BW Technologies GasAlertMicroClip Series User Manual

Page 1
User Manual
GasAlertMicroClip
1, 2, 3, and 4 Gas Detectors
H2S, CO, O2, Combustibles
Page 2
BW Technologies LP (BW) warrants this product to be free from defects in material and workmanship under normal use and service for a period of two years, beginning on the date of shipment to the buyer. This warranty extends only to the sale of new and unused products to the original buyer. BW’s warranty obligation is limited, at BW’s option, to refund of the purchase price, repair, or replacement of a defective product that is returned to a BW authorized service center within the warranty period. In no event shall BW’s liability hereunder exceed the purchase price actually paid by the buyer for the Product. This warranty does not include:
a) fuses, disposable batteries or the routine replacement of parts due to the normal wear and tear of the product arising from use; b) any product which in BW’s opinion, has been misused, altered, neglected or damaged by accident or abnormal conditions of operation, handling or use; c) any damage or defects attributable to repair of the product by any person other than an authorized dealer, or the installation of unapproved parts on the
product; or
The obligations set forth in this warranty are conditional on:
a) proper storage, installation, calibration, use, maintenance and compliance with the product manual instructions and any other applicable
recommendations of BW;
b) the buyer promptly notifying BW of any defect and, if required, promptly making the product available for correction. No goods shall be returned to BW
until receipt by the buyer of shipping instructions from BW; and
c) the right of BW to require that the buyer provide proof of purchase such as the original invoice, bill of sale or packing slip to establish that the product is
within the warranty period.
THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. BW SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT OR RELIANCE OR ANY OTHER THEORY. Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
BW Technologies by Honeywell BW Technologies by Honeywell BW Technologies by Honeywell Corporate Headquarters America Europe 2840 – 2nd Ave. SE 3279 West Pioneer Parkway 5 Canada Close Calgary, AB Arlington, TX Banbury, Oxfordshire Canada T2A 7X9 USA 76013 UK OX16 2RT
Page 3
Table of Contents
Title Page
Introduction............................................................................................................................................1
Contacting BW Technologies by Honeywell.........................................................................................2
Safety Information - Read First.............................................................................................................2
Getting Started......................................................................................................................................6
Activating the Detector........................................................................................................................10
Self-Test......................................................................................................................................10
Self-Test Pass.............................................................................................................................15
Self-Test Fail ...............................................................................................................................15
Battery Test.................................................................................................................................15
Backlight......................................................................................................................................16
Deactivating the Detector....................................................................................................................16
Options................................................................................................................................................16
Detector Identification..................................................................................................................17
Sensor Options............................................................................................................................17
User Options Menu ..................................................................................................................... 20
Language Menu...........................................................................................................................21
Alarms.................................................................................................................................................22
Computed Gas Exposures..........................................................................................................24
Viewing Gas Exposures..............................................................................................................25
Gas Alarm Setpoints ................................................................................................................... 26
Resetting Gas Alarm Setpoints...................................................................................................26
Stopping a Gas Alarm.................................................................................................................27
Sensor Alarm...............................................................................................................................27
i
Page 4
GasAlertMicroClip
User Manual
ii
Title Page
Low Battery Alarm .......................................................................................................................27
Automatic Shutdown Alarm.........................................................................................................27
Calibration...........................................................................................................................................28
Guidelines....................................................................................................................................28
Diagnostics Protection.................................................................................................................29
Applying Gas to the Sensors.......................................................................................................29
Calibration Procedure..................................................................................................................30
Datalogs ..............................................................................................................................................34
Event Logs ..........................................................................................................................................34
Downloading Datalogs and Event Logs......................................................................................34
Maintenance........................................................................................................................................34
Charging the Detector .................................................................................................................35
Replacing a Sensor or Sensor Filter ...........................................................................................36
Troubleshooting...................................................................................................................................39
Replacement Parts and Accessories..................................................................................................42
Specifications......................................................................................................................................43
General Datalogger Specifications..............................................................................................45
Page 5
List of Tables
Table Title Page
1. Gases Monitored ................................................................................................................... 1
2. International Symbols............................................................................................................5
3. GasAlertMicroClip Detector...................................................................................................7
4. Display Elements...................................................................................................................8
5. Pushbutton.............................................................................................................................9
6. Alarms..................................................................................................................................22
7. Computed Gas Exposures .................................................................................................. 24
8. Gas Alarm Setpoints............................................................................................................26
9. Sample Factory Alarm Setpoints.........................................................................................26
10. Applying Gas to the Sensors...............................................................................................29
11. Replacing a Sensor or Sensor Filter ................................................................................... 37
12. Troubleshooting Tips...........................................................................................................39
13. Replacement Parts and Accessories .................................................................................. 42
iii
Page 6
GasAlertMicroClip
User Manual
iv
Page 7
List of Figures
Figure Title Page
1. GasAlertMicroClip Detector...................................................................................................7
2. Display Elements...................................................................................................................8
3. Applying Gas to the Sensors...............................................................................................29
4. Connecting the Charging Adapter.......................................................................................35
5. Replacing a Sensor or Sensor Filter ................................................................................... 36
6. Replacing the Rear Shell.....................................................................................................38
v
Page 8
GasAlertMicroClip
User Manual
vi
CAUTION: FOR SAFETY REASONS, THIS EQUIPMENT MUST BE OPERATED AND SERVICED BY QUALIFIED PERSONNEL ONLY. READ AND UNDERSTAND INSTRUCTION MANUAL COMPLETELY BEFORE OPERATING OR SERVICING.
GasAlertMicroClip Multi-Gas Detector with User Downloadable Datalogger and Event Log
Standard instrument is equipped with integral concussion­proof boot, internal vibrator alarm, and provides full-time continuous datalogging while the instrument is operating. Data is saved to memory and can be downloaded by the user via an IR device to a PC. Data is imported into standard office software (Microsoft® Excel, Access, etc.). Wraparound memory ensures the most recent data is always saved.
Page 9
GasAlertMicroClip
Introduction
a Warning
To ensure your personal safety, read “Safety Information” before you use the detector.
The GasAlertMicroClip gas detector (“the detector”) warns of hazardous gas at levels above user-selectable alarm setpoints.
The detector is a personal safety device. It is your responsibility to respond properly to the alarm.
Table 1 lists the gases monitored.
Table 1. Gases Monitored
Gas Detected Unit of Measure
Hydrogen sulfide (H2S) parts per million (ppm) Carbon monoxide (CO)
parts per million (ppm) Oxygen (O2) percent by volume (%) Combustible gases (LEL)
Field selectable for:
a) percent of lower
explosive limit (% LEL)
b) percent by volume
methane 0-5.0% v/v
1
Page 10
GasAlertMicroClip
User Manual
Contacting BW Technologies by Honeywell
To contact BW Technologies by Honeywell, call: USA: 1-888-749-8878
Canada: 1-800-663-4164 Europe: +44 (0) 1295 700300 Other countries: +1-403-248-9226
Address correspondence to:
BW Technologies by Honeywell Corporate office 2840 – 2 Avenue S.E. Calgary, AB T2A 7X9 CANADA
Email us at:
info@bwtnet.com
Visit BW Technologies by Honeywell’s web site at:
www.gasmonitors.com
ISO 9001
Safety Information - Read First
Use the detector only as specified in this manual, otherwise the protection provided by the detector may be impaired.
International symbols used on the detector and in this manual are explained in Table 2.
Read the
Cautions on the following pages before using the
detector.
ec Note
This instrument contains a lithium polymer battery. Do not mix with the solid waste stream. Spent batteries should be disposed of by a qualified recycler or hazardous materials handler.
2
Page 11
GasAlertMicroClip
Safety Information - Read First
aCautions
Warning: Substitution of components may impair Intrinsic Safety.
Caution: For safety reasons, this equipment must be operated and serviced by qualified personnel only. Read
and understand the user manual completely before operating or servicing.
Do not use the detector if it is damaged. Before you use the detector, inspect the case. Look for cracks or
missing parts.
If the detector is damaged or something is missing, contact BW Technologies by Honeywell
immediately.
Use only sensor(s) specifically designed for your GasAlertMicroClip model. (See the section, Replacement Parts
and Accessories).
Calibrate the detector before first-time use and then on a regular schedule, depending on use and sensor
exposure to poisons and contaminants. BW recommends at least once every 180 days (6 months).
BW recommends to “bump test” the sensors, before each day’s use, to confirm their ability to respond to gas by
exposing the detector to a gas concentration that exceeds the alarm setpoints. Manually verify that the audible and visual alarms are activated. Calibrate if the readings are not within the specified limits.
It is recommended that the combustible sensor be checked with a known concentration of calibration gas after
any known exposure to catalyst contaminants/poisons (sulfur compounds, silicon vapors, halogenated compounds, etc).
The combustible sensor is factory calibrated to 50% LEL methane. If monitoring a different combustible gas in
the % LEL range, calibrate the sensor using the appropriate gas.
Caution: High off-scale readings may indicate an explosive concentration.
Only the combustible gas detection portion of this instrument has been assessed for performance by CSA
International.
3
Page 12
GasAlertMicroClip
User Manual
aCautions
Protect the combustible sensor from exposure to lead compounds, silicones, and chlorinated hydrocarbons.
Although certain organic vapors (such as leaded gasoline and halogenated hydrocarbons) may temporarily inhibit sensor performance, in most cases, the sensor will recover after calibration.
For use only in potentially explosive atmospheres where oxygen concentrations do not exceed 20.9% (v/v).
Any rapid up-scaling reading followed by a declining or erratic reading may indicate a gas concentration beyond
upper scale limit, which may be hazardous.
Calibrate only in a safe area that is free of hazardous gas.
Use only BW approved batteries for your GasAlertMicroClip model (see
Specifications
).
Charge the detector before first-time use. BW recommends the detector be charged after every workday.
Charge the GasAlertMicroClip using the recommended charging adapter only. Do not use any other charging
adapter. Failure to observe this precaution could lead to fire or explosion.
Extended exposure of the GasAlertMicroClip to certain concentrations of combustible gases and air may stress a
detector element, which can seriously affect its performance. If an alarm occurs due to high concentration of combustible gases, recalibration should be performed, or if needed, the sensor replaced.
Do not test the combustible sensor’s response with a butane cigarette lighter; doing so will damage the sensor.
Do not expose the detector to electrical shock and/or severe continuous mechanical shock.
Do not attempt to disassemble, adjust, or service the detector unless instructions for that procedure are
contained in the manual and/or that part is listed as a replacement part. Use only BW Technologies by Honeywell
replacement parts
.
The detector warranty will be voided if customer, personnel, or third parties damage the detector during repair
attempts. Non-BW Technologies by Honeywell repair/service attempts void this warranty.
4
Page 13
GasAlertMicroClip
Safety Information - Read First
Table 2. International Symbols
Symbol Meaning
n
Approved to both U.S. and Canadian Standards by CSA - International
g
European Explosives Protection
X
Conforms to European Union Directives
ATEX
Conforms to European ATEX Directives
IECEx
International Electrotechnical Commission Scheme for Certification to Standards for Electrical Equipment for Explosive Atmospheres
5
Page 14
GasAlertMicroClip
User Manual
Getting Started
The list below provides the standard items included with your detector. If the detector is damaged or something is missing, contact the place of purchase immediately.
x Sensors: H
2
S, CO, O2, and combustible (LEL);
x Calibration hose and cap; x Charging adapter; x Quick reference guide; x Quick reference operations key; and x CD.
To order replacement parts, see the section Replacement
Parts and Accessories.
The detector comes with sensors and rechargeable battery installed. If the battery needs to be replaced, contact
BW Technologies by Honeywell. If any of the sensors or the
sensor filter needs to be replaced, refer to
Replacing a
Sensor or Sensor Filter.
To become familiar with the features and functions of the detector, study the following figures and tables:
x Figure 1 and Table 3 describes the detector’s
components.
x Figure 2 and Table 4 describes the detector’s display
elements.
x Table 5 describes the detector’s pushbutton.
6
Page 15
GasAlertMicroClip
Getting Started
Figure 1. Parts of the GasAlertMicroClip
Table 3. Parts of the GasAlertMicroClip
Item Description
1
Visual alarm bars (LED)
2
Liquid crystal display (LCD)
3
Audible alarm
4
Sensors
5
Pushbutton
6
Charging connector / IR interface
7
Alligator clip
7
Page 16
GasAlertMicroClip
User Manual
Table 4. Display Elements
Item Description
1
Alarm condition
2
Automatically span sensor
3
Gas cylinder
4
Gas identifier bars
5
Battery life indicator
6
Stealth mode
7
Numeric value
8
Automatically zero sensor
Figure 2. Display Elements
8
Page 17
GasAlertMicroClip
Getting Started
Table 5. Pushbutton
Pushbutton Description
C
x To activate the detector press C. x To deactivate the detector, press and hold C until the OFF countdown is complete and the LCD
deactivates.
x To view the TWA, STEL, and maximum (MAX) hold readings, press
C twice. To clear the TWA, STEL,
and MAX readings, press
C when the LCD displays RESET.
x To initiate calibration, press and hold C as the detector executes the OFF countdown and continue to hold
C as the LCD briefly turns off and then executes the CAL countdown. Release C once the CAL
countdown is complete.
x To activate the backlight, press
C.
x To acknowledge latched alarms, press C. x To acknowledge a low alarm and disable the beeper, press C (if Low Alarm Acknowledge user option is
enabled).
9
Page 18
GasAlertMicroClip
User Manual
Activating the Detector
To activate the detector, press C in a normal clean atmosphere.
Self-Test
Once the detector is activated, it performs the following checks. Manually verify that all actions occur.
Battery Test
The detector administers a battery test during start-up. If the battery has insufficient power to start up, the liquid crystal display (LCD) displays the following screen before the detector deactivates.
Recharge the battery for 2-3 hours before restarting the detector (see
Charging the Detector
).
Audible/Visual Test
1. The LCD shows all the display elements as it beeps, flashes, vibrates, and briefly turns on the backlight.
Detector Version
2. The version number of the detector is then displays on the LCD.
10
Page 19
GasAlertMicroClip
Activating the Detector
Note
Users can enter a 2-line message, up to a maximum of 25 characters per line. The message appears on the LCD after the detector version displays. This option is disabled in the factory settings. The message can be enabled or changed in the Sensors tab in Soft Tools. See
GasAlertMicroClip Soft Tools
instruction sheet.
Sensors and Power Test
4. The detector then tests the sensors.
Alarm Setpoints
3. Next, the LCD shows the TWA, STEL, low, and high alarm setpoints.
Note
The alarm setpoints on a shipped detector may vary by region. See
Resetting Gas Alarm Setpoints
.
A screen then appears confirming that the test was successful (TEST OK) or which sensor(s) failed the test (e.g., ERROR H2S) before continuing with the self-test.
11
Page 20
GasAlertMicroClip
User Manual
Note
The sensors are continuously tested while the detector is activated.
Automatic Zero and O2 Calibration (optional)
5. The sensors (H2S, CO, and LEL) are automatically zeroed and the O
2
sensor is automatically calibrated (unless it is disabled in the detector’s options). The detector beeps twice to signal a successful zero and span.
If only the automatic oxygen calibration is enabled on the detector, the following screen appears.
Note
If ambient air is set to be measured 20.8% vol., the automatic oxygen calibration screen displays 20.8 instead of 20.9.
Calibration Due Date (optional)
6. The LCD displays the number of days remaining until the next calibration is due.
Note
If the calibration interval
is set to 0, then the
calibration due date start-up feature is bypassed.
If any sensor is past its calibration due date, the detector displays CAL DUE todAY as the detector beeps, flashes, and vibrates.
12
Page 21
GasAlertMicroClip
Activating the Detector
Press C to acknowledge the warning and enter normal operation, or continue with the self-test if applicable.
Forced Calibration Enabled
If forced calibration is enabled (refer to Force Calibration
When Overdue) and a sensor is past its calibration due date,
calibration is mandatory before the detector can enter normal operation.
If C is pressed, the detector launches the calibration procedure (refer to
Calibration
). If C is not pressed within 2
minutes, the detector deactivates.
Note
Calibration can also be performed by using BW Technologies by Honeywell’s MicroDock II.
Cal Lock Enabled
If the Cal Lock option is enabled, an infrared (IR) device is required to calibrate the detector.
IR--LOCK displays as the detector tries to establish a connection with an IR device (e.g., BW Technologies by Honeywell’s MicroDock II or IR Link adapter).
Once IR communication has been established, the calibration procedure is executed. If IR communication is not established within 2 minutes or C is pressed, the detector deactivates.
13
Page 22
GasAlertMicroClip
User Manual
Bump Check (optional)
Note
Bump check is not performed if the detector was just calibrated (step #6).
BW recommends to “bump check” the sensors, before each day’s use, to confirm their ability to respond to gas by exposing the detector to a gas concentration that exceeds the alarm setpoints.
If the
bump interval
is set to 0, the bump check
start-up feature is bypassed.
7. If bump check is enabled and the bump checks have passed, or none of the sensors are due for a bump check, a warning screen displays the number of days remaining until a bump check needs to be performed.
If any of the sensors need to be bump checked, the detector displays the following screen:
If force bump or is not enabled, press C to enter normal operation.
Note
BUMPCHK todAY also displays if a bump check was executed but failed.
Force Bump Enabled
If force bump is enabled, (refer to Force Bump When
Overdue) and a sensor is past its bump interval, a bump
check is mandatory before the detector can enter normal operation.
14
Page 23
GasAlertMicroClip
Activating the Detector
Self-Test Fail
While the LCD displays BUMPCHK todAY, apply gas to the detector or press C to deactivate the detector. If gas is applied to the detector, the detector needs to enter alarm for the bump check to be accepted. Once the bump check has been accepted, the detector enters normal operation.
Once the detector enters normal operation, the LCD displays which sensor has failed the self-test by displaying Err above the target gas bar. (Refer to the
Troubleshooting section.)
Bump Check Successful
If the bump check is successful, the detector enters normal operation. The detector remains in alarm until the bump gas dissipates. Verify that the audible and visual alarms are functioning properly (refer to
Alarms
).
Self-Test Pass
Battery Test
If the detector passes the self-test, the detector begins normal operation. The LCD displays the ambient gas readings.
The battery is tested on activation and continuously thereafter. A newly charged battery should operate for 10-12 hours typically before automatically shutting down. The battery power is continually displayed during normal operation. If the battery power is low,
flashes as the detector emits one beep and one flash (LED) every 5 seconds.
Note
If the confidence beep is enabled, the audible alarm beeps if the battery has sufficient power and stops if the battery power is low. See the section,
Confidence Beep
.
The detector starts recording the maximum gas exposure (MAX) and calculating the short-term exposure level (STEL) and time-weighted average (TWA) exposures.
15
Page 24
GasAlertMicroClip
User Manual
Backlight
The LCD’s backlight automatically activates when there is an alarm condition. Pressing C activates the backlight in any condition.
Deactivating the Detector
To deactivate the detector, press and hold C while it beeps and flashes to the corresponding countdown.
At the end of the countdown, the detector emits one short beep before completely deactivating.
Note
If C is released before the countdown is complete, the detector will not deactivate.
Options
To access the user options, connect the detector to the IR Link adapter and use the GasAlertMicroClip Soft Tools software. Refer to the GasAlertMicroClip Soft Tools manual for complete instructions.
The following are the available user options in Soft Tools:
1. Detector Identification: Start-up message;
2. Carbon Monoxide (CO), Hydrogen Sulphide
(H2S), Oxygen (O2), and Combustible/LEL:
sensor disabled, calibration gas concentration, calibration interval, bump interval, low alarm, high alarm, TWA alarm, STEL alarm, STEL interval, auto-zero on start-up, O
2
auto-calibration on start-
up, and LEL by vol CH
4
;
3. User Options: confidence beep, latching alarms, safe mode, stealth mode, IR stealth mode, low alarm acknowledge, force calibration when overdue, calibration lock, force bump when overdue,
4. Language: English, French (Français), German (Deutch), Spanish (Español), and Portuguese (Portugês).
16
Page 25
GasAlertMicroClip
Options
Sensor Disabled
Detector Identification
a Warning
Start-up Message
Disabling an installed sensor configures the detector to a 1, 2, or 3-gas unit. No protection is now provided for the gas targeted by that sensor(s). Disabling a sensor should be performed with extreme caution.
This option allows you to personalize your detector. In the Detector Identification group there are two fields available for you to enter a message:
Startup Message Top Line; and
Use the Disabled function to disable a sensor. If a sensor is disabled, it can no longer be viewed on the LCD during normal operation. All sensors are enabled upon shipment. If a sensor is disabled, it may be enabled again at any time
Startup Message Bottom Line. If a message is entered, it appears during the start-up
sequence after the detector version is displayed. The message either
If a sensor fails, disabling the sensor deactivates the sensor alarm (refer to Alarms
). The sensor should be replaced and
enabled as soon as possible (see
Replacing a Sensor or
Sensor Filter).
a) displays on the LCD for 3 seconds (if the entire
message fits on the LCD); or
b) scrolls by twice on the LCD.
The detector functions normally with the remaining enabled sensors.
Sensor Options
The sensor options tab allows you to change the factory settings for the H
2
S, CO, O2, and LEL sensors.
17
Page 26
GasAlertMicroClip
User Manual
Bump Interval
Calibration Gas Concentration
The Bump Interval determines the number of days before a sensor needs to be bump checked. A different interval value can be selected for each sensor. Set the interval value to 0 to disable this option. This option is disabled upon shipment.
The calibration span gas option (Cal Gas) allows you to input a new gas concentration for each sensor. Refer to the values listed on your calibration gas cylinder.
Note
Note
BW recommends that the calibration concentration value be set between specific ranges. Refer to the
Calibration
section.
BW recommends to “bump check” the sensors, before each day’s use, to confirm their ability to respond to gas by exposing the detector to a gas concentration that exceeds the alarm setpoints.
Calibration Interval
Low Alarm
The calibration interval (Cal Interval) allows you to choose the number of days before the next calibration should be performed. You can select the minimum number of days (1 day) to the maximum number of days (365 days) before the next calibration. A different calibration interval can be selected for each sensor. Select 0 to disable the calibration interval. The detector is shipped with a default of 180 days.
Low Alarm option allows you to change the low alarm setpoint for each of the sensors. Refer to
Resetting Gas
Alarm Setpoints for the factory alarm setpoints.
High Alarm
High Alarm option changes the high alarm setpoint for each of the sensors. Refer to
Resetting Gas Alarm Setpoints
for
the factory alarm setpoints.
Note
BW recommends that the detector be calibrated once every 180 days (6 months).
18
Page 27
GasAlertMicroClip
Options
Automatic O2 Calibration
TWA Alarm
When the O2 Auto-Calibration on Startup option is enabled, it forces the detector to automatically calibrate the oxygen sensor upon activating the detector. The automatic calibration option is enabled upon shipment.
The TWA Alarm option allows you to select the time­weighted average (TWA) alarm setpoint for either the H
2
S or
CO sensor. Refer to
Resetting Gas Alarm Setpoints
for the
factory alarm setpoints.
LEL By Vol CH4
STEL Alarm
The LEL By Vol CH4 option is only applicable to the LEL sensor. If it is enabled, it shows the LEL reading in %vol. assuming a methane environment.
The STEL Alarm option changes the short-term exposure limit (STEL) alarm setpoint for either the H
2
S or CO sensor.
Refer to
Resetting Gas Alarm Setpoints
for the factory alarm
setpoints.
STEL Interval
The STEL interval determines the rate at which STEL is calculated. The detector is shipped with the STEL Interval set to 15 minutes. This value can be adjusted between 5 and 15 minutes.
Auto Zero on Start-up
When enabled this option (CO/H2S/LEL Auto-Zero on Startup) allows the detector to auto zero the chosen sensor
during activation. The auto zero option for each sensor is disabled upon shipment.
19
Page 28
GasAlertMicroClip
User Manual
User Options Menu
Confidence Beep
The Confidence Beep notifies that the detector is on and the battery has sufficient power to respond to a hazardous level of gas to emit an alarm. Instead of beeping when the battery power is low, the audible alarm beeps once every 10 seconds to advise that the battery has sufficient power. The confidence beep stops when the battery power is low. This option is disabled upon shipment.
Latching Alarms
The detector is shipped with the Latching Alarm function disabled. If the low and high gas alarms are set to latch, the audible, visual, and vibrator alarms persist during an alarm. The alarm deactivates when it is acknowledged by pressing C and the alarm condition is no longer present.
Safe Mode
When enabled, the Safe Mode function advises that normal ambient conditions prevail and no monitored gas hazard exist.
SAFE is constantly displayed when all gas levels are
normal or below the alarm setpoints.
The safe mode option is disabled upon shipment.
Stealth Mode
The detector is shipped with Stealth Mode disabled. When it is activated, stealth mode disables the beepers, backlight, and alarm LEDs and
is constantly displayed on the LCD.
IR Stealth Mode
If the IR Stealth Mode option is enabled, it disables the audible alarm and the backlight. The vibrator and the IR LEDs are enabled and
is constantly displayed on the
LCD.
20
Page 29
GasAlertMicroClip
Options
Low Alarm Acknowledge Force Bump When Overdue
When Low Alarm Acknowledge is enabled, the audible alarm can be disabled during a low alarm if C is pressed, but the vibrator and alarm LEDs continue to be enabled. This option is not applicable to the O
2
sensor.
When enabled, the Force Bump When Overdue option forces a bump check if the sensor has exceeded its bump check interval (see
Bump Interval
). If a successful bump
check has not been performed, the detector will deactivate.
Force Calibration When Overdue
Language Menu
Enabling the Force Calibration When Overdue option forces the detector to enter the calibration process if a sensor is overdue for calibration upon start-up. If the detector is not calibrated immediately, the detector will deactivate. The detector is shipped with this option disabled.
The detector is shipped with English as the default language displayed. You can choose to view the detector in these additional languages.
x French (Français); x German (Deutsch);
Cal Lock
x Spanish (Español); and
If Cal Lock is enabled, calibration can only be completed with an IR device like Soft Tools or the MicroDock II Base Station. Though Cal Lock is enabled, the detector will still perform an auto zero when activated. The cal lock option is disabled upon shipment.
x Portuguese (Português).
21
Page 30
GasAlertMicroClip
User Manual
If more than one type or level of alarm exists at the same time, a multi-gas alarm will result.
Alarms
The following table describes the detector alarms and shows how the LCD looks for each alarm.
To change the factory-set alarm setpoints, refer to the GasAlertMicroClip Soft Tools instruction sheet.
During an alarm condition, the detector activates the backlight, audible/visual/vibrator alarms, and the LCD shows the current ambient gas reading.
Table 6. Alarms
Alarms Display` Alarms Display
Low Alarm:
x Slow siren x Slow alternating flash
x
L and target gas bar flash
x Vibrator alarm activates
High Alarm:
x Fast siren x Fast alternating flash
x
L and target gas bar flash
x Vibrator alarm activates
TWA Alarm:
x Slow siren x Slow alternating flash
x
L and target gas bar flash
x Vibrator alarm activates
STEL Alarm:
x Fast siren x Fast alternating flash
x
L and target gas bar flash
x Vibrator alarm activates
22
Page 31
GasAlertMicroClip
Alarms
Table 6. Alarms
Alarms Display Alarms Display
Multi-Gas Alarm:
x Alternating low and high alarm
siren and flash
x L and target gas bars flash x Vibrator alarm activates
Over Range (OL) Alarm:
x Fast siren and alternating flash x L and target gas bar flash x Vibrator alarm activates
Sensor Alarm: x Displays Err
Confidence Beep:
x One beep every 10 seconds
Low Battery Alarm:
(Confidence beep disabled) x One beep and one flash every
5 seconds
x
and L flashes
Automatic Shutdown Alarm:
x Eight beeps and eight flashes x LOW BAT and L display x Vibrator alarm temporarily
activates
x Displays OFF before turning off
23
Page 32
GasAlertMicroClip
User Manual
Table 6. Alarms
Alarms Display
Normal Shutdown:
x Four beeps and flashes x Vibrator alarm temporarily
activates
Note
Alarms can be set to be latching or non-latching. To confirm this setting, access the
Latching Alarms
option in the GasAlertMicroClip Soft Tools. Local regulations may require a latching alarm.
If the detector is in stealth mode, it only vibrates in alarm mode (the audible and visual alarms are disabled). If the detector is in IR stealth mode, the IR LEDs and vibrator activate in alarm mode (the audible alarm is disabled).
If the low alarm acknowledge option is enabled, the audible alarm can be disabled only during a low alarm by pressing C. The vibrator and LEDs remain activated. If the alarm escalates to a high, TWA, or STEL alarm, the audible alarm reactivates.
The backlight activates during any alarm condition.
Computed Gas Exposures
a Warning
To avoid possible personal injury, do not deactivate the detector during a work shift. TWA, STEL, and MAX readings reset once the detector is deactivated.
Table 7. Computed Gas Exposures
Gas Exposure Description
TWA (H
2
S and CO
only)
Time-weighted average (TWA) based on accumulated exposure to toxic gases averaged over a workday according to OSHA method.
OSHA: 8 hour moving average
STEL (H
2
S and CO
only)
Short-term exposure limit (STEL) to gas based on a 5-15 minute user selectable period.
Maximum* (peak)
Maximum (MAX) concentration encountered during work shift.
* For oxygen, it is the highest or the lowest concentration encountered.
24
Page 33
GasAlertMicroClip
Alarms
Viewing Gas Exposures
To view the TWA, STEL, and maximum (MAX) hold readings, press
C twice. The LCD first displays the TWA
gas exposures.
Then the LCD displays the STEL gas exposures.
Finally the LCD displays the MAX readings.
Clearing Gas Exposures
a Caution
Follow all safety procedures as defined by your employer.
Confirm with your supervisor before clearing TWA and STEL alarms.
To clear the TWA, STEL, and MAX exposure readings, press C when the LCD displays RESET.
25
Page 34
GasAlertMicroClip
User Manual
Gas Alarm Setpoints
The detector's gas alarm setpoints trigger the gas alarms that are described in the table below.
Table 8. Gas Alarm Setpoints
Alarm Condition
Low alarm Toxics and combustibles: Ambient
gas level above low alarm setpoint. Oxygen: Ambient gas level may be
set to above or below 20.9% (or 20.8%).
High alarm Toxics and combustibles: Ambient
gas level above high alarm setpoint. Oxygen: Ambient gas level may be
set to above or below 20.9% (or 20.8%).
TWA alarm Toxic only: Accumulated value
above the TWA alarm setpoint.
STEL alarm Toxic only: Accumulated value
above the STEL alarm setpoint.
Multi-gas alarm Two or more gas alarm conditions.
Resetting Gas Alarm Setpoints
Note
To change the alarm setpoints, you must connect the detector to the IR Link adapter and use the GasAlertMicroClip Soft Tools software. Refer to the GasAlertMicroClip Soft Tools manual for complete instructions.
Standard factory alarm setpoints will vary by region.
Occupational Safety and Health Association (OSHA) standard settings are used as an example.
The following table lists the factory alarm setpoints.
Table 9. Sample Factory Alarm Setpoints
Gas TWA STEL Low High
O
2
N/A N/A 19.5% vol. 23.5% vol. LEL N/A N/A 10% LEL 20% LEL CO 35 ppm 50 ppm 35 ppm 200 ppm H2S 10 ppm 15 ppm 10 ppm 15 ppm
Note
You can disable an alarm by setting the alarm setpoint to 0.
26
Page 35
GasAlertMicroClip
Alarms
Stopping a Gas Alarm
The low and high alarms stop when the ambient gas level returns to the acceptable range.
Note
If alarms are set to latch, press
C to reset the
alarms.
The detector computes the TWA value based on OSHA standards and the STEL value based on a user selectable 5 to 15 minute period (see
STEL Interval
).
To stop a TWA or STEL alarm, perform one of the following:
1. Deactivate (see Deactivating the Detector) and reactivate (see
Activating the Detector
) the
detector.
2. Reset the TWA/STEL/MAX exposure readings (see
Viewing Gas Exposures
).
Sensor Alarm
The detector tests for a missing or defective sensor on activation and continuously thereafter. If a sensor fails the self-test, Err appears above the gas bar of the failed sensor (refer to the
Troubleshooting
section).
Low Battery Alarm
The detector tests the battery on activation and continuously thereafter. Battery power is continually displayed during normal operation. If the battery voltage is low, the detector activates the low battery alarm.
The low battery alarm continues until you recharge the battery. If the battery voltage drops too low, the detector executes an automatic shutdown.
Note
If the confidence beep is on, the audible alarm does not beep during a low battery alarm (see
Confidence Beep
). Typically, the low battery alarm continues for 30 minutes before an automatic shutdown.
Automatic Shutdown Alarm
If the battery voltage is in immediate danger of dropping below the minimum operating voltage, LOW BAT and L display, the audible alarm beeps eight times, and the visual alarm flashes eight times. The LCD then displays OFF before it deactivates and the detector stops normal operation. Charge the battery. See the section,
Charging the
Detector.
27
Page 36
GasAlertMicroClip
User Manual
Calibration
x Calibrate the detector at least once every 180 days,
depending on use and sensor exposure to poisons and contaminants.
Guidelines
When calibrating the detector, adhere to the following guidelines:
x Calibrate the detector if the ambient gas display varies
at start-up.
x Recommended gas mixture:
CO: 50 to 500 ppm balance N
2
H
2
S: 10 to 100 ppm balance N
2
LEL: 10 to 100% LEL or 0.5 to 5% by vol. methane balance air O
2
: clean air, 20.9 % (or 20.8%)
x It is best to calibrate the sensor before changing the
alarm setpoints.
x Calibrate only in a safe area that is free of hazardous
gas.
x Do not calibrate the detector during or immediately after
charging is complete.
x CG-Q58-4 and CG-Q34-4 calibration gas (4-gas mix)
are available from BW Technologies by Honeywell. See the section,
Replacement Parts and Accessories
.
x The oxygen sensor can be automatically calibrated each
time upon activation (if this feature is enabled). Activate the detector in a normal (20.9%/20.8% oxygen) atmosphere.
x Calibration accuracy is never better than the calibration
gas accuracy. BW Technologies by Honeywell recommends a premium-grade calibration gas. Gases with the National Institute of Standards and Technology (NIST) traceable accuracy improve the validity of the calibration. Do not use a gas cylinder beyond its expiration date.
x The detector should be allowed to stabilize for 1 minute
after activation, prior to calibration or bump test.
x If you require a certified calibration, contact
BW Technologies by Honeywell
.
x Calibrate a new sensor before use. Install the sensor,
activate the detector, and allow the sensor to stabilize before starting calibration (used: 60 seconds; new: 5 minutes).
28
Page 37
GasAlertMicroClip
Calibration
Wind currents may cause false readings and poor calibrations.
Diagnostics Protection
The detector tests the ambient air (auto zero) and the test gas that is applied (auto span) to ensure it meets expected values.
If an unexpected level of target gas is present during an auto zero, the auto zero will fail for that sensor. The detector advises which sensor(s) failed the auto zero before proceeding to the auto span.
In auto span, if any target gas is not present or does not meet expected values, the display will advise you of that failure. The sensor(s) that failed to span will retain its previous set value and not continue with the calibration process.
Figure 3. Applying Gas to the Sensors
Applying Gas to the Sensors
Table 10. Applying Gas to the Sensors
The calibration cap and hose, which is shipped with the detector, simplifies sensor testing and calibration. Figure 3 and Table 10 show how to use it when applying gas to the sensors.
Item Description
1 Detector and calibration cap 2 Calibration hose 3 Regulator and gas cylinder
Note
The calibration cap should only be used during the calibration span process.
29
Page 38
GasAlertMicroClip
User Manual
Calibration Procedure
To calibrate the detector perform the following procedure.
Note
Calibrate O
2
in clean air.
If C is pressed at any time during the calibration procedure, calibration is ABORTED. The detector emits a quick beep and the calibration due dates are displayed before the detector returns to normal operation.
Start Calibration
Note
Verify that the calibration gas you are using matches the span concentration value(s) in the detector. See the section,
Calibration Gas
Concentration.
1. In a clean atmosphere, press and hold C as the detector beeps, flashes, and vibrates to the corresponding OFF countdown. Continue to hold C as the detector briefly deactivates.
2. The detector then reactivates and performs the CAL countdown. Continue to hold C until the CAL countdown is complete to enter calibration.
Note
If C is not held for the entire CAL countdown, the detector will deactivate.
30
Page 39
GasAlertMicroClip
Calibration
Auto Zero and Oxygen Sensor Calibration Auto Span
4. The LCD now prompts you to apply calibration gas to the sensors.
Note
Do not apply calibration gas at this point, otherwise auto zero may fail for that sensor and then it cannot be spanned.
3. The LCD flashes
while the detector automatically zeroes the toxic and combustible sensors and calibrates the oxygen sensor. The LCD notifies you if the auto zero has failed for a sensor.
5. K flashes and
remains displayed as you attach the calibration cap and apply gas to the sensors at a flow rate of 250 to 500 ml/min. (refer to Figure 3. Applying Gas to the Sensors).
Auto Zero Sensor Fail
The audible alarm then beeps twice.
31
Page 40
GasAlertMicroClip
User Manual
No Gas Detected
When the detector senses a sufficient amount of gas concentration (approximately 30 seconds), the audible alarm beeps once,
flashes, and K remains lit while the detector completes the span (approximately 2 minutes).
The LCD displays FAILURE if the detector does not detect any gas within 2 minutes before proceeding with the calibration procedure.
Note
Successful Span
The detector will not span a sensor if:
If the sensors have spanned successfully, the audible alarm beeps before the calibration procedure continues.
x
You do not apply gas to the sensor.
x
The sensor fails to detect at least one-half of the expected gas concentration in the first 30 seconds.
Unsuccessful Span
If the sensors fail the span, the LCD displays FAILURE before proceeding with the calibration procedure.
x
The gas concentration drops below one-half of the expected gas level during the 2 minute span.
32
Page 41
GasAlertMicroClip
Calibration
If you apply gas to a sensor and the detector fails to span the sensor, repeat the calibration procedure using a new gas cylinder. If the sensor fails the span a second time, replace the sensor. See the section,
Replacing a Sensor or Sensor
Filter.
Calibration Due Date
Note
If a sensor does not successfully span, its calibration due date will not be reset.
6. After span is complete, the LCD displays the following calibration due date screens before returning to normal operation.
Note
The number of days shown in the last screen represents the earliest date a calibration needs to be performed.
If a sensor failed to successfully span and it is past its calibration due date, the LCD displays the following three screens.
Press C to acknowledge the warning before returning to normal operation.
33
Page 42
GasAlertMicroClip
User Manual
Verification
x Peak exposure level (ppm or %);
After calibration is complete and the detector is in normal operating mode, test it using a gas cylinder other than the one used in calibration. The gas concentration should not exceed the sensor's detection range. Confirm that the LCD shows the expected concentration. To ensure the readings are accurate, apply the test gas for the same amount of time as was applied to the sensor when it was calibrated.
x Alarm duration in seconds; and x Alarm type.
Downloading Datalogs and Event Logs
The datalog and event log files can only be downloaded to a PC using an IR device (e.g., IR Link Adapter or MicroDock II Base Station).
Datalogs
Maintenance
The datalogger allows the detector to record various information so you can compile a report. The detector has a sample rate of 15-second intervals. The detector is capable of storing 16 hours of information. When the memory is full, the detector replaces the oldest data with the most recent data.
To keep the detector in good operating condition, perform the following basic maintenance as required:
x Calibrate, bump check, and inspect the detector at
regular intervals.
Event Logs
x Keep an operations log of all maintenance, bump checks,
calibrations, and alarm events. The event logging feature keeps a record of the ten most recent gas alarm events. The information that is recorded from an event is as follows:
x Clean the exterior with a soft damp cloth. Do not use
solvents, soaps, or polishes.
x Serial number; x Gas type; x Start time of event;
34
Page 43
GasAlertMicroClip
Maintenance
Charging the Detector
a Warning
The detector must be charged in a non-
hazardous atmosphere of 32°F - 113°F (0°C - 45°C).
Charge the detector using only the
recommended charging adapter. Do not use any other charging adapter. Failure to observe this precaution can lead to fire or explosion.
The charging adapter is voltage specific to North
America. Use of the charging adapter outside of North America will damage the charger and detector.
The battery can only be replaced by the
manufacturer. Not complying with this might cause explosion danger.
Note
To preserve battery life, deactivate the detector when you are not using it.
To charge the detector, execute the following steps:
1. Deactivate the detector and attach the GasAlertMicroClip charging adapter as shown in
the following figure. The detector can be activated during charging.
Figure 4. Connecting the Charging Adapter
35
Page 44
GasAlertMicroClip
User Manual
Each sensor has a high degree of resistance to common vapors and gases. A sensor will most likely clear itself if you move the detector to a clean environment and wait 10 to 30 minutes, but it could take up to a couple of days depending upon the gas that came in contact with the sensor. Do not expose a sensor to the vapors of inorganic solvents such as paint fumes or organic solvents. The
Troubleshooting
section describes problems caused by a sensor in need of calibration or replacement.
2. Allow the detector to charge for 2-3 hours. The charging indicator flashes on the LCD while the detector is being charged.
3. Once charging is complete, the LCD continues to display
. Remove the adapter and activate the
detector.
Note
If the battery indicator does not display within 5 minutes, refer to the
Troubleshooting
section.
To replace a sensor or sensor filter, refer to the following figure, table, and set of instructions.
To reach full battery capacity, allow the battery to fully charge and fully discharge three times.
Charging the detector in temperatures above 113°F will greatly reduce the number of charges the detector can accept.
The detector may be hot immediately following charging. This is normal.
Replacing a Sensor or Sensor Filter
a Warning
To avoid personal injury, use only sensors specifically designed for the detector. See the section
Replacement Parts and Accessories
.
Figure 5. Replacing a Sensor or Sensor Filter
36
Page 45
GasAlertMicroClip
Maintenance
3. Remove the two screws on the PCB and remove the PCB.
Table 11. Replacing a Sensor or Sensor Filter
Item Description
1 Front shell 2 LEL sensor 3 PCB 4 PCB screws (2) 5 Rear shell 6 Machine screws (6) 7 Sensor filter 8O
2
sensor
9H
2
S sensor 10 CO sensor 11 Sealing rib
Note
Ensure the battery does not get damaged once the PCB is removed.
4. Remove the old sensor filter or slide/pull out the old sensor(s).
5. Insert a new sensor filter or sensor(s).
Note
When inserting a new sensor filter, ensure the white side is facing the sensors (i.e., the black side is facing out).
Detectors that are configured for 1, 2, or 3 gases may contain a dummy sensor in one of the four sensor locations.
1. Deactivate the detector.
2. Remove the six machine screws on the rear shell and remove the back cover.
37
Page 46
GasAlertMicroClip
User Manual
6. Re-assemble the detector. When assembling the detector, be aware of the following:
x Ensure the charging contact pins inside the
rear shell are lined up to its appropriate hole before inserting the rear shell in place. If the contact pins are bent, the battery cannot charge properly.
x Verify that the battery PCB is in the same
position as when the rear shell was removed (refer back to step #2).
x Visually inspect the battery to ensure that it
has not been damaged before replacing the rear shell.
x Ensure the rib on the interior of the rear shell
fits between the battery pack and the PCB as shown in Figure 6.
x Ensure the rear shell sealing rib is not folded
when replacing the rear shell (see Figure 5).
x Torque the screws to 3-4 in-lbs in a crisscross
pattern to ensure a proper environment seal.
Figure 6. Replacing the Rear Shell
Calibrate the detector after changing any sensor. See the section,
Calibration
.
38
Page 47
GasAlertMicroClip
Troubleshooting
Troubleshooting
If you are still unable to correct the problem, contact
BW Technologies by Honeywell
.
The detector’s electronics are protected from variations in humidity and corrosive atmospheres. If you encounter a problem, try the solutions listed in the following table.
Table 12. Troubleshooting Tips
Problem Possible Cause Solution
The detector does not activate.
o Depleted battery
o Damaged or defective detector
o Charge battery (see Charging the
Detector)
o Contact BW Technologies by
Honeywell
The detector enters alarm mode immediately when activated.
o Sensor needs to stabilize
o Low battery alarm
o Detector requires calibration
o Used sensor: wait 60 seconds
New sensor: wait 5 minutes
o Charge battery (see
Charging the
Detector)
o Calibrate detector (see Calibration
)
The activation self-test fails.
o General fault o Sensor failure
o Contact BW Technologies by
Honeywell
o Replace sensor (see Replacing a
Sensor or Sensor Filter)
39
Page 48
GasAlertMicroClip
User Manual
Table 12. Troubleshooting Tips (cont.)
Problem Possible Cause Solution
Detector does not display normal ambient gas reading after activation self-test.
o Sensor not stabilized
o Detector requires calibration
o Target gas is present
o Used sensor: wait 60 seconds
New sensor: wait 5 minutes
o Calibrate detector (see
Calibration
)
o Detector is operating properly. Use
caution in suspect areas
Detector does not respond to pushbutton.
o Battery is depleted
o Detector is performing operations that
do not require user input
o Charge battery (see Charging the
Detector)
o Pushbutton operation restored
automatically when the operation ends
Detector does not accurately measure gas.
o Detector requires calibration
o Detector is colder/hotter than
ambient gas
o Sensor filter is blocked
o Calibrate sensor (see
Calibration
)
o Allow the detector to acquire ambient
temperature before use
o Clean the sensor filter (see
Replacing a Sensor or Sensor Filter
)
Detector has been charging for over 3 hours and the charging indicator still indicates that the battery has not finished charging.
o Battery is trickle charging o Battery is full and ready for operation
40
Page 49
GasAlertMicroClip
Troubleshooting
Table 12. Troubleshooting Tips (cont.)
Problem Possible Cause Solution
Detector does not enter alarm.
o Alarm setpoint(s) are set incorrectly
o Alarm setpoint(s) set to zero
o Detector is in calibration mode
o Reset alarm setpoints (see
Resetting
Gas Alarm Setpoints)
o Reset alarm setpoints (see Resetting
Gas Alarm Setpoints)
o Complete the calibration procedure
Detector intermittently enters alarm without apparent reason.
o Ambient gas levels are near alarm
setpoint or the sensor is exposed to a puff of the target gas
o Alarms set incorrectly
o Detector requires calibration o Missing or faulty sensor
o Detector is operating normally. Use
caution in suspect areas. Check maximum gas exposure reading.
o Reset alarm setpoints (see
Resetting
Gas Alarm Setpoints)
o Calibrate sensors (see Calibration
)
o Replace sensor (see Replacing a
Sensor or Sensor Filter)
Detector automatically deactivates.
o Automatic shutdown feature activated
due to weak battery
o Charge detector (see Charging the
Detector)
Battery indicator does not display when charging.
o Detector may be depleted below
normal levels
o Charge detector for approximately
8 hours. Detector LEDs may light during first 5 hours. This is normal.
o If the battery indicator does not light
after charging for 8 hours, contact
BW Technologies by Honeywell
.
41
Page 50
GasAlertMicroClip
User Manual
Replacement Parts and Accessories
a Warning
To avoid personal injury or damage to the detector, use only the specified replacement parts.
To order parts or accessories listed in the following table, contact
BW Technologies by Honeywell
.
Table 13. Replacement Parts and Accessories
Model No. Description Qty
SR-W-MC Replacement combustible sensor 1 SR-X-MC Replacement O2 sensor 1 SR-M-MC Replacement CO sensor 1
SR-H-MC Replacement H2S sensor 1 REG-0.5 Regulator (0.5 l/min) 1 CG-Q58-4 Quad gas cylinder: CH4-2.5%, O2-
18.0%, H
2
S-25 ppm,
CO-100 ppm, bal. N
2
(58 l)
1
CG-Q34-4 Quad gas cylinder: CH4-2.5%, O2-
18.0%, H
2
S-25 ppm,
CO-100 ppm, bal. N
2
(34 l)
1
CG-T34 Dual gas cylinder: 50% LEL
(CH
4
-2.5%) O2-20.9%, bal. N
2
(34 l)
1
Model No. Description Qty
G0042-H25 Single gas cylinder: H2S 25 ppm,
bal. N
2
(58 l)
1
CG2-M-200­103
Single gas cylinder: CO 200 ppm, bal N
2
(103 l)
1
CG-BUMP1 Bump alarm gas aerosol
(CH
4
-2.5%, O2-10%, H2S-40 ppm,
CO-200 ppm)
1
CK-Q34-4 Quad calibration kit with regulator,
quad gas cylinder (CG-Q34-4), hose, and carrying case
1
CK-Q58-4 Quad calibration kit with regulator,
quad gas cylinder (CG-Q58-4), hose, and carrying case
1
MC-XWHM­Y-NA-07
Confined space kit with manual aspirator pump
1
DOCK2-0­1C1L-00-N
GasAlertMicroClip MicroDock II docking module with charging cable
1
MC­SCREW-K1
Screw kit (5 sets of screws and screwdriver)
1
MC-AF-K1 Auxiliary filter kit 1 MC-SS-AF-K1Replacement auxiliary filters 10
MC-SS Sensor filters 2 MC-PA-1 Replacement charging adapter 1
42
Page 51
GasAlertMicroClip
Specifications
Model No. Description Qty
MC-VC-K1 Vehicle charging kit 1 MC-C01-
MC5*
GasAlertMicroClip MultiCharger 1
GA-USB1-IR IR connectivity kit 1 GA-PA-1* Charging adapter 1 MC-TC-1 Calibration cap and hose (1 ft./3 m) 1 GA-AG-3 Alligator clip (stainless steel) 1
*Add suffix (-UK) for United Kingdom mains plug, (-EU) for European mains plug, (-AU) for Australian mains plug.
Specifications
Instrument dimensions: 10.75 x 6.00 x 2.73 cm (4.2 x 2.4 x 1.1 in.)
Weight: 160 g (5.7 oz.) Operating temperature: -20°C to +58°C (-4°F to +136°F)
+50°C to +58°C is certified by CSA-International on the combustible sensor with r5% accuracy
Storage temperature: -40°C to +50°C (-40°F to +122°F) Operating humidity: 0% to 95% relative humidity
(non-condensing)
Alarm setpoints: May vary by region and are user-settable
Detection range:
H
2
S: 0 – 100 ppm (1 / 0.1 ppm increments)
H
2
S: 0 – 200 ppm (1 ppm increments) CO: 0 – 500 ppm (1 ppm increments) CO: 0 – 1000 ppm (1 ppm increments) O
2
: 0 – 30.0% vol. (0.1% vol. increments) Combustible (LEL): 0 – 100% LEL (1% LEL increments) or 0 – 5.0% v/v methane
Sensor type:
H
2
S, CO, O2: Single plug-in electrochemical cell Combustibles: Plug-in catalytic bead
O measuring principle:
2
Capillary controlled concentration
sensor Alarm conditions: TWA alarm, STEL alarm, low alarm,
high alarm, multi-gas alarm, over range (OL) alarm, low battery alarm, confidence beep, automatic shutdown alarm
Audible alarm: 95 dB+ at 30 cm (100 dB typical) variable pulsed beeper
Visual alarm: Red light-emitting diodes (LED) Display: Alphanumeric liquid crystal display (LCD) Backlight: Activates when the pushbutton is pressed and
deactivates after 5 seconds. Also activates during an alarm condition.
Self-test: Initiated upon activation Calibration: Automatic zero and automatic span Oxygen sensor: Automatic span upon activation
(selectable)
43
Page 52
GasAlertMicroClip
User Manual
Approvals:
Approved by CSA to both U.S. and Canadian Standards CAN/CSA C22.2 No. 157 and C22.2 152 ANS/UL – 913 and ANSI/ISA – S12.13 Part 1
CSA Class I, Division 1, Group A, B, C, and D ATEX CE 0539 g II 1 G EEx ia IIC T4
KEMA 06ATEX0056
IECEx Ex ia IIC T4
User field options: Start-up message, confidence beep,
latching alarm, enable/disable safe display mode, oxygen measurement, combustible sensor measurement, sensor disable, set calibration interval, force calibration, calibration due lock, force bump, bump interval, stealth mode, IR stealth mode, low alarm acknowledge, language selection, enable/disable automatic oxygen calibration, enable/disable auto zero at start-up, set alarm setpoints, set span concentration values, set STEL calculation period.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules and ICES-003 Canadian EMI requirements. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one of more of the following measures:
Battery operating time:
1 rechargeable lithium polymer: 10-12 hours (typical) Year of manufacture: The detector’s year of manufacture is
determined from the serial number. The second and third number after the first letter determines the year of manufacture.
E.g., H304
-Y000001 = 2004 year of manufacture
Approved battery: North America
Approved batteries for product (standards EN50020, UL913, C22.2 No. 157)
Rechargeable battery Temperature code
Lithium polymer T4 Battery charger: GasAlertMicroClip charging adapter
x Reorient or relocate the receiving antenna. x Increase the separation between the equipment and
receiver.
First-time charge: 2-3 hours Normal charge: 2-3 hours
x Connect the equipment into an outlet on a circuit different
from that to which the receiver is connected.
Warranty: 2 years including sensors
x Consult the dealer or an experienced radio/TV technician
for help.
44
Page 53
GasAlertMicroClip
Specifications
General Datalogger Specifications
Storage: 16 hours at 15-second intervals Memory type: Wraparound memory ensures most recent
data is always saved
Sample rate: One reading every 15 seconds Data recorded: All sensor readings, all alarm conditions,
calibrations, event flags, battery status, sensor status, confidence beep activation, and detector status along with the time and date for each reading and unit serial number
Operation: Requires no user intervention (automatic) Compatible with: Desktop PC computer or laptop Operating system: Windows 2000 or higher Download via: IR device (e.g., IR Link adapter, MicroDock
II Base Station, etc.) Software required: Soft Tools with spreadsheet/database
compatible with comma-separated-value (CSV) text files (Excel, Access, Quattro, etc.)
45
Page 54
GasAlertMicroClip
User Manual
46
Page 55
D5908/2 English
iERP: 124291 © BW Technologies 2007, all rights reserved. Printed in Canada. All product names are trademarks of their respective companies.
Page 56
Printing Information
iERP # 125109
Size: 3 x 5 inches
Stock: 80 lb cougar cover card stock
Print Color: Black
Language: English
Corporate Headquarters
Calgary, Alberta
ERRATA CARD
125109
GasAlertMicroClip
5 in.
3 in.
Pages 9 & 10
a Caution
Only activate the detector in a safe area that is free of hazardous gas.
Page 19
Auto Zero on Start-up
The auto zero option for each sensor is enabled upon shipment.
The following information has changed in the user manual (124291 D5908/2)
Page 57
DAILY BUMP TEST
iERP: 125538
BW recommends to “bump test” the
sensor(s), before each day’s use to confi rm
their ability to respond to gas by expos-
ing the detector to a gas concentration that
exceeds the alarm setpoints. Manually verify
that the audible and visual alarms are
activated.
If any sensor fails the bump test, contact
your local BW service department immedi-
ately.
a IMPORTANT NOTE
Loading...