing functions. (See Lock Levels, this section.) This key is also used to access the Clinician Bolus while the pump is running.
is the “Help” key. It is used to display help for a screen or an alarm message. (See Getting Help, this section.)
is used to enter, or save, a new value in the pump’s memory when programming new doses or new pump settings. It is also used to select an item from the Options Menu (Section 4) or Biomed Toolbox Menu (Section 5).
is used to move from one programming screen to the next without
Section 1: General Description
changing the setting or value displayed. It is also used to return from the Biomed Toolbox Menu to the Options Menu, or from the Options Menu to the main screen. (See Sections 4 and 5.)
Description
General
Í ‹ ´
Î
allows the patient to deliver a programmed amount of medication upon request.
is used to access the Options Menu, which contains such features as time, date, and the Event Log. (See Section 4, Options.)
allows you to answer “yes” to a question on the pump’s display, “scroll up” or increase a value (for example, a dose amount), or scroll through items on a menu.
allows you to answer “no” to a question on the pump’s display, “scroll down” or decrease a value, scroll through items on a menu, or cancel printing.
Power jack
You may plug a CADD™ External Power Source (EPS) system power pack or an AC Adapter into the Power jack as an alternate source of power.
Data In/Out jack
The Data In/Out jack is used for attaching the following accessories:
Interface Cable for printing reports
Remote Dose Cord for remote operation of the dose key
Modem Cable for communications
For more information on the Remote Dose Cord, printing or communications, see the instructions for use provided with those products.
Air Detector Port Cover
This encloses the Air Detector port when the Air Detector is not attached.
Air Detector accessory (optional)
The Air Detector attaches to the pump in the area shown in the diagram. If air is detected in the part of the tubing that passes through the Air Detector, an alarm sounds and delivery stops. (See Section 6 for Air Detector specifications.) The pump may be customized to require an Air Detector. (See Section 5, Biomed Toolbox.) If an Air Detector is attached but not required, it may be turned off.
WARNING: When the Air Detector is not installed, or is installed but turned
5
General
Description
Section 1: General Description
off, the pump will not detect air in the fluid path. It is recommended that you periodically inspect the fluid path and remove any air to prevent air embolism. Air embolism could result in death or serious injury to the patient.
Cassette
The cassette is the part of the Medication Cassette Reservoir or CADD Administration Set that attaches to the bottom of the pump. The following single-use products are compatible with the CADD-Prizm® pump:
Medication Cassette Reservoir (50 or 100 ml), used with a CADD Extension Set
CADD
Administration Set
WARNING: Follow the Instructions for Use provided with the Medication Cassette Reservoir and CADD™ Extension Set, or the CADD™ Administration Set, paying particular attention to all warnings and cautions associated with their use. Incorrect preparation and/or use of these products could result in serious patient injury or death.
Polemount Bracket recess
The optional Polemount Bracket slides into the recess on the back of the pump, allowing you to attach the pump to an IV pole.
Battery compartment
The 9 volt battery fits into this compartment. The 9 volt battery serves as the primary source of power, or as a backup when an EPS System power pack or AC Adapter is in use.
Cassette latch
This attaches the cassette to the pump. The pump detects whether the cassette is latched properly. Delivery will stop and an alarm will occur if the cassette becomes unlatched.
Cassette lock
This allows you to secure the cassette to the pump using the key provided. The cassette must be latched before it can be locked. The cassette must be locked onto the pump or the pump will not run.
6
Section 1: General Description
Other Features Not Shown
Downstream Occlusion Sensor: The pump contains a downstream occlusion sensor. When a downstream occlusion between the pump and patient access site is detected, an alarm will sound, delivery will stop, and the display will show “High Pressure.”
Upstream Occlusion Sensor: (Model 6101 only) The pump contains an up­stream occlusion sensor. This feature may be turned on or off (see Section 5, Biomed Toolbox). When the sensor is turned on, and an upstream occlusion (between pump and fluid container) is detected, an alarm will sound, delivery will stop, and the display will show “Upstream Occlusion.”
WARNING: When the Upstream Occlusion Sensor is turned Off, the pump will not detect occlusions upstream (between pump and fluid container). It is recommended that you periodically inspect the fluid path for kinks, a closed clamp, or other upstream obstructions. Upstream occlusions may result in under- or non-delivery of medications to the patient. If undetected, these occlusions could lead to death or serious injury to the patient.
Reservoir Volume Alarm: Reservoir Volume is a feature that indicates when the
fluid in the fluid container is low or depleted. Each time you change the fluid container, you may reset the Reservoir Volume to the originally programmed volume. Then, as medication is delivered, the Reservoir Volume automatically decreases. When the pump calculates that 5 ml remain in the fluid container, beeps sound and “Reservoir Volume Low” appears. This alarm recurs at every subsequent decrease of 1 ml until the Reservoir Volume reaches 0 ml, at which point the pump stops.
Description
General
7
General
Description
Section 1: General Description
The Main Screen
The main screen is the starting point for programming or viewing the pump’s settings. The following information may be displayed:
Active delivery mode
When running:
Power source display
The current
*** PCA 6210X *** Battery Low RUNNING Res Vol 50.0 ml Press NEXT to advance
Reservoir Volume
If no keys are pressed for a period of time (2 minutes when the pump is stopped, 1 minute when running), the display reverts to the main screen. When the 9 volt battery is low, “Battery Low” appears on the main screen. You can configure the pump so the main screen always displays the type of power source in use. (See Power Source Display, Section 5.)
running, stopped, or dosing
A reminder that the
key lets you
advance (to program or review settings)
Whether the pump is
8
Section 1: General Description
Description
General
Getting Help Using the
If you have a question about a screen, press the tion of the screen will appear along with instructions for pump operations you may be trying to perform. The following is an example of help for the Reser- voir Volume screen:
This is the calculated amount of fluid left in the reservoir. ?≥
To page through all the help screens, press original screen will reappear when no further help is available.
To exit help, press any key (other than the bring you back to the original screen.
If a help screen tells you to press a certain key, first exit help, then press that key.
Help screens are lock level dependent. If the pump’s current lock level prevents access to a certain function, the function will not be described in the help screens.
ÅÅ
Å
ÅÅ
Key
key for help. A descrip-
Å
The symbol in the lower right corner means there are additional help screens. Press the again to see the next help screen.
repeatedly. The
Å
key). This will
Å
Å
key
9
Section 1: General Description
Lock Levels
General
Description
Lock levels are used to limit patient access to certain programming and operat­ing functions. The table on the next page lists the functions that are accessible in Lock Level 0 (LL0), Lock Level 1 (LL1), and Lock Level 2 (LL2). When a function is accessible, the key associated with the function beeps when pressed. If a function is not accessible, the pump ignores the key press and a beep does not sound. Section 2, Pump Setup and Programming, describes how to change the lock level.
AutoLock
AutoLock is one of the Options. This feature automatically changes the lock level from LL0 to LL1 or LL2 when the pump is started (instead of requiring you to manually change the lock level before giving the pump to the patient). See Section 4 for more information on using AutoLock.
Security Codes
The following security codes are preset by the manufacturer for the clinician’s use:
** Text omitted from Online version **
WARNING: Do not disclose to the patient the pump’s security codes or any other information that would allow the patient complete access to all program­ming and operating functions. Improper programming could result in death or serious injury to the patient.
Customizing the Security Codes
If it becomes necessary to change the Lock Level Code and Biomed Toolbox Code to ensure that a patient will be unable to access these features, you may customize the Lock Level Code in the Biomed Toolbox. (See Section 5.) Cus­tomizing the Lock Level Code will not affect the Clinician Bolus Code.
10
Section 1: General Description
Lock Level Table
This table lists the operations that are accessible in each lock level while the pump is stopped and running. LL0 permits complete access to all programming and operating functions. LL1 permits limited control of pump programming and operations. LL2 permits only minimal control of the pump.
Description
General
Pump Operations and Programming LL0 LL1 LL2 Any Lock Level
Stop/Start the pump Yes Yes Yes Yes
View Help screens Yes Yes Yes Yes
Print Yes Yes Yes Yes
Reset Reservoir Volume Yes Yes Yes
Change the lock level Yes, w/code Yes, w/code Yes, w/code
Start a Demand Dose
Start a Clinician Bolus
Change Units Yes
Change Concentration Yes
Change Continuous Rate Yes Up to LL0 value
Change Demand Dose Yes Up to LL0 value
Clear Dose Counters Yes Yes
Clear Given amount Yes Yes
Add New Patient Marker Yes
Pump Options
No No No
No No No
Stopped Running
No
No
Yes
Yes, w/code
No No No
No No No
No No
No No
No No
No No
No No No
Prime Yes Yes
Extended History (view) Yes Yes Yes Yes
AutoLock Yes View only View only View only
Time Yes View only View only View only
Date Yes View only View only View only
Air Detector On/Off Yes View only View Only View Only
Event Log (view) Yes Yes Yes Yes
Biomed Toolbox Yes, w/code
No No No
No No
11
General
Description
Section 1: General Description
12
Section 2: Pump Setup and Programming
Section 2: Pump Setup and Programming
Installing the Battery
Use a new, 9 volt alkaline or lithium battery such as the DURACELL® Alkaline MN 1604, the EVEREADY® ENERGIZER® Alkaline #522 or the ULTRA­LIFE® Lithium U9VL battery. The pump retains all programmed values while the battery is removed. If the pump is running, you may connect an external power source to keep the pump running for 3 minutes while you change the battery.
Dispose of used batteries in an environmentally safe manner, and according to any regulations which may apply.
WARNING:
• Do not use rechargeable NiCad or nickel metal hydride (NiMH) batteries. Do not use carbon zinc (“heavy duty”) batteries. They do not provide sufficient power for the pump to operate properly, which could result in death or serious injury to the patient.
• Always have new batteries available for replacement. If power is lost, non­delivery of drug will occur and, depending on the type of drug being admin­istered, could result in death or serious injury to the patient.
• There is no pump alarm to alert users that a battery has not been properly installed or has become dislodged. An improperly installed or dislodged battery could result in loss of power and non-delivery of drug and, depend­ing on the type of drug being administered, could result in death or serious injury to the patient.
Programming
Setup &
• If the pump is dropped or hit against a hard surface, the battery door may become broken or damaged. DO NOT USE the pump if it has been dam­aged in this way because the battery will not be properly secured; this may result in loss of power, non-delivery of drug, and, depending on the type of drug being administered, death or serious injury to the patient.
13
Section 2: Pump Setup and Programming
Ope
t
To install a battery
Setup &
Programming
1. Make sure the pump is stopped. Press the button on the battery door and slide the battery door forward. Remove the used battery.
2. Match the + and – markings on the new battery with the mark­ings on the pump. Insert the battery. The pump will beep if the battery is inserted correctly.
3. Replace the battery door. The pump will begin to power up.
NOTE: If you put the battery in backwards, the display will remain blank. Reinsert the battery, making sure to match the + and – markings.
atch
ery»
/Button»
CAUTION: Do not store the pump for prolonged periods with the battery installed. Battery leakage could damage the pump.
NOTE:
Battery life is dependent on the amount of medication delivered, delivery rate, battery age, temperature, frequent screen display and backlighting and frequent printing.
The power of the battery will be quickly depleted at temperatures below +10˚C (50˚F).
14
Section 2: Pump Setup and Programming
Watching Power Up
When you install a battery, the pump will start its power up sequence during which it performs self-tests and displays programmed values. Watch for the following:
Pump model number, last error code (“lec”) if any, and serial number (“sn”) will appear.
The delivery mode contained in the pump and its software version will appear.
The display will turn completely on. Look for any stripes, which would indicate a faulty display.
If no Air Detector is attached, “No Air Detector attached” will appear. The pump’s program screens will appear, followed by screens showing the lock level setting, AutoLock setting (if in use), Air Detector status (if an Air Detector is attached), time, and date. You may need to confirm certain settings before power up will continue. If messages appear, see the Alarms and Messages Table in Section 6 of this manual for further explanation and instruction.
Programming
Setup &
NOTE:
The pump will briefly pause. Then a message will appear showing that the PCA delivery mode is currently active.
When power up is complete, “Power Up Successful” will appear, six beeps will sound, and the pump will be stopped.
When the pump is powered up in Lock Level 0 with an Air Detector attached, the pump will automatically turn on the Air Detector (the Air Detector setting in Options will change to “Turned On.”)
To move quickly through the power up screens, press edly. To skip the automatic review entirely, press Î.
repeat-
15
Section 2: Pump Setup and Programming
Changing to Lock Level 0 (LL0)
Before programming the pump, make sure the lock level is LL0. LL0 allows the clinician to access all programming and operating functions.
To change the lock level
Setup &
Programming
WARNING: Do not disclose to the patient the pump’s security codes or any other information that would allow the patient complete access to all program­ming and operating functions. Improper programming of the pump could result in death or serious injury to the patient.
1. Make sure the pump is stopped. Press appear. (If the lock level is already LL0, press
2. Press
3. Press
NOTE: If <Custom> appears on the screen, the Lock Level Code has been customized. Enter the custom Lock Level Code in the next step.
4. Press ´ or
** Text Omitted **
. The current lock level will
Œ
to exit.)
or
´
Œ
until “LL0” appears.
Î
again. “000” will appear.
until
Î
Lock Level
| LL2
Lock Level
LL0
Lock Level Code
000
Lock Level Code
***
5. Press Watch the display to verify that the correct lock level is being entered. If you do not see this message, the lock level has not changed. Repeat the above steps.
NOTE: To check the lock level, press To return to the screen you were on, press
16
to set the new lock level.
Œ
Œ
Lock Level
LL0
<Changing...>
. The current lock level will appear.
.
Section 2: Pump Setup and Programming
Programming the Pump: General Instructions
The procedure for changing a programmed setting is similar for most program­ming screens. The following example of the Reservoir Volume screen illustrates the typical features of a programming screen:
The | symbol means you
may use the
key to change the setting
´
or
Î
Reservoir Volume
| 10.0 ml
Screen name
Programming
Setup &
“Range” shows the
minimum and
maximum settings
Make sure the pump is stopped and in Lock Level 0.
To begin programming, start at the main screen and press
To change a setting, press pears. (Press and hold these keys to change values with increasing speed.) Then press will appear automatically.
To leave a setting unchanged, press
¤
<Range: 1 - 9999>
or
´
to save the new setting. The next screen
until the desired setting ap-
Î
to go to the next screen.
Setting
Messages you may see during programming
During programming, the following messages may appear:
Press ENTER to save” will appear 10 seconds after you change a setting to remind you to press
“Entering...” “Changing...” or “Resetting...” means the new setting is being entered into the pump’s memory. The pump will display this message, then automatically go to the next screen.
Change — to...?” may appear for the following reasons:
¤
.
.
you entered a new setting that must be confirmed,
entry is required because you changed Units or Concentration, or
you changed a setting and pressed a key other than
Press 20 seconds, or if you press setting.PCA Delivery Method
to confirm the setting. If you do not press ´ within
´
, the screen will revert to the previous
Î
¤
.
17
Section 2: Pump Setup and Programming
PCA Delivery Method
The PCA delivery mode provides the following methods of delivery:
Continuous Rate
Demand Dose, activated by the patient
Clinician Bolus, a dose activated by the clinician.
Setup &
Programming
You may program each of the methods individually or in combination with each other. The Continuous Rate and Demand Dose are programmed as described in this section. The Clinician Bolus feature is described in Section 3, Operating the Pump. Ranges and programming increments are listed in the Specifications in Section 6.
Clinician Bolus (used here as a loading dose)
Demand Doses
Continuous Rate
Dosage
18
Time
Section 2: Pump Setup and Programming
Programming Screens for PCA Delivery
The following figure illustrates the programming screens that are available in the PCA delivery mode:
Reservoir Volume
Units
Concentration (ml, mg or mcg)
Continuous Rate
Demand Dose
Reservoir Volume
| 100.0 ml
<Range: 1 - 9999>
Units
| Milligrams
<Range: mg or ml>
Concentration
| 1.0 mg/ml
<Range: 0.1 - 100>
Continuous Rate
| 5.00 mg/hr
<Range: 0 - 30.00>
Demand Dose
| 2.50 mg
<Range: 0 - 9.90>
Programming
Setup &
Demand Dose Lockout Time
Max Doses per Hour
Demand Dose Lockout
| 15 Min
<Range: 5 min-24 hr>
Max Doses Per Hour
| 2
<Range: 1 - 4>
19
Section 2: Pump Setup and Programming
Dose Counter
Given
Setup &
Programming
Air Detector (review)
Dose Counters Given/Attempt: 0/ 0 since 06/08/05 10:35 Press ENTER to clear
Milligrams Given
0.00 mg since 06/08/05 10:35 Press ENTER to clear
Air Detector
Required
<Review Only>
New Patient Marker (optional)
To insert New Patient Marker and clear Extended History press ENTER
Reservoir Volume
Enter the volume of fluid contained in the fluid container. The Reservoir Volume value decreases as the pump delivers fluid or you use the priming feature. When you change the fluid container and reset the Reservoir Volume, the value resets to the value entered on this screen. If you do not wish to use the Reservoir Volume feature, scroll down to “Not In Use” (located before 1 and after 9999 in the range of values).
Units
Enter the programming units. Possible settings are milliliters and milligrams. Micrograms will also be one of the choices if the Micrograms settings in the Biomed Toolbox is “On.” When you change the Units, the pump requires you to enter or verify the Continuous Rate and Demand Dose. If the units are mg or mcg, you must also enter the Concentration. Changing the Units clears the amount Given and the Extended History.
20
Loading...
+ 72 hidden pages