BW Technologies GasAlertQuattro 3, GasAlertQuattro 2, GasAlertQuattro 1, GasAlertQuattro 4 Operator's Manual

Page 1
1, 2, 3, and 4-Gas Detector
Operator’s Manual
Page 2
GasAlertQuattro
Introduction
The detector is a personal safety device. It is your responsibility to respond properly to the alarm.
The detector is shipped with English as the default
displayed language. Additional languages provided
are French, German, Portuguese, and Spanish. The
screens for the additional languages are displayed on
the detector and in the corresponding operator s
manual.
Note
Zeroing the Sensors
To zero the sensors, refer to steps #1-3 in Calibration on page 8.
Safety Information - Read First
Use the detector only as specified in this operator’s manual and the technical reference guide, otherwise protection provided by the detector may be impaired. Read the following Cautions before using the detector.
a
Cautions
Warning: Substitution of components may impair Intrinsic
Safety.
• Before using the detector, refer to Sensor Poisons and Contaminants.
• Protect the combustible sensor from exposure to lead compounds, silicones, and chlorinated hydrocarbons. Although certain organic vapors (such as leaded gasoline and halogenated hydrocarbons) may temporarily inhibit sensor performance, in most cases the sensor will recover after calibration.
• Caution For safety reasons, this equipment must be operated and serviced by qualified personnel only. Read and understand the technical reference guide completely before operating or servicing.
1
Page 3
GasAlertQuattro
Operator’s Manual
• Charge the detector before first-time use. BW Technologies by Honeywell recommends the detector be charged after every workday.
• Calibrate the detector before first-time use and then on a regular schedule, depending on use and sensor exposure to poisons and contaminants. BW recommends that the sensors must be calibrated regularly and at least once every 180 days (6 months).
• Calibrate only in a safe area that is free of hazardous gas in an atmosphere of 20.9% oxygen.
• The combustible sensor is factory calibrated to 50% LEL methane. If monitoring a different combustible gas in the % LEL range, calibrate the sensor using the appropriate gas.
• Only the combustible gas detection portion of this instrument has been assessed for performance by CSA International.
• BW recommends that the combustible sensor be checked with a known concentration of calibration gas after any exposure to contaminants/poisons such as sulfur compounds, silicon vapors, halogenated compounds, etc.
• BW recommends to bump test the sensors before each day’s use to confirm their ability to respond to gas by exposing the detector to a gas concentration that exceeds the alarm setpoints. Manually verify that the audible, visual, and vibrator alarms are activated. Calibrate if the readings are not within the specified limits.
• Caution High off-scale LEL readings may indicate an explosive concentration.
2
• Any rapid up scaling reading followed by a declining or erratic reading may indicate a gas concentration beyond upper scale limit, which can be hazardous.
• For use only in potentially explosive atmospheres where oxygen concentrations do not exceed 20.9% (v/v).
• Extended exposure of the GasAlertQuattro to certain concentrations of combustible gases and air may stress the detector element that can seriously affect its performance. If an alarm occurs due to a high concentration of combustible gases, a calibration should be performed, or if needed, the sensor replaced.
Warning: The lithium battery (QT-BAT-R01) may present a risk of fire or chemical burn hazard if misused. Do not disassemble, heat above 212°F (100°C), or incinerate.
Warning: Do not use any other lithium batteries with the GasAlertQuattro detector. Use of any other cell can cause fire and/or explosion. To order and replace the QT-BAT-R01 lithium battery, contact BW Technologies by Honeywell.
Warning: Lithium polymer cells exposed to heat at 266°F (130°C) for 10 minutes can cause fire and/or explosion.
ec Warnin g: This instrument contains a lithium polymer battery. Dispose of used lithium cells immediately. Do not disassemble and do not dispose of in fire. Do not mix with the solid waste stream. Spent batteries must be disposed
of by a qualified recycler or hazardous materials handler.
• Keep lithium cells away from children.
• Deactivating the detector by removing the battery pack may cause improper operation and harm the detector.
Page 4
Parts of the GasAlertQuattro
Parts of the GasAlertQuattro
Item Description Item Description Item Description Item Description
Confidence/compliance
1
indicator (green LED)
Visual alarm indicator
2
(red LED)
Hydrogen sulfide
3
(H2S) sensor
4
Pushbutton
Combustible (LEL)
5
sensor
Carbon monoxide
6
(CO) sensor
Liquid crystal display
7
(LCD)
8
Audio alarm
9
Oxygen (O2) sensor
10
11
12
GasAlertQuattro
Alligator clip
Battery pack
Charging connector and IR interface
3
Page 5
GasAlertQuattro
Operator’s Manual
Screen Elements
Calibration gas cylinder
Bump test gas cylinder
Indicates pass for startup, sensors, calibrations, and bump tests
Indicates fail for startup, sen­sors, calibrations, and bump tests
Pushbutton displays when screen provides an option to end or skip
Displays when the Stealth option is enabled
Display during startup to indicate audio and visual alarm pass or fail during a MicroDock II bump test
Displays during an operation such as charging or auto­zeroing
Displays when the most recent calibration failed but a previous calibration is still valid within the calibration due interval
4
Reading displays with white background during normal operation
Reading displays with black background when the sensor is in alarm
Displays when the detector is in alarm (not applicable to TWA and STEL)
Displays when there is a warning, failure, error, or low battery
Heartbeat pulses continually during normal operation to verify the detector is operating correctly
Displays for STEL alarms and setpoints
Displays for TWA alarms and setpoints
Displays when maximum concentration encountered during work shift displays
Displays if calibration is initi­ated and the IR Lock option is enabled
Displays during calibration and when startup is complete
Battery — full charge
Battery — half charge
Low battery warning
Displays when the detector is connected to an IR Link
Grey check box displays dur­ing bump tests or calibration when a gas is not due
Displays when the detector’s fi rmware is being updated
Page 6
Pushbutton
Pushbutton Description
• To activate the detector, press and hold C in a safe area that is free of hazardous gas and
in an atmosphere of 20.9% oxygen.
• To deactivate the detector, press and hold C during the powering off countdown. Release
C when OFF displays.
• To view the date/time, current battery power, calibration due date, bump test due date,
TWA, STEL, and peak readings, press C twice rapidly. To clear the TWA, STEL, and peak readings, press C when the LCD displays Hold C to reset peaks, TWA, STEL.
C
• To initiate calibration, press and hold C while the detector performs the OFF countdown. Continue holding C while the LCD briefly deactivates and then reactivates to begin the calibra ion countdown. Release C when Calibration started displays.
• To activate the backlight, press C and release.
• To acknowledge latching alarms, press C.
• To acknowledge a low alarm and disable the audible alarm, press C. The Low Alarm Acknowledge option must be enabled in Fleet Manager II.
• To acknowledge any of the “due today” messages (calibration and bump test) press C. If enabled, the force calibration and force bump features cannot be bypassed.
GasAlertQuattro
Pushbutton
5
Page 7
GasAlertQuattro
Operator’s Manual
Sensor Poisons and Contaminants
Several cleaners, solvents, and lubricants can contaminate and cause permanent damage to sensors. Before using cleaners, solvents, and lubricants in close proximity to the detector sen­sors, read the following caution and table.
a
Caution
Use only the following BW Technologies by Honeywell recommended products and procedures
• Use water based cleaners.
• Use non-alcohol based cleaners.
• Clean the exterior with a soft, damp cloth.
• Do not use soaps, polishes, or solvents.
6
The following table lists common products to avoid using around sensors.
Cleaners and
Lubricants
Brake cleaners
Lubricants
Rust inhibitors
Window and glass cleaners
Dishsoaps Mold releasing agents
Citrus based cleaners
Alcohol based cleaners
Hand sanitizer
Anionic detergents
Methanol (fuels and antifreezes)
Silicones Aerosols
Silicone cleaners and protectants
Silicone based adhesives, sealants, and gels
Hand/body and medicinal creams that contain silicone
Tissues containing silicone
Polishes
Bug repellents and sprays
Lubricants
Rust inhibitors
Window cleaners
Page 8
Connecting the Gas Cylinder to the Detector
Connecting the Gas Cylinder to the Detector
GasAlertQuattro
7
Page 9
GasAlertQuattro
Operator’s Manual
Calibration
Calibration is performed to adjust the sensitivity levels of sensors to ensure accurate responses to gas.
This calibration procedure is written as the procedure is intended. If an error or alarm screen displays, refer to Calibra ion Trouble­shooting in the GasAlertQuattro Technical Reference Guide.
a Caution
Calibrate only in a safe area that is free of hazardous gas in an atmosphere of 20.9% oxygen.
If performing single gas calibration, calibrate O
Note
The maximum hose length for calibration is 3 ft. (1 m).
The following steps are written for use with a standard quad gas cylinder.
Calibration can only be aborted after the sensors have been zeroed. If C is pressed to abort, CALIBRATION cancelled displays.
1. Press and hold C as the detector performs the Powering off countdown.
8
first.
2
Continue holding C when OFF displays and the detector briefly deactivates.
2. The detector activates again and performs the calibration countdown. Continue holding until Starting Calibration displays.
3. The detector enters the zero function. zeroing displays while the detector zeros all of the sensors.
a
Caution
If a sensor fails to zero, it cannot be calibrated. Refer to Startup Self-Test Troubleshooting in the
GasAlertQuattro Technical Reference Guide.
Page 10
If the IR Lock option is enabled, the following screen displays to indicate calibration can only be performed using an IR device (MicroDock II or
R Link).
4. When the following screen displays, attach the calibration cap and apply calibration gas at a flow rate of 250-500 ml/min. Refer to
Connecting the Gas Cylinder to the Detector.
If a sensor is not yet due for calibration, its box will have a greyed out checkmark.
5. The detector initially tests for gas. When a sufficient amount of
gas is detected, displays beside each gas that is detected.
6. The detector then begins calibrating the sensors. The following activities occur during the span:
calibrating displays at the bot­tom of the screen.
• Gas values adjust during the span.
• Target gas values that are defined in Fleet Manager II dis­play above or below the adjust­ing gas value.
To abort calibration after the sensors have been zeroed, press C.
7. When the following screen displays, close the valve on the gas cylinder and remove the calibration cap from the detector.
A check mark displays beside each sensor that has calibrated successfully.
GasAlertQuattro
Calibration
9
Page 11
GasAlertQuattro
Operator’s Manual
8. When calibration is complete, the following screen displays.
Note
The calibration due date cannot be reset for a sensor that fails calibration. If a sensor fails or an error screen displays, refer to Calibration Troubleshooting in the GasAlertQuattro Technical Reference Guide.
9. All successfully calibrated sensors automatically reset to the number of days defined in the Cal Interval field in Fleet Manager II.
The calibration due dates can be changed in Fleet Manager II.
10. The detector now enters normal operation.
10
Bump Test
A bump test applies test gas to force the detector into alarm. A bump test must be performed regularly to confirm the sensors are responding correctly to gas, and that the audible, visual and vibrator alarms activate during an alarm condition.
BW recommends to bump test the sensors before each day’s use to confirm their ability to respond to gas by exposing the sensors to a gas concentration that exceeds alarm setpoints.
1. Connect the calibration hose to the 0.5 l/min regulator
on the gas cylinder. Refer to Connecting the Gas Cyl-
inder to the Detector.
To bump test using the MicroDock II station, refer to the MicroDock II User Manual.
2. Connect the calibration hose to the intake inlet on the
calibration cap. Arrows on the calibration cap indicate the direction of gas flow.
3. Attach and tighten the calibration cap onto the
detector and apply gas. Verify the visual, audible, and vibrator alarms activate.
4. Close the regulator and remove the calibration cap.
The detector temporarily remains in alarm until the gas clears from the sensors.
a
Caution
Page 12
GasAlertQuattro
Alarms
Alarms
Refer to the following table for information about alarms and corresponding screens. For more information about alarms refer to the GasAlertQuattro Technical Reference Guide.
Alarm Screen Alarm Screen
Low Alarm
• Slow siren (upward tone)
• Slow flash
• Black box around gas flashes
• Vibrator alarm activates
High Alarm
• Fast siren (downward tone)
• Fast flash
• Black box around gas flashes
• Vibrator alarm activates
TWA Alarm
• Fast siren (downward tone)
• Fast flash
• Black box around gas flashes
• Vibrator alarm activates
STEL Alarm
• Fast siren (downward tone)
• Fast flash
• Black box around gas flashes
• Vibrator alarm activates
11
Page 13
GasAlertQuattro
Operator’s Manual
Alarm Screen Alarm Screen
Multi Alarm
• Alternating low and high alarm siren and flash
• Black box around gas flashes
• Type of alarm alternates
• Vibrator alarm activates
Over Limit (OL) Alarm
• Fast siren (downward tone)
• Fast flash
• Black box around gas flashes
• Vibrator alarm activates
Note: LCD may also display an under limit reading (-OL)
Sensor Alarm
• displays
If enabled, during an alarm condition the Latching Alarms option causes the low and high gas alarms (audible, visual, and vibrator) to persist until the alarm is acknowledged by pressing C and the gas concentration is below the low alarm setpoint. The peak concentration values display continually until the alarm no longer exists. Enable/disable Latching Alarms in Fleet Manager II. Local regulations may require the Latching Alarms option be enabled.
Normal Deactivation
• Sequence of alternating beeps and alternating flashes
• Vibrator alarm activates
• Countdown initiates
OFF displays
Note
12
Page 14
Alarm Screen Alarm Screen
Low Battery Alarm
• Sequence of 10 rapid sirens and alternating flashes with 7 seconds of silence in between (continues for 15 minutes)
flashes
• Vibrator alarm pulses
• After 15 minutes of the low battery alarm sequence, the detector enters critical battery alarm (see Critical Battery Alarm below)
Critical Battery Alarm
• Fifteen minutes after low battery alarm activates, sequence of 10 rapid sirens and alternating flashes with 1 second of silence in between (sequence reactivates seven times)
• Vibrator alarm pulses
Low Battery Powering Off dis­plays and the detector deactivates
GasAlertQuattro
Alarms
Confidence Beep
• One beep every 1-120 seconds (beep frequency is defined with Confidence Beep Interval option)
Confidence Flash
(default: one flash every 1 second)
• One flash every 1-120 seconds (flash frequency is defined with
Confidence Flash Interval option)
Heartbeat
pulses once every second to verify detector is operating correctly
Note
If the Low Alarm Acknowledge option is enabled, the audible alarm can be disabled during a low alarm condition. The LED and visual alarm indicators remain active until the alarm condition changes or the detector deactivates. Press C to acknowledge the low alarm and deactivate the audible alarm. If the alarm escalates to a high, TWA, or STEL alarm, the audible alarm reactivates.
Note: Confidence beep and confidence flash automatically disables during a low battery alarm.
13
Page 15
GasAlertQuattro
Operator’s Manual
User Options and Sensor Configuration
The following items are required to modify the user options and sensor configuration:
• Detector
• IR Link adapter or MicroDock II
• Fleet Manager II software
Refer to the GasAlertQuattro Technical Reference Guide and Fleet Manager II Operator s Manual for complete information.
Maintenance
To maintain the detector in good operating condition, perform the following basic maintenance as required.
• Calibrate, bump test, and inspect the detector on a regular schedule.
• Maintain an operations log of all maintenance, bump tests, calibrations, and alarm events.
• Clean the exterior with a soft damp cloth. Do not use solvents, soaps, or polishes. Refer to Sensor Poisons and
Contaminants.
14
Battery Pack Retaining Screw
The retaining screw (QAQD-20x) provided with the detector must be used to lock the battery pack on all European and
ECEx scheme detectors, and on all Canadian and U.S. Zone
Certified detectors.
A hex tool is required to tighten and loosen the retaining screw. Tighten the screw 1-2 turns using 3-4 in-lbs of torque. Do not overtighten the screw.
Page 16
Charging the Lithium Battery
a Warni ng
To avoid personal injury and/or damage to the detector, adhere to following
Charge only in a safe area that is free of hazardous gas within temperatures of 32°F to 104°F (0°C to 40°C).
Charge the battery immediately when the detector emits a low battery alarm.
Charge the lithium battery pack using the BW supplied charger and charger adapter only. The charging adapter is specific to your region. Use of the charging adapter outside your region will damage the charger and the detector. Failure to adhere to this caution can lead to fire and/or explosion.
Charge the lithium battery after each workday.
Change the lithium battery pack only in safe area that is free of hazardous gas.
1. Press and hold C to deactivate the detector, then
plug the charger into an AC outlet.
Note
The time required to charge will increase if the detector is activated.
GasAlertQuattro
Maintenance
2. Connect the charging adapter to the detector IR receptacle. Refer to the following illustration.
3. The lithium battery may require 6 hours to reach full capacity.
Replacing the Lithium Battery
1. Press and hold C to deactivate the detector.
2. Loosen the retaining screw if required. Push the bat­tery release latch toward the top of the detector to release the battery pack.
3. From the top of the battery pack, lift upward to remove.
4. Insert a fully charged battery pack. Insert the bottom of the battery pack first, then lower the top into place. Press until the release tab engages. Tighten the retaining screw if required.
15
Page 17
GasAlertQuattro
Operator’s Manual
Replacing the Alkaline Batteries
a Warni ng
To avoid personal injury and/or damage to the detector, use only BW recommended alkaline batteries. Refer to
Specifications
Change the alkaline batteries only in safe area that is free of hazardous gas.
1. Press and hold C to deactivate the detector.
2. Loosen the retaining screw if required. Remove the
.
16
alkaline battery pack. Refer to steps #2
ing the Lithium Battery.
3. Unhook the ejector bar from the release clasp. Pull the tab on the ejector bar up and then move the ejec­tor bar left and right to release the batteries. Remove the spent batteries.
4. Push the ejector bar down until the release clasp engages.
5. Insert the new batteries. Position the positive end of the battery at a 30° angle and insert into the battery pack before pushing the negative end down. Ensure the batteries are not inserted over the tab.
6. Replace the battery pack by inserting the bottom first, then lower top into place. Ensure the tab is tucked in before replacing the battery pack.
sPress until the release tab engages. Tighten the retaining screw if required.
-3 in Replac-
Page 18
Replacing the Sensors
GasAlertQuattro
Maintenance
17
Page 19
GasAlertQuattro
Operator’s Manual
a Warni ng
To avoid personal injury and/or property damage, only use sensors that are specifically designed for the detector.
Detectors that are configured for 1, 2, or 3 gases may contain a dummy sensor in one of the four sensor locations.
To replace a sensor or sensor filter, refer to the illustration
Replacing the Sensors
1. Press and hold C to deactivate the detector. Press
the release latch, and remove the battery pack.
2. Remove the six machine screws from the rear shell.
3. Remove the front shell.
4. Remove the spent sensor(s). Ensure no damage occurs to the LCD.
5. Insert the new sensor(s).
6. Reassemble the detector. Press the front and rear shells together firmly to ensure a proper seal. Ensure the front and rear shells have a tight, uniform 1/16 in. (1.5 mm) seal on all sides of the detector.
7. Replace the six machine screws using 3-4 in. lbs torque. Do not overtighten the screws. Replace the battery pack.
8. New sensors must be calibrated prior to use. Calibrate the new sensor(s) immediately. Refer to
Calibration
Note
and the following steps #1-8.
.
18
Replacing the Sensor Filter
To replace the filter, refer to the illustration Replacing the Sen-
sors and the following steps #1-6.
1. Press and hold C to deactivate the detector. Press
the release latch, and remove the battery pack.
2. Remove the six machine screws from the rear shell.
3. Remove the front shell. Remove the sensor filter.
4. Refer to the following illustration before inserting the new filter. Ensure the filter is laying flat and that the holes are correctly aligned over the filter posts.
5. Replace the front shell. Press the front and rear shells together firmly to ensure a proper seal. Ensure the front and rear shells have a tight, uniform 1/16 in. (1 mm) seal on all sides of the detector.
6. Replace the six machine screws using 3-4 in. lbs torque. Do not overtighten the screws. Replace the battery pack.
Page 20
Specifications
Instrument dimensions 12.9 x 8.0 x 3 8 cm (5.1 x 3.2 x 1.5 in.)
Weight 317 g (11.2 oz.)
Operating temperature -20°C to +50°C (-4°F to +122°F)
Storage temperature -40°C to +60°C (-40°F to +140°F)
Operating humidity 10% to 100% relative humidity
(non-condensing)
Dust and moisture ingress IP66/67 (with screw engaged)
Alarm setpoints May vary by region and are user-defined
Detection range
H
S: 0 - 200 ppm (1 / 0.1 ppm increments)
2
CO: 0 - 1000 ppm (1 ppm increments) O
: 0 - 30.0% vol. (0.1% vol. increments)
2
Combustible (LEL): 0 - 100% (1% LEL increments) or 0 - 5.0% v/v methane
Sensor type
H
S, CO, O2: Single plug-in electrochemical cell
2
Combustibles: Plug-in catalytic bead
O
measuring principle Capillary controlled concentration
2
sensor
Bump test specified limits BW recommends using a gas cyl­inder that will ensure the combustible sensor has an accuracy of -0 to +20% of actual reading
Alarm conditions TWA alarm, STEL alarm, low alarm, high alarm, multi alarm, over limit (OL) alarm, low battery alarm, crit­ical low battery alarm, confidence flash, confidence beep
Audible alarm 95 dB at 30 cm (12 in.) variable pulsed beeper
GasAlertQuattro
Specifications
Visual alarm Red light-emitting diodes (LEDs)
Confidence flash Green light-emitting diode. Flash
frequency is user-defined with confidence flash interval option
Confidence beep Audible beep from variable pulsed beeper. Beep frequency is user-defined with confidence beep interval option
Display Alphanumeric liquid crystal display (LCD) with flip display (0° or 180°) capability (user-defined in Fleet Manager II)
Backlight Activates upon startup and deactivates when self­test is complete. Activates when the pushbutton is pressed and deactivates after 10 seconds. Also activates during an alarm condition and remains lit until alarm ceases
Internal vibrator Vibrates during activation, deactivation, and all alarms
Self-test Initiated during activation, self-test runs continuously while detector is operational
Calibration Zero and automatic span
User options Startup message, lockout on self-test error, safe
mode, confidence flash, confidence beep, latching alarms, force calibration, force bump, calibration IR lock, flip display stealth mode, datalog interval, confidence flash interval, confi­dence beep interval, and language selection
Sensor options Sensor enable/disable, calibration gas values, calibration interval, bump test interval, alarm setpoints (low/high/TWA/STEL), STEL interval, TWA period, auto zero at startup enable/disable, LEL correction factor, 10% (of reading) over-span, low alarm acknowledge, O measurement, %vol methane measurement
Year of manufacture The detector's year of manufacture is determined from the serial number. The second and third num-
measurement, LEL gas
2
19
Page 21
GasAlertQuattro
Operator’s Manual
ber after the second letter determines the year of manufacture. E g., QA109
Approved lithium battery for GasAlertQuattro product Lithium-ion polymer (QT-BAT-R01) as per standards UL913, EN60079-11, EN60079-0, C22.2 No. 157
Rechargeable battery (QT-BAT-R01) Temperature code
Lithium polymer -20°C Ta +50°C T4
Lithium battery operating time One rechargeable lithium polymer battery provides the following operating runtimes:
20 hours at 20°C (68°F) 18 hours at -20°C (-4°F)
Approved alkaline battery pack for GasAlertQuattro (QT-B AT-A01) as per standards UL913, EN60079-11,
EN60079-0, C22 2 No. 157
-001000 = 2009 year of manufacture
Approved alkaline batteries for GasAlertQuattro product
Duracell MN1500 -20°C Ta +50°C T4 (129 9°C) Energizer E91VP -20°C Ta +50°C T3C (135 3°C)
AA Alkaline battery operating time
14 hours at 20°C (68°F)
Battery charger Charging adapter
First-time charge 6 hours
Normal charge 6 hours
Warranty 2 years including sensors
20
Approvals
Approved by CSA to both U.S. and Canadian Standards CAN/CSA C22.2 No. 157 and C22.2 152 ANS/UL - 913 and ANSI/ISA - S12.13 Part 1
CSA Class I, Division 1, Group A, B, C, D ATE X CE 0539 g II 1 G Ga Ex ia IIC T4 for Zone 0
IECEx Ga Ex ia IIC T4 IECEx CSA 09 0006
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules and ICES-003 Canadian EMI requirements. These limits are designed to provide reasonable protection against harmful interfer­ence in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that inter­ference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Group IIC KEMA 09 ATEX 0137 EN 60079-0, EN 60079-11, EN 60079-26
IEC 60079-0, IEC 60079-11, IEC 60079-26
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for
help.
Loading...