To ensure personal safety, read Safety Information - Read
First and the Cautions before using the detector.
The GasAlertMicroClip XT, XL and X3 (“the detector”) warns of
hazardous gas at levels above user-defined alarm setpoints.
The detector is a personal safety device. It is your responsibility to
respond properly to the alarm.
lists the gases monitored.
Table 1.
Zeroing the Sensors
To zero the sensors, refer to steps #1-3 in Connecting to the IR Link.
Table 1. Gases Monitored
Gas DetectedUnit of Measure
Hydrogen sulfide (H2S)parts per million (ppm)
Carbon monoxide (CO)parts per million (ppm)
Oxygen (O
Combustible gases (LEL)
Field selectable for:
)percent by volume (%)
2
1. percent of lower explosive
limit (% LEL)
2. percent by volume met
0-
5.0% v/v
hane
1
GasAlertMicroClip
User Manual
Contacting BW Technologies by Honeywell
To contact BW Technologies by Honeywell, call
USA: 1-888-749-8878
Canada: 1-800-663-4164
Europe: 00800-333-22244
Other countries: +1-403-248-9226
Address correspondence to
BW Technologies by Honeywell
Suite 110 4411-6 Street SE
Calgary Alberta
Canada, T2G 4E8
Email: info@gasmonitors.com
BW Technologies by Honeywell's website: www.honeywellanalytics
ISO 9001
.com
Safety Information - Read First
Use the detector only as specified in this guide and the operator’s
manual, otherwise the protection provided by the detector may be
impaired.
International symbols on the detector and in this manual are explained in
Table 2.
Read the Cautions on the following pages before using the detector.
ec Warning
This instrument contains a lithium polymer battery. Dispose of
lithium cells immediately. Do not disassemble and do not
dispose of in fire. Do not mix with the solid waste stream.
Spent batteries must be disposed of by a qualified recycler or
hazardous materials handler.
a Cautions
• Warning: Substitution of components may impair Intrinsic Safety.
• Before using the detector, refer to Sensor Poisons
Contaminants.
• Warning: For safety reasons, this equipment must be
a
nd serviced by qualified personnel only. Read and understan
the user
• Do not use the detector if it is damaged. Inspect the detector
before using. Look for cracks and/or missing parts.
• If the detector is damaged or parts are missing, contact BW
Technologies by Honeywell immediately.
• Only use sensor(s) that are specifically designed for t
Gas
manual completely before operating or servicing.
AlertMicroClip. Refer to Replacement Parts and Accessories
and
operated
d
he
.
2
GasAlertMicroClip
Safety Information - Read First
• Calibrate the detector before first-time use and then on a regular
schedule
and contaminants. BW recommends calibrating at least
every
• BW recommends to “bump test” the sensors before each day’
se to confirm their ability to respond to gas by exposing the
u
detector to a gas concentration that exceeds the alarm setpoints.
Manually verify that the audible and visual alarms are activated.
Calibrate if the readings are not within the specified limits.
• BW recommends the combustible sensor be checked with a
known concentration of calibration gas after any known
exposure to catalyst contaminants/poisons (sulfur compound
si
• The combustible sensor is factory calibrated to 50% LEL
methane. If monitoring a different combustible gas in the % LEL
range, calibrate the sensor using the appropriate gas.
• Caution: High off-scale readings may indicate an explosi
concentra
• Only the combustible gas detection portion of this instrument
has been assessed for performance by CSA International.
• Protect the combustible sensor from exposure to l
ompounds, silicones, and chlorinated hydrocarbons.
c
• Sensor exposure to certain organic vapors (such as leaded
gasoline and halogenated hydrocarbons) may temporarily inhibit
sensor performance. After exposure, a bump test or calibration is
recommended.
• For use only in potentially explosive atmospheres where oxygen
concentrations do not exceed 20.9% (v/v).
, depending on use and sensor exposure to poisons
once
180 days (6 months).
s
s,
licon vapors, halogenated compounds, etc).
ve
tion.
ead
• Any rapid up-scaling reading followed by a declining or errati
ading may indicate a gas concentration beyond upper scal
re
lim
it, which may be hazardous.
• Only calibrate the detector in a fresh air environment and in
s
afe area.
• Use only BW approved batteries for the GasAlertMicroClip
detector. Refer to Specifications
• Charge the detector before first-time use. BW recommends th
detector be
• Charge the detector using the recommended charging adap
on
ly. Do not use any other charging adapter. Failure to adhere to
this caution can lead to fire and/or explosion.
• Extended exposure of the detector to certain concentrations of
combustible gases and air may stress a detector ele
ca
n seriously affect its performance. If an alarm occurs due to
high concentration of combustible gases, calibrate the detector
f necessary, replace the sensor.
I
• Do not test the combustible sensor’s response with a butan
ci
garette lighter; doing so will damage the sensor.
• Do not expose the detector to electrical shock and/or se
ontinuous mechanical shock.
c
• Do not attempt to disassemble, adjust, or service the detector
unless instructions for that procedure are provided in
t
echnical reference guide, and/or that part is listed as
placement part. Use only BW Technologies by Honeywell
re
replacement parts. Refer to Replacement Parts and Accessories
• The detector warranty will be voided if customers, personnel, or
third parties damage the detector during repair attemp
No
n-BW Technologies by Honeywell repair/service attempts voi
his warranty.
t
charged after every workday.
.
ment that
a
c
e
a
e
ter
e
vere
the
ts.
.
.
d
3
GasAlertMicroClip
User Manual
Table 2. International Symbols
SymbolsDescription
n
g
X
ATEX
IECEx
Approved to both U.S. and Canadian Standards by CSA International
European Explosive Protection
Conforms to European Union Directives
Conforms to European ATEX Directives
International Electrotechnical Commission Scheme for Certification to Standards for Electrical Equipment for Explosive
Atmospheres
Conforms to Korea Testing Laboratory (KTL) Certification
Natural Institute of Metrology, Quality, and Technology. Conforms to Brazilian INMETRO Certification.
Aust
ralian Regulatory Compliance Mark
4
GasAlertMicroClip
Sensor Poisons and Contaminants
Sensor Poisons and Contaminants
Several cleaners, solvents, and lubricants can contaminate and cause permanent damage to sensors. Before using cleaners, solvents, and lubricants
in close proximity to the detector sensors, read the following cautions and refer to the lists below.
a Caution
Use only the following BW Technologies by Honeywell recommended products and procedures:
• Use water based cleaners.
• Use non-alcohol based cleaners.
• Clean the exterior of the detector with a soft, damp cloth.
• Do not use soaps, polishes, or solvents.
Below are common products to avoid using around sensors.
Cleaners and Lubricants
• Brake cleaners
• Lubricants
• Rust inhibitors
• Window and glass cleaners
• Dishsoaps
• Citrus based cleaners
• Alcohol based cleaners
• Hand sanitizers
• Anionic detergents
• Methanol (fuels and antifreezes)
5
GasAlertMicroClip
User Manual
Silicones
• Silicone cleaners and protectants
• Silicone based adhesives, sealants, and gels
• Hand/body and medicinal creams containing silicone
• Tissues containing silicone
• Mold releasing agents
• Polishes
Aerosols
• Bug repellents and sprays
• Lubricants
• Rust inhibitors
• Window cleaners
Getting Started
The list below provides the standard items included with the detector. If
the detector is damaged or parts are missing, contact the place of
purchase immediately.
• Sensors: H
• Calibration cap and hose
• Charging adapter
• Printed Operator‘s Manual
• Supplementary Booklet, including a Quick Reference Card
• CD-ROM, including translated operator‘s manuals
Configuration Software: The detector is configured with Fleet Manager
II software. It can be downloaded for free from BW Technologies by
Honeywell website: www.gasmonitors.com
6
S, CO, O2, and combustible (LEL)
2
.
Fleet Manager II CD-ROM is shipped with the MicroDock II base station
and IR Link kit.
The detector is shipped with the sensors and rechargeable battery
installed.
Battery Replacement: To replace the battery, contact BW Technologies
by Honeywell. The battery can only be replaced by the manufacturer.
Charge Battery and Replace Sensors: To charge the battery and
replace the sensors and/or sensor filter, refer to the following:
•
Battery Cautions
• Replacing a Sensor or Sensor Filter
To order replacement parts, refer to Replacement Parts and
Accessories.
To become oriented with the features and functions of the detector, refer
to the following figures and tables:
• Figure 1.
• Figure 2.
•Table 5.
and Table 3. describes the detector’s components.
and Table 4. describes the detector’s display elements.
describes the detector’s pushbutton.
GasAlertMicroClip
Parts of the GasAlertMicroClip
Parts of the GasAlertMicroClip
Figure 1. Parts of the GasAlertMicroClip
Table 3. Parts of the GasAlertMicroClip
ItemDescription
1IntelliFlash
2Visual alarm indicators (LEDs)
3Alligator clip
4Charging connector / IR interface
5Pushbutton (C)
6Carbon monoxide (CO) sensor
7
8
9Combustible (LEL) sensor
10Audible alarm
11Liquid crystal display (LCD)
Hydrogen sulfide (H
Oxygen (O
) sensor
2
S) sensor
2
7
GasAlertMicroClip
1
8
7
6
5
4
3
2
Display Elements
Display Elements
Figure 2. Display Elements
Table 4. Display Elements
ItemDescription
1Alarm condition
2Automatically zero sensor
3Numeric value
4Stealth mode
5Battery life indicator
6Gas identifier bars
7Gas cylinder
8Automatically span sensor
8
Pushbutton
GasAlertMicroClip
Pushbutton
Table 5. Pushbutton
PushbuttonDescription
• To activate the detector press C.
• To deactivate the detector, press and hold C until the OFF countdown is complete and the LCD
deactivates.
• To view the TWA, STEL, and peak (maximum) readings, press C twice. To clear the TWA, STEL,
and peak readings, press C when the LCD displays RESET.
C
• To initiate calibration, deactivate the detector. Press and hold C while the detector performs the
OFF countdown. Continue holding C while the LCD briefly deactivates. The LCD reactivates and
then begins the CAL countdown. Release C when the CAL countdown is complete.
• To activate the backlight in normal operation, press C.
• To acknowledge latched alarms, press C.
• To acknowledge a low alarm and disable the audible alarm, press C (if the Low Alarm Acknowledge option is enabled).
9
GasAlertMicroClip
User Manual
Activating the Detector
a Caution
Only activate the detector in a fresh air environment and in a
safe area.
To activate the detector, press
Self-Test
The following startup tests are written as startup performance is
intended. If an error occurs, refer to Startup Troubleshooting
When the detector is activated, it performs several startup tests. Confirm
the following tests occur.
Battery Test
The detector performs a battery test during startup. If the battery has
insufficient power to operate, the following screen displays.
Charge the battery for 2-3 hours before restarting the detector. Refer to
Charging the Battery
.
C.
.
Audible/Visual Test
1. All of the LCD elements display simultaneously as the detector
beeps, flashes, vibrates, and activates the backlight.
Detector Version
2. The current firmware version of the detector then displays on the
LCD.
Startup Message
3. If data is entered in the Startup Message option (25 characters
maximum) of Fleet Manager II, that data will display during the
startup self-test. To enter a startup message, refer to Detector
Identification or the Fleet Manager II Operator’s Manual.
10
GasAlertMicroClip
Activating the Detector
Alarm Setpoints
4. Next, the TWA, STEL, low, and high alarm setpoints display.
Note
Alarm setpoints may vary by region. Refer to Factory Gas Alarm
Setpoints.
Sensor and Power Test
5. The detector then tests the sensors.
After testing the sensors, the following screen displays to verify
all sensors have passed.
If an error message displays, refer to Startup Troubleshooting
Note
The sensors are tested continuously while the detector is activated.
.
11
GasAlertMicroClip
User Manual
Automatic Zero and O2 Calibration (optional)
6. Auto-Zero on Startup: If enabled, the H
S, CO, and LEL
2
sensors are automatically zeroed during startup. Each sensor is
enabled individually. The auto-zero option is enabled for each
sensor upon shipment.
O
Auto-Calibration on Startup: If enabled, the O2 sensor is
2
automatically calibrated during startup. The auto-calibration
option is enabled for the O
If the O
Auto-Calibration on Startup option is enabled, and the
2
sensor upon shipment.
2
Auto-Zero on Startup option is disabled for all sensors, the
following screen displays.
Note
If oxygen is configured to measure 20.8% vol., the oxygen
calibration screen displays 20.8% O
.
2
Calibration Due Date (optional)
7. The following screen displays the number of days remaining
before calibration is due. The number of days that displays is
when the next sensor calibration should be performed.
Note
If the Calibration Interval option is defined as 0, the calibration
due date is bypassed during startup.
12
GasAlertMicroClip
Activating the Detector
Last Calibration Failed (optional)
If any sensor failed the last calibration, CAL FAILURE displays
on the screen.
Note
When the CAL FAILURE displays, the previous calibration has
failed but the calibration is still valid until the next calibration due
date.
Overdue Calibration
If any sensor is past due for calibration, the detector beeps,
flashes, vibrates, and the following screen displays.
If calibration is overdue and the Force Calibration When Overdue option is enabled, a calibration must be performed to
enter normal operation. Refer to Calibration
.
Note
If calibration is not performed, or
C is not pressed within
2 minutes, the detector automatically deactivates.
If the Force Calibration When Overdue is disabled, press
acknowledge the warning. The detector continues with the
startup self-tests and then enters normal operation.
Cal IR Lock
If the Cal Lock option is enabled, the following screen displays.
Refer to Startup Troubleshooting
.
Bump Test
Note
A bump test cannot be conducted if the detector has just been
calibrated. If the Bump Interval
option is defined as 0 in Fleet
Manager II, the bump test is bypassed.
BW recommends to “bump test” the sensors, before each day’s
use, to confirm their ability to respond to gas by exposing the
detector to a gas concentration that exceeds the ala rm set poi nts.
C to
13
GasAlertMicroClip
User Manual
Last Bump Test Failed
If any sensor failed the last calibration, CAL FAILURE displays
on the screen.
Note
When the BUMPCHK FAILURE displays, the previous
calibration has failed but the calibration is still valid until the next
bump test due date.
Force Bump (optional)
8. If the Force Bump option is enabled and the sensors are due
have a bump test, the following screen displays.
A bump test must be performed to enter normal operation. Apply
gas to the sensors. Ensure the visual, audible, and vibrator
alarms activate. When the gas is removed, the detector briefly
remains in alarm until the gas has cleared from the sensors.
When the sensors successfully pass the bump check, the
following screen displays showing the number of days remaining
until the next bump check is due (I d = 1 day).
If Force Bump is disabled, press
self-tests.
If BUMPCHK todAY displays again after performing a bump
check, refer to Startup Troubleshooting
C to continue with the startup
Note
.
14
GasAlertMicroClip
Activating the Detector
Self-Test Pass
When the detector has passed all startup self-tests, it enters normal
operation. The LCD displays the ambient gas readings.
The detector automatically begins
• recording the peak (maximum) gas exposure,
• calculating the short-term exposure level (STEL), and
• calculating the time-weighted average (TWA) exposures.
Self-Test Fail
If the following error message displays after entering normal operation,
refer to Startup Troubleshooting
.
Battery Test
The battery is tested when the detector is activated and continuously
thereafter. See Specifications for Battery Runtimes.
• Battery power is continually displayed during normal operation. If
battery power is low,
sequence of 10 rapid sirens and alternating flashes with 7 seconds
of silence in between (continues for 15 minutes).
• If battery power becomes critically low,
The detector performs a sequence of 10 rapid sirens with 1 second
of silence in between (sequence reactivates seven times). The
detector then displays
If enabled, Confidence Beep and IntelliFlash automatically
deactivate during a low battery alarm. Refer to Confidence Beep
L and flashes. The detector performs a
L and LOW BAT display.
OFF and the detector deactivates.
Note
Backlight
The backlight automatically activates
• during startup,
• when the pushbutton is pressed (then deactivates after 5 seconds),
and
• when there is an alarm condition (unless Stealth is enabled).
15
GasAlertMicroClip
User Manual
Deactivating the Detector
To deactivate the detector, press and hold C. The detector
• performs a sequence of two sirens with alternating flashes,
•vibrates,
• initiates the deactivation countdown, and
•displays
If
will not deactivate.
OFF.
Note
C is released before the countdown is complete, the detector
Installing Fleet Manager II
Fleet Manager II is required to configure the detector and sensors. And
IR Link is also required. To purchase contact BW Technologies by
Honeywell.
To install Fleet Manager II complete the following:
1. Install Fleet Manager II using the Fleet Manager CD-ROM
(available with MicroDock II and the IR Link), or download (at no
cost) from BW Technologies by Honeywell website:
www.gasmonitors.com
2. Follow the installation wizard.
3. When installation is complete, open Fleet Manager II.
4. Click Administration located on the left toolbar.
5. Click the Login/Logout button.
.
16
6. When the Password dialog box displays, enter Admin (password
is case sensitive).
GasAlertMicroClip
Installing Fleet Manager II
7. Click OK.
8. From the Devices toolbar, click Configure Device via IR Link.
9. When the Device Selection dialog box displays, select
GasAlertMicroClip Series and click OK.
Fleet Manager II displays the Sensors tab that includes the
following sections:
•
Detector Identification
•CO, O2, H2S, LEL Sensor Configuration
• User Options
• Language Menu
17
Loading...
+ 47 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.