BW Technologies GasAlert MicroClip Operating Manual

Page 1
GasAlertMicroClip
H2S, CO, O2, Combustibles
1, 2, 3, and 4 Gas Detectors
User Manual
Page 2
Limited Warranty & Limitation of Liability
BW Technologies LP (BW) warrants this product to be free from defects in material and workmanship under normal use and service for a period of two years, beginning on the date of shipment to the buyer. This warranty extends only to the sale of new and unused products to the original buyer. BW’s warranty obligation is limited, at BW’s option, to refund of the purchase price, repair, or replacement of a defective product that is returned to a BW authorized service center within the warranty period. In no event shall BW’s liability hereunder exceed the purchase price actually paid by the buyer for the Product. This warranty does not include:
a) fuses, disposable batteries or the routine replacement of parts due to the normal wear and tear of the product arising from use; b) any product which in BW’s opinion, has been misused, altered, neglected or damaged by accident or abnormal conditions of operation, handling or use; c) any damage or defects attributable to repair of the product by any person other than an authorized dealer, or the installation of unapproved parts on the
product; or
The obligations set forth in this warranty are conditional on:
a) proper storage, installation, calibration, use, maintenance and compliance with the product manual instructions and any other applicable
recommendations of BW;
b) the buyer promptly notifying BW of any defect and, if required, promptly making the product available for correction. No goods shall be returned to BW
until receipt by the buyer of shipping instructions from BW; and
c) the right of BW to require that the buyer provide proof of purchase such as the original invoice, bill of sale or packing slip to establish that the product is
within the warranty period.
THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. BW SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT OR RELIANCE OR ANY OTHER THEORY. Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
BW Technologies by Honeywell BW Technologies by Honeywell BW Technologies by Honeywell Corporate Headquarters America Europe 2840 – 2nd Ave. SE 3279 West Pioneer Parkway 5 Canada Close Calgary, AB Arlington, TX Banbury, Oxfordshire Canada T2A 7X9 USA 76013 UK OX16 2RT
Page 3

Table of Contents

Title Page
Introduction............................................................................................................................................1
Contacting BW Technologies by Honeywell.........................................................................................2
Safety Information - Read First.............................................................................................................2
Getting Started......................................................................................................................................6
Activating the Detector........................................................................................................................10
Self-Test......................................................................................................................................10
Self-Test Pass.............................................................................................................................15
Self-Test Fail ...............................................................................................................................15
Battery Test.................................................................................................................................15
Backlight......................................................................................................................................16
Deactivating the Detector....................................................................................................................16
Options................................................................................................................................................16
Detector Identification..................................................................................................................17
Sensor Options............................................................................................................................17
User Options Menu .....................................................................................................................20
Language Menu...........................................................................................................................21
Alarms.................................................................................................................................................22
Computed Gas Exposures..........................................................................................................24
Viewing Gas Exposures..............................................................................................................25
Gas Alarm Setpoints ...................................................................................................................26
Resetting Gas Alarm Setpoints...................................................................................................26
Stopping a Gas Alarm.................................................................................................................27
Sensor Alarm...............................................................................................................................27
i
Page 4
GasAlertMicroClip
User Manual
Title Page
Low Battery Alarm .......................................................................................................................27
Automatic Shutdown Alarm.........................................................................................................27
Calibration...........................................................................................................................................28
Guidelines....................................................................................................................................28
Diagnostics Protection.................................................................................................................29
Applying Gas to the Sensors.......................................................................................................29
Calibration Procedure..................................................................................................................30
Datalogs..............................................................................................................................................34
Event Logs ..........................................................................................................................................34
Downloading Datalogs and Event Logs......................................................................................34
Maintenance........................................................................................................................................34
Charging the Detector .................................................................................................................35
Replacing a Sensor or Sensor Filter ...........................................................................................36
Troubleshooting...................................................................................................................................39
Replacement Parts and Accessories..................................................................................................42
Specifications......................................................................................................................................43
General Datalogger Specifications..............................................................................................45
ii
Page 5

List of Tables

Table Title Page
1. Gases Monitored ...................................................................................................................1
2. International Symbols............................................................................................................5
3. GasAlertMicroClip Detector...................................................................................................7
4. Display Elements...................................................................................................................8
5. Pushbutton.............................................................................................................................9
6. Alarms..................................................................................................................................22
7. Computed Gas Exposures ..................................................................................................24
8. Gas Alarm Setpoints............................................................................................................26
9. Sample Factory Alarm Setpoints.........................................................................................26
10. Applying Gas to the Sensors...............................................................................................29
11. Replacing a Sensor or Sensor Filter ...................................................................................37
12. Troubleshooting Tips...........................................................................................................39
13. Replacement Parts and Accessories ..................................................................................42
iii
Page 6
GasAlertMicroClip
User Manual
iv
Page 7

List of Figures

Figure Title Page
1. GasAlertMicroClip Detector...................................................................................................7
2. Display Elements...................................................................................................................8
3. Applying Gas to the Sensors...............................................................................................29
4. Connecting the Charging Adapter.......................................................................................35
5. Replacing a Sensor or Sensor Filter ...................................................................................36
6. Replacing the Rear Shell.....................................................................................................38
v
Page 8
GasAlertMicroClip
User Manual
CAUTION: FOR SAFETY REASONS, THIS EQUIPMENT MUST BE OPERATED AND SERVICED BY QUALIFIED PERSONNEL ONLY. READ AND UNDERSTAND INSTRUCTION MANUAL COMPLETELY BEFORE OPERATING OR SERVICING.
GasAlertMicroClip Multi-Gas Detector with User Downloadable Datalogger and Event Log
Standard instrument is equipped with integral concussion­proof boot, internal vibrator alarm, and provides full-time continuous datalogging while the instrument is operating. Data is saved to memory and can be downloaded by the user via an IR device to a PC. Data is imported into standard office software (Microsoft® Excel, Access, etc.). Wraparound memory ensures the most recent data is always saved.
vi
Page 9
GasAlertMicroClip

Introduction

a Warning
To ensure your personal safety, read “Safety Information” before you use the detector.
The GasAlertMicroClip gas detector (“the detector”) warns of hazardous gas at levels above user-selectable alarm setpoints.
The detector is a personal safety device. It is your responsibility to respond properly to the alarm.
Table 1 lists the gases monitored.
Table 1. Gases Monitored
Gas Detected Unit of Measure
Hydrogen sulfide (H2S) parts per million (ppm) Carbon monoxide (CO) Oxygen (O2) percent by volume (%) Combustible gases (LEL)
Field selectable for:
parts per million (ppm)
a) percent of lower explosive limit (% LEL) b) percent by volume methane 0-5.0% v/v
1
Page 10
GasAlertMicroClip
User Manual

Contacting BW Technologies by Honeywell

To contact BW Technologies by Honeywell, call: USA: 1-888-749-8878
Canada: 1-800-663-4164 Europe: +44 (0) 1295 700300 Other countries: +1-403-248-9226
Address correspondence to:
BW Technologies by Honeywell Corporate office 2840 – 2 Avenue S.E. Calgary, AB T2A 7X9 CANADA
Email us at: Visit BW Technologies by Honeywell’s web site at:
www.gasmonitors.com
ISO 9001
info@bwtnet.com

Safety Information - Read First

Use the detector only as specified in this manual, otherwise the protection provided by the detector may be impaired.
International symbols used on the detector and in this manual are explained in Table 2.
Read the detector.
Cautions on the following pages before using the
ec Note
This instrument contains a lithium polymer battery. Do not mix with the solid waste stream. Spent batteries should be disposed of by a qualified recycler or hazardous materials handler.
2
Page 11
GasAlertMicroClip
Safety Information - Read First
aCautions
Warning: Substitution of components may impair Intrinsic Safety. ⇒ Caution: For safety reasons, this equipment must be operated and serviced by qualified personnel only. Read
and understand the user manual completely before operating or servicing.
Do not use the detector if it is damaged. Before you use the detector, inspect the case. Look for cracks or
missing parts.
If the detector is damaged or something is missing, contact Use only sensor(s) specifically designed for your GasAlertMicroClip model. (See the section,
and Accessories).
Calibrate the detector before first-time use and then on a regular schedule, depending on use and sensor
exposure to poisons and contaminants. BW recommends at least once every 180 days (6 months).
BW recommends to “bump test” the sensors, before each day’s use, to confirm their ability to respond to gas by
exposing the detector to a gas concentration that exceeds the alarm setpoints. Manually verify that the audible and visual alarms are activated. Calibrate if the readings are not within the specified limits.
It is recommended that the combustible sensor be checked with a known concentration of calibration gas after
any known exposure to catalyst contaminants/poisons (sulfur compounds, silicon vapors, halogenated compounds, etc).
The combustible sensor is factory calibrated to 50% LEL methane. If monitoring a different combustible gas in
the % LEL range, calibrate the sensor using the appropriate gas.
Caution: High off-scale readings may indicate an explosive concentration. Only the combustible gas detection portion of this instrument has been assessed for performance by CSA
International.
BW Technologies by Honeywell immediately.
Replacement Parts
3
Page 12
GasAlertMicroClip
User Manual
aCautions
Protect the combustible sensor from exposure to lead compounds, silicones, and chlorinated hydrocarbons.
Although certain organic vapors (such as leaded gasoline and halogenated hydrocarbons) may temporarily inhibit sensor performance, in most cases, the sensor will recover after calibration.
For use only in potentially explosive atmospheres where oxygen concentrations do not exceed 20.9% (v/v). Any rapid up-scaling reading followed by a declining or erratic reading may indicate a gas concentration beyond
upper scale limit, which may be hazardous.
Calibrate only in a safe area that is free of hazardous gas. Use only BW approved batteries for your GasAlertMicroClip model (see Charge the detector before first-time use. BW recommends the detector be charged after every workday. Charge the GasAlertMicroClip using the recommended charging adapter only. Do not use any other charging
adapter. Failure to observe this precaution could lead to fire or explosion.
Extended exposure of the GasAlertMicroClip to certain concentrations of combustible gases and air may stress a
detector element, which can seriously affect its performance. If an alarm occurs due to high concentration of combustible gases, recalibration should be performed, or if needed, the sensor replaced.
Do not test the combustible sensor’s response with a butane cigarette lighter; doing so will damage the sensor. Do not expose the detector to electrical shock and/or severe continuous mechanical shock. Do not attempt to disassemble, adjust, or service the detector unless instructions for that procedure are
contained in the manual and/or that part is listed as a replacement part. Use only BW Technologies by Honeywell
replacement parts.
The detector warranty will be voided if customer, personnel, or third parties damage the detector during repair
attempts. Non-BW Technologies by Honeywell repair/service attempts void this warranty.
Specifications).
4
Page 13
GasAlertMicroClip
Safety Information - Read First
Table 2. International Symbols
Symbol Meaning
n
g
X
ATEX
IECEx
Approved to both U.S. and Canadian Standards by CSA - International
European Explosives Protection
Conforms to European Union Directives Conforms to European ATEX Directives
International Electrotechnical Commission Scheme for Certification to Standards for Electrical Equipment for Explosive Atmospheres
5
Page 14
GasAlertMicroClip
User Manual

Getting Started

The list below provides the standard items included with your detector. If the detector is damaged or something is missing, contact the place of purchase immediately.
Sensors: H
Calibration hose and cap;
Charging adapter;
Quick reference guide;
Quick reference operations key; and
CD.
S, CO, O2, and combustible (LEL);
2
To order replacement parts, see the section
Parts and Accessories.
The detector comes with sensors and rechargeable battery installed. If the battery needs to be replaced, contact
Technologies by Honeywell. If any of the sensors or the
sensor filter needs to be replaced, refer to
Sensor or Sensor Filter.
To become familiar with the features and functions of the detector, study the following figures and tables:
Figure 1 and Table 3 describes the detector’s components.
Figure 2 and Table 4 describes the detector’s display elements.
Table 5 describes the detector’s pushbutton.
Replacement
BW
Replacing a
6
Page 15
GasAlertMicroClip
Getting Started
Table 3. Parts of the GasAlertMicroClip
Item Description
1 2 3 4 5 6 7
Visual alarm bars (LED) Liquid crystal display (LCD) Audible alarm Sensors Pushbutton Charging connector / IR interface Alligator clip
Figure 1. Parts of the GasAlertMicroClip
7
Page 16
GasAlertMicroClip
User Manual
Figure 2. Display Elements
Table 4. Display Elements
Item Description
1 2 3 4 5 6 7 8
Alarm condition Automatically span sensor Gas cylinder Gas identifier bars Battery life indicator Stealth mode Numeric value Automatically zero sensor
8
Page 17
GasAlertMicroClip
Getting Started
Table 5. Pushbutton
Pushbutton Description
To activate the detector press C.
To deactivate the detector, press and hold
deactivates.
To view the TWA, STEL, and maximum (MAX) hold readings, press and MAX readings, press
C
To initiate calibration, press and hold
C as the LCD briefly turns off and then executes the CAL countdown. Release C once the CAL countdown is complete.
To activate the backlight, press
To acknowledge latched alarms, press
To acknowledge a low alarm and disable the beeper, press C (if Low Alarm Acknowledge user option is
enabled).
C when the LCD displays RESET.
C.
C until the OFF countdown is complete and the LCD
C twice. To clear the TWA, STEL,
C as the detector executes the OFF countdown and continue to hold
C.
9
Page 18
GasAlertMicroClip
User Manual

Activating the Detector

To activate the detector, press C in a normal clean atmosphere.

Self-Test

Once the detector is activated, it performs the following checks. Manually verify that all actions occur.

Battery Test

The detector administers a battery test during start-up. If the battery has insufficient power to start up, the liquid crystal display (LCD) displays the following screen before the detector deactivates.
Recharge the battery for 2-3 hours before restarting the detector (see
Charging the Detector).

Audible/Visual Test

1. The LCD shows all the display elements as it beeps, flashes, vibrates, and briefly turns on the backlight.

Detector Version

2. The version number of the detector is then displays on the LCD.
10
Page 19
GasAlertMicroClip
Activating the Detector
Note
Users can enter a 2-line message, up to a maximum of 25 characters per line. The message appears on the LCD after the detector version displays. This option is disabled in the factory settings. The message can be enabled or changed in the Sensors tab in Soft Tools. See
GasAlertMicroClip Soft Tools instruction sheet.

Alarm Setpoints

Sensors and Power Test

4. The detector then tests the sensors.
3. Next, the LCD shows the TWA, STEL, low, and high alarm setpoints.
Note
The alarm setpoints on a shipped detector may vary by region. See
Resetting Gas Alarm Setpoints.
A screen then appears confirming that the test was successful (TEST OK) or which sensor(s) failed the test (e.g., ERROR H2S) before continuing with the self-test.
11
Page 20
GasAlertMicroClip
User Manual
Note
The sensors are continuously tested while the detector is activated.

Automatic Zero and O2 Calibration (optional)

5. The sensors (H2S, CO, and LEL) are automatically zeroed and the O calibrated (unless it is disabled in the detector’s options). The detector beeps twice to signal a successful zero and span.
sensor is automatically
2
Note
If ambient air is set to be measured 20.8% vol., the automatic oxygen calibration screen displays 20.8 instead of 20.9.

Calibration Due Date (optional)

6. The LCD displays the number of days remaining until the next calibration is due.
If only the automatic oxygen calibration is enabled on the detector, the following screen appears.
12
If the
calibration interval is set to 0, then the
calibration due date start-up feature is bypassed.
If any sensor is past its calibration due date, the detector displays CAL DUE todAY as the detector beeps, flashes, and vibrates.
Note
Page 21
GasAlertMicroClip
Activating the Detector
Note
Calibration can also be performed by using BW Technologies by Honeywell’s MicroDock II.
Cal Lock Enabled
If the Cal Lock option is enabled, an infrared (IR) device is required to calibrate the detector.
Press C to acknowledge the warning and enter normal operation, or continue with the self-test if applicable.
IR--LOCK displays as the detector tries to establish a connection with an IR device (e.g., BW Technologies by Honeywell’s MicroDock II or IR Link adapter).
Forced Calibration Enabled
If forced calibration is enabled (refer to
Force Calibration When Overdue) and a sensor is past its calibration due date,
calibration is mandatory before the detector can enter normal operation.
If C is pressed, the detector launches the calibration procedure (refer to minutes, the detector deactivates.
Calibration). If C is not pressed within 2
Once IR communication has been established, the calibration procedure is executed. If IR communication is not established within 2 minutes or C is pressed, the detector deactivates.
13
Page 22
GasAlertMicroClip
User Manual
Bump Check (optional)
Note
Bump check is not performed if the detector was just calibrated (step #6).
BW recommends to “bump check” the sensors, before each day’s use, to confirm their ability to respond to gas by exposing the detector to a gas concentration that exceeds the alarm setpoints.
If the
bump interval is set to 0, the bump check
start-up feature is bypassed.
7. If bump check is enabled and the bump checks have passed, or none of the sensors are due for a bump check, a warning screen displays the number of days remaining until a bump check needs to be performed.
If any of the sensors need to be bump checked, the detector displays the following screen:
If force bump or is not enabled, press C to enter normal operation.
Note
BUMPCHK todAY also displays if a bump check was executed but failed.
Force Bump Enabled
If force bump is enabled, (refer to
Overdue) and a sensor is past its bump interval, a bump
check is mandatory before the detector can enter normal operation.
Force Bump When
14
Page 23
GasAlertMicroClip
Activating the Detector
While the LCD displays BUMPCHK todAY, apply gas to the detector or press C to deactivate the detector. If gas is applied to the detector, the detector needs to enter alarm for the bump check to be accepted. Once the bump check has been accepted, the detector enters normal operation.
Bump Check Successful
If the bump check is successful, the detector enters normal operation. The detector remains in alarm until the bump gas dissipates. Verify that the audible and visual alarms are functioning properly (refer to
Alarms).

Self-Test Pass

If the detector passes the self-test, the detector begins normal operation. The LCD displays the ambient gas readings.
The detector starts recording the maximum gas exposure (MAX) and calculating the short-term exposure level (STEL) and time-weighted average (TWA) exposures.

Self-Test Fail

Once the detector enters normal operation, the LCD displays which sensor has failed the self-test by displaying Err above the target gas bar. (Refer to the
Troubleshooting section.)

Battery Test

The battery is tested on activation and continuously thereafter. A newly charged battery should operate for 10-12 hours typically before automatically shutting down. The battery power is continually displayed during normal operation. If the battery power is low, detector emits one beep and one flash (LED) every 5 seconds.
Note
If the confidence beep is enabled, the audible alarm beeps if the battery has sufficient power and stops if the battery power is low. See the section,
Confidence Beep.
flashes as the
15
Page 24
GasAlertMicroClip
User Manual

Backlight

The LCD’s backlight automatically activates when there is an alarm condition. Pressing C activates the backlight in any condition.

Deactivating the Detector

To deactivate the detector, press and hold C while it beeps and flashes to the corresponding countdown.
At the end of the countdown, the detector emits one short beep before completely deactivating.
Note
If C is released before the countdown is complete, the detector will not deactivate.

Options

To access the user options, connect the detector to the IR Link adapter and use the GasAlertMicroClip Soft Tools software. Refer to the GasAlertMicroClip Soft Tools manual for complete instructions.
The following are the available user options in Soft Tools:
1. Detector Identification: Start-up message;
2. Carbon Monoxide (CO), Hydrogen Sulphide (H2S), Oxygen (O2), and Combustible/LEL: sensor disabled, calibration gas concentration, calibration interval, bump interval, low alarm, high alarm, TWA alarm, STEL alarm, STEL interval, auto-zero on start-up, O up, and LEL by vol CH
3. User Options: confidence beep, latching alarms, safe mode, stealth mode, IR stealth mode, low alarm acknowledge, force calibration when overdue, calibration lock, force bump when overdue,
4. Language: English, French (Français), German (Deutch), Spanish (Español), and Portuguese (Portugês).
auto-calibration on start-
2
;
4
16
Page 25
GasAlertMicroClip
Options

Detector Identification

Start-up Message
This option allows you to personalize your detector. In the Detector Identification group there are two fields available for you to enter a message:
Startup Message Top Line; and Startup Message Bottom Line.
If a message is entered, it appears during the start-up sequence after the detector version is displayed. The message either
a) displays on the LCD for 3 seconds (if the entire
message fits on the LCD); or
b) scrolls by twice on the LCD.

Sensor Options

The sensor options tab allows you to change the factory settings for the H
S, CO, O2, and LEL sensors.
2
Sensor Disabled
a Warning
Disabling an installed sensor configures the detector to a 1, 2, or 3-gas unit. No protection is now provided for the gas targeted by that sensor(s). Disabling a sensor should be performed with extreme caution.
Use the Disabled function to disable a sensor. If a sensor is disabled, it can no longer be viewed on the LCD during normal operation. All sensors are enabled upon shipment. If a sensor is disabled, it may be enabled again at any time
If a sensor fails, disabling the sensor deactivates the sensor alarm (refer to enabled as soon as possible (see
Sensor Filter).
The detector functions normally with the remaining enabled sensors.
Alarms). The sensor should be replaced and
Replacing a Sensor or
17
Page 26
GasAlertMicroClip
User Manual
Calibration Gas Concentration
The calibration span gas option (Cal Gas) allows you to input a new gas concentration for each sensor. Refer to the values listed on your calibration gas cylinder.
Note
BW recommends that the calibration concentration value be set between specific ranges. Refer to the
Calibration section.
Calibration Interval
The calibration interval (Cal Interval) allows you to choose the number of days before the next calibration should be performed. You can select the minimum number of days (1 day) to the maximum number of days (365 days) before the next calibration. A different calibration interval can be selected for each sensor. Select 0 to disable the calibration interval. The detector is shipped with a default of 180 days.
Note
BW recommends that the detector be calibrated once every 180 days (6 months).
Bump Interval
The Bump Interval determines the number of days before a sensor needs to be bump checked. A different interval value can be selected for each sensor. Set the interval value to 0 to disable this option. This option is disabled upon shipment.
Note
BW recommends to “bump check” the sensors, before each day’s use, to confirm their ability to respond to gas by exposing the detector to a gas concentration that exceeds the alarm setpoints.
Low Alarm
Low Alarm option allows you to change the low alarm setpoint for each of the sensors. Refer to
Alarm Setpoints for the factory alarm setpoints.
Resetting Gas
High Alarm
High Alarm option changes the high alarm setpoint for each of the sensors. Refer to the factory alarm setpoints.
Resetting Gas Alarm Setpoints for
18
Page 27
GasAlertMicroClip
Options
TWA Alarm
The TWA Alarm option allows you to select the time­weighted average (TWA) alarm setpoint for either the H CO sensor. Refer to factory alarm setpoints.
Resetting Gas Alarm Setpoints for the
S or
2
STEL Alarm
The STEL Alarm option changes the short-term exposure limit (STEL) alarm setpoint for either the H Refer to setpoints.
Resetting Gas Alarm Setpoints for the factory alarm
S or CO sensor.
2
STEL Interval
The STEL interval determines the rate at which STEL is calculated. The detector is shipped with the STEL Interval set to 15 minutes. This value can be adjusted between 5 and 15 minutes.
Auto Zero on Start-up
When enabled this option (CO/H2S/LEL Auto-Zero on Startup) allows the detector to auto zero the chosen sensor
during activation. The auto zero option for each sensor is disabled upon shipment.
Automatic O2 Calibration
When the O2 Auto-Calibration on Startup option is enabled, it forces the detector to automatically calibrate the oxygen sensor upon activating the detector. The automatic calibration option is enabled upon shipment.
LEL By Vol CH4
The LEL By Vol CH4 option is only applicable to the LEL sensor. If it is enabled, it shows the LEL reading in %vol. assuming a methane environment.
19
Page 28
GasAlertMicroClip
User Manual

User Options Menu

Confidence Beep
The Confidence Beep notifies that the detector is on and the battery has sufficient power to respond to a hazardous level of gas to emit an alarm. Instead of beeping when the battery power is low, the audible alarm beeps once every 10 seconds to advise that the battery has sufficient power. The confidence beep stops when the battery power is low. This option is disabled upon shipment.
Latching Alarms
The detector is shipped with the Latching Alarm function disabled. If the low and high gas alarms are set to latch, the audible, visual, and vibrator alarms persist during an alarm. The alarm deactivates when it is acknowledged by pressing C and the alarm condition is no longer present.
The safe mode option is disabled upon shipment.
Stealth Mode
The detector is shipped with Stealth Mode disabled. When it is activated, stealth mode disables the beepers, backlight, and alarm LEDs and
is constantly displayed on the LCD.
Safe Mode
When enabled, the Safe Mode function advises that normal ambient conditions prevail and no monitored gas hazard
SAFE is constantly displayed when all gas levels are
exist. normal or below the alarm setpoints.
20
IR Stealth Mode
If the IR Stealth Mode option is enabled, it disables the audible alarm and the backlight. The vibrator and the IR LEDs are enabled and LCD.
is constantly displayed on the
Page 29
GasAlertMicroClip
Options
Low Alarm Acknowledge
When Low Alarm Acknowledge is enabled, the audible alarm can be disabled during a low alarm if C is pressed, but the vibrator and alarm LEDs continue to be enabled. This option is not applicable to the O
sensor.
2
Force Calibration When Overdue
Enabling the Force Calibration When Overdue option forces the detector to enter the calibration process if a sensor is overdue for calibration upon start-up. If the detector is not calibrated immediately, the detector will deactivate. The detector is shipped with this option disabled.
Cal Lock
If Cal Lock is enabled, calibration can only be completed with an IR device like Soft Tools or the MicroDock II Base Station. Though Cal Lock is enabled, the detector will still perform an auto zero when activated. The cal lock option is disabled upon shipment.
Force Bump When Overdue
When enabled, the Force Bump When Overdue option forces a bump check if the sensor has exceeded its bump check interval (see check has not been performed, the detector will deactivate.
Bump Interval). If a successful bump

Language Menu

The detector is shipped with English as the default language displayed. You can choose to view the detector in these additional languages.
French (Français);
German (Deutsch);
Spanish (Español); and
Portuguese (Português).
21
Page 30
GasAlertMicroClip
User Manual

Alarms

The following table describes the detector alarms and shows how the LCD looks for each alarm.
During an alarm condition, the detector activates the backlight, audible/visual/vibrator alarms, and the LCD shows the current ambient gas reading.
Table 6. Alarms
Alarms Display` Alarms Display
Low Alarm:
Slow siren
Slow alternating flash
L and target gas bar flash
Vibrator alarm activates
TWA Alarm:
Slow siren
Slow alternating flash
L and target gas bar flash
Vibrator alarm activates
If more than one type or level of alarm exists at the same time, a multi-gas alarm will result.
To change the factory-set alarm setpoints, refer to the GasAlertMicroClip Soft Tools instruction sheet.
High Alarm:
Fast siren
Fast alternating flash
L and target gas bar flash
Vibrator alarm activates
STEL Alarm:
Fast siren
Fast alternating flash
L and target gas bar flash
Vibrator alarm activates
22
Page 31
GasAlertMicroClip
Alarms
Table 6. Alarms
Alarms Display Alarms Display
Multi-Gas Alarm:
Alternating low and high alarm siren and flash
L and target gas bars flash
Vibrator alarm activates
Sensor Alarm:
Displays Err
Over Range (OL) Alarm:
Fast siren and alternating flash
L and target gas bar flash
Vibrator alarm activates
Confidence Beep:
One beep every 10 seconds
Low Battery Alarm:
(Confidence beep disabled)
One beep and one flash every 5 seconds
and L flashes
Automatic Shutdown Alarm:
Eight beeps and eight flashes
LOW BAT and L display
Vibrator alarm temporarily
activates
Displays OFF before turning off
23
Page 32
GasAlertMicroClip
User Manual
Table 6. Alarms
Alarms Display
Normal Shutdown:
Four beeps and flashes
Vibrator alarm temporarily
activates
Note
Alarms can be set to be latching or non-latching. To confirm this setting, access the option in the GasAlertMicroClip Soft Tools. Local regulations may require a latching alarm.
If the detector is in stealth mode, it only vibrates in alarm mode (the audible and visual alarms are disabled). If the detector is in IR stealth mode, the IR LEDs and vibrator activate in alarm mode (the audible alarm is disabled).
If the low alarm acknowledge option is enabled, the audible alarm can be disabled only during a low alarm by pressing C. The vibrator and LEDs remain activated. If the alarm escalates to a high, TWA, or STEL alarm, the audible alarm reactivates.
The backlight activates during any alarm condition.
Latching Alarms

Computed Gas Exposures

a Warning
To avoid possible personal injury, do not deactivate the detector during a work shift. TWA, STEL, and MAX readings reset once the detector is deactivated.
Table 7. Computed Gas Exposures
Gas Exposure Description
TWA
S and CO
(H
2
only)
STEL
S and CO
(H
2
only) Maximum*
(peak) * For oxygen, it is the highest or the lowest concentration
encountered.
Time-weighted average (TWA) based on accumulated exposure to toxic gases averaged over a workday according to OSHA method.
OSHA: 8 hour moving average Short-term exposure limit (STEL) to gas
based on a 5-15 minute user selectable period.
Maximum (MAX) concentration encountered during work shift.
24
Page 33
GasAlertMicroClip
Alarms

Viewing Gas Exposures

To view the TWA, STEL, and maximum (MAX) hold readings, press
C twice. The LCD first displays the TWA
gas exposures.
Then the LCD displays the STEL gas exposures.
Finally the LCD displays the MAX readings.
Clearing Gas Exposures
a Caution
Follow all safety procedures as defined by your employer.
Confirm with your supervisor before clearing TWA and STEL alarms.
To clear the TWA, STEL, and MAX exposure readings, press C when the LCD displays RESET.
25
Page 34
GasAlertMicroClip
User Manual

Gas Alarm Setpoints

The detector's gas alarm setpoints trigger the gas alarms that are described in the table below.
Table 8. Gas Alarm Setpoints
Alarm Condition
Low alarm Toxics and combustibles: Ambient
gas level above low alarm setpoint. Oxygen: Ambient gas level may be
set to above or below 20.9% (or 20.8%).
High alarm Toxics and combustibles: Ambient
gas level above high alarm setpoint. Oxygen: Ambient gas level may be
set to above or below 20.9% (or 20.8%).
TWA alarm Toxic only: Accumulated value
above the TWA alarm setpoint.
STEL alarm Toxic only: Accumulated value
above the STEL alarm setpoint.
Multi-gas alarm Two or more gas alarm conditions.

Resetting Gas Alarm Setpoints

Note
To change the alarm setpoints, you must connect the detector to the IR Link adapter and use the GasAlertMicroClip Soft Tools software. Refer to the GasAlertMicroClip Soft Tools manual for complete instructions.
Standard factory alarm setpoints will vary by region. Occupational Safety and Health Association
(OSHA) standard settings are used as an example.
The following table lists the factory alarm setpoints.
Table 9. Sample Factory Alarm Setpoints
Gas TWA STEL Low High
O LEL N/A N/A 10% LEL 20% LEL CO 35 ppm 50 ppm 35 ppm 200 ppm H2S 10 ppm 15 ppm 10 ppm 15 ppm
N/A N/A 19.5% vol. 23.5% vol.
2
Note
You can disable an alarm by setting the alarm setpoint to 0.
26
Page 35
GasAlertMicroClip
Alarms

Stopping a Gas Alarm

The low and high alarms stop when the ambient gas level returns to the acceptable range.
Note
If alarms are set to latch, press alarms.
The detector computes the TWA value based on OSHA standards and the STEL value based on a user selectable 5 to 15 minute period (see
To stop a TWA or STEL alarm, perform one of the following:
1. Deactivate (see reactivate (see detector.
2. Reset the TWA/STEL/MAX exposure readings
Viewing Gas Exposures).
(see
STEL Interval).
Deactivating the Detector) and
Activating the Detector) the
C to reset the

Sensor Alarm

The detector tests for a missing or defective sensor on activation and continuously thereafter. If a sensor fails the self-test, Err appears above the gas bar of the failed sensor (refer to the
Troubleshooting section).

Low Battery Alarm

The detector tests the battery on activation and continuously thereafter. Battery power is continually displayed during normal operation. If the battery voltage is low, the detector activates the low battery alarm.
The low battery alarm continues until you recharge the battery. If the battery voltage drops too low, the detector executes an automatic shutdown.
Note
If the confidence beep is on, the audible alarm does not beep during a low battery alarm (see
Confidence Beep). Typically, the low battery alarm
continues for 30 minutes before an automatic shutdown.

Automatic Shutdown Alarm

If the battery voltage is in immediate danger of dropping below the minimum operating voltage, LOW BAT and L display, the audible alarm beeps eight times, and the visual alarm flashes eight times. The LCD then displays OFF before it deactivates and the detector stops normal operation. Charge the battery. See the section,
Detector.
Charging the
27
Page 36
GasAlertMicroClip
User Manual

Calibration

Guidelines

When calibrating the detector, adhere to the following guidelines:
Recommended gas mixture: CO: 50 to 500 ppm balance N
S: 10 to 100 ppm balance N2
H
2
LEL: 10 to 100% LEL or 0.5 to 5% by vol. methane balance air
: clean air, 20.9 % (or 20.8%)
O
2
CG-Q58-4 and CG-Q34-4 calibration gas (4-gas mix) are available from BW Technologies by Honeywell. See the section,
Replacement Parts and Accessories.
Calibration accuracy is never better than the calibration gas accuracy. BW Technologies by Honeywell recommends a premium-grade calibration gas. Gases with the National Institute of Standards and Technology (NIST) traceable accuracy improve the validity of the calibration. Do not use a gas cylinder beyond its expiration date.
Calibrate a new sensor before use. Install the sensor, activate the detector, and allow the sensor to stabilize before starting calibration (used: 60 seconds; new: 5 minutes).
2
Calibrate the detector at least once every 180 days, depending on use and sensor exposure to poisons and contaminants.
Calibrate the detector if the ambient gas display varies at start-up.
It is best to calibrate the sensor before changing the alarm setpoints.
Calibrate only in a safe area that is free of hazardous gas.
Do not calibrate the detector during or immediately after charging is complete.
The oxygen sensor can be automatically calibrated each time upon activation (if this feature is enabled). Activate the detector in a normal (20.9%/20.8% oxygen) atmosphere.
The detector should be allowed to stabilize for 1 minute after activation, prior to calibration or bump test.
If you require a certified calibration, contact
BW Technologies by Honeywell.
28
Page 37
GasAlertMicroClip
Calibration

Diagnostics Protection

The detector tests the ambient air (auto zero) and the test gas that is applied (auto span) to ensure it meets expected values.
If an unexpected level of target gas is present during an auto zero, the auto zero will fail for that sensor. The detector advises which sensor(s) failed the auto zero before proceeding to the auto span.
In auto span, if any target gas is not present or does not meet expected values, the display will advise you of that failure. The sensor(s) that failed to span will retain its previous set value and not continue with the calibration process.

Applying Gas to the Sensors

The calibration cap and hose, which is shipped with the detector, simplifies sensor testing and calibration. Figure 3 and Table 10 show how to use it when applying gas to the sensors.
Note
The calibration cap should only be used during the calibration span process.
Wind currents may cause false readings and poor calibrations.
Figure 3. Applying Gas to the Sensors
Table 10. Applying Gas to the Sensors
Item Description
1 Detector and calibration cap 2 Calibration hose 3 Regulator and gas cylinder
29
Page 38
GasAlertMicroClip
User Manual

Calibration Procedure

To calibrate the detector perform the following procedure.
Note
Calibrate O
in clean air.
2
If C is pressed at any time during the calibration procedure, calibration is ABORTED. The detector emits a quick beep and the calibration due dates are displayed before the detector returns to normal operation.
Start Calibration
Note
Verify that the calibration gas you are using matches the span concentration value(s) in the detector. See the section,
Concentration.
Calibration Gas
1. In a clean atmosphere, press and hold
C as the detector beeps, flashes, and vibrates to the corresponding OFF countdown. Continue to hold C as the detector briefly deactivates.
2. The detector then reactivates and performs the CAL countdown. Continue to hold C until the CAL countdown is complete to enter calibration.
Note
If C is not held for the entire CAL countdown, the detector will deactivate.
30
Page 39
GasAlertMicroClip
Calibration
Auto Zero and Oxygen Sensor Calibration
Note
Do not apply calibration gas at this point, otherwise auto zero may fail for that sensor and then it cannot be spanned.
3. The LCD flashes automatically zeroes the toxic and combustible sensors and calibrates the oxygen sensor. The LCD notifies you if the auto zero has failed for a sensor.
while the detector
Auto Zero Sensor Fail
The audible alarm then beeps twice.
Auto Span
4. The LCD now prompts you to apply calibration gas to the sensors.
5. K flashes and remains displayed as you attach the calibration cap and apply gas to the sensors at a flow rate of 250 to 500 ml/min. (refer to Figure 3. Applying Gas to the Sensors).
31
Page 40
GasAlertMicroClip
User Manual
When the detector senses a sufficient amount of gas concentration (approximately 30 seconds), the audible alarm beeps once,
flashes, and K remains lit while the detector completes the span (approximately 2 minutes).
Successful Span
If the sensors have spanned successfully, the audible alarm beeps before the calibration procedure continues.
Unsuccessful Span
If the sensors fail the span, the LCD displays FAILURE before proceeding with the calibration procedure.
No Gas Detected
The LCD displays FAILURE if the detector does not detect any gas within 2 minutes before proceeding with the calibration procedure.
The detector will not span a sensor if:
You do not apply gas to the sensor.
The sensor fails to detect at least one-half of the
expected gas concentration in the first 30 seconds.
The gas concentration drops below one-half of
the expected gas level during the 2 minute span.
Note
32
Page 41
GasAlertMicroClip
Calibration
If you apply gas to a sensor and the detector fails to span the sensor, repeat the calibration procedure using a new gas cylinder. If the sensor fails the span a second time, replace the sensor. See the section,
Filter.
Replacing a Sensor or Sensor
Calibration Due Date
Note
If a sensor does not successfully span, its calibration due date will not be reset.
6. After span is complete, the LCD displays the following calibration due date screens before returning to normal operation.
Note
The number of days shown in the last screen represents the earliest date a calibration needs to be performed.
If a sensor failed to successfully span and it is past its calibration due date, the LCD displays the following three screens.
Press C to acknowledge the warning before returning to normal operation.
33
Page 42
GasAlertMicroClip
User Manual
Verification
After calibration is complete and the detector is in normal operating mode, test it using a gas cylinder other than the one used in calibration. The gas concentration should not exceed the sensor's detection range. Confirm that the LCD shows the expected concentration. To ensure the readings are accurate, apply the test gas for the same amount of time as was applied to the sensor when it was calibrated.

Datalogs

The datalogger allows the detector to record various information so you can compile a report. The detector has a sample rate of 15-second intervals. The detector is capable of storing 16 hours of information. When the memory is full, the detector replaces the oldest data with the most recent data.

Event Logs

The event logging feature keeps a record of the ten most recent gas alarm events. The information that is recorded from an event is as follows:
Serial number;
Gas type;
Peak exposure level (ppm or %);
Alarm duration in seconds; and
Alarm type.

Downloading Datalogs and Event Logs

The datalog and event log files can only be downloaded to a PC using an IR device (e.g., IR Link Adapter or MicroDock II Base Station).

Maintenance

To keep the detector in good operating condition, perform the following basic maintenance as required:
Calibrate, bump check, and inspect the detector at
regular intervals.
Keep an operations log of all maintenance, bump checks,
calibrations, and alarm events.
Clean the exterior with a soft damp cloth. Do not use
solvents, soaps, or polishes.
Start time of event;
34
Page 43
GasAlertMicroClip
Maintenance

Charging the Detector

a Warning
The detector must be charged in a non-
hazardous atmosphere of 32°F - 113°F (0°C - 45°C).
Charge the detector using only the
recommended charging adapter. Do not use any other charging adapter. Failure to observe this precaution can lead to fire or explosion.
The charging adapter is voltage specific to North
America. Use of the charging adapter outside of North America will damage the charger and detector.
The battery can only be replaced by the
manufacturer. Not complying with this might cause explosion danger.
Note
To preserve battery life, deactivate the detector when you are not using it.
To charge the detector, execute the following steps:
1. Deactivate the detector and attach the GasAlertMicroClip charging adapter as shown in
the following figure. The detector can be activated during charging.
Figure 4. Connecting the Charging Adapter
35
Page 44
GasAlertMicroClip
User Manual
2. Allow the detector to charge for 2-3 hours. The charging indicator flashes on the LCD while the detector is being charged.
3. Once charging is complete, the LCD continues to display detector.
If the battery indicator does not display within 5 minutes, refer to the
To reach full battery capacity, allow the battery to fully charge and fully discharge three times.
Charging the detector in temperatures above 113°F will greatly reduce the number of charges the detector can accept.
The detector may be hot immediately following charging. This is normal.
. Remove the adapter and activate the
Note
Troubleshooting section.

Replacing a Sensor or Sensor Filter

a Warning
To avoid personal injury, use only sensors specifically designed for the detector. See the section
Replacement Parts and Accessories.
Each sensor has a high degree of resistance to common vapors and gases. A sensor will most likely clear itself if you move the detector to a clean environment and wait 10 to 30 minutes, but it could take up to a couple of days depending upon the gas that came in contact with the sensor. Do not expose a sensor to the vapors of inorganic solvents such as paint fumes or organic solvents. The describes problems caused by a sensor in need of calibration or replacement.
To replace a sensor or sensor filter, refer to the following figure, table, and set of instructions.
Troubleshooting section
Figure 5. Replacing a Sensor or Sensor Filter
36
Page 45
GasAlertMicroClip
Maintenance
Table 11. Replacing a Sensor or Sensor Filter
Item Description
1 Front shell 2 LEL sensor 3 PCB 4 PCB screws (2) 5 Rear shell 6 Machine screws (6) 7 Sensor filter 8 O2 sensor
9 H2S sensor 10 CO sensor 11 Sealing rib
1. Deactivate the detector.
2. Remove the six machine screws on the rear shell and remove the back cover.
3. Remove the two screws on the PCB and remove the PCB.
Note
Ensure the battery does not get damaged once the PCB is removed.
4. Remove the old sensor filter or slide/pull out the old sensor(s).
5. Insert a new sensor filter or sensor(s).
Note
When inserting a new sensor filter, ensure the white side is facing the sensors (i.e., the black side is facing out).
Detectors that are configured for 1, 2, or 3 gases may contain a dummy sensor in one of the four sensor locations.
37
Page 46
GasAlertMicroClip
User Manual
6. Re-assemble the detector. When assembling the detector, be aware of the following:
Ensure the charging contact pins inside the
rear shell are lined up to its appropriate hole before inserting the rear shell in place. If the contact pins are bent, the battery cannot charge properly.
Verify that the battery PCB is in the same
position as when the rear shell was removed (refer back to step #2).
38
Visually inspect the battery to ensure that it
has not been damaged before replacing the rear shell.
Ensure the rib on the interior of the rear shell
fits between the battery pack and the PCB as shown in Figure 6.
Ensure the rear shell sealing rib is not folded
when replacing the rear shell (see Figure 5).
Torque the screws to 3-4 in-lbs in a crisscross
pattern to ensure a proper environment seal.
Figure 6. Replacing the Rear Shell
Calibrate the detector after changing any sensor. See the section,
Calibration.
Page 47
GasAlertMicroClip
Troubleshooting
Troubleshooting
The detector’s electronics are protected from variations in humidity and corrosive atmospheres. If you encounter a problem, try the solutions listed in the following table.
Table 12. Troubleshooting Tips
Problem Possible Cause Solution
The detector does not activate.
The detector enters alarm mode immediately when activated.
The activation self-test fails.
Depleted battery
Damaged or defective detector
Sensor needs to stabilize
Low battery alarm
Detector requires calibration General fault Sensor failure
If you are still unable to correct the problem, contact
BW Technologies by Honeywell.
Charge battery (see
Detector)
Contact
Honeywell
Used sensor: wait 60 seconds
New sensor: wait 5 minutes
Charge battery (see
Detector)
Calibrate detector (see Contact
Honeywell
Replace sensor (see
Sensor or Sensor Filter)
BW Technologies by
BW Technologies by
Charging the
Charging the
Calibration)
Replacing a
39
Page 48
GasAlertMicroClip
User Manual
Problem Possible Cause Solution
Detector does not display normal ambient gas reading after activation self-test.
Detector does not respond to pushbutton.
Detector does not accurately measure gas.
Detector has been charging for over 3 hours and the charging indicator still indicates that the battery has not finished charging.
Table 12. Troubleshooting Tips (cont.)
Sensor not stabilized
Detector requires calibration
Target gas is present Battery is depleted
Detector is performing operations that
do not require user input
Detector requires calibration
Detector is colder/hotter than
ambient gas
Sensor filter is blocked Battery is trickle charging Battery is full and ready for operation
Used sensor: wait 60 seconds
New sensor: wait 5 minutes
Calibrate detector (see Detector is operating properly. Use
caution in suspect areas
Charge battery (see
Detector)
Pushbutton operation restored
automatically when the operation ends
Calibrate sensor (see Allow the detector to acquire ambient
temperature before use
Clean the sensor filter (see
Replacing a Sensor or Sensor Filter)
Calibration)
Charging the
Calibration)
40
Page 49
GasAlertMicroClip
Troubleshooting
Table 12. Troubleshooting Tips (cont.)
Problem Possible Cause Solution
Detector does not enter alarm.
Detector intermittently enters alarm without apparent reason.
Detector automatically deactivates.
Battery indicator does not display when charging.
Alarm setpoint(s) are set incorrectly
Alarm setpoint(s) set to zero
Detector is in calibration mode Ambient gas levels are near alarm
setpoint or the sensor is exposed to a puff of the target gas
Alarms set incorrectly
Detector requires calibration Missing or faulty sensor
Automatic shutdown feature activated
due to weak battery
Detector may be depleted below
normal levels
Reset alarm setpoints (see
Gas Alarm Setpoints)
Reset alarm setpoints (see
Gas Alarm Setpoints)
Complete the calibration procedure Detector is operating normally. Use
caution in suspect areas. Check maximum gas exposure reading.
Reset alarm setpoints (see
Gas Alarm Setpoints)
Calibrate sensors (see Replace sensor (see
Sensor or Sensor Filter)
Charge detector (see
Detector)
Charge detector for approximately
8 hours. Detector LEDs may light during first 5 hours. This is normal.
If the battery indicator does not light
after charging for 8 hours, contact
BW Technologies by Honeywell.
Resetting
Resetting
Resetting
Calibration)
Replacing a
Charging the
41
Page 50
GasAlertMicroClip
User Manual

Replacement Parts and Accessories

a Warning
To avoid personal injury or damage to the detector, use only the specified replacement parts.
To order parts or accessories listed in the following table,
BW Technologies by Honeywell.
contact
Table 13. Replacement Parts and Accessories
Model No. Description Qty
SR-W-MC Replacement combustible sensor 1 SR-X-MC Replacement O2 sensor 1 SR-M-MC Replacement CO sensor 1
SR-H-MC Replacement H2S sensor 1 REG-0.5 Regulator (0.5 l/min) 1 CG-Q58-4 Quad gas cylinder: CH4-2.5%, O2-
18.0%, H CO-100 ppm, bal. N
CG-Q34-4 Quad gas cylinder: CH4-2.5%, O2-
18.0%, H CO-100 ppm, bal. N
CG-T34 Dual gas cylinder: 50% LEL
(CH (34 l)
S-25 ppm,
2
S-25 ppm,
2
-2.5%) O2-20.9%, bal. N2
4
(58 l)
2
(34 l)
2
Model No. Description Qty
G0042-H25 Single gas cylinder: H2S 25 ppm,
(58 l)
bal. N
2
CG2-M-200­103
Single gas cylinder: CO 200 ppm,
(103 l)
bal N
2
CG-BUMP1 Bump alarm gas aerosol
-2.5%, O2-10%, H2S-40 ppm,
(CH
4
1
1
1
CO-200 ppm)
CK-Q34-4 Quad calibration kit with regulator,
1 quad gas cylinder (CG-Q34-4), hose, and carrying case
CK-Q58-4 Quad calibration kit with regulator,
1 quad gas cylinder (CG-Q58-4), hose, and carrying case
MC-XWHM­Y-NA-07
DOCK2-0­1C1L-00-N
1
MC­SCREW-K1
1
MC-AF-K1 Auxiliary filter kit 1
Confined space kit with manual aspirator pump
GasAlertMicroClip MicroDock II docking module with charging cable
Screw kit (5 sets of screws and screwdriver)
1
1
1
MC-SS-AF-K1Replacement auxiliary filters 10
1
MC-SS Sensor filters 2 MC-PA-1 Replacement charging adapter 1
42
Page 51
GasAlertMicroClip
Specifications
Model No. Description Qty
MC-VC-K1 Vehicle charging kit 1 MC-C01-
GasAlertMicroClip MultiCharger 1
MC5* GA-USB1-IR IR connectivity kit 1 GA-PA-1* Charging adapter 1 MC-TC-1 Calibration cap and hose (1 ft./3 m) 1 GA-AG-3 Alligator clip (stainless steel) 1
*Add suffix (-UK) for United Kingdom mains plug, (-EU) for European mains plug, (-AU) for Australian mains plug.

Specifications

Instrument dimensions: 10.75 x 6.00 x 2.73 cm (4.2 x 2.4 x 1.1 in.)
Weight: 160 g (5.7 oz.) Operating temperature: -20°C to +58°C (-4°F to +136°F)
+50°C to +58°C is certified by CSA-International on the combustible sensor with ±5% accuracy
Storage temperature: -40°C to +50°C (-40°F to +122°F) Operating humidity: 0% to 95% relative humidity
(non-condensing)
Alarm setpoints: May vary by region and are user-settable
Detection range:
H2S: 0 – 100 ppm (1 / 0.1 ppm increments)
S: 0 – 200 ppm (1 ppm increments)
H
2
CO: 0 – 500 ppm (1 ppm increments) CO: 0 – 1000 ppm (1 ppm increments)
: 0 – 30.0% vol. (0.1% vol. increments)
O
2
Combustible (LEL): 0 – 100% LEL (1% LEL increments) or 0 – 5.0% v/v methane
Sensor type:
S, CO, O2: Single plug-in electrochemical cell
H
2
Combustibles: Plug-in catalytic bead
O measuring principle:2 Capillary controlled concentration
sensor Alarm conditions: TWA alarm, STEL alarm, low alarm,
high alarm, multi-gas alarm, over range (OL) alarm, low battery alarm, confidence beep, automatic shutdown alarm
Audible alarm: 95 dB+ at 30 cm (100 dB typical) variable pulsed beeper
Visual alarm: Red light-emitting diodes (LED) Display: Alphanumeric liquid crystal display (LCD) Backlight: Activates when the pushbutton is pressed and
deactivates after 5 seconds. Also activates during an alarm condition.
Self-test: Initiated upon activation Calibration: Automatic zero and automatic span Oxygen sensor: Automatic span upon activation
(selectable)
43
Page 52
GasAlertMicroClip
User Manual
User field options: Start-up message, confidence beep, latching alarm, enable/disable safe display mode, oxygen measurement, combustible sensor measurement, sensor disable, set calibration interval, force calibration, calibration due lock, force bump, bump interval, stealth mode, IR stealth mode, low alarm acknowledge, language selection, enable/disable automatic oxygen calibration, enable/disable auto zero at start-up, set alarm setpoints, set span concentration values, set STEL calculation period.
Battery operating time:
1 rechargeable lithium polymer: 10-12 hours (typical) Year of manufacture: The detector’s year of manufacture is
determined from the serial number. The second and third number after the first letter determines the year of manufacture.
E.g., H3
Approved battery: North America
Approved batteries for product (standards EN50020, UL913, C22.2 No. 157)
Rechargeable battery Temperature code
Lithium polymer T4
Battery charger: GasAlertMicroClip charging adapter First-time charge: 2-3 hours Normal charge: 2-3 hours Warranty: 2 years including sensors
04-Y000001 = 2004 year of manufacture
Approvals: Approved by CSA to both U.S. and Canadian Standards CAN/CSA C22.2 No. 157 and C22.2 152 ANS/UL – 913 and ANSI/ISA – S12.13 Part 1
CSA Class I, Division 1, Group A, B, C, and D ATEX CE 0539 g II 1 G EEx ia IIC T4
KEMA 06ATEX0056 IECEx Ex ia IIC T4
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules and ICES-003 Canadian EMI requirements. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one of more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and
receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
44
Page 53
GasAlertMicroClip
Specifications

General Datalogger Specifications

Storage: 16 hours at 15-second intervals Memory type: Wraparound memory ensures most recent
data is always saved
Sample rate: One reading every 15 seconds Data recorded: All sensor readings, all alarm conditions,
calibrations, event flags, battery status, sensor status, confidence beep activation, and detector status along with the time and date for each reading and unit serial number
Operation: Requires no user intervention (automatic) Compatible with: Desktop PC computer or laptop Operating system: Windows 2000 or higher Download via: IR device (e.g., IR Link adapter, MicroDock
II Base Station, etc.) Software required: Soft Tools with spreadsheet/database
compatible with comma-separated-value (CSV) text files (Excel, Access, Quattro, etc.)
45
Page 54
GasAlertMicroClip
User Manual
46
Page 55
D5908/2 English
iERP: 124291 © BW Technologies 2007, all rights reserved. Printed in Canada. All product names are trademarks of their respective companies.
Page 56
Corporate Headquarters Calgary, Alberta
The following information has changed in the user manual (124291 D5908/2)
Pages 9 & 10
a Caution Only activate the detector in a safe area that is free of hazardous gas.
Page 19
Auto Zero on Start-up
The auto zero option for each sensor is enabled upon shipment.
¾
in.
3

ERRATA CARD

5½ in.
GasAlertMicroClip
125109
Printing Information
iERP # 125109 Size: 3¾ x 5½ inches Stock: 80 lb cougar cover card stock Print Color: Black Language: English
Page 57
DAILY BUMP TEST
iERP: 125538
a IMPORTANT NOTE
BW recommends to “bump test” the sensor(s), before each day’s use to confi rm their ability to respond to gas by expos­ing the detector to a gas concentration that exceeds the alarm setpoints. Manually verify that the audible and visual alarms are activated.
If any sensor fails the bump test, contact your local BW service department immedi­ately.
Loading...